510(k) Data Aggregation
(219 days)
NobelActive® S TiUltra™, NobelActive® S TiUnite®
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function.
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are indicated for single or multiple unit restorations in splinted or non-splinted applications.
This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
NobelReplace® S TiUltra™, NobelReplace® S TiUnite®
NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are indicated for single or multiple unit restorations.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants can be used in splinted or non-splinted applications.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.
NobelParallel™ S TiUltra™, NobelParallel™ S TiUnite®
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function.
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are indicated for single or multiple restorations in splinted or non-splinted applications.
This can be achieved by 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
Implants with <7 mm length are for delayed loading only when appropriate stability has been achieved.
The Subject Device, Nobel Biocare S Series Implants, comprises of six Device Lines NobelActive S TiUltra™, NobelParallel™ S TiUltra™, NobelReplace S TiUltra™, NobelActive S TiUnite, NobelParallel™ S TiUnite, NobelReplace S TiUnite.
Nobel Biocare S Series Implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental prostheses to restore chewing function. The Subject Device features a consistent diameter at the implant-abutment interface, regardless of the implant size or type. This means that the prosthetic components, such as abutments and restorative elements, have uniform dimensions and connections across all implant sizes within the system. The connection used is the Narrow Platform (NP) conical connection.
The implant bodies of the Subject Devices and Predicate devices are the same. This allows use of the same drills and drilling protocols, rescue procedures and rescue tools. By using the existing Narrow Platform connection, the same prosthetic workflows, prosthetic components, and lab components can be used.
The implants are available in different sizes as listed in Table 1.
The Nobel Biocare S Series Implants are manufactured from commercially pure titanium and incorporate a conical connection (Size Narrow Platform (NP)) with a hex interface in the collar region, which combines with Nobel Biocare's existing prosthetic Abutment portfolio cleared in K071370, K083100, K132746, K233208, K202344, K200040, K132749, K161416, K202452, K240346, K041236, K133731, K061003 and K161435.
Table 2 outlines which single-unit restoration abutments with an NP connection can be used with the subject device.
Table 3 outlines which multi-unit restoration abutments and bridges with an NP connection can be used with the subject devices.
Table 4 outlines the temporary and healing restorations with an NP connection that can be used with the subject devices.
Table 5 lists the corresponding submissions where compatible Class II accessories with an NP connection were previously identified.
The On1™ Base/On1™ Base Xeal must not be used in combination with NobelActive® S, NobelParallel™ S, NobelReplace® S implants.
NobelActive® S, NobelParallel™ S, NobelReplace® S implants in combination with NobelProcera® Angulated Screw Channel Abutment and NobelProcera® FCZ Implant Crown are indicated for maxillary lateral and mandibular central/lateral incisors only.
NobelActive® S, NobelParallel™ S, NobelReplace® S implants may be used in combination with NobelProcera® Angulated Screw Channel Abutment and NobelProcera® FCZ Implant Crown up to a maximum angulation of 20°
NobelActive® S, NobelParallel™ S, NobelReplace® S implants may be used in combination with GoldAdapt Engaging CC NP abutments in straight (0°) configurations only.
No angulation is allowed for Universal Base Conical Connection Abutments when mated with the NobelActive® S, NobelParallel™ S, NobelReplace® S implants with diameters greater than 3.5mm.
N/A
Ask a specific question about this device
(115 days)
IPDmilled Blanks are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed CAD/CAM customizations for the IPDmilled Blanks are to be designed and manufactured according to digital dentistry workflow or to be sent to an IPD validated milling center for manufacture. The workflow system integrates multiple components of the digital dentistry workflow: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, milling machine and associated tooling and accessories. Compatible Implant Systems: Dentium Co., Ltd Implantium, Friadent Implant Systems, Neoss ProActive Implant, SPI® Dental Implant, Inicell®, BEGO Semados® S-Line, ANKYLOS® C/X Implant System, MIS V3 Conical Connection Dental Implant System, Conical Connection Implants (MIS® C1), MIS C1 Narrow Platform Conical Connection Implant System, MIS C1 Wide Platform Conical Connection Abutments, PRAMA White Implant Systems, Altatec GmbH CAMLOG®/CONELOG® PROGRESSIVE-LINE Implants, Straumann TLX Implant System, Kontact Dental Implant System, ICX-Implant System, Tapered Pro Conical Implant System, Straumann® Tissue Level, Zimmer Tapered Screw-Vent®, Nobel Biocare® Nobel Active®, Straumann® Bone Level Tapered Implants, Neodent Implant System – GM Line, Osstem Implant System, Xpeed AnyRidge Internal Implant System, 3i Osseotite® Certain® Dental Implants, Astra Tech Implant System (Osseospeed®), OsseoSpeed™ Plus, BioHorizons Tapered Internal Implant System, MIS Internal Hex Dental Implant System (MIS® Seven®), Straumann BLX Ø3.5 mm Implants, Straumann® BLX Implant System, ALTATEC Camlog Screwline Implant System, Replace TiUnite Endosseous Implant.
The purpose of this submission is to expand abutment categories of IPD Dental Implant Abutments, which were previously cleared.
IPDmilled Blanks is a dental implant abutment system composed of dental abutments and screws intended to be placed into dental implants to provide support for dental prosthetic restorations. Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems. IPDmilled Blanks have a pre-manufactured connection interface that fits directly with a pre-specified dental implant. The customized shape of the abutment is intended to be designed and manufactured according to a digital dentistry workflow or to be sent to an IPD validated milling center for manufacture. IPDmilled Blank is delivered non-sterile and the final restoration, including the screw, is intended to be sterilized at the dental clinic before it is placed in the patient.
The IPDmilled Blanks are blistered together with their specific screw. The screws are intended to attach the prosthesis to the dental implant.
The metallic components of the subject abutments and screws are made of titanium alloy conforming to ISO 5832-3 "Implants for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy".
The design and manufacturing of the custom abutment will be conducted using a digital dentistry workflow requiring the use of the following equipment and software:
Intraoral Scanner: 3Shape TRIOS A/S Series
Desktop Scanner: 3Shape E Series.
Design Software: 3Shape Abutment Designer Software, K151455.
AbutmentCAD, K193352.
Milling unit: CORiTEC 350i PRO / CORiTEC 350i Loader PRO with MillBox CAM software.
