    Intended Use

    OZARK Cervical Plate System is indicated for use in anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylisthesis, trauma(including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusion (Pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).
    PYRENEES and BLUE RIDGE Cervical Plate System are indicated for use in anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radioaraphic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).
    The EVEREST Spinal System may be used in conjunction with the RANGE® (MESA® and DENALI®) Spinal Systems, all of which are cleared for the following indications: Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis); tumor, pseudoarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system the EVEREST Spinal System may also be used for the same indications as an adjunct to fusion. When used for posterior non-cervical pedicle screw fixation in pediatric patients the EVEREST Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scolosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
    RANGE (MESA and DENALI) and ARI are cleared for the following indications: Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis); tumor, pseudoarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion. Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
    The Caspian OCT/MESA Mini Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Caspian OCT/MESA Mini Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Caspian OCT/MESA Mini/Spinal System may be connected to Range/MESA/DENALI Spinal System and Everest Spinal System components via the rod to rod connectors or transition rods.
    The Caspian OCT/MESA Mini/DENALI Mini Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Caspian OCT/MESA Mini/DENALI Mini Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Caspian OCT/MESA Mini/DENALI Mini Spinal System may be connected to Range/MESA/DENALI Spinal System and Everest Spinal System components via the rod to rod connectors or transition rods.
    RANGE (MESA and DENALI) and ARI are cleared for the following indications: Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis); tumor, pseudoarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion. Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal System for pediatric use is intended to be used with autograft and/or allograff. Pediatric pedicle screw fixation is limited to a posterior approach.
    The Caspian OCT/MESA Mini/DENALI Mini Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Caspian OCT/MESA Mini/DENALI Mini Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Caspian OCT/MESA Mini/DENALI Mini Spinal System may be connected to Range/MESA/DENALI Spinal System and Everest Spinal System components via the rod to rod connectors or transition rods.
    The YUKON OCT Spinal System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/ or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The YUKON OCT Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the YUKON OCT Spinal System may be connected to Everest Spinal System components via the rod to rod connectors or transition rods.
    The Xia® 3 Spinal System is intended for use in the non-cervical spine. When used as an anterior/ anterolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the Xia® 3 Spinal System is intended to provide additional support during fusion using auto graft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), Spondylolisthesis, Trauma (i.e. fracture of dislocation), Spinal stenosis, Curvatures (i.e., scoliosis, kyphosis, and/or lordosis), Tumor, Pseudarthrosis, Failed previous fusion. The 5.5 mm rods from the Stryker Spine Radius™ Spinal System and 6.0 mm Vitallium rods from the Xia® Spinal System are intended to be used with the other components of the Xia® 3 Spinal System. When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/ or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
    The EVEREST Spinal System may be used in conjunction with the RANGE® (MESA® and DENALI®) Spinal Systems, all of which are cleared for the following indications: Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis); tumor; pseudarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system the EVEREST Spinal System may also be used for the same indications as an adjunct to fusion. When used for posterior non-cervical pedicle screw fixation in pediatric patients the EVEREST Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
    The CASCADIA lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiquous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy. Additionally, the CASCADIA lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. CASCADIA lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine. The CASCADIA hyperlordotic lateral lumbar implants (≥ 22°), are intended for levels L2-L5 and are to be used with CAYMAN United plates in addition to posterior supplemental fixation. The CASCADIA non-hyperlordotic lateral lumbar implants may optionally be used with CAYMAN United plates, in addition to supplemental spinal fixation systems. The CASCADIA cervical implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with cervical disc disease (DDD) at one level or two contiguous levels from C2 to T1. These patients should be skeletally mature and have had six weeks of non-operative treatment. The CASCADIA cervical implants are also to be used with supplemental fixation; the hyperlordotic CASCADIA cervical implants (i.e., ≥ 10°) are required to be used with an anterior cervical plate as the form of supplemental fixation.
    The CAYMAN Buttress Plates are intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications. The CAYMAN Thoracolumbar Plates are indicated for use via the lateral or anterolateral surgical approach in the treatment of thoracic and thoracolumbar (T1-L5) spine and for use as an anteriorly placed supplemental fixation device for the lumbosacral level below the bifurcation of the vascular structures (L5-S1). The Cayman Thoracolumbar Plate System is intended to provide temporary stabilization during fusion using autograph or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities and deformities: a) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), b) pseudoarthrosis, c) spondylolysis, d) spondylolisthesis, e) fracture, f) neoplastic disease, g) unsuccessful previous fusion surgery, h)lordotic deformities of the spine, i) thoracolumbar or lumbar scoliosis, j) deformity (i.e., scoliosis, kyphosis, and/or lordosis) associated with deficient posterior elements such as that resulting from laminectomy.
    The CAYMAN Buttress Plates are intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications. The CAYMAN Thoracolumbar Plates are indicated for use via the lateral or anterolateral surgical approach in the treatment of thoracic and thoracolumbar (T1-L5) spine and for use as an anteriorly placed supplemental fixation device for the lumbosacral level below the bifurcation of the vascular structures (L5-S1). The Cayman Thoracolumbar Plate System is intended to provide temporary stabilization during fusion using autograph or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities and deformities: a) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), b) pseudoarthrosis, c) spondylolysis, d) spondylolisthesis, e) fracture, f) neoplastic disease, g) unsuccessful previous fusion surgery, h )lordotic deformities of the spine, i) thoracolumbar or lumbar scoliosis, j) deformity (i.e., scoliosis, kyphosis, and/or lordosis) associated with deficient posterior elements such as that resulting from laminectomy.
    The CAYMAN LP Plate System is intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications.

