Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K223714
    Date Cleared
    2023-12-29

    (382 days)

    Product Code
    Regulation Number
    872.3640
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UniFit Dental Implant System is intended for surgical placement in the maxillary and or mandibular arch to support crowns, bridges or overdentures in partially or completely edentulous patients in order to restore masticatory function. UniFit Dental Implants may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading.

    UniFit short implants (6 mm L) are intended to be used only with straight abutments.

    All digitally designed custom abutments for use with Ti Base abutments or Pre-milled Blank abutments are to be sent to an Adin Dental validated milling center for manufacture.

    Device Description

    Adin's UniFit Dental Implant System is a new model that provides an additional connection platform to Adin's legally marketed Dental Implant Systems, indicated for use in surgical and restorative applications for placement in the maxillary and/or mandibular arch to provide support for prosthetic devices such as crowns, bridges, or overdentures in order to restore masticatory function.

    This new UniFit model is identical to Adin's cleared Touareg™ CloseFit (RP- Regular Platform, and WP- Wide Platform) and Adin's Touareg™-S implant systems (cleared under K112585 and K081751, respectively) except for the "Star" (Torx) connection and new 6mm length for specific diameters.

    The UniFit Dental Implant System includes dental implants, screws and prosthetic components for CAD/CAM restorations.

    The UniFit dental implants are tapered core implants with a spiral tap, and a dome apex and double lead thread design.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the UniFit Dental Implant System.

    Please Note: The document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets strict numerical acceptance criteria against a pre-defined performance target. The studies described are primarily to ensure safety, functionality, and equivalence to already cleared devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a submission focused on substantial equivalence to predicate devices rather than meeting a specific performance target for AI, the "acceptance criteria" here refer to meeting the established performance standards of the predicate devices and general safety/performance testing for the type of medical device. The document does not provide a traditional table of AI-specific acceptance criteria (e.g., sensitivity, specificity, AUC) or corresponding numerical performance for the device's AI component. Instead, it describes a series of engineering and biocompatibility tests to ensure the new device is as safe and effective as previously cleared devices.

    The "device performance" reported is that the device successfully meets the internal protocols and relevant ISO/ASTM standards, indicating it performs comparably to the predicate devices and is safe for its intended use.

    No AI is mentioned in the provided text for the UniFit Dental Implant System. The document describes a physical dental implant system and components, not a software or AI-driven medical device. Therefore, the subsequent questions regarding AI-specific criteria (sample size for test/training sets, experts, MRMC studies, standalone performance) are not applicable based on this document.

