Search Results
Found 4 results
510(k) Data Aggregation
(267 days)
The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.
The Millipede 088 Access Catheter consists of the catheter, a rotating hemostasis valve (RHV) and a valve crossing tool. The catheter, RHV and valve crossing tool are provided sterile. They are sterilized by ethylene oxide (EO).
The Millipede 088 Access Catheter is a single lumen, coil-reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The valve crossing tool is used to open the valve of the access sheath and to facilitate insertion of the Millipede 088 Access Catheter through the access sheath without damage. The RHV is assembled onto the hub of the Millipede 088 Access Catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of other devices through the Millipede 088 Access Catheter.
The provided document is a 510(k) summary for the Millipede 088 Access Catheter. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing detailed acceptance criteria and a standalone clinical study to prove the device meets those criteria.
Therefore, the document does not contain the information requested in points 2, 3, 4, 5, 8, and 9, and only partially addresses points 1, 6, and 7.
Here's a breakdown of what can be extracted:
1. Acceptance Criteria and Reported Device Performance
The document lists various performance tests and their conclusions, indicating that the device met "established specifications" or was "suitable for its intended use." However, the specific quantitative acceptance criteria for each test are not provided. The reported device performance is qualitative rather than quantitative in most cases.
Test | Test Method | Reported Device Performance (Conclusions) |
---|---|---|
Dimensional Inspection | Device dimensions were measured to confirm conformance to the specifications. | The device met established specifications. |
Tip Stiffness | Test specimens were tested for tip flexibility and compared to predicate and reference devices. | The device met established specifications. |
Visual Inspection | Device surface characteristics were assessed to confirm freedom from defects. | The device surface characteristics are suitable for its intended use. |
Simulated Use Testing | Deliverability and compatibility with accessory devices were evaluated in a neurovascular model. | The device performs as intended under simulated use conditions. |
Hydrophilic Coating Integrity | The integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles. | The hydrophilic coating integrity is suitable for its intended use. |
Particulate Recovery | Quantify the particulate size and count generated by simulated use of the test article. | The particulate size and count were similar to control devices. |
Tensile Strength | The tensile strength was evaluated for the bonds between sections of the catheter. | The device met established specifications. |
Air Leakage | Tested per ISO 10555-1:2013 Annex D. | The device integrity is suitable for its intended use. |
Liquid Leakage | Tested per ISO 10555-1:2013 Annex C. | The device integrity is suitable for its intended use. |
Static Burst | Tested per ISO 10555-1:2013 Annex F. | The device integrity is suitable for its intended use. |
Luer Integrity | The luers were evaluated for compliance to relevant standards. | The luers on the device are suitable for their intended use. |
Kink Resistance | Test specimen segments were formed into a defined bend diameter to evaluate kink resistance. | The device met established specifications. |
Torque Strength | The test specimens were rotated in a simulated use model to evaluate integrity after rotation. | The device met established specifications. |
Flow Rate Characterization | The flow rate of saline and a contrast-saline solution was characterized when injected through the catheter. | The flow rate was characterized. |
Radiopacity | Radiopacity of the device was evaluated in an animal model under fluoroscopy. | The radiopacity of the Millipede 088 Access Catheter was similar to a control device. |
Biocompatibility (various tests) | ISO 10993-1, 10993-4 (Hemocompatibility), 10993-5 (Cytotoxicity), 10993-10 (Irritation, Sensitization), 10993-11 (Systemic Toxicity, Pyrogenicity) | Non-cytotoxic, no sensitization response, met intracutaneous reactivity, met acute systemic injection, non-pyrogenic, not a complement activator, not an intrinsic coagulation pathway activator, non-hemolytic, similar thromboresistance to controls. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample sizes for the individual performance tests beyond implicitly suggesting multiple samples were tested (e.g., "test specimens," "test articles"). For the Animal Testing, it states "two studies in a porcine model."
