Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K210336
    Date Cleared
    2021-08-13

    (189 days)

    Product Code
    Regulation Number
    872.3640
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K151328, K101945, K182620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.

    The CR Abutment is an abutment placed over Neodent Zirconia Implants in order to provide support for prosthetic restorations, such as copings or crowns. It may be used for single-unit restorations that are cement-retained in esthetical areas over implants installed in maxilla or mandible.

    Device Description
    • Intended for single use;
    • Provided sterile via Ethylene Oxide;
    • . All devices of Zirconia Implant System are manufactured of Y-TZP (Yttrium-stabilized zirconium dioxide);
    • . The CR Abutment for Zirconia system is supplied along with a titanium alloy screw (ASTM F136);
    • ZiLock prosthetic interface with internal indexer;
    • Apically tapered implant with trapezoidal thread profile;
    • Implant provided with sand-blasted, acid etched surface finish;
    • . The subject Implants are compatible with the Zirconia Bases (already cleared per market per K201491) and CR Abutments (subject of this submission);
    • . The subject CR Abutments are compatible with the Zirconia Implants 4.3 mm (already cleared per market per K201491) and Zirconia Implants 3.75 mm (subject of this submission).
    AI/ML Overview

    This document describes the Neodent Implant System - Zirconia Implant System (K210336), which includes Zirconia Implants and CR Abutments. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly refer to "acceptance criteria" in the traditional sense of numerical thresholds for a study endpoint. Instead, the performance data presented aims to demonstrate equivalence to predicate devices through various tests.

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityCompliance with ISO 10993-1 and ISO 10993-18. Not representing a new worst-case compared to K201491.Biological Safety Assessment guided by ISO 10993-1 performed. Chemical characterization per ISO 10993-18 performed. Subject devices do not represent a new worst-case, relying on results from K201491.
    Dynamic Fatigue TestEquivalent fatigue strength to predicate devices per ISO 14801 and FDA Guidance.Performed according to ISO 14801. Devices exhibit a level of performance equivalent to that reviewed for the predicate devices. SEM images after fatigue show similar wear to K201491.
    Torsion TestSufficient resistance to static torsional loading.Performed to evaluate the Zirconia Implant System under static torsional loading. (Specific quantitative results not provided in this summary).
    Insertion TestAcceptable insertion torque into different bone types.Performed to evaluate insertion torque into sawbones material representing bone types I, II, III, and IV. (Specific quantitative results not provided in this summary).
    Screw Torque TestingNo visible failure after torque application and removal.Performed with torque application and visual analysis. No visible failure identified after applying torque and removing items. Subject screws considered equivalent to those cleared per K201491.
    Sterilization ValidationSterility Assurance Level (SAL) of 1x10^-6 via Ethylene Oxide (ISO 11135-1). Not representing a new worst-case compared to K201491.Sterilized by Ethylene Oxide per ISO 11135-1 (over-kill method). Achieved an SAL of 1x10^-6. Subject devices do not represent a new worst-case, adopting the method from K201491.
    Bacterial Endotoxin TestingResults
    Ask a Question

    Ask a specific question about this device

    K Number
    K210220
    Date Cleared
    2021-06-18

    (142 days)

    Product Code
    Regulation Number
    872.3640
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kontact™ Dental Implant System is indicated for use in partially or fully edentulous patients to support maxillary or mandibular single unit, multiple-unit, or overdenture dental restorations. Kontact Dental Implant System is indicated for immediate loading when good primary stability is achieved and the occlusal loading is appropriate. Kontact 3 mm diameter implants and prosthetics components are indicated for use in surgical and restorative applications in the maxillary lateral incisor or mandibular incisor regions.

