    Intended Use

    MSFX MIKRON SPINAL FIXATION SYSTEM is intended for use in the noncervical spine. When used as a posterior, noncervical pedicle and non- pedicle fixation system, the MSFX MIKRON SPINAL FIXATION SYSTEM is intended to provide additional support during fusion using autograft in the treatment of the treatment of the following acute and chronic instabilities or deformities:

    -Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).

    -Spondylolisthesis

    -Trauma (i.e. fracture or dislocation)

    -Spinal stenosis

    • -Curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
      -Tumor

    -Pseudoarthrosis; and failed previous fusion

    Device Description

    Msfx Mikron System is a top-loading, multiple component, posterior spinal fixation system consisting of polyaxial pedicle screws, monoaxial pedicle screws, cannulated and spondylolisthesis screws, rods (Straight and curved) , connectors , hooks and setscrews. It allows surgeons and functions to build a spinal implant construction to stabilize and promote spinal fusion. The System components are supplied non-sterile, for single use and fabricated from titanium alloy (Ti6Al4V-EL) that conforms to ASTM F136. Various sizes of these components are available.

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for a spinal fixation system (MSFX MIKRON SPINAL FIXATION SYSTEM). It does not describe the acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML medical device.

    The document primarily focuses on the regulatory aspects of a traditional medical device (spinal implants), including:

    • Device Name: MSFX Mikron Spinal Fixation System
    • Regulation Number/Name: 21 CFR 888.3070, Thoracolumbosacral Pedicle Screw System
    • Regulatory Class: Class II
    • Product Code: NKB, KWP
    • Indications for Use: Provides support during fusion for various spinal conditions (degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, failed previous fusion).
    • Predicate Devices: K171497 (Mikron Spinal Fixation system), K212220 (Artfx Spinal Fixation System), K091445 (CD HORIZON SPINAL SYSTEM), K071373 (XIA 3 Spinal System).
    • Device Description: Multi-component, posterior spinal fixation system made of titanium alloy.
    • Testing: Biomechanical tests performed according to ASTM F2193 and ASTM F1717, demonstrating substantial equivalence to the predicate device.

    Therefore, I cannot provide the requested information about acceptance criteria for an AI/ML device, its performance, study details (sample size, data provenance, expert ground truth, adjudication), or MRMC/standalone studies, as this document is not about an AI/ML medical device.

    To answer your request, I would need a document pertaining to an AI/ML medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180220

    Validate with FDA (Live)

    Date Cleared
    2018-05-21

    (116 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K113666, K071373, K091445, K994121, K111492

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SAGICO OSI SPINAL SYSTEM by Osimplant is intended for use in the non-cervical spine.

    When used for anterior/anterolateral and posterior, non-cervical pedicle fixation system, the SAGICO OSI SPINAL SYSTEM by Osimplant is intended to provide additional support during fusion using autograft or allograft in Skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:

    • · Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
    • · Spondylolisthesis;
    • · Trauma (i.e. fracture or dislocation);
    • · Spinal stenosis;
    • · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis);
    • · Tumor:
    • · Pseudoarthrosis; and
    • Failed previous fusion

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the SAGICO OSI SPINAL SYSTEM by Osimplant implants are indicated as an adjunct to treat adolescent idiopathic scoliosis.

    The SAGICO OSI SPINAL SYSTEM by Osimplant for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The SAGICO OSI SPINAL SYSTEM by Osimplant constructs of monoaxial screws, uniplanar screws, polyaxial screws, reduction screws, locking cap set screws, rods, hooks, monoaxial and muttiaxial transverse connectors and associated surgical instruments. The SACICO OSI SPINAL SYSTEM by Osimplants are available in a variety of sizes to accommodate individual patient anatomy and pathology conditions. SAGICO OSI SPINAL SYSTEM by Osimplant implants are designed to adapt 5.5mm diameter rods; the implants are manufactured from Ti6A14V alloy and offered in a sterile package option.

    AI/ML Overview

    The provided text is a 510(k) summary for the SAGICO OSI SPINAL SYSTEM by Osimplant, a medical device for spinal fixation. It details the device's indications for use, technological characteristics, and performance data used to establish substantial equivalence to predicate devices. However, this document does not describe an AI/ML device or algorithm. It is a clearance for a traditional physical medical implant system consisting of screws, rods, and associated instruments.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, or comparative effectiveness studies for an AI device cannot be extracted from this document, as it pertains to a mechanical spinal implant system.

