Search Results
Found 8 results
510(k) Data Aggregation
(90 days)
DESS® dental implants are indicated for surgical placement in the upper or lower jaw in edentulous or partially edentulous patients for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. They are designed to support single or multi-unit restorations in splinted or non-splinted applications, as well as to support overdenture attachment systems. DESS® dental implants may be used for immediate or early implantation following extraction or loss of natural teeth, and may be used for immediate or delayed loading techniques. Implants may be loaded immediately when good primary stability is achieved and occlusal loading is appropriate.
DESS® NEO GM Dental Implants are compatible with DESS® Dental Smart Solutions abutments having the identical NEO GM connection manufactured by Terrats Medical SL.
DESS® NEO GM Dental Implants with a diameter of 3.5 mm are indicated for use in reduced interdental spaces, where there is not enough alveolar bone for a larger diameter implant. The use of 3.5 mm implants is intended only for rehabilitation of the anterior region of the mouth.
DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
This submission includes dental implants that are compatible with DESS® Dental Smart Solutions abutments having the identical NEO GM connection manufactured by the sponsor of this submission, Terrats Medical SL, and cleared under K212628, K222288, K233316, K240208, and K242340. No claims of compatibility between the subject device implants and abutments from any OEM other than DESS® Dental Smart Solutions will be made. This submission also includes DESSLoc abutments that are compatible with eight (8) dental implant lines from five (5) OEM manufacturers. Also included in this submission is one (1) Pre-Milled Blank abutment that is compatible with three (3) implant lines manufactured by Alpha Dent Implants GmbH.
This submission includes one implant line, the Dental Implant NEO GM, a series of self-tapping, threaded, root-form dental implants to be placed at bone level. The subject device implants are provided in body diameters of 3.55 mm, 3.75 mm, 4.0 mm, 4.3 mm, 5.0 mm, 6.0 mm, and 7.0 mm. The subject device implant body diameters will be labeled as 3.5 mm, 3.75 mm, 4.0 mm, 4.3 mm, 5.0 mm, 6.0 mm, and 7.0 mm. Implant with body diameters ranging from 3.55 mm to 5.0 mm are provided in overall lengths of 7.9 mm, 9.9 mm, 11.4 mm, 12.9 mm, 15.9 mm, and 17.9 mm. Implants with body diameters 6.0 mm and 7.0 mm are provided in overall lengths of 7.9 mm, 9.9 mm, and 11.4 mm. The subject device implant lengths will be labeled as 8 mm, 10 mm, 11.5 mm, 13 mm, 16 mm, and 18 mm. All subject device implants, regardless of body diameter, have an internal Morse taper connection with a 16° included angle and 2.99 mm diameter opening at the top of the implant. This NEO GM connection is identical to the connection for abutments cleared previously in K242340.
All subject device implants are made of unalloyed titanium conforming ASTM F67 and ISO 5832-2. The entire endosseous surface, except for a small coronal bevel, features a grit blasted and double acid etched (SLA) surface, which is identical to the surface treatment for DESS® implants that were cleared in K212538.
This submission also includes DESSLoc Abutments designed for overdenture retention. The subject device DESSLoc Abutments are straight, non-engaging abutments that attach directly to the implant and are compatible with eight (8) dental implant lines from five (5) OEM manufacturers.
The subject device DESSLoc Abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136. The subject device DESSLoc abutments have a zirconium nitride (ZrN) coating produced by a physical vapor deposition (PVD) process. The ZrN coating is applied to increase the surface hardness and reduce wear of the abutment surface. The ZrN coating for the subject device DESSLoc Abutments is identical to the ZrN coating applied to DESSLoc Abutments cleared in K242340, K240208, K222288, K191986, and K170588.
This submission also includes one (1) Pre-Milled Blank Abutment that is compatible with three (3) implant lines manufactured by Alpha Dent Implants GmbH, including Implant Active Konus, Implant Classic Konus, and Implant Active Bio, cleared in cleared in K210499. Reverse engineering compatibility analysis of the Alpha Dent implants, abutments, and abutment screws and Terrats Medical SL abutments and abutment screws was provided in the prior Terrats Medical SL submission K243212.
The Pre-Milled Blank Abutment has a maximum (before milling) diameter of 10 mm and a solid cylindrical design and an engaging implant connection. The Pre-Milled Blank Abutment is manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136. All patient-specific custom abutment fabrication for the Pre-Milled Blank Abutment is by prescription on the order of the clinician and will be done at a Terrats Medical validated milling center under FDA quality system regulations. The design parameters for the CAD-CAM fabrication of a custom abutment from the Pre-Milled Blank Abutment are:
- Minimum wall thickness – 0.45 mm
- Minimum post height for single-unit restoration – 4.0 mm (post height measured above the gingival height of the final patient-matched design)
- Minimum gingival height – 0.5 mm
- Maximum gingival height – 6.0 mm
- Pre-Milled Blanks are for straight abutments only
The abutment screw for use with the Pre-Milled Blank Abutment and the Alpha Dent implants was cleared previously in K243212.
