    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Atlantis Abutment & Atlantis Abutment Milling
    The Atlantis Abutment and Atlantis Abutment Milling are intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The Atlantis Abutment screw is intended to secure the Atlantis Abutment to the endosseous implant.

    Atlantis Crown Abutment
    The Atlantis Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The Atlantis Abutment screw is intended to secure the Atlantis Crown Abutment to the endosseous implant.

    Atlantis Conus Abutment
    The Atlantis Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is an attachment-retained by friction fit to the abutment. The Atlantis Abutment screw is intended to secure the Atlantis Conus Abutment to the endosseous implant.

    Atlantis Healing Abutment
    The Atlantis Healing Abutment is intended for use with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The Atlantis Abutment screw is intended to secure the Atlantis Healing Abutment to the endosseous implant.

    Implant Manufacturer: Institut Straumann
    Implant System: Straumann BLX
    Implant Diameter (in mm): 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5
    Abutment Platform Diameter (in mm): 2.69

    Implant System: Neodent GM
    Implant Diameter (in mm): 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0
    Abutment Platform Diameter (in mm): 3.0

    Device Description

    The Atlantis® Abutments in Titanium are patient-specific dental abutments that are intended for attachment to dental implants in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function. The design of the Atlantis® Abutments in Titanium is derived from patient dental models and is completed by Dentsply Sirona using computer-assisted design (CAD) technology according to the clinician's prescription. The final CAD design of the Atlantis® Abutment in Titanium is fabricated using computer-assisted manufacturing (CAM) to produce a customized, patient-specific device. Design and fabrication of the Atlantis® Abutment in Titanium is completed in internal Dentsply Sirona manufacturing facilities.

    Alternatively, the CAD design, according to the clinician's prescription, can be performed by a laboratory or clinician in an FDA cleared abutment design software (3Shape Abutment Designer Software, K151455) within the design envelope of the Atlantis® Abutments which is codified in the validated and locked design library of the cleared software. Fabrication of the Atlantis® Abutment is then completed in internal Dentsply Sirona manufacturing facilities.

    The Atlantis® Abutment in Titanium serves as a connection of the prosthetic construction and the endosseous implant. The lower part of the abutment is designed to fit with the specific implant geometry it is compatible, and the upper part design is according to the patient's specific anatomy. The Atlantis® Abutment, including the Conus and Healing abutments further described below, are available in Titanium or Gold-shaded (titanium nitride layer applied using PVD (Physical Vapor Deposition)) Titanium. The Crown Abutment which is another Atlantis® Abutment in Titanium, is only available in Titanium.

    The Atlantis® Abutment design envelope became the basis for the other more specific designs that make up the Atlantis® Abutments in Titanium. The Atlantis® Abutment is intended for use with an endosseous implant and for single tooth restoration.

    The Atlantis® Crown Abutment in Titanium is also a custom fabricated abutment using the same internal CAD/CAM technology. The Atlantis® Crown Abutment in Titanium incorporates a design that is a combination of an abutment and an anatomically accurate crown to constitute the final finished device. It functions as a substructure that also serves as the final abutment, in a partially or completely edentulous patient.

    The Atlantis® Conus Abutment supports a removable prosthesis (bridges and overdentures) which is retained by friction fit to the abutment. The abutment connects to the prosthesis via caps embedded in the prosthesis.

    The Atlantis® Healing Abutment is used with the compatible implants for temporary use during soft tissue healing after one-stage and two-stage surgeries. It is designed based on the planned final Atlantis® Abutment or Atlantis® Crown Abutment, using the same emergence profile as those abutments to achieve an aesthetic outcome during the soft tissue healing phase.

    The proposed Atlantis® Abutments in Titanium for Straumann BLX Implant are compatible with the Straumann BLX Implants. This implant system is available in diameters of 3.5, 3.75, 4.0, 4.5 5.0, 5.5 and 6.5 mm and in lengths of 8, 10, 12, 14, 16, and 18 mm all with identical mating geometry.

    The proposed Atlantis® Abutments in Titanium for Neodent GM Implant are compatible with the Neodent GM Implants. This implant system is available in diameters of 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, and 7.0 mm and in lengths of 7, 8, 9, 10, 11, 11.5, 13, 15, 16, 17 and 18 mm, all with identical mating geometry.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252384

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-23

    (84 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add abutments to the DESS Dental Smart Solutions system, which includes abutments cleared previously in K242340 and K240208. The subject device abutment designs include Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), and abutment screws.

    Pre-Milled Blank Abutments are designed for custom abutment fabrication by a CAD-CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. The Pre-Milled Blank Abutments have a maximum (before milling) diameter of 10 mm or 14 mm and a solid cylindrical design with an engaging or non-engaging implant connection. They are available in versions with a straight screw channel and with an angled screw channel for esthetic considerations. The Pre-Milled Blank Abutments are manufactured from titanium alloy (Ti 6Al-4V).

    The design parameters for the CAD-CAM fabrication of a custom abutment from the Pre-Milled Blank Abutment are:

    • Minimum wall thickness – 0.45 mm
    • Minimum cementable post height for single-unit restoration – 4.0 mm (minimum cementable post height for single-unit restoration is defined as the height above the restorative margin)
    • Minimum gingival height – 0.5 mm
    • Maximum gingival height – 6.0 mm
    • Pre-Milled Blanks are for straight abutments only

    Multi-Unit Abutments: Straight and Angled are designed for attachment of multi-unit screw-retained restorations and are provided in three (3) designs, straight (0°), angled 17°, and angled 30°. The designs of the subject Multi-Unit Abutments are similar to the designs of Multi-Unit Abutments cleared in K242340 and K240208. All Multi-Unit Abutments are manufactured from titanium alloy (Ti-6Al-4V).

    The subject device straight Multi-Unit Abutments have a non-engaging, threaded design that attaches directly to the implant. The subject device straight (0°) Multi-Unit Abutments are included in this submission are similar in design to straight Multi-Unit Abutments cleared in K242340, K230143, and K222288. Subject device straight Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 1 mm to 5.5 mm.

    The subject device angled Multi-Unit Abutments are provided only in an engaging design that requires an abutment screw. The subject device Multi-Unit Abutments angled 17° and 30° are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 2.5 to 4.5 mm.

    Select straight and angled Multi-Unit Abutments have an added coating of zirconium nitride (ZrN). This coating is identical to the ZrN coating used on abutments with ZrN cleared in K242340.

    DESS® Dental Smart Solutions Screws are designed to attach the abutment to the implant or the prosthesis to the abutment. There are a total of seventeen (17) subject device screws compatible with the subject device components or previously cleared components. The new screws have designs that are similar to those of screws cleared in K242340 and K240208. Screws are made of titanium alloy (Ti-6Al-4V). Select subject device screws are available with DLC (Diamond-like Carbon) coating and are identical to that used on screws cleared in K240208.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250967

    Validate with FDA (Live)

    Device Name
    Sherlock
    Manufacturer
    Date Cleared
    2025-06-26

    (87 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K212664, K193335, K163194, K180536, K201225, K150704

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sherlock abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.

    All digitally designed CAD/CAM customizations for Sherlock abutments are to be sent to an Open Implants-validated milling center for manufacture.

    Sherlock abutments are compatible with the implant systems listed in the Compatibility Table:

    Compatibility Table

    Compatible Implant SystemsImplant Body Diameter (mm)Implant Platform Diameter (mm)
    Neodent Helix GM®, Drive GM®, Titamax GM®3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.03.0
    Device Description

    Sherlock is a dental implant abutment system that is being expanded to include a new compatible implant system, Neodent GM®. The Subject device implant platform diameter is 3.0 mm, and the corresponding compatible implant body diameters range from 3.5 mm to 7.0 mm.

    The abutment designs are Titanium Base, Titanium Blank, Straight Multi-Unit, Multi-Unit Angled 17°, and Multi-Unit Angled 30° Abutments. These abutment designs were previously cleared in the sponsor's K220482 Predicate device and K212664 Reference device submissions. All abutment designs are provided with corresponding abutment screws.