IPDmilled Blank libraries' have built-in design limitations, and the user isn't allowed to exceed these limitations. The custom abutment design limitation specifications are as follows:
Minimum gingival height: 0.5 mm
Maximum gingival height: 6.0 mm
Minimum wall thickness: 0.4 mm
Minimum post height: 4.0 mm
All IPDmilled Blanks are for straight abutments only.
N/A
Ask a specific question about this device
(307 days)
NizPlant Implants are a one-piece implant or two-piece assembly for single-stage surgical procedures intended for surgical implantation in edentulous or partially edentulous mandibles or maxilla for attachment of complete denture prostheses, or for fixed or removable bridgework for multi-unit tooth replacement.
NizPlant Implants can be indicated for immediate function when initial implant stability has been achieved and with appropriate occlusal loading.
Short (<9mmL) Implants: Indicated for multiple-unit restorations of partially or fully edentulous mandibles and maxillae where vertical bone volume is limited.
The NizPlant Implants have a 2.5mmD internal hex within its most coronal portion for the insertion tool placement used in the full seating of the NizPlant Implant in the osteotomy. Implant placement, which is designed to receive multiple abutment variations, expands its restorative options allowing for the implant to be used for:
• Support of attachment-retained overdentures
• Support of screw-retained prostheses
The NizPlant Dental Implant System consists of a root form Endosseous dental implant and various other prosthetic/restorative components (medical devices) to plan, implant, and support restorations for edentulous patients.
NizPlant Implants are a one-piece implant or two-piece assembly for single-stage surgical procedures intended for surgical implantation in edentulous or partially edentulous mandibles or maxilla for attachment of complete denture prostheses, or for fixed or removable bridgework multi-unit tooth replacement.
The top one-third of the NizPlant Implant body has a straight machined neck that is placed in crestal bone. The depth gauge lines at 1mm, 2mm and 2.5mm from the top of the straight machined neck portion of the implant body facilitate the placement level with the crest of the ridge or 1mm above the crest of the ridge. Above the straight neck of the implant body is a dual-function platform which acts as an overdenture or a multi-unit abutment.
The lower two-thirds of the Implant is tapered with double-lead progressively deeper reverse buttress threads for increased bone contact. The two full-length cutting grooves are self-tapping and extend over the tapered portion of the implant's body.
The NizPlant dental implant body is available in six diameter sizes. This assortment of implants in various diameters and lengths are to be matched with an extensive componentry offering. NizPlant Implants lengths are measured from the apex of the implant to the midpoint of the gold anodized zone, that is to the base of the abutment portion of the NizPlant Implant.
| Body Diameters | Platform Diameters | External Platform Diameter | Lengths |
|---|---|---|---|
| 3.0mm | 3.5mm | 3.9mm | 9mm, 11mm, 12.5mm, 14mm |
| 3.7mm | 3.5mm | 3.9mm | 7mm, 9mm, 11mm, 12.5mm, 14mm |
| 4.2mm | 3.5mm | 3.9mm | 7mm, 9mm, 11mm, 12.5mm, 14mm |
| 4.7mm | 3.5mm | 3.9mm | 7mm, 9mm, 11mm, 12.5mm, 14mm |
| 5.2mm | 3.5mm | 3.9mm | 7mm, 9mm, 11mm, 12.5mm, 14mm |
| 5.7mm | 3.5mm | 3.9mm | 7mm, 9mm, 11mm, 12.5mm, 14mm |
NizPlant Implants have a portion of their coronal surface anodized a gold color. The gold color aids aesthetically as it is the portion that extends transmucosal from bone level, through to the oral cavity by masking of the underlying color of the non-anodized standard titanium grey. Due to its length, the 7mmL and 9mmL implants have one less circumferential groove and are replaced with an additional external thread to increase fixation within the bone.
NizPlant Implants are specifically designed with a combination of 0.5mm and/or the 1mm height circumferential grooves above the blasted implant surface (7mmL - 9mmL implant does not have the additional 0.5mm circumferential score line). These visual demarcations aid the clinician by providing the option for vertical height variability during placement. For restorations where the vertical position of the surrounding bone and/or soft tissue dictate that, part of the implant's anodized portion within the straight machined neck of the NizPlant Implant body may remain supra-crestal after placement (in addition to the 3mm prosthetic option of the dual function platform).
The NizPlant Implants are anodized a gold color with a portion of the Implant going through a surface treatment (blasting) to create a microtextured surface roughness. These implants (and the devices included in the packaging) are sold sterile. The NizPlant Implants are manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant.
The various devices included in the NizPlant Implant System (such as Endosseous dental implant abutments, cover screws, fixation screw, etc.) and various other prosthetic/restorative components used to plan, implant, and support restorations for fully and partially edentulous patients are offered in compatible platform dimensions to enable a complete dental restoration.
The Cover Screws are available in 5.5mm diameter with a platform diameter of 3.5mm. They are manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant).
The Titanium Non-Engaging Temporary Abutments are available in 5.0mm diameter with a platform diameter of 5.5mm. This is a straight configuration. They are manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant) and anodized a gold color.
The Non-Engaging Angled Screw Channel (ASC) Abutments are available in 4.7mm diameter with a platform diameter of 5.0mm and length of 0.5mm. This is a straight configuration. They are manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant) and anodized a gold color.
The Straight Multi-Unit (MUA) Abutments are available in 3.2mm diameter with a platform diameter of 4.8mm and length of 3.2mm. This is a straight configuration. They are manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant) and anodized a gold color.
The Fixations Screws are all manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant).
N/A
Ask a specific question about this device
(90 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes dental implants, abutments, and prosthetic components cleared previously in various submissions. The previously cleared abutments and prosthetic components are compatible with a variety of original equipment manufacturer (OEM) dental implants as well as DESS Dental Smart Solutions dental implants.
This submission adds various abutments to the DESS and OEM implant lines as summarized on the following pages in Table 1 Summary of Subject Device Abutment Designs, and Table 2 Summary of Subject Device Abutment Sizes.
The subject device abutment designs include Cover Screws, Healing Abutments, Temporary Abutments, Ti Base Abutments, AURUM Base Abutments (Ti Base abutments with a screw channel design that allows easier instrument access to the abutment screw), CoCr Base Abutments, Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), DESSLoc Abutments (Locator-type abutments), and abutment screws.
This document is a 510(k) summary for the DESS Dental Smart Solutions, a line of endosseous dental implant abutments. It details the device's characteristics, intended use, and a comparison to predicate devices to establish substantial equivalence.
