(89 days)
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
The AXS Catalyst 7 Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in the removal of thrombus from the neuro vasculature using continuous aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostasis Valve (RHV), Tuohy Borst Valve with Sideport, and two peel-away introducer sheaths. The RHV and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst.
When used as part of the aspiration system, the AXS Catalyst Distal Access Catheter requires a minimum vacuum pressure of -68 kPa [-20.08 in Hg] from an external device (Medela Dominant Flex Pump). Bench and Animal testing utilizing the Medela Dominant Flex Pump and Suction Jar, AXS Universal Aspiration Tubing, and AXS Universal Liner Set were conducted to successfully demonstrate aspiration.
The provided document is a 510(k) summary for the AXS Universal Aspiration System (AXS Catalyst 7 Distal Access Catheter). It details the device's characteristics, comparison to predicate devices, and a summary of performance data (bench, animal, and clinical) to demonstrate substantial equivalence.
Based on the document, here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document presents its "acceptance criteria" implicitly through the "Conclusions" column in Table 2 (Performance Data - Bench Testing), Table 3 (Performance Data - Animal Testing), and Table 4 (Biocompatibility Studies). The acceptance criteria are "meets acceptance criteria," "acceptable," "equivalent," or "PASS."
| Test Type | Acceptance Criteria (Implicit from "Conclusions") | Reported Device Performance ("Conclusions" column) |
|---|---|---|
| Design Verification Bench Testing | ||
| Dimensional Verification | Meets acceptance criteria | Dimensional verification meets acceptance criteria. |
| Tip Configuration | Meets acceptance criteria | Tip configuration meets acceptance criteria. |
| Surface Integrity | Meets acceptance criteria | Surface integrity meets acceptance criteria. |
| Tip Buckling | Meets acceptance criteria | Tip buckling meets acceptance criteria. |
| Catheter lubricity and durability | Meets acceptance criteria | Coating lubricity and durability meets acceptance criteria. |
| Particulate characterization | Acceptable | Particulate generation was acceptable. |
| Coating integrity | Acceptable | Coating integrity was acceptable. |
| Trackability | Meets acceptance criteria | Track advance force meets acceptance criteria. |
| Tensile Strength | Meets acceptance criteria | Tensile strength meets acceptance criteria. |
| Liquid Leak Resistance | Meets acceptance criteria | Liquid leak resistance of catheter meets acceptance criteria. |
| Air Leak Resistance | Meets acceptance criteria | Air leak resistance of catheter meets acceptance criteria. |
| Catheter Torsional Bond Strength | Meets acceptance criteria | Catheter torsional bond strength meets acceptance criteria. |
| Flexural Fatigue | Meets acceptance criteria | Flexural fatigue meets acceptance criteria. |
| Catheter Kink Radius | Meets acceptance criteria | Catheter kink radius meets acceptance criteria. |
| Catheter Tip and Lumen Integrity (Direct Aspiration) | Meets acceptance criteria | Catheter tip and lumen integrity during direct aspiration meets acceptance criteria. |
| Chemical Compatibility | Meets acceptance criteria | Chemical compatibility meets acceptance criteria. |
| Hub Gauging | Meets acceptance criteria | Hub gauging meets acceptance criteria. |
| Aspiration Flow Rate Bench Testing | ||
| Aspiration Flow Rate | Meets acceptance criteria | All test samples meet acceptance criteria. |
| Design Validation Bench Testing | ||
| In-vitro Usability Study | Meets acceptance criteria | All test samples meet acceptance criteria. |
| Animal Study | Equivalent to predicate in safety | Subject Device is equivalent to the Additional Predicate Device in safety. |
| Biocompatibility Studies | PASS | MEM Elution Cytotoxicity: PASS (No cytotoxicity or cell lysis) |
| Hemolysis Extract/Direct Contact Method: PASS (Non-hemolytic) | ||
| USP Physiochemical <661>: PASS | ||
| Chemistry (Heptane) Analysis: PASS (Below level of detection) | ||
| FTIR (ISO 10993-18): PASS | ||
| Natural Rubber Latex ELISA Inhibition Assay for Antigenic Protein ASTM D6499-12: PASS (Below level of detection) |
2. Sample sizes used for the test set and the data provenance
The document does not explicitly state the numerical sample sizes (e.g., number of catheters tested) for each bench test. It mentions qualitative assessments like "Visually inspect" or "Prepare sample for test." For the animal study, it mentions "porcine test subjects" but does not give the exact number.