    Device Description

    The previously cleared devices consist of a variety of plate and screw systems designed to provide support across implanted levels in the cervical, thoracolumbar, and lumbosacral spine until fusion is achieved. The primary purpose of this submission is to update previously cleared MR safety information, establish an MR Conditional labeling claim, update cleaning, disinfection and sterilization instructions.

    AI/ML Overview

    This appears to be a 510(k) summary for various spinal plate and screw systems. The document focuses on establishing substantial equivalence to predicate devices, primarily through updating MRI safety information, cleaning/disinfection/sterilization instructions, and labeling.

    Unfortunately, this document does not contain the information requested about acceptance criteria and study results for an AI/software-based medical device. The listed devices are physical implants, and the "performance data" section refers to MR compatibility testing, not the performance of an AI algorithm in classification or detection tasks.

    Therefore, I cannot extract the following information from the provided text:

    • A table of acceptance criteria and the reported device performance (for an AI/software device)
    • Sample size used for the test set and the data provenance
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    • Adjudication method for the test set
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    • The type of ground truth used
    • The sample size for the training set
    • How the ground truth for the training set was established

    The document primarily addresses the safety and efficacy of physical spinal implants based on their design, materials, and established predicate devices, and MRI compatibility. It is not about an AI-powered diagnostic or therapeutic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180220
    Date Cleared
    2018-05-21

    (116 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K113666, K071373, K091445, K994121, K111492

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SAGICO OSI SPINAL SYSTEM by Osimplant is intended for use in the non-cervical spine.

    When used for anterior/anterolateral and posterior, non-cervical pedicle fixation system, the SAGICO OSI SPINAL SYSTEM by Osimplant is intended to provide additional support during fusion using autograft or allograft in Skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:

    • · Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
    • · Spondylolisthesis;
    • · Trauma (i.e. fracture or dislocation);
    • · Spinal stenosis;
    • · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis);
    • · Tumor:
    • · Pseudoarthrosis; and
    • Failed previous fusion

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the SAGICO OSI SPINAL SYSTEM by Osimplant implants are indicated as an adjunct to treat adolescent idiopathic scoliosis.

    The SAGICO OSI SPINAL SYSTEM by Osimplant for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The SAGICO OSI SPINAL SYSTEM by Osimplant constructs of monoaxial screws, uniplanar screws, polyaxial screws, reduction screws, locking cap set screws, rods, hooks, monoaxial and muttiaxial transverse connectors and associated surgical instruments. The SACICO OSI SPINAL SYSTEM by Osimplants are available in a variety of sizes to accommodate individual patient anatomy and pathology conditions. SAGICO OSI SPINAL SYSTEM by Osimplant implants are designed to adapt 5.5mm diameter rods; the implants are manufactured from Ti6A14V alloy and offered in a sterile package option.

    AI/ML Overview

    The provided text is a 510(k) summary for the SAGICO OSI SPINAL SYSTEM by Osimplant, a medical device for spinal fixation. It details the device's indications for use, technological characteristics, and performance data used to establish substantial equivalence to predicate devices. However, this document does not describe an AI/ML device or algorithm. It is a clearance for a traditional physical medical implant system consisting of screws, rods, and associated instruments.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, or comparative effectiveness studies for an AI device cannot be extracted from this document, as it pertains to a mechanical spinal implant system.

    The information below is a general template for response if the input data would contain information for an AI/ML powered device, as the prompt is not expecting the type of the device to be considered.