All subject device implants are provided sterile, and all subject device abutments are provided non-sterile. The Pre-Milled Blank Abutment is supplied with the previously-cleared non-sterile abutment screw for attachment to the corresponding compatible implant. All subject device components are provided in single-unit packages for single-patient, single-use only.
The provided FDA 510(k) Clearance Letter for DESS® Dental Implants (K251280) does not contain information about acceptance criteria or a specific study proving the device meets those criteria in the context of AI/ML device performance.
The document is a clearance letter for a dental implant system, which is a physical medical device, not an AI/ML-driven diagnostic or treatment planning software. Therefore, the questions related to AI/ML device performance (such as sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth) are not applicable to the content provided.
The document primarily focuses on demonstrating substantial equivalence to predicate devices based on:
- Intended Use and Indications for Use
- Technological Characteristics (design, materials, surface treatment, sterilization methods, dimensions)
- Performance Data (non-clinical data such as MRI review, sterilization validation, bacterial endotoxin testing, shelf life, biocompatibility, and mechanical testing according to ISO 14801).
Since the request specifically asks for AI/ML-related performance evaluation, and the provided document is not for an AI/ML device, I cannot fulfill the request as stated with the given input.
However, I can extract the non-clinical performance data and the comparison to predicate devices, which serve as the "study" and "acceptance criteria" for this type of physical medical device in the context of a 510(k) submission.
Non-Clinical Performance Data (Serving as "Study" for Substantial Equivalence):
-
Worst-Case MRI Review:
- Method: Scientific rationale and published literature (T.O. Woods, J.G. Delfino, and S. Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Evaluation Volume 49, No. 2 (March/April 2021): 783–795).
- Scope: Evaluated all compatible implant bodies, abutments, and fixation screws and their material composition.
- Parameters: Magnetically induced displacement force and torque, per FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment."
- Outcome: Rationale addressed parameters; device components found suitable for the MR environment.
-
Gamma Irradiation Sterilization Validation (for implants):
- Method: Selecting and substantiating a 25 kGy dose using method VDmax25, according to ISO 11137-1 and ISO 11137-2.
- Acceptance Criteria: Sterility Assurance Level (SAL) of $10^{-6}$.
- Outcome: Analysis showed the subject device implants met the SAL and did not create a new worst case compared to predicate device K212538.
-
Bacterial Endotoxin Testing:
- Method: Limulus amebocyte lysate (LAL) testing according to ANSI/AAMI ST72.
- Acceptance Criteria: Limit of
Ask a specific question about this device
(80 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 and K240982 to allow additional options of zirconia material, scanners, CAM software, and milling machines to the digital dentistry workflow. The subject devices are to be sent to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. There are no changes to the abutment design, implant compatibilities, or design parameters. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows (also referred to as point of care) under K221301 and K240982.
The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.
The provided 510(k) summary for DESS Dental Smart Solutions focuses on demonstrating substantial equivalence to predicate devices for dental implant abutments. It primarily addresses the expansion of compatible materials, scanners, CAM software, and milling machines within an existing digital dentistry workflow. The document does not describe an AI/ML-based device that would typically have acceptance criteria related to diagnostic performance.
Therefore, many of the requested items related to AI/ML device performance (like acceptance criteria for diagnostic metrics, sample size for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this submission.
The acceptance criteria and supporting "study" (non-clinical data) for this device are related to its mechanical performance, biocompatibility, and integration within the digital workflow, demonstrating that the expanded components maintain the safety and effectiveness of the previously cleared predicate devices.
Here's a breakdown based on the information provided and the non-applicability of AI/ML-specific questions:
1. A table of acceptance criteria and the reported device performance
Since this is not an AI/ML diagnostic device, the acceptance criteria are not in terms of traditional diagnostic metrics (sensitivity, specificity, AUC). Instead, they are related to material properties, mechanical integrity, and the digital workflow's accuracy.
Acceptance Criteria Category | Reported Device Performance (Summary from Submission) |
---|---|
Mechanical Integrity / Compatibility | - Fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions. (Implies successful completion to similar or better standards than predicate.) |
Biocompatibility | - Biocompatibility testing according to ISO 10993-5 and ISO 10993-12. (Implies successful demonstration of biocompatibility.) |
Sterilization | - Sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937. (Implies successful validation for sterilization.) |
CAD Design Restrictions | - Software verification included testing of restrictions that prevent design of components outside of the stated design parameters. |
- Abutment design library validated to demonstrate established design limitations are locked and cannot be modified by the user. (Implies successful implementation and verification of design constraints.) |
| CAM Restriction Zones / Manufacturing Accuracy | - Validation testing of CAM restriction zones conducted, including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in CAM software. (Implies successful validation to ensure manufacturing precision and prevent damage.) |
| Material Conformance | - Zirconia materials conform to ISO 6872. - Titanium alloy conforms to ASTM F136.
