K Number
K162848

Validate with FDA (Live)

Date Cleared
2017-09-22

(346 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures.

The Straumann CARES Golden Ti/TiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained.

Device Description

The previously cleared Straumann CARES Titanium Abutments are placed onto dental implants to support prosthetic reconstructions such as crowns and bridges. The abutment allows for individual customization regarding function and esthetics. The design step is typically performed by the dental laboratory to the specifications from a clinician. The abutments are produced by a Straumann Manufacturing milling center.

The patient-specific abutment is designed by a traditional waxup and subsequent scanning or by scanning of the intraoral setting and designing the element via CAD software. The design information is sent via an Internet portal to Straumann. Straumann verifies the design against the validated design parameters. These design limits include a maximum angulation of 30° and a minimum surface area as follows:

Tooth Position, Minimum surface area mm²
7,10, 24,25, 37
4,5,12,13,20,21,28,29, 47
26,11,2,27,8,9, 43
1-3, 14-16, 17-19,30-32, 56

The abutments are manufactured from solid grade 4 titanium at the Straumann milling center. The abutment is delivered to the dental laboratory for final processing before delivery to the clinician. The abutments are attached to the implant with a titanium alloy basal screw. The occlusal restoration is affixed to the titanium abutment using dental cement.

The proposed devices are a modification of the previously licensed devices. The titanium abutments that serve as the base of the CARES Golden Ti/TiN Abutments (the uncoated abutments) are identical in every respect to the reference devices. The modification consists of the addition of a titanium nitride (TiN) coating on the emergence profile and coronal aspect of the abutment. The apical portion of the abutment consisting of the implant-to-abutment interface is masked during the coating process to assure that the interface remains uncoated.

AI/ML Overview

The provided text describes the Straumann® CARES® Golden Ti/TiN Abutments, a dental device, and its substantial equivalence to predicate devices. It outlines the design, materials, and performance data used to support its clearance.

However, the document is a 510(k) summary for a medical device and therefore, it does not contain information about an AI/ML powered device, a study proving the device meets specific acceptance criteria related to AI/ML performance, or details about ground truth, expert consensus, or multi-reader multi-case studies.

Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance related to an AI/ML device, nor can I provide information on sample sizes for test/training sets, data provenance, expert numbers/qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment relevant to AI/ML.

The document focuses on demonstrating substantial equivalence based on material, design, indications for use, and a series of standard performance tests for dental abutments.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or a bird in flight, composed of three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 22, 2017

Institut Straumann AG % Jennifer Jackson Director of Regulatory Affairs and Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810

Re: K162848

Trade/Device Name: Straumann® CARES® Golden Ti/TiN Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: August 25, 2017 Received: August 28, 2017

Dear Jennifer Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mary S. Runner -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K162848

Device Name

Straumann® CARES® Golden Ti/TiN Abutments

Indications for Use (Describe)

Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures.

The Straumann CARES Golden TiTiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.510(k) SummaryK162848
Submitter:Straumann USA, LLC (on behalf of Institut Straumann AG)60 Minuteman RoadAndover, MA 01810
Contact Person:Jennifer M. Jackson, MSDirector, Regulatory Affairs and Quality+1 (978) 747-2509
Prepared By &Secondary Contact:Christopher KlaczykHead of Regulatory Affairs & ComplianceInstitut Straumann AG+41 61 965 1260
Date Prepared:September 22, 2017
Product Code(s):NHA (21 CFR 872.3630)
Device Class:II (21 CFR 872.3630)
Classification Panel:Dental
Classification Name:Endosseous dental implant abutment (21 CFR 872.3630)
Proprietary Name:Straumann® CARES® Golden Ti/TiN Abutment
Primary PredicateDevice:K082764, Straumann CARES WN Ti Abutments (Institut Straumann)
Reference Device(s):K052272, Straumann CARES RN Ti Abutments (Institut Straumann)K072151, Straumann CARES RC Ti Abutments (Institut Straumann)K081005, Straumann CARES NC Ti Abutments (Institut Straumann)K992334, 3i TiN Coated Implants and Abutments (Implants Innovations, Incorporated (3i))
Device Description:The previously cleared Straumann CARES TitaniumAbutments are placed onto dental implants to support prostheticreconstructions such as crowns and bridges. The abutmentallows for individual customization regarding function andesthetics. The design step is typically performed by the dentallaboratory to the specifications from a clinician. The abutmentsare produced by a Straumann Manufacturing milling center.

{4}------------------------------------------------

The patient-specific abutment is designed by a traditional waxup and subsequent scanning or by scanning of the intraoral setting and designing the element via CAD software. The design information is sent via an Internet portal to Straumann. Straumann verifies the design against the validated design parameters. These design limits include a maximum angulation of 30° and a minimum surface area as follows:

Tooth PositionMinimum surface area mm²
7,10, 24,2537
4,5,12,13,20,21,28,2947
26,11,2,27,8,943
1-3, 14-16, 17-19,30-3256

The abutments are manufactured from solid grade 4 titanium at the Straumann milling center. The abutment is delivered to the dental laboratory for final processing before delivery to the clinician. The abutments are attached to the implant with a titanium alloy basal screw. The occlusal restoration is affixed to the titanium abutment using dental cement.