    The Subject device Titanium Base abutments are intended to be used as a two-piece abutment composed of the base bottom-portion (prefabricated titanium base component) with a cemented/bonded CAD-CAM fabricated zirconia top-portion (superstructure) where the final two-piece abutment (base component and cemented superstructure) is the finished device used for the prosthetic restoration. Each patient-specific zirconia superstructure is individually prescribed by the clinician and manufactured by an authorized milling center.

    All Subject device prefabricated titanium base components are provided in a straight design with no angulation in the titanium base post. They are provided with either an indexed/engaging implant connection for crowns or a non-engaging/nonindexed implant connections for bridges. The standard prefabricated titanium base components are provided in gingival heights ranging from 0.8 mm to 3.0 mm and abutment post lengths of 8 mm or 10 mm. The ASC prefabricated titanium base components are provided in gingival heights ranging 0.8 mm to 2.5 mm and abutment post length of 8 mm. Additional gingival height may be provided for both abutment designs in the zirconia superstructure. ASC prefabricated titanium base components are provided with a cutout in the prosthetic post to accommodate a restoration with an angled screw channel when clinically necessary. Standard prefabricated titanium base components and ASC prefabricated titanium base components posts may be reduced to 4 mm to accommodate individual patient occlusion. The zirconia mesostructure may contain an angled post within the established design parameters.

    The overall design parameters for the two-part Standard and ASC CAD/CAM prefabricated titanium base components with zirconia mesostructure are:

    • Minimum Zirconia Wall Thickness – 0.5 mm
    • Minimum Post Height for single-unit abutment* – 4.0 mm
    • Minimum Overall Gingival Height – 0.8 mm (titanium base plus zirconia)
    • Maximum Overall Gingival Height – 5 mm
    • Maximum Correction Angle – 30°

    The required cement for bonding the zirconia superstructure to the Subject device Titanium Bases to create the final two-piece abutment is Kuraray Noritake Dental PANAVIA™ V5 cleared in K150704.

    Titanium Blank abutments, sometimes referred to as "Pre-mill" or "Ti-Blank" abutments are one-part abutments intended for use in a CAD/CAM workflow. Each Subject device Titanium Blank implant abutment has a pre-manufactured indexed implant connection interface with a cylindrical customization section and a milling retention geometry section. The retention geometry holds the component in a milling machine fixture while the patient-specific portion above the implant interface is milled in a dental milling machine. All patient-specific Titanium Blank abutment fabrication is by prescription on the order of the clinician.

    The overall design parameters for the Titanium Blank customized abutments are:

    • Minimum Wall Thickness – 0.75 mm
    • Minimum Post Height for single-unit abutment* – 4.0 mm
    • Minimum Overall Gingival Height – 0.8 mm
    • Maximum Overall Gingival Height – 5 mm
    • Maximum Correction Angle – 30°

    All digitally designed zirconia mesostructures for use with the Subject device titanium base abutments and digitally designed Subject device titanium blank abutments will be fabricated at an Open Implants validated milling center under FDA quality system regulations.

    Multi-Unit Abutments (MUAs) are intended for use with multi-unit restorations. They are considered two-part abutments. The base portion of the MUA is connected directly to the implant either with an integral screw (straight MUA) or with a separate multi-unit abutment screw (angulated abutments). Straight MUAs have a non-indexed connection with the dental implant. The angulated MUAs have an indexed connection with the dental implant. The second part of the MUA is a mating coping which is retained with a prosthetic screw. Multi-Unit Abutments are available in Straight, 17° Angulated and 30° Angulated configurations. The coping and prosthetic screw is compatible with each MUA design/configuration.

    All Subject device abutments and corresponding abutment screws are pre-manufactured from Ti-6Al-4V ELI (Grade 23) titanium conforming to ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) and are provided non-sterile to the user. The mesostructure/copings for Titanium base abutments are fabricated from zirconia conforming to ISO 13356, Implants for surgery — Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).

    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for Sherlock Dental Implant Abutment does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria.

    This document describes the device's indications for use, its substantial equivalence to predicate devices, and the non-clinical data reviewed by the FDA for clearance. However, it does not detail specific acceptance criteria or an associated study demonstrating compliance with those criteria.

    The "Performance Data" section lists several types of non-clinical data submitted, such as static and fatigue testing according to ISO 14801, a reverse engineering study, biocompatibility testing, cleaning and sterilization validation, and an MRI review. These are the studies performed to support the substantial equivalence claim, but the FDA letter does not specify quantitative acceptance criteria for these tests nor does it report specific performance outcomes that directly address such criteria.

    Therefore, I cannot populate the requested table or provide details for points 1-9 based solely on the provided text. The document focuses on demonstrating substantial equivalence to previously cleared devices rather than presenting a performance study against predefined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243009

    Validate with FDA (Live)

    Device Name
    MIST IC
    Date Cleared
    2025-06-24

    (271 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MIST IC abutments are intended for use to support a prosthetic device in a partially or completely edentulous patient. They are intended to support a single-unit or multi-unit, cement retained prosthesis in the mandible or maxilla. MIST IC abutments are compatible for use with the following implants:

    Compatible Implant SystemImplant Body Diameter, mmImplant Platform Name
    Straumann Bone Level (CrossFit® Morse Taper)2.9SC
    3.3NC
    4.1, 4.8RC
    Straumann® Tissue Level (CrossFit® Connection)3.3, 4.1, 4.8RN
    4.8WN
    Straumann BLX (TorcFit™ Internal Hexalobular)3.5, 3.75, 4.0, 4.5RB
    5.0, 5.5, 6.5WB
    Neodent GM (Morse taper)3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0Grand Morse (GM)

    All digitally designed custom abutments for use with MIST IC abutments are to be sent to an Imagine Milling Technologies validated milling center for manufacture.

    Device Description

    MIST IC from Imagine Milling Technologies, LLC is a line of Ti-base and machinable blank abutments to interface with compatible dental implants from four (4) manufacturers, and a total of eight (8) implant-abutment interface compatibilities. The subject device prosthetic platform diameters range from 3.8 mm to 6.9 mm. All stock subject device components (abutments and abutment screws) are made of titanium alloy conforming to ASTM F136. The subject device MIST IC L-LINK abutments have a TiN coating achieved through a physical vapor deposition (PVD) process that is identical to the process used for TiN coating of Imagine Milling Technologies, LLC devices cleared in K222368. The PVD cathodic arc evaporation process is a high current, low voltage process in which material evaporated from the cathode (Ti) is ionized, transported through the vacuum chamber with reactive gas (N2) and deposited as a non-porous, thin film on the titanium substrate. Each abutment is supplied with the non-sterile abutment screw designed for attachment to the corresponding compatible OEM implant.

    All patient-specific abutment fabrication for all MIST IC abutments is by prescription on the order of the clinician. All MIST IC abutments are intended to be milled at an Imagine Milling Technologies, LLC validated milling center under FDA quality system regulations.

    MIST IC L-LINK abutments are two-piece abutments to be used as a base when fabricating a CAD-CAM customized restoration where the superstructure produced will compose the second part of the two-piece abutment; the assembly becoming a final finished medical device after cementation on the subject device abutment. They are provided with engaging and non-engaging connections.

    The L-LINK abutments and corresponding zirconia superstructure are provided to the clinician either with the superstructure cemented to the abutment by the dental laboratory, or separately for the clinician to bond together chairside using the cement required in the labeling (RelyX RMGIP bonding cement, cleared in K022476).

    The design parameters for L-LINK patient-specific abutments are:

    • Minimum wall thickness – 0.5 mm
    • Minimum cementable post height for single-unit restoration – 4.0 (minimum cementable post height for single-unit restoration is defined as the height above the restorative margin)
    • Minimum gingival height – 0.5 mm
    • Maximum gingival height (for all except Straumann BLX L-Link) – 5.0 mm
    • Maximum gingival height (for Straumann BLX L-LINK) – 6.0 mm
    • Maximum angle – 20°

    All zirconia copings (superstructures) for use with the subject device MIST IC L-LINK abutments will conform to ISO 13356.