- Provenance: The studies were conducted by Stryker Neurovascular ("Submitter Name, Address, and Content" section) and likely took place at their facilities or certified labs. The animal study was conducted "in compliance with applicable requirements in the GLP regulation (21 CFR Part 58)," which specifies good laboratory practices, but not geographic provenance. The clinical data was not from a study conducted by the submitter but rather a "review was conducted considering published clinical study articles that featured the Predicate devices." This implies that the clinical data is retrospective from existing literature, and its origin would vary based on those publications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The studies described are primarily bench and animal tests, along with a review of published clinical articles. There's no mention of expert consensus or human readers being used to establish ground truth for a test set in the context of device performance, as one might find for an AI/imaging device. The "In-vitro Usability Study" mentions "multiple physician users" but doesn't specify their number or qualifications, nor is it described as establishing "ground truth" for the device's technical performance, but rather "evaluat[ing] aspiration integrity, ability to restore flow and the durability and kink resistance."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable/not provided. Adjudication methods are typically used in studies involving human readers or expert panels evaluating diagnostic outputs (like images), especially in the context of establishing ground truth where there might be disagreement. The studies described here are primarily engineering (bench) and animal model tests, where outcomes are determined by physical measurements or biological responses, not human interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a "Percutaneous Catheter" (specifically, a distal access catheter for stroke treatment), not an AI-based diagnostic or assistive software. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance would not be relevant to this device's submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical medical instrument, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the bench and animal tests, the "ground truth" is established through physical measurements, established engineering test methods (e.g., EN ISO 10555-1, EN 1707, EN ISO 14971), and biological evaluations (histopathology in the animal study). For the biocompatibility tests, it's based on established ISO standards (EN ISO 10993-1 and sub-parts) and analytical chemistry results. There is no mention of expert consensus or pathology reports serving as "ground truth" in the way it might for a diagnostic imaging AI. The "clinical performance" relies on a review of published clinical outcomes of predicate devices, leveraging their established safety and efficacy rather than generating new outcomes data for the subject device.
8. The sample size for the training set
This is not applicable/not provided. The device is a physical medical device, not a machine learning model. Therefore, there is no "training set."
9. How the ground truth for the training set was established
This is not applicable. As there is no training set for a physical device, there's no ground truth established for one.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 13, 2019
Stryker Neurovascular Germaine Fu. PhD Associate Project Manager, Regulatory Affairs 47900 Bayside Parkway Fremont, California 94538
Re: K183464
Trade/Device Name: AXS Universal Aspiration System (AXS Catalyst 7 Distal Access Catheter) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: December 13, 2018 Received: December 14, 2018
Dear Germaine Fu, PhD:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed
by Xiaolin Zheng
Date: 2019.03.13
18:49:44 -04'00'
Lin Zheng-s
Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K183464
Device Name
AXS Universal Aspiration System (AXS Catalyst 7 Distal Access Catheter)
Indications for Use (Describe)
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K183464
Introduction:
According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for a determination of substantial equivalence.
