(500 days)
The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc discase (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic sudies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.
With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The CD HORIZON® Pediatric Spinal System is intended to be used with autograft and/or allografi. Pediatric pedicle screw fixation is limited to a posterior approach.
The CD HORIZON SPIRE™ Plate is a posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.
In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.
The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Meduronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates, and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
Certain components within the CD HORZON® Spinal System are specifically excluded for use in pediatric patients. These include PEEK rods, CD HORIZON® SPIRE spinous process plate devices, Shape Memory Alloy Staples, DYNALOK® bolts, TSRH® screws and washers. All screws used in pediatric cases are only cleared for use via a posterior approach. All of the components used in pediatric cases are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy and medical grade cobaltchromium-molybdenum alloy.
Certain implant components from other Medtronic spinal systems can be used with the CD HORIZON® Spinal System in non-pediatric cases. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers, GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK® PLUS and DYNALOK CLASSIC® bolts along with rod/bolt connectors; and Medtronic Multi-Axial rods and screws. Please note that certain components are specifically designed to connect to 3.5mm, 4.5mm, 5.5mm rods or 6.35mm rods, while other components can connect to both 5.5mm rods and 6.35mm rods. Care should be taken so that the correct components are used in the spinal construct.
CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and associated screws are intended for anterios use only. However, for patients of smaller stature and pediatric patients, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.
The CD HORIZON® Spinal System implant components are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, medical grade cobaltchromium-molybdenum alloy, or medical grade PEEK OPTIMA-LT1. Certain CD HORIZON® Spinal System components may be coated with hydroxyapatite.
Never use stainless steel and titanium implant components in the same construct.
Medical grade titanium, titanium alloy and/or medical grade cobalt-chromiummolybdenum alloy may be used together. Never use titanium, titanium alloy and/or medical grade cobalt-chromium-molybdenum alloy with stainless steel in the same construct.
The CD HORIZON® Spinal System also includes anterior staples made of Shape Memory Alloy (Nitinol - NiTi). Shape Memory Alloy is compatible with titanium, titanium alloy and cobalt-chromium-molybdenum alloy. Do not use with stainless steel. These staples are not to be used in pediatric patients.
PEEK OPTIMA-LT1 implants may be used with stainless steel, titanium or cobaltchromium-molybdenum alloy implants. CD HORIZON® PEEK Rods are not to be used with CROSSLINK® Plates or in pediatric patients.
To achieve best results, do not use any of the CD HORIZON® Spinal System implant components with components from any other system or manufacturer unless specifically allowed to do so in this or another Medtronic document. As with all orthopaedic and neurosurgical implants, none of the CD HORIZON® Spinal System components should ever be reused under any circumstances.
The purpose of this 510(k) submission is to expand the indications of use to allow for use of pedicle screw based constructs to treat pediatric patients.
The provided text is a 510(k) summary for the CD HORIZON® Spinal System, focusing on expanding its indications for use in pediatric patients. It does not describe an AI medical device or a study proving its performance against acceptance criteria in the manner typically seen for AI/ML devices.
Therefore, most of the requested information regarding acceptance criteria, device performance, sample sizes for test and training sets, expert involvement, adjudication methods, MRMC studies, or standalone algorithm performance is not applicable to this document.
However, I can extract information related to the device's intended use and the "clinical assessment" provided in support of its expanded indication.
Here's the breakdown based on the provided text, noting where information is not available or not applicable:
1. A table of acceptance criteria and the reported device performance
This document describes a medical device, not an AI algorithm. It does not provide explicit acceptance criteria in terms of performance metrics (like sensitivity, specificity, or AUC) that would be evaluated by a study, nor does it report device performance in that way. Instead, it refers to the collection of clinical data to support the expanded indication.
| Acceptance Criteria (Not explicitly stated for performance, but implied clinical safety and effectiveness) | Reported Device Performance (Clinical Data Summary) |
|---|---|
| Implied: Demonstrate safety and effectiveness for pediatric pedicle screw fixation in adolescent idiopathic scoliosis, consistent with existing similar devices and clinical practice. | Published clinical data from: - More than 600 pediatric patients treated with pedicle screw constructs alone. - More than 900 patients treated with a hybrid construct (pedicle screws and hooks). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not explicitly detailed in the 510(k) summary for a "test set" in the context of an AI device. The document mentions "Published clinical data of pediatric patients" but does not specify the design (retrospective/prospective) or country of origin.
