K Number
K193529

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-03-05

(76 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ATLANTIS® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment screw is intended to secure the ATLANTIS® Abutment to the endosseous implant.

The ATLANTIS® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS® Crown Abutment to the endosseous implant.

The ATLANTIS® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis. in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment screw is intended to secure the ATLANTIS® Conus Abutment to the endosseous implant.

The ATLANTIS® Healing Abutment can be used with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The abutment screw is intended to secure the ATLANTIS® Healing Abutment to the endosseous implant.

ATLANTIS® Abutment is compatible with MIS Conical Connection implant from MIS Implant System.

ATLANTIS® products are compatible with the implants shown in the table below.

Implant manufacturer - MIS-IMPLANT TECHNOLOGIES INC
Trade NameAbutment Platform Diameter
Atlantis Abutment for MIS V3 NPØ2.765mm
Atlantis Abutment for MIS C1 NPØ2.76mm
Atlantis Abutment for MIS C1 & V3 SPØ3.16mm
Atlantis Abutment for MIS C1 WPØ4.01mm

Implant Diameter V3: Ø3.30 mm C1: Ø3.30 mm C1: Ø3.75,4.2 mm, V3: Ø3.90,4.3,5.0 mm C1: Ø5.0 mm

Device Description

The proposed ATLANTIS® Abutments for MIS Conical Connection Implant are endosseous dental implant abutments.

The proposed devices are compatible with:

  • MIS V3 conical connection narrow and standard dental implant diameters Ø3.3, 3.9, 4.3 and a. 5.0 mm (K163349)
  • b. MIS C1 narrow, standard and wide platform conical connection implant diameters Ø3.3, 3.75, 4.2, and 5.0 (K172505, K112162)

Refer to Table 5.1 for the implants the proposed ATLANTIS® Abutments for MIS Conical Connection Implant are compatible with.

Table 5.1 Implant systems which proposed ATLANTIS® Abutments for MIS Conical ConnectionImplant are compatible with.
Trade NameAbutment Platform DiameterImplant Diameter
Atlantis Abutment for MIS V3 NPØ2.765mmV3: Ø3.30 mm
Atlantis Abutment for MIS C1 NPØ2.76mmC1: Ø3.30 mm
Atlantis Abutment for MIS C1 & V3 SPØ3.16mmC1: Ø3.75, 4.2 mmV3: Ø3.90, 4.3, 5.0 mm
Atlantis Abutment for MIS C1 WPØ4.01mmC1: Ø5.0 mm

The abutments are available in four (4) designs:

  • ATLANTIS® Abutment for MIS Conical Connection Implant, a.
  • ATLANTIS® Crown Abutment for MIS Conical Connection Implant, b.
  • ATLANTIS® Conus Abutment (Custom or Overdenture) for MIS Conical Connection C. Implant
  • d. ATLANTIS® Healing Abutment for MIS Conical Connection Implant

The materials composition of the proposed devices are described below in Table 5.2.

Table 5.2 ATLANTIS® Abutment for MIS Conical Connection Implant Materials
AbutmentATLANTIS®Abutment forMIS ConicalConnectionImplantATLANTIS®CrownAbutment forMIS ConicalConnectionImplantATLANTIS®ConusAbutment(Custom) forMIS ConicalConnectionImplantATLANTIS®ConusAbutment(Overdenture)for MISConicalConnectionImplantATLANTIS® HealingAbutmentfor MIS ConicalConnection Implant
MaterialsTitanium, Gold-shaded Titanium (Gold-Hue)TitaniumTitanium, Gold-shaded Titanium (Gold-Hue)TitaniumTitanium, Gold-shaded Titanium (Gold-Hue)

The maximum abutment height is 15 mm above implant interface and the minimum abutment height is 4 mm above the trans-mucosal collar. The abutments are provided straight and up to 30° of angulation.

All proposed abutments are patient-specific abutments fabricated using CAD/CAM technology by Dentsply Sirona Implants. Each abutment is designed according to prescription instructions from the clinician to support a screw-retained, cement-retained or friction fit prosthesis.