The data provenance for the in vitro and animal studies isn't explicitly stated beyond "Good Laboratory Practices" for the animal studies, which is a standard of conduct rather than a geographic origin. No human data is presented.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided as the submission relies on objective performance testing, biocompatibility studies, and animal studies rather than expert-derived ground truth from human data for this type of device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The evaluation methods described are objective performance tests and animal studies, not human data requiring expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an access catheter, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI assistance is not relevant to its evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the performance testing described (mechanical, dimensional, coating integrity, etc.) and the animal studies represent "standalone" evaluations of the device's physical and functional properties without human interpretation of data in a clinical context.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the performance and biocompatibility tests, the "ground truth" is defined by the established specifications, relevant ISO standards, and comparison to control devices or predicate devices. For example:
- "The device met established specifications."
- "The test article is non-cytotoxic."
- "The particulate size and count were similar to control devices."
- "The radiopacity of the Millipede 088 Access Catheter was similar to a control device."
- For animal testing, "Usability, radiopacity, thromboresistance, and vessel injury were assessed," and "The results for the subject device were comparable to a control device."
8. The sample size for the training set
Not applicable. This device is a medical catheter and does not involve a "training set" in the context of machine learning or AI. Its performance is evaluated through engineering and biological testing.
9. How the ground truth for the training set was established
Not applicable (as above).
Ask a specific question about this device
(316 days)
The Minjie Catheter System is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
The Minjie Catheter System consists of the Minjie Catheter, Introducer Peel Away Tool and packaging hoop with Elbow Flush Luer. The Minjie Catheter is a single lumen. flexible, variable stiffness composite catheter with a Nitinol structure. A radiopaque marker band on the distal tip of the catheter is used for visualization under fluoroscopy. The distal section of the catheter is coated with a hydrophilic coating, to reduce friction during intravascular use. The Minjie Catheter System dimensions are included in the individual device label. The devices are supplied sterile and are intended for single use only.
The provided FDA 510(k) summary for the Minjie Catheter System does not contain acceptance criteria or reported device performance in the format requested for AI/algorithm-based devices. This document describes a medical device (a catheter system) and its performance through non-clinical (bench and animal) testing to demonstrate substantial equivalence to a predicate device, rather than the performance of an AI algorithm.
Therefore, many of the requested fields related to AI algorithm evaluation (sample sizes for test/training sets, data provenance, number/qualification of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) are not applicable to the information provided in this document.
However, I can extract information related to the device's non-clinical testing and its conclusions, which serve as the "proof" that the device meets some form of acceptance criteria for medical devices of this type.
Here's a summary based on the provided document, adapting the requested structure where possible for a non-AI device:
Acceptance Criteria and Study for the Minjie Catheter System
The Minjie Catheter System is a physical medical device (catheter) and its performance is evaluated against engineering specifications, biocompatibility standards, and functional capabilities rather than AI algorithm metrics. The studies performed are non-clinical, including bench testing and an animal study, to demonstrate substantial equivalence and safety/effectiveness for its intended use.
1. Table of Acceptance Criteria and Reported Device Performance
Note: The document states that "All devices met acceptance criteria" for each bench test, but it does not explicitly define the numeric acceptance criteria themselves. The reported performance is therefore a qualitative statement of compliance.