    This document does not describe an AI/ML powered medical device. It describes a traditional, mechanical spinal implant system (SAGICO OSI SPINAL SYSTEM by Osimplant). Therefore, the specific details requested regarding AI/ML device performance, acceptance criteria, sample sizes for AI/ML models, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance are not available in this document.

    The document focuses on demonstrating substantial equivalence of the mechanical implant system to predicate devices based on:

    • Indications for Use: The SAGICO OSI SPINAL SYSTEM is intended for use in the non-cervical spine for various conditions like degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients. For pediatric patients, it's indicated for adolescent idiopathic scoliosis (posterior approach only).
    • Technological Characteristics: The device consists of monoaxial screws, uniplanar screws, polyaxial screws, reduction screws, locking cap setscrews, rods, hooks, monoaxial and multiaxial transverse connectors, and associated surgical instruments. Implants are made from Ti6Al4V alloy and designed for 5.5mm diameter rods.
    • Performance Data (Non-clinical):
      • ASTM F1717-04: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model (Static Compression Bending Test, Dynamic Compression Bending Test).
      • ASTM F1798-13: Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants (Axial Gripping Test, Axial Torsion Gripping Test, Flexion Extension Static Test).
    • Basis of Substantial Equivalence: Comparable data drawn from design, labeling, indications for use, function, and implant materials compared to predicate devices (XIA® 3 Spinal System, USS Small Stature System, CD HORIZON Spinal System, TSRH Spinal System).

    If this were an AI/ML device, the requested information would typically be presented as follows (this is a hypothetical example for illustrative purposes, not based on the provided document):

    1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical for an AI/ML device)

    Performance MetricAcceptance Criteria (e.g., for detecting spinal fractures)Reported Device Performance (e.g., from a clinical study)
    Sensitivity≥ 90%92.5% (95% CI: 90.1% - 94.4%)
    Specificity≥ 80%85.3% (95% CI: 83.0% - 87.3%)
    AUC≥ 0.900.93

    2. Sample Size and Data Provenance for Test Set (Hypothetical)

    • Sample Size: 500 spinal imaging studies (e.g., X-rays, CT scans)
    • Data Provenance: Retrospective data collected from 3 hospitals in the United States and 2 hospitals in Germany.

    3. Number and Qualifications of Experts for Ground Truth (Hypothetical)

    • Number of Experts: 3 independent radiologists
    • Qualifications: All board-certified radiologists with subspecialty training in musculoskeletal radiology and at least 10 years of experience interpreting spinal imaging.

    4. Adjudication Method for Test Set (Hypothetical)

    • Adjudication Method: 2+1. Initial assessment by two radiologists. Discrepancies were resolved by a third, senior radiologist.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Hypothetical)

    • Was an MRMC study done? Yes.
    • Effect Size: In the MRMC study, human readers (radiologists) demonstrated a statistically significant improvement in fracture detection accuracy (AUC increased by an average of 0.05, p
    Ask a Question

    Ask a specific question about this device

    K Number
    K143683
    Manufacturer
    Date Cleared
    2015-05-22

    (149 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APEX-DL Spine System with APEX Spine System Components is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudoarthrosis).

    The APEX-DL Spine System with APEX Spine System Components is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.

    The APEX-DL Spine System with APEX Spine System Components is also a sacraliliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).

    When used in a percutaneous posterior approach with AIM MIS instrumentation, the APEX-DL Spine System with APEX Spine System Components is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.

    When used for posterior non-cervical pediatic patients, the APEX-DL Spine System implants with APEX Spine System Components are indicated as an adjunct to treat adolescent idiopathic scoliosis. The APEX DL Spine System is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The APEX-DL Spine System includes Monoaxial, Uniplanar, and Polyaxial Double Lead Thread Screws in the cannulated and non-cannulated versions and in regular and reduction (extended tab) versions. APEX-DL Spine System is a low profile thoracolumbar implant for use with wide range of patient statures. The APEX-DL Spine System Polyaxial screws feature a friction head, which is designed to provide precise reduction mechanism as a result of easier rod capturing. The APEX-DL Spine System also includes Lordosed Percutaneous Rods. The APEX-DL Spine System is compatible with the APEX Spine System 5.5mm and 6.0mm rods, hooks, side-by-side connectors, iliac connectors, cross connectors, and washers.