The coronal portion of the ATLANTIS® Abutments for MIS Conical Connection Implant can be fabricated as a conventional abutment for prosthesis attachment (ATLANTIS® Abutment for MIS Conical Connection Implant or ATLANTIS® Conus Abutment for MIS Conical Connection Implant), fabricated as a single tooth final restoration onto which porcelain is added ATLANTIS® Crown Abutment for MIS Conical Connection Implant) or fabricated with a short core for soft tissue healing (ATLANTIS® Healing Abutment).

AI/ML Overview

This document describes the Dentsply Sirona ATLANTIS® Abutment for MIS Conical Connection Implants and its substantial equivalence to predicate devices, based on non-clinical performance data. There is no information provided about an AI/ML device in this document, so the questions regarding AI/ML device performance are not applicable.

Here's an analysis based on the provided text, focusing on the mechanical and material aspects of the device, and addressing the questions where applicable:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in a quantitative manner with reported performance metrics to directly compare. Instead, it relies on demonstrating substantial equivalence to predicate devices through various tests and comparisons. The performance is assessed against established standards and predicate device characteristics.

However, based on the non-clinical performance data section, we can infer some criteria and the types of performance reported:

Acceptance Criteria (Inferred from tests)Reported Device Performance (ATLANTIS® Abutment for MIS Conical Connection Implants)
Fatigue Resistance (according to ISO 14801)Fatigue test results were compared to the predicate and reference devices. The proposed device's fatigue testing results are substantially equivalent to those of the ATLANTIS® Abutment for HIOSSEN ET Implant (K160626).
Geometric Compatibility with specified MIS Conical Connection ImplantsGeometric compatibility analysis was conducted on OEM implant bodies, OEM abutments, and OEM screws. This supported the dimensional compatibility of the ATLANTIS® Abutment for MIS Conical Connection Implant with MIS V3 implants (Narrow: Ø3.3mm and Standard: Ø3.9, 4.3, 5.0mm) and MIS C1 implants (Narrow Ø3.3mm, Standard: Ø3.75, 4.2mm and Wide: Ø5.0, 6.0 mm).
Sterilization Efficacy (according to ISO 17665-1)Sterilization validation of the proposed device is referenced by equivalency to the sterilization validation of predicate (K172225), which was validated according to ISO 17665-1.
Biocompatibility (based on material composition and manufacturing process)The material composition and manufacturing processing methods of the proposed device are identical to the predicate device (K172225). Therefore, no additional biocompatibility data was included to support substantial equivalence.
Design Specifications (e.g., angulation, height, width)The final design of the proposed device adheres to the same design limitations as the predicate (K172225), including: max abutment angulation of 30°, max abutment height of 15 mm above the implant interface, max abutment post height above the trans-mucosal collar of 4 mm, and final abutment width of 3.3 mm to 6.5 mm.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample size for the "test set" in terms of number of abutments or implants used, nor does it specify the country of origin or whether the data was retrospective or prospective for the non-clinical tests. It refers to "fatigue testing" and "geometric compatibility analysis," which are bench tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable as the document describes a dental abutment, not an AI/ML device requiring expert ground truth for interpretation of images or other data. The "ground truth" for this device would be its physical and mechanical properties, assessed through engineering and materials testing, not expert review.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 typically refer to expert review processes for clinical or image interpretation studies, which are not relevant to the non-clinical bench testing described for this dental abutment.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document is about a dental implant abutment, not an AI-assisted diagnostic or therapeutic device. There are no human readers or AI involved in its function.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance relies on established engineering standards and material properties, and comparison to legally marketed predicate devices' performance.

  • Fatigue Resistance: Ground truth would be defined by the specified load cycles and failure criteria outlined in ISO 14801.
  • Geometric Compatibility: Ground truth is the precise dimensional specifications of the OEM implant bodies, OEM abutments, and OEM screws.
  • Sterilization: Ground truth is the successful validation according to ISO 17665-1.
  • Biocompatibility: Ground truth is the conformity of materials to ASTM F136 and the established biocompatibility of the predicate device (K172225) using identical materials and processes.
  • Design Specifications: Ground truth lies in the adherence to predefined maximums and minimums for angulation, height, and width.