Submitter Name, Address and Content:
| Submitter: | Stryker Neurovascular47900 Bayside ParkwayFremont, CA 94538-6515(FDA Registration Number: 3008853977) |
|---|---|
| Contact: | Germaine FuAssociate Project Manager, Regulatory AffairsPhone: 510-413-2862Fax: 510-413-2588Email: germaine.fu@stryker.com |
| Date Prepared: | December 13, 2018 |
| Device Name and Classification: | |
| Trade/Proprietary Name: | AXS Universal Aspiration System (AXS Catalyst 7 Distal AccessCatheter) |
| Common Name: | Catheter, Thrombus Retriever |
| Classification Name: | Percutaneous Catheter, 21CFR 870.1250 – Class II |
| Product Code: | NRY |
{4}------------------------------------------------
| Legally MarketedPredicateDevice(s): | Primary Predicate Device | Additional Predicate Device |
|---|---|---|
| AXS Catalyst Distal AccessCatheter(K173841) | Penumbra System ACE 68Reperfusion Catheter(K161640) |
Device Description
The AXS Catalyst 7 Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in the removal of thrombus from the neuro vasculature using continuous aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostasis Valve (RHV), Tuohy Borst Valve with Sideport, and two peel-away introducer sheaths. The RHV and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst.
When used as part of the aspiration system, the AXS Catalyst Distal Access Catheter requires a minimum vacuum pressure of -68 kPa [-20.08 in Hg] from an external device (Medela Dominant Flex Pump). Bench and Animal testing utilizing the Medela Dominant Flex Pump and Suction Jar, AXS Universal Aspiration Tubing, and AXS Universal Liner Set were conducted to successfully demonstrate aspiration.
Indications for Use
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
Technological Characteristics and Product Feature Comparison
Stryker Neurovascular has demonstrated the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system is substantially equivalent to the Primary Predicate device (K173841) and Additional Predicate device (K161640) based on the same or similar materials, similar design, and the same fundamental operating principles. A comparison of the Subject device with the Predicate devices is summarized in Table 1 below.
{5}------------------------------------------------
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | |||
|---|---|---|---|
| Detail | Submission Subject Device | Primary Predicate Device | Additional Predicate Device |
| AXS Catalyst 7 Distal Access Catheter | AXS Catalyst 5 and 6 Distal Access Catheter | Penumbra System ACE 68 Reperfusion Catheter | |
| Manufacturer | Stryker Neurovascular | Stryker Neurovascular | Penumbra, Inc. |
| 510(k) Number | K183464 | K173841 | K161640 |
| Device Trade Name | AXS Catalyst® 7 Distal Access Catheter | AXS Catalyst® 5 and 6 Distal Access Catheter | Penumbra System ACE 68 Reperfusion Catheter |
| Regulation Number | 21 CFR 870.1250 | Same | Same |
| Regulation Name | Percutaneous Catheter | Same | Same |
| Classification | II | Same | Same |
| Product Code | NRY | Same | Same |
| Intended Use/ Indication for Use | The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 | Same | As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, |
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | |||
| Detail | Submission Subject Device | Primary Predicate Device | Additional Predicate Device |
| AXS Catalyst 7 Distal Access Catheter | AXS Catalyst 5 and 6 Distal Access Catheter | Penumbra System ACE 68 Reperfusion Catheter | |
| hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment. | and vertebral arteries) within 8 hours of symptom onset. | ||
| Device Description | The AXS Universal Aspiration System is composed of the following components:• AXS Catalyst Distal Access Catheter• AXS Universal Aspiration Tubing• AXS Universal Liner Set• Medela Dominant Flex PumpThe AXS Universal Aspiration System is designed to remove thrombus from the neurovasculature using continuous aspiration. | Same | The Penumbra System ACE 68 Reperfusion Catheter is a component to the currently available Penumbra System. The ACE 68 Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel in the neurovasculature. ACE 68 is provided sterile, non-pyrogenic, and intended for single use only. |