    Test/Evaluation CategoryAcceptance Criteria (Implicit from text)Reported Device Performance
    Material BiocompatibilityCompliance with ISO 10993-1, ISO 7405, FDA guidance, ISO 10993-5 (Cytotoxicity: non-cytotoxic), ISO 10993-11 (Pyrogenicity), ISO 10993-18 (Chemical extractions).- In-vitro Cytotoxicity test met criteria (non-cytotoxic for worst-case combinations).
    • Pyrogenic Material-Mediated and chemical extractions conducted.
    • Biocompatibility tests were conducted in accordance with referenced standards. |
      | Connection Design Effectiveness | Stable connection and all measurements within component's specification, following tolerance analysis. | All tested specimens showed stable connection and all measurements were found to be within each component's specification. |
      | Maximal Applicable Torque Resistance (Implant) | Withstand maximal insertion torque as predefined in implant specification, considering safety margins and device design. | Test conducted in accordance with internal protocol; implies meeting criteria (no explicit numerical value reported as metric). |
      | Tool to Implant Connection Lifetime Performance | Maximal applied torque and deformation test met predefined acceptance criteria; no mechanical or visual damages observed at end of reusable tool's recommended lifetime. | Met predefined acceptance criteria; no mechanical or visual damages observed. |
      | Implant-Abutment Connection Degree of Rotation | Predefined acceptance criterion for degree of free rotation after applying closure torque. | All tested specimens met the predefined acceptance criterion. |
      | Abutment Screw Performance - Maximum Applicable Torque | Torque resistance of connecting screw such that breakage occurs above acceptance criteria and at the desired location (under the screw head/groove). | All tested screws failed in values above the acceptance criteria and at the desired location. |
      | Screw Loosening and Abutment Extraction by Retrieval Tool | Ability to screw and loosen with no deformation or fracture; abutment easily retrieved from implant under lateral pressure; screw retrieval torque withstands predefined torque. | All tested screws and abutments were successfully unscrewed and retrieved; screw retrieval torque withstood predefined torque. |
      | Regular Platform Dynamic Loading (Fatigue) | Meet requirements of ISO 14801:2016 for fatigue load (fatigue limit) for endosseous dental implants under "worst case" conditions and prosthetic components. | All of the success criteria were met. |
      | Torsion Testing | Meet acceptance criteria for torsional performance as per YY0315:2016 for torsional yield strength and maximum torque of worst-case implant/connecting part joints. | All tested specimens met acceptance criteria for UniFit implant system torsional performance per YY0315:2016. |
      | 6.00mm Implants Pull-Out Test | Axial pull-out strength comparable to or better than predicate devices (MIS short implants K103089), per ASTM F543. | The test met its acceptance criteria (comparison with legally marketed device). |
      | Comparative Surface Area Analysis (before treatment) | UniFit Short Implant's actual surface area (before surface treatment) greater than or equivalent to legally marketed reference item (MIS short implants K103089) at worst-case implant variation. | The total surface area measurements of the tested items were greater than the legally marketed reference item, the acceptance criteria was met. |
      | Comparative Bone to Implant Contact (BIC) Surface Area Analysis | Test article (Adin's UniFit short implant, 6.0 mm) BIC volume and surface area comparable to or exceeding reference item (MIS SEVEN MF7-06420) for Hard Bone (Type I) and Soft Bone (Type IV) conditions. | Test article, at both conditions of the bone, exceeded the MIS reference item in terms of B.I.C. volume and surface area. |
      | Single TMA Fatigue Rationale | Fatigue limit of Single TMA abutments assembled to designated Dental Implants (including compatible screw) conforms to ISO 14801:2016, and does not present new worst-case compared to existing and approved TMA system. | The Single TMA and designated superstructures do not present any new worst-case when compared to the existing and approved TMA system (Rationale provided). |
      | Fatigue testing of TiBase Abutments | Worst-case combination of subject device implants, TiBase abutments, and zirconia superstructure is strong enough for clinical application per ISO 14801. | Tested according to ISO 14801; implies meeting strength requirements (no explicit numerical value reported as metric). |
      | Gamma Irradiation Sterilization | Assurance of SAL (Sterility Assurance Level) of 10^-6 using VDmax 20kGy according to ISO 11137-2:2013 and ISO/TS 13004. | Sterilization validated in accordance with ISO 11137-2:2013, assuring SAL of 10^-6. |
      | Cleaning and Steam Sterilization (End-User) | Assurance of SAL of 10^-6 for steam sterilization (for non-sterile components) using full cycle approach as detailed in Annex D4 of ISO 17665-1:2006, in accordance with ISO 17665-1:2006 and ISO 17665-2:2009. | Steam sterilization validated in accordance with ISO 17665-1:2006 and ISO 17665-2:2009, assuring SAL of 10^-6. |
      | Disinfection (End-User) | Disinfection process validated in accordance with ISO 17664-1:2021 and AAMI TIR12:2020. | Disinfection process validation conducted with accordance with ISO 17664-1:2021 and AAMI TIR12:2020. |
      | Shelf-Life, Packaging and Transportation | Ensure 5 years shelf life under accelerated and real-time aging, and confirmed sterility throughout shelf life and after packaging/transportation per ISO 11607-1. | Shelf-life tests conducted to ensure 5 years shelf life, confirming sterility after packaging and transportation. |

    2. Sample Size Used for the Test Set and Data Provenance

    As no AI component is described, this question is not directly applicable. The performance data section describes various engineering tests.

