(56 days)
VENUS Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use:
- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis;
- trauma (i.e., fracture or dislocation);
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); .
- tumor;
- stenosis, and
- failed previous fusion (pseudoarthrosis)
The VENUS Spinal Fixation System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the VENUS Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudoarthrosis).
This system is comprised of screws, set screws, rods, crosslink and connectors. The components of this system are manufactured of Titanium alloy (Titanium-6Aluminum-4 Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537). The screws are available from 4.0 to 8.5mm diameters with lengths ranging from 20-150mm.
This document describes the design and testing of the VENUS Spinal Fixation System, a non-cervical spinal fixation device.
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Static Compression Bending (ASTM F1717-10) | Performance must be equivalent to predicate devices. | The VENUS Spinal Fixation system performs equivalently to the predicates in static compression bending. |
| Static Torsion (ASTM F1717-10) | Performance must be equivalent to predicate devices. | The VENUS Spinal Fixation system performs equivalently to the predicates in static torsion. |
| Dynamic Compression Bending (ASTM F1717-10) | Performance must be equivalent to predicate devices. | The VENUS Spinal Fixation system performs equivalently to the predicates in dynamic compression bending. |
| Gripping-Push Down (ASTM F1798) | Performance must be equivalent to predicate devices. | The VENUS Spinal Fixation system performs equivalently to the predicates in gripping-push down. |
Note: The document explicitly states that the acceptance criterion for all performance tests is "equivalent to the predicates." The specific numerical or qualitative thresholds for this equivalence are not provided in this summary but would be detailed in the full test reports.
2. Sample Size and Data Provenance for Test Set
- Sample Size: Not explicitly stated for each test. The document refers to "bench testing results," implying that multiple samples of the VENUS Spinal Fixation System components were subjected to each test.
- Data Provenance: The tests are "bench testing" conducted by L&K BIOMED, Co., Ltd. The country of origin for the testing would be South Korea, where L&K BIOMED is located. The data is prospective, generated specifically for this 510(k) submission.
3. Number of Experts and Qualifications for Ground Truth of Test Set
This section is not applicable as the provided documentation describes a medical device (spinal fixation system) and its mechanical performance testing, not an AI/diagnostic device that would require expert-established ground truth for a test set. The "ground truth" here is the physical measurement of mechanical properties.
4. Adjudication Method for Test Set
This section is not applicable for the same reasons as point 3. Mechanical test results do not typically involve adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This section is not applicable as the provided documentation describes a medical device (spinal fixation system) and its mechanical performance testing, not an AI or diagnostic device that would involve human readers or MRMC studies.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
This section is not applicable as the provided documentation describes a medical device (spinal fixation system) and its mechanical performance testing, not an AI or algorithm-based device.
7. Type of Ground Truth Used
The ground truth used for performance evaluation is based on standardized mechanical testing results (e.g., measurements of strength, stiffness, and fatigue performance) as per established ASTM standards (F1717-10 and F1798). The "ground truth" performance is then compared to that of legally marketed predicate devices.
8. Sample Size for the Training Set
This section is not applicable as the provided documentation describes a medical device (spinal fixation system) and its mechanical performance testing, not a machine learning or AI device that would have a training set.
9. How the Ground Truth for the Training Set Was Established
This section is not applicable for the same reasons as point 8.
{0}------------------------------------------------
ge 1 of 2.