    The information below is a general template for response if the input data would contain information for an AI/ML powered device, as the prompt is not expecting the type of the device to be considered.

    This document does not describe an AI/ML powered medical device. It describes a traditional, mechanical spinal implant system (SAGICO OSI SPINAL SYSTEM by Osimplant). Therefore, the specific details requested regarding AI/ML device performance, acceptance criteria, sample sizes for AI/ML models, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance are not available in this document.

    The document focuses on demonstrating substantial equivalence of the mechanical implant system to predicate devices based on:

    • Indications for Use: The SAGICO OSI SPINAL SYSTEM is intended for use in the non-cervical spine for various conditions like degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients. For pediatric patients, it's indicated for adolescent idiopathic scoliosis (posterior approach only).
    • Technological Characteristics: The device consists of monoaxial screws, uniplanar screws, polyaxial screws, reduction screws, locking cap setscrews, rods, hooks, monoaxial and multiaxial transverse connectors, and associated surgical instruments. Implants are made from Ti6Al4V alloy and designed for 5.5mm diameter rods.
    • Performance Data (Non-clinical):
      • ASTM F1717-04: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model (Static Compression Bending Test, Dynamic Compression Bending Test).
      • ASTM F1798-13: Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants (Axial Gripping Test, Axial Torsion Gripping Test, Flexion Extension Static Test).
    • Basis of Substantial Equivalence: Comparable data drawn from design, labeling, indications for use, function, and implant materials compared to predicate devices (XIA® 3 Spinal System, USS Small Stature System, CD HORIZON Spinal System, TSRH Spinal System).

    If this were an AI/ML device, the requested information would typically be presented as follows (this is a hypothetical example for illustrative purposes, not based on the provided document):

    1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical for an AI/ML device)

    Performance MetricAcceptance Criteria (e.g., for detecting spinal fractures)Reported Device Performance (e.g., from a clinical study)
    Sensitivity≥ 90%92.5% (95% CI: 90.1% - 94.4%)
    Specificity≥ 80%85.3% (95% CI: 83.0% - 87.3%)
    AUC≥ 0.900.93

    2. Sample Size and Data Provenance for Test Set (Hypothetical)

    • Sample Size: 500 spinal imaging studies (e.g., X-rays, CT scans)
    • Data Provenance: Retrospective data collected from 3 hospitals in the United States and 2 hospitals in Germany.

    3. Number and Qualifications of Experts for Ground Truth (Hypothetical)

    • Number of Experts: 3 independent radiologists
    • Qualifications: All board-certified radiologists with subspecialty training in musculoskeletal radiology and at least 10 years of experience interpreting spinal imaging.

    4. Adjudication Method for Test Set (Hypothetical)

    • Adjudication Method: 2+1. Initial assessment by two radiologists. Discrepancies were resolved by a third, senior radiologist.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Hypothetical)

    • Was an MRMC study done? Yes.
    • Effect Size: In the MRMC study, human readers (radiologists) demonstrated a statistically significant improvement in fracture detection accuracy (AUC increased by an average of 0.05, p < 0.001) when assisted by the AI device compared to reading without AI assistance. This translated to a 15% reduction in missed fractures among the assisted group.

    6. Standalone Algorithm Performance (Hypothetical)

    • Was a standalone study done? Yes. The performance metrics in the table above reflect the standalone performance of the algorithm on the test set.

    7. Type of Ground Truth Used for Test Set (Hypothetical)

    • Type of Ground Truth: Expert consensus (as established by the 2+1 adjudication method based on expert review of the imaging studies). In cases where available, pathology reports and follow-up clinical outcomes were used to corroborate expert diagnoses.

    8. Sample Size for Training Set (Hypothetical)

    • Sample Size: 5,000 spinal imaging studies (e.g., X-rays, CT scans).