    MIST IC PREFIT abutments are cylindrical abutments designed for patient-specific abutment fabrication by a CAD-CAM process and machined into a one-piece, all titanium abutment. The portion of the abutment available for milling is either 9.9 mm in diameter by 20 mm in length or 13.9 mm in diameter by 20 mm in length. MIST IC PREFIT abutments have an engaging connection.

    The design parameters for PREFIT patient-specific abutments are:

    • Minimum wall thickness – 0.5 mm
    • Minimum cementable post height for single-unit restoration – 4.0 (minimum cementable post height for single-unit restoration is defined as the height above the restorative margin)
    • Minimum gingival height – 0.5 mm
    • Maximum gingival height (for all except Straumann BLX PREFIT) – 5.0 mm
    • Maximum gingival height (for Straumann BLX PREFIT) – 6.0 mm
    • Maximum angle – 30°
    AI/ML Overview

    The provided document is a 510(k) clearance letter for a dental device, specifically dental implant abutments. It details the device's administrative information, predicate devices, indications for use, subject device description, and performance data used to demonstrate substantial equivalence.

    However, the document does not contain any information regarding acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML algorithm. The performance data section refers to non-clinical analyses, such as MR compatibility, sterilization validation, biocompatibility, mechanical properties testing, reverse engineering, and static/dynamic compression-bending testing conforming to ISO 14801. These are standard tests for mechanical dental devices, not AI/ML performance evaluation.

    Therefore, I cannot provide the requested information for acceptance criteria and a study proving an AI device meets them based on the given text. The device described, "Mist IC," is a physical dental implant abutment, not an artificial intelligence/machine learning (AI/ML) powered medical device.

    To directly answer your prompt based on the provided text, the following information is missing or not applicable:

    1. A table of acceptance criteria and the reported device performance: Not applicable. The document describes mechanical and material performance testing for a physical device, not AI/ML performance metrics like sensitivity, specificity, or AUC against defined acceptance criteria for an AI algorithm.
    2. Sample sizes used for the test set and the data provenance: Not applicable. There is no AI/ML test set. The document refers to testing of physical components (OEM implants, abutments, screws, subject device constructs).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No AI/ML ground truth establishment.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" for this device's performance would be the physical properties and mechanical integrity verified by engineering standards (e.g., ISO 14801).
    8. The sample size for the training set: Not applicable. No AI/ML training set mentioned.
    9. How the ground truth for the training set was established: Not applicable. No AI/ML training set mentioned.

    The document focuses solely on the substantial equivalence of the "Mist IC" dental implant abutments to legally marketed predicate devices based on design, materials, manufacturing, and conventional mechanical/material performance testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243530

    Validate with FDA (Live)

    Device Name
    Dynamic TiBase
    Date Cleared
    2025-05-30

    (197 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamic TiBase abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Compatible Implant Systems:

    Compatible Implant System (Connection)Implant Body Diameter, mmImplant Platform, mm
    Osstem® TS Hiossen® ET (Internal Taper)3.5Mini
    4.0, 4.5, 5.0, 5.5, 6.0, 7.0Regular
    Neodent GM (Morse taper)3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0GM
    Nobel Active (Conical Connection)3.5NP
    4.3, 5.0RP
    5.5WP
    Straumann Bone Level (CrossFit® Morse Taper)3.3NC
    4.1, 4.8RC
    Straumann BLX (TorcFit™ Internal Hexalobular)3.5, 3.75, 4.0, 4.5RB
    5.0, 5.5, 6.5WB

    All digitally designed custom abutments for use with Dynamic TiBase abutments are to be sent to a Talladium Medical validated milling center for manufacture.

    Device Description

    Dynamic TiBase abutments are two-piece abutments composed of a CAD-CAM fabricated zirconia superstructure and a prefabricated titanium base component where the final two-piece abutment (base and cemented superstructure) is the finished device used for the prosthetic restoration. All subject device bases are made of titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3 and ASTM F136. The Dynamic TiBase abutments are provided in engaging and non-engaging designs for single-unit and multi-unit restorations, respectively.

    For each of the compatible OEM implant lines, the prefabricated titanium base components are provided with a gingival height (in the titanium base) ranging from 0.3 mm to 4 mm, and a platform diameter ranging from 4.30 mm to 5.50 mm. Angulation and additional gingival height may be provided in the zirconia superstructure. All Dynamic TiBase prefabricated titanium base components have a post with a cut-out to accommodate a restoration with an angled channel for screw access when clinically necessary. The post height of the prefabricated titanium base component ranges from 3.8 mm to 5.40 mm, and from 2.3 mm to 3.8 mm (cut-out height). The cementable post height of the final patient-matched abutment design, measured above the total combined gingival collar, shall be no less than 4 mm.

    All zirconia superstructures (copings) used to complete the final two-piece subject device Dynamic TiBase abutment will be made at a Talladium España, SL validated milling center under FDA quality system regulations, and the material will conform to ISO 13356.

    The design parameters for the CAD-CAM zirconia superstructure for the Dynamic TiBase vary slightly among the compatible OEM implants. The design parameters for the CAD-CAM zirconia superstructure are summarized in the following table:

    Implant Compatibility (1)Minimum Wall Thickness, mmMaximum Gingival Height, mmMinimum Gingival Height (2), mmMaximum Angulation
    Osstem® TS Hiossen® ET (Internal Taper)0.325.181.225º
    Neodent GM (Morse taper)0.325.201.230º
    Nobel Active (Conical Connection)0.356.200.30
    Straumann Bone Level (CrossFit® Morse Taper)0.325.181.130º
    Straumann BLX (TorcFit™ Internal Hexalobular)0.325.341.530º

    (1) for the compatible sizes shown in Table 1
    (2) minimum gingival height in the titanium base, not the zirconia superstructure

    The required cement for bonding the zirconia superstructure to the Dynamic TiBases to create the final two-piece abutment is Nova Resin Cement cleared in K213609.

    Also, the subject of this submission are seven (7) abutment screws for use with the subject abutments.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a dental implant abutment — the Dynamic TiBase — not an AI-powered diagnostic device or software. Therefore, the information typically requested about acceptance criteria and study designs for validating AI/ML-based medical devices (such as sample size, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details) is not applicable to this document.

    The "performance data" section in this 510(k) summary refers to traditional engineering and biocompatibility testing for a physical medical device, not performance metrics for an algorithm.

    Here's how to interpret the provided document in the context of "acceptance criteria" and "proof":

    Acceptance Criteria and Reported Device Performance (as inferred for a physical device):

    For a physical device like the Dynamic TiBase, the "acceptance criteria" are generally met through demonstrating substantial equivalence to a previously cleared predicate device. This involves validating material properties, manufacturing processes, functional performance (e.g., mechanical strength, compatibility), and biocompatibility.

    The "study that proves the device meets the acceptance criteria" refers to the non-clinical testing performed to demonstrate substantial equivalence.