    AI/ML Overview

    The APEX-DL Spine System is a spinal implant for immobilization and stabilization of spinal segments. The document indicates that no clinical studies were performed. The acceptance criteria and testing are based on non-clinical (mechanical) tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance CriteriaReported Device Performance
    ASTM F1717 (Vertebrectomy Model)
    Static Compression BendingNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Static TorsionNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Dynamic Compression BendingNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    ASTM F1798 (Interconnection Mechanisms)
    Static Axial Gripping CapacityNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Axial Torque Gripping CapacityNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Static Flexion-ExtensionNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"
    Dynamic Flexion-ExtensionNot explicitly stated (comparison to predicates)Results were "equal or higher than the predicate systems"

    Note: The acceptance criteria are implicitly defined as demonstrating mechanical properties that are "equal or higher" than the identified predicate devices, based on the testing standards.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in the provided document. For each of the ASTM tests listed, a specific number of samples (implants/constructs) would have been used, but this detail is not provided.
    • Data Provenance: The data is from non-clinical (mechanical) testing conducted presumably by SpineCraft, LLC or a contracted testing facility. It is not patient or human data; therefore, country of origin or retrospective/prospective does not apply in the typical sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a non-clinical, mechanical testing study, not a study requiring expert clinical assessment or ground truth establishment in a medical imaging or diagnostic context. The "ground truth" here is the objective measurement of mechanical properties according to established ASTM standards.

    4. Adjudication Method for the Test Set

    Not applicable. This is a non-clinical, mechanical testing study. Adjudication methods like 2+1 or 3+1 are used for expert consensus in clinical or imaging studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is a clinical study involving multiple human readers interpreting medical cases. The provided document explicitly states, "No clinical studies were performed."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical spinal implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on objective mechanical measurements according to recognized industry standards (ASTM F1717 and ASTM F1798). The performance is then compared to "predicate systems" as a benchmark for substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device undergoing mechanical testing, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As described in point 8, there is no "training set" for this type of device evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150417
    Manufacturer
    Date Cleared
    2015-05-19

    (90 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ASTRA System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudo-arthrosis).

    The ASTRA Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 to S1), and for whom the device is intended to be removed after solid fusion is attained.

    The ASTRA System is also a sacral/iliac screw fixation system of the non-cervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudo-arthrosis).

    When used in a percutaneous, posterior approach with AVANT Spine MIS instrumentation, the ASTRA Spine System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis), tumor, pseudo-arthrosis, and failed previous fusion in skeletally mature patients . Levels of fixation are for the thoracic, lumbar and sacral spine.

    Device Description

    The ASTRA Spine System is a top loading, multiple component, posterior spinal fixation system which consists of rods, cannulated and non-cannulated monoaxial, uniplanar and polyaxial screws, hooks, iliac connectors, rod connectors, and cross connectors. Most of the components are available in a variety of sizes to more closely match the patient's anatomy.

    Materials:
    Titanium alloy
    CoCr alloy

    AI/ML Overview

    Here's an analysis of the provided text regarding the ASTRA Spine System, focusing on acceptance criteria and study details.

    Important Note: The provided document is a 510(k) summary for a medical device (Spine System), not an AI/ML device. Therefore, many of the requested fields related to AI/ML specific studies (like sample sizes for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance) are not applicable to this type of medical device submission. This document focuses on demonstrating substantial equivalence to predicate devices through non-clinical (mechanical) testing.


    1. Table of Acceptance Criteria and Reported Device Performance

    For this medical device, the "acceptance criteria" are implicitly defined by performance equivalence to predicate devices in standardized mechanical tests. The "reported device performance" is a statement of comparative equivalence.

    Acceptance Criteria (Implicit)Reported Device Performance (Summary)
    Static Compression Bending: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Static Torsion: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Dynamic Compression Bending: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Axial Gripping: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Torsional Gripping: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Static Flexion-Extension: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Dynamic Flexion-Extension: Must perform at least equivalent to predicate systems."The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems."
    Overall Substantial Equivalence: Must be substantially equivalent in intended use, design, material, performance, and function."The ASTRA Spine System is substantially equivalent to the predicate devices in terms of indications for use, design, material, performance and function."

    2. Sample Size Used for the Test Set and the Data Provenance

    This is a physical device, and the "test set" refers to the tested device components.

    • Sample Size: Not explicitly stated in terms of number of components tested for each test, but standard engineering practices for medical device testing would involve a sufficient number (e.g., n=5 or n=10 per test) to ensure statistical significance, though the exact numbers are not provided in this summary.
    • Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting. There is no patient data provenance (e.g., country of origin, retrospective/prospective) since no human data was used for these mechanical tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not Applicable. For mechanical testing of a physical device, ground truth is established by standardized testing protocols (e.g., ASTM standards) and measured physical properties, not by expert consensus on clinical data. Engineers and lab technicians perform and analyze the tests according to these standards.