- Sample Size: "More than 600 pediatric patients treated with pedicle screw constructs alone" and "more than 900 patients treated with a hybrid construct." (Total >1500 pediatric patients). This is referred to as clinical data supporting the expanded indication, not a "test set."
- Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). It is referred to as "Published clinical data."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as the document describes a spinal implant system, not a device requiring ground truth established by experts in an AI context. The "ground truth" would be the clinical outcomes and diagnoses of the patients, established by treating clinicians.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. The document refers to clinical patient data and outcomes, not an adjudication process for labeling an AI test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a spinal implant system, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as it is not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the clinical data provided, the "ground truth" implicitly refers to the clinical outcomes and diagnoses of the pediatric patients treated for adolescent idiopathic scoliosis, as determined by their treating physicians. This would encompass aspects like the initial diagnosis, surgical success, fusion rates, complication rates, and long-term functional outcomes, which are standard for clinical studies of medical devices.
8. The sample size for the training set
This is not applicable as it is not an AI algorithm.
9. How the ground truth for the training set was established
This is not applicable as it is not an AI algorithm.
{0}------------------------------------------------
CD HORIZON® Spinal System 510(k) Summary - K091445
September 2010
: SEP 2 7 : 2010
I. Company: Medtronic Sofamor Danek USA 1800 Pyramid Place Memphis, Tennessec 38132 Telephone: (901) 396-3133 Fax: (901) 346-9738
Contact: Lee Grant Principal, Regulatory Affairs Specialist
- Proposed Proprietary Trade Name: CD HORIZON® Spinal System II.
- III. Classification Name(s): Spinal Interlaminal Fixation and Spinal Intervertebral Fixation Orthosis and/or Pedicle Screw Spinal System (per 21CFR Section 888.3050, 888.3060 and/or 888.3070); Product Code(s): OSH, KWP, KWQ, MNH, MNI and NKB
- IV. Description: The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Meduronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates, and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
Certain components within the CD HORZON® Spinal System are specifically excluded for use in pediatric patients. These include PEEK rods, CD HORIZON® SPIRE spinous process plate devices, Shape Memory Alloy Staples, DYNALOK® bolts, TSRH® screws and washers. All screws used in pediatric cases are only cleared for use via a posterior approach. All of the components used in pediatric cases are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy and medical grade cobaltchromium-molybdenum alloy. ¿
Certain implant components from other Medtronic spinal systems can be used with the CD HORIZON® Spinal System in non-pediatric cases. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers, GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK® PLUS and DYNALOK CLASSIC® bolts along with rod/bolt connectors; and Medtronic Multi-Axial rods and screws. Please note that certain components are specifically designed to connect to $3.5mm, $4.5mm, $5.5mm rods or $6.35mm rods, while other components can connect to both $5.5mm rods and @6.35mm rods. Care should be taken so that the correct components are used in the spinal construct.
{1}------------------------------------------------
CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORİZON® ECLIPSE® rods and associated screws are intended for anterios use only. However, for patients of smaller stature and pediatric patients, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.
The CD HORIZON® Spinal System implant components are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, medical grade cobaltchromium-molybdenum alloy, or medical grade PEEK OPTIMA-LT1. Certain CD HORIZON® Spinal System components may be coated with hydroxyapatite. No warranties express, or implied, are made: Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded. See the MDT Catalog for further information about warranties and limitations of liability.
