K Number
K173968

Validate with FDA (Live)

Date Cleared
2018-07-17

(200 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components directly connected to the endosseous dental implants are indicated for use as an aid in prosthetic rehabilitations. The patient-specific prosthetic restoration (bridge or over-denture) can be cemented on the Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Straumann® Variobase™ for Bridge/Bar Cylindrical and patient-specific restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed Straumann® Variobase® for Bridge/Bar Cylindrical prosthetic components are intended to be sent to Straumann for manufacture at a validated milling center.

Device Description

The Straumann® Variobase™ for Bridge/Bar Cylindrical, see Figure 1 and Table 1, are non-engaging (without rotational lock) abutments made from Ti-6Al-7Nb (TAN) that support a bridge or bar reconstruction (framework or full contour) on two or more dental implants. The corresponding basal screw is delivered with the abutment for connecting the abutment to the implant. A dental laboratory technician designs and manufactures the bridge/bar reconstruction via their preferred workflow using traditional or CAD/CAM methods.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called "Straumann® Variobase™ for Bridge/Bar Cylindrical". It does not describe a study that uses software, AI, or machine learning. Therefore, I cannot provide the requested information about acceptance criteria, device performance, sample sizes, ground truth establishment, or multi-reader multi-case studies as these concepts are not applicable to the provided document.

The document focuses on demonstrating substantial equivalence to predicate devices through technical characteristics and bench testing (dynamic fatigue, software validation, sterilization validation, and biocompatibility testing) for a dental implant abutment. The "Software validation conforming to the requirements of IEC 62304" mentioned in the Performance Data section refers to the validation of software used in the manufacture of the device (specifically, for digitally designed components intended to be sent to a validated milling center), not a clinical performance or diagnostic AI component of the device itself.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 17, 2018

Institut Straumann AG Jennifer Jackson Director, Regulatory Affairs and Quality 60 Minuteman Road Andover, Massachusetts 01810

Re: K173968

Trade/Device Name: Straumann® Variobase™ for Bridge/Bar Cylindrical Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 14, 2018 Received: June 18, 2018

Dear Jennifer Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K173968

Device Name

Straumann® Variobase™ for Bridge/Bar Cylindrical

Indications for Use (Describe)

Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components directly connected to the endosseous dental implants are indicated for use as an aid in prosthetic rehabilitations. The patientspecific prosthetic restoration (bridge or over-denture) can be cemented on the Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Straumann® Variobase™ for Bridge/Bar Cylindrical and patient-specific restorations may be placed into occlusion when the implant is fully osseointegrated.

All digitally designed Straumann® Variobase® for Bridge/Bar Cylindrical prosthetic components are intended to be sent to Straumann for manufacture at a validated milling center.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)

| | Over-The-Counter Use (21 CFR 801

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to espond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)Equivalent

{3}------------------------------------------------

510(k) Summary

Submitter's Contact Information
Submitter:Straumann USA, LLC (on behalf of Institut Straumann AG)60 Minuteman RoadAndover, MA 01810Registration No.: 1222315Owner/Operator No.: 9005052
Contact Person:Jennifer JacksonDirector, Regulatory Affairs & Quality
Prepared By:Dr. Gordon DoddsManager Design Control QMEtkon GmbH
Date Prepared:July 16, 2018
Product Code(s):NHA (21 CFR 872.3630)
Device Class:II (21 CFR 872.3630)
Classification Panel:Dental
Name of the Device
Classification Name:Endosseous dental implant abutment (21 CFR 872.3630)
Proprietary Name:Straumann® Variobase™ for Bridge/Bar Cylindrical
Predicate Device(s):
Predicate Device(s):K151157 Straumann Variobase for Bridge/Bar K142890 Straumann Variobase Abutments K120822 Straumann CARES Variobase AbutmentNNC, RN, WN, NC, RC K132844 Straumann CARES Bone Level ScrewRetained Bars, Straumann CARES Bone LevelScrew Retained Bridges

{4}------------------------------------------------

  • K041295 RN Synocta UCLA Gold Abutment for . the Straumann Dental Implant

Device Description

The Straumann® Variobase™ for Bridge/Bar Cylindrical, see Figure 1 and Table 1, are non-engaging (without rotational lock) abutments made from Ti-6Al-7Nb (TAN) that support a bridge or bar reconstruction (framework or full contour) on two or more dental implants. The corresponding basal screw is delivered with the abutment for connecting the abutment to the implant. A dental laboratory technician designs and manufactures the bridge/bar reconstruction via their preferred workflow using traditional or CAD/CAM methods.