| Detail | Submission SubjectDevice | Primary PredicateDevice | Additional PredicateDevice |
| AXS Catalyst 7 DistalAccess Catheter | AXS Catalyst 5 and 6Distal Access Catheter | Penumbra SystemACE 68 ReperfusionCatheter | |
| AccessoryDevicesProvided (notin directcontact withpatient) | The AXS CatalystDistal Access Catheterdelivers aspirationfrom the MedelaDominant Flex Pumpdirectly to the site ofthe occlusion toremove the clot. TheAXS Catalyst DistalAccess Catheter is asterile, single lumen,variable stiffnesscatheter. | Same | Peelable sheath,rotating hemostasisvalve, shapingmandrel |
| Materials | |||
| Outer Jacket | Pebax with Nylon,Tecoflex | Same | Vestamid, Pebax,Tecoflex, Pellethane |
| Reinforcement | Nitinol wire andpolymer fiber | Stainless Steel,Nitinol wire, andpolymer fiber | Stainless steel andNitinol wire |
| Strain Relief | Thermoplastic rubber(Polyolefin) | Thermoplastic rubber(Santoprene) | Stainless steel, 304 |
| Inner Layer | PTFE | Same | Same |
| Catheter Hub | Nylon | Same | Proximal HubGrilamid(TR55-LX) |
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | |||
| Detail | Submission SubjectDevice | Primary PredicateDevice | Additional PredicateDevice |
| AXS Catalyst 7 DistalAccess Catheter | AXS Catalyst 5 and 6Distal Access Catheter | Penumbra SystemACE 68 Reperfusion | |
| Catheter | |||
| Marker Band | Platinum/Iridium | Same | Same |
| Adhesive | Cyanoacrylate | Same | Not available |
| Outer JacketCoating | Hydrophilic Coating | Same | SRDX Harmony(proprietary) coating |
| Features | |||
| Labeled Shaft | Distal OD: | Distal OD: | Distal OD: |
| OuterDiameter | 6.2F (0.082 in.) | 5.3F (0.0696 in.) | 0.084 in. |
| Proximal OD: | 5.4F (0.0709 in.) | Proximal OD: | |
| 6.3F (0.0825 in.) | 0.084 in. | ||
| Proximal OD: | |||
| 5.6F (0.0735 in.) | |||
| 6.0F (0.0787 in.) | |||
| Effective | 115cm | CAT 5:115cm, 132cm | 115cm |
| Lengths | 125cm132cm | 120cm | |
| CAT 6: 132cm | 125cm | ||
| 127cm | |||
| 132cm | |||
| Distal ID | 0.068 in. | CAT 5: 0.058 in.CAT 6: 0.060 in. | 0.068 in. |
| Proximal ID | 0.068 in. | CAT 5: 0.058 in.CAT 6: 0.060 in. | 0.068 in. |
| PackagingMaterials | Polyethylene Tubeand HDPEPackaging Card | Same | Polyester/Polyethylene/Tyvek and |
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | |||
| Detail | Submission Subject Device | Primary Predicate Device | Additional Predicate Device |
| andConfiguration | Polyethylene Packaging Card | ||
| Sterilization Method | EO Sterilization | Same | Same |
| How Supplied | Single Use/Sterile | Same | Same |
| Principles of Operation | The AXS Catalyst Distal AccessCatheter is advancedinto the neurovasculature by aphysician trained ininterventionalendovascularprocedures using acompatible sheath orguide catheter, andover anappropriately sizedguide wire. A peelaway sheath isprovided in thepackage to providesupport and facilitatethe introduction ofthe AXS CatalystCatheter tip into thesheath/guidecatheter valve. Oncethe catheter isinserted, the peelaway sheath can beremoved. Underfluoroscopic | Same | Confirm vesseldiameter andselect anappropriate sizePenumbraReperfusionCatheter. Thecatheter tip maybe shaped usingthe steam shapingmandrel provided.Attach therotatinghemostasis valveprovided to thecatheter. Insertthe catheter intothe rotatinghemostasis valveconnected to theproximal hub of aguide catheter. If aguide catheter isnot used, insertthe catheterthrough the valveof the longfemoral sheath |
| AXS Catalyst 7 DistalAccess Catheter | AXS Catalyst 5 and 6Distal Access Catheter | Penumbra SystemACE 68 ReperfusionCatheter | |
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | |||
