(80 days)
The Synthes USS (including USS Side-Opening, USS Dual-Opening, USS Small Stature (which includes small stature and pediatric patients), USS VAS variable axis components, USS Fracture, Click' X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degencrative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), turnor, stenosis, and failed previous fusion (pseudoarthrosis).
When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.
When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click' X, Click' X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) can be linked to the CerviFix System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral).
In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) can be interchanged with all USS 6.0 mm rods and transconnectors.
The Synthes USS Polyaxial and Synthes USS Iliosacral Systems are additions to Synthes' existing posterior thoracolumbar spine systems. The Synthes USS Polyaxial and Synthes USS Iliosacral Systems consist of a family of screws and connectors designed to facilitate construct assembly and complex spine manipulation. The systems are intended for additional rod fixation in the ilium and sacrum. The systems contain a variety of different connectors for attachment to the ilium and to the S2 pedicle. The implants are all manufactured from either Titanium Aluminum Niobium (Ti-6Al-7Nb) ASTM F1295 or Commercially Pure Titanium (CpTi), the same as the predicates.
This document is a 510(k) summary for the Synthes USS Polyaxial System and Synthes USS Iliosacral System. It states that non-clinical performance and conclusions were sufficient, and clinical data was not required for this device. Therefore, a study proving the device meets acceptance criteria, as per the detailed request, was not performed or deemed necessary.
The provided information focuses on demonstrating substantial equivalence to predicate devices through bench testing, rather than presenting a performance study with detailed acceptance criteria and reported device performance in a clinical or simulated clinical setting.
Here's a breakdown based on the document's content, noting where information is not available:
1. Table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Not Explicitly Stated) | Reported Device Performance |
|---|---|
| Likely based on performance of predicate devices for design, function, and material equivalence. | "Bench testing results demonstrate that the Synthes USS Polyaxial and Synthes USS Iliosacral Systems are substantially equivalent to the predicate devices." (No specific quantitative performance metrics are provided). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size: Not specified. This refers to non-clinical bench testing.
- Data Provenance: Not specified. The testing is non-clinical bench testing, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable as no clinical data or ground truth requiring expert consensus was used for this 510(k) submission. The evaluation was based on engineering bench tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable; no expert adjudication was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case comparative effectiveness study was not done. This device is a surgical implant, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable; this is a physical medical device, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not applicable for the performance evaluation. The "ground truth" for substantial equivalence was the performance of the predicate devices in terms of design, function, and material properties, evaluated through bench testing.
8. The sample size for the training set:
- Not applicable; there was no training set as no AI/algorithm was developed or evaluated.
9. How the ground truth for the training set was established:
- Not applicable; there was no training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for SYNTHES Spine. The word "SYNTHES" is in all caps and bolded. Below that is the word "Spine" in a smaller, cursive font.
6 510(k) Summary
NOV 2 4 2008
| 510(k) Summary | |
|---|---|
| Name of Firm: | Synthes Spine1302 Wrights Lane EastWest Chester, PA 19380 |
| 510(k) Contact: | Jason LipmanRegulatory Affairs SpecialistTelephone: 610-719-5629Facsimile: 610-719-5102Email: Lipman.Jason@Synthes.com |
| Date Prepared: | September 3, 2008 |
| Trade Name: | Synthes USS Polyaxial SystemSynthes USS Iliosacral System |
| Classification: | 21 CFR 888.3070 - Pedicle screw spinal systemClass IIIOrthopaedic and Rehabilitation Devices PanelProduct Code: NKB, MNH, MNI, KWQ, KWP |
| Predicates: | Synthes USS Fracture (K010658)Synthes Universal Sacral System (K963045)Synthes USS VAS (K002517)Synthes Click'X (K992739, K031175)Synthes Pangea (K052123)Synthes USS (K010108, K022949) |
