Search Results
Found 3 results
510(k) Data Aggregation
(98 days)
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion.
The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft.
The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
The Polaris Spinal System is a non-cervical spinal fixation device. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This submission is a line extension to Polaris Spinal System to add other styles of multiaxial screw, including a screw with translating feature, and to provide the Polaris Spinal System sterile packed.
The provided document describes a 510(k) summary for the Biomet Spine Polaris Spinal System - Translation Screw, which is a non-cervical spinal fixation device.
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical Testing: | |
Static compression bending per ASTM 1717-11a | Met all mechanical test requirements based on worst-case construct testing. |
Static torsion per ASTM 1717-11a | Met all mechanical test requirements based on worst-case construct testing. |
Dynamic compression bending fatigue per ASTM 1717-11a | Met all mechanical test requirements based on worst-case construct testing. |
Substantial Equivalence: | Reported as substantially equivalent to predicate devices regarding material, design, operational principles, intended use, indications, and fundamental technology. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in the context of clinical data for algorithmic performance. The testing performed was mechanical testing of the device hardware.
- Test Set Sample Size: Not applicable in the context of clinical data or algorithm performance.
- Data Provenance: Not applicable. The "data" comes from physical mechanical tests conducted on the device components.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth, in the context of algorithm performance, typically refers to expert-validated clinical diagnoses or outcomes. This submission focuses on mechanical performance data of a physical medical device.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods are relevant for resolving discrepancies in expert opinions for clinical ground truth. Mechanical testing has well-defined pass/fail criteria per ASTM standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
No, an MRMC comparative effectiveness study was not done. This device is a physical spinal fixation system, not an AI or imaging diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. This submission is for a physical medical device.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The "ground truth" for the mechanical testing was the established pass/fail criteria defined by ASTM 1717-11a standard test methods for static and fatigue testing of spinal implant constructs in a vertebrectomy model.
8. The Sample Size for the Training Set
Not applicable. This device is a physical implant, not an AI algorithm requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for a physical device.
Ask a specific question about this device
(115 days)
The Dali Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all of the following indications regardless of the intended use:
- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis;
- trauma (i.e. fracture or dislocation);
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis);
- tumor;
- stenosis, and;
- failed previous fusion (pseudoarthrosis)
This system is comprised of screws, set screws, rods, crosslink and connectors. The components of this system are manufactured of Titanium alloy (Titanium-6Aluminum-4Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537). The screws are available from 4.0 to 8.5mm diameters with lengths ranging from 20-150mm.
The provided text describes the "Dali Spinal Fixation System," a Class III medical device for spinal fixation. Here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical Performance | Bench testing results demonstrate that Dali Spinal Fixation System performs equivalently to the predicates in: |
- Static compression bending
- Static torsion
- Dynamic compression bending (in accordance with ASTM F1717-10)
- Gripping-push down (in accordance with ASTM F1798) |
| Material Equivalence | The components of this system are manufactured of Titanium alloy (Titanium-6Aluminum-4Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537), which are the same materials used in predicate devices. |
| Design Equivalence (substantially similar) | The Dali Spinal Fixation System shares technological characteristics similar to the predicate devices, including similar design. |
| Intended Use / Indications for Use | The device's intended use and indications for use as a non-cervical spinal fixation device (posterior pedicle fixation T1-S2/ilium, or anterolateral fixation T8-L5) for skeletally mature patients as an adjunct to fusion for conditions like degenerative disc disease, spondylolisthesis, trauma, deformities, tumor, stenosis, and failed previous fusion are deemed substantially equivalent to legally marketed predicate devices. |
| Overall Safety and Effectiveness | The Dali Spinal Fixation System is substantially equivalent to the device referenced above and is therefore safe and effective for its intended use. |
2. Sample size used for the test set and the data provenance
The document explicitly states: "Clinical data and conclusions were not needed for this device." This means there was no clinical test set involving human subjects. The evaluation was based on non-clinical (bench) testing. Therefore, there is no sample size for an clinical test set or data provenance (country of origin, retrospective/prospective clinical data).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. As no clinical data was required or used, there was no need for experts to establish ground truth for a clinical test set. The acceptance was based on engineering and material equivalence to predicate devices.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. No clinical test set with human data requiring adjudication was performed.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a spinal fixation system (physical implant), not an AI-assisted diagnostic or imaging device. Therefore, an MRMC study related to human reader performance with or without AI assistance is irrelevant and was not performed.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a physical medical implant, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical performance evaluation, the "ground truth" was established by standardized mechanical testing protocols (e.g., ASTM F1717-10 for dynamic compression bending, and ASTM F1798 for gripping-push down) and material specifications (ASTM F136 for Titanium alloy, ASTM F1537 for CoCrMo alloy). The "truth" was that the device met or exceeded the performance of the predicate devices under these controlled, quantifiable conditions.
