Search Results
Found 3 results
510(k) Data Aggregation
(393 days)
The BTI Dental Implant System UnicCa® for oral implant surgery is to be used for the partial or total replacement of teeth in edentulate patients. Once attached to the bone, the implants act as an anchor for various fixed or removable prosthetic solutions that can be used to improve or restore a patient's mastication function.
In the case of 5.5 - 6.5mm long UnicCa® implants should be used in a two-stage surgical procedure. These implants are indicated for delayed loading. These implants are indicated only for straight abutments and to support permanently fixed restorations.
In the case of Tiny® 3.0 UnicCa® implants: These implants shall be used only to replace maxillary lateral incisors and mandibular lateral and central incisors. Immediate loading is recommended when there is good primary stability and an appropriate occlusal load.
BTI Dental Implant System UnicCa® is a self-tapping, threaded, root form dental titanium implant provided in two types of connections: external (i.e., Externa®) and interna®), in a variety of platforms and range of diameters (3.0 – 6.0 mm) and lengths (5.5 – 18.0 mm). BTI Dental Implant System UnicCa® features an implant surface treatment that improves the hydrophilicity of the implant.
The purpose of this 510(k) is to expand the product offering and include a new dental implant platform (Interna Narrow), and compatible abutments, together with additional implants for the already existing Universal Plus platform. Interna Narrow implants will be offered with implant diameters ranging from 3.3 to 4.75 mm and lengths of 5.5 to 15 mm. Additionally, new implants for the existing Universal Plus platform of diameter 6.0 mm, and lengths of 7.5 to 11.5 mm are included.
This document describes the BTI Interna Narrow/Plus Dental Implant System UnicCa®, which is a dental implant system. The information provided is for a 510(k) premarket notification to the FDA, asserting substantial equivalence to previously cleared predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state "acceptance criteria" in the typical sense of numerical thresholds for outcomes. Instead, it demonstrates substantial equivalence to predicate devices through various non-clinical tests and comparisons. The "performance" is largely reported as demonstrating "substantial equivalence" or being "identical" or "similar" to the predicate devices.
Feature / Test Category | Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance |
---|---|---|
Product Classification | Identical product codes (DZE, NHA) and regulation number (21 CFR 872.3640), Class II. | Identical. |
Indications for Use | Identical to primary predicate device (K151391) for general use in partial/total replacement of teeth, specific uses for 5.5-6.5mm implants (two-stage, delayed loading, straight abutments, fixed restorations), and Tiny® 3.0 implants (maxillary lateral, mandibular lateral/central incisors, immediate loading with good primary stability/appropriate occlusal load). | Identical to primary predicate device. Includes specific indications for 5.5-6.5mm long UnicCa® implants and Tiny® 3.0 UnicCa® implants that match the predicate. |
Implant Design | Similar designs, sterilization, and packaging as identified predicate devices (K022258, K053355, K061383) for use as dental implants. | Similar designs, sterilization and packaging are included in subject and the identified predicate devices, intended for similar use. |
Platform Diameter (mm) | Range of implant platform diameters covered by identified predicate devices (e.g., Interna: 4.1, 5.5; Externa: 3.5, 4.1, 5.5). | Similar. A new platform diameter for the Interna connection (Interna Narrow, 3.5 mm) is included, but the range is covered by predicate devices. Interna Universal Plus is identical to predicate. |
Implant Length (mm) | 5.5 - 18 mm, matching predicate K151391. | Identical (Interna Narrow: 5.5 – 15 mm; Interna Universal Plus: 7.5 – 11.5 mm, which falls within the predicate's 5.5-18 mm range). |
Implant Body Diameter (mm) | 3.0 - 6.0 mm, matching predicates K151391, K022258, K053355. | Identical (Interna Narrow: 3.3 – 4.75 mm; Interna Universal Plus: 6.0 mm, which falls within the predicate's 3.0-6.0 mm range). |
Implant Material | Commercially pure titanium (Grade 4). | Identical. |
Implant Surface Treatment | Calcium surface treatment. | Identical to primary predicate (K151391). |
Restoration | Single and multiple restorations. | Identical. |
Connection Design | Engaging, non-engaging. | Identical. |
Abutment Prosthetic Diameter (mm) | Range of diameters covered by identified predicate devices (K022258, K053355, K061383), e.g., 3.5 - 6.5 mm. | Similar. The subject device's range (3.5 - 5.1 mm) is covered. |
Abutment Gingival Height (mm) | Range of gingival heights covered by identified predicate and reference devices (K022258, K053355, K061383, K130808), e.g., 1 - 4 mm. | Similar. The subject device's range (0.5 - 4 mm) is covered. |
Abutment Angulation | Straight (0°) with divergence allowance supported by bench testing. | Identical. Divergence allowance in subject device is supported by bench testing provided. |
Abutment Material | Commercially pure titanium (Grade 4). | Identical to identified predicate devices. |
Prosthetic Screws Material | Ti6Al4V, or other materials in predicate (e.g., gold/palladium alloy in K022258). | Identical to K053355 (Ti6Al4V). |
Packaging, Sterilization, Shelf-Life | Similar packaging and sterilization processes, sterile by gamma irradiation or non-sterile for end-user sterilization. | Similar packaging as the identified predicate devices. Healing abutments/Transepithelial abutments: sterile by gamma irradiation, thermoform tray with peel top lid. Temporary Titanium abutments, Titanium abutments, Healing Caps, Aesthetic Interfaces and Screws: non-sterile, to be sterilized by end user (moist heat), thermosealed bag. |