    4. Adjudication Method for the Test Set

    • Not Applicable. There is no adjudication in the sense of reconciling human expert opinions for clinical images or data. The results of mechanical tests are objective measurements against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • Not Applicable. This is a physical spinal implant, not an AI/ML diagnostic or assistive device. No MRMC studies were performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This is a physical spinal implant; there is no algorithm or standalone performance.

    7. The Type of Ground Truth Used

    • Mechanical Test Standards and Predicate Device Performance: The "ground truth" for the non-clinical tests is based on the established performance characteristics and safety profiles of legally marketed predicate devices, as defined by FDA regulations for substantial equivalence, and adherence to relevant ASTM standards (ASTM F1717 and ASTM F1798).

    8. The Sample Size for the Training Set

    • Not Applicable. This is a physical device; there is no AI/ML model or "training set" in the context of machine learning. The device design and materials are based on engineering principles and knowledge of predicate devices.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set for an AI/ML model, this question is not relevant. The "ground truth" for developing the physical device is based on established biomechanical and medical understanding of spinal fixation and the performance of existing, cleared devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142381
    Manufacturer
    Date Cleared
    2014-10-21

    (56 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    CD HORIZON® Spinal System: K140276, Stryker Spine Xia® 3 Spinal System: K113666, Stryker Spine Xia® 4.5

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xia® 3 Spinal System is intended for use in the non-cervical spine. When used as an anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the Xia® 3 Spinal System is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:
    • Degenerative Disc Disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    • Spondylolisthesis
    • Trauma (i.e. fracture of dislocation)
    • Spinal stenosis
    • Curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
    • Tumor
    • Pseudarthrosis
    • Failed previous fusion
    The 5.5 mm rods from the Stryker Spine Radius™ Spinal System and 6.0 mm Vitallium rods from the Xia® Spinal System are intended to be used with the other components of the Xia® 3 Spinal System.
    When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The Xia® 4.5 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications:
    • Degenerative Disc Disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    • Spondylolisthesis
    • Trauma (i.e. fracture of dislocation)
    • Spinal stenosis
    • Curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
    • Tumor
    • Pseudarthrosis
    • Failed previous fusion
    The Stryker Spine DIAPASON™ Spinal System, Opus™ Spinal System, and Xia® 4.5 Spinal System can be linked to the Xia® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion.
    Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the Xia® 4.5 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis. kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 4.5 Spinal System is intended to treat pediatic patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description
    1. Xia® 3 Spinal System
      The Xia® 3 Spinal System is comprised of screws, blockers, and hooks that affix rods and connectors to vertebrae of the spinal column for purposes of stabilization, or corrective action through the application of force.

    2. Xia® 4.5 Spinal System
      The Xia® 4.5 Spinal System is comprised of monoaxial and polyaxial bone and reduction screws, hooks, dual staples, and blockers that affix rods, rod-to-rod connectors, and cross connectors to vertebrae of the spinal column.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Stryker Spine Xia® 3 and Xia® 4.5 Spinal Systems. It does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria through testing or data analysis.

    Instead, this document focuses on demonstrating substantial equivalence to previously cleared predicate devices for an expansion of indications for use.

    Here's why and what information is available:

    The document explicitly states:

    • "This submission seeks to build upon predicate Xia® 3 and Xia® 4.5 Spinal Systems through expansion of indications only. Therefore, no additional performance data is necessary." (Page 6, "Summary of Performance Data")

    This means the submission is not presenting new performance studies or acceptance criteria for a novel device or a significantly modified device. Instead, it's arguing that the expanded indications for the Xia® 3 and Xia® 4.5 Spinal Systems are substantially equivalent to the indications already cleared for a primary predicate device (Medtronic Sofamor Danek, CD HORIZON® Spinal System cleared under K140276) and previous versions of the Xia® systems.