    • Test Set Description: The "test set" for the engineering evaluations consists of physical samples of the UniFit Dental Implant System.
    • Sample Sizes:
      • For most tests (connection design effectiveness, torque resistance, tool life, rotation, screw performance, screw loosening/extraction, torsion testing, fatigue testing of TiBase abutments), the document states "All tested specimens" or "All tested screws," implying all manufactured or a representative sample of components tested met the criteria. It does not provide specific numerical counts for these "specimens" or "screws."
      • For the UniFit 6.00mm Implants Pull-Out test, comparison was made to "proposed predicate devices MIS short implants (K103089)." The sample size for UniFit is not explicitly stated but implies sufficient numbers for comparison.
      • For Comparative Surface Area Analysis and Comparative Bone to Implant Contact Surface Area Analysis, comparison was made against "legally marketed devices (MIS short implants cleared under 510(k) K103089) at worst case implant variation" and "MIS's SEVEN MF7-06420." Again, specific sample sizes for the UniFit components are not given, but "tested items" is used.
    • Data Provenance: The data comes from internal testing conducted by Adin Dental Implant Systems Ltd. (implied by "Adin internal protocol") and potentially external accredited labs for biocompatibility and sterilization (e.g., "Tests have been conducted at MDT Medical Device Testing GmbH"). The tests are in accordance with international standards (ISO, ASTM, YY) and FDA guidance. The implants are compared to predicate devices already marketed in the US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This question is not applicable as the document does not describe the use of experts to establish a "ground truth" for an AI algorithm's performance on a test set. The validation is through engineering, materials, and biocompatibility testing against established standards and comparisons to predicate devices, which generally do not involve human "experts" establishing ground truth in the AI sense.


    4. Adjudication Method for the Test Set

    Not applicable, as there is no AI component or subjective human interpretation of results requiring adjudication for ground truth.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable, as no AI component is described.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as no AI component is described.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For this device, "ground truth" is established through:

    • Compliance with International Standards: Adherence to ISO and ASTM standards for material properties, mechanical performance, fatigue, sterility, and biocompatibility.
    • Comparison to Predicate Devices: Performance characteristics (e.g., implant design, connection types, material, dimensions, sterilization methods, clinical indications) are compared against several legally marketed predicate and reference devices (e.g., Adin's Touareg™ CloseFit Dental Implant System, MIS C1 implants, Straumann BLX System, etc.). The "ground truth" implicitly aligns with the documented safety and effectiveness of these predicate devices.
    • Internal Protocols: Performance data is generated and evaluated against Adin's internal protocols and specifications.

    8. The Sample Size for the Training Set

    Not applicable, as no AI component or machine learning model is described that would require a training set.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no AI component or machine learning model is described that would require a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212108
    Device Name
    Dynamic TiBase
    Date Cleared
    2021-09-24

    (80 days)

    Product Code
    Regulation Number
    872.3630
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K182246, K034014, K072933, K191919, K162021

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.

    Device Description

    The purpose of this submission is to obtain marketing clearance for Dynamic TiBase, a series of twopiece titanium base abutments that require the fabrication of patient-specific custom superstructures using CAD-CAM technology. In final, finished form, the subject device abutments are intended to be used as a two-piece abutment composed of the base bottom-half (titanium base) with a bonded CAD-CAM zirconia top-half. Each patient-specific zirconia superstructure is individually prescribed by the clinician and manufactured by an authorized milling center.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Talladium España, SL Dynamic TiBase, which is an endosseous dental implant abutment. The primary purpose of this submission is to demonstrate substantial equivalence to legally marketed predicate devices, not to prove clinical effectiveness or safety in the same way clinical trials do. Therefore, the information provided focuses on non-clinical data and comparisons to existing devices, rather than clinical study results with human subjects.

    Based on the provided document, here's a breakdown of the acceptance criteria and the study that proves the device meets them, specifically in the context of a 510(k) submission:

    1. A table of acceptance criteria and the reported device performance

    For a 510(k) submission for a device like this, the "acceptance criteria" are not clinical performance metrics in the traditional sense (e.g., sensitivity, specificity for a diagnostic device). Instead, they revolve around demonstrating substantial equivalence to predicate devices in terms of:

    • Intended Use: The device performs the same function for the same patient population.
    • Technological Characteristics: The device has similar design, materials, and operating principles.
    • Performance (Non-Clinical): The device performs similarly to or better than the predicate in engineering/bench tests relevant to its function and safety. Any differences in technological characteristics must not raise new questions of safety or effectiveness.

    The reported device performance is primarily based on non-clinical (bench) testing and comparison tables, not clinical studies in human subjects.