Based on the provided text, the device itself is a physical medical device (dental implant abutments and associated components), not a software or AI-driven system. Therefore, the questions related to AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, and ground truth establishment for AI would not be directly applicable to this product as described.
The document focuses on establishing substantial equivalence to previously cleared predicate devices through:
- Identical Intended Use: The device is intended for "functional and esthetic rehabilitation of the edentulous mandible or maxilla" by providing support for prosthetic restorations, which is the same as the predicate devices.
- Similar Technological Characteristics: The device utilizes similar designs, materials (Ti-6Al-4V, Co-Cr-Mo alloy), manufacturing processes, and sterilization methods as its predicates.
- Performance Data: Non-clinical testing (mechanical testing per ISO 14801, MR environment assessment) and comparison to existing data from predicate devices are used to demonstrate safety and effectiveness.
Therefore, many of the requested elements for describing AI acceptance criteria and studies are not present or applicable in this document.
However, I can extract information relevant to the device's performance assessment and criteria for its type of submission.
Here's an attempt to answer the questions based solely on the provided text, acknowledging that the nature of the device (a physical implant component) means many AI-specific questions will be answered as "Not Applicable" (N/A):
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantifiable "acceptance criteria" in a table format with corresponding "reported device performance" metrics for this specific 510(k) submission. Instead, it relies on demonstrating substantial equivalence to existing predicate devices.
The underlying "acceptance criteria" for demonstrating substantial equivalence for this device type are primarily through:
- Mechanical Testing (ISO 14801): This is a critical performance standard for dental implants and abutments. The document states that mechanical testing was "conducted according to ISO 14801 to support the performance." The acceptance criteria would be successful completion of these tests, demonstrating the device's mechanical strength and fatigue resistance are comparable to or better than predicate devices. The reported performance is simply that the tests supported the performance.
- Material Conformance: Materials must conform to specific ASTM standards (e.g., ASTM F136 for Ti-6Al-4V, ASTM F1537 for Co-Cr-Mo). The reported performance is that the materials conform to these standards.
- Biocompatibility: While not detailed in this excerpt, the mention of "biocompatibility" in relation to predicates implies conformance to relevant biocompatibility standards (e.g., ISO 10993 series). The reported performance is that it is compatible.
- Sterilization Validation: Demonstrated sterility assurance level (SAL) of 10⁻⁶ via validated methods (moist heat or gamma irradiation). The reported performance is that validation was performed and met this SAL.
- Dimensional Compatibility: The abutments must fit the corresponding OEM implants correctly. The reported performance is that reverse engineering dimensional analysis confirmed compatibility.
Due to the nature of the document being a 510(k) summary focusing on substantial equivalence rather than a full study report, specific numerical performance results for the device tests are not provided in this text.
2. Sample sizes used for the test set and the data provenance
- Sample Size for Mechanical Testing: The document states "mechanical testing conducted according to ISO 14801." For such tests, ISO 14801 typically specifies minimum sample sizes (e.g., 10-11 samples for static strength, typically more for fatigue). The exact number of samples used for this specific submission is not explicitly stated, but it would have followed the standard's requirements.
- Data Provenance: The mechanical testing and material analyses are assumed to be "non-clinical data submitted or referenced" by the manufacturer, Terrats Medical SL, based in Barcelona, Spain. The "reverse engineering dimensional analysis" was done by Terrats Medical SL or through contractual agreement. This is prospective testing performed to support the 510(k). The document itself does not specify the country of origin for the underlying OEM implant data used for reverse engineering, although the OEM companies are listed (e.g., Astra Tech AB, BioHorizons).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A. This is a physical device. Ground truth, in the context of AI, refers to validated labels for data used to train and test an algorithm. For a physical device, performance is evaluated through engineering and biocompatibility testing against defined standards. There are no "experts" establishing ground truth in the AI sense. Testing would be performed by qualified engineers and technicians.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
N/A. Adjudication methods are typically used in clinical studies involving interpretation (e.g., by radiologists) to resolve discrepancies. This document describes non-clinical performance testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A. This product is a dental implant abutment, not an AI software intended to assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
N/A. This is a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
N/A. For engineering tests of physical devices, the "ground truth" is typically derived from established engineering principles, international standards (e.g., ISO 14801 for mechanical properties, ASTM for materials), and the physical properties of the materials and designs themselves. There isn't "expert consensus" or "pathology" in the AI or clinical trials sense.
8. The sample size for the training set
N/A. This is a physical device; there's no "training set" in the machine learning sense. The device is manufactured based on established engineering designs and material specifications.
9. How the ground truth for the training set was established
N/A. No training set for AI. For device manufacturing, the "ground truth" for design and production parameters comes from established engineering best practices, prior successful device designs (predicate devices), and adherence to quality systems regulations (21 CFR Part 820).
Ask a specific question about this device
(98 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system by a change in sterilization status to provide products sterile to the end user that were previously cleared to be provided non-sterile. The subject device abutments and abutment screws were cleared previously to be provided non-sterile to the end user in K170588, K191986, K212628, and K22288. All subject device components will now be provided sterile.
The subject device components include Healing Abutments, Multi-Unit Abutments (0, 17°, and 30°), and abutment screws.
The provided text is a 510(k) summary for the DESS Dental Smart Solutions, an endosseous dental implant abutment. It details the device, its intended use, and its substantial equivalence to previously cleared predicate and reference devices. However, this document does not contain the acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML medical device.
The 510(k) submission for this dental abutment focuses on establishing substantial equivalence based on:
- Design and Material: The subject device components are identical in design, material (Ti-6Al-4V alloy, DLC coating), and technological characteristics to previously cleared devices.
- Manufacturing: The manufacturing process is consistent with previously cleared devices.
- Biocompatibility: Referenced from previous K-clearances.
- Sterilization: The main change in this submission is expanding the system to provide products sterile to the end-user via gamma irradiation, which was validated by referencing a previous K-clearance (K212538).
- MR Environment Testing: Non-clinical analysis was performed to evaluate the subject devices in the MR environment, referencing published literature and FDA guidance.
- Shelf Life Testing: Referenced from a previous K-clearance (K212538) for samples after accelerated aging.
Therefore, I cannot fulfill your request to describe the acceptance criteria and a study proving an AI/ML device meets them based on the provided text, as this document is not about an AI/ML medical device. It's about dental implant abutments and establishes substantial equivalence through non-clinical performance data and comparison to predicate devices, not through AI/ML performance metrics.