    Therefore, for the information requested:

    1. A table of acceptance criteria and the reported device performance: This information is not provided in the document as no new performance data was deemed necessary for this submission. The device is implicitly accepted based on its substantial equivalence to predicate devices, which would have undergone performance testing during their original clearance.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as no new performance studies were conducted for this 510(k). The document only mentions predicate device clearances.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a spinal fixation system, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
    8. The sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, this 510(k) relies on the safety and effectiveness demonstrated by its predicate devices for its original indications and argues that the expanded indications are within the scope of what was already established for similar, legally marketed devices. It is a regulatory submission for a spinal implant, which typically doesn't involve the kind of performance studies (e.g., diagnostic accuracy, reader studies) that would generate the detailed data requested in your prompt regarding acceptance criteria, sample sizes, and expert ground truth for an AI/diagnostic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140678
    Date Cleared
    2014-06-19

    (93 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CosmoLock Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The CosmoLock Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The CosmoLock Pedicle Screw System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

    Materials:

    Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

    AI/ML Overview

    This document is a 510(k) Summary for the CosmoLock Pedicle Screw System. It describes the device and its intended use and claims substantial equivalence to predicate devices. However, it does not contain the detailed information needed to answer many of your questions, as it explicitly states that no clinical studies were performed.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test)Reported Device Performance
    Static compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Dynamic compression per ASTM F1717Equivalent to predicate devices (implied: met criteria)
    Static torsion per ASTM F1717Equivalent to predicate devices (implied: met criteria)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The studies mentioned are non-clinical (mechanical tests), and details about sample sizes for these tests are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. No clinical studies were performed, and therefore no ground truth established by experts is relevant to the data presented.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No clinical studies were performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This document states, "No clinical studies were performed."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a mechanical pedicle screw system, not an AI or algorithm-based device. No standalone performance study in this context was performed.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth for the non-clinical tests would be the established mechanical testing standards (ASTM F1717) and the performance of the predicate devices. There is no biological or expert-defined "ground truth" in the typical medical imaging/diagnosis sense, as this is a mechanical implant.

    8. The sample size for the training set

    This information is not applicable/provided. No clinical studies were performed, and there is no mention of an algorithm or AI requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. No training set or associated ground truth was established, as no clinical studies or AI development are described. The focus is on demonstrating mechanical equivalence to predicate devices.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthrosis; and Failed Previous Fusion.

    The Radius® Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation system, the Radius® Spinal system is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Radius® Spinal System can also be linked to the XIA® Titanium Spinal System via the Ø5.5mm to Ø6.0mm Radius® rod-to-rod connector.

    The Stryker Spine TRIO® Plate System is intended for posterior, noncervical (T10-S1) pedical and nonpedical fixation of the spine for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.

    The Stryker Spine TRIO® Spinal Fixation System is intended for posterior, noncervical pedicle and non-pedicle fixation of the spine. The Stryker Spine TRIO® Spinal Fixation System is indicated for: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® Spinal Fixation Sytem is intended to be used in conjunction with the OSS Diapason Rods, Opus Spinal System Rods, and the Multi-Axis Cross Connectors.

    The Stryker Spine TRIO® Spinal System is intended for posterior, noncervical pedicle and nonpedicle fixation of the spine to provide immobilization and stabilization of spinal segements in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® + Spinal System is intended to be used in conjunction with the OSS/Diapason Rods or Opus Rods, XIA® Pre-bent Rods, and the Multi-Axis Cross Connectors.

    The Stryker Spine TRIO® TRAUMA Spinal System is intended for percutaneous, posterior, non-cervical pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.

    The XIA® Spinal System and XIA® 4.5 Spinal System are intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DDD) (Defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e., fracture or dislocation); Spinal Stenosis; Curvatures (i.e., Scoliosis, Kyphosis, and/or Lordosis); Tumor; Pseudoarthrosis and; Failed previous fusion. The 6mm diameter rods from the DIAPASON® Spinal System and OPUS® Spinal System are intended to be used with the other components of the XIA® Titanium Spinal System. The Titanium Multi-Axial Cross Connector are intended to be used with the other components of the XIA® Titanium Spinal System.

    The XIA® 3 Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle fixation system, the XIA® 3 Spinal System is intended to provide additional support during autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor, Pseudoarthorisis; and Failed Pervious Fusion. The Ø5.5mm rods from the Stryker Spine Radius® Spinal System and the Ø6.0mm Vitallium Rods from the XIA® Spinal System are intended to be used with the other components of the XIA® 3 Spinal System. When used for posterior noncervical pedicle screw fixation in pediatric patients, the XIA® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. The XIA® 3 Spinal System for pediatic use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The XIA® 4.5 Spinal System is intended for anterior/anteriolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. Except for the staples, the XIA® 4.5 Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical pedicle screw fixation in pediatric patients. The XIA® 4.5 Spinal System for pediatric use in intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The STRYKER Spine thoraco-lumbar spinal fixation systems, subject of this 510(k), are non-cervical, pedicle and non-pedicle fixation systems comprised of screws, rods, plates, hooks, connectors, washers and staples. The components are manufactured from either Titanium (Titanium Alloy and CP Titanium), Stainless Steel or Cobalt-Chromium-Molybdenum Alloy (Vitallium®). This Special 510(k) submission seeks clearance for sterile labeling of the listed STRYKER Spine thoraco-lumbar spinal fixation systems. All of the components of the subject devices will be sterilized by gamma radiation, a traditional sterilization method as per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.