| Detail | Submission SubjectDeviceAXS Catalyst 7 DistalAccess Catheter | Primary PredicateDeviceAXS Catalyst 5 and 6Distal Access Catheter | Additional PredicateDevicePenumbra SystemACE 68 ReperfusionCatheter |
| catheter can beadvanced throughthe vasculature tothe desired location.The catheter isdesigned to removethrombus from thevasculature usingaspiration providedby an externalsource, the MedelaDominant FlexPump. The aspirationsystem is attached tothe catheter via theRHV (or TuohyBorst), and thevacuum is turned on.The aspiration tubingclamp is opened toapply aspiration.Upon completion ofthe procedure, theAXS Catalyst DistalAccess Catheter, theAXS UniversalAspiration Tubing,and the AXSUniversal Liner Setare disposed of perthe institution'sprocedures. TheMedela DominantFlex Pump is cleaned | sheath. Afterinserting thecatheter, removethe peelablesheath from thevascular sheath,and peel from thecatheter shaft.Usingconventionalcatheterizationtechniques underfluoroscopicguidance, advancethe catheter intothe target vesselover anappropriateneurovascularguidewire.Position thecatheter proximalto the thrombus.Remove theguidewire fromthe catheter. | ||
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | |||
| Detail | Submission Subject Device | Primary Predicate Device | Additional Predicate Device |
| AXS Catalyst 7 DistalAccess Catheter | AXS Catalyst 5 and 6Distal Access Catheter | Penumbra SystemACE 68 ReperfusionCatheter | |
| per the instructionsprovided with thepump. | |||
| Aspiration System Components | |||
| AspirationPump | Medela AG DominantFlex PumpK170329 Adjustable vacuumpressure dial Three selectable airflow rates:40,50,60 L/min Non-Sterile Reusable Cleanable Minimum VacuumPressure foraspiration:-68 kPa [-20.08inHg] | Same | Penumbra MAXPump K161640 Adjustable vacuumpressure dial Non-adjustable airflow rate Non-Sterile Reusable Cleanable Minimum VacuumPressure foraspiration:-68 kPa [-20.08inHg] |
| AspirationTubing | Medela AG AXSUniversal AspirationTubing Class II Exempt EO Sterilized Non-reusable ID: 0.218" Length: 300cm Flow clamp | Same | PenumbraAspiration Tubing K161640 EO Sterilized Non-reusable ID: 0.110" Length: 284.5cm Flow switch |
| AspirationCanister | Medela AG DominantFlex Pump utilizes areusablepolycarbonate suction | Same | Penumbra MAXCanister is a non- |
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | |||
| Detail | Submission SubjectDevice | Primary PredicateDevice | Additional PredicateDevice |
| AXS Catalyst 7 DistalAccess Catheter | AXS Catalyst 5 and 6Distal Access Catheter | Penumbra SystemACE 68 ReperfusionCatheter | |
| jar and a non-reusableliner. The AXSUniversal Liner Setconsists of 3individually packagedpoly-bags containingone canister liner andone ClotFinderspecimen cup. TheAXS Universal LinerSet is offered with andwithout a desiccant.• Class I Exempt• Non-Sterile | reusablepolycarbonate jar.• Class I Exempt• Non-Sterile |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
The differences between the devices are not critical as demonstrated above and through the testing referenced below.
Risk Assessment
Risk assessment of the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system has been conducted in accordance with EN ISO 14971. A warning is included in the instructions for use to instruct users as to appropriate use of aspiration. Results of testing are appropriate for determining that the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system is substantially equivalent to the legally marketed Predicate devices.
Testing Summary
Performance Data – Bench Testing
{13}------------------------------------------------
The results of design verification and design validation testing conducted on the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system demonstrates that it performs as designed, is suitable for the indication for use, and is substantially equivalent to the legally marketed Predicate devices. Testing was conducted in accordance with EN ISO 10555-1 and EN 1707. The design verification and design validation bench testing are summarized in Table 2 below.