    Device Description

    The Kontact™ Dental Implant System comprises endosseous root-form dental implants, abutments, abutment screws, and other associated components for single-unit, multi-unit, and overdenture restorations. Kontact dental implants are provided in five body diameters: 3.0 mm, 3.6 mm, 4.2 mm, 4.8 mm, and 5.4 mm. The platform diameter for each size is the same as the (maximum) body diameter. The implants are provided in lengths ranging from 8 mm to 16 mm. The implants have a recessed internal section for abutment indexing, and an internal threaded section for mating to the corresponding subject device cover screw, or abutment screw. All implants are manufactured from titanium alloy conforming to ASTM F136 and ISO 5832-3. The endosseous threaded surface of the subject device implants is grit-blasted with non-resorbable aluminum oxide (Al2O3) particles. Kontact conventional and prosthetic components include cover screws, abutment screws, temporary abutments, straight abutments, angled abutments, prepable abutments, multi-unit abutments, healing caps, and overdenture abutments. Most device screws and all subject device abutments are manufactured from titanium alloy conforming to ASTM F136 and ISO 5832-3. Select subject device abutments and screws manufactured from titanium alloy are anodized using standard electrolytic passivation processing to impart a distinctive surface color. The subject device abutment screw intended for 3 mm diameter implants and the prosthesis screw intended for conical abutments are manufactured from cobalt alloy, conforming to ASTM F1058 and ISO 5832-7. The abutment screw is coated with titanium nitride (TiN); the prosthesis screw is coated with chromium nitride (CrN). Subject device healing caps intended for use with UniPost abutments, and IsoPost abutments are manufactured from polyetheretherketone (PEEK). Subject device healing caps intended for NanoPost abutments and IsoPost abutments are manufactured from polymethyl methacrylate (PMMA).

    AI/ML Overview

    The product, Kontact™ Dental Implant System, has not reported any clinical studies. The only data provided is non-clinical performance data.

    Here's the breakdown of the information requested, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/Evaluation TypeAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityConformance to ISO 10993-5 and ISO 10993-12Device demonstrates biocompatibility according to ISO 10993-5 (tests for in vitro cytotoxicity) and ISO 10993-12 (sample preparation and reference materials). Specific results are not detailed but compliance with these standards indicates meeting requirements for biological safety.
    Gamma SterilizationConformance to ISO 11137-1 and ISO 11137-2Validation performed for subject device implants and cover screws. Specific Sterility Assurance Levels (SAL) are not stated, but conformance to these standards implies achieving a validated sterile state (e.g., SAL of 10^-6).
    Bacterial Endotoxin Testing (BET)≤ 20 EU/deviceLimulus amebocyte lysate (LAL) test performed according to ANSI/AAMI ST72 and USP 43-NF38:2020 . Ongoing monitoring and control ensure BET levels meet or are below ≤ 20 EU/device. This indicates the devices are free from pyrogenic contamination within accepted limits.
    Shelf Life ValidationConformance to ASTM F1980 and ISO 11607-1Packaging stability validated. Conformance to ASTM F1980 (standard guide for accelerated aging of sterile medical device packages) and ISO 11607-1 (packaging for terminally sterilized medical devices) indicates the device maintains its integrity and sterility over its claimed shelf life. Specific shelf life duration is not provided.
    Moist Heat Sterilization (End-user)Conformance to ISO 17665-1 and ISO TS 17665-2Instructions and validation for end-user moist heat sterilization provided. Conformance to ISO 17665-1 (sterilization of health care products - moist heat - requirements for the development, validation and routine control of a sterilization process for medical devices) and ISO TS 17665-2 (guidance on the application of ISO 17665-1) ensures proper sterilization by the end-user. Specific parameters are not detailed.
    Static Compression & Compression Fatigue TestingConformance to ISO 14801Mechanical performance testing performed on the subject device according to ISO 14801 (dentistry - implants - dynamic fatigue test for endosseous dental implants). The fatigue limit data for abutments compatible with 3.0 mm diameter implants was supported by reference device K183518, and "constructs of all other subject device abutments in combination with all other subject device implants have sufficient strength for their intended use." This indicates the devices meet mechanical strength requirements for their intended application. The specific criteria for "sufficient strength" are not explicitly quantified (e.g., number of cycles survived, load at failure) but implied by conformance to the standard.
    Endosseous Surface CharacterizationDemonstrated by SEM and EDSThe grit-blasted surface (with non-resorbable aluminum oxide) was validated by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) characterization. This confirms the presence and composition of the surface modification intended to facilitate osseointegration.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for any of the non-clinical tests.
    • Data Provenance: All data is non-clinical and derived from laboratory testing of the device and its components, primarily referencing international and industry standards (ISO, ASTM, ANSI/AAMI, USP). Information on the country of origin of the test data is not provided, but the manufacturer is Biotech Dental, SAS, based in France. The data is retrospective in the sense that it's generated as part of a premarket submission, not from a prospective clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. This device relies on pre-defined engineering and sterility standards, not expert-established ground truth from clinical cases. There were no clinical studies.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    • Not applicable. There was no test set for clinical performance requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • Not applicable. No clinical studies were conducted, and this product does not involve AI or human reader interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This product is a physical dental implant system, not a software algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. Ground truth, in the clinical sense, was not used as there were no clinical studies. The "ground truth" for non-clinical testing is adherence to established international and industrial standards (e.g., ISO, ASTM) for material properties, mechanical performance, and sterility.