    9. How Ground Truth for Training Set was Established (Hypothetical)

    • Ground Truth Establishment: Annotated by a team of trained radiology residents and fellows under the supervision of experienced attending radiologists. A subset of the annotations (10%) was independently reviewed by a senior radiologist for quality control, and discrepancies were resolved through discussion.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143683

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-05-22

    (149 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APEX-DL Spine System with APEX Spine System Components is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).

    The APEX-DL Spine System with APEX Spine System Components is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.

    The APEX-DL Spine System with APEX Spine System Components is also a sacraliliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).

    When used in a percutaneous posterior approach with AIM MIS instrumentation, the APEX-DL Spine System with APEX Spine System Components is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.

    When used for posterior non-cervical pediatic patients, the APEX-DL Spine System implants with APEX Spine System Components are indicated as an adjunct to treat adolescent idiopathic scoliosis. The APEX DL Spine System is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The APEX-DL Spine System includes Monoaxial, Uniplanar, and Polyaxial Double Lead Thread Screws in the cannulated and non-cannulated versions and in regular and reduction (extended tab) versions. APEX-DL Spine System is a low profile thoracolumbar implant for use with wide range of patient statures. The APEX-DL Spine System Polyaxial screws feature a friction head, which is designed to provide precise reduction mechanism as a result of easier rod capturing. The APEX-DL Spine System also includes Lordosed Percutaneous Rods. The APEX-DL Spine System is compatible with the APEX Spine System 5.5mm and 6.0mm rods, hooks, side-by-side connectors, iliac connectors, cross connectors, and washers.

    AI/ML Overview

    The APEX-DL Spine System is a spinal implant for immobilization and stabilization of spinal segments. The document indicates that no clinical studies were performed. The acceptance criteria and testing are based on non-clinical (mechanical) tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance CriteriaReported Device Performance
    ASTM F1717 (Vertebrectomy Model)
    Static Compression BendingNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Static TorsionNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Dynamic Compression BendingNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    ASTM F1798 (Interconnection Mechanisms)
    Static Axial Gripping CapacityNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Axial Torque Gripping CapacityNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Static Flexion-ExtensionNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Dynamic Flexion-ExtensionNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"

    Note: The acceptance criteria are implicitly defined as demonstrating mechanical properties that are "equal or higher" than the identified predicate devices, based on the testing standards.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in the provided document. For each of the ASTM tests listed, a specific number of samples (implants/constructs) would have been used, but this detail is not provided.
    • Data Provenance: The data is from non-clinical (mechanical) testing conducted presumably by SpineCraft, LLC or a contracted testing facility. It is not patient or human data; therefore, country of origin or retrospective/prospective does not apply in the typical sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a non-clinical, mechanical testing study, not a study requiring expert clinical assessment or ground truth establishment in a medical imaging or diagnostic context. The "ground truth" here is the objective measurement of mechanical properties according to established ASTM standards.

    4. Adjudication Method for the Test Set

    Not applicable. This is a non-clinical, mechanical testing study. Adjudication methods like 2+1 or 3+1 are used for expert consensus in clinical or imaging studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is a clinical study involving multiple human readers interpreting medical cases. The provided document explicitly states, "No clinical studies were performed."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical spinal implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on objective mechanical measurements according to recognized industry standards (ASTM F1717 and ASTM F1798). The performance is then compared to "predicate systems" as a benchmark for substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device undergoing mechanical testing, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As described in point 8, there is no "training set" for this type of device evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150417

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-05-19

    (90 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ASTRA System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudo-arthrosis).

    The ASTRA Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.

    The ASTRA System is also a sacral/iliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).

    When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis), tumor, pseudo-arthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.

    Device Description

    The ASTRA Spine System is a top loading, multiple component, posterior spinal fixation system which consists of rods, cannulated and non-cannulated monoaxial, uniplanar and polyaxial screws, hooks, iliac connectors, rod connectors, and cross connectors. Most of the components are available in a variety of sizes to more closely match the patient's anatomy.

    Materials:
    Titanium alloy
    CoCr alloy

    AI/ML Overview

    Here's an analysis of the provided text regarding the ASTRA Spine System, focusing on acceptance criteria and study details.

    Important Note: The provided document is a 510(k) summary for a medical device (Spine System), not an AI/ML device. Therefore, many of the requested fields related to AI/ML specific studies (like sample sizes for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance) are not applicable to this type of medical device submission. This document focuses on demonstrating substantial equivalence to predicate devices through non-clinical (mechanical) testing.