    AI/ML Overview

    This is a 510(k) Premarket Notification for sterile packaging of a range of spinal systems. The key claim is substantial equivalence to previously cleared non-sterile versions of the same devices. Therefore, the "device" in question for this submission is specifically the sterile packaging and the sterilization process, not the spinal implant itself as a medical device performing a physiological function. The "performance" being evaluated is the ability to maintain sterility and the structural integrity of the device within its packaging over its shelf life.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Sterilization EffectivenessDemonstrated by gamma radiation, a traditional sterilization method per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.
    Sterile Barrier Packaging SystemDemonstrated effective in maintaining sterility for the recommended 5-year shelf-life.
    Impact on Device CharacteristicsNo design modifications were incorporated to facilitate sterile packaging. The sterile-packed implant components maintain the same technological characteristics (design, technical requirements, materials of construction, indications/intended use) as the non-sterile predicate devices.
    Shelf LifeDemonstrated a 5-year shelf life in sterile packaging.
    Safety and EffectivenessConcluded to be safe and effective, retaining the same intended use, indications, technological characteristics, and mode of operation as the predicate non-sterile devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses the sterilization process and packaging rather than clinical performance of the implant itself. Therefore, typical "test set" in the context of clinical or diagnostic AI/software devices (e.g., patient data) is not applicable here.

    • Sterilization and Packaging Validation Data: The document mentions "accelerated aging data" for validating the 5-year shelf life. It does not specify the sample size for this testing, nor the country of origin or whether it was retrospective or prospective. However, these types of tests are typically prospective laboratory studies conducted in controlled environments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This submission is about the sterility and packaging of medical implants, not a diagnostic or clinical performance study that would require expert ground truth labeling. The "ground truth" here is the scientific validation of sterilization methods and shelf life according to recognized standards and FDA guidance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used in clinical studies for establishing ground truth, which is not relevant to this submission on sterile packaging.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this submission relies on:

    • Scientific Validation of Sterilization: Adherence to established scientific principles and FDA guidance for sterilization (gamma radiation), demonstrating a Sterility Assurance Level (SAL).
    • Accelerated Aging Data: Laboratory testing to simulate real-time aging and demonstrate that the sterile barrier packaging system maintains sterility for the stated shelf life.
    • Material and Design Specification Conformity: Confirmation that the components are manufactured from specified materials and that their design, technical requirements, and intended use are identical to the non-sterile predicate devices.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not a machine learning or AI device, there is no training set and therefore no ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130932
    Manufacturer
    Date Cleared
    2013-08-07

    (125 days)

    Product Code
    Regulation Number
    888.3070
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle, and nonpedicle fixation (T1-S2/Ilium). Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications:

    • . degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
    • . spondy lolisthesis,
    • . trauma (i.e., fracture or dislocation),
    • . spinal stenosis,
    • . deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
    • . tumor;
    • . pseudoarthrosis, and
    • failed previous fusion .
      When used for fixation to the ilium, the offset connectors of the Firebird Spinal Fixation System must be used in conjunction with pedicle screws placed at the S1 or S2 spinal level.
      The Phoenix MIS Fixation System when used with the Firebird Spinal Fixation System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
      The Firebird Spinal Fixation System components are used with certain components of the Orthofix Spinal Fixation System, including rods, rod connectors and cross-connectors.
      When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Firebird Spinal Fixation System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Firebird Spinal Fixation System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
    Device Description

    The Firebird Spinal Fixation System is a temporary, titanium alloy, multiple component system comprised of a variety of non-sterile, single use components, made of titanium alloy or cobalt chrome alloy, that allow the surgeon to build a spinal implant construct. The system is attached to the vertebral body and ilium by means of screw or hook fixation to the non-cervical spine. The Firebird Spinal Fixation System consists of an assortment of rods, multi-axial and mono-axial pedicle screws, set screws, lateral offsets, bone screws, screw bodies, hooks and iliac connectors. The expansion of indications for the Firebird Spinal Fixation System is proposed for the inclusion of adolescent idiopathic scoliosis alone, and not other indications for a pediatric population.
    A subset of the Firebird Spinal Fixation System and Phoenix MIS System components may be used in pediatric patients. These components consist of a variety of screws ranging in diameters from 4.5mm to 7.5mm and lengths ranging from 25mm to 60mm.

