Search Results
Found 4 results
510(k) Data Aggregation
(79 days)
NobelProcera Zirconia Implant Bridge (previously cleared per K202452) The NobelProcera® Zirconia Implant Bridge are indicated for use as a bridge anatomically shaped and/or framework in the treatment of partially edentulous jaws for the purpose of restoring chewing function.
TiUltra Implants and Xeal Abutments (previously cleared per K202344) NobelActive TiUltra NobelActive TiUltra implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. Nobel Active Tilltra implants are indicated for single or multiple unit restorations in splinted applications. This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. NobelActive TiUltra 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central incisor in the mandible. Nobel Active TiUltra 3.0 implants are indicated for single-unit restorations only. NobelReplace CC TiUltra NobelReplace CC TiUltra implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The NobelReplace CC TiUltra implants are indicated for single or multiple unit restorations. The NobelReplace CC Tilltra implants can be used in splinted or non-splications. The NobelReplace CC TiUltra implant may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied. NobelParallel CC TiUltra NobelParallel CC TiUltra implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function. NobelParallel CC TiUltra implants are indicated for single or multiple restorations in splinted applications. This can be achieved by a 2-stage or 1-stage surgical techniques in combination with immediate, early of delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. Implants with
Not Found
This document is a 510(k) premarket notification decision letter from the FDA to Nobel Biocare AG regarding their Dental Implant Systems Portfolio - MR Conditional. It explicitly states that the letter covers indications for use and general controls, but does not contain information about acceptance criteria or performance studies for the device itself.
Therefore, I cannot provide the requested information for the following reasons:
- Acceptance Criteria and Performance Data: The document is a regulatory clearance letter, not a clinical study report. It does not contain acceptance criteria for device performance, nor does it present any data from studies proving the device meets particular criteria. The letter confirms substantial equivalence to legally marketed predicate devices, which means the FDA has determined the device is as safe and effective as a previously cleared device, not that specific performance metrics were tested and met in a new study.
- Study Details (Sample size, data provenance, experts, adjudication, MRMC, Standalone, Ground Truth, Training Set): Since no performance study data is included in this FDA 510(k) clearance letter, none of these details can be extracted. The document refers to various previously cleared predicate devices (e.g., K202452, K202344, K181869), but it doesn't describe the studies that led to their clearance.
In summary, the provided text does not contain the information necessary to describe acceptance criteria or a study proving the device meets those criteria.
Ask a specific question about this device
(79 days)
The Panthera Dental Milled Bar is indicated for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient for purpose of restoring chewing function. It is intended for use to support multiple tooth prostheses in the mandible or maxilla. The prostheses can be screw retained.
The Panthera Dental Milled Bars are indicated for compatibility with the Zimmer Tapered Screw-Vent System for sizes 3.5 and 4.5.
The Panthera Dental Milled Bar is a metallic dental restorative device which is intended for attaching by screw retention to dental implants to aid in the treatment of partial and totally edentulous patients for the purpose of restoring chewing function.
The Panthera Dental milled bars for which clearance is requested, are included in one of the following bar types, which have distinct design specifications.
The Type I bars are specific for removable overdenture and include:
- Panthera Dental Dolder Bar, Hader Bar, Milled Bar, REBourke Bar and Paris Bar.
The Type II bars are specific for fixed prostheses and include:
- Panthera Dental Wrap-around Bar, Montreal Bar, Montreal Bar with metallic lingual, Pin Lingual Bar and Pin Wrap-Around Bar.
The Panthera Dental Milled Bar is designed to match an individual patient. Panthera Dental designs the bar from a three-dimensional optical and/or digital scanner system that scans the casting of a patient's impression: the model cast is prepared beforehand by the dental professional using standard dental techniques. The designed bar is then machined using a computer-aided design/ computer-aided manufacturing software system (CAD/CAM). The bar is milled from titanium (Ti-6AI-4V grade 5). CAD/CAM fabrication is only performed by Panthera Dental, within our manufacturing control and not by the dental laboratory.
