Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    K Number
    K243066

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-16

    (80 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K161497, K233231, K163194, K191191, K072878, K130808

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ditron's Dental Implants and Abutments are indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the chewing function of patients with fully or partially edentulous:

    • Two stages: MPI, ULT, API, CPI, MPC, APC, CPC and UPC models .
    • . One stage: OPI and TPI models

    The 3.3mm and 3.0mm diameter implants (OPI and TPI. MPI and API) are intended only for the incisors and cuspids of the maxilla and mandible. They are also indicated for denture stabilization using multiple implants.

    The Two stages and One stage implants are indicated for temporary or long-term use. They are selftapping titanium threaded screws indicated for long term intra bony applications. They permit immediate splint stability and long-term fixation of new or existing crown, bridge and prosthesis and protection of graft sites.

    MPI, ULT, API, CPI, OPI, TPI, MPC, APC, CPC and UPC models are indicated for immediate loading (except for MPI. and API 6mm length) in single- and multi-tooth restoration, when good primary stability is achieved and with appropriate occlusal loading.

    The 30-degree multi-unit abutments must be used within 45 degrees of parallelism for a splinted restoration.

    The 17-degree multi-unit abutments must be used within 32 degrees of parallelism for a splinted restoration.

    The Temporary Titanium and PEEK Abutments are indicated to be used on Ditron implants to provide temporary support for prosthesis structure for up to 6 months. They can be used in one or two stage procedures and also immediate load when there is good primary stability.

    Device Description

    This submission covers changes related to Ditron's Dental implants and abutments. The inclusion of additional product variations aims to provide dental surgeons with a broader range of implant and abutment options for patient treatment.

    The requested additions to Ditron's Dental Implants and Abutments within this 510(k) are hereby described:

    . An additional implant types:

    • MPC (Conical Connection Molecular Precision Implant) o
    • O APC (Conical Connection Advanced Precision Implant)
    • UPC (Conical Connection Ultimate Precision Implant) O
    • CPC (Conical Connection Cylindrical Precision Implant) O

    The Conical Connection (CC) implants include a conical connection platform (Morse taper). The intended use, materials, drilling protocol, thread designs and bone interface of the subject Conical Connection implants (MPC, UPC, APC and CPC) are identical to Ditron's cleared Internal Hex connection platform implants (MPI, ULT, API and CPI, respectively). The CC implants include color anodization for aesthetic purposes. The CC compatible Cover and Prosthetic Screws are color anodized as well.

    The MPC/APC/CPC/UPC Implants are available in diameters of 3.5mm, 4.2mm, 5.0mm, and 6.0mm, with lengths of 8.0mm, 11.5mm, 12mm, 13mm, 14mm, and 16mm. Additionally, a 6.5mm length is offered for MPC/APC/CPC implants in 4.2mm, 5.0mm, and 6.0mm diameters, and a 7.0mm length is available for UPC implants in 4.2mm, 5.0mm, and 6.0mm diameters.

    • An additional Abutment type: CC Healing Caps and SRA Healing Caps Abutments The CC Healing Caps are color-anodized for aesthetic purposes. They are used for the maintenance of the soft tissue during the osseointegration phase of Ditron CC implants to be rehabilitated using technique. They are available in diameters of 3.5, 4.5 and 5.5mm and the following gingival heights: 0.8, 1.5, 2.5, 3.5 and 4.5mm.
    • An additional Abutment type: CC Cement Retained Abutments the CC Anatomic/Universal . straight/angulated abutments are Cement-Retained Abutments. They are intermediary prosthetic components to be installed onto Ditron's CC implants to support the final prosthesis. They are available as straight (0°) or angulated (15° and 25°) in diameters: 3.5, 4.5 and 5.5mm and the following gingival heights: 0.8, 1.5, 2.5, 3.5, 4.5 mm.
    • . An additional Abutment type: CC Single-Unit and Multi-Unit Abutments - The CC Screw-retained Single Unit Abutment and Multi Unit Abutments are intermediary prosthetic components to be installed onto CC Ditron's implants to support the final prosthesis. They are available as straight (0°) or angulated (17° and 30°) and the following gingival heights: 0.8, 1.5, 2.5, 3.5, 4.5, 5.0 mm. The Single Unit abutment height is up to 4.5mm only.
    • An additional Abutment type: CC Temporary Titanium Abutments The CC Temporary Titanium Abutments are temporary intermediary prosthetic components to be installed onto the Ditron's CC implants to support the provisional prosthesis for up to 6 months. They are available as straight (0°) in diameters: 3.5 and 4.5mm and the following gingival heights: 0.8, 1.5, 2.5, 3.5, 4.5mm.
    • An additional Abutment type: CC Temporary PEEK Abutments The CC Temporary PEEK Abutments are temporary intermediary prosthetic components to be installed onto Ditron's CC implants to support the provisional prosthesis for up to 6 months. They are composed of a customizable cylindrical body made of PEEK and a non-customizable base made of titanium. They are available in diameters of 4.5 and 6.0mm and the following gingival heights: 0.8, 1.5, 2.5, 3.5, 4.5, 5.5mm.
    • An additional Abutment type: CC Liberator Overdenture Abutment The CC Liberator Overdenture Abutment is similar in design to Ditron's cleared Liberator abutment, except for the addition of Titanium Nitride (TiN) coating on the subject device. It is available in the following gingival heights: 0.5, 1.0, 2.0, 3.0, 4.0, 5.0 and 6.0mm.
    • An additional Abutment type: SRA Titanium Coping Abutments The SRA Titanium Coping . Abutment is suitable for use with the CC Screw-retained single Abutments. It connects to the abutment using one of the designated abutment screws. They are available in short (L4.85mm) and long (L11mm) versions.
    • Modification to Cleared Titanium Abutments Addition of Laser Marking for identification as an option for all of the Ditron Dental cleared Titanium Abutments (K140728, K161497, K233231).
    • Modification to Cleared Liberator and TPI Overdenture Abutments Addition of TiN coating to . Ditron's cleared Liberator abutment (K161497) and the cleared TPI Liberator Overdenture Abutment (K233231).
    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving that an AI/ML device meets those criteria. The document is a 510(k) summary for Dental Implants and Abutments, focusing on demonstrating substantial equivalence to predicate devices based on technological characteristics and non-clinical performance data. It does not mention any AI/ML components or related studies.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size used for the test set and data provenance.
    3. Number of experts and their qualifications for ground truth.
    4. Adjudication method.
    5. MRMC comparative effectiveness study results (effect size).
    6. Standalone performance study for an algorithm.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212364

    Validate with FDA (Live)

    Date Cleared
    2022-08-10

    (376 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BLT Dental Implant Systems are intended to be placed in the upper and lower jaw to support prosthetic devices and to restore a patient's chewing function. BLT Dental Implant systems are also indicated for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants through the corresponding components (abutments). In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.

    Device Description

    The proposed devices, BLT Dental Implant Systems, are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients. The proposed devices can also be used for immediate or early implantation following extraction or loss of natural teeth. The proposed device contains dental implant, healing cap and abutment. The dental implant system is available in two types, NC type and RC type.

    AI/ML Overview

    The request asks for acceptance criteria and details of the study proving these criteria are met. The provided text is a 510(k) Summary for a dental implant system. This type of submission focuses on demonstrating substantial equivalence to a predicate device, primarily through non-clinical testing and comparison of characteristics, rather than a clinical study with detailed acceptance criteria and performance metrics typically found for AI/ML devices. Therefore, much of the requested information (like effect size, ground truth, sample size for training/test sets, expert qualifications, and adjudication methods) is not applicable or unavailable from this document.

    However, I can extract information related to the device's performance based on non-clinical testing and the acceptance criteria implicitly defined by compliance with recognized standards and demonstrated equivalence to the predicate device.