    8. The Sample Size for the Training Set

    • Not applicable. There was no training set, as this is a physical medical device and not an AI/ML algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. There was no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192053
    Manufacturer
    Date Cleared
    2019-10-29

    (90 days)

    Product Code
    Regulation Number
    872.3640
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K151328, K180477

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TAV Medical's W Zirconia Implants are intended for surgical placement in the patient's upper and lower jaw to provide support for prosthetic devices, such as artificial teeth and in order to restore the patient chewing function. The implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. The ø3.6mm reduced diameter implants are recommended for central and lateral incisors only.

    Device Description

    TAV Medical's W Zirconia Implants are dental implants, composed of the following implant models: W One Piece - monotype implant with integrated abutment. W Two Piece – implant for screw retained Abutment. The implants are tissue level designed and includes a body portion and a neck (1.8mm). The implant body portion is configured to extend into the bone and osseo-integrate with the alveolar bone. The neck should be positioned 1.8mm above the bone level. TAV Medical's W Zirconia Implants are made of Yttria stabilized tetragonal zirconia (Y-TZP). This material conforms with ISO 13356:2015 standard for Implants for surgery - Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP). The Titanium abutments are going through anodizing process using an electrolytic process that adjusts the oxide level of the metal surface. This adjustment changes the spectrum of light, resulting in perceived color. By controlling the surface oxide level, an entire range of colors can be achieved. The Subject Device includes the following dimensions: W One-Piece implants: Diameter of 3.6mm for lengths of 8mm, 10mm, 12mm and 14mm and; Diameters of 4.1mm and 4.8mm for length of 14mm. W Two-Piece implants: Diameters of 4.1mm and 4.8mm for lengths of 8mm, 10mm, 12mm and 14mm. Cover Screw are screwed into the implant to protect the inner configuration of the implant during the healing phase in cases of submucosal healing protocol. Cover screws are made of Titanium alloy Ti 6Al 4V ELI & Anodize and are available in diameters of 4.1mm and 4.8mm. Titanium Healing Caps are intended to protect the 2-piece implants during the healing phase. The healing caps also support the emergence profile and keep the implant shoulder ideal for the impression phase. The healing caps are available in different geometrical features such as height and diameter. The Titanium Healing Caps manufactured from Ti 6Al 4V ELI & Anodize and are available in diameters of 4.1mm and 4.8mm. Each diameter is provided with height dimensions of 1mm, 2mm, 3mm and 4mm. PEEK Healing Caps are designed to protect the two-piece implant during the healing phase. The healing caps also support the emergence profile and keep the implant shoulder ideal for the impression phase. Peek healing caps are available in different geometrical features such as height and diameter and available in 2 different designs, one design for the one-piece implant and second design for the two-piece implant. Both designs are manufactured from PEEK material and are indicated to be placed in patients' mouth for a maximum duration of 180 days. The PEEK healing caps are available in diameters of 4.1mm and 4.8mm and height dimensions of 1mm, 2mm, 3mm and 4mm. PEEK Temporary Caps serves as a basis for temporary restoration. TAV Medical temporary restoration caps available in two configurations, temporary Cap for Crown and Temporary Cap for Bridge. The caps are manufactured from PEEK material and are indicated to be placed in patients' mouth for a maximum duration of 180 days. The PEEK temporary caps are available in diameter of 3.6mm. Titanium Abutments for the W Two Piece Zirconia Implants are intended for use as an adapter between the implant and the crown. The abutments are characterized by distinct geometrical features such as length and angulation. The titanium abutments are manufactured from ASTM F136-13 compatible Titanium Ti 6Al 4V ELI & Anodize. TAV Medical abutments are intended to be connected to 2-piece W Zirconia implants with titanium screw, to provide support for prosthetic reconstructions. The abutments are available in the following dimensions: Length of 5.0mm and 6.0mm and Angles of 0°, 5°, 10° and 15°. The 0º serves also for bridges.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the TAV Medical Ltd. W Zirconia Implants. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a standalone study with a defined test set and ground truth in the way one might evaluate a diagnostic AI system.

