K Number
K171409

Validate with FDA (Live)

Device Name
OT EQUATOR
Manufacturer
Date Cleared
2018-04-09

(329 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OT Equator is designed as an endosseous dental implant retentive component used to retain a complete or partial denture. The OT Equator is screwed into an endosseous implant in the mandible or maxilla.

The OT Equator abutments are indicated for use with the following implant systems: (list of compatible implant systems follows)

Device Description

The OT Equator Implant Anchor abutment for endosseous dental implants is appropriate for use with overdentures or partial denture retained in whole or in part by endosseous implants in the mandibula or maxilla. The device is intended for use in healthcare facilities/hospitals.

The sphere OT Equator technology represents a prosthetic abutment solution for "overdenture". The emerging profile, specifically designed to accommodate the relative interchangeable copings OT Equator, allows the coupling with a removable prosthesis or "overdenture", which is properly built only if it is properly profiled and based on the patient's gingival arch. Therefore, the removable prosthesis unloads the chewer loads on the natural gum of the intra oral arch, making the OT Equator abutments passive with lacking of solicitation.

The abutments are for straight use only and are recommended for placement on straight implant bodies with no divergence or need for angle corrections.

The OT Equator is manufactured of titanium Ti-6Al-4V (meets ASTM Standard F-136) and it is designed to retain any prosthetic restoration.

The OT Equator overdenture system is comprised of a semispherical head, a base, and a screw.

AI/ML Overview

This document is a 510(k) Premarket Notification from the U.S. Food and Drug Administration (FDA) for the OT Equator dental implant abutment. It is a regulatory document and does not contain information about the acceptance criteria or study performance of an AI/ML device.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and the reported device performance
  2. Sample size used for the test set and the data provenance
  3. Number of experts used to establish the ground truth for the test set and their qualifications
  4. Adjudication method for the test set
  5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study effect size
  6. Standalone performance
  7. Type of ground truth used
  8. Sample size for the training set
  9. How the ground truth for the training set was established

The document focuses on demonstrating substantial equivalence of the OT Equator device to previously cleared predicate devices based on materials, design, intended use, and manufacturing processes, rather than presenting a study related to an AI/ML device and its performance metrics. The non-clinical tests mentioned (biocompatibility, sterilization validation, reverse engineering analysis) are performed on the physical dental implant abutment and its interaction with implant systems, not on an AI system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the words "U.S. Food & Drug Administration" on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA text is in a blue sans-serif font.

Rhein'83 SRL Claudia Nardi President Rhein'83 SRL Via E. Zago, 10/abc Bologna, 40128 ITALY

Re: K171409

Trade/Device Name: OT Equator Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 6, 2018 Received: March 8, 2018

Dear Claudia Nardi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

April 9, 2018

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mary S. Runner -S

forTina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K 171409 Device Name

OT Equator

Indications for Use (Describe)

The OT Equator is designed as an endosseous dental implant retentive component used to retain a complete or partial

denture. The OT Equator is screwed into an endosseous implant in the mandible or maxilla.

The OT Equator abutments are indicated for use with the following implant systems:

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

ImplantManufacturerImplant Family NameImplant NameImplantDiameter (mm)
NOBEL BIOCARE ABNobelActiveNobelActive NP3.5
NOBEL BIOCARE ABNobelActiveNobelActive RP4.3 and 5
NOBEL BIOCARE USALLCNobelreplace Tapered ConicalConnectionReplace Select TC NP3.5
NOBEL BIOCARE UASINCReplace Tiunite EndosseousImplantReplace Select Straight TiUnitRP4.3
NOBEL BIOCARE UASINCReplace Tiunite EndosseousImplantReplace Select Straight TiUnitWP5
NOBEL BIOCARE ABVarious Branemark SystemImplants-Immediate FunctionIndicationBrånemark System Mk IIITiUnite NP3.3
NOBEL BIOCARE ABVarious Branemark SystemImplants-Immediate FunctionIndicationBrånemark System Mk IIIShorty RP4.1
NOBEL BIOCARE ABVarious Branemark SystemImplants-Immediate FunctionIndicationBrånemark System Mk IIITiUnite WP5 and 6
STRAUMANN USAStraumann Bone Level TaperedImplantBone level NC3.3
STRAUMANN USAStraumann Bone Level TaperedImplantBone level RC4.1 and 4.8
STRAUMANN USAITI Dental Implant System(Tissue Level TitaniumImplants)Tissue level RN4.8
STRAUMANN USAITI Dental Implant System(Tissue Level TitaniumImplants)Tissue level WN6.5
ZIMMER DENTAL INC.Tapered Screw-Vent ImplantSystemTapered Screw-Vent3.5, 4.5 and 5.7
ZIMMER DENTAL INCSpline Dental Implant SystemSpline3.25, 3.75, 4 and5
ASTRA TECH ABAstra Tech Implant SystemOsseoSpeed™ TX3.5, 4, 4.5 and 5
DENTSPLYINTERNATIONAL, INC.Ankylos C/X Dental ImplantSystemAnkylos C/3.5
DENTSPLYINTERNATIONAL, INC.Frialit Plus, Xive S Plus, Xive TgPlus, Ankylos Plus, DentalImplant SystemsFrialit plus Stepped Screw3.4, 3.8, 4.5, 5.5and 6.5
BIOMET 3iFull OSSEOTITE® Certain® IIDental ImplantCertain3.25, 4.1, 5 and 6
ImplantManufacturerImplant Family NameImplant NameImplant Diameter(mm)
IMPLANT DIRECT LLCLegacy Dental ImplantsLegacy3, 3.75 and 4.7
IMPLANT DIRECT LLCReactive Dental Implant SystemTri-Lobe3.5, 4.3, 5 and 6
IMPLANT DIRECT LLCReplus Dental ImplantsTri-Lobe3.5, 4.3, 5 and 6
IMPLANT DIRECTSYBRONMANUFACTURING LLCInteractive/ Swishplus2Implant SystemSwish Plus3.7, 4.1 and 4.8
NEODENT USA, INC.Neodent Implant SystemCM Titamax3.5, 3.75, 4, 4.3 and5
NEODENT USA, INC.Neodent Implant SystemCM Drive3.5, 3.75, 4, 4.3 and5
MIS - IMPLANTTECHNOLOGIES LTD.Seven Implants; BiocomImplants; Lance ImplantsSeven3.3, 3.75, 4.2, 5 and6
MIS - IMPLANTTECHNOLOGIES LTD.Conical Connection ImplantsC13.3, 3.75, 3.9, 4.2,4.3 and 5
MEGAGEN CO., LTD.Anyridge Internal ImplantSystemAnyridge3.5, 4, 4.5, 5 and 5.5
NEOSS, LTD.Neoss Proactive ImplantProActive3.5, 4, 4.5, 5, 5.5 and6
BIOHORIZONS IMPLANTSYSTEMS, INC.Biohorizons Laser-Lok 3.0Implant SystemLaser-lok®3.0
BIOHORIZONS IMPLANTSYSTEMS, INC.Biohorizons Tapered InternalImplant SystemTapered internal3.5, 4.5 and 5.7
KEYSTONE DENTAL INC./ LIFECOREBIOMEDICAL INCPrimaconnex InternalConnection Implant System,Primaconnex InternalConnection ProstheticsPrimaconnex InternalConnection Implant SystemSD3.5
KEYSTONE DENTAL INC./ LIFECOREBIOMEDICAL INCPrimaconnex InternalConnection Implant System,Primaconnex InternalConnection ProstheticsPrimaconnex InternalConnection Implant SystemRD4.1
KEYSTONE DENTAL INC./ LIFECOREBIOMEDICAL INCPrimaconnex InternalConnection Implant System,Primaconnex InternalConnection ProstheticsPrimaconnex InternalConnection Implant SystemWD5
TRI DENTAL IMPLANTSINT. AGTRI® Dental Implant SystemTRI Narrow3.3
TRI DENTAL IMPLANTSINT. AGTRI® Dental Implant SystemTRI Vent3.75, 4.1 and 4.7
TRI DENTAL IMPLANTSINT, AGTRI® Dental Implant SystemTRI OCTA3.75, 4.1 and 4.7
ImplantManufacturerImplant Family NameImplant NameImplant Diameter(mm)
ALTATEC GMBHCamlog Implant SystemModified Implants andAbutmentsCamlog3.3 , 3.8 , 4.3 , 5 and 6
ALTATEC GMBHConelog Implant SystemConelog3.3 , 3.8 , 4.3 and 5
DENTIS CO.Dentis Dental Implant Systemi-Clean3.7 , 4.3 and 4.8
DENTIS CO.OneQ-SL Implant SystemCommoni-Clean3.7 , 4.3 and 4.8
DENTIS CO.OneQ-SL s-Clean ImplantSystems-Clean Tapered II RBM ImplantSystems-clean3.7 , 4.3 and 4.8
DENTIS CO.Dentis Dental Implant Systeme-Clean3.5 , 4.1 and 5.1
DENTSPLY IMPLANTSMANUFACTURING GMBHAstra Tech Implant SystemOsseoSpeed Profile EV3 , 3.6 , 4.2 and 4.8
MEGAGEN IMPLANT CO.Anyone Internal ImplantSystemAnyone3.6 , 4.0 , 4.5 , 5 and 6