To provide the information you requested, I would need a document detailing the validation of an AI/ML medical device, which would include definitions of acceptance criteria (e.g., accuracy, sensitivity, specificity), details of training and test datasets, ground truth establishment, and potentially clinical effectiveness studies if applicable.
Ask a specific question about this device
(118 days)
MIST IC abutments are intended for use to support a prosthetic device in a partially or completely edentulous patient. They are intended to support a single-unit or multi-unit, cement retained prosthesis in the mandible or maxilla. MIST IC abutments are compatible for use with the following implants:
| Compatible Implant Systems | Implant Body Diameter, mm | Implant Platform, mm |
|---|---|---|
| Biomet 3i OSSEOTITE® Certain® | 3.25 | 3.4 |
| 4.0 | 4.1 | |
| 5.0 | 5.0 | |
| 6.0 | 6.0 | |
| NobelActive® (conical connection) | 3.5 | 3.5 (NP) |
| 4.3, 5.0 | 3.9 (RP) | |
| 5.5 | 5.1 (WP) | |
| NobelReplace Conical Connection | 3.5 | 3.5 (NP) |
| 4.3, 5.0 | 3.9 (RP) | |
| NobelParallel Conical Connection | 3.75 | 3.5 (NP) |
| 4.3, 5.0 | 3.9 (RP) | |
| 5.5 | 5.1 (WP) | |
| Replace Select Tapered TiUnite | 3.5 | 3.5 (NP) |
| 4.3 | 4.3 (RP) | |
| 5.0 | 5.0 (WP) | |
| 6.0 | 6.0 | |
| Replace Select Tapered PMC | 3.5 | 3.5 (NP) |
| 4.3 | 4.3 (RP) | |
| 5.0 | 5.0 (WP) | |
| 6.0 | 6.0 | |
| Replace Select TC | 3.5 | 3.5 (NP) |
| 4.0 | 4.3 (RP) | |
| Zimmer Screw-Vent® | 3.7 | 3.5 |
| 4.7 | 4.5 | |
| Zimmer Tapered Screw-Vent® | 3.7, 4.1 | 3.5 |
| 4.7 | 4.5 | |
| 6.0 | 5.7 |
All digitally designed custom abutments for use with MIST IC abutments are to be sent to an Imagine Milling Technologies validated milling center for manufacture.
MIST IC abutments for Biomet 3i Certain 3.25 mm implant bodies are indicated for maxillary lateral and mandbular central/lateral incisors only.
MIST IC from Imagine Milling Technologies, LLC is a line of Ti-base and machinable blank abutments to interface with compatible dental implants from 3 manufacturers, a total of 14 implant-abutment interface compatibilities. The compatible implant body diameters range from 3.25 mm to 6.0 mm and the corresponding implant platform diameters range from 3.4 mm. The subject device prosthetic platform diameters range from 3.8 mm. All stock subject device components (abutments and abutment screws) are made of titanium alloy conforming to ASTM F136. The subject device MIST IC L-LINK abutments have a TiN coating achieved through a physical vapor deposition (PVD) process that is identical to the process used for TiN coating of Imagine Milling Technologies, LLC devices cleared in K 182246. The PVD cathodic arc evaporation process is a high current, low voltage process in which material evaporated from the cathode (Ti) is ionized, transported through the vacuum chamber with reactive gas (N2) and deposited as a non-porous, thin film on the titanium substrate.
Each abutment is supplied with the non-sterile abutment screw designed for attachment to the corresponding compatible OEM implant.
All patient-specific abutment fabrication for all MIST IC abutments is by prescription on the order of the clinician. All MIST IC abutments are intended to be milled at an Imagine Milling Technologies, LLC validated milling center under FDA quality system regulations.
MIST IC L-LINK abutments are two-piece abutments to be used as a base when fabricating a CAD-CAM customized restoration where the superstructure produced will compose the second part of the two-piece abutment; the assembly becoming a final finished medical device after cementation on the subject device abutment. They are provided in straight designs, with engaging and non-engaging connections. The L-LNK abutments and corresponding zirconia superstructure are provided to the clinician either with the superstructure cemented to the abutment by the dental laboratory, or separately for the clinician to bond together chairside using the cement recommended in the labeling (RelyX RMGIP bonding cement, cleared in K022476).
Design parameters for the L-LINK zirconia superstructure are: Minimum wall thickness - 0.5 mm Minimum post height for single-unit restoration - 4.0 mm Minimum gingival height of the coping - 0 mm (all L-LINK bases have minimum gingival height of 0.5 mm) Maximum gingival height – 5.0 mm Maximum angle - 20°
All zirconia copings (superstructures) for use with the subject device MIST IC L-LINK abutments will conform to ISO 13356.
MIST IC PREFIT abutments are cylindrical abutments designed for patient-specific abutment fabrication by a CAD-CAM process and machined into a one-piece, all titanium abutment. The portion of the abutment available for milling is either 9.9 mm in diameter by 20 mm in diameter by 20 mm in diameter by 20 mm in length. MIST IC PREFIT abutments have an engaging connection.
Design parameters for the PREFIT patient specific abutment are: Minimum wall thickness - 0.5 mm Minimum post height for single-unit restoration - 4.0 mm Minimum gingival height - 0.5 mm Maximum gingival height - 5.0 mm Maximum angle - 30°
The provided text describes the regulatory clearance of a dental device, "MIST IC." It outlines the device's intended use and compares it to a predicate device. However, it does not contain explicit information about acceptance criteria or a specific study proving the device meets those criteria, especially in the context of AI performance.
This document focuses on establishing substantial equivalence for regulatory purposes, relying heavily on non-clinical testing and comparison to an existing predicate device. The information provided is primarily relevant to the safety and mechanical performance of the dental implant abutments, not AI/algorithm performance.
Therefore, many of the requested fields cannot be directly answered from the provided text.