    However, I can extract the information related to performance testing that supports the claim of substantial equivalence.

    Here's an analysis of the "acceptance criteria" (which are more akin to performance benchmarks against predicate devices and regulatory standards in this context) and the study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission, the "acceptance criteria" are implicitly meeting the performance of predicate devices and adherence to relevant ISO standards and FDA guidance documents. The document doesn't explicitly list numerical acceptance criteria with pass/fail thresholds in the same way a diagnostic AI study would. Instead, it states that the subject device's performance was "substantially equivalent" to the predicate devices and conformed to standards.

    Acceptance Criteria (Implied)Reported Device Performance
    Material ConformanceMaterial (Yttria stabilized tetragonal zirconia, Y-TZP) conforms with ISO 13356:2015.
    BiocompatibilityEstablished through identical manufacturing methods, facility, raw materials as the primary predicate device (K172668),
    and compliance with ISO 10993-1, 10993-5 and FDA Guidance for Use of ISO 10993-1. No additional tests were conducted, as it was concluded biocompatibility was already established.
    Sterilization (Implants)Validation leveraged from primary predicate K172668, conducted in accordance with ANSI/AAMI/ISO 11137-2:2013. Achieved SAL of 10^-6.
    Sterilization (Abutments)Steam heat sterilization validation performed in compliance with FDA Guidance (Jan 21, 2016) and ANSI/AAMI/ISO 17665-1:2006. Achieved SAL of 10^-6.
    Shelf LifeAccelerated aging applied to final packaging, followed by real-time aging validating implants packaging. (No specific duration or results reported in this summary, but implies positive validation).
    Mechanical Performance (Fatigue Testing)Performed according to FDA guidance (May 2004) and ISO 14801:2016. Results indicated substantial equivalence to the primary predicate device.
    Mechanical Performance (Implant-to-Abutment Connection Wear)Bench tests conducted after fatigue testing to assess wear of implant body, abutment, and fixation screw, and screw loosening. Data concluded 'comparable behavior' to reference devices.
    Overall EquivalenceSame intended use, indications for use, mode of operation, materials, manufacturing technology, and body contact as predicate devices, demonstrating substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a medical device (dental implants), not an AI algorithm. Therefore, the concept of a "test set" for an AI model (like images with ground truth labels) does not directly apply.

    • Mechanical Testing: While not explicitly stated as "sample size," mechanical tests (like fatigue testing and wear assessment) would involve a certain number of physical implant samples. This information is typically detailed in the full test reports, which are not included in this 510(k) summary. The summary only states that tests were performed according to ISO standards, which define methodologies including sample numbers.
    • Data Provenance: Not applicable in the context of an AI test set. The provenance of the device components (materials, manufacturing) is Israel.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable as this is a physical medical device, not a diagnostic AI system requiring expert-established ground truth for a test set. Evaluation relies on established engineering principles, material science, and bio-compatibility standards through laboratory testing.

    4. Adjudication Method for the Test Set

    Not applicable for a physical medical device. Decisions are based on objective measurements against engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is not an AI diagnostic device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an AI algorithm.

    7. The Type of Ground Truth Used

    For a physical medical device, "ground truth" refers to established scientific and engineering principles, validated test methodologies, and regulatory standards.

    • Material Properties: Conformance to ISO 13356:2015.
    • Biocompatibility: Demonstrated through adherence to ISO 10993-1, 10993-5, and FDA guidance.
    • Sterilization: Demonstrated through adherence to ANSI/AAMI/ISO 11137-2:2013 and ANSI/AAMI/ISO 17665-1:2006, and FDA guidance.
    • Mechanical Performance: Demonstrated through adherence to FDA guidance (May 2004) and ISO 14801:2016. Testing involved objective measurements of fatigue life, wear, and screw loosening.

    The underlying "ground truth" is that the device must meet the performance and safety requirements outlined in these standards and guidance documents, demonstrating substantial equivalence to legally marketed predicate devices.

    8. The Sample Size for the Training Set

    This is not applicable as this is a physical medical device, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable as this is a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180477
    Date Cleared
    2019-01-04

    (316 days)

    Product Code
    Regulation Number
    872.3640
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Straumann PURE Ceramic Implant is indicated for the restoration of single-tooth gaps and in edentulous or partially edentulous jaws. The prosthetic restorations used are single crowns, fixed partial or full dentures, which are connected to the implants through the corresponding components.

