Found 6 results

510(k) Data Aggregation

    K Number
    K251471
    Date Cleared
    2025-08-20

    (99 days)

    Product Code
    Regulation Number
    872.3630
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.

    Device Description

    IPD Dental Implant Abutments is a dental implant abutment system composed of dental abutments, screws, as well as other dental abutment accessories, intended to be placed into dental implants to provide support for dental prosthetic restorations.

    Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems.

    IPD Dental Implant Abutments includes the following categories of dental abutment designs:

    • Titanium base (Interface) abutments (INC3D);
    • Multi-Unit abutments (MUA);
    • Overdenture Abutments (PSD);
    • Temporary Abutments (PP);
    • Healing Abutments (TC).

    The system also includes the use of the corresponding screws intended to attach the prosthesis to the dental implant. Specifically:

    • Ti Screw (TT): Used during restoration fabrication.
    • TiN Screw (TTN): Used in finished restorations, with TiN coating.
    • TPA Screw (TPA): Used in finished angulated restorations, with TiN coating.

    The metallic components of the subject abutments and screws are made of titanium alloy conforming to ISO 5832-3 "Implant for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy".

    The purpose of this submission is to expand IPD Dental Implant Abutments offerings with:
    • New IPD's compatible dental implant systems,
    • New angulations available abutment-category specific.
    • New in-house TiN coating.

    IPD dental implant abutments and screws are compatible with the following commercially available dental implant systems:
    (Table 2. Summary of IPD abutments categories with compatibilized OEM Implant/Abutment Systems with specific reference to maximum angulation specifically included in this submission. provided in original text)

    Ti Base (Interface) abutments are attached (screw-retained) to the implant/abutment and cemented to the zirconia superstructure.

    The Ti Base is a two-piece abutment composed of the titanium component, as the bottom-half, and the zirconia superstructure, as the top-half. It consists of a pre-manufactured prosthetic component in Titanium alloy per ISO 5832-3, as well as the supporting digital library file for FDA-cleared design software (3Shape Abutment Designer™ Software, cleared under K151455) which enables the design of a patient-specific superstructure by the laboratory/clinician and which will be manufactured in FDA-cleared Zirconia (e.g., DD Bio Z, K142987) according to digital dentistry workflow at the point of care, or at a dental laboratory.

    The design and fabrication of the zirconia superstructure for Ti Base (Interface) will be conducted using a digital dentistry workflow requiring the use of the following equipment, software and materials:
    Scanner: 3D Scanner D850.
    Design Software: 3Shape Abutment Designer Software, K151455.
    Zirconia Material: DD Bio Z, K142987.
    Milling machine/Brand: Dental Concept System Model: DC1 Milling System.
    Cement: Multilink® Automix, K123397.

    Ti Base (Interface) abutment design parameters for the zirconia superstructure are defined as follows:
    Minimum gingival height: 1.5 mm
    Minimum wall thickness: 0.43 mm
    Minimum post height for single-unit restorations: 4.75 mm (1)
    Maximum gingival height: 6.0 mm
    Maximum angulation of the final abutment 30° (2)

    The resulting final prosthetic restoration is screwed to the dental implant. All subject abutments are single-use and provided non-sterile. Final restoration (which includes the corresponding screw) is intended to be sterilized at the dental clinic before it is placed in the patient.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to IPD Dental Implant Abutments, a medical device, not an AI/ML-driven software product. Therefore, the information requested regarding acceptance criteria and study data for an AI/ML device (e.g., sample size for test/training sets, expert ground truthing, MRMC studies, standalone performance) is not applicable to this document.

    The document describes the device, its intended use, comparison to predicate devices, and the non-clinical performance testing conducted to demonstrate substantial equivalence. These tests are physical and chemical in nature, not related to the performance of an AI/ML algorithm.

