    K Number
    K140678
    Date Cleared
    2014-06-19

    (93 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The CosmoLock Pedicle Screw System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials:

    Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

    AI/ML Overview

    This document is a 510(k) Summary for the CosmoLock Pedicle Screw System. It describes the device and its intended use and claims substantial equivalence to predicate devices. However, it does not contain the detailed information needed to answer many of your questions, as it explicitly states that no clinical studies were performed.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test)Reported Device Performance
    Static compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Dynamic compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Static torsion per ASTM F1717Equivalent to predicate devices (implied: met criteria)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The studies mentioned are non-clinical (mechanical tests), and details about sample sizes for these tests are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. No clinical studies were performed, and therefore no ground truth established by experts is relevant to the data presented.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No clinical studies were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This document states, "No clinical studies were performed."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a mechanical pedicle screw system, not an AI or algorithm-based device. No standalone performance study in this context was performed.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth for the non-clinical tests would be the established mechanical testing standards (ASTM F1717) and the performance of the predicate devices. There is no biological or expert-defined "ground truth" in the typical medical imaging/diagnosis sense, as this is a mechanical implant.

    8. The sample size for the training set

    This information is not applicable/provided. No clinical studies were performed, and there is no mention of an algorithm or AI requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. No training set or associated ground truth was established, as no clinical studies or AI development are described. The focus is on demonstrating mechanical equivalence to predicate devices.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthrosis; and Failed Previous Fusion.

    The Radius® Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation system, the Radius® Spinal system is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Radius® Spinal System can also be linked to the XIA® Titanium Spinal System via the Ø5.5mm to Ø6.0mm Radius® rod-to-rod connector.

    The Stryker Spine TRIO® Plate System is intended for posterior, noncervical (T10-S1) pedical and nonpedical fixation of the spine for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.

    The Stryker Spine TRIO® Spinal Fixation System is intended for posterior, noncervical pedicle and non-pedicle fixation of the spine. The Stryker Spine TRIO® Spinal Fixation System is indicated for: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® Spinal Fixation Sytem is intended to be used in conjunction with the OSS Diapason Rods, Opus Spinal System Rods, and the Multi-Axis Cross Connectors.

    The Stryker Spine TRIO® Spinal System is intended for posterior, noncervical pedicle and nonpedicle fixation of the spine to provide immobilization and stabilization of spinal segements in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® + Spinal System is intended to be used in conjunction with the OSS/Diapason Rods or Opus Rods, XIA® Pre-bent Rods, and the Multi-Axis Cross Connectors.

    The Stryker Spine TRIO® TRAUMA Spinal System is intended for percutaneous, posterior, non-cervical pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.

    The XIA® Spinal System and XIA® 4.5 Spinal System are intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DDD) (Defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e., fracture or dislocation); Spinal Stenosis; Curvatures (i.e., Scoliosis, Kyphosis, and/or Lordosis); Tumor; Pseudoarthrosis and; Failed previous fusion. The 6mm diameter rods from the DIAPASON® Spinal System and OPUS® Spinal System are intended to be used with the other components of the XIA® Titanium Spinal System. The Titanium Multi-Axial Cross Connector are intended to be used with the other components of the XIA® Titanium Spinal System.

    The XIA® 3 Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle fixation system, the XIA® 3 Spinal System is intended to provide additional support during autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor, Pseudoarthorisis; and Failed Pervious Fusion. The Ø5.5mm rods from the Stryker Spine Radius® Spinal System and the Ø6.0mm Vitallium Rods from the XIA® Spinal System are intended to be used with the other components of the XIA® 3 Spinal System. When used for posterior noncervical pedicle screw fixation in pediatric patients, the XIA® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. The XIA® 3 Spinal System for pediatic use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The XIA® 4.5 Spinal System is intended for anterior/anteriolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. Except for the staples, the XIA® 4.5 Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical pedicle screw fixation in pediatric patients. The XIA® 4.5 Spinal System for pediatric use in intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The STRYKER Spine thoraco-lumbar spinal fixation systems, subject of this 510(k), are non-cervical, pedicle and non-pedicle fixation systems comprised of screws, rods, plates, hooks, connectors, washers and staples. The components are manufactured from either Titanium (Titanium Alloy and CP Titanium), Stainless Steel or Cobalt-Chromium-Molybdenum Alloy (Vitallium®). This Special 510(k) submission seeks clearance for sterile labeling of the listed STRYKER Spine thoraco-lumbar spinal fixation systems. All of the components of the subject devices will be sterilized by gamma radiation, a traditional sterilization method as per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.