Here's a breakdown of what can and cannot be extracted from the document based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in terms of numerical thresholds or performance metrics for an AI algorithm. Instead, it lists non-clinical tests performed to demonstrate substantial equivalence to the predicate device and ensure safety and efficacy.
| Acceptance Criteria Category | Specific Test/Evaluation | Reported Device Performance / Outcome |
|---|---|---|
| Material & Biocompatibility | Biocompatibility according to ISO 10993-5 and ISO 10993-12 | Found compatible |
| Sterilization | Sterilization validation according to ISO 17665-1 and ISO 17665-2 | Validated |
| Mechanical Strength | Static and dynamic compression-bending testing according to ISO 14801 | Fatigue limit data demonstrated sufficient strength for intended use |
| Magnetic Resonance (MR) Safety | Non-clinical analysis of metallic subject devices and compatible OEM implant bodies in MR environment (per FDA guidance) | Addressed parameters including magnetically induced displacement force and torque |
| Compatibility | Reverse engineering of OEM implant bodies, OEM abutments, and OEM abutment screws | Confirmed compatibility for each of the compatible OEM implant system interfaces |
2. Sample Size Used for the Test Set and Data Provenance
This information is not provided in the document. The document refers to "non-clinical testing data" but does not detail the sample sizes for these tests or the provenance of any data beyond indicating it's for dental implant abutment compatibility and mechanical strength. There is no mention of an "AI test set."
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided. The document describes mechanical performance and material compatibility testing, not a study requiring expert-established ground truth for an AI algorithm.
4. Adjudication Method for the Test Set
This information is not provided. As no "test set" for AI or expert review is mentioned, adjudication methods are not applicable here.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical data were included in this submission." This type of study typically involves human readers and is clinical in nature.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The document describes a physical medical device (dental implant abutments) and its mechanical and material properties. There is no mention of an AI algorithm or standalone algorithm performance.
7. The Type of Ground Truth Used
For the non-clinical tests mentioned, the "ground truth" is established by objective engineering standards and validated testing protocols (e.g., ISO 14801 for mechanical testing, ISO 10993 for biocompatibility). There is no "expert consensus," "pathology," or "outcomes data" as ground truth for an AI in this context.
8. The Sample Size for the Training Set
This information is not provided. There is no mention of a training set as the document does not concern an AI algorithm.
9. How the Ground Truth for the Training Set Was Established
This information is not provided as there is no mention of a training set or AI algorithm in the provided text.
Summary of what the document IS about:
The document is an FDA 510(k) clearance letter for the MIST IC dental abutment. It establishes substantial equivalence to a predicate device (K182246) based on:
- Non-clinical testing: Biocompatibility, sterilization, MR compatibility, reverse engineering for compatibility, and static/dynamic compression-bending.
- Material properties: Both subject and predicate devices are made of Titanium Alloy (ASTM F136), with TiN coating for L-LINK abutments, and use zirconia copings (ISO 13356) and the same bonding cement (RelyX RMGIP).
- Design similarities: Similar CAD-CAM fabrication process, restoration types (single/multi-unit, cement-retained), and design parameters (minimum wall thickness, post height, gingival height, angulation).
- Intended Use: Supports a prosthetic device in partially or completely edentulous patients, compatible with various specified implant systems.
The text does not discuss, describe, or evaluate any AI component of the device.
Ask a specific question about this device
(504 days)
The DSP Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
DSP Implant System is composed of the External Hex (HE) implant line is composed of implants and related prosthetic components available in multiple designs. DSP implants are made of commercially pure titanium (ASTM F67). DSP prosthetic components are made of commercially pure titanium or cobalt-molybdenum alloy (ASTM F1537). DSP implant system screws (abutment and prosthetic screws) are made of titanium alloy (ASTM F136). The HE implant line is named after the External Hex implant-to-abutment interface. It is subdivided in External Hex SLIM (HE SLIM), External Hex BIOFIT), External Hex SOULFIT (HE SOULFIT), External Hex PROPFIT (HE PROPFIT) and External Hex WAYFIT). The HE implants are bone-level implants. The subject device abutments components mate exclusively with the subject implants of the same line.
This is a 510(k) summary for the DSP Implant System, a dental implant device. It establishes substantial equivalence to various predicate devices rather than providing performance data against specific acceptance criteria for the new device. Therefore, it does not present a typical study design with acceptance criteria and results in the way a clinical trial or device performance study would.
However, based on the provided text, we can infer the approach taken by the manufacturer to demonstrate substantial equivalence, which serves as their "proof" that the device meets regulatory requirements.
Here's an interpretation of the requested information based on the 510(k) summary:
1. A table of acceptance criteria and the reported device performance
Since this is a 510(k) submission focusing on substantial equivalence, there isn't a table of discrete acceptance criteria and reported device performance in terms of clinical outcomes or specific quantitative measures as one might find in a standalone performance study. Instead, the "acceptance criteria" are implied by the characteristics of the legally marketed predicate devices, and the "reported device performance" is the demonstration that the subject device shares these characteristics or has equivalent performance as validated through non-clinical testing.
The acceptance is based on demonstrating that the DSP Implant System is substantially equivalent to the predicate devices in terms of:
- Intended Use: The indications for use statement of the subject device is equivalent to the primary predicate device K101207, and within the scope of reference devices.
- Technological Characteristics: The subject device's design features, materials (CPTi Gr4, CoCr, POM, Ti6Al4V-ELI), sterilization methods (irradiation), and surface treatments (grit-blasted and acid-etched, machined collar) are compared to the predicate and reference devices. Minor differences are justified as not raising new questions of safety or effectiveness.
- Performance: Non-clinical testing demonstrates that the device performs comparably to the predicate devices in relevant areas for dental implants.