- Co-Cr-Mo alloy conforms to ASTM F15337. (Implies materials meet standards.) |
| Physical Dimensions | - Device encompasses the same range of physical dimensions as the predicate device. (Implies dimensional equivalence.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated in terms of a "test set" for diagnostic performance. The validation involved physical testing of components (e.g., fatigue testing) and software verification. The specific number of abutments or digital design instances used for these non-clinical tests is not detailed in this summary.
- Data Provenance: Not applicable in the context of patient data for an AI/ML device. The "data" here refers to engineering and material testing results, likely conducted in controlled lab environments (implied to be in accordance with international standards like ISO and ASTM). The manufacturer is Terrats Medical SL, in Spain, so testing would likely originate from their facilities or contracted labs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device relates to engineering specifications and material science.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a diagnostic device involving expert review adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For this device, the "ground truth" is based on:
- Engineering Specifications: Defined design parameters (e.g., minimum wall thickness, post height, angulation limits).
- Material Standards: Conformance to international standards such as ASTM F136, ISO 6872.
- Benchmarking/Predicate Equivalence: Performance is assessed against established performance of the predicate devices and OEM implant systems.
- Software Validation Logic: Verification that software correctly enforces design rules and CAD/CAM restrictions.
8. The sample size for the training set
Not applicable. This device does not involve a machine learning training set.
9. How the ground truth for the training set was established
Not applicable. This device does not involve a machine learning training set.
Ask a specific question about this device
(138 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 to add the ability for the subject device Base Abutments and Pre-milled (Blank) Abutments to be designed using AbutmentCAD software in the digital dentistry workflow, as well as add angulation to some of the Pre-Milled (Blank) Abutments. The subject devices are to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. The proposed change is to allow the subject devices to be designed using AbutmentCAD by exocad GmbH, the current workflow allows only the use of 3Shape by 3Shape A/S for the design software. Another purpose of this submission is to expand the design parameters to allow angulation (up to 30°) on Pre-milled (Blank) Abutments that are compatible with Neodent Grand Morse, Nobel Active/Nobel Parallel Conical 3.0 mm, and Straumann BLX implants. There are no changes to the abutment design or implant compatibilities. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows in under K221301.
The subject device DESS Dental Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, titanium and ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The digital workflow includes the following products (not subject devices of this submission):
- Ceramic material: VITA YZ ST and VITA YZ XT (K180703)
- . Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436)
- . Intraoral Scanner: 3Shape TRIOS A/S Series Intraoral Scanner (510(k) exempt under 21 CFR 872.3661)
- Desktop scanner: 3Shape D900 Dental Lab Scanner (510(k) exempt under 21 CFR 872.3661)
- Abutment design software: 3Shape Abutment Designer Software (K151455) and AbutmentCAD ● (K193352)
- . Milling machine: VHF R5 by vhf camfacture AG with DentalCAM and DentalCNC 7 software
The provided text describes a 510(k) premarket notification for DESS Dental Smart Solutions, which are dental implant abutments. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device's effectiveness through clinical performance studies with specific statistical acceptance criteria for accuracy, sensitivity, or specificity.
Therefore, the document does not contain the information requested regarding:
- A table of acceptance criteria and reported device performance (in terms of clinical metrics like accuracy, sensitivity, specificity).
- Sample size used for the test set or its provenance.
- Number of experts used to establish ground truth or their qualifications.
- Adjudication method for the test set.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study or its effect size.
- Standalone (algorithm only) performance.
- Type of ground truth used (expert consensus, pathology, outcomes data).
- Sample size for the training set.
- How the ground truth for the training set was established.
The study described in this document focuses on non-clinical performance data to demonstrate substantial equivalence, specifically:
- Sterilization validation: According to ISO 17665-1, ISO 17665-2, and ISO 14937.
- Biocompatibility testing: According to ISO 10993-5 and ISO 10993-12.
- Fatigue testing and reverse engineering analysis: Of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility. This includes fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions.
- MR Conditional labeling.
- Validation testing of CAM restriction zones: Including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in the CAM software.
- Software verification: Included testing of restrictions that prevent design of components outside of the stated design parameters. The abutment design library was validated to demonstrate that established design limitations and specifications are locked and cannot be modified by the user.