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

K 171409 Separate page

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

Indications for Use

K 171409 Separate page

{6}------------------------------------------------

510(k) Summary

Applicant:

Company Name:Rhein'83 srl
Company Address:via E. Zago 10/abc - 40128 Bologna - Italy
Company Phone:+39 051244510
Company Fax:+39 051245238
Company e-mailinfo@rhein83.com
Official Contact for Correspondence:Claudia Nardi - President Rhein'83 srl
Phone:+39 051244510
E-mail:info@rhein83.com

Date Summary Prepared:

April 4, 2018

DEVICE IDENTIFICATION

Trade name:OT EQUATOR
Generic/ Common Name:Implant Abutment
Classification:21 CFR 872.3630
Classification name:Endosseous dental implant AbutmentClass II
Product Code:NHA
Panel:Dental

PREDICATE DEVICES:

Primary predicate: K142211, OT Equator, Rhein'83 srl Reference device: K160382, OT Equator, Rhein'83 srl

DEVICE DESCRIPTION

The OT Equator Implant Anchor abutment for endosseous dental implants is appropriate for use with overdentures or partial denture retained in whole or in part by endosseous implants in the mandibula or maxilla. The device is intended for use in healthcare facilities/hospitals.

The sphere OT Equator technology represents a prosthetic abutment solution for "overdenture". The emerging profile, specifically designed to accommodate the relative interchangeable copings OT

{7}------------------------------------------------

Equator, allows the coupling with a removable prosthesis or "overdenture", which is properly built only if it is properly profiled and based on the patient's gingival arch. Therefore, the removable prosthesis unloads the chewer loads on the natural gum of the intra oral arch, making the OT Equator abutments passive with lacking of solicitation.

The abutments are for straight use only and are recommended for placement on straight implant bodies with no divergence or need for angle corrections.

The OT Equator is manufactured of titanium Ti-6Al-4V (meets ASTM Standard F-136) and it is designed to retain any prosthetic restoration.

The OT Equator overdenture system is comprised of a semispherical head, a base, and a screw.

The part number, critical dimensions and materials for the OT Equator Abutment and accessories subject to this 510(k) are summarized below:

Component name anddrawingPart NumberCritical DimensionsMaterial
OT EQUATOR ABUTMENTImage: OT EQUATOR ABUTMENT030HEAD DIAMETER: 2.5mmHEAD HEIGHT: 2.0 MMCUFF HEIGHT: From 1 to 7mmTHREAD DIAMETER:Congruent with implant ThreadedCONNECTIONS: Different implantplatformsTITANIUM Medical Grade 5E.L.I. (Ti6AL4V)Surface treatment: TiNcoating (Nitride)ONLY IN THE HEAD OF THEABUTMENT
OT EQUATOR - VIOLETRETENTIVE CAPImage: OT EQUATOR - VIOLET RETENTIVE CAP140CEVEXTERNAL Ø 3.8mmINTERNAL Ø 2.5mmKEPITAL
OT EQUATOR - WHITERETENTIVE CAPImage: OT EQUATOR - WHITE RETENTIVE CAP140CETEXTERNAL Ø 3.8mmINTERNAL Ø 2.5mmRILSAN BMNO
OT EQUATOR - PINKRETENTIVE CAPImage: OT EQUATOR - PINK RETENTIVE CAP140CEREXTERNAL Ø 3.8mmINTERNAL Ø 2.5mmRILSAN BMNO
OT EQUATOR - YELLOWRETENTIVE CAPImage: OT EQUATOR - YELLOW RETENTIVE CAP140CEGEXTERNAL Ø 3.8mmINTERNAL Ø 2.5mmPEBAX

{8}------------------------------------------------

Component name anddrawingPart NumberCritical DimensionsMaterial
STAINLESS STEELHOUSING FOR CAPImage: Stainless steel housing for cap141CAEINTERNAL Ø 3.84STAINLESS STEEL AISI303

INTENDED USE /INDICATIONS FOR USE

The OT Equator is designed as an endosseous dental implant retentive component used to retain a complete or partial denture. The OT Equator is screwed into an endosseous implant in the mandible or maxilla.

