K Number
K170588

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-08-08

(161 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

All digitally designed custom abutments for use with TiBase or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture.

Device Description

DESS Dental Smart Solutions is a dental implant abutment system that includes seven abutment design types (Healing, Temporary, Straight, TiBase, Pre-milled Blank, DESS LOC, Multi-Unit), and ten abutment connections compatible with eleven implant systems. Platform diameters range from 3.3 mm to 5.7 mm. Corresponding implant body diameters range from 3.25 mm to 6.0 mm. Abutments are offered in a variety of connection types to enable compatibility with a large number of currently marketed implants. Straight, Temporary and TiBase abutments have a SelectGrip® surface. DESS LOC Abutments have a ZrN coating. Selected DESS screws include DLC coating. DESS Dental Smart Solutions abutments are straight abutments. All abutments are provided non-sterile.

AI/ML Overview

This document describes the premarket notification for the DESS Dental Smart Solutions, an endosseous dental implant abutment. The information provided is for regulatory clearance based on substantial equivalence, not a standalone performance study of the device. Consequently, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/algorithm-based medical device are not applicable or find direct answers within this document.

However, I can extract the relevant information from this regulatory document concerning the device's performance evaluation to demonstrate its substantial equivalence to predicate devices, which serves as its "acceptance criteria" for market clearance.

Here's an analysis of the provided text in relation to your questions, focusing on the available "performance data":

1. A table of acceptance criteria and the reported device performance

The "acceptance criteria" for this device are based on demonstrating substantial equivalence to legally marketed predicate devices, rather than meeting specific performance metrics of an AI algorithm like sensitivity, specificity, or accuracy. The "performance" is shown through engineering and dimensional analysis, and material compatibility, to ensure it functions similarly to the predicate devices.

Acceptance Criteria (based on Substantial Equivalence)Reported Device Performance (as demonstrated by non-clinical testing)
Sterilization: Ensure sterilization of the final finished device.Sterilization validation to an SAL of 10^-6 according to ISO 17665-2.
Biocompatibility: Demonstrate devices are non-cytotoxic.Biological evaluation according to ISO 10993-1 and cytotoxicity testing according to ISO 10993-5 for all surfaces.
Surface Suitability: Demonstrate suitability of specialized surfaces (ZrN and SelectGrip).SEM evaluation and measurement of the ZrN and SelectGrip surfaces.
Compatibility with OEM Implant Systems: Confirm compatibility with various OEM implant bodies, abutments, and abutment fixation screws.Engineering and dimensional analysis of OEM implant bodies, OEM abutments, and OEM abutment fixation screws.
Design, Function, Material, Size, and Indications for Use: Be substantially equivalent to predicate devices.Detailed comparison tables demonstrating similarity in Indications for Use and Technological Characteristics (Prosthesis Attachment, Restoration, Abutment Platform Diameter, Abutment Angle, Abutment/Implant Interface, Material for Abutment and Screw).

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not explicitly stated in terms of a "test set" as one would define it for an AI/algorithm study. The "test set" here refers to the actual physical devices (parts, materials) subjected to engineering, dimensional, and biological evaluations. The document does not specify the number of individual abutments or components tested for each non-clinical evaluation.
  • Data Provenance: The document states that Terrats Medical SL is located in Montcada i Reixac, Spain. The testing was non-clinical. No information is provided regarding the country of origin of testing data beyond the manufacturer's location.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not Applicable: This type of expertise for establishing ground truth is relevant for AI/algorithm studies involving image interpretation or clinical diagnosis. For a dental implant abutment, "ground truth" is based on established engineering and material science standards (e.g., ISO standards, dimensional specifications, material composition), which do not typically involve human expert consensus in the same way as an AI diagnostic study.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not Applicable: Adjudication methods like 2+1 or 3+1 are used in studies where multiple human readers interpret data (e.g., medical images) and their discrepancies need to be resolved. This document pertains to the physical and material properties of a medical device, which are evaluated against defined engineering and biological standards, not subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable: No MRMC study was performed. This document concerns the regulatory clearance of a physical medical device (dental abutment), not an AI-assisted diagnostic tool. The document explicitly states: "No clinical data were included in this submission."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable: This is not an AI algorithm. The device "DESS Dental Smart Solutions" refers to a system of dental implant abutments, which are physical components. The document indicates that "All digitally designed custom abutments... are to be sent to a Terrats Medical validated milling center for manufacture," implying a digital design process, but the device itself is a physical product, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for this device's performance relies on established engineering standards, material specifications, and biological safety standards (e.g., ISO 17665-2 for sterilization, ISO 10993-1/5 for biocompatibility), and dimensional compatibility with OEM implant systems. It is a physical and material "ground truth" rather than a clinical or diagnostic "ground truth."