    Here's a breakdown of why an AI/ML-focused response is not possible, based on the provided text:

    • Device Type: The device is "IPD Dental Implant Abutments," which are physical components used in dentistry (titanium alloy abutments, screws, designed for zirconia superstructures). It is not software, a diagnostic imaging tool, or an AI/ML algorithm.
    • Purpose of Submission: The submission aims to expand compatibility with new dental implant systems and include new angulations and in-house TiN coating. This is a modification of a physical medical device, not a new AI/ML development.
    • Performance Data (Section VII): This section explicitly lists non-clinical performance testing such as:
      • Sterilization validation (ISO 17665-1)
      • Biocompatibility testing (Cytotoxicity, Sensitization, Irritation per ISO 10993)
      • Reverse engineering and dimensional analysis for compatibility
      • Validation of the digital workflow and software system (but this refers to the CAD/CAM software used to design the physical abutments, not an AI/ML diagnostic tool)
      • Static and dynamic fatigue testing (ISO 14801)
      • Modified Surfaces Information
      • MRI safety review

    Conclusion:

    The provided document describes a 510(k) clearance for a physical dental implant component. It does not contain any information about the acceptance criteria or study design for an AI/ML driven medical device. Therefore, a table of acceptance criteria and reported device performance related to AI/ML, sample sizes for test/training sets, details on expert ground truthing, MRMC studies, or standalone performance of an algorithm cannot be extracted from this text.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 and K240982 to allow additional options of zirconia material, scanners, CAM software, and milling machines to the digital dentistry workflow. The subject devices are to be sent to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. There are no changes to the abutment design, implant compatibilities, or design parameters. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows (also referred to as point of care) under K221301 and K240982.

    The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.

    Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.

    All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine and associated tooling and accessories.

    The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

    AI/ML Overview

    The provided 510(k) summary for DESS Dental Smart Solutions focuses on demonstrating substantial equivalence to predicate devices for dental implant abutments. It primarily addresses the expansion of compatible materials, scanners, CAM software, and milling machines within an existing digital dentistry workflow. The document does not describe an AI/ML-based device that would typically have acceptance criteria related to diagnostic performance.

    Therefore, many of the requested items related to AI/ML device performance (like acceptance criteria for diagnostic metrics, sample size for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this submission.

    The acceptance criteria and supporting "study" (non-clinical data) for this device are related to its mechanical performance, biocompatibility, and integration within the digital workflow, demonstrating that the expanded components maintain the safety and effectiveness of the previously cleared predicate devices.

    Here's a breakdown based on the information provided and the non-applicability of AI/ML-specific questions:

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML diagnostic device, the acceptance criteria are not in terms of traditional diagnostic metrics (sensitivity, specificity, AUC). Instead, they are related to material properties, mechanical integrity, and the digital workflow's accuracy.

    Acceptance Criteria CategoryReported Device Performance (Summary from Submission)
    Mechanical Integrity / Compatibility- Fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions. (Implies successful completion to similar or better standards than predicate.)
    Biocompatibility- Biocompatibility testing according to ISO 10993-5 and ISO 10993-12. (Implies successful demonstration of biocompatibility.)
    Sterilization- Sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937. (Implies successful validation for sterilization.)
    CAD Design Restrictions- Software verification included testing of restrictions that prevent design of components outside of the stated design parameters.
    • Abutment design library validated to demonstrate established design limitations are locked and cannot be modified by the user. (Implies successful implementation and verification of design constraints.) |
      | CAM Restriction Zones / Manufacturing Accuracy | - Validation testing of CAM restriction zones conducted, including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in CAM software. (Implies successful validation to ensure manufacturing precision and prevent damage.) |
      | Material Conformance | - Zirconia materials conform to ISO 6872.
    • Titanium alloy conforms to ASTM F136.
    • Co-Cr-Mo alloy conforms to ASTM F15337. (Implies materials meet standards.) |
      | Physical Dimensions | - Device encompasses the same range of physical dimensions as the predicate device. (Implies dimensional equivalence.) |

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated in terms of a "test set" for diagnostic performance. The validation involved physical testing of components (e.g., fatigue testing) and software verification. The specific number of abutments or digital design instances used for these non-clinical tests is not detailed in this summary.
    • Data Provenance: Not applicable in the context of patient data for an AI/ML device. The "data" here refers to engineering and material testing results, likely conducted in controlled lab environments (implied to be in accordance with international standards like ISO and ASTM). The manufacturer is Terrats Medical SL, in Spain, so testing would likely originate from their facilities or contracted labs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device relates to engineering specifications and material science.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a diagnostic device involving expert review adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this device, the "ground truth" is based on:

    • Engineering Specifications: Defined design parameters (e.g., minimum wall thickness, post height, angulation limits).
    • Material Standards: Conformance to international standards such as ASTM F136, ISO 6872.
    • Benchmarking/Predicate Equivalence: Performance is assessed against established performance of the predicate devices and OEM implant systems.
    • Software Validation Logic: Verification that software correctly enforces design rules and CAD/CAM restrictions.

    8. The sample size for the training set

    Not applicable. This device does not involve a machine learning training set.

    9. How the ground truth for the training set was established

    Not applicable. This device does not involve a machine learning training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    Device Description

    The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 to add the ability for the subject device Base Abutments and Pre-milled (Blank) Abutments to be designed using AbutmentCAD software in the digital dentistry workflow, as well as add angulation to some of the Pre-Milled (Blank) Abutments. The subject devices are to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. The proposed change is to allow the subject devices to be designed using AbutmentCAD by exocad GmbH, the current workflow allows only the use of 3Shape by 3Shape A/S for the design software. Another purpose of this submission is to expand the design parameters to allow angulation (up to 30°) on Pre-milled (Blank) Abutments that are compatible with Neodent Grand Morse, Nobel Active/Nobel Parallel Conical 3.0 mm, and Straumann BLX implants. There are no changes to the abutment design or implant compatibilities. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows in under K221301.

    The subject device DESS Dental Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.

    Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.

    All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, titanium and ceramic material, milling machine and associated tooling and accessories.

    The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The digital workflow includes the following products (not subject devices of this submission):

    • Ceramic material: VITA YZ ST and VITA YZ XT (K180703)
    • . Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436)
    • . Intraoral Scanner: 3Shape TRIOS A/S Series Intraoral Scanner (510(k) exempt under 21 CFR 872.3661)
    • Desktop scanner: 3Shape D900 Dental Lab Scanner (510(k) exempt under 21 CFR 872.3661)
    • Abutment design software: 3Shape Abutment Designer Software (K151455) and AbutmentCAD ● (K193352)
    • . Milling machine: VHF R5 by vhf camfacture AG with DentalCAM and DentalCNC 7 software
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for DESS Dental Smart Solutions, which are dental implant abutments. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device's effectiveness through clinical performance studies with specific statistical acceptance criteria for accuracy, sensitivity, or specificity.

    Therefore, the document does not contain the information requested regarding:

    • A table of acceptance criteria and reported device performance (in terms of clinical metrics like accuracy, sensitivity, specificity).
    • Sample size used for the test set or its provenance.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for the test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study or its effect size.
    • Standalone (algorithm only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    The study described in this document focuses on non-clinical performance data to demonstrate substantial equivalence, specifically:

    1. Sterilization validation: According to ISO 17665-1, ISO 17665-2, and ISO 14937.
    2. Biocompatibility testing: According to ISO 10993-5 and ISO 10993-12.
    3. Fatigue testing and reverse engineering analysis: Of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility. This includes fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions.
    4. MR Conditional labeling.
    5. Validation testing of CAM restriction zones: Including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in the CAM software.
    6. Software verification: Included testing of restrictions that prevent design of components outside of the stated design parameters. The abutment design library was validated to demonstrate that established design limitations and specifications are locked and cannot be modified by the user.

    The acceptance criteria and reported "performance" for this submission are based on these engineering and design validations, ensuring the device meets safety and performance standards equivalent to the predicate device, K221301. The key "performance" metrics are about maintaining physical and material integrity and compatibility.

    The core of the submission is to expand the DESS Dental Smart Solutions abutment system to:

    • Allow design using AbutmentCAD software (in addition to 3Shape software).
    • Add angulation (up to 30°) to some Pre-milled (Blank) Abutments for specific implant systems.