    1. Table of Acceptance Criteria and Reported Device Performance

    For this medical device, the "acceptance criteria" are implicitly defined by performance equivalence to predicate devices in standardized mechanical tests. The "reported device performance" is a statement of comparative equivalence.

    Acceptance Criteria (Implicit)Reported Device Performance (Summary)
    Static Compression Bending: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Static Torsion: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Dynamic Compression Bending: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Axial Gripping: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Torsional Gripping: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Static Flexion-Extension: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Dynamic Flexion-Extension: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Overall Substantial Equivalence: Must be substantially equivalent in intended use, design, material, performance, and function."The ASTRA Spine System is substantially equivalent to the predicate devices in terms of indications for use, design, material, performance and function."

    2. Sample Size Used for the Test Set and the Data Provenance

    This is a physical device, and the "test set" refers to the tested device components.

    • Sample Size: Not explicitly stated in terms of number of components tested for each test, but standard engineering practices for medical device testing would involve a sufficient number (e.g., n=5 or n=10 per test) to ensure statistical significance, though the exact numbers are not provided in this summary.
    • Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting. There is no patient data provenance (e.g., country of origin, retrospective/prospective) since no human data was used for these mechanical tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not Applicable. For mechanical testing of a physical device, ground truth is established by standardized testing protocols (e.g., ASTM standards) and measured physical properties, not by expert consensus on clinical data. Engineers and lab technicians perform and analyze the tests according to these standards.

    4. Adjudication Method for the Test Set

    • Not Applicable. There is no adjudication in the sense of reconciling human expert opinions for clinical images or data. The results of mechanical tests are objective measurements against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • Not Applicable. This is a physical spinal implant, not an AI/ML diagnostic or assistive device. No MRMC studies were performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This is a physical spinal implant; there is no algorithm or standalone performance.

    7. The Type of Ground Truth Used

    • Mechanical Test Standards and Predicate Device Performance: The "ground truth" for the non-clinical tests is based on the established performance characteristics and safety profiles of legally marketed predicate devices, as defined by FDA regulations for substantial equivalence, and adherence to relevant ASTM standards (ASTM F1717 and ASTM F1798).

    8. The Sample Size for the Training Set

    • Not Applicable. This is a physical device; there is no AI/ML model or "training set" in the context of machine learning. The device design and materials are based on engineering principles and knowledge of predicate devices.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set for an AI/ML model, this question is not relevant. The "ground truth" for developing the physical device is based on established biomechanical and medical understanding of spinal fixation and the performance of existing, cleared devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140678

    Validate with FDA (Live)

    Date Cleared
    2014-06-19

    (93 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The CosmoLock Pedicle Screw System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials:

    Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

    AI/ML Overview

    This document is a 510(k) Summary for the CosmoLock Pedicle Screw System. It describes the device and its intended use and claims substantial equivalence to predicate devices. However, it does not contain the detailed information needed to answer many of your questions, as it explicitly states that no clinical studies were performed.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test)Reported Device Performance
    Static compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Dynamic compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Static torsion per ASTM F1717Equivalent to predicate devices (implied: met criteria)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The studies mentioned are non-clinical (mechanical tests), and details about sample sizes for these tests are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. No clinical studies were performed, and therefore no ground truth established by experts is relevant to the data presented.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No clinical studies were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This document states, "No clinical studies were performed."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a mechanical pedicle screw system, not an AI or algorithm-based device. No standalone performance study in this context was performed.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth for the non-clinical tests would be the established mechanical testing standards (ASTM F1717) and the performance of the predicate devices. There is no biological or expert-defined "ground truth" in the typical medical imaging/diagnosis sense, as this is a mechanical implant.

    8. The sample size for the training set

    This information is not applicable/provided. No clinical studies were performed, and there is no mention of an algorithm or AI requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. No training set or associated ground truth was established, as no clinical studies or AI development are described. The focus is on demonstrating mechanical equivalence to predicate devices.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthrosis; and Failed Previous Fusion.

    The Radius® Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation system, the Radius® Spinal system is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Radius® Spinal System can also be linked to the XIA® Titanium Spinal System via the Ø5.5mm to Ø6.0mm Radius® rod-to-rod connector.