The proposed devices are a modification of the previously licensed devices. The titanium abutments that serve as the base of the CARES Golden Ti/TiN Abutments (the uncoated abutments) are identical in every respect to the reference devices. The modification consists of the addition of a titanium nitride (TiN) coating on the emergence profile and coronal aspect of the abutment. The apical portion of the abutment consisting of the implant-to-abutment interface is masked during the coating process to assure that the interface remains uncoated.

Indications For Use: Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures.

The Straumann CARES Golden Ti/TiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained.

  • Materials: The substrate material for all Straumann CARES Golden Ti/TiN Abutments is commercially pure Titanium (grade 4) conforming to ISO 5832-2. The surface of the emergence profile and coronal aspect of the abutments are coated with titanium nitride (TiN). The implant-to-abutment interface is masked and remains uncoated.

{5}------------------------------------------------

Technological Characteristics:A comparison of the relevant technological characteristics between the subject and primary predicate devices is provided in the table that follows.
Performance Data:Test data to support the evaluation of the subject Straumann CARES Golden Ti/TiN abutments is as follows:Sterilization validation in accordance with ISO 17665-1 and ISO/TS 17665-2 Biocompatibility assessment per ISO 10993-1 Cytotoxicity testing per ISO 10993-5 Chemical characterization per ISO 10993-18 Dynamic fatigue testing per ISO 14801 by reference to K052272, K072151, K081005 and K082764 Coating characterization including chemical composition, thickness, porosity, mean volume percent voids, SEM images at 100X Adhesion testing by static shear bonding strength per ASTM F1044 employing minimum shear stress acceptance criteria per ASTM F2068 Adhesion testing by dynamic shear bonding strength per ASTM F1160 Abrasion testing consistent with the method described by Tamura, et al, entitled Mechanical Properties of Surface Nitrided Titanium for Abrasion Resistant Implant Materials . No animal or human clinical studies were conducted.
Conclusions:Based upon our assessment of the design and applicable performance data, the subject devices have been determined to be substantially equivalent to the identified predicate devices.

{6}------------------------------------------------

FeaturePrimary Predicate DevicesCARES Titanium Abutments(K082764)Subject DevicesCARES Titanium Gold AbutmentsEquivalenceDiscussion
Indications For UseAbutments are placed into the dentalimplants to provide support for prostheticrestoration such as crowns, bridges andoverdentures.The Straumann Titanium Abutment isindicated for single tooth replacementsand multiple tooth restorations. Theprosthetic restoration is cement-retained.Abutments are placed into the dentalimplants to provide support for prostheticrestoration such as crowns, bridges oroverdentures.The Straumann CARES Golden Ti/TiNAbutment is indicated for single toothreplacements and multiple toothrestorations. The prosthetic restoration iscement-retained.Identical
Compatible ImplantsStraumann Bone Level implants havingthe NC and RC implant-to-abutmentinterface geometries.Straumann Tissue Level implants havingthe NN, RN and WN implant-to-abutment interface geometries.Straumann Bone Level implants havingthe NC and RC implant-to-abutmentinterface geometries.Straumann Tissue Level implants havingthe NN, RN and WN implant-to-abutment interface geometries.Identical
MaterialCommercially pure grade 4 Titaniumper ISO 5832-2Commercially pure grade 4 Titaniumper ISO 5832-2Identical
Surface FinishUncoated (raw titanium)Titanium nitride (TiN) coating onemergence profile and coronal surfacesEquivalentThe TiN coating is being added to impartgold color to the emergence profile andcoronal surfaces of the abutment in orderto improve the esthetics of the finishedrestoration. The presence of the coatingdoes not change the intended uses orindications of the devices. This use ofTiN coating was originally cleared tomarket per K992334.
ConstructionOne-piece solid devices.One-piece solid devices.Identical
FeaturePrimary Predicate DevicesCARES Titanium Abutments(K082764)Subject DevicesCARES Titanium Gold AbutmentsEquivalenceDiscussion
Design Limits:Maximum Angulation: 30°Minimum Surface Area:Pos, 7,10, 24,25 37 mm²Pos, 4,5,12,1320,21,28,29 47 mm²Pos, 26,11,2,27,8,9 43 mm²Pos, 1-3, 14-16, 17-19,30-32 56 mm²Maximum Angulation: 30°Minimum Surface Area:Pos, 7,10, 24,25 37 mm²Pos, 4,5,12,1320,21,28,29 47 mm²Pos, 26,11,2,27,8,9 43 mm²Pos, 1-3, 14-16, 17-19,30-32 56 mm²Identical
PackagingThe abutment and basal screw inreclosable poly bags, co-packaged in akraft board box.The abutment and basal screw inkraft board box.Identical
SterilityProvided non-sterile.Terminally sterilized via moist heat(autoclave) to a Sterility Assurance Levelof 10-6. Validated per ISO 17665-1.Provided non-sterile.Terminally sterilized via moist heat(autoclave) to a Sterility Assurance Levelof 10-6. Validated per ISO 17665-1.Identical

{7}------------------------------------------------

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)