    The document explicitly states: "No clinical data were included in this submission." and "The subject device, the predicate device, and reference devices have the same intended use, technological characteristics, and are materials. The subject device, the predicate device, and reference devices encompass the same range of physical dimensions, manufactured by similar methods, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above."

    Therefore, this FDA submission is for a physical medical device (dental implant abutment) and its manufacturing/design software modifications, not an AI or diagnostic device that would involve clinical performance metrics like sensitivity or specificity.

    Ask a Question

    Ask a specific question about this device

    K Number
    K234112
    Date Cleared
    2024-05-09

    (134 days)

    Product Code
    Regulation Number
    872.3630
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K140878, K112160, K013227

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Customized Abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Customized Abutments are indicated for screwretained single restorations or cement-retained single or multi-unit restorations. The Customized Abutment will be attached to a dental implant using the included abutment screw.

    Patient-Specific Abutment is compatible with following Implant Systems:

    1. Zimmer Tapered Screw-Vent®
      / Implant diameter size (mm): 3.7, 4.1, 4.7, 6.0/ Restorative Platform diameter (mm): 3.5, 4.5, 5.7
    2. Straumann® Bone Level Tapered Implant
      / Implant diameter size (mm): 3.3, 4.1, 4.8/ Restorative Platform diameter (mm): 3.1, 3.7, 4.4

    All digitally-designed Customized abutments are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture.

    Device Description

    Patient-specific abutment is made from Ti-6Al-4V Eli conforming to ASTM F136 to be used in fabricating patient-specific abutments. The subject devices are indicated for cemented or screwand cement retained prosthesis (SCRP) restorations. Each patient-specific abutment is individually prescribed by the clinician.

    The diameters of patient-specific abutment are 3.3, 3.7, 4.1, 4.7, 4.8, 6.0 mm and Hex and square connection design.

    Customized abutments are supplied with an abutment screw and provided non-sterile.

    AI/ML Overview

    The provided text describes the regulatory clearance (K234112) for the ARUM DENTISTRY Co., Ltd. Customized Abutment, particularly focusing on its substantial equivalence to a predicate device (K223634). The information provided does not outline acceptance criteria for an AI/ML powered device, nor does it detail a study proving the device meets AI/ML specific acceptance criteria.

    The information provided pertains to a traditional medical device (dental abutments) and focuses on engineering and material performance rather than AI/ML algorithm performance.

    Therefore, many of the requested fields regarding AI/ML specific criteria cannot be extracted from the given text.

    However, I can provide information on the non-clinical tests conducted for the Customized Abutment device, which are analogous to acceptance criteria and performance data for this type of medical device.

    1. Table of Acceptance Criteria and the Reported Device Performance (Non-AI/ML):

    Non-Clinical Test/Acceptance CriteriaReported Device Performance
    Material BiocompatibilityDemonstrated by the reference ARUM DENTISTRY submission, K223634, using Ti-6Al-4V Eli conforming to ASTM F136 and the same materials and manufacturing processes. Implied to meet ISO 10993-5 and ANSI/AAMI ST72.
    Sterilization ValidationThe Customized Abutment is delivered non-sterile for end-user sterilization. The recommended sterilization has been validated according to ISO 17655-1 and applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015." The worst-case construct was tested, and results demonstrated equivalence to the predicate devices.
    MRI SafetyNon-clinical worst-case MRI Review was performed. Evaluation was based on scientific rationale and published literature, considering the entire system (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
    Mechanical Performance (Static and Fatigue)Demonstrated through static and fatigue testing according to ISO 14801. (Specific numerical results are not provided in the text, but the successful completion of these tests indicates meeting performance requirements for the device type.)
    Compatibility with Implant SystemsReverse engineering of OEM implant bodies, OEM abutments, and OEM abutment screws was performed to confirm compatibility with listed implant systems (Zimmer Tapered Screw-Vent® and Straumann® Bone Level Tapered Implant).
    Manufacturing Process ControlAll digitally-designed Customized abutments are intended to be sent to an ARUM DENTISTRY validated milling center for manufacture. This implies adherence to established quality control for manufacturing precision.