    Acceptance Criterion (Inferred for a Dental Abutment)Reported Device Performance / Proof
    Biocompatibility (Device materials are safe for biological contact)Referenced from K232151 (provided in K212108): Biocompatibility testing according to ISO 10993-5 (cytotoxicity) for abutment materials ASTM F136 (titanium alloy) and ISO 13356 (zirconia). This indicates the materials did not show cytotoxic effects when tested according to the standard.
    Mechanical Performance/Strength (Abutment can withstand oral forces without failure)Provided in this submission: Mechanical testing conducted according to ISO 14801 to support the performance of the subject device abutments in conjunction with the compatible OEM implants. This standard specifically evaluates the fatigue strength of endosseous dental implants. The successful completion of this testing, validated against the predicate device, implies the device meets required load-bearing capacities.
    MRI Safety (Device does not pose undue risk in MRI environment)Provided in this submission: Non-clinical worst-case MRI review to evaluate subject device components in the MR environment using scientific rationale and published literature. This included addressing magnetically induced displacement force and torque parameters as per FDA guidance. The review concluded that the device is safe for MRI under specified conditions.
    Sterilization Efficacy (Device supplied non-sterile can be effectively sterilized by user)Referenced from K232151 (provided in K212108): Moist heat sterilization validated to a sterility assurance level (SAL) of 10-6 by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2. This demonstrates that the specified sterilization method for end-users achieves the required sterility.
    Dimensional Compatibility & Fit (Device correctly interfaces with specified implants)Provided in this submission: Reverse engineering dimensional analysis (of OEM implant bodies, OEM abutments, and OEM abutment screws) to demonstrate that the subject device abutments are compatible with listed OEM Implant Systems (Osstem® TS/ Hiossen® ET, Neodent GM, NobelActive, Straumann Bone Level, and Straumann BLX Implant Systems). This confirms the physical design correctly matches the intended implant systems.
    Material Conformance (Materials meet specified standards)Materials confirmed to conform to: Ti-6Al-4V (ASTM F136, ISO 5832-3) for titanium components; Zirconia (ISO 13356) for superstructure; Nova Resin Cement (cleared in K213609) for bonding. This demonstrates the use of medical-grade, standard-compliant materials.
    Manufacturing Quality (Device is produced under controlled conditions)Stated that zirconia superstructures will be made at a Talladium España, SL validated milling center under FDA quality system regulations. This implies adherence to Good Manufacturing Practices (GMP).

    Regarding the specific questions about an AI/ML context:

    1. A table of acceptance criteria and the reported device performance: Provided above, adapted for a physical medical device.
    2. Sample size used for the test set and the data provenance: Not applicable. The "test set" for this physical device refers to the number of physical abutment samples or material samples subjected to mechanical, biocompatibility, and MRI testing. The document does not specify exact sample numbers for these engineering tests, only the standards used (e.g., ISO 14801 typically specifies minimum sample sizes). Data provenance is "non-clinical data" generated from laboratory testing.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a dental abutment's performance is established by engineering standards, material specifications, and physical testing, not by expert human interpretation of images or clinical outcomes in the same way as an AI diagnostic.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This is a concept used in evaluating human reader performance in AI studies.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for AI-assisted diagnostic devices.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This refers to an AI algorithm's performance.
    7. The type of ground truth used: For a physical device, "ground truth" is defined by adherence to established engineering standards (e.g., ISO 14801 for mechanical strength), material properties, and biocompatibility standards. "Reverse engineering dimensional analysis" served as a form of "ground truth" for compatibility. No pathology or outcomes data was used for this premarket notification.
    8. The sample size for the training set: Not applicable. There is no "training set" as this is not an AI/ML device.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, this 510(k) clearance is for a physical medical device (dental abutment), and the "performance data" section details the engineering and material testing conducted to demonstrate its safety and effectiveness, primarily through substantial equivalence to previously cleared devices. It does not involve AI/ML validation methodologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242542

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-22

    (239 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PreFace abutment, TI-Forms abutment, Titanium base 2nd generation, and Titanium base ASC Flex are intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. Abutment-level prosthetic components (Multi-unit Titanium Base, Multi-unit Titanium Cap, MedentiBASE Titanium Base) are intended for use as a support for multi-unit screw-retained bridges and bars in the maxilla or mandible of a partially or fully edentulous patient.

    All digitally designed abutments for use with PreFace abutment, TI-Forms abutment, Titanium base 2nd generation, Titanium base ASC Flex, Multi-unit Titanium Base, Multi-unit Titanium Cap, and MedentiBASE Titanium Base are intended to be sent to an FDA-registered Medentika validated milling center for manufacture or to be manufactured according to the digital dentistry workflow, which integrates multiple components: Scans from desktop and intra oral scanners, CAD and CAM software and milling machine with associated accessories.

    Medentika abutments for the Nobel Biocare Nobel Active® 3.0 mm, Dentsply Sirona Astra Tech OsseoSpeed EV® 3.0 mm and TX® 3.0 mm, Straumann Bone Level 2.9 implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.

    Device Description

    The subject devices are Medentika CAD/CAM Abutments, which primarily expand the options for fabricating patient-specific final abutments from a "validated milling center" to a "digital dentistry workflow". This workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machines, and associated tooling and accessories. The devices include Titanium Base abutments, Titanium base ASC Flex abutments, and PreFace and TI-Form (blanks) abutments.

    The abutments are made of titanium alloy (Ti-6Al-4V ELI). Titanium base abutments also include a zirconia superstructure. The specified zirconia materials for milling superstructures are Ivoclar Vivadent IPS e.max ZirCAD Prime, Ivoclar Vivadent IPS e.max ZirCAD Prime Esthetic, Amann Girrbach Zolid Bion, Amann Girrbach Zolid Gen-X, and Institut Straumann AG n!ce Zirkonia HT. The specified cement for bonding superstructures is Multilink Hybrid Abutment Cement from Ivoclar Vivadent AG.

    Key design parameters for CAD/CAM zirconia superstructures (on Titanium base and Titanium base ASC Flex) include: minimum wall thickness of 0.5 mm, minimum cementable post height of 4.0 mm for single unit restorations, maximum gingival margin height of 5.0 mm, minimum gingival margin height of 0.5 mm, and maximum angulation of the final abutment of 30°.

    PreFace and TI-Forms abutments (blanks) are used by dental laboratories to fabricate customized abutments from titanium alloy. Their design parameters include: minimum wall thickness of 0.4 mm, minimum cementable post height of 4.0 mm, maximum gingival margin height of 5.0 mm, minimum gingival margin height of 0.5 mm, and maximum angulation of 30°.

    Prosthetic-level components (Multi-unit Titanium Base, Multi-unit Titanium Cap, MedentiBASE Titanium Base) are provided for use with previously cleared Medentika multi-unit abutments and MedentiBASE abutments.

    All abutments are provided non-sterile with appropriate abutment screws. The screws attach the abutment to the implant or the prosthesis to the abutment.

    AI/ML Overview

    The provided 510(k) clearance letter and summary describe a medical device, Medentika CAD/CAM Abutments, and its substantial equivalence to predicate devices based on non-clinical performance data. The document does not contain information about acceptance criteria or performance data for an AI/ML-based device, nor does it detail a clinical study involving human readers or expert consensus for ground truth.

    Therefore, for the information requested in your prompt, I can only extract what is presented in the document, which pertains to the non-AI aspects of device acceptance and testing. Many of the points specifically refer to AI/MRMC studies, which are not applicable to this document.

    Here's an analysis based on the provided text:

    Device Description and Purpose:
    The device is "Medentika CAD/CAM Abutments," which are dental implant abutments. The primary purpose of this submission is to expand the fabrication options for patient-specific final abutments from a "validated milling center" to a "digital dentistry workflow" that integrates CAD/CAM software and milling machines. It also adds new sizes and OEM compatibilities.

    Study Type:
    This is a pre-market notification (510(k)) submission seeking substantial equivalence to existing legally marketed devices. It relies heavily on non-clinical performance data to demonstrate that the new manufacturing workflow and expanded compatibilities do not raise new questions of safety or effectiveness.


    Analysis of Requested Information (based on the provided document):

    1. A table of acceptance criteria and the reported device performance:

    The document outlines various performance tests conducted to demonstrate substantial equivalence, but it does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance." Instead, it states that the tests demonstrate sufficient strength or ensure accuracy and reliability.