8. The sample size for the training set

Not applicable as this is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable as this is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

March 5, 2020

Dentsply Sirona Karl Nittinger Vice President, Corporate Regulatory Affairs 221 West Philadelphia Street, Suite 60W York, Pennsylvania 17401

Re: K193529

Trade/Device Name: ATLANTIS® Abutment for MIS Conical Connection Implants Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: December 19, 2019 Received: December 20, 2019

Dear Karl Nittinger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193529

Device Name

ATLANTIS® Abutment for MIS Conical Connection Implants

Indications for Use (Describe)

The ATLANTIS® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment screw is intended to secure the ATLANTIS® Abutment to the endosseous implant.

The ATLANTIS® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS® Crown Abutment to the endosseous implant.

The ATLANTIS® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis. in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment screw is intended to secure the ATLANTIS® Conus Abutment to the endosseous implant.

The ATLANTIS® Healing Abutment can be used with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The abutment screw is intended to secure the ATLANTIS® Healing Abutment to the endosseous implant.

ATLANTIS® Abutment is compatible with MIS Conical Connection implant from MIS Implant System.

ATLANTIS® products are compatible with the implants shown in the table below.

Implant manufacturer - MIS-IMPLANT TECHNOLOGIES INC
Trade NameAbutment Platform Diameter
Atlantis Abutment for MIS V3 NPØ2.765mm
Atlantis Abutment for MIS C1 NPØ2.76mm
Atlantis Abutment for MIS C1 & V3 SPØ3.16mm
Atlantis Abutment for MIS C1 WPØ4.01mm

Implant Diameter V3: Ø3.30 mm C1: Ø3.30 mm C1: Ø3.75,4.2 mm, V3: Ø3.90,4.3,5.0 mm C1: Ø5.0 mm

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Dentsply Sirona Inc. 221 West Philadelphia Street Suite 60W York, PA 17401

Image /page/4/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized, abstract shape on the left, resembling a curved, segmented leaf or a stylized letter 'S'. To the right of the shape are the words "Dentsply" on the top line and "Sirona" on the bottom line. The text is in a simple, sans-serif font and is aligned with the right edge of the abstract shape.

510(k) SUMMARY K193529 ATLANTIS Abutment for MIS Conical Connection Implants

    1. Submitter Information:
      Dentsply Sirona Inc. 221 West Philadelphia Street Suite 60W York, PA 17401
Contact Person:Karl Nittinger
Telephone Number:717-849-4424
Fax Number:717-849-4343
Date Prepared:05 March 2020

2. Device Identification:

  • . Proprietary Name: ATLANTIS® Abutment for MIS Conical Connection Implants ● Classification Name: Endosseous dental implant abutment ● CFR Number: 21 CFR 872.3630 Device Class: Class II ●
  • . Product Code: NHA

3. Predicate Device:

Predicate Device Name510(k)Company Name
ATLANTIS® Abutment for MISImplantK172225MIS Implants Technologies LTD(Dentsply Sirona)
Reference Devices:
ATLANTIS® Abutment forHIOSSEN ET ImplantK160626Dentsply Sirona Implants
MIS V3 Conical Connection DentalImplant SystemK163349MIS Implants Technologies LTD(Dentsply Sirona)
MIS C1 Narrow Platform ConicalConnection Implant System, MIS C1Wide Platform Conical ConnectionAbutmentK172505MIS Implants Technologies LTD(Dentsply Sirona)
MIS C1 Standard Platform ConicalConnection Dental Implant systemMIS C1 Wide Platform ConicalConnection Dental Implant systemK112162MIS Implants Technologies LTD(Dentsply Sirona)

The proposed ATLANTIS® Abutments for MIS Conical Connection Implant are compatible with Implant Systems cleared in the reference devices K172505, K112162, and K163349.

{5}------------------------------------------------

    1. Device Description
      The proposed ATLANTIS® Abutments for MIS Conical Connection Implant are endosseous dental implant abutments.

The proposed devices are compatible with:

  • MIS V3 conical connection narrow and standard dental implant diameters Ø3.3, 3.9, 4.3 and a. 5.0 mm (K163349)
  • b. MIS C1 narrow, standard and wide platform conical connection implant diameters Ø3.3, 3.75, 4.2, and 5.0 (K172505, K112162)

Refer to Table 5.1 for the implants the proposed ATLANTIS® Abutments for MIS Conical Connection Implant are compatible with.