    2. Sample size used for the test set and the data provenance:

    • Mechanical Testing (Static and Fatigue): The document mentions "The worst-case construct was tested" for sterilization and that "static and fatigue testing according to ISO 14801" was conducted. Specific sample sizes are not provided.
    • Data Provenance: The non-clinical tests were conducted internally or by contractors to ARUM DENTISTRY. The document does not specify a country of origin for the data beyond ARUM DENTISTRY Co., Ltd. being in the Republic of Korea. These are most likely prospective tests conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable and not provided as the approval is for a physical dental abutment, not an AI/ML algorithm requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable and not provided. "Adjudication" typically refers to resolving discrepancies in expert labeling or diagnoses for AI/ML ground truth, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. No MRMC study was conducted as the device is a physical dental abutment, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a physical product, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the non-clinical tests, the "ground truth" is defined by the standards and specifications used (e.g., ASTM F136 for material, ISO 14801 for mechanical testing, ISO 17655-1 for sterilization, FDA guidance for MRI safety). The device's performance is compared against the requirements of these established standards.

    8. The sample size for the training set:

    This is not applicable. The device is a physical product, not an AI/ML model that requires a training set. The device design parameters are "trained" through engineering design principles and verified through physical testing.

    9. How the ground truth for the training set was established:

    This is not applicable. There is no training set as it's not an AI/ML algorithm. The design of the abutment relies on established engineering principles, material science, and compatibility specifications with existing implant systems.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200355
    Date Cleared
    2020-07-22

    (160 days)

    Product Code
    Regulation Number
    872.3630
    Why did this record match?
    Reference Devices :

    NobelParallel™ Concial Connection, K050258, Groovy Implants, K011028, Screw-Vent Dental Implant System, K112160

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Duranext Abutments are intended for use with dental implants as a support for single or multiple toth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

    Device Description

    Duranext Abutments from Elegant Direct Corp. are a line of machinable blanks incorporating interface features compatible with eleven (11) endosseous dental implant system platforms (three (3) designs from two (2) manufacturers) and intended to be milled at an Elegant Direct Corp. validated milling center to produce patient-specific dental implant abutments. The subject device platform diameters range from 3.0 mm to 6.0 mm, and the corresponding compatible implant body diameters range from 3.0 mm to 6.0 mm.

    Duranext Abutments are designed for fabrication of custom titanium alloy dental implant abutments by a CAD/CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. The portion of each abutment available for milling is 9.5 mm in diameter and 20 mm long. The apical end is premanufactured to fit the compatible implant platform, as shown above, and is available in an engaging (anti-rotation) design. A feature at the coronal end of the abutment is provided to interface with the milling equipment. Each abutment is provided with a screw designed to fit the compatible implant. The patient-specific abutment is intended to support a cement-retained single crown or multi-unit restoration.

    AI/ML Overview

    The provided document is a 510(k) summary for the Duranext Abutments, a dental device. It does not describe a study involving an AI algorithm or human-in-the-loop performance. Therefore, I cannot extract the information required for questions about AI performance, multi-reader multi-case studies, or specific details of ground truth establishment for AI training/testing.

    However, I can provide information about the acceptance criteria and the non-clinical study that proves the device meets those criteria, as well as general device information.

    Here's what can be extracted based on the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" in the format of a diagnostic test (e.g., sensitivity, specificity). Instead, substantial equivalence is claimed based on performance testing and comparison to predicate devices for mechanical properties, biocompatibility, and sterilization. The acceptance criteria are implicitly met by successful completion of these tests in accordance with relevant ISO standards.

    Acceptance Criterion (implicitly met by standard)Reported Device Performance (Summary)
    SterilizationSterilization according to ISO 17665-1 (demonstrated readiness for sterilization)
    BiocompatibilityBiocompatibility according to ISO 10993-12 (for general requirements), and specifically ISO 10993-5 (cytotoxicity)
    Compatibility with OEM ImplantsReverse engineering of OEM implant bodies, OEM abutments, and OEM abutment screws (confirmed compatibility)
    Mechanical Strength / Dynamic FatigueStatic compression and compression fatigue testing according to ISO 14801 (sufficient strength for intended use)
    Design Parameters (for CAD/CAM abutments)Minimum wall thickness: 0.5 mm
    Minimum post height: 4.0 mm
    Maximum abutment height from prosthetic platform: 20.0 mm
    Maximum gingival height: 4.0 mm
    Minimum gingival height: 0.5 mm
    Angulation: 0° to 30°