    Closure and Healing caps are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Closure and Healing caps should be used only with suitable implant connections. Do not use healing components for longer than 6 months.

    The provisional components are intended to serve as a base for temporary crown or bridge restoration out of occlusion for the Straumann® PURE Ceramic Implant System. The Straumann® Temporary Abutment VITA CAD-Temp® for the Straumann® PURE Ceramic Implant is indicated for temporary usage of up to 180 days.

    CI RD Straumann PUREbase abutment is a titanium base placed onto Straumann ceramic dental implants to provide support for customized prosthetic restorations and is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Straumann® Variobase Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description

    The Straumann PURE Ceramic Implant is a dental implant made out of yttrium-stabilized zirconium dioxide (Y-TZP). The Straumann PURE Ceramic Implant is based on features of the Straumann PURE Ceramic implant Monotype. Straumann PURE Ceramic Implants can be placed using the existing instruments using the same osteotomy preparation protocol as for Bone Level implants. The subject implant is a two -piece implant and is available in Ø4.1 mm with lengths of 8, 10, 12, and 14 mm.

    The Closure and Healing caps are screws machined as one piece and come in three gingival heights to accommodate individual gingival thickness. The material of the devices is titanium grade 4. The Closure caps are screwed into the implant to protect the inner configuration and shoulder of the implant during the healing phase in cases of submerged (submucosal) healing protocols and do not support a prosthetic restoration. Healing caps are screwed into the implant to protect the inner configuration in cases of transmucosal healing protocols and are placed out of occlusion and do not support a prosthetic restoration.

    The Temporary Abutments are used to serve as a temporary crown or bridge restoration for the Straumann PURE Ceramic Implant System.

    The CI RD Straumann PUREbase Abutment for Ceramic implant is a premanufactured (stock) abutment (the first piece of the two-piece abutment), sometimes referred to as "Tibase", and is used as a base when fabricating a CAD/CAM customized restoration (the second piece of the two-piece abutment). The assembly of the two-pieces becomes a finished medical device after cementation of the CAD/CAM customized restoration on the PURE base abutment.

    AI/ML Overview

    The Straumann PURE Ceramic Implant System did not present specific, quantified acceptance criteria with corresponding performance data in the provided text. Instead, the document focuses on demonstrating substantial equivalence to predicate and reference devices through various tests.

    However, based on the provided "Performance Data 5.7" section and the comparison tables, we can infer the types of acceptance criteria and the nature of the study that proves the device meets those criteria. The acceptance criteria generally revolve around demonstrating equivalence to legally marketed predicate and reference devices in terms of safety and effectiveness, as per FDA guidelines for 510(k) submissions.

    Here's an attempt to structure the information based on the request, interpreting "acceptance criteria" as the performance standards expected for demonstrating substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since explicit numerical acceptance criteria were not stated, this table describes the type of performance evaluated and the outcome in relation to the predicate/reference devices.

    Acceptance Criterion (Inferred from regulatory requirements and comparative studies)Reported Device Performance
    Biocompatibility: Device materials are safe for biological contact and do not cause adverse reactions. (In accordance with ISO 10993-1:2009).The subject device materials are identical to the predicate and reference device materials. Therefore, no new issues regarding biocompatibility were raised.
    Sterilization Efficacy (for sterile components): Sterilization process reliably achieves a Sterility Assurance Level (SAL) of 10-6. (Per ISO 11135).Sterilization validation was performed using the Half Cycle Overkill Approach, demonstrating the process can reliably and consistently sterilize to a minimum SAL of 10-6.
    Pyrogenicity: Device does not contain pyrogens above acceptable limits. (Based on FDA Guidance, testing limit of 20 EU/device for blood contacting/implanted devices).The device meets pyrogen limit specifications as determined by LAL Endotoxin Analysis, with a testing limit of 20 EU/device.
    Reprocessing Validation (for non-sterile components to be reprocessed by end-user): Reprocessing methods (e.g., autoclaving) are effective. (Per ISO 17665-1:2006 and FDA guidance).The sterilization process for the Straumann PURE temporary abutment and CI RD Straumann PUREbase abutments (as recommended in labeling) was validated according to applicable recommendations in FDA guidance and ISO 17665-1:2006.
    Mechanical Strength/Fatigue: Device can withstand dynamic loading forces typical of oral function without failure. (Per ISO 14801:2016 and FDA guidance).Dynamic fatigue tests demonstrated the Straumann PURE Ceramic Implant, Straumann PURE temporary abutments, and CI RD Straumann PUREbase abutments are equivalent to the predicate and reference devices.
    Wear on Implant-to-Abutment Connection: Connection surfaces exhibit comparable wear behavior to predicate/reference devices.Post-fatigue bench tests concluded comparable behavior of the subject device to the reference devices in terms of wear on the implant-to-abutment connection.
    Screw Loosening: Risk of screw loosening is comparable to predicate/reference devices.Post-fatigue bench tests concluded comparable behavior of the subject device to the reference devices in terms of screw loosening. A literature review was also provided to evaluate the risk of screw loosening with ceramic abutments.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes bench testing for mechanical performance, biocompatibility, and sterilization. It does not refer to a "test set" in the context of clinical data or data from human subjects.