    Here's a summary of the performance tests and their implied purpose:

    Performance Test CategoryPurpose / Implied Acceptance CriteriaReported Device Performance (Summary)
    Biocompatibility TestingTo ensure the device materials (titanium alloy, zirconia, cement) are safe for use in the human body."Biocompatibility testing of final finished devices... according to ISO 10993-1, ISO 10993-5, and ISO 10993-12" was provided. Implied: The device passed these tests.
    Mechanical Testing (ISO 14801)To demonstrate that the abutments, including zirconia and cement, in combination with compatible implants, have sufficient strength for intended use."Mechanical testing conducted according to ISO 14801... to demonstrate that the subject Medentika CAD/CAM Abutments... have sufficient strength for the intended use" was provided. Implied: The device demonstrated sufficient strength.
    Reverse Engineering Dimensional AnalysisFor new OEM compatibilities, to show that the subject device abutments are compatible with the respective OEM implants."Reverse engineering dimensional analysis for the OEM compatibilities... to demonstrate that the subject device abutments are compatible with the respective OEM implants" was provided. Implied: Compatibility was demonstrated.
    CAD Software ValidationTo demonstrate that maximum and minimum design parameters for the subject devices are locked into the design software and available libraries."Validation of CAD software to demonstrate that the maximum and minimum design parameters... are locked into the design software and available libraries" was provided. Implied: Software validation confirmed design parameter locking.
    CAM Software & Milling Machine ValidationTo ensure the accuracy and reliability of the milling process (verified NC file imports, milling tools, materials, milling strategies, post-processing)."Validation of CAM software and milling machines to ensure the accuracy and reliability of the milling process" was provided. Implied: Accuracy and reliability were confirmed.
    CAM Restriction Zones ValidationTo show avoidance of damage or modification of the connection geometry and locking of restriction zones from user editing in the CAM software."Validation testing of CAM restriction zones to show avoidance of damage or modification of the connection geometry and locking of restriction zones..." was provided. Implied: Restriction zones prevent damage.
    MR Environment AnalysisTo evaluate device compatibility in a Magnetic Resonance (MR) environment."Non-clinical analysis and testing to evaluate the metallic subject devices and compatible dental implants in the MR environment" was referenced from K180564. Implied: Device is compatible or safe in MR environment.
    Sterilization ValidationTo ensure non-sterile devices can be sterilized by the end-user to a specific sterility assurance level."Moist heat sterilization for subject devices provided non-sterile to the end user, validated to a sterility assurance level of 10-6 by the overkill method according to ISO 17665-1 and ISO TR 17665-2" was referenced. Implied: Sterilization method is effective.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document does not specify numerical sample sizes for any of the non-clinical tests (e.g., how many abutments were mechanically tested, how many software validation tests were run). It simply states that "testing was conducted" or "validation was performed."
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of pre-market non-clinical testing for medical devices, these are typically prospective laboratory tests conducted by the manufacturer or accredited testing facilities. The manufacturer is Medentika® GmbH (Huegelsheim, Germany), suggesting the testing likely occurred in Germany or at internationally recognized labs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to the provided document. The ground truth for this device is established through engineering specifications, material standards (e.g., ASTM F136), and validated manufacturing processes, not through human expert consensus on diagnostic images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable to the provided document, as it describes non-clinical engineering and manufacturing validation, not a multi-reader clinical study for AI.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable to the provided document. This device is a physical dental abutment and its associated CAD/CAM workflow, not an AI-based diagnostic tool that would require human reader studies. The document explicitly states: "No clinical data were included in this submission."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable to the provided document. While the device utilizes CAD/CAM software and milling machines, it is a physical product manufactured through a workflow, not a standalone AI algorithm whose performance needs to be assessed in isolation. The software functions as a design and manufacturing aid, not a diagnostic or decision-making algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" in this context refers to established engineering and material standards:

    • Standards Compliance: Adherence to ISO standards (e.g., ISO 14801 for mechanical strength, ISO 10993 for biocompatibility, ISO 17665 for sterilization).
    • Dimensional Accuracy: Verification against established design parameters and compatibility specifications for dental implants (e.g., OEM implant body and abutment dimensions).
    • Material Specifications: Conformance to ASTM F136 for titanium alloy and specifications for zirconia and cement.
    • Software Design Parameters: The "ground truth" for the CAD software validation is the pre-defined maximum and minimum design parameters that the software must enforce.

    8. The sample size for the training set:

    This information is not applicable to the provided document. The "device" in question is a physical dental abutment and its manufacturing workflow, not an AI/ML model that requires a training set. The CAD/CAM software itself is validated, not "trained" on a dataset in the AI sense.

    9. How the ground truth for the training set was established:

    This information is not applicable to the provided document for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243478

    Validate with FDA (Live)

    Date Cleared
    2025-02-19

    (103 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Patient-specific abutment restorations, milled from Pre-milled Abutment Blanks (PMABs), are indicated for single tooth replacement and multiple tooth restorations. They are directly connected to various endosseous dental implant systems using a basal screw. Patient-specific abutment restorations milled from Pre-milled Abutment Blanks are to be digitally designed and milled using the Straumann InLab Validated Workflow. The Straumann InLab Validated Workflow is indicated for the design and fabrication of single or multiple-unit implant-borne prosthetics for the restoration of partially or fully edentulous mandibles and maxillae. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners or Extra-Oral Scanners. CAD software, CAM software, pre-milled abutment blanks, milling machines and associated tooling and accessories.

    Device Description

    The Straumann InLab Validated Workflow is similar to the primary predicate K171649. It employs optical impression files that document the topographical characteristics of teeth, traditional dental impressions, or stone models. The 3Shape CAD software then allows the design of the desired restorations.

    The CAM software converts the digital restoration design into the tooling and tool path commands needed to fabricate the restoration. When choosing the Straumann Validated workflow, the user will only see the available and cleared components which were tested and demonstrated as part of the validated workflow. The milling command file is encrypted prior to transfer to the Roland DWX-42W Plus milling System; this encryption ensures that files generated using other CAD or CAM software cannot be used with the Straumann InLab Validated Workflow. The user will then load the milling command file into the Roland DWX-42W Plus milling System where it is decoded. The user loads the appropriate dental material blank and initiates the milling operation.

    This premarket notification includes restorations (one-piece metal patient-specific abutment restorations) manufactured from various Pre-milled Abutment Blanks (PMABs) from the Straumann Group companies : Institut Straumann AG and Neodent PMABs. The digital workflow using the Straumann InLab Validated Workflow includes the use of the following products: Dental Scan of the patient's situation, CAD Software, CAM Module, Milling System, Abutment Milling Blanks.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Straumann InLab Validated Workflow, a device for designing and manufacturing patient-specific dental abutments from pre-milled abutment blanks (PMABs). The document describes the device, its intended use, technological characteristics, and performance testing.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria and reported device performance in a consolidated format as typically seen in a clinical study report. However, it mentions key performance areas and states that requirements were met.

    Acceptance Criteria (Implicit from tests performed):

    • Sterilization Validation: Meet ISO 17665-1 and FDA guidance for reprocessing medical devices.
    • Biocompatibility: In line with ISO 10993-1 and FDA guidance for biological evaluation.
    • Electromagnetic Compatibility (EMC): Considered MR Conditional.
    • Milling Accuracy: Dimensions of milled restoration are the same as the intended CAD design, met within tool's expected lifetime.
    • Simulated Use Validation: Correct implementation of PMABs, design constraints, and workflow restrictions in software and libraries.
    • Implant-Abutment Connection Protection: No damage to the connection geometry during milling.
    • Dynamic Fatigue: Performance consistent with FDA guidance for dental implants/abutments.