Test Name | Test Method Summary | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|---|
Catheter Visual Inspection | Inspection for dents, kinks, cracks, damage, or anomalies. | No damage or anomalies affecting function. | All devices met acceptance criteria. |
Coating Integrity | Inspection under magnification for coating defects before and after simulated use and particulate testing. | No coating defects. | All devices met acceptance criteria. |
Particulate Testing | Evaluation for particulate generation under simulated use in a tortuous anatomical model. | Meets particulate generation limits. | All devices met acceptance criteria. |
Simulated Use | Repeated navigation through a tortuous benchtop model to assess compatibility with accessories, stability, and navigation to M1 and M2 segments of MCA. | Successful navigation, compatibility, and stability. | All devices met acceptance criteria. |
Coating Frictional Forces / Durability | Evaluation of frictional forces and durability via repeated navigation through simulated use test model. | Acceptable frictional forces and durability. | All devices met acceptance criteria. |
Distal Tip Buckling | Testing distal tip under compressive loads at 5mm, 10mm, and 20mm to evaluate stiffness. | Meets specified stiffness requirements. | All devices met acceptance criteria. |
Kink Resistance | Wrapping device around mandrels of clinically relevant diameters and inspecting for kinks. | No kinks. | All devices met acceptance criteria. |
Liquid Leakage under Pressure | Tested per ISO 10555-1, Annex C. | No liquid leakage. | All devices met acceptance criteria. |
Hub Air Aspiration Leak | Tested per ISO 10555-1, Annex D. | No hub air aspiration leak. | All devices met acceptance criteria. |
Torque to Failure | Tested in a simulated use model to determine number of rotations to failure. | Withstands specified torque without failure. | All devices met acceptance criteria. |
Manual Injection / Peak Pressure | Tested with manual syringe injection of worst-case contrast media after simulated use. | Withstands generated pressures. | All devices met acceptance criteria. |
Static / Dynamic Burst | Tested under full-length static conditions to burst per ISO 10555-1, Annex F. | Meets burst pressure requirements. | All devices met acceptance criteria. |
Luer Hub Compatibility | Tested per ISO 80369-7 and ISO 80369-20. | Compatible with luer hubs. | All devices met acceptance criteria. |
Dimensional Inspection | Measurement of usable length, proximal/distal inner/outer diameters. | Meets specified dimensional tolerances. | All devices met acceptance criteria. |
Lumen Patency | Mandrel of required size must pass from proximal hub to distal tip. | Mandrel passes freely. | All devices met acceptance criteria. |
Shaft Peak Tensile Force | Tested to failure at distal tip section and each joint per ISO 10555-1, Annex B. | Withstands specified tensile forces. | All devices met acceptance criteria. |
Hub Peak Tensile Force | Tested to failure per ISO 10555-1, Annex B. | Withstands specified tensile forces. | All devices met acceptance criteria. |
Physician Usability Testing | Navigated through a tortuous benchtop model to assess compatibility, stability, injection ability, and navigation to M1/M2 segments of MCA. | Usable and effective for intended functions. | All devices met acceptance criteria. |
Biocompatibility Test Name | Test Results Conclusion |
---|---|
Cytotoxicity | Non-cytotoxic |
Sensitization | Non-sensitizer |
Intracutaneous Irritation | Non-irritant |
Acute Systemic Toxicity | No acute systemic toxicity |
Material Mediated Pyrogen | Non-pyrogenic |
Hemocompatibility (Hemolysis) | Non-hemolytic |
Hemocompatibility (Complement Activation) | Non-activator of the complement system |
Hemocompatibility (PTT) | Non-activator of coagulation |
Hemocompatibility (Platelet & Leukocyte Counts) | Non-activator of platelet and leukocyte |
Sterilization/Shelf-Life Test | Acceptance Criteria (Implied) | Reported Performance |
---|---|---|
EO Sterilization Validation | SAL of 10^-6 per ISO 11135:2014. | Achieved SAL of 10^-6. |
Endotoxin Testing | Meets FDA guidance for pyrogen and endotoxins. | Compliant. |
Packaging Integrity | Maintains sterility and integrity for 24 months. | Met all acceptance criteria. |
Product Stability (Aging) | Device remains functional for 24 months. | Met all acceptance criteria. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Bench Testing: The document does not specify the exact sample size for each bench test (e.g., "All devices met acceptance criteria" implies multiple devices were tested). There is no explicit mention of data provenance (e.g., country of origin). This is non-clinical, prospective testing.
- Biocompatibility Testing: Not specified for individual tests.
- Sterilization and Shelf-Life: Not specified, but validation studies would involve multiple units.