MAR 2 6 2012
L&K BIOMED
510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
| 1. | Submitter: | L&K BIOMED Co., Ltd.#1104, Ace High-end Tower 3 cha, 371-50, Gasan-Dong,Geumcheon-gu, Seoul 153-803 Republic of Korea |
|---|---|---|
| Contact Person: | Hee Kyeong JooTelephone. 82-2-2624-1471FAX .82-2-2624-1477Email.hkjoo83@gmail.com | |
| Prepared Date | January 25, 2012 |
Device Identification 2.
| Trade Name: | VENUS Spinal Fixation System |
|---|---|
| Common Name: | Spinal Fixation Appliances |
| Product Code: | NKB, KWP, KWQ, MNH, MNI |
| Class | Class III |
| ClassificationName: | 1) Spinal Interlaminal Fixation Orthosis, 21 CFR §888.30502) Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.30603) Pedicle Screw Spinal System, 21 CFR §888.3070 |
3. Identification of Legally Marketed Devices
L&K BIOMED: VENUS Spinal Fixation System (K 100706,K 103085) GS Medical: Anyplus Spinal Fixation System (K091717) Spine Craft: APEX Spine System (K062513,K092825,K102488) Stryker Spine: Xia®III Spinal System (K071373,K083393,K091291) Synthes Spine: USS Polyaixal and Synthes USS Iliosacral System (K022949,K082572)
4. Device Description
This system is comprised of screws, set screws, rods, crosslink and connectors. The components of this system are manufactured of Titanium alloy (Titanium-6Aluminum-4 Vanadium ELI, per ASTM F136) and CoCrMo alloy (Cobalt-28Chromium-6Molybdenum, per ASTM F1537). The screws are available from 4.0 to 8.5mm diameters with lengths ranging from 20-150mm.
5. Intended use / Indications for Use
VENUS Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use:
{1}------------------------------------------------
- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis;
- trauma (i.e., fracture or dislocation);
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); .
- tumor;
●
- stenosis, and
- failed previous fusion (pseudoarthrosis)
The VENUS Spinal Fixation System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the VENUS Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudoarthrosis).
6. Comparison of the Technology Characteristics
The VENUS Spinal Fixation System shares technological characteristics similar to the predicate devices. These characteristics include similar design, the same materials, substantially equivalent performance characteristics and the same intended use.
7. Performance Data
Non-Clinical Performance and Conclusions:
Bench testing results demonstrate that VENUS Spinal Fixation system performs equivalently to the predicates in static compression bending, static tersion, static torsion, dynamic compression bending (in accordance with ASTM F1717-10) and gripping-push down (in accordance with ASTM F1798).
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device.
8. Conclusion
The VENUS Spinal Fixation System is substantially equivalent to the device referenced above and is therefore safe and effective for its intended use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
L&K BIOMED Co., Ltd. % Hee Kyeong Joo Manager #1104, Ace High-end Tower 3 cha, Gasan-Dong Geumcheon-gu Seoul 153-803 Republic of Korea
MAR 2 6 2012
Re: K120270
Trade/Device Name: VENUS Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, KWP, KWQ Dated: January 25, 2012 Received: January 30, 2012
Dear Hee Kyeong Joo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Hee Kyeong Joo
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific advice to: Jour Critices/CDRH/CDRHOffices/ucm115809.htm for go to mep.//www.tadiological Health's (CDRH's) Office of Compliance. Also, please the comer for Devices that reading by reference to premarket notification" (21CFR Part note the regulation regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
for Peter R
Der
Mark N. Meller
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
L&K BIOMED
Indications for Use Statement
510(k) Number (if known):
Device Name: VENUS Spinal Fixation System
Indications For Use:
VENUS Spinal Fixation System is non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), or as an anterolateral fixation system (T8-L.5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use:
- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis; ●
- trauma (i.e., fracture or dislocation); .
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); .
- . tumor;
- . stenosis, and
- failed previous fusion (pseudoarthrosis) .
The VENUS Spinal Fixation System is a pedicle screw system indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the VENUS Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudoarthrosis).
| Prescription Use | ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ |
|---|---|
Over-The-Counter Use
(Part 21 CER801 Subpart D)
(21 CER801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
| Concurrence of CDRH, Office of Device Evaluation (OED) | |
|---|---|
| (Division Sign-Off) | |
| VENUS Spinal Fixation System 510(K) | Page 10/ 126 |
| Division of Surgical, Orthopedic, and Restorative Devices |
K120270 S10(k) Number
N/A