The OT Equator abutments are indicated for use with the following implant systems:

ImplantManufacturerImplant Family NameImplant NameImplantDiameter (mm)
NOBEL BIOCARE ABNobelActiveNobelActive NP3.5
NOBEL BIOCARE ABNobelActiveNobelActive RP4.3 and 5
NOBEL BIOCARE USALLCNobelreplace Tapered ConicalConnectionReplace Select TC NP3.5
NOBEL BIOCARE UASINCReplace Tiunite EndosseousImplantReplace Select Straight TiUnitRP4.3
NOBEL BIOCARE UASINCReplace Tiunite EndosseousImplantReplace Select Straight TiUnitWP5
NOBEL BIOCARE ABVarious Branemark SystemImplants-Immediate FunctionIndicationBrånemark System Mk IIITiUnite NP3.3
NOBEL BIOCARE ABVarious Branemark SystemImplants-Immediate FunctionIndicationBrånemark System Mk IIIShorty RP4.1
NOBEL BIOCARE ABVarious Branemark SystemImplants-Immediate FunctionIndicationBrånemark System Mk IIITiUnite WP5 and 6
STRAUMANN USAStraumann Bone Level TaperedImplantBone level NC3.3
STRAUMANN USAStraumann Bone Level TaperedImplantBone level RC4.1 and 4.8
STRAUMANN USAITI Dental Implant System(Tissue Level TitaniumImplants)Tissue level RN4.8
STRAUMANN USAITI Dental Implant System(Tissue Level TitaniumImplants)Tissue level WN6.5
ZIMMER DENTALINC.Tapered Screw-Vent ImplantSystemTapered Screw-Vent3.5, 4.5 and 5.7
ZIMMER DENTALINCSpline Dental Implant SystemSpline3.25,3.75,4 and 5
ASTRA TECH ABAstra Tech Implant SystemOsseoSpeed™ TX3.5 , 4 , 4.5 and 5
DENTSPLYINTERNATIONAL, INC.Ankylos C/X Dental ImplantSystemAnkylos C/3.5
DENTSPLYINTERNATIONAL, INC.Frialit Plus, Xive S Plus, Xive TgPlus, Ankylos Plus, DentalImplant SystemsFrialit plus Stepped Screw3.4, 3.8, 4.5, 5.5and 6.5
BIOMET 3iFull OSSEOTITE® Certain® IIDental ImplantCertain3.25, 4.1, 5 and 6
IMPLANT DIRECT LLCLegacy Dental ImplantsLegacy3, 3.75 and 4.7
IMPLANT DIRECT LLCReactive Dental Implant SystemTri-Lobe3.5, 4.3, 5 and 6
IMPLANT DIRECT LLCReplus Dental ImplantsTri-Lobe3.5, 4.3, 5 and 6
IMPLANT DIRECTSYBRONMANUFACTURING LLCInteractive/ Swishplus2Implant SystemSwish Plus3.7, 4.1 and 4.8
NEODENT USA, INC.Neodent Implant SystemCM Titamax3.5, 3.75, 4, 4.3and 5
NEODENT USA, INC.Neodent Implant SystemCM Drive3.5, 3.75, 4, 4.3 and5
MIS - IMPLANTTECHNOLOGIES LTD.Seven Implants; BiocomImplants; Lance ImplantsSeven3.3, 3.75, 4.2, 5 and6
MIS - IMPLANTTECHNOLOGIES LTD.Conical Connection ImplantsC13.3, 3.75, 3.9, 4.2,4.3 and 5
MEGAGEN CO., LTD.Anyridge Internal ImplantSystemAnyridge3.5, 4, 4.5, 5 and5.5
NEOSS, LTD.Neoss Proactive ImplantProActive3.5, 4, 4.5, 5, 5.5and 6
BIOHORIZONS IMPLANTSYSTEMS, INC.Biohorizons Laser-Lok 3.0Implant SystemLaser-lok®3.0
BIOHORIZONS IMPLANTSYSTEMS, INC.Biohorizons Tapered InternalImplant SystemTapered internal3.5, 4.5 and 5.7
KEYSTONE DENTAL INC./ LIFECOREBIOMEDICALINCPrimaconnex InternalConnection Implant System,Primaconnex InternalConnection ProstheticsPrimaconnex InternalConnection Implant System SD3.5
KEYSTONE DENTAL INC./ LIFECOREBIOMEDICALINCPrimaconnex InternalConnection Implant System,Primaconnex InternalConnection ProstheticsPrimaconnex InternalConnection Implant System RD4.1
KEYSTONE DENTAL INC./ LIFECOREBIOMEDICALINCPrimaconnex InternalConnection Implant System,Primaconnex InternalConnection ProstheticsPrimaconnex InternalConnection Implant System WD5
TRI DENTAL IMPLANTSINT. AGTRI® Dental Implant SystemTRI Narrow3.3
TRI DENTAL IMPLANTSINT. AGTRI® Dental Implant SystemTRI Vent3.75, 4.1 and 4.7
TRI DENTAL IMPLANTSINT. AGTRI® Dental Implant SystemTRI OCTA3.75, 4.1 and 4.7
ALTATEC GMBHCamlog Implant SystemModified Implants andAbutmentsCamlog3.3, 3.8, 4.3, 5 and6
ALTATEC GMBHConelog Implant SystemConelog3.3, 3.8, 4.3 and 5
DENTIS CO.Dentis Dental Implant Systemi-Clean3.7, 4.3 and 4.8
DENTIS CO.OneQ-SL Implant SystemCommoni-Clean3.7, 4.3 and 4.8
DENTIS CO.OneQ-SL s-Clean ImplantSystems-Clean Tapered II RBM ImplantSystems-clean3.7, 4.3 and 4.8
DENTIS CO.Dentis Dental Implant Systeme-Clean3.5, 4.1 and 5.1
DENTSPLY IMPLANTSMANUFACTURINGGMBHAstra Tech Implant SystemOsseoSpeed Profile EV3, 3.6, 4.2 and 4.8
MEGAGEN IMPLANT CO.Anyone Internal Implant SystemAnyone3.6, 4.0, 4.5, 5 and6