    Here's the summary based on the provided document:

    Acceptance Criteria and Study Details for BLT Dental Implant System (K212364)

    The "acceptance criteria" for this device are implicitly defined by its compliance with various ISO and ASTM standards, and its demonstrated substantial equivalence to a legally marketed predicate device (K150388) and several reference devices. The "study" proving these criteria is a series of non-clinical tests and a comparison of technological characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicit via Standards/Equivalence)Reported Device PerformanceStudy that Proves this Performance
    Mechanical PerformanceNo significant difference in dynamic fatigue performance compared to predicate device. (Aligned with ISO 14801)"The test result does not show any significant difference." (between proposed device and predicate)Mechanical test performed according to FDA guidance and ISO 14801.
    BiocompatibilityMaterials are biocompatible and do not cause adverse effects. (Aligned with ISO 10993 standards)"The subject device is as safe... as the legally marketed predicate device..." as materials are identical to previously cleared device (K192274).Leveraged prior biocompatibility testing for K192274, as materials are identical in formulation, processing, and sterilization. Formal compliance with ISO 10993-1, -5, -10, -11 mentioned.
    SterilizationAchieves a Sterility Assurance Level (SAL) of $10^{-6}$. Maintain package integrity. (Aligned with ISO 11137-2)"Sterilized by irradiation to achieve a SAL of $10^{-6}$." "Vaccuum leak test and sterility test were provided to verify the package integrity."Sterilization method validated in accordance with ISO 11137-2. Package process is the same as K150388. Complies with ISO 11137-1, ISO 11607-1, -2, ASTM D3078-02(2013), ASTM F88/F88M-15, ASTM F1886/F1886M-16, ASTM F1929-15, ASTM F2096-11.
    Material CompositionComplies with relevant material specifications (Titanium alloys).Made of Pure Titanium (for implant), Titanium Alloy (for abutment/healing cap). Meets ASTM F136-13, ASTM F67-13.Chemical analysis leveraged from K150388 as surface modification is the same. Test reports for ASTM standards cited.
    Bacterial EndotoxinMeets bacterial endotoxin limits."Bacteria endotoxin limit were evaluated for the each lot device."USP <85> Bacterial Endotoxin Test.
    PackagingPackaging integrity for sterile barrier systems.Packaging system maintains sterility.Complies with ISO 11607-1:2019, ISO 11607-2:2019, ASTM D3078-02(2013), ASTM F88/F88M-15, ASTM F1886/F1886M-16, ASTM F1929-15, ASTM F2096-11 (Reapproved 2019), ASTM F1980-07(2011).
    Shelf Life5 years.5 years.(Details not explicitly stated in document, but assumed based on predicate equivalence and compliance with accelerated aging standards like ASTM F1980-07(2011) related to packaging.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for specific non-clinical tests. Non-clinical tests like mechanical testing (ISO 14801) typically involve a specific number of samples for destructive testing, but the exact count isn't provided in the summary.
    • Data Provenance: The tests were conducted by the manufacturer, Bio Concept Co., Ltd. and leverage data from previously cleared devices where materials or processes are identical. The document does not specify the country of origin of the raw test data (e.g., where the mechanical tests were physically performed) beyond the manufacturer's location in China. The data is retrospective in the sense that prior clearance data (K192274, K150388) is referenced for aspects like biocompatibility and surface analysis.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This is not applicable as this is a non-clinical submission for a physical medical device (dental implant system), not an AI/ML device relying on expert-annotated ground truth. The "ground truth" for non-clinical tests comes from physical measurements against established standards.

    4. Adjudication Method for the Test Set

    • Not applicable as this is a non-clinical submission for a physical medical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical dental implant system, not an AI/ML device, so no MRMC study was performed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical device, not an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for establishing device performance in this context is derived from:
      • Scientific Standards: Compliance with recognized international and national standards (ISO 14801 for mechanical properties, ISO 10993 for biocompatibility, ISO 11137-2 for sterilization, ASTM standards for materials and packaging).
      • Predicate Device Equivalence: Direct comparison of technological characteristics, materials, and intended use to an existing legally marketed device (K150388) and numerous reference devices.
      • Laboratory Testing: Results from mechanical tests, sterilization validation, bacterial endotoxin tests, and packaging integrity tests.

    8. The Sample Size for the Training Set

    • Not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211952

    Validate with FDA (Live)

    Date Cleared
    2022-07-21

    (393 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTI Dental Implant System UnicCa® for oral implant surgery is to be used for the partial or total replacement of teeth in edentulate patients. Once attached to the bone, the implants act as an anchor for various fixed or removable prosthetic solutions that can be used to improve or restore a patient's mastication function.

    In the case of 5.5 - 6.5mm long UnicCa® implants should be used in a two-stage surgical procedure. These implants are indicated for delayed loading. These implants are indicated only for straight abutments and to support permanently fixed restorations.

    In the case of Tiny® 3.0 UnicCa® implants: These implants shall be used only to replace maxillary lateral incisors and mandibular lateral and central incisors. Immediate loading is recommended when there is good primary stability and an appropriate occlusal load.

    Device Description

    BTI Dental Implant System UnicCa® is a self-tapping, threaded, root form dental titanium implant provided in two types of connections: external (i.e., Externa®) and interna®), in a variety of platforms and range of diameters (3.0 – 6.0 mm) and lengths (5.5 – 18.0 mm). BTI Dental Implant System UnicCa® features an implant surface treatment that improves the hydrophilicity of the implant.

    The purpose of this 510(k) is to expand the product offering and include a new dental implant platform (Interna Narrow), and compatible abutments, together with additional implants for the already existing Universal Plus platform. Interna Narrow implants will be offered with implant diameters ranging from 3.3 to 4.75 mm and lengths of 5.5 to 15 mm. Additionally, new implants for the existing Universal Plus platform of diameter 6.0 mm, and lengths of 7.5 to 11.5 mm are included.

    AI/ML Overview

    This document describes the BTI Interna Narrow/Plus Dental Implant System UnicCa®, which is a dental implant system. The information provided is for a 510(k) premarket notification to the FDA, asserting substantial equivalence to previously cleared predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" in the typical sense of numerical thresholds for outcomes. Instead, it demonstrates substantial equivalence to predicate devices through various non-clinical tests and comparisons. The "performance" is largely reported as demonstrating "substantial equivalence" or being "identical" or "similar" to the predicate devices.