8. The sample size for the training set

  • Not Applicable: This device is a physical product, not an AI model that requires a "training set."

9. How the ground truth for the training set was established

  • Not Applicable: As there is no training set for an AI model, this question is not relevant.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/0/Picture/11 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 8, 2017

Terrats Medical SL % Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K170588

Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 6, 2017 Received: July 7, 2017

Dear Linda Schulz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mary S. Runner -S

for Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170588

Device Name

DESS Dental Smart Solutions

Indications for Use (Describe)

DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

All digitally designed custom abutments for use with TiBase or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture.

Implant System CompatibilityImplant Diameter (mm)Platform Diameter (mm)
3i Certain®3.25, 4.0, 5.03.4, 4.1, 5.0
3i OSSEOTITE®3.25, 3.75, 4.0, 5.03.4, 4.1, 5.0
OsseoSpeed™3.5, 4.0, 5.03.5/4.0, 4.5/5.0
FRIADENT XiVE3.4, 3.8, 4.53.4, 3.8, 4.5
NobelActive®3.5, 4.3, 5.0NP, RP
NobelReplace Conical3.5, 4.3, 5.0NP, RP
Nobel Replace Trilobe3.5, 4.3, 5.0NP, RP, WP
Brånemark3.5, 3.75/4.0, 5.0NP, RP, WP
Straumann® Bone Level3.3, 4.1, 4.8NC, RC
Straumann® Tissue Level3.3, 4.1, 4.8RN, WN
Tapered Screw-Vent®3.7, 4.1, 4.7, 6.03.5, 4.5, 5.7

Compatible Implant Systems

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

{3}------------------------------------------------

510(k) Summary Terrats Medical SL

DESS Dental Smart Solutions

K170588 August 8, 2017

ADMINISTRATIVE INFORMATION

Manufacturer NameTerrats Medical SL
Avenida La Ferrería (Pol Ind La Ferrería) 62
Montcada i Reixac, 08110
Spain
Telephone +34 93 564 60 06
Fax +34 93 564 73 17
Official ContactRoger Terrats, COO
Representative/ConsultantLinda Schulz, BSDH, RDH
Kevin Thomas, PhD
PaxMed International, LLC
12264 El Camino Real, Suite 400
San Diego, CA 92130
Telephone: +1-858-792-1235
Fax: +1-858-792-1236
Email: LSchulz@paxmed.com
KThomas@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameDESS Dental Smart Solutions
Common NameDental implant abutment
Classification NameEndosseous dental implant abutment
Classification Regulations21 CFR 872.3630, Class II
Product CodeNHA
Classification PanelDental Products Panel
Reviewing BranchDental Devices Branch

PREDICATE DEVICE INFORMATION

Primary Predicate K120414, OsseoSpeed™ Plus, Astra Tech AB

Reference Predicates K072878, Modification to: Locator Implant Anchor, Zest Anchors, Inc. K092341, Low Profile Abutment, Biomet 3i K 150203, Medentika CAD/CAM Abutments, Medentika GmbH K150367, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários SA

{4}------------------------------------------------

K0633413i OSSEOTITE Certain® Dental ImplantsImplant Innovations, Inc.
K063286OSSEOTITE® Dental ImplantsImplant Innovations, Inc.
K101732OsseoSpeed™Astra Tech AB
K073075FRIADENT Implant SystemsDENTSPLY International, Inc.
K142260NobelActive®Nobel Biocare AB
K073142NobelReplace Hexagonal ImplantsNobel Biocare AB
K050705TiUnite® ImplantsNobel Biocare AB
K050406NOBELSPEEDY™ ImplantsNobel Biocare USA LLC
K022562Various Brånemark System Implants -Immediate Function IndicationNobel Biocare AB
K140878Straumann® Bone Level Tapered ImplantsStraumann USA, LLC
K062129P.004 ImplantsInstitut Straumann AG
K130222Straumann® Dental Implant System SLActive andRoxolid Product FamiliesStraumann USA, LLC
K112160Tapered Screw-Vent® X ImplantZimmer Dental, Incorporated

Compatible Implant System Predicates

INDICATIONS FOR USE

DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

All digitally designed custom abutments for use with TiBase or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture.