    Reported Device Performance:

    Performance AreaReported Device Performance
    Sterilization ValidationThe recommended sterilization method (moist heat/steam) was validated according to ISO 17665-1 and applicable FDA guidance on reprocessing medical devices. Sterilization parameters are equivalent.
    Biocompatibility TestingBiological assessment performed per ISO 10993-1 and FDA guidance. Device materials and manufacturing processes are equivalent to predicate, raising no new biocompatibility issues.
    Electromagnetic CompatibilityNo significant changes from currently marketed predicate devices; no new EMC issues raised. Device can be considered MR Conditional.
    Milling AccuracyBlack-box validation confirmed that dimensions of the milled restoration are the same as the intended CAD design. The accuracy requirement was met considering the tool's expected lifetime.
    Simulated Use ValidationConfirmed that relevant PMABs, design constraints, and workflow restrictions are correctly implemented (adequately written and locked into compatible design software and available libraries).
    Implant-Abutment Connection ProtectionA simulated implant-abutment connection protection test was conducted to mitigate potential risk of damaging the implant-abutment connection geometry during milling.
    Dynamic FatigueThe device design and performance testing submitted or referenced were conducted according to FDA guidance for root-form endosseous dental implants and abutments.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for the test sets in the performance studies (e.g., dynamic fatigue, milling accuracy, simulated use validation). It refers to these as "Performance Testing - Bench."

    Data Provenance: The studies are described as "nonclinical tests" and "bench" testing, suggesting they were conducted in a laboratory or simulated environment, rather than involving patient data. The document does not specify country of origin for the data (beyond the applicant being Institut Straumann AG in Switzerland and JJGC Indústria e Comércio de Materiais Dentários AS in Brazil, with contact in Andover, MA, USA). The studies appear to be prospective, designed specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish ground truth for the test set. The validation methods described (e.g., milling accuracy, simulated use) imply technical verification against design specifications and functional requirements rather than expert consensus on diagnostic or clinical outcomes.

    4. Adjudication Method for the Test Set

    No adjudication method is mentioned, as there is no indication of multiple reviewers or subjective assessments requiring adjudication in the described bench tests.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned. The device is a workflow for manufacturing dental restorations, not an AI diagnostic or assistive tool for human readers in the traditional sense of comparing human performance with and without AI assistance for interpretation.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)

    The milling accuracy test ("black-box validation confirming that the dimensions of the milled restoration are the same as the intended CAD design") and the simulated use validation implicitly test the algorithm and workflow in a standalone manner, separate from a human operator's interpretative judgment. The "Straumann InLab Validated Workflow" itself is largely an automated process once the design is finalized. It is designed to ensure that the digital design translates accurately into a physical product.

    7. Type of Ground Truth Used

    The ground truth used for these performance tests appears to be:

    • Design Intent/Specifications: For milling accuracy, the ground truth is the CAD design itself. For simulated use, the ground truth refers to the correct implementation of design constraints and workflow restrictions within the software and libraries.
    • Industry Standards and Regulatory Guidance: For sterilization, biocompatibility, and dynamic fatigue, the ground truth is defined by recognized international standards (ISO) and FDA guidance documents.
    • Functional Requirements: For implant-abutment connection protection, the ground truth is the prevention of damage to the connection geometry.

    8. Sample Size for the Training Set

    The document does not describe the use of machine learning or AI models in a way that would involve a distinct "training set" in the context of typical software or diagnostic AI submissions. The workflow involves CAD/CAM software where parameters and constraints are defined by engineering and design principles rather than learned from a large training dataset.

    9. How the Ground Truth for the Training Set Was Established

    As no specific training set for a machine learning model is described, there is no information on how its ground truth would have been established. The workflow relies on validated engineering designs, material properties, and manufacturing processes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242064

    Validate with FDA (Live)

    Date Cleared
    2025-01-17

    (186 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Omnibut is a pre-manufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

    Device Description

    The Omnibut™ is a transmucosal abutment used to support screw-retained prostheses on four or more implants. The subject device has a premanufactured connection for the platforms listed in Table 1 Compatible Implant Systems.

    The system involves a ball abutment attached to an implant. A retention attachment allows for angle corrections of up to 30° off the implant axis. The ball abutment is inserted into the attachment is adjusted to the desired angle using an orientation screw. The abutment supports prostheses that connect via titanium cylinders, which are incorporated into resin or ceramic prostheses. Finally, the prostheses are retained to the abutment by prosthetic screws.

    The subject device abutments and system components are manufactured from Ti-6Al-4V alloy conforming to ASTM F136. The subject device is a single use device is provided nonsterile and intended to be sterilized by the user prior to placement in the patient.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Omni-Directional Multi-unit Abutment System (Omnibut™), based on the provided FDA 510(k) summary:

    Description of the Device

    The Omnibut™ is a pre-manufactured prosthetic component directly connected to endosseous dental implants. It is intended for use as an aid in prosthetic rehabilitation, specifically for supporting screw-retained prostheses on four or more implants. The system includes a ball abutment attached to an implant, with a retention attachment allowing for angle corrections of up to 30° off the implant axis. It supports prostheses that connect via titanium cylinders, which are incorporated into resin or ceramic prostheses. The device is made from Ti-6Al-4V alloy and is provided non-sterile, requiring user sterilization.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Test)Description of TestReported Device Performance (Results)
    Biological Evaluation
    ISO 10993-1:2018; FDA Guidance on the Use of ISO 10993-1, 2023Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management ProcessPASS – All biological endpoint testing performed on the device, along with the analysis on the physical and chemical information, returned passing results. All biological endpoint testing suggests that the Omnibut is biocompatible and does not present a foreseen biological risk to those patient populations it is intended for.
    ISO 10993-5 (2009)Biological Evaluation of Medical Devices – Part 5: Tests for in vitro cytotoxicityPASS – The test article showed no evidence of causing cell lysis or toxicity. The test article met the requirements of the test since the grade was grade 0 (no reactivity).
    Sterilization Validation
    AAMI TIR12:2020; ANSI/AAMI/ISO 17665-1:2026/(R)2013; ANSI/AAMI ST79:2017Designing, Testing, And Labeling Medical Devices Intended For Processing By Health Care Facilities: A Guide For Device Manufacturers, Overkill method according to Section 5.7PASS – Results from testing have demonstrated that the Omnibut was able to achieve a 10-6 SAL when using the recommended parameters in the Instructions for Use (IFU).
    Mechanical Performance
    ISO 14801:2016; FDA Guidance Document, "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" (May 12, 2004) Section 8Dentistry — Implants — Dynamic loading test for endosseous dental implantsPASS – The results conclude that when evaluated in a manner consistent with ISO 14801:2016, the Omnibut met all predetermined acceptance criteria.
    No specific standard (Compatibility)Reverse engineering dimensional analysis of OEM implant bodies, OEM abutments, and OEM abutment screws were performed to demonstrate that the Omnibut abutments are compatible with the noted implant systems. Compatibility must be demonstrated to ensure proper device functionality.PASS – The engineering and dimensional analysis concluded that each Omnibut design is compatible with the applicable implant connection. PASS - Omnibut prosthetic side components were demonstrated to be compatible.
    No specific standard (Retention Force Testing)The Omnibut has a retention attachment. Retention Force testing was performed via tensile push-out to ensure that the attachment will not detach during clinical use.PASS – The Omnibut retention attachment did not detach at a predetermined acceptable force.
    No specific standard (Simulated-Use Testing)Simulated Use of four Omnibuts with components connected to a Titanium Bar under a clinically relevant cyclic load.PASS – The Omnibut and components did not yield, deform, or fracture after fatigue testing.
    No specific standard (Simulated Cleaning Testing)Simulated cleaning of Omnibuts in a fixture with a clinically worst case cleansibility construction.PASS – All parts of the Omnibuts were clean of soil indicators after six soilage and cleaning cycles.
    MR Safety
    FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (May 2021)Non-clinical analysis to evaluate the subject device in the MR environment using scientific rationale and published literature (TO Woods, JG Delfino, and S Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Evaluation, Volume 49, No. 2, 2021, pp. 783-795).The analysis addressed parameters per the FDA guidance, including magnetically induced displacement force and torque. The document does not explicitly state the results (e.g., "MR Safe" or "MR Compatible"), but implies that the analysis was sufficient to support safety in the MR environment based on the scientific rationale and literature used.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify the exact sample sizes for each individual non-clinical test (e.g., number of abutments tested for dynamic fatigue, retention force, simulated use, or cleaning). It refers to the testing as "non-clinical" bench testing.