Implicit Acceptance Criteria and Reported Device Performance (based on substantial equivalence justification):
| Acceptance Criteria (Implied from Predicate Device Characteristics) | Reported Device Performance (DSP Implant System's Characteristics & Non-Clinical Testing) |
|---|---|
| Intended Use: Support prosthetic devices (artificial teeth) to restore chewing function; single-stage or two-stage procedures; single/multiple unit restorations; immediate loading with good primary stability and appropriate occlusal loading. | Equivalent: "The DSP Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading." (Matches K101207, with one phrase "multiple tooth application may be rigidly splinted" removed as it's a technique choice). |
| Implant-to-Abutment Connection: e.g., External Hex (HE). | Equivalent: HE connection for all subject implants (HE SLIM, HE BIOFIT, HE SOULFIT, HE WAYFIT, HE PROPFIT) and abutment components (CoCr Base UCLA, CoCr UCLA, Titanium Abutments, Mini Conical Abutments, Anatomical Abutments, O'ring Abutment, Temporary Coping, CoCr Base Coping, One Step Hybrid Coping). |
| Raw Materials: Commercially Pure Titanium (CPTi Gr4), Titanium alloys (ASTM F136), Cobalt-Chromium alloys, Polyoxymethylene (POM). | Equivalent: Implants are CPTi Gr4. Prosthetic components use CPTi Gr4, CoCr alloy, CoCr/POM, and Ti alloy (ASTM F136). These match or are comparable to predicate/reference device materials. |
| Surface Treatment: Grit-blasted and acid-etched (and/or machined collar). | Equivalent: Implants are treated. Grit-blasted and acid-etched. Machined collars are specified (e.g., h 1.0 mm for HE SLIM, h 0.3 mm for HE BIOFIT). EDS analysis confirmed no contaminating metal particles. |
| Sterilization: Sterile by irradiation (or ethylene oxide for predicates). | Equivalent/Validated: Provided sterile by irradiation. Sterilization validated according to ISO 11137-1, ISO 17665-1, ISO 17665-2. Sterile barrier shelf life testing (ASTM F1980), package integrity (ASTM F1929, ASTM F88/F88M), sterility of content (ISO 11737-2). |
| Biocompatibility: Materials are biocompatible. | Validated: Biocompatibility of materials supported by testing according to ISO 10993-5 (likely ISO 10993-1 as well, given the guidance document reference). |
| Pyrogenicity: Device is non-pyrogenic. | Validated: Pyrogenicity monitoring according to AAMI/ANSI ST72. Bacterial Endotoxin Test (BET) performed according to USP <85> using photometric technique. Representative samples selected for testing. |
| Dimensions/Geometry: Range of diameters and lengths for implants and abutments. | Equivalent: Tables 5.2-5.17 detail the diameter and length ranges for various implants and abutments, comparing them to predicate/reference devices. The text explicitly states that the subject implants and abutments are "substantially equivalent" to predicate/reference devices in design and range of dimensions, with specific predicate/reference devices covering particular dimensions or features. For example, K101207 covers 17mm length for HE SLIM, while K170398 covers other dimensions and body geometrical features. |
| Functional Equivalence: e.g., Screw-retained, Cement-retained, Overdenture prosthesis attachment; Single-unit, Multi-unit restoration. | Equivalent: All subject implants and abutments support screw-retained, cement-retained, and/or overdenture prosthesis attachment, and single-unit or multi-unit restorations, consistent with the predicate devices. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not applicable for a typical clinical or performance study as presented in this 510(k). The "testing" referred to is non-clinical bench testing (sterilization, biocompatibility, pyrogenicity, material analysis). These tests typically use a representative sample size specified by the relevant standard (e.g., a certain number of devices for sterility testing). The document does not specify these individual sample sizes.
- Data Provenance: The data is non-clinical (bench testing) and generated by the manufacturer to support device characteristics. The 510(k) is submitted by a Brazilian manufacturer (D.S.P. Industrial Eireli) and prepared by a Brazilian regulatory affairs specialist (PR Servicos Regulatorios Administrarivos Ltda). The location of the actual testing labs is not specified. The studies are not clinical, so concepts like "retrospective" or "prospective" do not apply.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This is a non-clinical submission. Ground truth, in the sense of expert opinion on clinical data, is not established. The "ground truth" for the non-clinical tests is established by the specifications of the referenced international and national standards (ISO, ASTM, USP, AAMI/ANSI).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. Adjudication methods are typically for evaluation of conflicting expert opinions or clinical outcomes, which is not the nature of this non-clinical submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a submission for a dental implant system (hardware), not an AI/software device. An MRMC study is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- No. This is a physical medical device. Not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- For the non-clinical performance data, the "ground truth" is defined by the requirements and test methodologies of the referenced industry standards (e.g., ISO 11137 for sterilization, ISO 10993-5 for biocompatibility, AAMI/ANSI ST72 and USP <85> for pyrogenicity, SEM/EDS for material analysis). The demonstration is that the device conforms to these standards and shares equivalent characteristics with legally marketed predicate devices.
8. The sample size for the training set
- Not applicable. This is a physical medical device, not an AI/software device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(247 days)
The Panthera Dental Milled Bar is indicated for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient for purpose of restoring chewing function. It is intended for use to support multiple tooth prostheses in the mandible or maxilla. The prostheses can be screw retained.
The Panthera Dental Milled Bars are indicated for compatibility with:
- · NobelActive: NP Ø3.5; RP Ø4.3 / Ø5.0; WP Ø5.5
- NobelParallel CC: NP Ø3.75: RP Ø4.3 / Ø5.0: WP Ø5.5
- NobelReplace: NP Ø3.5; RP Ø4.0 / Ø4.3 / Ø5.0; WP Ø5.0; 6 Ø6.0
- NobelSpeedy: RP Ø4.0 / Ø5.0; WP Ø5.0 / Ø6.0
- Brånemark: NP Ø3.3; RP Ø3.75 / Ø4.0; WP Ø5.0
The Panthera Dental Milled Bar is a metallic dental restorative device that is intended for attaching by screw retention to dental implants to aid in the treatment of partial and totally edentulous patients for the purpose of restoring chewing function.
The Panthera Dental milled bars for which clearance is requested, are included in one of the following bar types, which have distinct design specifications.
The Type I bars are specific for removable overdenture and include:
- . Panthera Dental Dolder Bar, Hader Bar, Milled Bar, REBourke Bar and Paris Bar.
The Type II bars are specific for fixed prostheses and include:
- Panthera Dental Wrap-around Bar, Montreal Bar, Montreal Bar with metallic lingual, Pin Lingual Bar and Pin Wrap-Around Bar.
The Panthera Dental Milled Bar is designed to match an individual patient. Panthera Dental designs the bar from a three-dimensional optical and/or digital scanner system that scans the patient's impression; the dental professional prepares the model cast beforehand. The designed bar is then machined using a computer-aided design/ computer-aided manufacturing (CAD/CAM) software system. The bar is milled from titanium (Ti-6AI-4V grade 5). CAD/CAM fabrication is only performed by Panthera Dental, within our manufacturing control and not by the dental laboratory.
The Panthera Dental Milled Bar is packaged as non-sterile, and delivered to a dental laboratory for completion. Once received at the laboratory, the Panthera Dental Milled Bar is matched to a denture for final placement. The Panthera Dental Milled Bar provides retention and support for a removable or fixed denture made of standard laboratory dental materials such as resin composite.
Here's a breakdown of the acceptance criteria and the study information for the Panthera Dental Milled Bars, based on the provided document.