- Animal Testing: The document states "An animal study was performed," but does not specify the number of animals used. Data provenance is not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as this is a physical device and not an AI algorithm requiring expert ground truth for image interpretation or diagnosis. Bench testing results are typically objective measurements against engineering specifications. Animal study evaluations would be performed by qualified veterinary and pathology staff, but specific numbers and qualifications are not provided.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable for physical device testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic or therapeutic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For this physical device:
- Bench Testing: Engineering specifications, industry standards (e.g., ISO), and defined performance parameters serve as the "ground truth" or acceptance criteria.
- Biocompatibility Testing: Established biological response parameters outlined in ISO 10993 series and FDA guidance.
- Animal Testing: Angiographic and histological evaluations, which are considered objective medical findings.
8. The sample size for the training set
Not applicable. This is not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established
Not applicable. This is not an AI algorithm.
Ask a specific question about this device
(236 days)
The SOFIA® EX Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA® EX Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA® EX Catheter is not intended for use in coronary arteries.
The SOFIA® EX Catheter is a single-lumen, flexible catheter equipped with the coil and the braid reinforcement. The distal segment is designed to facilitate vessel selection with 55-65cm of distal shaft hydrophilic coating for navigation through the vasculatures. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. The device is provided sterile and for single use. The catheter is placed in a dispenser tube (HDPE) and is placed on a packaging card (polyethylene) that is provided in a sterile barrier tyvek pouch and placed in a carton box.
The provided document is a 510(k) premarket notification for the SOFIA® EX Intracranial Support Catheter. It describes the device, its intended use, comparison to predicate devices, and performance data used to demonstrate substantial equivalence.
Here's the breakdown of the acceptance criteria and the study proving the device meets them, as extracted from the document:
1. A table of acceptance criteria and the reported device performance:
The document presents the performance data in a table format under section "VII. Performance Data" (pages 7-8). Each "Test" listed implicitly defines an acceptance criterion (e.g., "Device met acceptance criteria"). Because the document is a 510(k) summary, the specific quantitative acceptance criteria values are not explicitly stated, but the results confirm that the "Device met acceptance criteria" for each test.
Test Name | Test Method Summary | Reported Device Performance |
---|---|---|
Dimensional Inspection | The usable length, proximal and distal outer diameters, distal length and inner diameters were measured and recorded. | Device met acceptance criteria for length, inner and outer diameters. The device size (5F) comparable to the 5F predicate and reference devices. The inner diameter is larger than the predicate device, while the outer diameter is larger, but still compatible with tested 6F guiding sheaths. |
Catheter Tip Stability | Simulated use of the delivery of a braided device is performed in a tortuous anatomical benchtop model and the movement of the subject device is measured and recorded. | Device met acceptance criteria for tip stability. The device is able to support the delivery of braided devices and stent-retrievers without losing distal tip position. The stability of the distal tip was better (less movement) than the predicate device. |
Simulated Use and Physician Simulated Use | The device is put through simulated use. The device is navigated through a tortuous benchtop model to assess preparation, introduction, tracking, and support of the device. | Device met acceptance criteria. |
Dynamic Burst Testing | Device hub is connected to a pressure control machine and was tested under pressures experienced during worst-case dynamic injections. | Device met acceptance criteria and was able to withstand pressures experienced during worst-case dynamic injections. |
Liquid Leakage | Device was tested per ISO 10555-1, Annex C liquid leakage testing. Device is connected at hub and is pressurized with fluid and maintains the pressure for a specified duration of time. | Device met acceptance criteria. |