Table 5.1 Implant systems which proposed ATLANTIS® Abutments for MIS Conical ConnectionImplant are compatible with.
Trade NameAbutment Platform DiameterImplant Diameter
Atlantis Abutment for MIS V3 NPØ2.765mmV3: Ø3.30 mm
Atlantis Abutment for MIS C1 NPØ2.76mmC1: Ø3.30 mm
Atlantis Abutment for MIS C1 & V3 SPØ3.16mmC1: Ø3.75, 4.2 mmV3: Ø3.90, 4.3, 5.0 mm
Atlantis Abutment for MIS C1 WPØ4.01mmC1: Ø5.0 mm

The abutments are available in four (4) designs:

  • ATLANTIS® Abutment for MIS Conical Connection Implant, a.
  • ATLANTIS® Crown Abutment for MIS Conical Connection Implant, b.
  • ATLANTIS® Conus Abutment (Custom or Overdenture) for MIS Conical Connection C. Implant
  • d. ATLANTIS® Healing Abutment for MIS Conical Connection Implant

The materials composition of the proposed devices are described below in Table 5.2.

Table 5.2 ATLANTIS® Abutment for MIS Conical Connection Implant Materials
AbutmentATLANTIS®Abutment forMIS ConicalConnectionImplantATLANTIS®CrownAbutment forMIS ConicalConnectionImplantATLANTIS®ConusAbutment(Custom) forMIS ConicalConnectionImplantATLANTIS®ConusAbutment(Overdenture)for MISConicalConnectionImplantATLANTIS® HealingAbutmentfor MIS ConicalConnection Implant
MaterialsTitanium, Gold-shaded Titanium (Gold-Hue)TitaniumTitanium, Gold-shaded Titanium (Gold-Hue)TitaniumTitanium, Gold-shaded Titanium (Gold-Hue)

{6}------------------------------------------------

The maximum abutment height is 15 mm above implant interface and the minimum abutment height is 4 mm above the trans-mucosal collar. The abutments are provided straight and up to 30° of angulation.

All proposed abutments are patient-specific abutments fabricated using CAD/CAM technology by Dentsply Sirona Implants. Each abutment is designed according to prescription instructions from the clinician to support a screw-retained, cement-retained or friction fit prosthesis.

The coronal portion of the ATLANTIS® Abutments for MIS Conical Connection Implant can be fabricated as a conventional abutment for prosthesis attachment (ATLANTIS® Abutment for MIS Conical Connection Implant or ATLANTIS® Conus Abutment for MIS Conical Connection Implant), fabricated as a single tooth final restoration onto which porcelain is added ATLANTIS® Crown Abutment for MIS Conical Connection Implant) or fabricated with a short core for soft tissue healing (ATLANTIS® Healing Abutment).

    1. Indications for Use
      The ATLANTIS® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS® Abutment to the endosseous implant.

The ATLANTIS® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS® Crown Abutment to the endosseous implant.

The ATLANTIS® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS® Conus Abutment to the endosseous implant.

The ATLANTIS® Healing Abutment can be used with an endosseous implant for temporary use during soft tissue healing after one-stage or two-stage surgeries. The abutment screw is intended to secure the ATLANTIS® Healing Abutment to the endosseous implant.

ATLANTIS® Abutment is compatible with MIS Conical Connection implant from MIS Implant System.

ATLANTIS® products are compatible with the implants shown in the table below.

Implant manufacturer - MIS-IMPLANT TECHNOLOGIES INC
Trade NameAbutment Platform DiameterImplant Diameter
Atlantis Abutment for MIS V3 NPØ2.765mmV3: Ø3.30 mm
Atlantis Abutment for MIS C1 NPØ2.76mmC1: Ø3.30 mm
Atlantis Abutment for MIS C1 & V3 SPØ3.16mmC1: Ø3.75,4.2 mm,V3: Ø3.90,4.3,5.0 mm
Atlantis Abutment for MIS C1 WPØ4.01mmC1: Ø5.0 mm

{7}------------------------------------------------

6. Substantial Equivalence Discussion

ATLANTIS® Abutment for MIS Conical Connection Implant is a patient specific restorative device designed by Dentsply Sirona Implants technicians and manufactured by Dentsply Sirona Implants using CAD/CAM technology.