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly present acceptance criteria in a tabular format with specific numerical targets. Instead, it describes performance through compliance with recognized standards and successful testing outcomes. The "performance" is primarily demonstrated through meeting mechanical strength requirements and biocompatibility standards, and by showing substantial equivalence to a predicate device.
| Acceptance Criteria Category | Description of Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Mechanical Strength | Meets the requirements for dynamic fatigue testing for endosseous dental implants. | The device (bar and cylinders) has the mechanical strength for its intended clinical application, as demonstrated by fatigue testing according to ISO 14801 and FDA guidance. |
| Biocompatibility | Biocompatible with tissue/bone/dentin for long-term contact. | Made from Titanium Ti-6Al-4V, a material used in the predicate. Cytotoxicity, extractable/leachable, toxicological risk assessment, and bacterial endotoxin testing compliant with ISO 10993-1, ISO 10993-5, ISO 10993-12, ANSI/AAMI ST72, and USP <85>. |
| Sterilization | Can be effectively sterilized to a Sterility Assurance Level (SAL) of 10^-6. | Sterilization validation (Pre-vacuum steam sterilization) for the primary predicate was successful, demonstrating an SAL of at least 10^-6. The proposed device uses the same materials and sterilization cycle, therefore no additional testing was required. |
| Design Accuracy/Fit | Ensures a perfect fit and proper seating with compatible implant systems, with no possible interference. | Reverse engineering (RE) confirmed that the critical parameters of the design for interface connection, including tolerance limits, fall within the measured sizes of the Nobel Biocare counterparts. |
| Manufacturing Reliability | Manufacturing process is reliable and produces consistent results. | A process capability study conducted for the primary predicate (K173466) is valid for the proposed device, as the manufacturing process is identical. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify the exact sample size for each test (e.g., how many bars were subjected to fatigue testing). It refers to "the bar itself" and "the bar cylinders" for fatigue testing. For reverse engineering, it mentions "each interface connection" and "each components per size and type" which implies multiple samples were analyzed to determine critical parameters and tolerance limits.
- Data Provenance: The document does not specify country of origin for data. The non-clinical testing was conducted by Panthera Dental Inc., a Canadian company. The tests are framed within the context of a 510(k) submission to the U.S. FDA. The testing conducted was for the primary predicate device (K173466) and determined to be applicable to the proposed device due to identical materials and manufacturing processes. The tests were retrospective in the sense that they were done on manufactured devices/components.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: This information is not explicitly stated. The document refers to compliance with ISO and FDA guidance documents for testing, which implies a standard methodology rather than expert consensus on a subjective test set.
- Qualifications of Experts: Not specified. Testing was performed by Panthera Dental Inc. or their designated testing facilities to meet recognized standards.
4. Adjudication Method for the Test Set
Not applicable. The tests mentioned (fatigue, biocompatibility, sterilization, reverse engineering, process capability) are objective physical, chemical, and mechanical tests, not subjective evaluations requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The Panthera Dental Milled Bars are a physical dental device, not an AI or imaging diagnostic tool that would typically involve a reader study.
6. Standalone (Algorithm Only) Performance Study
Not applicable. This is a physical medical device, not an algorithm or software. The mention of CAD/CAM refers to the design and manufacturing process, not an independent algorithm performance.
7. Type of Ground Truth Used
The "ground truth" for the device's performance is established through:
- Compliance with recognized standards: ISO 14801 for fatigue testing, ISO 10993 series and USP <85> for biocompatibility, and ANSI/AAMI-ST8 for sterilization.
- Measurement and verification: Reverse engineering to ensure precise fit with implant systems.
- Predicate device equivalence: Performance data for the predicate device (K173466) is directly applied due to identical materials, design principles, and manufacturing processes.
8. Sample Size for the Training Set
Not applicable. This is a physical medical device. There is no "training set" in the context of machine learning or AI. The design and manufacturing processes are established based on engineering principles and validated through testing.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no "training set" in the context of this device. The design specifications and manufacturing parameters are established through engineering design, material science, and testing against industry standards, not via a machine learning training process.
Ask a specific question about this device
(207 days)
The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid Base™ is used as an interface between a dental implant and a zirconia superstructure and will be attached to the implant using the included prosthetic screw and attached to the zirconia superstructure by cementing.
The Elos Accurate® Hybrid Base™ is compatible with the implant systems listed in Table 1.
All digitally designed zirconia superstructures for use with the Elos Accurate® Hybrid Base™ are only intended to be sent and manufactured at an FDA registered Elos Medtech approved milling facility.
The Elos Accurate Hybrid Base is a titanium base designed to interface with a dental implant and to support a patient-specific ceramic superstructure or a multi-unit direct restoration cemented to the base. The base and the superstructure form a patient-specific abutment that will support a definitive restoration, either a single crown or a multi-unit restoration. The Elos Accurate Hybrid Base Engaging is intended for single-unit restorations and Elos Accurate Hybrid Base Nonengaging is intended for multi-unit restorations. Alternatively, a definitive multi-unit restoration may be cemented directly to the Elos Accurate Hybrid Base.
Manufacture of the final finished device will be at an Elos Medtech-approved milling facility that is registered with FDA as a medical device manufacturer and is qualified as a contract manufacturer to Elos Medtech.
This submission includes a two-piece abutment (titanium base and zirconia superstructure) and abutment screws compatible with OEM implants from Dentsply Implants (Astra Tech TX, Astra Tech EV), Nobel Biocare (Nobel Active/Conical Connection and Nobel Replace) and Institut Straumann (Straumann Bone Level NC and RC).
Abutments compatible with Astra Tech implants are available in engaging (anti-rotational) designs, and all others are available in both engaging and non-engaging designs.
This document is a 510(k) Premarket Notification from the FDA regarding the "Elos Accurate® Hybrid Base" dental implant device. It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific performance criteria through a study with acceptance criteria.
Therefore, the information required to directly answer some of your questions, particularly those related to the performance study design for an AI/ML-based medical device (e.g., sample size for test/training sets, expert consensus, MRMC studies, ground truth establishment for AI models), is not present in this document. This is because the device in question is a physical dental implant component, not an AI/ML algorithm.
However, I can extract the relevant information regarding the performance data provided for this physical device.
Here's a breakdown based on the provided document:
1. A table of acceptance criteria and the reported device performance
The document does not present a formal table of acceptance criteria and reported device performance in the way an AI/ML device submission would. Instead, it states that substantial equivalence is demonstrated through:
- Biocompatibility evaluation and confirmatory cytotoxicity testing: According to ISO 10993-5 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity).
- Dynamic compression-bending testing: According to ISO 14801 (Dentistry - Implants - Dynamic fatigue test for endosseous dental implants).