The acceptance criteria and reported "performance" for this submission are based on these engineering and design validations, ensuring the device meets safety and performance standards equivalent to the predicate device, K221301. The key "performance" metrics are about maintaining physical and material integrity and compatibility.
The core of the submission is to expand the DESS Dental Smart Solutions abutment system to:
- Allow design using AbutmentCAD software (in addition to 3Shape software).
- Add angulation (up to 30°) to some Pre-milled (Blank) Abutments for specific implant systems.
The document explicitly states: "No clinical data were included in this submission." and "The subject device, the predicate device, and reference devices have the same intended use, technological characteristics, and are materials. The subject device, the predicate device, and reference devices encompass the same range of physical dimensions, manufactured by similar methods, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above."
Therefore, this FDA submission is for a physical medical device (dental implant abutment) and its manufacturing/design software modifications, not an AI or diagnostic device that would involve clinical performance metrics like sensitivity or specificity.
Ask a specific question about this device
(139 days)
CEREC Cercon 4D™ Abutment System is intended for use in partially or fully edentulous mandibles and maxillae in support of single cement-retained restorations.
The system comprises three parts:
- CEREC Cercon 4D™ Abutment Block
- TiBase
- CAD/CAM system
The CEREC Cercon 4D™ ceramic structure cemented to the TiBase is recommended for two-piece hybrid abutments for single tooth restorations and hybrid abutment crowns, used in conjunction with endosseous dental implants.
The CEREC Cercon 4D Abutment Blocks, which are used for fabrication of a ceramic structure, two-piece hybrid abutments (meso-structure and crown) and abutment crowns, that are cemented to a TiBase (titanium base) used with dental implant systems. The CEREC Cercon 4D Abutment Blocks are not provided as the finished, fully assembled dental implant medical devices. The abutment blocks are materials supplied to dental professionals that must be further processed/manufactured using CAD/CAM technology and they are not intended to be reused as in the context of direct patient-applied devices and materials.
CEREC Cercon 4D™ Abutment Block are Yttria-doped zirconia blocks suitable for chairside and lab side use in fabrication of single cement-retained restorations. CEREC Ceron 4D™ Abutment Block are designed with a pre-drilled screw access channel and anti-rotation feature. The design allows for fabrication of a ceramic structure, two-piece hybrid abutments (mesostructure and crown) and abutment crowns, that are cemented to theBase (Titanium base) used with dental implant systems.
The provided document describes the substantial equivalence of the CEREC Cercon 4D™ Abutment Blocks and System, primarily focusing on non-clinical performance and material characteristics, rather than an AI/ML-based device. Therefore, many of the requested elements pertaining to AI/ML device studies (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone performance, training set details) are not applicable or cannot be extracted from this document.
However, I can extract information related to the acceptance criteria and study that proves the device meets those criteria from the perspective of a medical device (specifically, a dental abutment system), even without AI elements.
Here's the information based on the provided text, with Not Applicable (N/A) for fields that relate to AI/ML studies and are not covered in this document.
Acceptance Criteria and Device Performance for CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Abutment System
The device under review is primarily a dental abutment system, and its performance is evaluated based on material properties, mechanical strength, and software integration, not on diagnostic accuracy or AI assistance.
1. Table of Acceptance Criteria and the Reported Device Performance
Test Performed | Test Method/Applicable Standards | Acceptance Criteria | Reported Performance (Results) |
---|---|---|---|
Flexural Strength | ISO 6872:2015 Amd 1. 2018 Dentistry-Ceramic Materials | >1,100 MPa | Pass |
Fatigue Testing | ISO 14801:2016 Dentistry-Implants-Dynamic loading test for endosseous dental implants | (Implied: Meets requirements) | Pass |
Sterilization Validation | ISO 17665-1 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices | Achieve a Sterility Assurance Level (SAL) of 10⁻⁶ | Validated |
Biocompatibility | ISO 10993 standard series (specifically ISO 10993-5, -10, -23) | Meets ISO 10993 requirements | Meets requirements |
Software Validation (Angulation) | Internal software integration requirements for the addition of the proposed device | Max angulation of 20° (User cannot proceed if outside) | Meets requirements |
Software Validation (Wall Thickness) | Internal software integration requirements for the addition of the proposed device | Minimal wall thickness of 0.5 mm (User cannot proceed if outside) | Meets requirements |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set:
- For Flexural Strength (Table 8.1): Not explicitly stated, but typically involves a certain number of samples to ensure statistical significance as per ISO 6872.
- For Fatigue Testing (Table 8.2): "New fatigue testing was conducted on the worst-case combinations relating to the greatest angulation, the platform size and the gingival height for the proposed Dentsply Sirona TiBase/Dentsply Sirona Implant Systems and Third Party TiBase/Third Party Implant Systems (Camlog) combinations." The exact number of samples per test condition is not specified in the document, but standardized tests like ISO 14801 would stipulate a minimum.