    Feature / Test CategoryAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
    Product ClassificationIdentical product codes (DZE, NHA) and regulation number (21 CFR 872.3640), Class II.Identical.
    Indications for UseIdentical to primary predicate device (K151391) for general use in partial/total replacement of teeth, specific uses for 5.5-6.5mm implants (two-stage, delayed loading, straight abutments, fixed restorations), and Tiny® 3.0 implants (maxillary lateral, mandibular lateral/central incisors, immediate loading with good primary stability/appropriate occlusal load).Identical to primary predicate device. Includes specific indications for 5.5-6.5mm long UnicCa® implants and Tiny® 3.0 UnicCa® implants that match the predicate.
    Implant DesignSimilar designs, sterilization, and packaging as identified predicate devices (K022258, K053355, K061383) for use as dental implants.Similar designs, sterilization and packaging are included in subject and the identified predicate devices, intended for similar use.
    Platform Diameter (mm)Range of implant platform diameters covered by identified predicate devices (e.g., Interna: 4.1, 5.5; Externa: 3.5, 4.1, 5.5).Similar. A new platform diameter for the Interna connection (Interna Narrow, 3.5 mm) is included, but the range is covered by predicate devices. Interna Universal Plus is identical to predicate.
    Implant Length (mm)5.5 - 18 mm, matching predicate K151391.Identical (Interna Narrow: 5.5 – 15 mm; Interna Universal Plus: 7.5 – 11.5 mm, which falls within the predicate's 5.5-18 mm range).
    Implant Body Diameter (mm)3.0 - 6.0 mm, matching predicates K151391, K022258, K053355.Identical (Interna Narrow: 3.3 – 4.75 mm; Interna Universal Plus: 6.0 mm, which falls within the predicate's 3.0-6.0 mm range).
    Implant MaterialCommercially pure titanium (Grade 4).Identical.
    Implant Surface TreatmentCalcium surface treatment.Identical to primary predicate (K151391).
    RestorationSingle and multiple restorations.Identical.
    Connection DesignEngaging, non-engaging.Identical.
    Abutment Prosthetic Diameter (mm)Range of diameters covered by identified predicate devices (K022258, K053355, K061383), e.g., 3.5 - 6.5 mm.Similar. The subject device's range (3.5 - 5.1 mm) is covered.
    Abutment Gingival Height (mm)Range of gingival heights covered by identified predicate and reference devices (K022258, K053355, K061383, K130808), e.g., 1 - 4 mm.Similar. The subject device's range (0.5 - 4 mm) is covered.
    Abutment AngulationStraight (0°) with divergence allowance supported by bench testing.Identical. Divergence allowance in subject device is supported by bench testing provided.
    Abutment MaterialCommercially pure titanium (Grade 4).Identical to identified predicate devices.
    Prosthetic Screws MaterialTi6Al4V, or other materials in predicate (e.g., gold/palladium alloy in K022258).Identical to K053355 (Ti6Al4V).
    Packaging, Sterilization, Shelf-LifeSimilar packaging and sterilization processes, sterile by gamma irradiation or non-sterile for end-user sterilization.Similar packaging as the identified predicate devices. Healing abutments/Transepithelial abutments: sterile by gamma irradiation, thermoform tray with peel top lid. Temporary Titanium abutments, Titanium abutments, Healing Caps, Aesthetic Interfaces and Screws: non-sterile, to be sterilized by end user (moist heat), thermosealed bag.
    BiocompatibilityConformance with ISO 10993-1, with satisfactory results for all biological endpoints relative to a long-term implant.All biological endpoints relative to a long-term implant as identified in the FDA guidance have been addressed with satisfactory results.
    Corrosion, Surface Hydrophilicity, TOF-SIMSLeveraged from predicate K151391. Performance should be comparable.Leveraged from K151391.
    Fatigue TestingConformance with ISO 14801 and FDA guidance document, showing substantial equivalence to predicate for intended use.Test results show that the proposed device is substantially equivalent to the predicate device for its intended use.
    Comparative Area Analyses (Bone Resorption, Bone-to-Implant Contact)Subject implants and reference device should show substantial equivalence in these aspects.Results showed that subject device is substantially equivalent.
    Sterilization ValidationSterility assurance level (SAL) of 10-6, according to ISO 11137-1, ISO 11137-2 and ISO 17665-1.Validation performed to a SAL of 10-6, according to specified ISO standards.
    Packaging/Sterile Barrier/Shelf-Life ValidationConformance with ISO 11607-1, ISO 11607-2, ASTM F88, ASTM F1886, ASTM 1929 and ASTM F2096.Validation performed according to specified ISO and ASTM standards.
    Transport ValidationConformance with ISTA 3A.Validation performed according to ISTA 3A.
    Endotoxin TestingConformance with ANSI/AAMI ST72.Testing performed according to ANSI/AAMI ST72.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document mentions "Comparative area analyses for full external area and for bone resorption of up to 3 mm have been performed for subject implants and reference device." However, specific numerical sample sizes for this or other non-clinical tests (e.g., fatigue testing, biocompatibility) are not provided.
    • Data Provenance: The document does not specify the country of origin for the data or whether it is retrospective or prospective. It refers to manufacturing in Spain (B.T.I. Biotechnology Institute, SL Jose Rivero Qualified Person Leonardo Da Vinci 14, Parque Tecnologico de Alava Minano, Alava 01510 SPAIN). The tests are non-clinical, implying laboratory or bench testing, not human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable and not provided as the submission relies on non-clinical, bench testing for substantial equivalence, not clinical study data requiring expert-established ground truth.

    4. Adjudication method for the test set:

    This information is not applicable and not provided as the submission relies on non-clinical, bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a physical dental implant system, not an AI-assisted diagnostic or therapeutic tool for which an MRMC study would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This is not applicable as the device is a physical dental implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this submission is based on established engineering and biological performance standards (e.g., ISO, ASTM standards, FDA guidance documents) and comparative data to already cleared predicate physical devices. For example:

    • Biocompatibility: Conformance with ISO 10993-1.
    • Fatigue Testing: Conformance with ISO 14801 and FDA guidance.
    • Sterilization Validation: Conformance with ISO 11137 series.
    • Physical measurements and material specifications are compared to those of the predicate devices.

    8. The sample size for the training set:

    This information is not applicable as there is no mention of a "training set" in the context of this physical device submission, which does not involve machine learning or AI models.

    9. How the ground truth for the training set was established:

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190654

    Validate with FDA (Live)

    Date Cleared
    2019-06-28

    (106 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K130808

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Healing Caps are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Healing Caps should be used only with suitable implant connections. The healing components are intended to be used up to 6 months.

    Device Description

    The Healing Caps are screws machined into 2 pieces and come in three gingival heights to accommodate individual gingival thickness. The material of the devices is yttria-stabilized tetragonal zirconia (Y-TZP) and titanium grade 4. Healing caps are screwed into the implant to protect the inner configuration in cases of transmucosal healing protocols and are placed out of occlusion and do not support a prosthetic restoration.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Straumann® CI RD Ceramic Healing Caps, asserting its substantial equivalence to previously marketed predicate devices. The document focuses on demonstrating that the new device is as safe and effective as existing legally marketed devices, rather than establishing efficacy against specific clinical outcomes. Therefore, the "acceptance criteria" and "device performance" in this context refer to engineering and biological performance characteristics compared against established standards for dental implant components.

    Here's a breakdown of the requested information based on the provided text, noting where specific details are not available in a 510(k) summary:

    1. A table of acceptance criteria and the reported device performance

    Since this is a 510(k) submission, the "acceptance criteria" are typically demonstrating compliance with relevant international standards and equivalence to predicate devices. The "reported device performance" refers to the results of engineering and biocompatibility testing.

    Acceptance Criteria CategorySpecific Criteria / Standard AppliedReported Device Performance
    SterilizationSterility Assurance Level (SAL): Minimum SAL of 10-6 per ISO 11135 (Ethylene Oxide sterilization).Endotoxin Limits: < 20 EU/device (based on FDA Guidance for blood-contacting/implanted devices).Packaging Integrity: Maintained after sterilization, handling, distribution, transport, and storage (demonstrated by ISTA 2A transportation study and packaging stability study).Sterility Assurance Level (SAL): Validation performed per ISO 11135 demonstrating reliable and consistent sterilization to a minimum SAL of 10-6.Endotoxin Limits: LAL Endotoxin Analysis performed on finished packages from a production lot after sterilization demonstrated compliance with the < 20 EU/device limit.Packaging Integrity: A transportation study per ISTA 2A demonstrated integrity of the sterile barrier system. A packaging stability study demonstrated integrity of the packaging system and sterile barrier after all relevant processes and for the defined shelf life.
    BiocompatibilityBiological safety assessed according to ISO 10993-1 guidelines.Cytotoxicity: Evaluated per ISO 10993-5.Chemical Characterization: Evaluated per ISO 10993-18.Biological Safety: Assessed according to ISO 10993-1.Cytotoxicity: Tests performed per ISO 10993-5.Chemical Characterization: Tests performed per ISO 10993-18.(The document states these tests were performed, implying they met the acceptance criteria, but does not explicitly state the specific results or numerical compliance for each test. In a 510(k) summary, positive results are generally implied for substantial equivalence.)
    Material EquivalenceThe device's material properties (Y-TZP and titanium grade 4) and dimensions should be comparable to predicate devices to establish substantial equivalence.The materials (Y-TZP and titanium grade 4) are consistent with materials used in the predicate and reference devices, or are well-established for this type of device. The technological characteristics table (Table 1) compares dimensions and materials, indicating fundamental similarity.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify exact sample sizes for the sterilization, endotoxin, packaging, or biocompatibility tests. It indicates that testing was performed on "production lot" or "completed package," which typically implies a statistically relevant sample size for validation, but the exact number isn't provided.
    • Data Provenance: Not specified. These are laboratory-based engineering and biological tests, not clinical studies involving human subjects or patient data. The manufacturer is Institut Straumann AG (based in Switzerland, submitted by Straumann USA, LLC), so the testing would likely have been conducted by them or a contract lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This is not applicable for this type of submission. The "ground truth" for engineering and biocompatibility tests is based on established scientific principles and international standards (e.g., ISO, FDA guidance), not on expert consensus from clinical cases.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are typically used in clinical studies when interpreting ambiguous patient data or images. The tests described are laboratory validations against objective criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a medical device for dental implants, not an AI/software device that involves human readers or image interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device (healing caps), not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the performance data (sterilization, endotoxin, biocompatibility, packaging) is based on pre-defined specifications and international standards. For example, a sterility assurance level of 10-6 is a defined standard, not an expert consensus or pathology result. Endotoxin limits are also defined by regulatory guidance.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of this 510(k) for a physical medical device. This term is relevant for machine learning or AI models.