    AI/ML Overview

    This is a 510(k) Premarket Notification for sterile packaging of a range of spinal systems. The key claim is substantial equivalence to previously cleared non-sterile versions of the same devices. Therefore, the "device" in question for this submission is specifically the sterile packaging and the sterilization process, not the spinal implant itself as a medical device performing a physiological function. The "performance" being evaluated is the ability to maintain sterility and the structural integrity of the device within its packaging over its shelf life.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Sterilization EffectivenessDemonstrated by gamma radiation, a traditional sterilization method per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.
    Sterile Barrier Packaging SystemDemonstrated effective in maintaining sterility for the recommended 5-year shelf-life.
    Impact on Device CharacteristicsNo design modifications were incorporated to facilitate sterile packaging. The sterile-packed implant components maintain the same technological characteristics (design, technical requirements, materials of construction, indications/intended use) as the non-sterile predicate devices.
    Shelf LifeDemonstrated a 5-year shelf life in sterile packaging.
    Safety and EffectivenessConcluded to be safe and effective, retaining the same intended use, indications, technological characteristics, and mode of operation as the predicate non-sterile devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses the sterilization process and packaging rather than clinical performance of the implant itself. Therefore, typical "test set" in the context of clinical or diagnostic AI/software devices (e.g., patient data) is not applicable here.

    • Sterilization and Packaging Validation Data: The document mentions "accelerated aging data" for validating the 5-year shelf life. It does not specify the sample size for this testing, nor the country of origin or whether it was retrospective or prospective. However, these types of tests are typically prospective laboratory studies conducted in controlled environments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This submission is about the sterility and packaging of medical implants, not a diagnostic or clinical performance study that would require expert ground truth labeling. The "ground truth" here is the scientific validation of sterilization methods and shelf life according to recognized standards and FDA guidance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used in clinical studies for establishing ground truth, which is not relevant to this submission on sterile packaging.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this submission relies on:

    • Scientific Validation of Sterilization: Adherence to established scientific principles and FDA guidance for sterilization (gamma radiation), demonstrating a Sterility Assurance Level (SAL).
    • Accelerated Aging Data: Laboratory testing to simulate real-time aging and demonstrate that the sterile barrier packaging system maintains sterility for the stated shelf life.
    • Material and Design Specification Conformity: Confirmation that the components are manufactured from specified materials and that their design, technical requirements, and intended use are identical to the non-sterile predicate devices.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not a machine learning or AI device, there is no training set and therefore no ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121342
    Manufacturer
    Date Cleared
    2012-06-22

    (49 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    Reference Devices :

    K050461, K060361, K060748, K092605

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XIA® 4.5 Spinal System is intended for anterior/anteriolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications:

    • Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
    • Spondylolisthesis:
    • Trauma (i.e. fracture or dislocation);
    • Spinal stenosis;
    • Curvatures (i.e. scoliosis, kyphosis, and/or lordosis);
    • Tumor;
    • Pseudoarthrosis: and
    • Failed previous fusion.

    The Stryker Spine DIAPASON™ Spinal System, OPUS™ Spinal System and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector.

    When used for posterior noncervical pedicle screw fixation in pediatric patients, the XIA 4.5 Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The XIA 4.5 Spinal System for pediatric use is intended to be used with autograft and/or allografi. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The XIA® 4.5 Spinal System consists of a variety of titanium Monoaxial & Polyaxial Bone Screws, Monoaxial & Polyaxial Reduction Screws, Hooks, Blockers, Rod Connectors, Dual Staples and Connectors. The existing components of the XIA 4.5 Spinal System were determined substantially equivalent through K050461 for class II indications consistent with 21 CFR 888.3070 (b) (1). The indications were expanded through K060361 to include class III indications per product code NKB, 21 CFR 888.3070 (b) (2). Subsequent line extensions have been cleared through K060748, and K092605.

    The expansion of indications for the XIA 4.5 Spinal System is proposed for the inclusion of adolescent idiopathic scoliosis alone, and not other indications for a pediation. As pediatric patients are unlikely to exhibit symptoms of degenerative disc disease (DDD) or stenosis due to the wear and tear on the spine necessary to develop these diseases, expansion of these indications to a pediation is not warranted.

    AI/ML Overview

    This document (K121342) is a 510(k) premarket notification for the Stryker XIA® 4.5 Spinal System, seeking clearance for expanded indications, specifically for adolescent idiopathic scoliosis.

    The acceptance criteria and study information indicate that the device's substantial equivalence relies on biomechanical testing and a clinical literature analysis, not on a clinical ground truth study as typically seen for AI/ML devices. Therefore, many of the requested fields are not applicable.