Biocompatibility | Conformance with ISO 10993-1, with satisfactory results for all biological endpoints relative to a long-term implant. | All biological endpoints relative to a long-term implant as identified in the FDA guidance have been addressed with satisfactory results. |
Corrosion, Surface Hydrophilicity, TOF-SIMS | Leveraged from predicate K151391. Performance should be comparable. | Leveraged from K151391. |
Fatigue Testing | Conformance with ISO 14801 and FDA guidance document, showing substantial equivalence to predicate for intended use. | Test results show that the proposed device is substantially equivalent to the predicate device for its intended use. |
Comparative Area Analyses (Bone Resorption, Bone-to-Implant Contact) | Subject implants and reference device should show substantial equivalence in these aspects. | Results showed that subject device is substantially equivalent. |
Sterilization Validation | Sterility assurance level (SAL) of 10-6, according to ISO 11137-1, ISO 11137-2 and ISO 17665-1. | Validation performed to a SAL of 10-6, according to specified ISO standards. |
Packaging/Sterile Barrier/Shelf-Life Validation | Conformance with ISO 11607-1, ISO 11607-2, ASTM F88, ASTM F1886, ASTM 1929 and ASTM F2096. | Validation performed according to specified ISO and ASTM standards. |
Transport Validation | Conformance with ISTA 3A. | Validation performed according to ISTA 3A. |
Endotoxin Testing | Conformance with ANSI/AAMI ST72. | Testing performed according to ANSI/AAMI ST72. |
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: The document mentions "Comparative area analyses for full external area and for bone resorption of up to 3 mm have been performed for subject implants and reference device." However, specific numerical sample sizes for this or other non-clinical tests (e.g., fatigue testing, biocompatibility) are not provided.
- Data Provenance: The document does not specify the country of origin for the data or whether it is retrospective or prospective. It refers to manufacturing in Spain (B.T.I. Biotechnology Institute, SL Jose Rivero Qualified Person Leonardo Da Vinci 14, Parque Tecnologico de Alava Minano, Alava 01510 SPAIN). The tests are non-clinical, implying laboratory or bench testing, not human patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable and not provided as the submission relies on non-clinical, bench testing for substantial equivalence, not clinical study data requiring expert-established ground truth.
4. Adjudication method for the test set:
This information is not applicable and not provided as the submission relies on non-clinical, bench testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The device is a physical dental implant system, not an AI-assisted diagnostic or therapeutic tool for which an MRMC study would be relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This is not applicable as the device is a physical dental implant system, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this submission is based on established engineering and biological performance standards (e.g., ISO, ASTM standards, FDA guidance documents) and comparative data to already cleared predicate physical devices. For example:
- Biocompatibility: Conformance with ISO 10993-1.
- Fatigue Testing: Conformance with ISO 14801 and FDA guidance.
- Sterilization Validation: Conformance with ISO 11137 series.
- Physical measurements and material specifications are compared to those of the predicate devices.
8. The sample size for the training set:
This information is not applicable as there is no mention of a "training set" in the context of this physical device submission, which does not involve machine learning or AI models.
9. How the ground truth for the training set was established:
This information is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(251 days)
N1™ TiUltra™ TCC Implant system is indicated for use in the maxilla or mandible for anchoring or supporting prosthetic teeth, in order to restore patient esthetics and chewing function. N1™ TiUltra™ TCC Implant system is indicated for single or multiple unit restorations in splinted or non-splinted applications using a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, given that sufficient primary stability and appropriate occlusal loading for the selected technique has been achieved.
N1 TiUltra TCC Implant system is composed of implant site preparation tools, abutments, and abutment screws.
The provided text describes the regulatory clearance of a dental implant system (N1™ TiUltra™ TCC Implant system) and outlines the testing conducted to demonstrate its substantial equivalence to predicate devices. However, it does not contain information typically found in a clinical study report that would detail acceptance criteria and a study proving the device meets those criteria in the context of diagnostic or screening performance (e.g., sensitivity, specificity, accuracy).
Instead, the document focuses on demonstrating substantial equivalence to legally marketed predicate devices through a comparison of technological characteristics, intended use, indications for use, and comprehensive non-clinical and clinical testing. The concept of "acceptance criteria" here refers to demonstrating that the new device is as safe and effective as the predicate, rather than meeting specific performance metrics for a diagnostic task.
Therefore, I will interpret "acceptance criteria" as the criteria for establishing substantial equivalence and "reported device performance" as the outcomes of the non-clinical and clinical studies conducted to support that equivalence.