Never use stainless steel and titanium implant components in the same construct.
Medical grade titanium, titanium alloy and/or medical grade cobalt-chromiummolybdenum alloy may be used together. Never use titanium, titanium alloy and/or medical grade cobalt-chromium-molybdenum alloy with stainless steel in the same construct.
The CD HORIZON® Spinal System also includes anterior staples made of Shape Memory Alloy (Nitinol - NiTi). Shape Memory Alloy is compatible with titanium, titanium alloy and cobalt-chromium-molybdenum alloy. Do not use with stainless steel. These staples are not to be used in pediatric patients.
PEEK OPTIMA-LT1 implants may be used with stainless steel, titanium or cobaltchromium-molybdenum alloy implants. CD HORIZON® PEEK Rods are not to be used with CROSSLINK® Plates or in pediatric patients.
To achieve best results, do not use any of the CD HORIZON® Spinal System implant components with components from any other system or manufacturer unless specifically allowed to do so in this or another Medtronic document. As with all orthopaedic and neurosurgical implants, none of the CD HORIZON® Spinal System components should ever be reused under any circumstances.
The purpose of this 510(k) submission is to expand the indications of use to allow for use of pedicle screw based constructs to treat pediatric patients.
- V. Indications for Use: The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.
{2}------------------------------------------------
With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The CD HORIZON® Pediatric Spinal System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The CD HORIZON SPIRE™ Plate is a posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.
In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.
-
V1. Clinical Assessment: Published clinical data of pediatric patients diagnosed with adolescent idiopathic scoliosis and treated specifically with CD HORIZON® Spinal System pedicle screw instrumentation was provided in support of this application. This data included results of more than 600 pediatric patients treated with pedicle screw constructs alone, along with more than 900 patients treated with a hybrid construct consisting of both pedicle screws and hooks.
( -
VII. Substantial Equivalence: The design features, material and indications for use of the CD HORIZON® Spinal System are substantially equivalent to the Orthobiom Spinal System (Paradigm Spine) previously cleared in K071668 (SE 07/02/2008) and to the Synthes Small Stature USS Spinal System applications K082572 (SE 11/24/08) and in K994121 (SE 10/17/00). No new implants were included in this application as this submission sought only to expand the indication for posterior pedicle screw constructs previously cleared by the FDA, including those in K091974, SE 09/02/09 for use in pediatric patients diagnosed with adolescent idiopathic scoliosis. The safety and effectiveness of the CD HORIZON® Spinal System for this expanded indication has been adequately supported by reported clinical results of this and similar devices which is contained within this premarket notification.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure, composed of curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2013
Medtronic Sofamor Danek % Mr. Lee Grant Principal, Regulatory Affairs 1800 Pyramid Place Memphis, Tennessee 38132
Re: K091445
Trade/Device Name: CD HORIZON® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP, KWQ Dated: September 01, 2010 Received: September 02, 2010
Dear Mr. Grant:
This letter corrects our substantially equivalent letter of September 27, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Lee Grant
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/4/Picture/7 description: The image shows the text "Erin Daeith" in a stylized font. The first name, "Erin," is written in a simple, bold font. The last name, "Daeith," is written in a more decorative font, with the letters overlapping and having a textured appearance. The overall design gives the text a unique and eye-catching look.
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
September 2010
Device Name: CD HORIZON® Spinal System
KD91445
SEP. 27, 2010
Indications for Use:
The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc discase (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic sudies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.
With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The CD HORIZON® Pediatric Spinal System is intended to be used with autograft and/or allografi. Pediatric pedicle screw fixation is limited to a posterior approach.
The CD HORIZON SPIRE™ Plate is a posterior, non-pedicle supplemental fixation device intended for use in the non-cervical spine (TI-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis, trauma; and/or tumor.
In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Surgical, Orthonedic, and Restorative Devices
Ko91445 510(k) Number_
N/A