| DeviceDescription: | The Synthes USS Polyaxial and Synthes USS Iliosacral Systems are additions toSynthes' existing posterior thoracolumbar spine systems. The Synthes USSPolyaxial and Synthes USS Iliosacral Systems consist of a family of screws andconnectors designed to facilitate construct assembly and complex spinemanipulation.The systems are intended for additional rod fixation in the ilium and sacrum. Thesystems contain a variety of different connectors for attachment to the ilium andto the S2 pedicle. The implants are all manufactured from either TitaniumAluminum Niobium (Ti-6Al-7Nb) ASTM F1295 or Commercially PureTitanium (CpTi), the same as the predicates. |
| Intended Use/Indications forUse: | The Synthes USS (including USS Side-Opening, USS Dual-Opening, USS SmallStature (which includes small stature and pediatric patients), USS VAS variableaxis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, PangeaMonoaxial, USS Polyaxial, and USS Iliosacral) are non-cervical spinal fixationdevices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateralfixation system (T8-L5). Pedicle screw fixation is limited to skeletally maturepatients with the exception of the Small Stature USS. These devices areindicated as an adjunct to fusion for all of the following indications regardless ofthe intended use: degenerative disc disease (defined as discogenic back pain withdegeneration of the disc confirmed by history and radiographic studies),spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures(i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor,stenosis, and failed previous fusion (pseudoarthrosis). |
| 510(k) Summary | |
| When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct. | |
| When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) can be linked to the CerviFix System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral).In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) can be interchanged with all USS 6.0 mm rods and transconnectors. | |
| Comparison ofthe device topredicatedevice(s): | The Synthes USS Polyaxial and Synthes USS Iliosacral Systems are a result of design modifications to the predicate devices. It is substantially equivalent to the predicates in design, function, material, and intended use. |
| PerformanceDate(Non-Clinicaland/or Clinical): | Non-Clinical Performance and Conclusions:Bench testing results demonstrate that the Synthes USS Polyaxial and Synthes USS Iliosacral Systems are substantially equivalent to the predicate devices.Clinical Performance and Conclusions:Clinical data and conclusions were not needed for this device. |
{1}------------------------------------------------
Image /page/1/Picture/5 description: The image shows the logo for Synthes Spine. The word "SYNTHES" is written in a bold, sans-serif font, with a registered trademark symbol to the right. Below the word "SYNTHES" is the word "Spine" in a smaller, italicized font.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 4 2008
Synthes Spinc % Mr. Jason Lipman Regulatory Affairs Specialist 1302 Wrights Lane East West Chester, Pennsylvania 19380
Re: K082572
Trade/Device Name: Synthes USS Polyaxial System and Synthes USS Iliosacral System Regulation Number: 21 CFR 888.3070 Regulation Names: Pedicle screw spinal system Regulatory Class: II Product Code: NKB, MNI, MNH, KWP, KWQ Dated: September 3, 2008 Received: September 9, 2008
Dear Mr. Lipman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Jason Lipman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device acverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SYNTHES Spine. The logo consists of a symbol on the left, followed by the word "SYNTHES" in bold, uppercase letters. Below "SYNTHES" is the word "Spine" in a smaller, cursive font. The logo is simple and professional, and it is likely used to represent a company that manufactures or sells spinal implants or other medical devices.
5 Indications for Use Statement
510(k) Number: K082572
Synthes USS Polyaxial System and Synthes USS Iliosacral System Device Name:
Indications for Use:
The Synthes USS (including USS Side-Opening, USS Dual-Opening, USS Small Stature (which includes small stature and pediatric patients), USS VAS variable axis components, USS Fracture, Click' X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degencrative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), turnor, stenosis, and failed previous fusion (pseudoarthrosis).
When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.
When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click' X, Click' X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) can be linked to the CerviFix System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral).
In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Polyaxial, and USS Iliosacral) can be interchanged with all USS 6.0 mm rods and transconnectors.
Prescription Use X (21 CFR 801 Subpart D) AND / OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---|---|
| Division of General, Restorative, | |
| and Neurological Devices | |
| 510(k) Number | K082572 |
CONFIDENTIAL Traditional 510(k) - Synthes USS Iliosacral and Polyaxial Systems
Page 19 of 46
N/A