    • Test Set Provenance: The data is generated from bench testing (laboratory studies), not from clinical data involving human patients. Therefore, information like "country of origin of the data" or "retrospective/prospective" is not applicable in the typical sense of clinical trials. The testing was performed in vitro.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the studies are non-clinical bench tests. The "ground truth" for these tests is based on established engineering standards (e.g., ISO 14801), biological evaluation standards (ISO 10993), and internal company protocols for mechanical and cleaning validation, not on expert clinical interpretation of patient data.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable as the studies are non-clinical bench tests. Adjudication methods are typically employed in studies involving human interpretation or clinical endpoints to resolve discrepancies in assessment.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices involving human readers/interpreters, which is not the case for this dental implant abutment. The device is a physical component, not a diagnostic AI tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    A standalone performance study was not done in the context of an algorithm. This question is relevant for AI/software devices; this device is a physical medical device. The "standalone" performance here refers to the device's mechanical and biological performance on its own, which is what the bench tests evaluate.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical performance evaluations relies on:

    • Established engineering and biological standards: e.g., ISO 14801:2016 for dynamic fatigue, ISO 10993 for biocompatibility, ANSI/AAMI/ISO for sterilization.
    • Predetermined acceptance criteria: For retention force, simulated use, and cleaning efficacy, the "ground truth" is defined by specific pass/fail criteria established during the test design based on expected clinical performance and safety.
    • Dimensional accuracy and compatibility models: For compatibility testing, the "ground truth" is established by the dimensions and specifications of OEM implant bodies and abutments.

    8. The Sample Size for the Training Set

    This information is not applicable because the device is a physical medical component, not a machine learning model or AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the same reason as above (not a machine learning model).

    Ask a Question

    Ask a specific question about this device

    K Number
    K241616

    Validate with FDA (Live)

    Date Cleared
    2024-12-17

    (195 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ALLONUS Tech Prosthetic is intended for use with dental implants as a support for single or multiple-unit prosthetic restorations in the maxilla or mandible of partially or fully edentulous patient. It is including; cemented retained, screw-retained, or overdenture restorations.

    It is compatible with the following systems:

    • · Astra OsseoSpeed EV(K130999) 3.0
    • · Astra OsseoSpeed EV(K120414) 3.6, 4.2, 4.8, 5.4 mm
    • Tapered Internal Implants (K071638) (K143022) 3.4. 3.8. 4.6. 5.8 mm
    • · BioHorizons Laser-Lok Implant System (K093321) 3.0 mm
    • · Conelog Screw-Line (K113779) 3.3, 3.8, 4.3, 5.0 mm
    • Osstem TSIII SA (K121995) 3.5 (3.7) , 4.0 (4.2) , 4.5 (4.6) , 5.0 (5.1), 6.0 (6.0), 7.0 (6.8) mm (Mini. Regular)
    • · Megagen AnyRidge Internal Implant System (K140091) 4.0, 4.4, 4.9, 5.4 (3.1)
    • · Neodent Implant System GM Helix (K163194, K180536) 3.5, 3.75, 4.0, 4.3, 5.0 (3.0) 6.0 (3.0)
    • · Nobel Active 3.0 (K102436) 3.0
    • · Nobel Active Internal Connection Implant (K071370) NP RP
    • · Nobelactive Wide Platform (Wp) (K133731) WP
    • Straumann BLX Implant (K173961, K181703, K191256) 3.5, 3.75, 4.5, 5.5, 6.5 (RB, WB)
    • · Straumann 02.9 mm Bone Level Tapered Implants, SC CARES Abutments (K162890) 2.9 (SC)
    • · Straumann® Bone Level Tapered Implants (K140878) 3.3, 4.1, 4.8 (NC, RC)
    • · Zimmer 3.1mmD Dental Implant System (K142082) 3.1 (2.9)
    • (Ti-base only) Screw Vent® and Tapered Screw Vent® (K013227) 3.7(3.5), 4.7(4.5), 6.0(5.7)

    All digitally designed abutments and/or coping for use with the abutments are intended to be sent to a ALLONUS Tech-validated milling center for manufacture.

    Device Description

    ALLONUS Tech Prosthetic is made of titanium alloy (Ti-6AI-4V ELI, ASTM F136) intended for use as an aid in prosthetic restoration. It consists of Pre-Milled Blank abutment, Ti-Base Abutment, and Multi-unit Abutment and Components (Multi-unit Healing Cap, Multi-unit Temporary cylinder, Multi-unit Ti-cylinder).

    Pre-Milled Blank has a pre-manufactured implant interface connection interface with a customizable cylindrical area-by CAD/CAM- above the implant-abutment interface.

    Ti Base consists of a two-piece abutment, where the titanium base is a pre-manufactured component of the abutment that will be used to support a CAD/CAM-designed zirconia superstructure (the second part of the two-piece abutment) that composes the final abutment.

    Multi-unit Abutment which are placed into the dental implant to provide support for the prosthetic restoration. The abutments are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F-136). Multi-unit Abutment includes abutments and components (Multi-unit Healing Cap, Multi-unit Temporary cylinder, Multi-unit Ti-cylinder). Multi-unit Abutment - are provided in various gingival cuff height ranging from 1 to 6 mm.

    All digitally designed abutments and/or coping for use with the abutments are intended to be sent to a ALLONUS Tech-validated milling center for manufacture.

    AI/ML Overview

    The provided text is a 510(k) Summary for the ALLONUS Tech Prosthetic, which is an endosseous dental implant abutment. It details the device's technical characteristics and compares it to predicate devices to establish substantial equivalence, rather than describing a study that proves the device meets specific acceptance criteria for a new and novel performance claim.

    Therefore, many of the requested categories for acceptance criteria and study details are not directly applicable or available in this type of submission. The information provided primarily focuses on demonstrating equivalence through comparison to existing legally marketed devices, material properties, and standard performance tests for similar devices.

    However, I can extract the relevant information that is available from the document for each type of device within the ALLONUS Tech Prosthetic family: Pre-Milled Blank, Ti Base, Multi-Unit Abutment, Multi-Unit Healing Cap, Multi-Unit Temporary Cylinder, and Multi-Unit Ti-Cylinder.


    General Information on Acceptance Criteria and Studies for ALLONUS Tech Prosthetic (as inferred from the 510(k) Summary):

    The ALLONUS Tech Prosthetic is claiming substantial equivalence to predicate devices, meaning it does not need to establish new performance criteria but rather demonstrate that it is as safe and effective as existing legally marketed devices. The "acceptance criteria" here are largely implied by the performance of the predicate devices and the relevant ISO standards for dental implant abutments. The studies performed are non-clinical bench tests.

    1. Table of Acceptance Criteria (Inferred from Comparison) and Reported Device Performance:

    The document presents comparisons to predicate devices to establish substantial equivalence rather than explicit acceptance criteria and corresponding performance metrics for novel claims. However, the design limits of the subject devices are compared to the design limits of the predicate devices, which act as de facto acceptance criteria in the context of substantial equivalence. The device's performance is demonstrated by meeting the standards in non-clinical testing.

