(149 days)
The APEX-DL Spine System with APEX Spine System Components is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX-DL Spine System with APEX Spine System Components is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.
The APEX-DL Spine System with APEX Spine System Components is also a sacraliliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous posterior approach with AIM MIS instrumentation, the APEX-DL Spine System with APEX Spine System Components is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
When used for posterior non-cervical pediatic patients, the APEX-DL Spine System implants with APEX Spine System Components are indicated as an adjunct to treat adolescent idiopathic scoliosis. The APEX DL Spine System is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The APEX-DL Spine System includes Monoaxial, Uniplanar, and Polyaxial Double Lead Thread Screws in the cannulated and non-cannulated versions and in regular and reduction (extended tab) versions. APEX-DL Spine System is a low profile thoracolumbar implant for use with wide range of patient statures. The APEX-DL Spine System Polyaxial screws feature a friction head, which is designed to provide precise reduction mechanism as a result of easier rod capturing. The APEX-DL Spine System also includes Lordosed Percutaneous Rods. The APEX-DL Spine System is compatible with the APEX Spine System 5.5mm and 6.0mm rods, hooks, side-by-side connectors, iliac connectors, cross connectors, and washers.
The APEX-DL Spine System is a spinal implant for immobilization and stabilization of spinal segments. The document indicates that no clinical studies were performed. The acceptance criteria and testing are based on non-clinical (mechanical) tests.
1. Table of Acceptance Criteria and Reported Device Performance
| Test Type | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| ASTM F1717 (Vertebrectomy Model) | ||
| Static Compression Bending | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
| Static Torsion | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
| Dynamic Compression Bending | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
| ASTM F1798 (Interconnection Mechanisms) | ||
| Static Axial Gripping Capacity | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
| Axial Torque Gripping Capacity | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
| Static Flexion-Extension | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
| Dynamic Flexion-Extension | Not explicitly stated (comparison to predicates) | Results were "equal or higher than the predicate systems" |
Note: The acceptance criteria are implicitly defined as demonstrating mechanical properties that are "equal or higher" than the identified predicate devices, based on the testing standards.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated in the provided document. For each of the ASTM tests listed, a specific number of samples (implants/constructs) would have been used, but this detail is not provided.
- Data Provenance: The data is from non-clinical (mechanical) testing conducted presumably by SpineCraft, LLC or a contracted testing facility. It is not patient or human data; therefore, country of origin or retrospective/prospective does not apply in the typical sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is a non-clinical, mechanical testing study, not a study requiring expert clinical assessment or ground truth establishment in a medical imaging or diagnostic context. The "ground truth" here is the objective measurement of mechanical properties according to established ASTM standards.
4. Adjudication Method for the Test Set
Not applicable. This is a non-clinical, mechanical testing study. Adjudication methods like 2+1 or 3+1 are used for expert consensus in clinical or imaging studies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. An MRMC study is a clinical study involving multiple human readers interpreting medical cases. The provided document explicitly states, "No clinical studies were performed."
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This device is a physical spinal implant, not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is based on objective mechanical measurements according to recognized industry standards (ASTM F1717 and ASTM F1798). The performance is then compared to "predicate systems" as a benchmark for substantial equivalence.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device undergoing mechanical testing, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As described in point 8, there is no "training set" for this type of device evaluation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of human faces in profile, arranged in a cascading manner. The faces are rendered in a simple, line-art style, with the lines forming the contours of the faces. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 22, 2015
SpineCraft, LLC Ms. Ami Akallal-Asaad Vice President, Regulatory Affairs & Quality Assurance 777 Oakmont Lane Westmont, Illinois 60559
Re: K143683
Trade/Device Name: APEX-DL Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI, KWP Dated: April 20, 2015 Received: April 24, 2015
Dear Ms. Akallal-Asaad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Ami Akallal-Asaad
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K143683
Device Name
APEX-DL SPINE SYSTEM
Indications for Use (Describe)
The APEX-DL Spine System with APEX Spine System Components is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX-DL Spine System with APEX Spine System Components is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.
The APEX-DL Spine System with APEX Spine System Components is also a sacraliliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous posterior approach with AIM MIS instrumentation, the APEX-DL Spine System with APEX Spine System Components is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
When used for posterior non-cervical pediatic patients, the APEX-DL Spine System implants with APEX Spine System Components are indicated as an adjunct to treat adolescent idiopathic scoliosis. The APEX DL Spine System is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
|X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
510(k) Summary
In accordance with 21 CFR 807.92 of the Federal Code of Requlations the following 510(k) Summary is submitted for the APEX-DL Spine System.