    AI/ML Overview

    The provided document is a 510(k) summary for the Orthofix Firebird Spinal Fixation System. This type of regulatory submission (510(k)) is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific acceptance criteria as you might see for a novel device or a PMA.

    Therefore, the document does not contain information about acceptance criteria and a study proving a device meets them in the way a diagnostic AI device or a novel therapeutic device would. Instead, it focuses on demonstrating equivalence to existing devices based on design, materials, indications for use, and mechanical performance.

    I will attempt to extract the closest equivalents to your requested information where possible, based on the context of a 510(k) submission.


    Analysis of the Provided Document Regarding Acceptance Criteria and Device Performance

    The Orthofix Firebird Spinal Fixation System is a Class III Preamendment Device (Pedicle screw spinal system) for which the manufacturer is seeking 510(k) clearance. A 510(k) clearance means the device is "substantially equivalent" to predicate devices. This type of submission generally relies on demonstrating that the new device has the same intended use, technological characteristics, and performs as safely and effectively as a legally marketed device. It typically does not involve traditional clinical studies with acceptance criteria for device classification as one would expect for a novel device or an AI/diagnostic product.

    Therefore, many of the requested points below are not directly applicable or quantifiable from this document.


    1. A table of acceptance criteria and the reported device performance

      Not applicable in the context of this 510(k) submission. The performance assessment is focused on "substantial equivalence" to predicate devices, primarily through mechanical testing and engineering analysis rather than specific clinical acceptance criteria.

      The document states:
      "Previous mechanical testing of the system included static and dynamic compression bending testing and static torsion testing per ASTM F1717-04 and finite element analysis."

      While these tests have criteria for compliance with the standard, the document does not present a table of these criteria alongside specific device performance data. The conclusion drawn is that the device is substantially equivalent based on these results.

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

      Not explicitly stated for the mechanical testing mentioned. When mechanical testing is performed, the "sample size" refers to the number of physical devices or components tested. The data provenance is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

      Not applicable. Mechanical testing does not involve "experts" establishing ground truth in the way a diagnostic study would. The standards for mechanical testing (e.g., ASTM F1717-04) define the methods and parameters.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

      Not applicable. Mechanical testing does not involve adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      Not applicable. This is a spinal fixation system, not a diagnostic or AI-powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      Not applicable. This is a spinal fixation system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      For mechanical testing, the "ground truth" is defined by the physical properties and performance limits set by engineering standards (e.g., ASTM F1717-04). For the expansion of indications (specifically for adolescent idiopathic scoliosis), the document mentions: "Published clinical results and engineering analysis supported expansion of indications." This implies that existing clinical literature for similar predicate devices, combined with engineering analysis (which would include the mechanical testing), served as the basis for justifying the expanded indications. Outcomes data or expert consensus from previously published studies might have indirectly informed this, but no new study for this submission is detailed.

    8. The sample size for the training set

      Not applicable. This is a spinal fixation system, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established

      Not applicable. This is a spinal fixation system, not a machine learning model.


    Summary regarding the Firebird Spinal Fixation System and 510(k) process:

    The K130932 submission for the Firebird Spinal Fixation System is a 510(k) premarket notification. The core of this submission is to demonstrate substantial equivalence to existing predicate devices. This is achieved by showing that the device:

    • Has the same intended use.
    • Has the same basic technological characteristics (design, materials: titanium alloy per ASTM F136 and cobalt chrome per ASTM F1537, as with predicates).
    • Performs as safely and effectively as the predicate devices.

    The "proof" the device meets acceptance criteria in this context relies on:

    • Mechanical Testing: Static and dynamic compression bending, and static torsion testing per ASTM F1717-04.
    • Finite Element Analysis (FEA).
    • Comparison to existing predicate devices: K081684, K082797, K100044, K093926, K122901 (Orthofix Inc.), K113666 (Stryker Spine), K091445 (Medtronic Sofamor Danek USA), K994121 (Synthes Spine), K121630 (K2M, Inc.), K111492 (Medtronic Sofamor Danek USA).

    The document states: "Previous testing performed on this device indicates that the Firebird Spinal Fixation System is substantially equivalent to predicate devices." and "Published clinical results and engineering analysis supported expansion of indications." This means the mechanical performance met relevant engineering standards, and the safety and efficacy for the expanded indications (specifically pediatric adolescent idiopathic scoliosis) were supported by existing clinical literature and the device's engineering characteristics being similar to already approved predicate devices.