    The Stryker Spine TRIO® Plate System is intended for posterior, noncervical (T10-S1) pedical and nonpedical fixation of the spine for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.

    The Stryker Spine TRIO® Spinal Fixation System is intended for posterior, noncervical pedicle and non-pedicle fixation of the spine. The Stryker Spine TRIO® Spinal Fixation System is indicated for: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® Spinal Fixation Sytem is intended to be used in conjunction with the OSS Diapason Rods, Opus Spinal System Rods, and the Multi-Axis Cross Connectors.

    The Stryker Spine TRIO® Spinal System is intended for posterior, noncervical pedicle and nonpedicle fixation of the spine to provide immobilization and stabilization of spinal segements in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® + Spinal System is intended to be used in conjunction with the OSS/Diapason Rods or Opus Rods, XIA® Pre-bent Rods, and the Multi-Axis Cross Connectors.

    The Stryker Spine TRIO® TRAUMA Spinal System is intended for percutaneous, posterior, non-cervical pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.

    The XIA® Spinal System and XIA® 4.5 Spinal System are intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DDD) (Defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e., fracture or dislocation); Spinal Stenosis; Curvatures (i.e., Scoliosis, Kyphosis, and/or Lordosis); Tumor; Pseudoarthrosis and; Failed previous fusion. The 6mm diameter rods from the DIAPASON® Spinal System and OPUS® Spinal System are intended to be used with the other components of the XIA® Titanium Spinal System. The Titanium Multi-Axial Cross Connector are intended to be used with the other components of the XIA® Titanium Spinal System.

    The XIA® 3 Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle fixation system, the XIA® 3 Spinal System is intended to provide additional support during autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor, Pseudoarthorisis; and Failed Pervious Fusion. The Ø5.5mm rods from the Stryker Spine Radius® Spinal System and the Ø6.0mm Vitallium Rods from the XIA® Spinal System are intended to be used with the other components of the XIA® 3 Spinal System. When used for posterior noncervical pedicle screw fixation in pediatric patients, the XIA® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. The XIA® 3 Spinal System for pediatic use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The XIA® 4.5 Spinal System is intended for anterior/anteriolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. Except for the staples, the XIA® 4.5 Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical pedicle screw fixation in pediatric patients. The XIA® 4.5 Spinal System for pediatric use in intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The STRYKER Spine thoraco-lumbar spinal fixation systems, subject of this 510(k), are non-cervical, pedicle and non-pedicle fixation systems comprised of screws, rods, plates, hooks, connectors, washers and staples. The components are manufactured from either Titanium (Titanium Alloy and CP Titanium), Stainless Steel or Cobalt-Chromium-Molybdenum Alloy (Vitallium®). This Special 510(k) submission seeks clearance for sterile labeling of the listed STRYKER Spine thoraco-lumbar spinal fixation systems. All of the components of the subject devices will be sterilized by gamma radiation, a traditional sterilization method as per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.

    AI/ML Overview

    This is a 510(k) Premarket Notification for sterile packaging of a range of spinal systems. The key claim is substantial equivalence to previously cleared non-sterile versions of the same devices. Therefore, the "device" in question for this submission is specifically the sterile packaging and the sterilization process, not the spinal implant itself as a medical device performing a physiological function. The "performance" being evaluated is the ability to maintain sterility and the structural integrity of the device within its packaging over its shelf life.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Sterilization EffectivenessDemonstrated by gamma radiation, a traditional sterilization method per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.
    Sterile Barrier Packaging SystemDemonstrated effective in maintaining sterility for the recommended 5-year shelf-life.
    Impact on Device CharacteristicsNo design modifications were incorporated to facilitate sterile packaging. The sterile-packed implant components maintain the same technological characteristics (design, technical requirements, materials of construction, indications/intended use) as the non-sterile predicate devices.
    Shelf LifeDemonstrated a 5-year shelf life in sterile packaging.
    Safety and EffectivenessConcluded to be safe and effective, retaining the same intended use, indications, technological characteristics, and mode of operation as the predicate non-sterile devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses the sterilization process and packaging rather than clinical performance of the implant itself. Therefore, typical "test set" in the context of clinical or diagnostic AI/software devices (e.g., patient data) is not applicable here.