    9. How the ground truth for the training set was established

    • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182313

    Validate with FDA (Live)

    Date Cleared
    2019-03-11

    (196 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BoneTrust® Dental Implants are medical devices intended to be surgically placed in the bone of the maxillary and/or mandibular arches to provide support for prosthetic restorations (crowns, bridges or overdenture) in edentulous or partially edentulous patients to restore a patients' chewing function.

    BoneTrust® implants can also be used for immediate loading when sufficient primary stability is achieved and with appropriate occlusal loading.

    BoneTrust® Short Dental Implants with length 6.5 mm are intended for delayed loading only

    BoneTrust® Abutments and Prosthetic parts are intended for use with Bone Trust Dental Implants in the maxillary and/or mandibular arches to provide support for crowns, bridges or overdentulous or partially edentulous patients.

    Device Description

    The BoneTrust® Implant System includes various sizes of threaded root-form dental implants and abutments intended to support prosthetic restorations in edentulous or partially edentulous patients.

    The BoneTrust® Implants are bone level, root form implants constructed of commercially pure titanium (Grade 4) per ISO 5832-2, with a sand-blasted, acid-etched surface treatment. BoneTrust® Implants are screw- shaped dental implants with a Hexagon or conical torx internal connection.

    BoneTrust® Dental Abutments are intended for cement-retained and screw-retained restorations. A cylindrical internal hexagon or conical torx allows connection to the BoneTrust® implant. BoneTrust® Dental Abutments are available in different designs

    AI/ML Overview

    This document is a 510(k) summary for the BoneTrust® Implant System. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a standalone study for novel performance metrics. Therefore, many of the requested sections about acceptance criteria, sample sizes, ground truth establishment, expert adjudication, and MRMC studies are not applicable.

    However, I can extract the information provided regarding non-clinical testing used to support substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria in terms of performance metrics (e.g., accuracy, sensitivity, specificity) for a novel AI device. Instead, it relies on demonstrating "substantial equivalence" through non-clinical testing to legally marketed predicate devices. The "reported device performance" is essentially that the device performed comparably to predicate devices in the described tests.

    Acceptance Criterion (Implicit)Reported Device Performance
    Biocompatibility in accordance with ISO 10993 seriesTests (including cytotoxicity per ISO 10993-5) support substantial equivalence to legally marketed predicate devices.
    Fatigue strength of the implant-abutment interfaceTested for representative samples in accordance with ISO 14801. Results support substantial equivalence to legally marketed predicate devices.
    Sterilization validation (for non-sterile devices)Validation conducted per ISO 17665-1 and ISO 17665-2 for pre-vacuum steam sterilization (wrapped cycle), demonstrating a sterility assurance level (SAL) of 10^-6.
    Sterilization validation (for sterile devices)Validation conducted per ISO 11137-1 and 11137-2 for Beta-radiation sterilization. Shelf life testing per ISO 11607-1 and ASTM F1980 (accelerated and real-time aging). LAL testing per USP 85 and FDA Guidance.
    Implant Surface AnalysisInvestigation using EDX, SEM, and BSE. Results support substantial equivalence to legally marketed predicate devices.
    Performance of BoneTrust® Short ImplantsComparative surface area analysis, comparative bone-to-implant contact study (using animal histology), and comparative pull-out strength test. Results support substantial equivalence to legally marketed predicate devices.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not provide specific sample sizes for the test sets in the non-clinical studies. It mentions "representative samples" for fatigue testing. Data provenance is not specified beyond being non-clinical testing performed to support substantial equivalence to legally marketed predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This is a 510(k) submission for a physical medical device (dental implants and abutments), not an AI/software device requiring ground truth establishment by human experts for diagnostic or similar tasks. The "ground truth" for the non-clinical tests would be the measurement results from the tests themselves (e.g., cytotoxicity levels, fatigue limits, physical characteristics).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as this is for a physical device and involves objective non-clinical tests rather than subjective human interpretation needing adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic or interpretation device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical studies performed, the "ground truth" would be objective measurements obtained through standardized testing methods, such as:

    • Biocompatibility: Laboratory results (e.g., cell viability in cytotoxicity tests).
    • Fatigue testing: Load cycles to failure or endurance limit under specified conditions, as per ISO 14801.
    • Sterilization: Sterility Assurance Level (SAL), microbial ingress testing, package integrity.
    • Implant Surface Analysis: Microscopic imagery (SEM, BSE) and elemental composition data (EDX).
    • Short Implants Performance: Quantitative measurements from comparative surface area analysis, histological examination for bone-to-implant contact (animal study), and force measurements for pull-out strength tests.

    8. The sample size for the training set

    Not applicable. There is no training set mentioned for this physical device.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180477

    Validate with FDA (Live)

    Date Cleared
    2019-01-04

    (316 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Straumann PURE Ceramic Implant is indicated for the restoration of single-tooth gaps and in edentulous or partially edentulous jaws. The prosthetic restorations used are single crowns, fixed partial or full dentures, which are connected to the implants through the corresponding components.

    Closure and Healing caps are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Closure and Healing caps should be used only with suitable implant connections. Do not use healing components for longer than 6 months.

    The provisional components are intended to serve as a base for temporary crown or bridge restoration out of occlusion for the Straumann® PURE Ceramic Implant System. The Straumann® Temporary Abutment VITA CAD-Temp® for the Straumann® PURE Ceramic Implant is indicated for temporary usage of up to 180 days.

    CI RD Straumann PUREbase abutment is a titanium base placed onto Straumann ceramic dental implants to provide support for customized prosthetic restorations and is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Straumann® Variobase Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description

    The Straumann PURE Ceramic Implant is a dental implant made out of yttrium-stabilized zirconium dioxide (Y-TZP). The Straumann PURE Ceramic Implant is based on features of the Straumann PURE Ceramic implant Monotype. Straumann PURE Ceramic Implants can be placed using the existing instruments using the same osteotomy preparation protocol as for Bone Level implants. The subject implant is a two -piece implant and is available in Ø4.1 mm with lengths of 8, 10, 12, and 14 mm.

    The Closure and Healing caps are screws machined as one piece and come in three gingival heights to accommodate individual gingival thickness. The material of the devices is titanium grade 4. The Closure caps are screwed into the implant to protect the inner configuration and shoulder of the implant during the healing phase in cases of submerged (submucosal) healing protocols and do not support a prosthetic restoration. Healing caps are screwed into the implant to protect the inner configuration in cases of transmucosal healing protocols and are placed out of occlusion and do not support a prosthetic restoration.

    The Temporary Abutments are used to serve as a temporary crown or bridge restoration for the Straumann PURE Ceramic Implant System.

    The CI RD Straumann PUREbase Abutment for Ceramic implant is a premanufactured (stock) abutment (the first piece of the two-piece abutment), sometimes referred to as "Tibase", and is used as a base when fabricating a CAD/CAM customized restoration (the second piece of the two-piece abutment). The assembly of the two-pieces becomes a finished medical device after cementation of the CAD/CAM customized restoration on the PURE base abutment.

    AI/ML Overview

    The Straumann PURE Ceramic Implant System did not present specific, quantified acceptance criteria with corresponding performance data in the provided text. Instead, the document focuses on demonstrating substantial equivalence to predicate and reference devices through various tests.

    However, based on the provided "Performance Data 5.7" section and the comparison tables, we can infer the types of acceptance criteria and the nature of the study that proves the device meets those criteria. The acceptance criteria generally revolve around demonstrating equivalence to legally marketed predicate and reference devices in terms of safety and effectiveness, as per FDA guidelines for 510(k) submissions.