- Engineering and dimensional analysis: To confirm compatibility with OEM implant bodies, OEM abutments, and OEM abutment fixation screws.
- Sterilization validation: In accordance with ISO 17665-1 and ISO 17665-2.
Reported Performance/Demonstration:
- "Substantial equivalence with regard to compatibility with OEM components is supported by engineering and dimensional analysis of OEM implant bodies, OEM abutments and OEM abutment fixation screws to confirm compatibility."
- "Substantial equivalence with regard to mechanical performance is supported by dynamic testing according to ISO 14801 Dentistry – Implants – Dynamic fatigue test for endosseous dental implants."
- "The coatings used on selected screws are identical to coatings on previously cleared devices."
- "Sterilization validation of the reprocessing instructions were conducted in accordance to ISO 17665-1 and ISO 17665-2."
- "Minor differences in the designs, dimensions, sizes, or compatible OEM implant lines... do not affect substantial equivalence. These minor differences... are related to the compatible OEM implant designs or are mitigated by the mechanical performance testing."
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not specified for any of the non-clinical tests (biocompatibility, dynamic fatigue, engineering analysis, sterilization). The document indicates "No clinical data were included in this submission," further confirming that there wasn't a "test set" in the context of patient data for AI performance evaluation.
- Data Provenance: Not applicable as it's not clinical data. The tests are laboratory-based.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This device is a physical component, not an AI/ML diagnostic tool requiring expert ground truth for image interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. As above, this is not a diagnostic study requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. No clinical or human-in-the-loop study was conducted or required for this physical device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an algorithm. The "performance data" refers to the physical and mechanical properties of the dental abutment.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the non-clinical performance data, the "ground truth" or reference standards are the ISO standards themselves (ISO 10993-5, ISO 14801, ISO 17665-1, ISO 17665-2), and the engineering specifications/dimensions of OEM implant components for compatibility analysis. This is a technical ground truth based on established material science and mechanical engineering principles, rather than medical expert consensus or patient outcomes.
8. The sample size for the training set
- Not applicable. There is no "training set" in the context of an AI/ML model for this physical device.
9. How the ground truth for the training set was established
- Not applicable. As above, no training set for an AI/ML model.
Ask a specific question about this device
(115 days)
The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the included Elos Prosthetic screw.
The Elos Accurate® Customized Abutments are compatible with the implant systems listed in table 1:
Table 1.
| Elos Accurate Customized Abutment – Model Type | Platform compatibility | Platform diameter [mm] | Implant Body diameter [mm] |
|---|---|---|---|
| AB-NBR43 | Nobel Replace RP | 4.3 | 4.3 |
| AB-NBR50 | Nobel Replace WP | 5 | 5 |
| AB-NBR60 | Nobel Replace 6.0 | 6 | 6 |
| AB-NBA35 | Nobel CC NP | 3.5 | 3.5 & 3.75 |
All digitally designed CAD/CAM customizations for the Elos Accurate® Customized Abutments are only intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility.
The Elos Accurate® Customized Abutment is a patient specific abutment intended for attaching to dental implants in order to provide basis for single- or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to the implant using the included Elos Prosthetic Screw and attached to the crown/coping manually by cementation. The Elos Accurate® Customized Abutment consists of an Abutment Blank used in fabricating of a full patient-specific abutment in Titanium alloy per ASTM F136. The Abutment Blank used in creation of the Elos Accurate® Customized Abutment has a pre-manufactured connection interface that fits directly to a pre-specified dental implant. The customized shape of the abutment is intended to be manufactured by an Elos Medtech approved milling facility. The Elos Accurate® Customized Abutment is delivered nonsterile and the final restoration including corresponding Elos Prosthetic Screw is intended to be sterilized at the dental clinic before it is placed in the patient. The Elos Accurate® Customized Abutment provides clinicians and laboratories with a prosthetic device that can be used in definitive (permanent) single- or multi restorations.
The provided text describes a medical device, the Elos Accurate® Customized Abutment, and its 510(k) submission to the FDA. However, the document does not contain information regarding objective acceptance criteria (e.g., performance metrics with defined thresholds), nor does it detail a study that proves the device meets such criteria.
Instead, the document focuses on demonstrating substantial equivalence to a previously cleared predicate device (K190299) based on similar intended use, operating principle, design, materials, and sterilization processes.
Therefore, I cannot provide a table of acceptance criteria and reported device performance or information about sample sizes for test sets, expert adjudication methods, MRMC studies, standalone performance, or ground truth details, as these are not present in the provided text.
The information that is available regarding testing is non-clinical and aims to show similarity and safety compared to the predicate device, rather than meeting specific performance thresholds against a defined clinical outcome.
Here's a summary of the non-clinical testing performed, which serves as the evidence for substantial equivalence:
Non-Clinical Testing Information (as provided):
- Type of Testing:
- Engineering and dimensional analysis of original manufacturers' components (abutments, implants & abutment screws) for determination of compatibility.
- Fatigue testing per ISO 14801 according to FDA guidance for Industry and FDA Staff "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" dated May 12, 2004.
- Biocompatibility testing for cytotoxicity according to ISO 10993-5.
- Tests Included for Biocompatibility:
- Non-coated prosthetic screw (representative for subject device, tested on primary predicate device K190299).
- Medicarb coated prosthetic screw (representative for subject device, tested on reference predicate device K171799).
- Elos Accurate® Customized Abutment (representative for subject devices, tested on reference predicate device K171799).
- Sterilization: The primary predicate device was tested according to ISO 17665-1 & ISO 17665-2, demonstrating a Sterility Assurance Level (SAL) of $10^{-6}$. No additional testing was deemed necessary for the subject device.
Missing Information (Not provided in the document):
- Specific quantitative acceptance criteria for performance metrics.
- A clinical study designed to demonstrate performance against such criteria.
- Sample size used for a test set (as no such test set for performance evaluation is described).
- Data provenance for a test set.
- Number of experts used to establish ground truth or their qualifications.
- Adjudication method for a test set.
- Multi-reader multi-case (MRMC) comparative effectiveness study results.
- Effect size of human reader improvement with AI vs. without AI assistance.
- Standalone (algorithm-only) performance.
- Type of ground truth used (expert consensus, pathology, outcomes data, etc.).
- Sample size for the training set.
- How ground truth for the training set was established.
The document's purpose is to establish substantial equivalence for regulatory clearance, not to report on a clinical performance study with predefined acceptance criteria.
Ask a specific question about this device