Implant System CompatibilityImplant Diameter (mm)Platform Diameter (mm)
3i Certain®3.25, 4.0, 5.03.4, 4.1, 5.0
3i OSSEOTITE®3.25, 3.75, 4.0, 5.03.4, 4.1, 5.0
OsseoSpeed™3.5, 4.0, 5.03.5/4.0, 4.5/5.0
FRIADENT XiVE3.4, 3.8, 4.53.4, 3.8, 4.5
NobelActive®3.5, 4.3, 5.0NP, RP
NobelReplace Conical3.5, 4.3, 5.0NP, RP
Nobel Replace Trilobe3.5, 4.3, 5.0NP, RP, WP
Brånemark3.5, 3.75/4.0, 5.0NP, RP, WP
Straumann® Bone Level3.3, 4.1, 4.8NC, RC
Straumann® Tissue Level3.3, 4.1, 4.8RN, WN
Tapered Screw-Vent®3.7, 4.1, 4.7, 6.03.5, 4.5, 5.7

Compatible Implant Systems

DEVICE DESCRIPTION

DESS Dental Smart Solutions is a dental implant abutment system that includes seven abutment design types (Healing, Temporary, Straight, TiBase, Pre-milled Blank, DESS LOC, Multi-Unit), and ten abutment connections compatible with eleven implant systems. Platform diameters range from 3.3 mm to 5.7 mm. Corresponding implant body diameters range from 3.25 mm to 6.0 mm. The following table outlines the body and platform diameters by abutment design and corresponding implant line.

{5}------------------------------------------------

Summary of Abutment DesignsDESS Abutment SystemOEM Implant System3i Certain®3.4 (NP), 4.1 (RP), 5.0 (WP)3i OSSEOTITE®3.4 (NP), 4.1 (RP), 5.0 (WP)OsseoSpeedTM3.5/4.0 (RP), 4.5/5.0 (WP)FRIADENT XIVE®3.4 (NP), 3.8 (RP), 4.5 (WP)NobelActive®NP (3.5), RP (3.9)NobelReplace® ConicalNP (3.5), RP (3.9)NobelReplace® TrilobeNP (3.5), RP (4.3), WP (5.0)Brånemark System®NP (3.5), RP (4.1), WP (5.1)Straumann® Bone LevelNC (3.3), RC (4.1)Straumann® Tissue LevelRN (4.8), WN (6.5)Tapered Screw-Vent3.4 (NP), 4.5 (RP), 5.7 (WP)
Multi-UnitNPNPRPNPNPNPNPNPNCRNNP
DESS LOCRPRPWPRPRPRPRPRPRCWNRP
Pre-milled BlankEngagingNPNPRPNPNPNPNPNPNCRNNP
Ti BaseEngagingRPRPWPRPRPRPRPRPRCWNRP
Ti BaseNon-EngagingNPNPRPNPNPNPNPNPNCRNNP
StraightRPRPWPRPRPRPRPRPRCWNRP
TemporaryEngagingNPNPRPNPNPNPNPNPNCRNNP
TemporaryNon-EngagingRPRPWPRPRPRPRPRPRCWNRP
HealingNP 3.4RP 4.1WP 5.0NP 3.4RP 4.1WP 5.0RP 3.5/4.0WP 4.5/5.0NP 3.4RP 3.8WP 4.5NP 3.5RP 3.9NP 3.5RP 3.9NP 3.5RP 4.3WP 5.0NP 3.5RP 4.1WP 5.1NC 3.3RC 4.1RN 4.8WN 6.5NP 3.4RP 4.5WP 5.7
Internal Hex “Click”
External Hex USA
Internal Hex Conic
Internal Hex FD
Active Hex
Tri-Lobe
External Hex Universal
Conical BL
Octagon
Internal Hex USA

{6}------------------------------------------------

Abutments are offered in a variety of connection types to enable compatibility with a large number of currently marketed implants. Straight, Temporary and TiBase abutments have a SelectGrip® surface. DESS LOC Abutments have a ZrN coating. Selected DESS screws include DLC coating. DESS Dental Smart Solutions abutments are straight abutments. All abutments are provided non-sterile.