| Table 2: Performance Data - Bench Testing | |||
|---|---|---|---|
| Test | Test Method Summary | Conclusions | |
| Design Verification Testing | |||
| Dimensional Verification | Purpose: To describe the procedure andtechnique of making dimensionalmeasurements using variousmeasurement equipment.Method: Verify dimensions usingspecified measurement tool. Recordmeasurements. | Dimensional verificationmeets acceptancecriteria. | |
| Tip Configuration | Purpose: To verify that the catheter tip issmooth, rounded, tapered or similarlyfinished in order to minimize trauma tovessels during use per EN ISO 10555-1.Method: Visually inspect distal tip at 10Xmagnification to verify distal tip end issmooth, rounded, tapered or similarlyfinished. Record results. | Tip configuration meetsacceptance criteria. | |
| Surface Integrity | Purpose: To determine if externalsurface of the catheter is free fromextraneous matter, process and surfacedefects, and does not have drops oflubricant fluids.Method: Visually inspect externalsurface of catheter for extraneousmatter, process and surface defects, anddrops of lubricant fluids. Record results. | Surface integrity meetsacceptance criteria. | |
| Table 2: Performance Data - Bench Testing | |||
| Test | Test Method Summary | Conclusions | |
| Tip Buckling | Purpose: To measure the maximumforce required to cause a test sample tobuckle. | Tip buckling meetsacceptance criteria. | |
| Method: Prepare sample for test. Usebuckling tester to measure the maximumforce required to cause a test sample tobuckle. Record results. | |||
| Catheter lubricity anddurability | Purpose: To determine the lubricity anddurability of the coating on the catheterouter shaft. | Coating lubricity anddurability meetsacceptance criteria. | |
| Method: Prepare sample for test. Usefriction tester to measure the frictionalforce of the device sample when pulledbetween two clamped pads. Record thepeak frictional force over 5 cycles. | |||
| Particulatecharacterization | Purpose: To evaluate particulates per theFDA Guidance Document: Class II SpecialControls for Certain PercutaneousTransluminal Coronary Angioplasty(PTCA) Catheters.Method: Use light obscuration particlecounting to measure the total number of | Particulate generationwas acceptable. | |
| particulates generated during simulateduse in each of three size ranges: ≥ 10µm, ≥ 25 µm, and ≥ 50 µm. | |||
| Coating integrity | Purpose: To evaluate coating consistencypre and post simulated use. | Coating integrity wasacceptable. | |
| Method: Visually inspect for coatinganomalies, defects or artifacts pre andpost simulated use using a simulated invitro test model. | |||
| Table 2: Performance Data - Bench Testing | |||
| Test | Test Method Summary | Conclusions | |
| Trackability | Purpose: To measure track advanceforce of catheter over microcatheter.Method: Place neurovascular model in are-circulating water bath at 37°C tosimulate human arterial circulation.Insert sample, attached to a motor driveassembly, through model over amicrocatheter. Advance catheterthrough model and determine peaktracking force. Record results. | Track advance forcemeets acceptancecriteria. | |
| Tensile Strength | Purpose: To determine tensile forcetensile force required to induce failure offused joints, shaft junctions, and markerband for non-hydratable catheters basedon EN ISO 10555-1.Method: Identify joint and preparesample for test. Use tensile tester todetermine applied peak tensile force.Record results. | Tensile strength meetsacceptance criteria. | |
| Liquid Leak Resistance | Purpose:1) To determine whether cathetermeets the freedom from leakage-liquidleak requirement 4.7.1 of EN ISO 10555-1.2) To determine if catheter hub meetsthe liquid leakage requirement 4.2.1 ofEN 1707.Method: Connect test hub sample tofixture and flush with water to expel air.Occlude distal tip. Apply pressure of300kPa minimum and maintain pressurefor 30s. Visually inspect catheter/hubjoint and catheter shaft for leaks.Record results. | Liquid leak resistance ofcatheter meetsacceptance criteria. | |
| Table 2: Performance Data - Bench Testing | |||
| Test | Test Method Summary | Conclusions | |
| Air Leak Resistance | Purpose:1) To determine whether catheter meetsthe freedom from leakage-air aspirationrequirement of 4.7.2 of EN ISO 10555-1.2) To determine if catheter hub meetsthe air leakage requirement 4.2.2 of EN1707.Method: Connect test hub sample to apartially filled syringe. With the nozzle ofthe syringe pointing down towards theground, withdraw the plunger to the10cc mark. Hold for 15 seconds andexamine the water in the syringe for theformation of air bubbles. Record results. | Air leak resistance ofcatheter meetsacceptance criteria. | |