The Panthera Dental Milled Bar is packaged as non-sterile, and delivered to a dental laboratory for completion. Once received at the laboratory, the Panthera Dental Milled Bar is matched to a denture for final placement.
The Panthera Dental Milled Bar provides retention and support for a removable or fixed denture made of standard laboratory dental materials such as resin composite.
The provided text is a 510(k) summary for a medical device (Panthera Dental Milled Bars), which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving acceptance criteria for an AI/CADe device.
Therefore, this document does not contain the information required to answer your request about acceptance criteria and a study proving device performance for an AI/CADe device.
Here's why and what kind of information would be needed:
- Acceptance Criteria & Reported Device Performance: This document only provides a comparison of design specifications between the applicant's device and a predicate device (Tables 5.1, 5.2, 5.3) for physical dimensions and material. It does not list "acceptance criteria" in the way one would for an AI/CADe system (e.g., sensitivity, specificity, AUC thresholds).
- Study That Proves the Device Meets Acceptance Criteria: The document explicitly states: "Human clinical study was not deemed necessary to support substantial equivalence." and "Panthera Dental Inc. performed no clinical testing." It discusses non-clinical testing like fatigue testing, biocompatibility, and process capability for the physical dental bar, not an AI algorithm.
- Missing Information for AI/CADe devices: The document lacks all the specific details you requested for an AI/CADe study, such as:
- Sample sizes for test sets (it mentions "thirty (30) cylinders" for process capability testing, not an AI test set).
- Data provenance (country of origin, retrospective/prospective).
- Number and qualifications of experts for ground truth.
- Adjudication method.
- Multi-reader multi-case (MRMC) comparative effectiveness study.
- Standalone algorithm performance.
- Type of ground truth (pathology, outcomes data, etc.).
- Training set sample size and ground truth establishment for the training set.
In summary, the provided document describes a physical dental device, not an AI/CADe system, and thus does not contain the information requested about AI performance metrics or related studies.
Ask a specific question about this device
(306 days)
Implant Direct Custom Bars are patient specific devices indicated for attachment to dental implants in the treatment of partially or fully edentulous jaws for the purpose of restoring chewing function.
Custom bars are compatible at the implant level with InterActive (3.4mm Platform) & SwishActive (3.4 Platform) System implants.
Custom bars are compatible at the abutment level with InterActive (3.4mm Platform) & SwishActive (3.4 Platform) system straight multi-unit abutments.
Implant Direct Custom Bars are computer-aided design/computer-aided manufacturing, individually designed prosthetic devices for partially or fully edentulous restorations. The bars and fixation screws are made from Wrought Titanium Alloy conforming to an FDA recognized consensus standard ASTM F136. The Custom Bars consist of two device types: fixed-detachable frameworks and overdenture bars. The bar systems consists of the bars as well as fixation screws to facilitate the attachment to the dental implants.
Implant Direct Custom Bars are attached directly to an implant to provide support for the fabricated denture. The implant specific interfaces on the Custom Bars are designed and manufactured with precise mating features in order to the mating surfaces of the implants or abutments.
The Implant Direct Custom Bars are designed and produced at Implant Direct under the direction of clinical professionals. The patient requirements and implant locations are obtained conventionally or digitally and a bar is designed, using CAD software, according to established customization parameters. Once the design is finished, the bar design is sent to the clinician for approval before manufacture. After bar design approval is received, the bar design is milled using a precise CAM system, inspected, cleaned, packaged non-sterile, and shipped to the customer where it is finished into the final restoration.
This document describes the premarket notification for "Custom Bars" (K162633), which are patient-specific dental devices. The device is a physical product, not an AI/ML powered medical device. Therefore, the questions related to AI/ML device performance, such as sample sizes for test and training sets, expert ground truth establishment, MRMC studies, standalone performance, and data provenance, are not applicable.