- For Sterilization Validation, Biocompatibility, and Software Validation: Not explicitly specified in terms of sample count in this summary.
- Data Provenance: The document does not specify the country of origin of the data. The tests are described as "non-clinical tests" and "performance bench testing," indicating laboratory-based studies. The document does not mention if the data is retrospective or prospective, as this distinction is more relevant for clinical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is not an AI/ML diagnostic tool requiring expert ground truth for image interpretation or similar. The "ground truth" (or more accurately, established performance standards) for this device is based on mechanical properties and ISO standards, which are objective and do not require expert human interpretation in the way an AI diagnostic system would.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no human expert interpretation or consensus review is involved in the performance testing of this device (e.g., physical strength, material composition).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic device; therefore, MRMC studies are irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm-based device. Its "system" aspect refers to the combination of the abutment block, TiBase, and CAD/CAM system for fabrication, not an AI algorithm. The performance described is of the physical components and the software's ability to constrain design parameters.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is established by international consensus standards (e.g., ISO 6872, ISO 14801, ISO 10993, ISO 17665-1) for dental materials and implants, along with internal software integration requirements. These are objective, quantitative measures rather than subjective human interpretations or clinical outcomes data in the context of diagnostic accuracy.
8. The sample size for the training set
Not applicable. This device does not have a "training set" in the context of machine learning.
9. How the ground truth for the training set was established
Not applicable. This device does not have a "training set" in the context of machine learning.
Ask a specific question about this device
(90 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes dental implants, abutments, and prosthetic components cleared previously in various submissions. The previously cleared abutments and prosthetic components are compatible with a variety of original equipment manufacturer (OEM) dental implants as well as DESS Dental Smart Solutions dental implants.
This submission adds various abutments to the DESS and OEM implant lines as summarized on the following pages in Table 1 Summary of Subject Device Abutment Designs, and Table 2 Summary of Subject Device Abutment Sizes.
The subject device abutment designs include Cover Screws, Healing Abutments, Temporary Abutments, Ti Base Abutments, AURUM Base Abutments (Ti Base abutments with a screw channel design that allows easier instrument access to the abutment screw), CoCr Base Abutments, Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), DESSLoc Abutments (Locator-type abutments), and abutment screws.
This document is a 510(k) summary for the DESS Dental Smart Solutions, a line of endosseous dental implant abutments. It details the device's characteristics, intended use, and a comparison to predicate devices to establish substantial equivalence.
Based on the provided text, the device itself is a physical medical device (dental implant abutments and associated components), not a software or AI-driven system. Therefore, the questions related to AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, and ground truth establishment for AI would not be directly applicable to this product as described.
The document focuses on establishing substantial equivalence to previously cleared predicate devices through:
- Identical Intended Use: The device is intended for "functional and esthetic rehabilitation of the edentulous mandible or maxilla" by providing support for prosthetic restorations, which is the same as the predicate devices.
- Similar Technological Characteristics: The device utilizes similar designs, materials (Ti-6Al-4V, Co-Cr-Mo alloy), manufacturing processes, and sterilization methods as its predicates.
- Performance Data: Non-clinical testing (mechanical testing per ISO 14801, MR environment assessment) and comparison to existing data from predicate devices are used to demonstrate safety and effectiveness.
Therefore, many of the requested elements for describing AI acceptance criteria and studies are not present or applicable in this document.
However, I can extract information relevant to the device's performance assessment and criteria for its type of submission.
Here's an attempt to answer the questions based solely on the provided text, acknowledging that the nature of the device (a physical implant component) means many AI-specific questions will be answered as "Not Applicable" (N/A):
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantifiable "acceptance criteria" in a table format with corresponding "reported device performance" metrics for this specific 510(k) submission. Instead, it relies on demonstrating substantial equivalence to existing predicate devices.
The underlying "acceptance criteria" for demonstrating substantial equivalence for this device type are primarily through:
- Mechanical Testing (ISO 14801): This is a critical performance standard for dental implants and abutments. The document states that mechanical testing was "conducted according to ISO 14801 to support the performance." The acceptance criteria would be successful completion of these tests, demonstrating the device's mechanical strength and fatigue resistance are comparable to or better than predicate devices. The reported performance is simply that the tests supported the performance.
- Material Conformance: Materials must conform to specific ASTM standards (e.g., ASTM F136 for Ti-6Al-4V, ASTM F1537 for Co-Cr-Mo). The reported performance is that the materials conform to these standards.
- Biocompatibility: While not detailed in this excerpt, the mention of "biocompatibility" in relation to predicates implies conformance to relevant biocompatibility standards (e.g., ISO 10993 series). The reported performance is that it is compatible.