With the exception of the abutment to implant interface geometry, the design of the ATLANTIS® Abutment for MIS Conical Connection Implant, ATLANTIS® Conus Abutment for MIS Conical Connection Implant, and ATLANTIS® Crown Abutment for MIS Conical Connection Implant are identical to the design of the predicate ATLANTIS® Abutment for MIS Implant (K172225). Both the proposed ATLANTIS® Abutment for MIS Conical Connection Implants and the predicate device (K172225) are composed of the identical titanium allov conforming to ASTM F136 (Standard Specification for Wrought Titanium-6 Alumin-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications) and are both offered in a "Gold Hue" variants which feature a titanium nitride coating to produce a gold-shaded external surface. The difference in implant connection interface geometry is due to the compatibility requirement of the proposed devices with the specified MIS Conical Connection Implants (versus that of the predicate device (K172225) which feature implant interface geometry designed for compatibility with MS Internal Hex Implants). Dynamic fatigue testing to international standard ISO 14801 (Dental-implants Dynamic Fatigue Test for Endosseous Dental Implants) has been conducted and is included in support of substantial equivalence with respect to the difference in implant interface geometry.

The abutment platform diameter range of the proposed devices is similar to that of the predicate device (K172225). Differences in abutment platform diameter do not impact substantial equivalence in that abutment platform diameter changes according to the corresponding platform connection. However, specifications, including abutment diameter, of the final, finished abutment are identical for all ATLANTIS® abutments.

The design of the ATLANTIS® Healing Abutment is similar to the design of the predicate. V3 Conical Connection Healing Caps (K163349). The differences in height above the margin do not affect substantial equivalence as the healing abutment is used only in the healing phase and does not carry any load. In the case of the subject ATLANTIS Healing Abutment, the healing cap diameter is the customized, patient-specific portion of the healing cap.

The design specifications of the final, finished patient-specific design of the proposed ATLANTIS® Abutment for MIS Conical Connections Implants are identical to those of the predicate ATLANTIS® Abutment for MIS Implant cleared under K172225. The final design of both the proposed and predicate (K172225) adhere to the same design limitations within the final patient-specific design, including: maximum abutment angulation of 30°, maximum abutment height of 15 mm above the implant interface, maximum abutment post height above the trans-mucosal collar of 4 mm, as well as, final abutment width of 3.3 mm to 6.5 mm.

The proposed devices have the identical intended use, are manufactured using the identical materials and processes, designed and manufactured in the identical facility, and are characterized by the identical fundamental product technology as the predicate device (K172225). The indications for use of the proposed device are similar to those of the predicate device (K172225) but differ in that they reference the proposed device's compatibility with the MIS Conical Connection Implants (while the predicate device (K172225) indications reference compatibility with MIS Internal Hex Implants). The proposed indications for use also include the indications for the ATLANTIS Healing Abutments intended as compatible with the MIS Conical Connection Implants in this premarket notification. Geometric compatibility analysis and bench testing have been included in support of substantial equivalence with respect to these differences in the proposed indications for use compared to the cleared indications for use of the predicate device (K172225).

Table 5.3 compares the proposed devices [ATLANTIS® Abutment, ATLANTIS® Crown Abutment, and ATLANTIS® Conus Abutment (Custom or Overdenture)]. Table 5.4 compares the proposed ATLANTIS® Healing Abutment with the reference device, K163349, V3 Conical Connection Healing Caps.