Liquid Leakage at Rated Burst Pressure | Device was tested per ISO 10555-2 Annex A, liquid leakage testing. Device is connected at hub and is pressurized with fluid and maintains rated burst pressure for a specified duration of time. | Device met acceptance criteria and does not leak fluids at rated burst pressure. |
Air Leakage | Device was tested per ISO 594-2. Device is connected at hub and subjected to negative pressure and any air leaking into the device is recorded. | Device met acceptance criteria. Device does not allow air to leak into the device when subjected to negative pressure. |
Static Burst | Device was connected at hub and tested under full-length static conditions to burst. | Device met acceptance criteria. All devices burst above the rated burst pressure and had better results than the predicate device. |
Tensile Strength | Device was tested per ISO 11070. The device is tensile tested to failure and the force at break is measured and recorded. | Device met acceptance criteria. |
Tip Buckling | Distal tip buckling force under compressive load was evaluated for stiffness. | Device met acceptance criteria. Device has a softer distal tip than the reference device. |
Torque Response | Device was tested for full-length torque response. The device is tracked through a tortuous benchtop model and the proximal hub is turned, the distal tip torque response is measured and recorded. | Device met acceptance criteria. Device has better torque response than the reference device. |
Radio-detectability | Device is put under fluoroscope to assess visibility. | Device met acceptance criteria. Device is visible under fluoroscopy. |
Coating Lubricity and Durability | Device coating was evaluated for frictional force and durability. | Device met acceptance criteria. Average friction is comparable to predicate device. |
Particulate Testing | Device was evaluated for particulate generation under simulated use in a representative tortuous anatomical model per USP. | Number of particulates generated met acceptance criteria and is within the limits per USP and is comparable to the predicate and reference devices. |
Kink Resistance | Device is evaluated for kink resistance by subjecting the device to bending experienced in tortuous anatomy. | Device met acceptance criteria. Results matched results of the predicate device. |
Corrosion Resistance | Device is tested per ISO 10555-1, Annex A and ISO 11070, Annex B to evaluate corrosion resistance. | Device met acceptance criteria. Device is resistant to corrosion. |
Cytotoxicity (ISO Medium Eluate Method) | 1x CMEM Cell Growth Medium (MEM supplemented with 10% fetal bovine serum extract) L929 Mouse Fibroblast Cell Line (Extracted at 37°C/24 hrs, 6.0 cm²/mL) | Non-cytotoxic. The test article is considered non-cytotoxic to cells. |
Sensitization (ISO Kligman Maximization Test) | Normal Saline and Vegetable (Cottonseed) Oil Extracts Hartley Guinea Pigs (Extracted at 50°C/72 hrs, 6.0 cm²/mL) | Non-sensitizing. The test article did not elicit a sensitization response. |
Irritation/Intracutaneous Toxicity | Normal Saline and Vegetable (Cottonseed) Oil Extracts (Extracted at 50°C/72 hrs, 6.0 cm²/mL) | Non-irritant. No evidence of irritation. |
Systemic Toxicity (ISO Systemic Injection Test) | Normal Saline and Vegetable (Cottonseed) Oil Extracts Albino Swiss Mice (Extracted at 50°C/72 hrs, 6.0 cm²/mL) | Non-cytotoxic (sic, likely meant non-toxic per the test type). No weight loss, mortality, or evidence of systemic toxicity from the extract exposure to the mice. |
Systemic Toxicity (ISO Rabbit Pyrogen Test) | Normal Saline New Zealand White Rabbits (2 Male and 2 Female - non-pregnant and nulliparous) (Extracted at 50°C/72 hrs, 6.0 cm²/mL) | Non-pyrogenic. All individual rabbits for both the test article and negative control showed a total rise in temperature of |
Ask a specific question about this device
(30 days)
The Reverse™M Micro Catheter is intended for use in neuro, peripheral, and coronary vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils.
The Reverse™ Micro Catheter is a single lumen catheter designed to be introduced over a steerable guide wire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate vessel navigation. The micro catheter incorporates a PTFE liner to facilitate movement or introduction of devices passed through its lumen. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter body also employs a hydrophilic coating to enhance lubricity and promote deliverability. The Reverse™ Micro Catheter incorporates a standard luer fitting to facilitate the attachment of ancillary devices. The catheter and included accessories (steam shaping mandrel and peel away introducer sheath) are provided sterile, nonpyrogenic, and are intended for single use only. To accommodate physician preference and anatomical variations, the Reverse™ Micro Catheter is available in two models, Reverse 021, with a 0.021" inner diameter (ID) and, Reverse 027, with a 0.027" ID.