    Here's an attempt to structure the information based on the request, interpreting "acceptance criteria" as the performance standards expected for demonstrating substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since explicit numerical acceptance criteria were not stated, this table describes the type of performance evaluated and the outcome in relation to the predicate/reference devices.

    Acceptance Criterion (Inferred from regulatory requirements and comparative studies)Reported Device Performance
    Biocompatibility: Device materials are safe for biological contact and do not cause adverse reactions. (In accordance with ISO 10993-1:2009).The subject device materials are identical to the predicate and reference device materials. Therefore, no new issues regarding biocompatibility were raised.
    Sterilization Efficacy (for sterile components): Sterilization process reliably achieves a Sterility Assurance Level (SAL) of 10-6. (Per ISO 11135).Sterilization validation was performed using the Half Cycle Overkill Approach, demonstrating the process can reliably and consistently sterilize to a minimum SAL of 10-6.
    Pyrogenicity: Device does not contain pyrogens above acceptable limits. (Based on FDA Guidance, testing limit of 20 EU/device for blood contacting/implanted devices).The device meets pyrogen limit specifications as determined by LAL Endotoxin Analysis, with a testing limit of 20 EU/device.
    Reprocessing Validation (for non-sterile components to be reprocessed by end-user): Reprocessing methods (e.g., autoclaving) are effective. (Per ISO 17665-1:2006 and FDA guidance).The sterilization process for the Straumann PURE temporary abutment and CI RD Straumann PUREbase abutments (as recommended in labeling) was validated according to applicable recommendations in FDA guidance and ISO 17665-1:2006.
    Mechanical Strength/Fatigue: Device can withstand dynamic loading forces typical of oral function without failure. (Per ISO 14801:2016 and FDA guidance).Dynamic fatigue tests demonstrated the Straumann PURE Ceramic Implant, Straumann PURE temporary abutments, and CI RD Straumann PUREbase abutments are equivalent to the predicate and reference devices.
    Wear on Implant-to-Abutment Connection: Connection surfaces exhibit comparable wear behavior to predicate/reference devices.Post-fatigue bench tests concluded comparable behavior of the subject device to the reference devices in terms of wear on the implant-to-abutment connection.
    Screw Loosening: Risk of screw loosening is comparable to predicate/reference devices.Post-fatigue bench tests concluded comparable behavior of the subject device to the reference devices in terms of screw loosening. A literature review was also provided to evaluate the risk of screw loosening with ceramic abutments.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes bench testing for mechanical performance, biocompatibility, and sterilization. It does not refer to a "test set" in the context of clinical data or data from human subjects.

    • Sample Size: Not explicitly stated for each test, but typically, mechanical and sterilization validations involve a statistically representative number of units or replicates to demonstrate consistency and meet statistical confidence levels defined in the relevant ISO standards (e.g., ISO 11135, ISO 14801). Biocompatibility is based on material composition equivalency, not a specific sample size.
    • Data Provenance: The studies are described as bench tests (laboratory studies), biocompatibility assessments based on material equivalency, and sterilization/reprocessing validations. There is no mention of country of origin of data in the context of test subjects, as these are not human studies for this specific assessment. The data is generated from laboratory testing and analytical comparisons.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is typically relevant for studies involving human interpretation (e.g., image analysis by radiologists). For this device, which involves material properties and mechanical performance, the "ground truth" is established by adherence to international standards (ISO) and FDA guidance documents, and the expertise lies in the validation engineers and materials scientists performing and interpreting these tests. There is no mention of "experts" in the sense of clinical adjudicators for a "test set" in the provided text.

    4. Adjudication Method for the Test Set

    Not applicable as this is not a study involving human interpretation of medical data (e.g., images) requiring adjudication of findings. Performance is evaluated against objective engineering and scientific standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an endosseous dental implant system, not an AI-powered diagnostic or assistive technology.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm or software. It is a physical medical device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluations (mechanical, sterilization, biocompatibility) is defined by the International Organization for Standardization (ISO) standards and FDA guidance documents to which these tests were conducted.

    • Biocompatibility: ISO 10993-1:2009.
    • Sterilization: ISO 11135 and FDA Guidance on "Submission and Review of Sterility Information...".
    • Reprocessing: ISO 17665-1:2006 and FDA guidance "Reprocessing Medical Devices in Health Care Settings...".
    • Mechanical Testing (Dynamic Fatigue): ISO 14801:2016 and FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
    • Wear and Screw Loosening: Bench tests demonstrating comparable behavior to reference devices, supported by a literature review for screw loosening risk.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This device is not an AI/machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173961

    Validate with FDA (Live)

    Date Cleared
    2018-06-05

    (159 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    ., K101545 Genesis Implant System ●, K153624 Neodent Implant System ., K170838 Medentika TiBases ., K130808

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® BLX Implants are suitable for endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. BLX Implants can be placed with immediate function on single-tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding abutment components.

    Straumann® Closure Caps and Healing Abutments are indicated to be placed in the patient's mouth at the end of the implant placement to protect the inner configuration of the implant and to shape, maintain and stabilize the soft tissue during the healing process. Closure caps and healing abutments should be used only with suitable implant connections. Straumann Closure Caps and Healing Abutments have a maximum duration of usage of 6 months.

    Prosthetic components directly or indirectly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. Temporary components can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion. Final abutments may be placed into occlusion when the implant is fully osseointegrated. BLX Temporary Abutments have a maximum duration of usage of 180 days.

    Straumann® Variobase® prosthetic components directly or indirectly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. The prosthetic restoration (crowns) can be cemented onto the Straumann® Variobase® prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they must be placed out of occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed copings and/or crowns for use with the Straumann® Variobase® Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description

    The Straumann BLX Implants are fully tapered implants manufactured utilizing the Roxolid material and are finished with SLActive® surface. The connection is identified as conical fitting with Torx style engaging feature. The prosthetic platforms are identified as RB (Regular Base) and WB (Wide Base). The implants with a RB platform have a "small top/head", and implants with WB platform have a "large top/head", whereas the internal connection is identical for both platform and all the implant diameters and lengths.

    The closure cap and healing abutments are manufactured from Titanium Grade 4, per ISO 5832-2 and ISO 5832-11, and are anodized signal violet for the parts compatible with the RB platform and brown for the parts compatible with the WB platform for identification purposes. Closure caps are screwed into the implant to protect the inner configuration and shoulder of the implant during the healing phase in cases of submerged (submucosal) healing. Healing abutments are screwed into the implant to protect the inner configuration in cases of transmucosal healing and are placed out of occlusion.

    The temporary abutments are manufactured from TAN and are anodized signal violet (RB platform) and brown (WB platform) for identification purposes. The temporary abutments are fixed in the implant with a basal screw which is also manufactured from TAN. The basal screw will be delivered with the temporary abutment.

    BLX Variobase abutments is a two-piece abutment ultimately composed by three components: Variobase™ Abutment (Ti-base), Prosthetic restoration (coping and/or crown), and Basal Screw. The BLX Variobase Abutments are manufactured and are delivered with the corresponding basal screw. The prosthetic restoration (crowns) can be cemented onto the Variobase prosthetic components.

    Basal screws are used to seat the temporary abutments and the BLX Variobase Abutments to the dental implant, and can be also be used during lab procedures to fix lab prosthetic parts on implant analogs. There is one basal screw for the RB platform and one for the WB platform. They have identical designs and differ only in color-coding (signal violet and brown) to ease the handling. They are provided along the abutments but they are also provided as standalone screws. The BLX basal screws are manufactured from TAN.

    AI/ML Overview

    This document describes the Straumann® BLX Implant System, a dental implant system, and its various components. The information provided is for a 510(k) premarket notification to the FDA. The goal is to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than setting and meeting specific, quantitative acceptance criteria for novel performance. Instead, the "acceptance criteria" are implied by successful comparative testing and adherence to existing standards and guidance documents. The "reported device performance" is the outcome of these tests.