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not mention "test sets" in the context of diagnostic data. The "testing" refers to non-clinical performance testing of the physical abutment blanks. The sample sizes for these specific engineering tests (e.g., number of abutments tested for fatigue) are not provided in this summary. The provenance is implied to be from the manufacturer's testing or a contracted lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable, as this is a physical medical device (dental abutment) and the evaluation is based on non-clinical performance testing against engineering standards, not diagnostic interpretation by experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in diagnostic studies, which is not what this document describes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental abutment, not an AI-powered diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by engineering standards and physical measurements. For example:

    • Biocompatibility: Established by adherence to ISO 10993 series standards, which involves methods like cytotoxicity testing.
    • Mechanical Strength: Established by dynamic fatigue testing according to ISO 14801, which defines acceptable load cycles and failure modes.
    • Dimensional Compatibility: Established by reverse engineering and direct measurement against OEM specifications.

    8. The sample size for the training set

    Not applicable. This device is not an AI algorithm, so there is no training set in the context of machine learning.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no training set. The "ground truth" for the device's design and manufacturing parameters is established via engineering specifications, material properties, and adherence to relevant ISO standards, rather than a labeled dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173908
    Manufacturer
    Date Cleared
    2018-06-15

    (175 days)

    Product Code
    Regulation Number
    872.3630
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

    All digitally designed custom abutments for use with Aurum™ Abutment or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture.

    Device Description

    DESS Dental Smart Solutions subject devices include four abutment design types (Aurum Base, Pre-milled Blank, CoCr Pre-milled Blank, CoCr Abutment) and one screw type (Aurum Base Screw). Abutments are provided in ten abutment connections compatible with eleven implant platform diameters range from 3.3 mm to 6.5 mm. Corresponding implant body diameters range from 3.25 mm to 6.0 mm. All abutments are provided non-sterile.

    AI/ML Overview

    The document describes the DESS Dental Smart Solutions, which are dental implant abutments. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

    Here's an analysis of the provided information regarding acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria in a typical tabular format with specific numerical targets. Instead, it relies on demonstrating equivalence to predicate devices through various performance tests and comparisons. The "acceptance criteria" are implied by the successful completion of these tests and the determination that the device is "substantially equivalent" to already marketed devices.

    However, some design parameters are mentioned as remaining the same or being comparable to the predicate devices, which can be seen as implicit performance criteria:

    Feature / Performance MetricAcceptance Criteria (Implied by Equivalence to Predicate)Reported Device Performance
    SterilizationSAL of 10^-6 according to ISO 17665-1 and ISO 17665-2"Sterilization to an SAL of 10^-6 according to ISO 17665-1 and ISO 17665-2 to ensure sterilization of the final finished device" (Successful)
    BiocompatibilityNon-cytotoxic according to ISO 10993-1 and ISO 10993-5"Biological evaluation according to ISO 10993-1 and cytotoxicity testing to ISO 10993-5 to demonstrate that all devices are non-cytotoxic" (Successful)
    Minimum Wall ThicknessAurum Abutment: 0.4 mm
    Pre-milled Blank: 0.45 mm
    CoCr Abutment: 0.4 mm (Same as primary predicate K170588)Aurum Abutment: 0.4 mm
    Pre-milled Blank: 0.45 mm
    CoCr Abutment: 0.4 mm (Stated to remain the same as primary predicate K170588, implying conformity to existing successful designs)
    Minimum Post Height (single-unit)Aurum Abutment: 4.0 mm
    Pre-milled Blank: 4.0 mm
    CoCr Abutment: 4.0 mm (Increased from previous version, but deemed substantially equivalent to K092341 at 4mm)Aurum Abutment: 4.0 mm (Stated as the new minimum for single-unit restorations, deemed substantially equivalent to K092341's titanium post height which also requires additional castable component to create 4mm)
    Pre-milled Blank: 4.0 mm
    CoCr Abutment: 4.0 mm (Stated to remain the same as primary predicate K170588)
    Maximum Gingival HeightAurum Abutment: 6.0 mm
    Pre-milled Blank: 6.0 mm
    CoCr Abutment: 6.0 mm (Same as primary predicate K170588)Aurum Abutment: 6.0 mm
    Pre-milled Blank: 6.0 mm
    CoCr Abutment: 6.0 mm (Stated to remain the same as primary predicate K170588)
    Maximum Total Abutment HeightPre-milled Blank: 19 mmPre-milled Blank: 19 mm
    Angulation CorrectionNo angulation correction (Same as primary predicate K170588)All abutments are for straight abutments only (implies no angulation correction, same as primary predicate K170588)
    Chemical CompositionMaterials used are comparable to predicate devices (Titanium Alloy, CoCr, Zirconia)Aurum Base (Titanium Alloy ASTM F136), CoCr Pre-milled Blank (CoCr ASTM 1537), Ti Pre-milled Blank (Titanium Alloy ASTM E136), CoCr Abutment (CoCr ASTM 1537). Screws are Titanium Alloy. (Found to be substantially equivalent to predicates)
    Physical DimensionsEncompasses the same range of physical dimensions including diameter and design of abutments as predicate devices."The subject device and predicate devices encompass the same range of physical dimensions, including diameter and design of the abutments." (Stated as a conclusion of equivalence)
    Packaging/Sterilization MethodsSimilar to predicate devices."The subject and predicate devices are packaged in similar materials and are to be sterilized using similar methods." (Stated as a conclusion of equivalence)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document states: "Non-clinical testing data submitted to demonstrate substantial equivalence included: sterilization... biological evaluation... and compatibility analysis by reference to K170588."

    • The report does not specify sample sizes for the sterilization or biocompatibility tests.
    • The data provenance is implied to be from the manufacturer's (Terrats Medical SL) internal testing as part of their submission for regulatory clearance. It's not explicitly stated whether the data is retrospective or prospective, or the country of origin of the raw data, beyond the manufacturer being from Spain. The "compatibility analysis by reference" means using existing data/information from the predicate device (K170588).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information – number and qualifications of experts for ground truth – is typically relevant for studies involving subjective assessments, like image interpretation in AI/ML medical devices. This document is for a physical dental implant abutment, and the tests performed are objective, non-clinical tests (sterilization, biocompatibility, mechanical properties based on design parameters). Therefore, this information is not applicable and not provided in the submission.

    4. Adjudication Method for the Test Set

    As the tests are objective non-clinical tests, an adjudication method for a "test set" (in the context of expert review) is not applicable and not mentioned. The results of the non-clinical tests would either meet or not meet the specified standards (e.g., SAL, non-cytotoxicity).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This type of study is specifically relevant for AI/ML-driven diagnostic or interpretative devices involving human readers. The device described here is a physical dental implant abutment, not an AI/ML diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No standalone algorithm performance study was done. This is not an AI/ML device. The "CAD/CAM process" mentioned refers to computer-aided design and manufacturing for custom abutments, which is a manufacturing process, not an analytical algorithm for diagnosis or interpretation.

    7. The Type of Ground Truth Used

    For the non-clinical tests conducted:

    • Sterilization: The ground truth is the scientific standard for sterility, defined as a Sterility Assurance Level (SAL) of $10^{-6}$ based on ISO 17665-1 and ISO 17665-2.
    • Biocompatibility: The ground truth is the absence of cytotoxicity, determined by adherence to ISO 10993-1 and ISO 10993-5.
    • Design Parameters/Mechanical Equivalence: The "ground truth" for design parameters (e.g., wall thickness, post height) is derived from engineering specifications and comparison to the proven safety and effectiveness of the legally marketed predicate devices (K170588 and other reference devices). The ultimate ground truth effectively is that the device performs equivalently to previously approved devices.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical device, not an AI/ML algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As this is not an AI/ML device, there is no training set or associated ground truth establishment process in that context.