    • Sterilization and Packaging Validation Data: The document mentions "accelerated aging data" for validating the 5-year shelf life. It does not specify the sample size for this testing, nor the country of origin or whether it was retrospective or prospective. However, these types of tests are typically prospective laboratory studies conducted in controlled environments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This submission is about the sterility and packaging of medical implants, not a diagnostic or clinical performance study that would require expert ground truth labeling. The "ground truth" here is the scientific validation of sterilization methods and shelf life according to recognized standards and FDA guidance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used in clinical studies for establishing ground truth, which is not relevant to this submission on sterile packaging.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this submission relies on:

    • Scientific Validation of Sterilization: Adherence to established scientific principles and FDA guidance for sterilization (gamma radiation), demonstrating a Sterility Assurance Level (SAL).
    • Accelerated Aging Data: Laboratory testing to simulate real-time aging and demonstrate that the sterile barrier packaging system maintains sterility for the stated shelf life.
    • Material and Design Specification Conformity: Confirmation that the components are manufactured from specified materials and that their design, technical requirements, and intended use are identical to the non-sterile predicate devices.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not a machine learning or AI device, there is no training set and therefore no ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121568

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-21

    (115 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dali Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all of the following indications regardless of the intended use:

    • degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);
    • spondylolisthesis;
    • trauma (i.e. fracture or dislocation);
    • deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis);
    • tumor;
    • stenosis, and;
    • failed previous fusion (pseudoarthrosis)
    Device Description

    This system is comprised of screws, set screws, rods, crosslink and connectors. The components of this system are manufactured of Titanium alloy (Titanium-6Aluminum-4Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537). The screws are available from 4.0 to 8.5mm diameters with lengths ranging from 20-150mm.

    AI/ML Overview

    The provided text describes the "Dali Spinal Fixation System," a Class III medical device for spinal fixation. Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Mechanical PerformanceBench testing results demonstrate that Dali Spinal Fixation System performs equivalently to the predicates in: - Static compression bending - Static torsion - Dynamic compression bending (in accordance with ASTM F1717-10) - Gripping-push down (in accordance with ASTM F1798)
    Material EquivalenceThe components of this system are manufactured of Titanium alloy (Titanium-6Aluminum-4Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537), which are the same materials used in predicate devices.
    Design Equivalence (substantially similar)The Dali Spinal Fixation System shares technological characteristics similar to the predicate devices, including similar design.
    Intended Use / Indications for UseThe device's intended use and indications for use as a non-cervical spinal fixation device (posterior pedicle fixation T1-S2/ilium, or anterolateral fixation T8-L5) for skeletally mature patients as an adjunct to fusion for conditions like degenerative disc disease, spondylolisthesis, trauma, deformities, tumor, stenosis, and failed previous fusion are deemed substantially equivalent to legally marketed predicate devices.
    Overall Safety and EffectivenessThe Dali Spinal Fixation System is substantially equivalent to the device referenced above and is therefore safe and effective for its intended use.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical data and conclusions were not needed for this device." This means there was no clinical test set involving human subjects. The evaluation was based on non-clinical (bench) testing. Therefore, there is no sample size for an clinical test set or data provenance (country of origin, retrospective/prospective clinical data).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no clinical data was required or used, there was no need for experts to establish ground truth for a clinical test set. The acceptance was based on engineering and material equivalence to predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No clinical test set with human data requiring adjudication was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a spinal fixation system (physical implant), not an AI-assisted diagnostic or imaging device. Therefore, an MRMC study related to human reader performance with or without AI assistance is irrelevant and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical medical implant, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance evaluation, the "ground truth" was established by standardized mechanical testing protocols (e.g., ASTM F1717-10 for dynamic compression bending, and ASTM F1798 for gripping-push down) and material specifications (ASTM F136 for Titanium alloy, ASTM F1537 for CoCrMo alloy). The "truth" was that the device met or exceeded the performance of the predicate devices under these controlled, quantifiable conditions.

    8. The sample size for the training set

    Not applicable. This device is a physical implant evaluated through bench testing and comparison to predicate devices, not an AI or machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there was no training set.