PERFORMANCE DATA

Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation to an SAL of 10° according to ISO 17665-2 to ensure sterilization of the final finished device; biological evaluation according to ISO 10993-1 and cytotoxicity testing according to ISO 10993-5 for all surfaces to demonstrate that all devices are non-cytotoxic; SEM evaluation and measurement of the ZrN and SelectGrip surfaces to demonstrate suitability of the surface; and engineering and dimensional analysis of OEM implant bodies, OEM abutments, and OEM abutment fixation screws to confirm compatibility.

No clinical data were included in this submission.

EQUIVALENCE TO MARKETED DEVICE

The subject device is substantially equivalent in indications and design principles to the predicate devices shown above. Below are summary tables comparing the Indications for Use and the technological characteristics of the subject device and the predicate devices.

Subject DeviceIndications for Use Statement
DESSDental SmartSolutionsTerrats Medical SLDESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in themaxillary or mandibular arch to provide support for prosthetic restorations.All digitally designed custom abutments for use with TiBase or Pre-milled Blank are to be sent to a Terrats Medicalvalidated milling center for manufacture.
Compatible Implant Systems
Implant System CompatibilityImplant Diameter (mm)Platform Diameter (mm)
3i Certain®3.25, 4.0, 5.03.4, 4.1, 5.0
3i OSSEOTITE®3.25, 3.75, 4.0, 5.03.4, 4.1, 5.0
OsseoSpeed™3.5, 4.0, 5.03.5/4.0, 4.5/5.0
FRIADENT XiVE3.4, 3.8, 4.53.4, 3.8, 4.5
NobelActive®3.5, 4.3, 5.0NP, RP
NobelReplace Conical3.5, 4.3, 5.0NP, RP
Nobel Replace Trilobe3.5, 4.3, 5.0NP, RP, WP
Brånemark3.5, 3.75/4.0, 5.0NP, RP, WP
Straumann® Bone Level3.3, 4.1, 4.8NC, RC
Straumann® Tissue Level3.3, 4.1, 4.8RN, WN
Tapered Screw-Vent®3.7, 4.1, 4.7, 6.03.5, 4.5, 5.7
Primary PredicateDeviceIndications for Use Statement
K120414OsseoSpeed™ PlusAstra Tech ABImplants:The Astra Tech Dental Implants are intended for both one- and two-stage surgical procedures in the following situations and with the following clinical protocols:• replacing single and multiple missing teeth in the mandible and maxilla,• immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge,• especially indicated for use in soft bone applications where implants with other implant surface treatments may be less effective,• immediate loading in all indications, except in single tooth situations on implants shorter than 8 mm, or in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate.The intended use for OsseoSpeed™ Plus 3.0S is limited to replacement of maxillary lateral incisors and mandibular incisors.Abutments:Astra Tech Implant System Plus abutments are intended to be used in conjunction with Astra Tech Implant System Plus in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.Atlantis Abutments:The Atlantis™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous; patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.The Atlantis™ Crown Abutment in Zirconia is intended for use with an endosseous; implant to function as a substructure that also serves as the final restoration, in partially or completely edentulous; patients. The prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous implant.
ReferencePredicate Devices
K072878Modification to:Locator ImplantAnchorZest Anchors, Inc.Not available
K092341Low ProfileAbutmentBiomet 3i, Inc.Biomet 3i Low Profile Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is screw retained to the abutment.
K150203MedentikaCAD/CAMAbutmentsMedentika GmbHMedentika Preface CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Implant System CompatibilitySeriesImplant Diameter (mm)Platform Diameter (mm)
Nobel Biocare Replace™ SelectE3.5, 4.3, 5.0, 6.03.5, 4.3, 5.0, 6.0
Nobel Biocare NobelActive™F3.0, 3.5, 4.3, 5.03.0, 3.5, 3.9(4.3), 3.9(5.0)
Biomet 3i Osseotite® Certain®H3.25, 4.0, 5.03.4, 4.1, 5.0
Biomet 3i Osseotite®I3.25, 3.75, 4.0, 5.03.4, 4.1, 5.0
Nobel Biocare BrånemarkK3.3, 3.75, 4.0, 5.03.5, 4.1, 4.1, 5.1
Straumann Bone LevelL3.3, 4.1, 4.83.3, 4.1, 4.8
Straumann StandardN3.3, 4.1, 4.83.5(NNC), 4.8, 6.5
Zimmer Tapered Screw-vent®R3.3, 3.7, 4.1, 4.7, 6.03.5, 4.5, 5.7
Astra Tech OsseoSpeed™S3.0, 3.5, 4.0, 4.5, 5.03.0, 3.5, 4.0, 4.5, 5.0
Dentsply Friadent® Frialit/XiVE®T3.4, 3.8, 4.5, 5.53.4, 3.8, 4.5, 5.5
Medentika PreFace is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center.
K150367Neodent ImplantSystemJJGC Indústria eComércio deMateriais DentáriosSATitanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.PreFace Abutment is a titanium abutment to be used in fabricating a full custom abutment and placed onto Neodent dental implants to provide support for customized prosthetic restorations. PreFace Abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations.