Image /page/4/Picture/4 description: The image shows five dental implant abutments labeled NC, RC, NNC, RN, and WN. The abutments are all silver in color, except for the NC abutment, which has a gold-colored base. The abutments are all different shapes and sizes, with the NC abutment being the smallest and the WN abutment being the largest. The abutments are used to connect dental implants to crowns or other dental prosthetics.

Figure 1 Straumann Variobase for Bridge/Bar Cylindrical

Table 1 - Listing of devices

Device NameDevice Number
Straumann® NC Variobase™ for Bridge/BarCylindrical022.0110 / 010.6085
Straumann® RC Variobase™ for Bridge/BarCylindrical022.0111 / 010.6086
Straumann® RN Variobase™ for Bridge/BarCylindrical048.378 / 010.6083
Straumann® WN Variobase™ for Bridge/BarCylindrical048.379 / 010.6084
Straumann® NNC Variobase™ for Bridge/BarCylindrical048.377 / 010.6082

Intended Use:

{5}------------------------------------------------

Straumann® Variobase™ for Bridge/Bar Cylindrical are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as bridges and over-dentures

Indications For Use

Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components directly connected to the endosseous dental implants are indicated for use as an aid in prosthetic rehabilitations. The patient-specific prosthetic restoration (bridge or overdenture) can be cemented on the Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Straumann® Variobase™ for Bridge/Bar Cylindrical and patient-specific restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed Straumann® Variobase® for Bridge/Bar Cylindrical prosthetic components are intended to be sent to Straumann for manufacture at a validated milling center.

{6}------------------------------------------------

Technological Chracteristics

The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following table.