However, I can extract information related to the performance data and the study conducted for this non-AI/ML device:
1. Table of Acceptance Criteria and Reported Device Performance:
Parameter | Acceptance Criteria | Reported Device Performance |
---|---|---|
Mechanical Testing | ||
- Fatigue Testing (ISO 14801) | Bar interface successfully completes endurance testing and is equivalent to the predicate device. | Results indicated that the bar interface successfully completed endurance testing and was equivalent to the predicate device. |
- Bending Strength (modified ASTM F382) | Device is equivalent to the predicate device. | Results indicated that the subject device is equivalent to the predicate device. |
Biocompatibility Testing | Device meets all biocompatibility requirements for its intended use (ISO 10993-1 and ISO 10993-5). | Cytotoxicity testing and comprehensive biocompatibility evaluation was conducted. Results indicate that the device met all biocompatibility requirements for its intended use. The material conforms to FDA recognized consensus standard ASTM F136 and is equivalent to the reference predicate. |
Steam Sterilization Validation | Acceptance criteria are met, and devices can be sterilized per Gravity Displacement and Pre-vacuum cycle parameters (ISO 17665-1 and ISO 17665-2). | Results indicated the acceptance criteria was met and the devices can be sterilized per Gravity Displacement and Pre-vacuum cycle parameters. (Performed on worst-case sample) |
Regarding the other requested information, since this is not an AI/ML medical device, the following points are not applicable or not provided in the document:
- Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for this type of mechanical and biocompatibility testing. The testing was performed on "worst-case configurations" and "worst-case sample" as described.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. "Ground truth" in the context of expert consensus, pathology, or outcomes data is typically for diagnostic/screening devices, not for physical dental prosthetics.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for physical device performance testing against laboratory standards.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/ML device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI/ML device.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): For this physical device, the "ground truth" or standard for acceptance is adherence to recognized international standards (ISO 14801, ASTM F382, ISO 10993-1, ISO 10993-5, ISO 17665-1, ISO 17665-2) and equivalence to predicate devices, as demonstrated through non-clinical laboratory testing.
- The sample size for the training set: Not applicable, as this is not an AI/ML device.
- How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.
Additional Information from the document:
- Study Type: Non-clinical testing (mechanical, biocompatibility, and cleaning/sterilization validation). Clinical performance testing was not performed and was deemed not required to support substantial equivalence.
- Predicate Device for Comparison: Dentsply International ISUS Implant Suprastructures (K122424) was the primary predicate. Nobel Biocare NobelProcera Overdenture Bar (K132749) and Biomet 3i CAM StructSURE Overdenture Bars (K101582) were used as references.
- Conclusion: The studies concluded that the Custom Bars are substantially equivalent to the legally marketed predicate devices based on the non-clinical testing performed.
Ask a specific question about this device
(195 days)
The TREFOIL System is used to restore chewing function in fully edentulous mandibles.
The three implants of the TREFOIL System are placed between the mental foramina in fully edentulous mandibles in a 1-stage surgical technique combined with an immediate function loading protocol, provided sufficient primary stability for the selected technique is achieved. In cases where sufficient primary stability for two implants or more is not reached, the implants along with the Framework may also be used with an early or delayed loading protocol.
The following prerequisites must be fulfilled:
- Adequate quantity of bone (minimum height of 13 mm and minimum width of 6-7 mm).
- Adequate mouth opening (minimum 40 mm) to accommodate the guided surgery instruments.
- Implant-supported prosthetics seated directly on dedicated implants
The TREFOIL System is a method of placing three dental implants in predetermined positions (between the mental foramina) and using a pre-designed prosthetic bar to act as a screw-retained framework seated on the implants. The TREFOIL System restores chewing function and esthetics in the mandible in completely edentulous patients.
The TREFOIL System consists of dental implants, surgical components necessary to place the implants in predetermined positions, and prosthetic components that are included in the prosthetic bar or are used in the dental lab during the creation of the prosthetic bar.
The dental implants are threaded endosseous implants made of CP4 titanium. The implant is parallel walled and has an internal conical abutment connection. The implant is available in 11.5 mm length with two available collar lengths (4.5, 6.0 mm). The apex of the implants have cutting chamber allowing for self-tapping. The implant bone interface has the TiUnite implant surface treatment. The TREFOIL System bar is made of titanium vanadium alloy. The TREFOIL System surgical tooling is made of stainless steel.