    Here's the information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Static compression bending (per ASTM F1717)Substantially Equivalent: Testing demonstrated equivalence to predicate devices.
    Dynamic compression bending (per ASTM F1717)Substantially Equivalent: Testing demonstrated equivalence to predicate devices.
    Static torsion testing (per ASTM F1717)Substantially Equivalent: Testing demonstrated equivalence to predicate devices.
    Interconnection strength testing (per ASTM F1798)Substantially Equivalent: Testing demonstrated equivalence to predicate devices.
    Clinical Literature AnalysisSubstantially Equivalent: Analysis supported equivalence to predicate devices.
    Dimensional ComparisonsSubstantially Equivalent: Comparisons supported equivalence to predicate devices.
    Engineering AnalysesSubstantially Equivalent: Analyses supported equivalence to predicate devices.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable. The submission relies on biomechanical testing and literature review rather than a clinical test set in the context of an AI/ML device.
    • Data Provenance: Not applicable for a clinical test set. Biomechanical testing would have been conducted in a lab environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable. Ground truth from experts is not mentioned for this type of submission.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable. There is no clinical test set requiring adjudication in this submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This is a spinal implant, not an AI/ML diagnostic or assistive device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Study: No. This product is a physical spinal implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: For the biomechanical testing, the "ground truth" would be the established performance characteristics of the predicate devices and the relevant ASTM standards for spinal implants. For the clinical literature analysis, the literature itself serves as the basis for comparison. No clinical "ground truth" as in pathology or outcomes data from a prospective study of this device.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable. This is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment for Training Set: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071921
    Manufacturer
    Date Cleared
    2007-11-16

    (127 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K050461, K022191

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INNOVASIS 'Excella® Spinal System, when used for pedicle screw fixation is intended only for patients:
    a) Having severe spondylolisthesis (Grade 3 & 4) at the L5-S1 joint;
    b) Who are receiving fusion using autogenous bone graft only;
    C) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below): and
    d) Who are having the device removed after the development of a solid fusion mass.

    The INNOVASIS 'Excella®' Spinal System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine:
    a) Degenerative spondylolisthesis with objective evidence of neurologic impairment;
    b) Fracture:
    C) Dislocation:
    d) Scoliosis:
    e) Kyphosis:
    f) Spinal tumor; and
    g) Previous failed fusion (pseudoarthrosis).

    The INNOVASIS 'Excella® Spinal System, when used for anterolateral nonpedicle fixation, is intended for the following indications:
    a) Deqenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies):
    b) Spinal stenosis;
    C) Spondylolisthesis;
    ರು) Spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis);
    e) Pseudoarthrosis;
    f) Tumor;
    g) Trauma (i.e. fracture or dislocation); and
    h) Previous failed fusion.

    The INNOVASIS 'Excella® Spinal System, when used for posterior non-pedicle screw fixation to the non-cervical spine, is intended for the following indications:
    a) Degenerative disc disease (defined as back pain of disogenic origin with degenerative disc confirmed by patient history and radiographic studies);
    b) Spinal stenosis;
    c) Spondylolisthesis;
    d) Spinal deformities (i.e. scoliosis, kyphosis and/or lordosis);
    e) Pseudoarthrosis;
    () Tumor:
    g) Trauma (i.e. fracture or dislocation);
    h) Previous failed fusion.

    Device Description

    The Excella® Spinal System broadens to include the Excella® -P bone screws, which feature a multi-axial joint for free motion of the screw head to allow the physician greater flexibility when placing the screws, and a series of 4.5mm diameter polyaxial bone screws of 30mm, 35mm, 40mm and 45mm lengths. The locking caps for Excella®-P are a two piece design which includes a threaded locking screw attached to an alignment cap. The alignment cap is attached to the bottom of the locking screw and rotates about the radial axis. The modified cross connectors come in various pre-assembled, adjustable lengths and allow for variation in vertical placement on two rods.

    AI/ML Overview

    This document is a 510(k) submission for a spinal system, not for an AI/ML device. Therefore, the requested information regarding acceptance criteria, study details, and AI-specific metrics (sample sizes, ground truth establishment, expert involvement, MRMC studies, standalone performance) for an AI device is not available in the provided text.

    The document discusses the mechanical performance testing for the Innovasis Excella® Spinal System to demonstrate substantial equivalence to predicate devices, but this is related to the physical properties of the spinal implants, not a software or AI/ML product.

    Therefore, I cannot provide the requested table and study details for an AI/ML device based on the provided text, as the device described is a physical spinal implant system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062853
    Manufacturer
    Date Cleared
    2006-10-24

    (29 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K050461 and K060361

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput-T3), the Stryker Spine OASYSTM System is intended for:

    • Degenerative Disc Disease (as defined by neck or back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies)
    • Spondvlolisthesis .
    • Spinal Stenosis .
    • Fracture/Dislocation ●
    • Atlanto/axial fracture with instability .
    • Occipitocervical dislocation .
    • Revision of previous cervical spine surgery .
    • Tumors .