    Here's a generalized table summarizing this approach:

    Feature/TestAcceptance Criteria (Based on Predicate/Standards)Reported Device Performance (Subject Device)
    MaterialsTi-6Al-4V ELI (ASTM F136) for metallic components, Zirconia Oxide for Ti-Base superstructureTi-6Al-4V ELI (ASTM F136) for Pre-Milled Blank, Multi-Unit Abutment, Healing Cap, Temporary Cylinder, Ti-Cylinder. Ti-6Al-4V ELI (ASTM F136) and Zirconia Oxide (InCoris Zi) for Ti Base.
    Fatigue Test (ISO 14801:2016)Met criteria of the standard.Met criteria of the standard. (Page 23)
    End-User Steam Sterilization (ISO 17665-1:2006, 17665-2:2009, ANSI/AAMI ST79:2010)Met criteria of the standard.Met criteria of the standard. (Page 23)
    Biocompatibility (ISO 10993-1:2009, -5:2009, -10:2010)Met criteria of the standard.Met criteria of the standard. (Page 23)
    Implant-to-Abutment CompatibilityDemonstrated by reverse engineering and assessment of OEM components.Demonstrated compatibility with various OEM implant systems (listed in Indications for Use). Reverse engineering included assessment of maximum and minimum dimensions of critical design aspects and tolerances of OEM implant body, OEM abutment, OEM abutment screw, along with cross-sectional images of the subject device and compatible implant body. (Page 24)
    MR Safety (FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment")Demonstrated substantial equivalence to predicate devices using scientific rationale and published literature for magnetically induced displacement force and torque.Non-clinical worst-case MRI review performed; rationale addressed parameters per FDA guidance. Results demonstrated substantial equivalence to predicate devices. (Page 24)
    Design Limits (examples)Pre-Milled Blank: Post Angle: 0-30°, Diameter: 3.0-7.0 mm (Predicate) Ti Base: Post Angle: 0-30°, Diameter: 3.0-7.0 mm (Predicate) Multi-Unit Abutment: Diameter: 4.8 mm, Gingival Height: 1.5-4.5 mm, Angle: 0, 17, 30° (Reference Device)Pre-Milled Blank: Design limits are "slightly different" but within diameter and angle range of predicate. (Page 11) Specific limits are provided in a table on page 7 for various compatible systems (e.g., Min Gingival Height: 0.5 mm, Max Gingival Height: 5.0-6.0 mm, Min Wall Thickness: 0.3-0.6 mm, Min Post Height: 4.0-4.3 mm). Ti Base: Post Angle: 0-15°, Gingival Height: 0.5-5.0 mm, Post Height: 4.0-6.0 mm, Diameter: 5.0-8.0 mm, Thickness: 0.4 mm. These are "slightly different" but within diameter and angle range of predicate. (Page 17) Multi-Unit Abutment: Diameter: 4.8 mm, Gingival Height: 1, 2, 3, 4, 5, 6 mm, Angle: 0, 17, 29°. Diameter is "slightly big," gingival height "slightly different," angles "within the predicate device's range." (Page 18)
    SterilityNon-sterile (Predicate)Non-sterile (All subject devices)
    Indications for Use / Intended UseSimilar to predicate devicesALLONUS Tech Prosthetic is intended for use with dental implants as a support for single or multiple-unit prosthetic restorations in the maxilla or mandible of partially or fully edentulous patients, including cemented-retained, screw-retained, or overdenture restorations. Compatible with various specific implant systems. (Pages 3, 9, 10, 16, 17, 18, 19, 20, 21, 22)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Test Set Sample Size:
      • The document does not specify the exact sample sizes used for the non-clinical bench tests (fatigue, sterilization, biocompatibility, or dimensional analysis). It only states that tests were performed "for the subject device" and "of the worst-case scenario through fatigue testing."
    • Data Provenance:
      • The 510(k) submission is from ALLONUS Tech Co., LTD. in the REPUBLIC OF KOREA. This implies the testing was likely conducted in or overseen by this entity.
      • The studies were non-clinical bench tests, not clinical studies involving human patients. Therefore, terms like "retrospective" or "prospective" clinical dataProvenance are not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the 510(k) Summary. For non-clinical bench testing, "ground truth" is typically established by recognized international standards (e.g., ISO, ASTM) and engineering principles, rather than expert consensus on clinical cases.
    • The document mentions "Dimensional analysis and reverse engineering... were performed" and "assessment of maximum and minimum dimensions... along with cross-sectional images." This suggests engineering expertise, but specific numbers or qualifications of experts are not stated.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable as the studies were non-clinical bench tests. Adjudication methods like 2+1 (two readers plus one adjudicator) are used in clinical studies, particularly for diagnostic imaging, to resolve discrepancies in expert interpretations of patient data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study (MRMC) is relevant for diagnostic AI devices that assist human interpretation of medical images or data. The ALLONUS Tech Prosthetic is a physical medical device (dental implant abutment), not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No, a standalone (algorithm-only) performance study was not done. This question is also typically relevant for AI/software as a medical device (SaMD). The ALLONUS Tech Prosthetic is a physical device that integrates with human dental procedures.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical tests (fatigue, sterilization, biocompatibility), the "ground truth" is adherence to recognized international standards (ISO, ASTM) and established engineering specifications.
    • For implant-to-abutment compatibility, the "ground truth" was based on dimensional analysis and reverse engineering of OEM implant bodies, abutments, and screws, comparing the subject device's design to established OEM specifications.

    8. The sample size for the training set:

    • This is not applicable. "Training set" refers to data used to train machine learning algorithms. The ALLONUS Tech Prosthetic is a physical medical device, not an AI/ML algorithm.

    9. How the ground truth for the training set was established:

    • This is not applicable as there is no training set for a physical medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242686

    Validate with FDA (Live)

    Date Cleared
    2024-12-04

    (89 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NeoConvert Solution: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The NeoConvert Solution may be used with single-stage or two-stage procedures, for temporary multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Direct Screw to MUA: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The Direct Screw to MUA may be used with single-stage or two-stage procedures, for screw-retained multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. All digitally designed personalized copings to be used with the Direct to Multi Unit Abutment Screw are intended to be sent to Straumann for manufacture at a validated milling center.

    GM Attachment TiN 30°: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The GM Attachment TiN 30° may be used with single-stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    This premarket notification includes the following additions to the Neodent Implant System: NeoConvert Solution, Direct Screw to MUA, GM Attachment TiN 30°. The Neodent Multi-Unit Abutments proposed on this submission are similar to devices already cleared in previous submissions of Neodent Implant System, according to predicate devices described above. This submission intends to expand the portfolio of Neodent Abutments with these new solutions for multi-unit restorations, to provide more treatment options to customers. The NeoConvert Solution is a technique comprised by some abutments (not subject of this submission) used for converting removable temporary prosthesis into a screw-retained fixed temporary prosthesis. The Direct Screw to MUA is a prosthetic component designed to fix planned and digitally designed multi-unit milled restorations directly onto mini or micro abutments, eliminating the need for using the passive seating cylinder to restore chewing function. The GM Attachment TiN 30° is a prosthetic component used to stabilize implant-mucosa-supported removable full prosthesis (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandible. Although all these three new solutions are compatible with the same implants and components of Neodent Implant System and used for multiple unit rehabilitation, they are not used in conjunction with each other.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Neodent Implant System," focusing on the NeoConvert Solution, Direct Screw to MUA, and GM Attachment TiN 30°. This document details the device's technical characteristics, intended use, and its substantial equivalence to previously cleared predicate devices.

    However, the provided text does not contain information about acceptance criteria and a study proving a device meets those criteria in the context of an AI/ML medical device submission. The document is a traditional 510(k) submission for dental implants and abutments, which are physical medical devices, not AI/ML software. Therefore, the specific requirements for an AI/ML device study (such as sample sizes for test and training sets, expert qualifications for ground truth, MRMC studies, or standalone performance) are not applicable or present in this document.

    The "Performance Testing" section (pages 15-16) describes bench testing (dynamic fatigue, torsion, SEM images), MRI compatibility testing, biocompatibility testing, sterilization validation, and shelf-life testing for the physical dental implant components. These are standard performance tests for this type of physical device, not AI/ML algorithm validation.

    Therefore, I cannot provide a table of acceptance criteria and a study proving the device meets acceptance criteria as typically outlined for an AI/ML medical device, as this information is not present in the provided document. The document focuses on demonstrating substantial equivalence of a physical dental device to predicate devices through engineering and material testing.