    Test TypeImplied Acceptance Criteria (Demonstrated by Equivalence/Compliance)Reported Device Performance
    Sterilization ValidationAchieves a Sterility Assurance Level (SAL) of 10⁻⁶ via gamma irradiation. Complies with ISO 11137-1:2006 and ISO 11137-2:2013. For end-user sterilized components, validation according to ISO 17665-1, ISO 17665-2, and FDA guidance "Reprocessing Medical Devices in Health Care Settings."BLX implants, closure caps, and healing abutments are sterile via gamma irradiation (25 kGy minimum), validated to SAL 10⁻⁶ per ISO 11137-1:2006 and ISO 11137-2:2013 (overkill bioburden method). BLX Temporary Abutments and BLX Variobase Abutments are end-user sterilized by moist heat (steam), validated per ISO 17665-1, ISO 17665-2, and FDA reprocessing guidance.
    Shelf LifePackaging ensures maintaining sterility for a specified period and materials are not adversely affected.Shelf life for sterile devices remains 5 years, as packaging is equivalent to predicate/reference devices and materials are not adversely affected by time.
    PyrogenicityMeets pyrogen limit specifications based on LAL Endotoxin Analysis (e.g., < 20 EU/device).Pyrogenicity determined by LAL Endotoxin Analysis, with a testing limit of 20 EU/device for blood contacting and implanted devices.
    BiocompatibilityComplies with ISO 10993-1:2009 and FDA guidance "Use of International Standard ISO 10993-1." Equivalent nature of body contact, contact duration, material formulation, and sterilization methods compared to predicates.Biological assessment performed per ISO 10993-1:2009 and FDA guidance. Subject devices have equivalent body contact, duration, material formulation, and sterilization methods compared to predicates. Differences in manufacturing assessed via Chemical Characterization and Cytotoxicity tests.
    Dynamic Fatigue & Static StrengthFatigue load limits are equivalent to reference devices, especially for worst-case configurations.Tests performed according to FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments." Both RB and WB platforms, for worst-case considerations, proved equivalent to reference devices.
    Insertion TorqueAllows reaching suitable implant insertion torques.Tests conducted on worst-case defined for each diameter (Ø4.5, Ø 5.5, Ø 6.5 mm) of the BLX portfolio (shortest, middle, longest implants). Demonstrated that BLX Implants and related cutting instruments allow reaching suitable implant insertion torques.
    Pullout Strength, Surface Area, B-I ContactEquivalent to reference devices.Performed for 6mm long BLX implants with Ø 4.5, 5.5, and 6.5 mm. Equivalency to reference devices was demonstrated.
    Software Verification & ValidationComplies with FDA guidance documents for CAD/CAM systems and general software validation. Software functions correctly within the digital workflow.Conducted and documented according to FDA guidance documents "Class II Special Controls Guidance Document: Optical Impression Systems for Computer Assisted Design and Manufacturing (CAD/CAM) of Dental Restorations" and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff." Software considered "moderate" level of concern. Implementation verification shows Variobase devices and modeling components work together within the digital workflow for copings/crowns.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state numerical sample sizes for most bench tests (e.g., fatigue, insertion torque, pullout strength). It refers to "worst case considerations" for selecting test configurations (e.g., shortest, middle, and longest implants for insertion torque; both RB and WB platforms for fatigue). This implies a representative selection but not a specific count.

    The data provenance is not explicitly stated as "country of origin." However, as this is an FDA submission for a device from "Institut Straumann AG" and "Straumann USA, LLC" (Andover, Massachusetts), the testing would typically be conducted in facilities compliant with US/international standards relevant to medical device approval, likely either in the US or Europe where Straumann has its main operations. The tests are retrospective, as they are being submitted to demonstrate the device meets prior requirements for market clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this document. The document describes a medical device (dental implants and components) and its engineering and biological testing, not an AI/diagnostic software. Therefore, there is no "ground truth" to be established by experts in the context of diagnostic interpretation, and no test set in that specific sense.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reason as point 3. There is no diagnostic "test set" requiring adjudication by multiple readers or experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This document describes a physical medical device (dental implants), not an AI-assisted diagnostic or clinical decision support tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This document describes a physical medical device and some associated software for design/milling, not a standalone AI algorithm with performance metrics.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the engineering tests (fatigue, strength, torque, pullout), the "ground truth" is defined by established engineering and material science principles, industry standards (e.g., ISO, ASTM, FDA guidance), and the performance of legally marketed predicate devices. The "truth" is whether the device meets the performance requirements set forth by these standards and demonstrates equivalence to existing, safe, and effective devices. For biocompatibility, the ground truth is compliance with ISO 10993 standards and a lack of adverse biological reactions.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical dental implant system, not an AI model that requires a "training set." The software mentioned is for CAD/CAM and is validated, not "trained" in the machine learning sense.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171757

    Validate with FDA (Live)

    Date Cleared
    2017-11-27

    (166 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K141871, K151247, K150814, K130808

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations.

    Device Description

    Straumann® Screw Retained Abutments include one-piece straight and angled (17° and 30°) abutments, basal screws and abutment carrier pin. The subject device is identical to the primary predicate, K133421, and subsequent premarket notification submissions K141871, K150814 and K151247 with the exception that the subject device will be delivered to the user sterile as opposed to non-sterile and a change in packaging to ensure sterility of the devices.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Straumann® Screw Retained Abutments. It aims to demonstrate substantial equivalence to previously cleared predicate devices. Since this is not a diagnostic device and doesn't involve an AI algorithm, the information requested in your prompt (acceptance criteria, study details, expert input, MRMC study, standalone performance, training set) typically applies to AI/ML-based medical devices or diagnostic tools.

    This submission focuses on engineering, material, and sterilization equivalence for a dental implant component. Therefore, the detailed AI/ML-specific questions in your prompt are not applicable to this document.

    However, I can extract information related to the device's technical characteristics and the claim of substantial equivalence.

    Here's a breakdown of what is applicable from your prompt, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in the format typically used for performance metrics of diagnostic algorithms. Instead, it relies on demonstrating substantial equivalence to predicate devices. This means that the device's characteristics must be similar enough to the predicate to ensure equivalent safety and effectiveness.

    The table provided (pages 5-6) serves as the primary evidence for this. It compares the Subject Device (Straumann® Screw Retained Abutments) to several Predicate Devices across various technical characteristics. The "performance" in this context is demonstrating that these characteristics are either identical or sufficiently similar to the legally marketed predicates.

    FeatureSubject Device PerformancePredicate Device Performance (K133421, K141871, K150814, K151247)
    Indications for UseStraumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations. (Identical to predicates)Straumann® Magellan™ Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations. (For K133421, K141871, K150814, K151247)
    Implant-to-Abutment ConnectionNarrow CrossFit (NC), Regular CrossFit (RC) (Identical to K133421, K151247; broader than K141871, K150814 which only list NC)Narrow CrossFit (NC), Regular CrossFit (RC) (K133421, K151247)Narrow CrossFit (NC) (K141871, K150814)
    Interface TypeEngaging (Identical to all predicates)Engaging (All predicates)
    Platform Diameter(s)NC Straight and Angled: Ø3.5 mm, Ø4.6 mmRC Straight and Angled: Ø4.6 mm (Comparable ranges; Subject device appears to cover the full range of predicates)NC Straight: Ø3.5 mm, Ø4.6 mmRC Straight and Angled: Ø4.6 mm (K133421)NC Angled: Ø3.5 mm, Ø4.6 mm (K141871)NC Straight: Ø3.5 mm, Ø4.6 mm (K150814)NC Angled: Ø3.5 mm, Ø4.6 mmRC Angled: Ø4.6 mm (K151247)
    Abutment Angulation(s)0°, 17°, 30° (Identical to K133421, K151247; broader than K141871 (17°, 30°) and K150814 (0°))0°, 17°, 30° (K133421)17°, 30° (K141871, K151247)(K150814)
    Gingival Height(s)NC and RC Straight: 1.0, 2.5 and 4.0 mmNC and RC Angled: 2.5, 4.0 and 5.5 mm (Comparable and comprehensive ranges)NC and RC Straight: 1.0, 2.5 and 4.0 mmRC Angled: 2.5 and 4.0 mm (K133421)NC Angled: 2.5 and 4.0 mm (K141871)NC Straight: 1.0 mm (K150814)NC and RC Angled: 5.5 mm (K151247)
    Orientation of Angulation to Engagement FeaturesType A (45°), Type B (0°) (Identical to K133421, K141871, K151247; not applicable for K150814 as it's 0°)Type A (45°), Type B (0°) (K133421, K141871, K151247)N/A (K150814)
    Device MaterialTi-6Al-7Nb titanium alloy (Identical to all predicates)Ti-6Al-7Nb titanium alloy (All predicates)
    Sterilization StatusEnd user receives product sterilized per Gamma Irradiation, 25 kGy minimum, Validated per ISO 11137-1 and ISO 11137-2 to an SAL of 10 x 10-6 (Key difference: Subject device is sterile upon delivery)End user to sterilize product per IFU Method: Autoclave moist heat fractionated vacuum or gravity displacement Conditions: 134°C (273°F) for 5 minutes (All predicates are non-sterile; user is responsible for sterilization)
    PackagingChanged to ensure sterility. Equivalent to reference predicate device, Straumann® Healing Abutments, Healing Caps, Closure Screws (K130808) (Difference from non-sterile predicates, similarity to sterile reference predicate)Not explicitly detailed for non-sterile predicates, but implied to be non-sterile packaging. K130808 (sterile reference predicate) has equivalent packaging.