Table 2: Substantial Equivalence Comparison

{7}------------------------------------------------

FEATURESubject DeviceStraumann® Variobase™ for Bridge/BarCylindricalSubject SubmissionPredicate DeviceStraumann® Variobase™Abutments for Bridge/Bar(K151157)Reference PredicateDevicesStraumann® Variobase™Abutments(K120822, K142890)Straumann® Screws(K132844, K041295)EquivalenceDiscussion
Indications forUseStraumann® Variobase™ for Bridge/BarCylindrical prosthetic components directlyconnected to the endosseous dental implantsare indicated for use as an aid in prostheticrehabilitations. The patient-specific prostheticrestoration (bridge or over-denture) can becemented on the Straumann® Variobase™ forBridge/Bar Cylindrical prosthetic components.A temporary restoration can be used prior tothe insertion of the final components tomaintain, stabilize and form the soft tissueduring the healing phase. Temporaryrestorations are indicated to be placed out ofocclusion. Straumann® Variobase™ forBridge/Bar Cylindrical and patient-specificrestorations may be placed into occlusionwhen the implant is fully osseointegrated.All digitally designed Straumann®Variobase® for Bridge/Bar Cylindricalprosthetic components are intended to besent to Straumann for manufacture at avalidated milling center.Straumann® Variobase® prostheticcomponents directly or indirectlyconnected to the endosseous dentalimplant are intended for use as an aidin prosthetic rehabilitations. Theprosthetic restoration (bridge oroverdenture) can be cemented on theStraumann® Variobase® prostheticcomponents. A temporary restorationcan be used prior to the insertion ofthe final components to maintain,stabilize and form the soft tissueduring the healing phase. They maynot be placed into occlusion. Finalabutments and restorations may beplaced into occlusion when theimplant is fully osseointegrated.Not applicableIdenticalSection regardingproduction methods hasbeen added. The sameproduction methods areused for the predicatedevice.
AbutmentMaterialTitanium-Aluminum-Niobium alloy (Ti-6Al-7Nb)Titanium-Aluminum-Niobium alloy (Ti-6Al-7Nb)Not applicableIdentical
FEATURESubject DeviceStraumann® Variobase™ for Bridge/BarCylindricalSubject SubmissionPredicate DeviceStraumann® Variobase™Abutments for Bridge/Bar(K151157)Reference PredicateDevicesStraumann® Variobase™Abutments(K120822, K142890)Straumann® Screws(K132844, K041295)EquivalenceDiscussion
AbutmentDiameter4.5 – 7.0 mm4.2 – 7.0 mmNot applicableEquivalentThe subject device NCinterface is 0.3 mm widerand gives more support
AbutmentHeight3.5 – 4.5 mm3.5 – 4.5 mmNot applicableIdentical
AbutmentApicalDesignNon-engaging (no rotational lock) Morse taperNon-engaging (no rotational lock)Morse taperNot applicableIdentical
AbutmentCoronalDesignStraight wall with detents10° or15° conical taperK120822, K142890 -Straight wall with detentsEquivalentThe coronal design isidentical to the referencepredicate devices and hasthe equivalent function asthe predicate devices.
Basal ScrewDesignNC/RC Conical Head Screw Bone LevelRN/WN Conical Head Screw Tissue LevelK132844, K041295 -NC/RC Conical Head ScrewBone LevelRN/WN Conical HeadScrew Tissue LevelIdentical
FEATURESubject DeviceStraumann® Variobase™ for Bridge/BarCylindricalSubject SubmissionPredicate DeviceStraumann® Variobase™Abutments for Bridge/Bar(K151157)Reference PredicateDevicesStraumann® Variobase™Abutments(K120822, K142890)Straumann® Screws(K132844, K041295)EquivalenceDiscussion
RestorationTypesSupportedPatient-specific prosthetic bridge or over-denturePatient-specific prosthetic bridge orover-dentureNot applicableIdentical
RestorationMaterialMaterials cleared by the FDA under 21 CFR872.6660 or exempt materials as describedunder 21 CFR 872.3060 (Noble metal alloys)and 21 CFR 872.3710 (Base metal alloys)Materials cleared by the FDA under21 CFR 872.6660 or exemptmaterials as described under 21 CFR872.3060 (Noble metal alloys) and 21CFR 872.3710 (Base metal alloys)Not applicableIdentical
DesignWorkflowWax-up or CADWax-up or CADNot applicableIdentical
ManufacturingWorkflowTraditional casting or Straumann Milling CenterTraditional casting, pressing orStraumann Milling CenterNot applicableIdentical
MechanicalStabilityDynamic fatigue testPassDynamic fatigue testPassNot applicableIdentical
SterilizationSteam autoclaveSteam autoclaveNot applicableIdentical
FEATURESubject DeviceStraumann® Variobase™ for Bridge/BarCylindricalSubject SubmissionPredicate DeviceStraumann® Variobase™Abutments for Bridge/Bar(K151157)Reference PredicateDevicesStraumann® Variobase™Abutments(K120822, K142890)Straumann® Screws(K132844, K041295)EquivalenceDiscussion
Mode of ActionScrew-retained or cement retainedScrew-retained or cement retainedNot applicableIdentical
ReusableNoNoNot applicableIdentical

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Performance Data

Per Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments dated May 12, 2004, the substantial equivalence of the subject device(s) are satisfactorily addressed via bench studies.

The following testing has been conducted or substantial equivalence to tested predicate devices has been demonstrated:

  • Dynamic fatigue testing conforming to FDA guidance and ISO 14801. ●
  • . Software validation conforming to the requirements of IEC 62304.
  • . Sterilization validation conforming to ISO 17665-1 and ISO/TS 17665-2 (K151157 Straumann Variobase for Bridge/Bar).
  • Biocompatibility testing conforming to ISO-10993-1 (K120822 Straumann CARES . Variobase Abutment NNC, RN, WN, NC, RC)

Conclusions

Based upon our assessment of the design and applicable performance data, the subject devices have been determined to be substantially equivalent to the identified predicate devices.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)