    Here's a table summarizing the "acceptance criteria" (or areas of equivalence demonstration) and the reported device performance as per the document:

    Acceptance Criterion (Area of Equivalence)Reported Device Performance (as demonstrated in 510(k))
    Intended UseDynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
    Demonstrated Equivalence: Stated to be substantially equivalent to primary predicate (K202026) and reference devices (K182246, K191919, K162021) in indications for use. Differences in wording do not affect intended use.
    Technological CharacteristicsDevice Type: CAD-CAM Titanium Base Abutments (two-piece, titanium base with bonded CAD-CAM zirconia superstructure).
    Materials: Titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and ISO 5832-3 for abutments and screws. Zirconia (ISO 13356) for copings.
    Design/Features: Straight design, 0.7 mm gingival height, cut-out for angled screw channel, specific post heights. Compatible with Thommen SPI® CONTACT dental implant fixtures.
    BiocompatibilityAll subject device abutments are biocompatible.
    Performance: Biocompatibility testing performed according to ISO 10993-12 (referenced) and confirmatory testing for finished devices (ISO 10993-5 and ISO 10993-12) was done.
    Sterilization ValidationThe device can be effectively sterilized by the end-user.
    Performance: Moist heat sterilization validation performed according to ISO 17665-1 and ISO 11737-2 (referenced). Device provided non-sterile for end-user sterilization.
    Mechanical Performance (Strength/Fatigue)The device is mechanically robust enough for its intended use.
    Performance: Static compression and compression fatigue testing performed according to ISO 14801 of worst-case constructs (abutments, zirconia superstructures, abutment screws, compatible implants). The fatigue limit data demonstrated sufficient strength. Differences in superstructure design parameters (wall thickness, gingival height) were mitigated by mechanical testing.
    Manufacturing ProcessThe manufacturing process is consistent and yields safe devices.
    Performance: Titanium alloy components manufactured from identical materials in identical facilities using identical processes as previously cleared products (K162021). Anodization process is also identical.
    Compatible Implant InterfaceThe device properly interfaces with specified implants.
    Performance: Established compatibility with Thommen SPI® CONTACT dental implant fixtures (cleared in K034014 and K072933) via business agreement and identical interface connections/platforms.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document explicitly states "No clinical data were included in this submission." The "studies" were non-clinical (bench) tests. While the ISO 14801 standard specifies the number of samples for fatigue and static testing (typically 5-10 for static, 10-12 for fatigue), the exact sample sizes used for each specific test (e.g., number of abutments tested at different loads) are not detailed in this summary. It only says "worst-case constructs" were tested.
    • Data Provenance: The data is non-clinical (bench testing) performed by the manufacturer or a contracted lab. The manufacturer is Talladium España, SL, located in Lleida, Spain. This indicates the testing was likely done in Spain or a European lab adhering to ISO standards. The data is retrospective in the sense that the tests were completed before the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For a 510(k) submission based on non-clinical data, there are no "experts" establishing a clinical "ground truth" through consensus on patient cases. The "ground truth" for non-clinical testing is compliance with established engineering and biocompatibility standards (e.g., ISO 14801, ISO 10993). The "experts" here would be the engineers and scientists who designed and conducted the tests and analyzed the results according to these international standards. Their qualifications are inherent in their ability to perform such testing and demonstrate compliance with these standards.

    4. Adjudication method for the test set

    Not applicable. There's no clinical test set requiring adjudication. The non-clinical tests (biocompatibility, sterilization, mechanical) are evaluated against predefined acceptance criteria established by the relevant ISO standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental implant abutment, not an AI-powered diagnostic imaging tool. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this 510(k) submission is compliance with established international engineering standards (e.g., ISO 14801 for mechanical performance, ISO 10993 for biocompatibility, ISO 17665-1 and ISO 11737-2 for sterilization) and the demonstration of substantial equivalence to predicate devices. This is achieved through:

    • Bench Test Results: Quantitative data from static compression and compression fatigue tests.
    • Material Specifications: Conformance to ASTM and ISO material standards.
    • Biocompatibility Test Reports: Results showing no adverse biological reactions.
    • Sterilization Validation Data: Proof of effective sterilization.
    • Direct Comparison: Tables and textual analysis comparing the subject device's features and performance to those of legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1