    The study proving the device meets its "acceptance criteria" (i.e., substantial equivalence) is the comparison to predicate devices and the verification/validation activities for the changes (primarily sterilization and packaging). The document explicitly states: "The subject device is identical to the primary predicate, K133421, and subsequent premarket notification submissions K141871, K150814 and K151247 with the exception that the subject device will be delivered to the user sterile as opposed to non-sterile and a change in packaging to ensure sterility of the devices." And "Packaging materials and configuration is equivalent to reference predicate device, Straumann® Healing Abutments, Healing Caps, Closure Screws, K130808."

    The substantial equivalence is primarily based on:

    • Identical Indications for Use.
    • Identical fundamental operating principles.
    • Identical materials (Ti-6Al-7Nb titanium alloy).
    • Comparable design features (Implant-to-Abutment Connection, Interface Type, Platform Diameter(s), Abutment Angulation(s), Gingival Height(s), Orientation of Angulation to Engagement Features).
    • Validation of the sterilization process according to recognized standards (ISO 11137-1 and ISO 11137-2) and equivalence of packaging to a previously cleared sterile device (K130808).

    The following points are not applicable to this 510(k) submission for a non-AI/ML dental abutment:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    8. The sample size for the training set
    9. How the ground truth for the training set was established

    These points would be relevant for devices that involve algorithms, image interpretation, or diagnostics where a "ground truth" needs to be established and performance metrics (e.g., sensitivity, specificity, AUC) are calculated based on a test set. This submission is for a physical medical device (dental abutment) where the primary evaluation is based on engineering design, material properties, and manufacturing processes, demonstrated through comparison to legally marketed predicate devices and validation of changes like sterilization.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163194

    Validate with FDA (Live)

    Date Cleared
    2017-07-14

    (241 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    12 - 150
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K150182, K153624, K093027, K151455, K130808, K061804

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use for GM implants and conventional abutments:

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Indications for Use for GM Titanium Base abutments:

    Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.

    All digitally designed copings and/or crowns for use with the Neodent Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.

    Indications for Use for GM Pro Peek Abutments:

    The Pro PEEK Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. They can be used in one or two stage procedures and also immediate load when there is good primary stability.

    Device Description

    The GM Line of the Neodent Implant System (GM Line) consists in an expansion of the Neodent Implant System previously cleared under K101945, K150182 and K153624, presenting a new design of Morse taper implant-to-abutment interface trade named GM line.

    The GM line comprises a range of endosseous dental implants as well as conventional and CAD/CAM abutments described as follows:

    The GM Line dental implants are threaded, self-tapping, root form with a Morse taper implantto-abutment interface with an internal hexagonal index exclusive to the GM line. They are made of titanium grade 4 conforming to ASTM F67. They are available in two types of surface treatment:

    NeoPoros: rough surface created using an abrasive particle jet concept with controlled grain oxides, followed by acid etching creating uniform cavities in the implant surface.

    Acqua: hydrophilic surface created from Neoporos rough surface that undergoes additional processing that renders hydrophilic surface.

    GM Titamax implants: cylindrical shape with double threads of pyramidal profile, rounded apex.

    GM Helix implants: conical shape, double threads with trapezoidal profile, a rounded apex.

    GM Drive implants: conical shape with double and progressive threads, cutting chambers in the counterclockwise direction, blade-shaped apical thread, rounded apex.

    The GM Line dental abutments are made of titanium alloy conforming to ASTM F136 and have a machined surface. The abutments are intended to be placed directly onto implants and are recommended according to the available interocclusal space, gingival height, and threedimensional position of the implant. They can be used in immediate or conventional rehabilitation procedures on the maxilla or mandible.

    GM Cover Screws: devices for temporary use during the implant healing phase.

    GM Healings: devices for temporary use for maintenance of the soft tissue during the osseointegration phase.

    GM Micro Abutments / GM Mini Conical Abutment / GM Exact Mini Conical Abutment: intermediary prosthetic components to be installed onto GM implants to support the final prosthesis.

    GM Exact Abutments: intermediary prosthetic components to be installed onto GM implants to support the final prosthesis.

    GM Exact Click Universal Abutments (straight and angled): intermediary prosthetic components to be installed onto GM implants to support the final prosthesis.

    GM Exact Titanium Base abutments: intermediary prosthetic components to be installed onto GM implants to support the final prosthesis; the Titanium Base two-piece abutments used as a base when fabricating a CAD/CAM customized restoration.

    GM Pro PEEK Abutment: temporary intermediary prosthetic components to be installed onto the Implants to support the provisional prosthesis up to 6 months.

    Protection Cylinders: prosthetic components with a tapered shape intended to be placed on abutments to protect them during the fabrication of the prosthesis and/or healing of the peri-implant tissues.

    Coping Screws: prosthetic devices intended for coping fixation on the corresponding GM (screw-retained) abutment.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "Neodent Implant System - GM Line." It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than an independent study proving the device meets specific acceptance criteria in a clinical or performance study context.

    Therefore, the information requested in your prompt (Acceptance Criteria, Study Design for proving performance, Sample Sizes, Expert consensus, MRMC studies, etc.) is not present in this document. This document details the comparison of the new device with existing ones based on:

    1. Indications for Use: The new device's indications align with or are subsumed by those of the predicates.
    2. Technological Characteristics: Materials, design features, sterilization methods, and compatibility are compared.
    3. Performance Data (Mechanical Testing and Biocompatibility): This section confirms that the materials meet established standards (ASTM, ISO) and that the device components (implants, abutments) have undergone mechanical fatigue testing according to ISO 14801 and FDA guidance. Biocompatibility has been assessed for the materials used. However, it does not provide specific acceptance criteria or detailed results of these tests, simply stating that the performance is "equivalent" to predicate devices.

    Summary of missing information based on your prompt:

    • Table of acceptance criteria and reported device performance: Not provided in a quantifiable manner for clinical performance. Only mentions "equivalent" mechanical performance based on ISO standards.
    • Sample sizes for test set and data provenance: No clinical test set involving human subjects is described. Mechanical tests likely involved specific numbers of devices, but these are not disclosed.
    • Number of experts and qualifications for ground truth: Not applicable as no clinical ground truth assessment is detailed.
    • Adjudication method for test set: Not applicable.
    • MRMC comparative effectiveness study: Not conducted or reported. This submission relies on substantial equivalence to existing devices, not a comparative clinical trial with human readers/outcomes.
    • Standalone performance (algorithm only): Not applicable as this is a physical medical device (dental implant system), not an AI algorithm.
    • Type of ground truth used: For regulatory purposes, the "ground truth" for this submission is established through comparison to legally marketed predicate devices and adherence to relevant industry standards (ISO, ASTM) and FDA guidance for mechanical and biocompatibility testing. No clinical "ground truth" derived from patient outcomes or expert consensus on images is mentioned regarding the device's function.
    • Sample size for training set: Not applicable as this is not an AI/ML device.
    • How ground truth for training set was established: Not applicable.