- Sterilization Validation: Demonstrated sterility assurance level (SAL) of 10⁻⁶ via validated methods (moist heat or gamma irradiation). The reported performance is that validation was performed and met this SAL.
- Dimensional Compatibility: The abutments must fit the corresponding OEM implants correctly. The reported performance is that reverse engineering dimensional analysis confirmed compatibility.
Due to the nature of the document being a 510(k) summary focusing on substantial equivalence rather than a full study report, specific numerical performance results for the device tests are not provided in this text.
2. Sample sizes used for the test set and the data provenance
- Sample Size for Mechanical Testing: The document states "mechanical testing conducted according to ISO 14801." For such tests, ISO 14801 typically specifies minimum sample sizes (e.g., 10-11 samples for static strength, typically more for fatigue). The exact number of samples used for this specific submission is not explicitly stated, but it would have followed the standard's requirements.
- Data Provenance: The mechanical testing and material analyses are assumed to be "non-clinical data submitted or referenced" by the manufacturer, Terrats Medical SL, based in Barcelona, Spain. The "reverse engineering dimensional analysis" was done by Terrats Medical SL or through contractual agreement. This is prospective testing performed to support the 510(k). The document itself does not specify the country of origin for the underlying OEM implant data used for reverse engineering, although the OEM companies are listed (e.g., Astra Tech AB, BioHorizons).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A. This is a physical device. Ground truth, in the context of AI, refers to validated labels for data used to train and test an algorithm. For a physical device, performance is evaluated through engineering and biocompatibility testing against defined standards. There are no "experts" establishing ground truth in the AI sense. Testing would be performed by qualified engineers and technicians.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
N/A. Adjudication methods are typically used in clinical studies involving interpretation (e.g., by radiologists) to resolve discrepancies. This document describes non-clinical performance testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A. This product is a dental implant abutment, not an AI software intended to assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
N/A. This is a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
N/A. For engineering tests of physical devices, the "ground truth" is typically derived from established engineering principles, international standards (e.g., ISO 14801 for mechanical properties, ASTM for materials), and the physical properties of the materials and designs themselves. There isn't "expert consensus" or "pathology" in the AI or clinical trials sense.
8. The sample size for the training set
N/A. This is a physical device; there's no "training set" in the machine learning sense. The device is manufactured based on established engineering designs and material specifications.
9. How the ground truth for the training set was established
N/A. No training set for AI. For device manufacturing, the "ground truth" for design and production parameters comes from established engineering best practices, prior successful device designs (predicate devices), and adherence to quality systems regulations (21 CFR Part 820).
Ask a specific question about this device
(98 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system by a change in sterilization status to provide products sterile to the end user that were previously cleared to be provided non-sterile. The subject device abutments and abutment screws were cleared previously to be provided non-sterile to the end user in K170588, K191986, K212628, and K22288. All subject device components will now be provided sterile.
The subject device components include Healing Abutments, Multi-Unit Abutments (0, 17°, and 30°), and abutment screws.
The provided text is a 510(k) summary for the DESS Dental Smart Solutions, an endosseous dental implant abutment. It details the device, its intended use, and its substantial equivalence to previously cleared predicate and reference devices. However, this document does not contain the acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML medical device.
The 510(k) submission for this dental abutment focuses on establishing substantial equivalence based on:
- Design and Material: The subject device components are identical in design, material (Ti-6Al-4V alloy, DLC coating), and technological characteristics to previously cleared devices.
- Manufacturing: The manufacturing process is consistent with previously cleared devices.
- Biocompatibility: Referenced from previous K-clearances.
- Sterilization: The main change in this submission is expanding the system to provide products sterile to the end-user via gamma irradiation, which was validated by referencing a previous K-clearance (K212538).
- MR Environment Testing: Non-clinical analysis was performed to evaluate the subject devices in the MR environment, referencing published literature and FDA guidance.
- Shelf Life Testing: Referenced from a previous K-clearance (K212538) for samples after accelerated aging.
Therefore, I cannot fulfill your request to describe the acceptance criteria and a study proving an AI/ML device meets them based on the provided text, as this document is not about an AI/ML medical device. It's about dental implant abutments and establishes substantial equivalence through non-clinical performance data and comparison to predicate devices, not through AI/ML performance metrics.
To provide the information you requested, I would need a document detailing the validation of an AI/ML medical device, which would include definitions of acceptance criteria (e.g., accuracy, sensitivity, specificity), details of training and test datasets, ground truth establishment, and potentially clinical effectiveness studies if applicable.