    • Sample Size: Not explicitly stated for each test, but typically, mechanical and sterilization validations involve a statistically representative number of units or replicates to demonstrate consistency and meet statistical confidence levels defined in the relevant ISO standards (e.g., ISO 11135, ISO 14801). Biocompatibility is based on material composition equivalency, not a specific sample size.
    • Data Provenance: The studies are described as bench tests (laboratory studies), biocompatibility assessments based on material equivalency, and sterilization/reprocessing validations. There is no mention of country of origin of data in the context of test subjects, as these are not human studies for this specific assessment. The data is generated from laboratory testing and analytical comparisons.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is typically relevant for studies involving human interpretation (e.g., image analysis by radiologists). For this device, which involves material properties and mechanical performance, the "ground truth" is established by adherence to international standards (ISO) and FDA guidance documents, and the expertise lies in the validation engineers and materials scientists performing and interpreting these tests. There is no mention of "experts" in the sense of clinical adjudicators for a "test set" in the provided text.

    4. Adjudication Method for the Test Set

    Not applicable as this is not a study involving human interpretation of medical data (e.g., images) requiring adjudication of findings. Performance is evaluated against objective engineering and scientific standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an endosseous dental implant system, not an AI-powered diagnostic or assistive technology.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm or software. It is a physical medical device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluations (mechanical, sterilization, biocompatibility) is defined by the International Organization for Standardization (ISO) standards and FDA guidance documents to which these tests were conducted.

    • Biocompatibility: ISO 10993-1:2009.
    • Sterilization: ISO 11135 and FDA Guidance on "Submission and Review of Sterility Information...".
    • Reprocessing: ISO 17665-1:2006 and FDA guidance "Reprocessing Medical Devices in Health Care Settings...".
    • Mechanical Testing (Dynamic Fatigue): ISO 14801:2016 and FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
    • Wear and Screw Loosening: Bench tests demonstrating comparable behavior to reference devices, supported by a literature review for screw loosening risk.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This device is not an AI/machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173116
    Date Cleared
    2018-06-28

    (272 days)

    Product Code
    Regulation Number
    872.3640
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K151328

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KJ ZIRCONIA Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. It is for delayed loading.

    Device Description

    KJ ZIRCONIA Implant System is a dental implant system made of ZIRCONIA intended to be surgically placed in the bone of the upper or lower jaw arches. It is one piece type implant which the implant and the abutment are manufactured as one piece. It offers two types: BL type and BS type, and the difference is only in the length of the abutment part.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA for a dental implant system. It primarily focuses on demonstrating substantial equivalence to previously marketed devices rather than presenting performance data from a clinical or algorithm-based study for an AI/ML device. Therefore, it does not contain the information needed to answer your request about acceptance criteria and study proving performance for an AI/ML device.

    Here's why the document doesn't provide the requested information:

    • Device Type: The "KJ ZIRCONIA Implant System" is a physical medical device (a dental implant), not an AI/ML software device.
    • Regulatory Pathway: A 510(k) submission for this type of device typically demonstrates equivalency to existing predicate devices based on design, materials, and intended use, supported by non-clinical testing (e.g., sterilization, biocompatibility, mechanical properties). It does not usually involve clinical performance studies in the way an AI/ML diagnostic or prognostic device would.
    • Absence of AI/ML Specifics: There is no mention of algorithms, machine learning models, test sets, ground truth establishment by experts, or MRMC studies, which are all pertinent to evaluating AI/ML product performance.

    Therefore, I cannot extract the requested information (acceptance criteria table, sample size, expert qualifications, adjudication, MRMC study, standalone performance, ground truth types, training set details) from this document as it pertains to an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1