{8}------------------------------------------------

Table 5.3 - Substantial Equivalence Comparison Summary (ATLANTIS Abutments, ATLANTIS Conus Abutments, ATLANTISCrown Abutments
ATLANTIS® Abutment for MIS Conical ConnectionImplantProposed deviceATLANTIS® Abutment for MIS ImplantK172225Predicate Device
The ATLANTIS® Abutment is intended for use with anendosseous implant to support a prosthetic device in a partiallyor completely edentulous patient. It is intended for use to supportsingle and multiple tooth prosthesis, in mandible or maxilla. Theprosthesis can be cemented or screw retained to the abutment.The abutment screw is intended to secure the ATLANTIS®Abutment to the endosseous implant.The ATLANTIS® Abutment is intended for use with an endosseousimplant to support a prosthetic device in a partially or completelyedentulous patient. It is intended for use to support single and multipletooth prosthesis, in mandible or maxilla. The prosthesis can becemented or screw retained to the abutment. The abutment screw isintended to secure the ATLANTIS® Abutment to the endosseousimplant.
The ATLANTIS® Crown Abutment is intended for use withan endosseous implant to function as a substructure that alsoserves as the final restoration, in a partially or completelyedentulous patient. The abutment screw is intended to secure theATLANTIS® Crown Abutment to the endosseous implant.The ATLANTIS® Crown Abutment is intended for use with anendosseous implant to function as a substructure that also serves as thefinal restoration, in a partially or completely edentulous patient. Theabutment screw is intended to secure the ATLANTIS® CrownAbutment to the endosseous implant.
The ATLANTIS® Conus Abutment is intended for use with anendosseous implant to support a prosthetic device in partially orcompletely edentulous patients. It is intended for use to support aremovable multiple tooth prosthesis, in the mandible or maxilla.The prosthesis is attachment-retained by friction fit to theabutment. The abutment screw is intended to secure theATLANTIS® Conus Abutment to the endosseous implant.The ATLANTIS® Conus Abutment is intended for use with anendosseous implant to support a prosthetic device in partially orcompletely edentulous patients. It is intended for use to support aremovable multiple tooth prosthesis, in the mandible or maxilla. Theprosthesis is attachment-retained by friction fit to the abutment. Theabutment screw is intended to secure the ATLANTIS® ConusAbutment to the endosseous implant.
The ATLANTIS® Healing Abutment can be used with anendosseous implant for temporary use during soft tissue healingafter one-stage or two-stage surgeries. The abutment screw isintended to secure the ATLANTIS® Healing Abutment to theendosseous implant.ATLANTIS® Abutment for MIS implant is compatible with MISimplant from MIS Implant System. MIS short implants (6mm) are to beused only with straight abutments.ATLANTIS® products are compatible with the implants shown in the
ATLANTIS® Abutment is compatible with MIS ConicalConnection implant from MIS Implant System.table below:Implant manufacturer : MIS IMPLANT TECHNOLOGIES LTD
ATLANTIS® products are compatible with the implants shownin the table below:
Implant manufacturer : MIS IMPLANT TECHNOLOGIES LTD
Trade NameAbutment PlatformDiameterImplant DiameterTrade NameAbutment PlatformDiameterImplant Diameter
Atlantis Abutment for MISV3 NPØ2.765mmV3: Ø3.30mmMIS Implant M4 &SEVEN NarrowPlatformØ3.30 mmØ3.30 mm
Atlantis Abutment for MISC1 NPØ2.76mmC1: Ø3.30mm
Atlantis Abutment for MISC1 & V3 SPØ3.16mmC1: Ø3.75, 4.2 mm,V3: Ø3.90, 4.3, 5.0 mmMIS Implant M4 &SEVEN StandardPlatformØ3.75 and4.2 mmØ3.75 and 4.2 mm
Atlantis Abutment for MISC1 WPØ4.01mmC1: Ø5.0 mmMIS Implant M4 &SEVEN WidePlatformØ5.0 and6.0 mmØ5.0 and 6.0 mm

{9}------------------------------------------------

Table 5.3 – Substantial Equivalence Comparison Summary (ATLANTIS Abutments, ATLANTIS Conus Abutments, ATLANTISCrown Abutments (continued).
ATLANTIS® Abutment for MIS ConicalConnection ImplantProposed deviceATLANTIS® Abutment for MIS ImplantK172225Predicate Device
ProsthesisattachmentScrew-retainedCement-retainedFriction FitScrew-retainedCement-retainedFriction Fit
RestorationSingle or Multi-unitSingle or Multi-unit
AbutmentplatformdiameterØ2.765, 2.76, 3.16, 4.01 mmØ 3.3, 3.75, 4.2, 5.0, 6.0 mm
AbutmentAngleStraight, up to 30°Straight, up to 30°
Max.AbutmentHeight15 mm15 mm
Min.AbutmentPost Height4 mm4 mm
DesigntypePatient specific designPatient specific design
ImplantConnectionConical ConnectionInternal hex connection
Material:AbutmentTitanium alloy, Gold Shaded Titanium (Gold Hue)[conforming to ASTM F136]- Titanium alloy, Gold Shaded Titanium (Gold Hue),[conforming to ASTM F136]- Zirconia
Material:ScrewTitanium alloy[conforming to ASTM F136]Titanium alloy[conforming to ASTM F136]