{9}------------------------------------------------

{10}------------------------------------------------

DISCUSSION OF NON CLINICAL TESTS

Biocompatibility

The submitted device, as the predicate devices already licensed (OT EQUATOR K142211 and K160382), is classified as permanent duration (> 30 days), mucosal membranes contacting device. The materials used to manufacture the submitted device are identical to the materials used for the identified predicate devices, the already licensed OT EQUATOR (K142211 and K160382). The materials used to manufacture the submitted device are identical to the materials used for the identified predicate devices and the same process is performed, including the TiN (Titanium Nitride) coating for the OT EQUATOR abutment; the modified surface of the abutment after the coating process is identical to the surface of the predicate devices. The type and duration of patient contact are the same as well.

Therefore, the results of the biocompatibility tests performed on the predicate devices OT EQUATOR apply also to the subject device and there was no need to perform new biocompatibility testing on the subject device OT EQUATOR. No additional biocompatibility testing is needed for the determination of substantial equivalence.

Sterilization Validation

Steam sterilization validation test, conducted on the predicate device OT EQUATOR (K142211) according to ISO 17664 and ISO 11737-1/-2 in order to demonstrate as SAL of 10-6 related to the OT Equator abutments, are considered still valid, since the materials and packing of the subject device are the same of the predicate.

Reverse engineering analysis

In order to ensure that the OT Equator abutment can be coupled with the related compatible implant a reverse engineering analysis was conducted on the submitted device, as for the predicate devices already licensed; 10 samples of OEM implant system were used in the reverse engineering analysis: 10 OEM implant bodies and 10 OEM screws were reversed as for the predicate and reference devices, and the results were reported in a chart form.

The implant dimensions were detected with a stereo microscope and with a wide range of calibrated plugs. The dimensions of the threaded holes and the related tolerance degree were detected with the "qo/no qo" threaded gauges. In the same way the maximum depth of the threaded hole was detected with the related threaded gauge, with the aid of the profile projector.

{11}------------------------------------------------

The results of nonclinical tests demonstrate that the device is equivalent to the predicate devices.