The provided text describes a 510(k) premarket notification for the TREFOIL System, an endosseous dental implant system. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study.
Therefore, much of the requested information about acceptance criteria, detailed study design, sample sizes, ground truth establishment, expert qualifications, and MRMC studies is not explicitly available in the provided text.
However, I can extract the information related to the performance data presented and what it aimed to establish.
Here's a summary based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state quantitative acceptance criteria in a table format, nor does it provide numerical performance results against such criteria. Instead, it describes types of testing performed and their general conclusions.
Acceptance Criterion (Inferred from testing) | Reported Device Performance |
---|---|
Surgical Accuracy (Angular and Translational) | Established the angular and translational accuracy errors generated by the surgical procedure, indicating the system's ability to maintain precision during implant placement. |
Framework Compensation Mechanism (Passive Fit) | The ability of the Trefoil framework compensation mechanism to adapt to expected angular and translational accuracy errors was established through 3D modeling. The modeling confirmed that the compensation mechanism ensures a passive fit of the framework bar on the implants. |
Fatigue Strength (Withstand Mastication Forces) | Fatigue testing was performed to establish that the TREFOIL System will withstand foreseeable mastication forces. The results demonstrated a higher fatigue strength than the predicate Branemark Novum (K000018). |
Sterilization Efficacy | For sterile devices, the sterilization method (Gamma radiation) is the same as the predicate and was validated in accordance with ANSI/AAMI/ISO 11137. For non-sterile devices (for end-user sterilization), pre-vacuum and gravity moist heat sterilization were validated in accordance with ANSI/AAMI/ISO ISO 17665-1. No additional testing was required. |
Packaging Integrity | The packaging for the subject device is the same as the predicate (titanium cylinder in plastic vial with PVC shrink-wrap and tamper-resistant strip). No additional testing was required. |
Shelf Life | The packaging for the subject device is the same as the predicate and has a 5-year expiration date, determined by real-time aging. No additional testing was required. |
Biocompatibility | The device is manufactured from the same material using the same method as the predicate, has the same intended use, and patient contact type/duration. No additional testing was required. |
Surface Treatment Qualification | The Trefoil implants use the same TiUnite surface treatment as the predicate NobelActive implants. This treatment was qualified through Auger analysis, IR spectrum analysis, and cytotoxicity testing. No additional testing was required. |
2. Sample Size for the test set and data provenance:
- Surgical Accuracy Test: Implants were placed "into bone models." The specific number of models or implants is not stated.
- Data Provenance: The nature of the studies suggests these were retrospective (laboratory-based) tests conducted by the manufacturer, rather than prospective clinical trials with patient data. The country of origin for the data is not specified but is presumably where Nobel Biocare AB conducts its research.
3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:
This information is not provided in the document. The tests described (surgical accuracy in bone models, 3D modeling for passive fit, fatigue testing) are physical or computational tests, not human-interpretation-based tests that would require expert ground truth.
4. Adjudication method for the test set:
This information is not applicable/not provided. The assessment of these non-clinical tests would have been based on measurement and engineering analysis, not expert adjudication of subjective findings.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The device is an endosseous dental implant system, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or mentioned.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This is not applicable. The device is a physical medical device (dental implant system), not an algorithm or software.
7. The type of ground truth used:
- Surgical Accuracy Test: Ground truth would have been based on physical measurements and engineering specifications of the implant placement within the bone models (e.g., deviation from planned position and angle).
- Framework Compensation Mechanism: Ground truth was established through three-dimensional modeling and simulation, confirming the design's ability to achieve a "passive fit" based on engineering principles.
- Fatigue Testing: Ground truth was based on material science and engineering standards (ISO 14801) and the device's ability to withstand a specified number of cycles at a given load.
- Other tests (Sterilization, Packaging, Shelf Life, Biocompatibility, Surface Treatment): Ground truth was established through adherence to recognized industry standards (e.g., ISO, ANSI/AAMI), analytical chemistry (Auger analysis, IR spectrum), and established biological testing methodologies (cytotoxicity).
8. The sample size for the training set:
This information is not applicable/not provided. The device is not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established:
This information is not applicable/not provided, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1