Ask a specific question about this device
(122 days)
The Sirona Dental CAD/CAM System is intended for use in partially or fully edentulous mandibles and maxillae in support of single or multiple-unit cement retained restorations. For the BH 3.0 S. SSO 3.5 L and SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible. The system consists of three major parts: TiBase. inCoris mesostructure, and CAD/CAM software. Specifically, the inCoris mesostructure and TiBase components make up a two-piece abutment which is used in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity. The inCoris mesostructure may also be used in conjunction with the Camlog Titanium base CAD/CAM (types K2244.XXX) (K083496) in the Camlog Implant System. The CAD/CAM software is intended to design and fabricate the inCoris mesostructure. The inCoris mesostructure and TiBase two-piece abutment is compatible with the following implant systems: (list of compatible implant systems follows).
The Sirona Dental CAD/CAM System which is the subject of this premarket notification is a modification to the Sirona Dental CAD/CAM System as previously cleared under K111421. The modifications represented in the subject device consist of the implementation of a new "chairside" CAD/CAM software version, CEREC SW version 4.6.1, in which additional functionality for the control of critical CAD/CAM abutment dimensions has been added. Additionally, the modified Sirona Dental CAD/CAM System that is the subject of this premarket notification includes a line extension to the existing offerings of the Sirona TiBase titanium base component offerings. These additional TiBase variants facilitate compatibility with additional implant systems. The modified Sirona Dental CAD/CAM System which is the subject of this premarket notification consists of: CEREC SW version 4.6.1, "chairside" CAD/CAM software; CEREC AC digital acquisition unit; CEREC AC Connect digital acquisition unit; CEREC Omnicam 3D digital intraoral scanner; CEREC MCXL product family of CAM milling units; Sirona TiBase titanium base components; inCoris ZI zirconium mesostructure blocks. As subject to this premarket notification, the Sirona Dental CAD/CAM System is utilized to digitally acquire and record the topographical characteristics of teeth, dental impressions, or physical stone models in order to facilitate the computer aided design (CAD) and computer aided manufacturing (CAM) of two-piece "CAD/CAM" abutments. The patient-specific two-piece abutments consist of pre-fabricated "TiBase" components which are designed with interface geometry to facilitate compatibility and connection with currently marketed dental implant system. The CEREC SW 4.6.1 CAD/CAM software is utilized to drive the specified acquisition unit hardware to acquire the intraoral dental scans and to design the mesostructure component of the CAD/CAM abutments. Following the completion of the design, the CEREC SW 4.6.1 drives the CAM fabrication of the mesostructure component in the "chairside" workflow by utilizing the CEREC MCXL milling equipment and the defined zirconium block materials. The completed mesostructure is cemented to the TiBase component using PANAVIA F 2.0 dental cement in order to complete the finished, two-piece CAD/CAM dental abutment.
Here is the information about the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device (K111421) rather than explicitly detailing numeric acceptance criteria for each test. However, it lists the types of non-clinical performance data and states that the results support substantial equivalence. The implied acceptance criterion for all tests is "conformity" with the referenced standards or successful validation/analysis, and the reported performance is that these criteria were met.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Conformity with IEC 60601-1 (medical electrical equipment basic safety and essential performance) | Testing conducted, results reported to support substantial equivalence. |
Conformity with IEC 60601-1-2 (electromagnetic compatibility) | Testing conducted, results reported to support substantial equivalence. |
Conformity with IEC 62304 (medical device software lifecycle processes) | Software validation conducted, deliverables compiled, and included with reference to FDA guidance. |
Dynamic fatigue testing of new TiBase variants (worst-case construct) according to ISO 14801 | Dynamic fatigue testing conducted, results reported to support substantial equivalence. |
Compatibility analyses of new TiBase interface geometries with OEM implant connection geometries | Compatibility analysis conducted (reverse engineering on OEM implant body, abutment, screw, or by manufacturing agreement), results reported to support substantial equivalence. |
System validation testing (design and fabrication workflow of CEREC SW 4.6.1) | System validation testing conducted to confirm design and fabrication workflow with defined scanning, acquisition, and milling equipment. |
Software verification and validation for abutment design library (including design restrictions) | Software verification and validation conducted to demonstrate that restrictions prevent mesostructure design outside limitations and that established design limitations in the encrypted library are locked and cannot be modified. Screenshots under user verification testing were included to demonstrate this. |
Validation of recommended steam sterilization parameters (EN ISO 17665-1, ANSVAAMI ST79:2010) | Recommended parameters for steam sterilization of TiBase components validated. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the numerical sample sizes for each specific test (e.g., number of abutments for fatigue testing, number of software test cases). It refers generally to "testing" and "analyses."