    When used with the occipital plate, the bone screws are limited to occipital fixation only. The bone screws are not intended to be used in the cervical spine.

    The use of the polyaxial screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. They are not intended to be placed in the cervical spine.

    The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.

    The Stryker Spine OASYSTM System can also be linked to the Xia System, SR90D System and Xia 4.5 Spinal System via the rod-to-rod connectors.

    Device Description

    This Special 510(k) submission is intended to introduce a line extension to the predicate OASYSTM System, which consists of two rod-to-rod connectors (axial and parallel versions) used to link the 3.5mm OASYS™ System rods with the Xia® 4.5 Spinal System rods (510(k) numbers #K050461 and #K060361). The Xia® 4.5 Spinal System rods will retain their original cleared name.

    AI/ML Overview

    This document is a Special 510(k) Premarket Notification for a line extension to the Stryker Spine OASYS™ System. It does not describe a study proving a device meets acceptance criteria in the traditional sense of a clinical or performance study with numerical targets. Instead, the submission focuses on demonstrating substantial equivalence to a predicate device based on material composition and validated mechanical properties.

    Therefore, the requested information elements related to clinical study design, ground truth, expert adjudication, sample sizes for test/training sets, and comparative effectiveness studies are not applicable in this context.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The primary "acceptance criterion" for this 510(k) is demonstrating equivalent mechanical properties to the predicate device.

    Acceptance CriterionReported Device Performance
    Equivalent mechanical properties to the predicate OASYS™ System (K032394)"Testing has demonstrated that the additional rod-to-rod connector components have equivalent mechanical properties to the predicate OASYS™ System (K032394)."
    Same indications for use as the predicate device"Both the new components and the existing system components are intended to address the same indications for use."
    Made from the same materials as the predicate device"Both the new components and the existing components are made from the same materials."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. This document describes a Special 510(k) submission for a line extension, not a clinical study involving human or animal subjects that would require a "test set" and associated data provenance in this manner. The "testing" mentioned refers to mechanical property validation on the device components themselves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. See point 2. The "ground truth" here is the established mechanical properties of the predicate device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. See point 2.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This submission is for spinal fixation components, not an AI or imaging device involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This submission is for spinal fixation components, not an algorithm or software device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is the established mechanical properties of the predicate device, as determined by engineering standards and testing, and the material composition of the predicate device.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of this type of device submission.

    9. How the ground truth for the training set was established

    • Not applicable. There is no "training set" in the context of this type of device submission. The "ground truth" for demonstrating substantial equivalence (as described in point 7) would have been established through engineering design, material specifications, and standardized mechanical testing processes.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052761
    Manufacturer
    Date Cleared
    2005-10-11

    (11 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K050461

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XIA® 4.5 Spinal System is intended for posterior noncervical pedicle fixation for the following indications: severe spondylolisthesis (i.e. Grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.

    The XIA® 4.5 Spinal System is also intended for anterolateral and posterior, non-cervical, nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.

    The XIA® 4.5 Spinal System is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients only.

    Device Description

    The Stryker Spine Xia" 4.5 Spinal System is comprised of Ø 4.5 mm rods, Polyaxial and Monoaxial bone screws, Blockers, Hooks. Dual Staples, and Connectors. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are fabricated from titanium alloy. The implants are provided non-sterile and are used for either posterior or anterolateral non-cervical spinal fixation.

    This submission is intended to address a line extension to Xia" 4.5 Spinal System. The line extension includes a new range of Titanium alloy connectors.

    AI/ML Overview

    This submission is a 510(k) premarket notification for a line extension to an existing spinal system. It does not describe a study involving an AI/ML device or its performance characteristics. Instead, it focuses on demonstrating substantial equivalence to a predicate device for a modified medical implant. Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, ground truth, expert involvement, and comparative effectiveness studies is not applicable to this document.

    The document describes the:

    • Identification of the device and sponsor.
    • Description of the device modification: A line extension including a new range of titanium alloy connectors for the Xia® 4.5 Spinal System.
    • Intended Use: Specifies the indications for both posterior noncervical pedicle fixation and anterolateral and posterior, non-cervical, non-pedicle fixation, for skeletally mature patients only.
    • Statement of Technological Comparison: Asserts that the new components share the same intended use, material, and basic design concepts as the predicate device (Xia® 4.5 Spinal System, K050461). It also states that "Mechanical testing also demonstrated comparable mechanical properties to the predicate device." This is the only mention of "testing" and it refers to mechanical properties, not clinical or algorithmic performance.
    • FDA's substantial equivalence determination letter: Confirming the device is substantially equivalent to legally marketed predicate devices.

    In summary, there is no clinical study or AI/ML device performance evaluation presented in this document. The submission is for a modification to a spinal implant and relies on demonstrating substantial equivalence through material, design, intended use, and mechanical properties.