K Number
K140678

Validate with FDA (Live)

Date Cleared
2014-06-19

(93 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

Device Description

The CosmoLock Pedicle Screw System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

Materials:

Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

AI/ML Overview

This document is a 510(k) Summary for the CosmoLock Pedicle Screw System. It describes the device and its intended use and claims substantial equivalence to predicate devices. However, it does not contain the detailed information needed to answer many of your questions, as it explicitly states that no clinical studies were performed.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Test)Reported Device Performance
Static compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
Dynamic compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
Static torsion per ASTM F1717Equivalent to predicate devices (implied: met criteria)

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The studies mentioned are non-clinical (mechanical tests), and details about sample sizes for these tests are not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable/provided. No clinical studies were performed, and therefore no ground truth established by experts is relevant to the data presented.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/provided. No clinical studies were performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This document states, "No clinical studies were performed."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a mechanical pedicle screw system, not an AI or algorithm-based device. No standalone performance study in this context was performed.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The ground truth for the non-clinical tests would be the established mechanical testing standards (ASTM F1717) and the performance of the predicate devices. There is no biological or expert-defined "ground truth" in the typical medical imaging/diagnosis sense, as this is a mechanical implant.

8. The sample size for the training set

This information is not applicable/provided. No clinical studies were performed, and there is no mention of an algorithm or AI requiring a training set.

9. How the ground truth for the training set was established

This information is not applicable/provided. No training set or associated ground truth was established, as no clinical studies or AI development are described. The focus is on demonstrating mechanical equivalence to predicate devices.

{0}------------------------------------------------

JUN 1 9 2014

510(k) Summary

for the CosmoLock Pedicle Screw System

In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the CosmoLock Pedicle Screw System

1. GENERAL INFORMATION

Date Prepared:June 16, 2014
Trade Name:CosmoLock Pedicle Screw System
Common Name:pedicle screw system
Classification Name:orthosis, spinal pedicle fixationorthosis, spondylolisthesis spinal fixation
Class:II
Product Code:MNIMNH
CFR section:21 CFR section 888.3070
Device panel:Orthopedic
Predicate Device:Legally Marketed Alphatec - ZODIAC™ Polyaxial Spinal Fixation System (K033090 / K042673 / K071890K093077 / K100685)Stryker Spine- Xia Spinal System (K050461 / K052761 / K060361 / K060748 / K071373K113666)Cross Medical - Synergy VLS (K940631 / 950099)DePuy MOTECH - Moss Miami SS (K950697)Cross Medical - PWB (now Synergy) (K920116)
Submitter:Kalitec Direct, LLC618 E. South Street, Suite 500Orlando, FL 32801407-545-2063 Tele407-358-5441 Fax
Contact:J.D. Webb1001 Oakwood BlvdRound Rock, TX 78681512-388-0199 Tele512-692-3699 Faxe-mail: jdwebb@orthomedix.net

2. DEVICE DESCRIPTION

The CosmoLock Pedicle Screw System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

Materials:

Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

{1}------------------------------------------------

SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES 3.

The CosmoLock Pedicle Screw System is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performances.

4. INTENDED USE

The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

NON-CLINICAL TEST SUMMARY 5.

The following tests were conducted:

  • � Static and dynamic compression per ASTM F1717
  • Static torsion per ASTM F1717 .

The results of this testing indicate that the CosmoLock Pedicle Screw System is equivalent to predicate devices.

CLINICAL TEST SUMMARY દ.

No clinical studies were performed

CONCLUSIONS NONCLINICAL AND CLINICAL 7.

Kalitec Direct, LLC considers the CosmoLock Pedicle Screw System to be equivalent to the predicate device listed above. This conclusion is based upon the device's similarities in principles of operation, technology, materials and indications for use

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 19, 2014

Kalitec Direct, LLC % Mr. J.D. Webb OrthoMedix Group, Incorporated 1001 Oakwood Boulevard Round Rock. Texas 78681

Re: K140678

Trade/Device Name: CosmoLock Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: May 23, 2014 Received: May 27, 2014

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Mr. J.D. Webb

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.him for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins

  • for
    Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140678

Devica Name CosmoLock Pedicle Screw System

Indications for Use (Describe)

The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Zane Wally Watt Division of Orthopedic Devices

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.