Here's a breakdown of the information based on the provided text, addressing your specific questions, with explanations for what is not present:
1. Table of acceptance criteria and the reported device performance
Since this is a submission for substantial equivalence, the "acceptance criteria" are demonstrating similarity to predicate devices and acceptable performance in various non-clinical and clinical tests.
Acceptance Criteria (for Substantial Equivalence) | Reported Device Performance (N1™ TiUltra™ TCC Implant system) |
---|---|
Substantial Equivalence in Intended Use and Indications for Use. | N1™ TiUltra™ TCC Implant system shares the same or similar intended use and indications for use as predicate devices, with differences adequately addressed. |
Sterilization Validation (SAL 10⁻⁶) | Successfully demonstrated sterility assurance level of 10⁻⁶ via gamma sterilization, confirming safety for use. |
Endotoxin Levels (meeting USP , , ANSI/AAMI ST72) | Testing confirmed endotoxin levels are within acceptable limits, demonstrating pyrogen-free status. |
End-User Cleaning & Sterilization Validation (for reusable components) | Validation performed per ISO 17665-1, AAMI TIR12, AAMI/ANSI ST79, ANSI/AAMI ST77, ISO 17664, ensuring safe reprocessing for reusable components. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Packaging Performance (maintaining sterility and device integrity) | Met standards (ISO 11607-1, ASTM D4169, D4332, F1886/F1886M, F2096, F1980), demonstrating packaging efficacy in maintaining sterility and integrity. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Biocompatibility (ISO 10993-1, -5, -12, -18) | All materials passed biocompatibility tests (cytotoxicity, chemical characterization), confirming safety for patient contact. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Mechanical Resistance (ISO 14801, FDA Guidance) | Demonstrated sufficient fatigue resistance and mechanical stability in dynamic loading tests, comparable to predicate devices. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Insertion Parameters (Heat Generation, Torque) | Comparative testing demonstrated equivalence in heat generation and insertion torque to the Primary Predicate, indicating similar surgical experience and bone response. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Osseointegration (Animal Study) | Animal study showed representative osseointegration behavior comparable to the predicate (TiUltra Implants and Xeal Abutments) at 13 ± 2 weeks, supporting the new drilling technique. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Mesostructure Design & Manufacturing (Software & E2E Validation) | Software verification for abutment design library ensured adherence to design limitations; validation confirmed locked specifications, and end-to-end (E2E) validation demonstrated proper manufacturing workflow. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
MR Conditional Labeling (Referencing K212125) | Met acceptable outcomes in MR conditional testing for safe use in specific MR environments, leveraging prior clearance. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Long-term Clinical Performance (MBLC, Survival, Safety, Tissue Health) | Primary Endpoint (MBLC): Non-inferiority to historical data for the predicate device (NobelActive implant system) was met with statistical significance (p |
Ask a specific question about this device
(79 days)
NobelProcera Zirconia Implant Bridge (previously cleared per K202452) The NobelProcera® Zirconia Implant Bridge are indicated for use as a bridge anatomically shaped and/or framework in the treatment of partially edentulous jaws for the purpose of restoring chewing function.
TiUltra Implants and Xeal Abutments (previously cleared per K202344) NobelActive TiUltra NobelActive TiUltra implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. Nobel Active Tilltra implants are indicated for single or multiple unit restorations in splinted applications. This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. NobelActive TiUltra 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central incisor in the mandible. Nobel Active TiUltra 3.0 implants are indicated for single-unit restorations only. NobelReplace CC TiUltra NobelReplace CC TiUltra implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The NobelReplace CC TiUltra implants are indicated for single or multiple unit restorations. The NobelReplace CC Tilltra implants can be used in splinted or non-splications. The NobelReplace CC TiUltra implant may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied. NobelParallel CC TiUltra NobelParallel CC TiUltra implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function. NobelParallel CC TiUltra implants are indicated for single or multiple restorations in splinted applications. This can be achieved by a 2-stage or 1-stage surgical techniques in combination with immediate, early of delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. Implants with
Not Found
This document is a 510(k) premarket notification decision letter from the FDA to Nobel Biocare AG regarding their Dental Implant Systems Portfolio - MR Conditional. It explicitly states that the letter covers indications for use and general controls, but does not contain information about acceptance criteria or performance studies for the device itself.
Therefore, I cannot provide the requested information for the following reasons:
- Acceptance Criteria and Performance Data: The document is a regulatory clearance letter, not a clinical study report. It does not contain acceptance criteria for device performance, nor does it present any data from studies proving the device meets particular criteria. The letter confirms substantial equivalence to legally marketed predicate devices, which means the FDA has determined the device is as safe and effective as a previously cleared device, not that specific performance metrics were tested and met in a new study.
- Study Details (Sample size, data provenance, experts, adjudication, MRMC, Standalone, Ground Truth, Training Set): Since no performance study data is included in this FDA 510(k) clearance letter, none of these details can be extracted. The document refers to various previously cleared predicate devices (e.g., K202452, K202344, K181869), but it doesn't describe the studies that led to their clearance.
In summary, the provided text does not contain the information necessary to describe acceptance criteria or a study proving the device meets those criteria.
Ask a specific question about this device
Page 1 of 1