The document you provided describes a medical device, the Reverse™ Micro Catheter, and its non-clinical performance evaluation for substantial equivalence to a predicate device. It does not describe a study involving an AI model or a specific algorithm. Therefore, I cannot provide information about acceptance criteria, device performance, sample sizes, ground truth, expert involvement, or comparative effectiveness studies in the context of AI.
The document focuses on demonstrating that the modified Reverse™ Micro Catheter is substantially equivalent to a previously cleared device (Reverse™ Medical Microcatheter, K130858) based on non-clinical performance data.
Here's a breakdown of what the document does provide, structured to address the spirit of your request where possible, but re-contextualized for a medical device rather than an AI study:
1. A table of acceptance criteria and the reported device performance
The document states that "All testing passed the acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria for each test. The performance results are reported as successful completion of tests without new safety or effectiveness issues.
Test Category | Specific Tests Evaluated | Reported Performance | Acceptance Criteria (Not explicitly quantified in document) |
---|---|---|---|
Bench Testing | - Surface inspection |
- Dimensional requirements
- Lubricity/friction
- Particulates
- Coating characterization
- Simulated use device compatibility
- Simulated use navigation and delivery
- Shelf-life verification (1 year accelerated aging) | "All testing passed the acceptance criteria."
"Comprehensive verification and validation activities were successfully completed; raising no new issues of safety or effectiveness." | Implicitly, meeting specifications and predicate equivalence. |
| Biological Safety | - MEM Elution Cytotoxicity - Guinea Pig Maximization Sensitization
- Intracutaneous Reactivity
- Acute System Injection
- Materials Mediated Rabbit Pyrogen
- Hemolysis Direct Contact
- Hemolysis Indirect Extract Method
- Complement Activation
- In vivo Thrombogenicity
- Genotoxicity (Ames Assay, in vitro Mouse Lymphoma, in vivo Mouse Micronucleus)
- USP Physicochemical | "Testing demonstrated that the Reverse™ Micro Catheter finished device, accessories, and packaging materials have no residual risk of biological hazards; and are therefore considered biocompatible for their intended use." | Compliance with ISO 10993-1 and related standards. |
| Microbial Assessments | - Bioburden - Endotoxin
- Sterility assurance | "Sterilization has been validated and is controlled... to provide a minimum SAL of 10-6." | Minimum SAL of 10-6; compliance with ISO 11135-1. |
| Packaging Validation | - Visual Inspection - Bubble Leak
- Seal Strength | "To evaluate integrity of the packaging configuration." (Implied successful evaluation) | Implicitly, maintaining package integrity after sterilization, environmental conditioning, and simulated shipping. |
2. Sample sized used for the test set and the data provenance
The document does not specify sample sizes for individual bench tests, biological safety tests, or packaging validation. It refers broadly to "comprehensive verification and validation activities." The data provenance is internal testing by the manufacturer, Medtronic Neurovascular, for a new medical device. This is prospective testing of the device being submitted for clearance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This concept is not applicable to this type of device submission. There isn't "ground truth" established by experts in the context of an AI algorithm's performance. Instead, the "truth" is determined by established engineering specifications, biocompatibility standards, and functional performance requirements for a physical medical device.
4. Adjudication method for the test set
Not applicable. There's no human-in-the-loop performance or diagnostic adjudication involved for this physical device's non-clinical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable, as this is not an AI device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable, as this is not an AI device. The testing described is "standalone" in the sense that it's the device itself being tested, not an algorithm.
7. The type of ground truth used
For this medical device, the "ground truth" for the non-clinical tests would be:
- Engineering Specifications: Conformance to pre-defined dimensional, material, and performance requirements (e.g., inner diameter, lubricity, navigation).
- International Standards: Adherence to established standards for biocompatibility (ISO 10993-1), sterilization (ISO 11135-1), and packaging integrity.
- Predicate Device Performance: Demonstrating that the subject device performs "at least as safe and effective as the predicate device."
8. The sample size for the training set
Not applicable, as this is not an AI device.
9. How the ground truth for the training set was established
Not applicable, as this is not an AI device.
Ask a specific question about this device
Page 1 of 1