Date Prepared: April 20, 2015
| 1. Submitter:SpineCraft, LLC777 Oakmont LaneWestmont, IL 60559 USATel: (630) 920-7300Fax: (630) 920-7310 | Contact Person:Ami Akallal-AsaadVP, Regulatory Affairs & QASpineCraft, LLCa.asaad@spinecraft.com |
|---|---|
| 2. Trade name: | APEX-DL Spine System |
| Common Name: | Pedicle Screw Spine System |
| Classification Product Code: | Pedicle screw spinal system per NKB 888.3070Class III |
| Subsequent Product Codes: | Pedicle screw spinal system, adolescent idiopathic scoliosis peOSH 888.3070Pedicle screw spinal system per MNI 888.3070Pedicle screw spinal system per MNH 888.3070Spinal interlaminal fixation orthosis per KWP 888.3050 |
3. Primary predicate or legally marketed device which is substantially equivalent:
- EXPEDIUM Verse Spine System (K142185) DePuy Synthes Spine o
Secondary predicate devices: 4.
- Viper Spine System (K090648 / K102701) DePuy Spine
- o APEX Spine System (K062513 / K092825 / K102488 / K110906 / K132603) SpineCraft
- Xia Spinal System (K113666 / K071373 / K060748) Stryker ●
- SPINE-TECH SILHOUETTE SPINAL FIXATION SYSTEM (K980288) SPINE-TECH, INC ●
- MOSS MIAMI SS (K950697 / K000536) DePuy Spine o
- o Synergy VLS - open - (K940631 / K950099) Cross Medical
- o Ti Expedium 4.5 Spine System (K081252) DePuy Spine
Description of the device: 5.
The APEX-DL Spine System includes Monoaxial, Uniplanar, and Polyaxial Double Lead Thread Screws in the cannulated and non-cannulated versions and in regular and reduction (extended tab) versions. APEX-DL Spine System is a low profile thoracolumbar implant for use with wide range of patient statures. The APEX-DL Spine System Polyaxial screws feature a friction head, which is designed to provide precise reduction mechanism as a result of easier rod capturing. The APEX-DL Spine System also includes Lordosed Percutaneous Rods. The APEX-DL Spine System is compatible with the APEX Spine System 5.5mm and 6.0mm rods, hooks, side-by-side connectors, iliac connectors, cross connectors, and washers.
-
- Material:
{4}------------------------------------------------
Titanium alloy (Ti-6Al-4V ELI) per ASTM F136 Cobalt chromium alloy (Co-28Cr-6Mo) per ASTM F1537
7. Substantial equivalence claimed to predicate devices:
APEX-DL Spine System is substantially equivalent to the predicate devices identified above in terms of intended use, design (fundamental scientific technology), materials, mechanical safety and/or performance, sterility, and biocompatibility.
8. Intended Use:
The APEX-DL Spine System with APEX Spine System Components is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudoarthrosis).
The APEX-DL Spine System with APEX Spine System Components is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.
The APEX-DL Spine System with APEX Spine System Components is also a sacral/iliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).
When used in a percutaneous posterior approach with AIM MIS instrumentation, the APEX-DL Spine System with APEX Spine System Components is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the APEX-DL Spine System implants with APEX Spine System Components are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The APEX DL Spine System is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.
9. Non-clinical Test Summary:
The following tests were conducted:
- ASTM F1717, "Standard Test Methods for Spinal Implant Constructs in a ● Vertebrectomy Model". Testing included Static Compression Bending Tests, Static Torsion Tests and Dynamic Compression Bending Tests.
- ASTM F1798 "Standard Guide for Evaluating the Static and Fatigue Properties of . Interconnection Mechanisms and Subassemblies used in Spinal Arthrodesis Implants". Testing included, Static Axial Gripping Capacity, Axial Torque Gripping Capacity, Static Flexion-Extension, and Dynamic Flexion-Extension.
The results of this testing were compared to predicate systems, with the results being equal or higher than the predicate systems.
10. Clinical Test Summary
{5}------------------------------------------------
No clinical studies were performed
11. Conclusions Nonclinical and Clinical
The APEX-DL Spine System is substantially equivalent to the predicate devices in terms of indications for use, design, material, performance, and function.
N/A