    In conclusion, this document demonstrates regulatory compliance and substantial equivalence through a comparison of device characteristics and a statement of meeting mechanical and biocompatibility standards, rather than presenting a detailed study proving the device meets specific, quantifiable clinical acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162890

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-08

    (142 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K120414 OsseoSpeed™ Plus, K130808 Straumann Healing Abutments, Healing Caps, and Closure Screws, K092814

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® Bone Level Tapered Implants Ø2.9 mm are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and aesthetic oral rehabilitation of patients with missing teeth. Straumann® Bone Level Tapered Implants 02,9 mm can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding components (abutments).

    The Straumann® Bone Level Tapered Implants 02.9 mm are indicated for reconstruction of missing incisors in the lower jaw and lateral incisors in the upper jaw.

    Straumann® Closure Caps and Healing Abutments are indicated to be placed in the dental implant after placement in the patient's jaw to protect the inner configuration of the implant and stabilize the soft tissue during the healing process. Closure Caps and Healing Abutments should be used only with the corresponding implant comection.

    Straumann® SC Temporary Abutments are indicated for use as an aid in prosthetic rehabilitations. Temporary components can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase. Strauman® SC Temporary Abutments have a maximum duration of usage of 180 days.

    Straumann® SC Variobase® abutments are indicated for use as an aid in prosthetic rehabilitations. The prosthetic restoration can be cemented on the Straumann® SC Variobase® prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the healing phase. Final abutments and restorations may be placed into occlusion when is fully osseointegrated. All digitally designed copings and/or crowns for use with the Straumann® Variobase® Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.

    Straumann® SC CARES® abutments are indicated for single-tooth replacements and multiple tooth restorations. The prosthetic restoration can be cemented or directly veneered/screw-retained.

    Device Description

    The Straumann Ø2.9mm Bone Level Tapered (BLT) Implants are apically tapered implants with an external diameter of Ø2.9 mm and lengths of 10 mm, 12 mm, and 14 mm. The implants are manufactured utilizing the Roxolid material and are finished with either the SLA® or SLActive® surface. The prosthetic platform is identified as SC (Small CrossFit®) which corresponds to a shoulder diameter of Ø2.9 mm.

    The closure cap and healing abutments are manufactured from Titanium Grade 4 and are anodized blue for identification purposes. The closure cap is conical and has a height of 0.5 mm. The healing abutments are seated in the implant with a basal screw which is manufactured from TAN. The healing abutments are oval in shape and are available in four different heights ranging between 2.0 mm and 6.5 mm.

    The temporary abutments are manufactured from TAN and are anodized blue for identification purposes. The temporary abutments are oval in shape in order to accommodate narrow interdental spaces and are available with three different gingival heights ranging between 1.0 mm and 3.0 mm. The temporary abutments are seated in the implant with a basal screw which is also manufactured from TAN.

    There are three components to the Straumann® SC Variobase™ Abutments:

    • Straumann® SC Variobase™ Abutments (Ti-base)
    • Prosthetic restoration (coping and/or crown)
    • Basal Screw

    The Straumann® SC Variobase® Abutments are manufactured from TAN. The abutments are oval in shape to accommodate narrow interdental spaces and are available with three different gingival heights ranging between 1.0 mm and 3.0 mm. The abutments will be delivered with the corresponding basal screw.

    The following is an overview of the possible prosthetic restoration (coping and/or crown) materials:

    • Cast materials:
      • Type 4 metals (ISO 22674)
      • Base metal alloys (e.g., cobalt-chromium (CoCr))
      • Noble metal alloys (e.g., gold alloy)
    • Press materials:
      • IPS e.max® Press Ceramic (K120053)
    • Digital materials:
      • coron®
      • zerion® LT
      • polycon® ae

    All digitally designed copings and/or crowns for use with the Straumann® Variobase® Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.

    The Straumann® SC CARES® Abutments are customized abutments manufactured from TAN. The abutments are designed by the customer by scanning the intraoral situation and designing of the shape with the CAD module of the integrated software. The design data is then sent to Straumann where the fabrication of the customized abutment is performed. The TAN alloy is capable of being directly veneered once the abutment is subject to a heat treatment step by the dental laboratory technician. A finished crown may also be cemented to the subject abutments rather than by direct veneer. The abutments will be delivered with the corresponding basal screw.

    AI/ML Overview

    The provided document (K162890) describes the regulatory submission for Straumann dental implants and associated components. However, it does not contain information about studies proving the device meets acceptance criteria for performance metrics related to diagnostic accuracy, such as sensitivity, specificity, or AUC, as these types of metrics are typically associated with diagnostic or AI-powered devices.

    Instead, the document focuses on demonstrating substantial equivalence to existing legally marketed predicate devices through various non-clinical tests. The "acceptance criteria" here relate to the successful completion of these engineering and safety tests, rather than clinical performance metrics for diagnostic capability.

    Therefore, the requested information cannot be fully provided in the format desired for diagnostic devices because this submission deals with dental implants, not a diagnostic AI.

    However, I can extract the information related to the performance testing that was conducted to support the substantial equivalence claim.


    Acceptance Criteria and Device Performance (Not applicable in the typical sense for diagnostic AI):

    The document does not present acceptance criteria or reported device performance in terms of diagnostic accuracy metrics (e.g., sensitivity, specificity, AUC) because the devices (dental implants and abutments) are not diagnostic AI devices. Instead, the "performance testing" focuses on engineering, biocompatibility, and sterilization validations to demonstrate substantial equivalence to predicate devices.

    Summary of Performance Testing Conducted (which can be seen as meeting "acceptance criteria" for safety and efficacy in this context):

    1. Sterilization Validation (Section 5.8.1):

      • Acceptance Criteria: Validation according to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015".
      • Reported Device Performance: The sterilization process for SC Temporary Abutments, SC Variobase Abutments, and SC CARES Abutments was validated.
    2. Biocompatibility Testing (Section 5.8.2):

      • Acceptance Criteria: Compliance with ISO 10993-1:2009 and the FDA Guidance document "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016". The subject devices should have identical nature of body contact, contact duration, material formulation, manufacturing processes, and sterilization methods compared to predicate devices, raising no new biocompatibility issues.
      • Reported Device Performance: The subject devices have identical biocompatibility characteristics to the primary and reference predicate devices. Therefore, no additional biocompatibility testing was required or performed.
    3. Software Verification and Validation Testing (Section 5.8.3): (Applicable to CARES Visual software for abutment design, which is part of the manufacturing process, not a diagnostic AI)

      • Acceptance Criteria: Compliance with FDA guidance documents "Class II Special Controls Guidance Document: Optical Impression Systems for Computer Assisted Design and Manufacturing (CAD/CAM) of Dental Restorations" and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff".
      • Reported Device Performance: Software verification and validation testing were conducted; the software for the device (CARES Visual) was considered a "moderate" level of concern.
    4. Bench Testing (Section 5.8.4):

      • Acceptance Criteria: Compliance with the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" for dynamic fatigue, static strength, and insertion torque tests.
      • Reported Device Performance: Dynamic fatigue, static strength, and insertion torque tests demonstrated the Straumann® Ø2.9 mm Bone Level Tapered implants, SC Temporary Abutments, SC Variobase Abutments, and SC CARES abutments are equivalent to the predicate and reference devices.

    Regarding the specific questions about diagnostic AI studies:

    1. A table of acceptance criteria and the reported device performance: As explained above, not applicable for diagnostic metrics. The performance relates to engineering and safety tests.
    2. Sample size used for the test set and the data provenance: Not applicable. The "tests" mentioned are non-clinical engineering and bench tests, not studies on patient data for diagnostic evaluation.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as there is no diagnostic test set or ground truth in this context.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted diagnostic device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to a diagnostic AI scenario. The "ground truth" for the bench tests would be the established engineering standards and physical properties of the materials.
    8. The sample size for the training set: Not applicable, as there is no AI training set.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2