8. The sample size for the training set
Not applicable. This device is a physical implant evaluated through bench testing and comparison to predicate devices, not an AI or machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there was no training set.
Ask a specific question about this device
(28 days)
The FOCUS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. And the FOCUS Spinal System can be used in an open approach and a percutaneous approach with MIS instrumentation.
The FOCUS Spinal System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and for lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.
When used in a percutaneous approach with MIS Instrumentation, the FOCUS Spinal System is intended for noncervical pedicle fixation for the following indications; degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and for lordosis); turnor; pseudarthrosis; and failed previous fusion in skeletally mature patients.
FOCUS Spinal System consists of cannulated polyscrews, straight rods, curved rods, crosslinks and set screw components that can be used via percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are fabricated from titanium alloy (ASTM F136). The implants will be provided non-sterile.
Here's an analysis of the provided text regarding acceptance criteria and performance study details:
The provided document describes a 510(k) submission for the FOCUS Spinal System. It is a Class III medical device, specifically a pedicle screw spinal system.
Based on the content, the acceptance criteria and the study proving the device meets these criteria are primarily focused on mechanical performance and substantial equivalence to a predicate device, rather than clinical performance or AI-assisted diagnostic capabilities.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria/Test | Performance Demonstrated |
---|---|---|
Mechanical Performance | Static Compression Bending (ASTM F1717) | Performed (details not given, but concluded as "performs as well as the predicate device") |
Static Tension (ASTM F1717) | Performed (details not given, but concluded as "performs as well as the predicate device") | |
Static Torsion (ASTM F1717) | Performed (details not given, but concluded as "performs as well as the predicate device") | |
Dynamic Compression Bending (ASTM F1717) | Performed (details not given, but concluded as "performs as well as the predicate device") | |
Substantial Equivalence | Equivalence in design, function, and intended use to legally marketed devices. | "No significant differences... which would adversely affect the use of the product. It is substantially equivalent to these other devices." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in the context of clinical data or patient studies. The "test set" for this device refers to physical samples of the FOCUS Spinal System components used for mechanical testing.
- Sample Size: The document states that "Static compression bending, tension, torsion and dynamic compression bending were performed according to ASTM F1717 on a worst-case, screw construct." The exact number of constructs tested is not specified.
- Data Provenance: The tests are mechanical in nature, performed on the device itself. There is no mention of human subjects, patient data, country of origin of data, or whether it was retrospective or prospective.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: This information is not provided as the study described is a mechanical performance study, not a clinical study requiring expert assessment for ground truth.
- Qualifications of Experts: Not applicable.
4. Adjudication Method
- Adjudication Method: Not applicable, as there is no clinical data or image interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done?: No. The provided text does not mention any MRMC study or any comparison of human readers with or without AI assistance. The device is a physical spinal fixation system, not an AI diagnostic tool.
- Effect Size of Human Readers Improvement with AI: Not applicable.
6. Standalone Performance Study (Algorithm Only)
- Standalone Study Done?: No. This device is a physical implant, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.
7. Type of Ground Truth Used
The ground truth for this device is based on mechanical engineering standards (ASTM F1717) and a comparison of the mechanical performance of the FOCUS Spinal System to a predicate device. There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for the device's substantial equivalence determination.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This is a mechanical device, not an AI model, and therefore does not involve a "training set" in the machine learning sense.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth Was Established: Not applicable, as there is no training set for an AI model. The "ground truth" for the device's safety and effectiveness relies on its mechanical properties meeting established standards and being comparable to legally marketed predicate devices.
Ask a specific question about this device
Page 1 of 1