Comparison of Indications for Use Statements

{7}------------------------------------------------

{8}------------------------------------------------

Subject DevicePrimaryPredicateReference Predicates
K120414K072878K092341K150203K150367
DESSDental SmartSolutionsOsseoSpeed™ PlusModification to:Locator ImplantAnchorLow ProfileAbutmentMedentikaCAD/CAMAbutmentsNeodent ImplantSystem
Terrats MedicalSLAstra Tech ABZest Anchors, Inc.Biomet 3i, Inc.MedentikaGmbHJJGC Indústria eComércio de MateriaisDentários SA
Design
Prosthesis AttachmentCement-retainedScrew-retainedCement-retainedScrew-retainedOverdentureattachmentScrew-retainedCement-retainedCement-retainedScrew-retained
RestorationSingle-unitMulti-unitSingle-unitMulti-unitOverdentureSingle-unitMulti-unitSingle-unitMulti-unitSingle-unitMulti-unit
Abutment PlatformDiameter (mm)3.4 - 5.73.0 - 5.42.5 - 6.53.4 - 5.03.0 - 7.03.0 - 6.0
Abutment AngleStraightStraight to 30°StraightStraight to 30°Straight to 30°Straight to 30°
Abutment/ImplantInterfaceInternal,ExternalInternalInternal, ExternalInternal,ExternalInternal,ExternalInternal
Material
AbutmentTi-6Al-4VTi-6Al-4VZirconia, Gold,PEEKTi-6Al-4VTi-6Al-4VTi-6Al-4VTi-6Al-4V
ScrewTi-6Al-4VTi-6Al-4VTi-6Al-4VTi-6Al-4VTi-6Al-4VTi-6Al-4V

Comparison of Technological Characteristics

DESS Dental Smart Solutions abutments are substantially equivalent in design, function, material, size, and Indications for Use to OsseoSpeed Plus (K120414) and Medentika CAD/CAM Abutments (K150203). All are intended for use with endosseous dental implants in the maxilla and mandible to provide prosthetic support. Digital files for DESS Dental Smart Solutions abutments and for Medentika CAD/CAM Abutments are to be sent to a validated milling center for manufacture. Differences in the type of restoration named or specific milling center stated in the Indications for Use statement do not affect the intended use.

Subject device abutment designs and function are substantially equivalent to design and function of abutments included in K120414, K072878, K092341, K150203 and K150367. Implant/abutment interface compatbility for the subject device is substantially equivalent to the compatible implant system predicates listed above.

CONCLUSION

The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject device devices encompass the same range of physical dimensions, including diameter and design of the abutments. The subject and predicate devices are packaged in similar materials and are to be sterilized using similar methods.

The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)