K Number
K090648

Validate with FDA (Live)

Device Name
VIPER 2 SYSTEM
Manufacturer
Date Cleared
2009-06-10

(91 days)

Product Code
Regulation Number
888.3070
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VIPER Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The VIPER Systems are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.

When used in a percutaneous approach with MIS Instrumentation, the VIPER Systems are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

The PEEK rods of EXPEDIUM Spine System and VIPER Systems are contraindicated for degenerative disc disease.

Device Description

The subject VIPER2 Spine System components consist of 5.5mm screws and rods and are available in various geometries and sizes.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the VIPER™ 2 System:

It's important to note that the provided documents (K090648 Special 510(k) Submission) are for a spinal implant system (VIPER™ 2 System), which is a physical device, not an AI/software medical device. Therefore, many of the typical acceptance criteria and study types requested in your prompt (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, training set details, ground truth for training set) are not applicable to this type of traditional medical device submission.

The "acceptance criteria" for a physical implant like this primarily revolve around mechanical performance and biocompatibility, demonstrating that the device is safe and effective for its intended use, and substantially equivalent to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Mechanical Performance (Strength, Durability)Performance data per ASTM F 1717 were submitted to characterize the subject VIPER2 System components.
Material BiocompatibilityManufactured from ASTM F 138 implant grade stainless steel, ASTM F 136 implant grade titanium alloy, and ASTM F 1537 implant grade cobalt-chromium-molybdenum alloy. (These are standard implant-grade materials, implying biocompatibility acceptance through material specification).
Substantial Equivalence to Predicate DevicesThe FDA reviewed the submission and determined the device is substantially equivalent to legally marketed predicate devices.
Consistency with Intended Use & IndicationsThe indications for use for the VIPER™ 2 System are consistent with those of the predicate devices.

Study that Proves the Device Meets Acceptance Criteria:

The key study mentioned is:

  • Performance data per ASTM F 1717: This is an in vitro mechanical test standard for spinal implant constructions in a vertebrectomy model. This standard is used to evaluate the fatigue life and static strength of spinal instrumentation systems.

Explanation of the Study and its Relevance to Acceptance:

The submission explicitly states that "Performance data per ASTM F 1717 were submitted to characterize the subject VIPER2 System components addressed in this notification." This means the manufacturer performed mechanical testing in accordance with this recognized industry standard.

  • ASTM F 1717 outlines specific test methodologies and conditions to simulate the loads and stresses that spinal implants experience in the body. The acceptance criteria for this type of testing would be a demonstration that the VIPER™ 2 System components (screws, rods) meet or exceed the established mechanical performance benchmarks (e.g., fatigue strength, static bending strength) for spinal fixation devices, and ideally, show comparable or superior performance to the predicate devices when tested under the same standard.
  • By meeting the requirements of ASTM F 1717, the manufacturer provides evidence that the device is structurally sound and can withstand the mechanical environment of the spine for its intended duration, thereby establishing its safety and effectiveness from a mechanical standpoint.

Regarding the specific questions that are generally not applicable to this type of device:

  1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable for a physical device mechanical test. ASTM F 1717 would involve a set number of physical prototypes for testing. The "data provenance" refers to the lab where the mechanical tests were performed.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable. "Ground truth" in this context is the physical behavior of the device under mechanical load, measured by engineering equipment, not expert opinion.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Mechanical test results are objective measurements.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a physical implant, not an AI system.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For mechanical testing, the "ground truth" is the objective measurement of mechanical properties as defined by the ASTM standard.
  7. The sample size for the training set: Not applicable (no "training set" for physical device mechanical testing).
  8. How the ground truth for the training set was established: Not applicable.

In summary, the K090648 submission for the VIPER™ 2 System relies on standardized mechanical engineering tests (ASTM F 1717) and material specifications to demonstrate product performance and substantial equivalence to predicate devices, which are the primary "acceptance criteria" for spinal implant components.

{0}------------------------------------------------

K090648

Special 510(k) Submission – Additions to VIPER™ System

5.510(K) SUMMARY
Submitter:DePuy Spine, Inc.325 Paramount DriveRaynham, MA 02767
JUN 10 2009
Contact Person:Frank JurczakRegulatory Affairs Associate
Voice:(508) 828-3288
Fax:(508) 828-3797
E-Mail:fjurczak@its.jnj.com
Date Prepared:March 9, 2009
Device Class:Class III
Classification Name:Spinal interlaminar fixation orthosisper 21 CFR §888.3050Spinal intervertebral body fixation orthosisper 21 CFR §888.3060Pedicle screw spinal fixationper 21 CFR §888.3070
Classification Panel:Orthopedics
FDA Panel Number:87
Product Code(s):NKB, KWQ, KWP, MNH, MNI
Proprietary Name:VIPER TM 2 System
Predicate Devices:EXPEDIUM TM Spine System (K082942, K033901,K041119, K073364)VIPER System (K071860, K061520, K041801)MOSS MIAMI Spine System (K933881, K962628,K983583)
Device Description:The subject VIPER2 Spine System components consist of5.5mm screws and rods and are available in variousgeometries and sizes.
Intended Use:The VIPER Systems are intended to provideimmobilization and stabilization of spinal segments inskeletally mature patients as an adjunct to fusion in the

DePuy Spine, Inc., a Johnson & Johnson Company

Page 5-1

of 2

{1}------------------------------------------------

KGA D648

Special 510(k) Submission - Additions to VIPERTM System

treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The VIPER Systems are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.

When used in a percutaneous approach with MIS Instrumentation, the VIPER Systems are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

The PEEK rods of EXPEDIUM Spine System and VIPER Systems are contraindicated for degenerative disc disease.

Materials:

Manufactured from ASTM F 138 implant grade stainless steel, ASTM F 136 implant grade titanium alloy, and ASTM F 1537 implant grade cobalt-chromiummolybdenum alloy.

Performance Data: Performance data per ASTM F 1717 were submitted to characterize the subject VIPER2 System components addressed in this notification.

Page 2 of 2

DePuy Spine, Inc., a Johnson & Johnson Company

Page 5-2

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" around the perimeter. In the center of the seal is an emblem of an eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DePuy Spine Incorporated % Mr. Frank Jurczak Associate II, Regulatory Affairs 325 Paramount Drive Raynham, Massachusetts 02767

Re: K090648

Trade/Device Name: VIPER™ 2 System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: III Product Code: NKB, KWQ, KWP, MNH, & MNI Dated: May 12, 2009 Received: May 13, 2009

Dear Mr. Jurczak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket. approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

JUN 1 0 2009

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing

{3}------------------------------------------------

Page 2-Mr. Frank Jurczak

(reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Barbara buerho

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE STATEMENT 4.

510(k) Number (if known): K090648

Device Name: VIPER™2 System

Indications For Use:

The VIPER Systems are intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The VIPER Systems are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients.

When used in a percutaneous approach with MIS Instrumentation, the VIPER Systems are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

The PEEK rods of EXPEDIUM Spine System and VIPER Systems are contraindicated for degenerative disc disease.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for vision Sign-Off)

Unvision of Surgical, Orthopedic, and Restorative Devices

19(k) Number K090648

Page 1 of 1

DePuy Spine, Inc., a Johnson & Johnson Company

N/A