SUBSTANTIAL EQUIVALENCE

The OT EQUATOR is same or similar in materials, design and intended use to the predicate devices. In further support of a substantial equivalence determination, hereunder is a comparison chart with the submitted device and the predicate devices.

FeatureRhein83 OT Equator(Submitted Product)PREDICATE DEVICEREFERENCE DEVICE
K numberN.A.K160382K142211
Proprietary /Trade NameOT EQUATOROT EQUATOROT EQUATOR
CFR Section872.3630872.3630872.3630
Pro-codeNHANHANHA
ClassificationnameENDOSSEOUS DENTALIMPLANTABUTMENTENDOSSEOUS DENTALIMPLANTABUTMENTENDOSSEOUS DENTALIMPLANTABUTMENT
Class II (special controls)Class II (special controls)Class II (special controls)
Indications ForUse / IntendedUseThe OT Equator is designed as anendosseous dental implant retentivecomponent used to retain acomplete or partial denture. The OTEquator isscrewed into anendosseousimplant in the mandibleor maxilla. (contains a list ofcompatible implant bodies)The OT Equator is designed as anendosseous dental implant retentivecomponent used to retain a completeor partial denture. The OT Equator isscrewed into an endosseous implantin the mandible or maxilla. (containsa list of compatible implant bodies)The OT Equator is designed as anendosseous dental implant retentivecomponentused to retain a completeor partial denture. The OT Equator isscrewed into an endosseous implantin the mandible or maxilla. (containsa list of compatible implant bodies)
Intended UsersAdequately trained dentaltechnicians and dentists.Adequately trained dental techniciansand dentists.Adequately trained dental techniciansand dentists.
MaterialTitanium 6Al-4VTitanium 6Al-4VTitanium 6Al-4V
Material(Male socket)Rilsan, Pebax, KepitalRilsan, Pebax, KepitalRilsan, Pebax, Kepital
Material(Housing)Stainless Steel AISI303Stainless Steel AISI303Stainless Steel AISI303
PlatformDiameterAccording to the compatible implant(see table above)According to the compatible implantAccording to the compatible implant
SurfacetreatmentTiN coating only for the head part ofthe abutmentTiN coating only for the head part ofthe abutmentTiN coating only for the head part ofthe abutment
Cuff Width1 mm - 2mm - 3mm - 4mm -5mm -6mm - 7mm1 mm - 2mm - 3mm - 4mm -5mm -6mm – 7mm1mm - 2mm - 3mm - 4mm -5mm -6mm – 7mm
Height2 mm2 mm2 mm
ComponentsOT EQUATOR PROFILE kitcontains: Ot Equator Profileabutment, Stainless Steel Housing,retentive caps, protective diskOT EQUATOR PROFILE kitcontains: Ot Equator Profileabutment, stainless steel housing ,retentive caps, protective disk.OT EQUATOR PROFILE kit contains:Ot Equator Profile abutment, stainlesssteel housing , retentive caps,protective disk
SterilizationMarketed non sterile, to be sterilizedbefore insertion in patient's mouth.Marketed non sterile, to be sterilizedbefore insertion in patient's mouth.Marketed non sterile, to be sterilizedbefore insertion in patient's mouth.
ReusableNoNoNo

{12}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE DISCUSSION:

The Indications for Use Statements shown above in the comparison chart a list of compatible implant bodies. The differences between the Indications for Use of the subject device and the primary predicate device is the addition of compatible implant bodies (Tri Dental Implants Int. AG, Altatec GMBH, Dentis Co., Dentsply Implants Manufacturing GMBH and Megagen Implant Co.).

As shown in the table above, the submitted device is substantial equivalent in all features: indications for use and intended users, materials, manufacturing process, cuff width, height, status, components.

The subject device varies from the predicate devices only for the list of compatible implant systems, thus the only difference is in the interface connections between the abutments and the compatible implant bodies, that have specific dimensions according to the compatible implant systems. As for the predicate devices, in order to ensure that the OT Equator abutment can be coupled with the related compatible implant a reverse engineering analysis was conducted on the submitted device.

There are no additional differences, thus it was concluded that the subject dental implant is substantially equivalent to the predicate devices.

CONCLUSIONS

Based on the available information, we conclude that the OT Equator is substantially equivalent to the existing legally marketed device under Federal Food, Drug and Cosmetic Act. Therefore, the subject device is determined to be equivalent to the predicate devices.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)