The data provenance is implied to be internal testing conducted by Dentsply Sirona, as the document details their testing efforts to support the 510(k) submission. No information about country of origin of the data is provided, nor whether it was retrospective or prospective, although typically such a submission would involve prospective testing designed to meet the specified standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This information is not provided in the document. The document describes compliance with recognized standards and internal validation processes but does not detail the involvement of external experts for establishing ground truth on a test set.
4. Adjudication Method for the Test Set:
This information is not provided. The non-clinical testing appears to rely on objective measurements against established engineering and regulatory standards rather than subjective expert adjudication of results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC comparative effectiveness study was done or reported. This device is a CAD/CAM system for designing and fabricating dental abutments, not an AI or imaging diagnostic tool that would typically involve human reader performance studies.
6. Standalone (Algorithm Only) Performance Study:
Yes, a standalone performance study in the form of non-clinical performance data and software verification and validation testing was performed. The document describes:
- Testing to verify conformity with various IEC and ISO standards for medical electrical equipment, electromagnetic compatibility, and dynamic loading.
- Compatibility analyses of new TiBase interface geometries.
- System validation testing for the CAD/CAM software's design and fabrication workflow.
- Software verification and validation testing specifically for the abutment design library to demonstrate design restrictions and locked specifications.
These tests focus on the technical performance and safety of the device components and software, independent of human clinical application for their evaluation.
7. Type of Ground Truth Used:
The ground truth used for non-clinical testing is based on:
- Engineering specifications and design requirements: For confirming the functionality and outputs of the CAD/CAM system and software.
- Recognized international standards: Such as IEC 60601-1, IEC 60601-1-2, IEC 62304, and ISO 14801 which define performance benchmarks and safety criteria.
- Original manufacturer's implant connection geometries: For compatibility analyses of TiBase interfaces.
8. Sample Size for the Training Set:
This information is not applicable and therefore not provided. The Sirona Dental CAD/CAM System, as described, is not an AI/ML-based diagnostic device that typically requires a "training set" in the context of machine learning model development. It's a system for computer-aided design and manufacturing within predetermined parameters.
9. How the Ground Truth for the Training Set Was Established:
This information is not applicable, as there is no mention or indication of a "training set" for an AI/ML model for this device. The software functions based on established CAD/CAM principles and predefined parameters rather than learning from a training dataset.
Ask a specific question about this device
(73 days)
The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.
This device is compatible with the following manufacturers' implant systems: BioHorizons The Prodigy System™ Dental Implants 3.5mm, 4.0mm, 5.0mm and 6.0mm Implants and BioHorizons The Maestro™ System Implants RBM and HA coated 3.5mm, 4.0mm and 5.0mm Implants.
Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.
Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.
The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations.
The Atlantis™ Abutments in Zirconia for BioHorizons Implant and abutment screws are made from biocompatible vttria-stabilized tetragonal zirconia polycrystals (Y-TZP) and meets ISO Standards 6972 & 13356. The abutment screws are made from titanium grade Ti-6A1-V ELI and meets ASTM Standard F-136. The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with BioHorizons Prodigy System™ 3.5mm, 4.0mm, 5.0mm and 6.0mm Implants and BioHorizons The Maestro™ System Implants RBM and HA coated 3.5mm, 4.0mm and 5.0mm Implants.
The provided text is a 510(k) summary for a pre-market notification of a dental implant abutment. It does not contain any information about acceptance criteria or a study that proves the device meets specific performance criteria.
The document focuses on establishing substantial equivalence to a predicate device, as required for 510(k) submissions. This typically involves demonstrating that the new device has the same intended use, technological characteristics (materials, design), and performs similarly to a legally marketed device, often through a comparison of engineering specifications and materials rather than extensive clinical efficacy studies with specific performance metrics.
Therefore, I cannot fulfill your request for details on acceptance criteria and a study to prove they are met from the provided input. The document explicitly states:
- "The Atlantis™ Abutments in Zirconia for BioHorizons Implants are substantially equivalent in intended use, material, design and performance to the Atlantis Abutments for Zimmer Interface cleared under K053373 and BioHorizons Prodigy System™ Dental Implants cleared under K042429."
This highlights that the basis for approval is substantial equivalence, not a standalone study demonstrating performance against a set of predefined acceptance criteria for the new device itself.
To answer your specific points based on the absence of information:
- A table of acceptance criteria and the reported device performance: This information is not provided in the document.
- Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): No test set or related performance data is described in the document.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as no test set or ground truth establishment is described.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no test set or adjudication is described.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This document pertains to a dental abutment, not an AI-assisted diagnostic tool, so an MRMC study is not relevant and not mentioned.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical medical device, not a software algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable, as no performance study requiring ground truth is described.
- The sample size for the training set: Not applicable, as no algorithm training is described.
- How the ground truth for the training set was established: Not applicable, as no algorithm training is described.
Ask a specific question about this device
Page 1 of 1