K Number
K994121

Validate with FDA (Live)

Manufacturer
Date Cleared
2000-10-17

(316 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

When used as a posterior pedicle screw fixation system, the Synthes Small Stature USS Is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients (including small stature) as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the theracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scallosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition, the Synthes Small Stature USS is intended for treatment of severe spondylofisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients (including small stature) and pediatric patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2.

When used as a posterior non-pedicle screw fixation system in skeletally mature patients (including small stature) and pediatric patients, the Synthes Small Stature USS is intended for scollotic, lordotic, or kyphotic deformities (such as scoliosis, Schuermann's disease), degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), and fractures of the posterior thoracolumbar spine.

The anterior components of the Synthes Small Stature USS when used in skeletally mature patients (Including small stature) and pediatric patients are intended for anterclateral screw and/or staple fixation for the correction of anterolateral lordotic deformities for the spine, lumbar scoliosis, pseudoarthrosis, and fracture or dislocation of the thoracolumbar spine-(levels 78-L5),

In addition, when used with 3.5/5.0mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Universal Spinal System

Device Description

The Synthes Small Stature USS consists of a variety of rods, hooks, station is also, avgileble The Synthes Sthair Stature OOO Consisted of a variety of Feat, Normentation is also available to facilitate implantation of the device components.

Certain implant components from the Synthes Small Stature USS can be used to connect the se Geltain Implant components from the Synthes Synthes posterior spinal systems (USS or CerviFix).

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Synthes Small Stature USS), not a study report or clinical trial. Therefore, the document does not contain the requested information about acceptance criteria, device performance, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, training set details, or type of ground truth.

The document is a regulatory approval letter from the FDA, stating that the device is substantially equivalent to a predicate device already on the market. It outlines the device description, indications for use, and regulatory classification. It does not present data from any specific study proving the device meets acceptance criteria.

To answer your request, I would need a different type of document, such as a clinical study report, a scientific paper detailing performance evaluation, or a submission that includes detailed testing results.

{0}------------------------------------------------

K994121

OCT 1 7 2000

Synthes Small Stature USS Summary of Safety and Effectiveness 12/2/99

SYNTHES (U.S.A.) 1690 Russell Road Paoli, PA 19301 (610) 647-9700

Device

Synthes Small Stature USS

Product Description

Product Description
The Synthes Small Stature USS consists of a variety of rods, hooks, station is also, avgileble The Synthes Sthair Stature OOO Consisted of a variety of Feat, Normentation is also available to facilitate implantation of the device components.

Certain implant components from the Synthes Small Stature USS can be used to connect the se Geltain Implant components from the Synthes Synthes posterior spinal systems (USS or CerviFix).

Indications

When used as a posterior pedicle screw fixation system, the Synthes Small Stature USS is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients (including small stature) as an adjunct to fusion in the freatment of the following acute patients (including Small Stature) as an adjunct to factor in the sacral spine. degenerative and Chrome Instablities of delomition of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition, the Synthes Small Stature USS is intended for treatment of severe spondyclishesis In addition, the Synthes Onlar Octore Occasion in the patients (including small status) and (Grade 3 and 4) of the Lo-of Veltebran by autogenous bone graft having implants attached to the pealating patients receiving fusion by autogue to the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2.

When used as a posterior non-pedicle screw fixation system in skeletally mature patients (including small stature) and pediatic patients, the Synthes Small Stature USS intended for scoliotic, lordotic, or kyphotic deformities (such as scoliosis, Schuerman's disease), degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed degencrative und radiographic studies), and fractures of the posterior thoracolumbar spine.

The anterior components of the Synthes Small Stature USS when used in skeletally mature patients (including small stature) and pediatic patients are intended for antercoping lumber patientis (including onlial other of anterolateral lordotic deformities for the spine, humbar andror stable fixation for the ocerceden of the thoracolumbar spine (levels T8-L5).

In addition, when used with 3.5/5.0mm parallel connectors, the Synthes Small Stature USS can in addition, when used with b.org. only parallel connectors, the Synthes Small Stature USS can be linked to the Universal Spinal System

3

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle's head and neck, with three stylized lines representing the eagle's feathers.

OCT 1 7 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jonathan M. Gilbert Senior Regulatory Affairs Associate SYNTHES (USA) 1690 Russell Road Post Office Box 1766 Paoli, Pennsylvania 19301

Re: K994121

Trade Name: Small Statue USS Regulatory Class: II Product Code: MNI, MNH and KWP Dated: July 19, 2000 Received: July 20, 2000

!

Dear Mr. Gilbert:

We have reviewed your Section 510(k) notification of intent to market the device referenced we have have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the stated in the oneround) to ceet Device Amendments, or to devices that have been reclassified in enceditions with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general thereof of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remance ripproval), in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to nspotions, and 1 000 and may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{2}------------------------------------------------

Page 2 - Mr. Jonathan Gilbert

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at additionally 809.10 for m Yriro diagnostions on the promotion and advertising of your device, (301) 394-4659. Additionally, for questions on the presses note the regulation please contact the Office or Comphanee at (301) 0 - 1 (30 ) - 1 (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small information on your responsionnes ander and rise no rio not (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Mark M. Milken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications For Use Statement

Page 1 of 1

510(k) Number (if Known): NA K99412)

Device Name: Synthes Small Stature USS

Indications for Use:

When used as a posterior pedicle screw fixation system, the Synthes Small Stature USS Is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients (including small stature) as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the theracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scallosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition, the Synthes Small Stature USS is intended for treatment of severe spondylofisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients (including small stature) and pediatric patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2.

When used as a posterior non-pedicle screw fixation system in skeletally mature patients (including small stature) and pediatric patients, the Synthes Small Stature USS is intended for scollotic, lordotic, or kyphotic deformities (such as scoliosis, Schuermann's disease), degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), and fractures of the posterior thoracolumbar spine.

The anterior components of the Synthes Small Stature USS when used in skeletally mature patients (Including small stature) and pediatric patients are intended for anterclateral screw and/or staple fixation for the correction of anterolateral lordotic deformities for the spine, lumbar scoliosis, pseudoarthrosis, and fracture or dislocation of the thoracolumbar spine-(levels 78-L5),

In addition, when used with 3.5/5.0mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Universal Spinal System

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (OBE)

Prescription Use (Per 21 CFR 801.109) OR

Over-the-Counter Use

Mark n Milkman

(Division Sign-C (Division Sign oral Restorative 510(k) Number.

N/A