{10}------------------------------------------------

Table 5.4-Difference and Similarities of proposed, predicate and reference devices (Healing Abutments)
Trade NameATLANTIS® Abutment for MIS ConicalConnection Implant Healing AbutmentProposed DeviceV3 Conical Connection Healing CapsK163349Reference Device
Material(s)Titanium alloy, Gold Shaded Titanium (Gold Hue)Titanium 6Al-4V ELI per ASTM F136
Interface DiameterØ2.76, 2.765, 3.16, 4.01 mmCover screws: 2.77,3.15 mmHealing caps: 3.3,3.9,4.0,4.8,5.8 mm
Gingival height -healing caps0.5mm to 4mm (above the margin)2,3,4,5,6,8
Surface Treatment/N/APolished and anodized after machined
Connection TypeConical ConnectionConical connectionwithout indexes
PlatformNP/SP/WPNP/SP
Sterilization MethodSteam sterilized by end userRadiation

7. Non-Clinical Performance Data

Non-clinical testing data submitted, referenced or relied upon to demonstrate substantial equivalence includes:

  • Fatigue testing was conducted according to ISO 14801: Dental-implants Dynamic Fatigue Test for . Endosseous Dental Implants. Fatigue test results were compared to the predicate and reference devices.
  • Geometric compatibility analysis was conducted on OEM implant bodies, OEM abutments, and ● OEM screws to support the dimensional compatibility of the ATLANTIS® Abutment for MIS Conical Connection Implant with the MIS V3 implants (Narrow: Ø3.3mm and Standard: Ø3.9, 4.3, 5.0mm) and MIS C1 implants (Narrow Ø3.3mm, Standard: Ø3.75, 4.2mm and Wide: Ø5.0, 6.0 mm).
  • Sterilization validation of the proposed device is referenced by equivalency to the sterilization validation of predicate (K172225). Predicate device was validated according to ISO 17665-1: Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
  • The material composition and manufacturing processing methods of the proposed ATLANTIS® Abutment for MIS Conical Connection Implant are identical to the predicate device, ATLANTIS® Abutment for MIS Implant (K172225). Therefore, no additional biocompatibility data is included to support substantial equivalence.

{11}------------------------------------------------

Clinical Performance Data 8.

No human clinical data was included in this premarket notification to support substantial equivalence of the proposed modification to predicate device ATLANTIS® Abutment for MIS Implant (K172225).

9 Conclusion Regarding Substantial Equivalence

The proposed ATLANTIS® Abutments for MIS Conical Connection Implant are endosseous dental implant abutments which are intended to support a prosthetic device in partially or completely edentulous patients. The proposed ATLANTIS® Abutment for MIS Conical Connection Implant has the same intended use, incorporates the same fundamental technology, and has similar indications for use as the predicate device ATLANTIS® Abutment for MIS Implant (K172225). The proposed device fatigue testing results are substantially equivalent to those of the ATLANTIS® Abutment for HIOSSEN ET Implant (K160626) and therefore this predicate was included as a reference device. The proposed healing abutments are also technologically substantially equivalent to those of the V3 Conical Connection Healing Caps (K163349) and therefore this predicate was also included as a reference device.

In addition, the proposed ATLANTIS® Abutments are compatible with the following implant systems: MIS C1 Narrow Platform Conical Connection Implant System (K172505), MIS C1 Standard and Wide Platform Conical Connection Abutment (K112162), MIS V3 Conical Connection Dental Implant System (K163349). Non-clinical bench testing has been conducted and are included in this premarket notification to demonstrate the performance of the proposed ATLANTIS® Abutments for MIS Conical Connection Implant in support of substantial equivalence. The comparison of the indications for use, technological characteristics, with the inclusion of the results of nonclinical testing, support a conclusion of substantial equivalence of the proposed ATLANTIS® Abutments for MIS Conical Connection Implant to the predicate devices.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)