| Catheter Torsional BondStrength | Purpose: To measure the strength of acatheter shaft when torque is applied.Torque strength is defined as number ofrotations before failure occurs.Method: Prepare test sample and insertinto torsional bond strength test fixturewith tortuous path model. Apply torqueto catheter shaft and observe number of360-degree rotations before failureoccurs. Record results. | Catheter torsional bondstrength meetsacceptance criteria. |
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
| Table 2: Performance Data - Bench Testing | ||
|---|---|---|
| Test | Test Method Summary | Conclusions |
| Flexural Fatigue | Purpose: To determine the flexuralfatigue on the catheter shaft.Method: Prepare test sample. Advanceentire assembly of guide wire,microcatheter, and test sample into testmodel and track it through test model.While holding the guide wire,microcatheter, and test sample, pull thewhole assembly back proximally until itexits the models. Repeat for nine moreruns. After run number ten, removeguide wire and microcatheter out of testsample and inspect sample for kink ordamage. Record results. | Flexural fatigue meetsacceptance criteria. |
| Catheter Kink Radius | Purpose: To measure the kink radius of acatheter at its distal and specific mid-shaft joint section.Method: Prepare test sample. Threadtest sample through fixture loop and lockdown test sample. Pull both ends of testsample until test sample kinks. Calculatekink radius using measurement of 2nd tofinal loop OD and sample OD. Recordresults. | Catheter kink radiusmeets acceptancecriteria. |
| Table 2: Performance Data - Bench Testing | ||
| Test | Test Method Summary | Conclusions |
| Catheter Tip and Lumen Integrity(Direct Aspiration) | Purpose: To test resistance to tip and lumen collapse during direct aspiration and test tip integrity to tears and missing material.Method: Prepare test sample and simulated use model. Insert plug in catheter tip. Place test sample in the model to a specified location following procedural instructions outlined in the Instructions for Use. Aspirate test sample using aspiration pump. Visually inspect test sample to verify absence of tip or lumen collapse. Record results. | Catheter tip and lumen integrity during direct aspiration meets acceptance criteria. |
| Chemical Compatibility | Purpose: To determine visual and dimensional integrity of catheter following exposure to saline and non-ionic contrast liquids.Method: Prepare sample for test. Flush sample with appropriate chemical. Measure ID and OD using RAM optical measurement system. Insert mandrel through sample to verify inner lumen integrity. Repeat with second mandrel and record results. Visually inspect distal end of sample for any chemical effects on the shaft, inner lumen and cross-sectional areas. Record results. | Chemical compatibility meets acceptance criteria. |
| Hub Gauging | Purpose: To determine if catheter hub meets gauging requirement 4.1 of EN 1707.Method: Using the appropriate gauge, apply gauge to the conical fitting with a total axial force of 5N without use of torque. Remove axial load and inspect sample. | Hub gauging meets acceptance criteria. |
| Aspiration Flow Rate Bench Testing | ||
| Table 2: Performance Data - Bench Testing | ||
| Test | Test Method Summary | Conclusions |
| Aspiration Flow Rate | Purpose: To determine the aspiration flow rate and lumen integrity of the system when no occlusion is present.Method: Measure the volume of water an aspiration system (catheter + associated pump and accessories) can aspirate in 20s. Divide this volume in mL by 20 to give the flow rate in mL/s. | All test samples meet acceptance criteria. |
| Design Validation Testing | ||
| In-vitro Usability Study | Purpose: To evaluate aspiration integrity, ability to restore flow and the durability and kink resistance of the Subject and Predicate devices in a tortuous anatomical model with multiple physician users.Method: Track the devices to the target site to perform simulated neurothrombectomy procedure in the neurovascular model that replicates the tortuosity, diameter and location of the arteries in the neurovasculature. | All test samples meet acceptance criteria. |
{18}------------------------------------------------
{19}------------------------------------------------
{20}------------------------------------------------
Performance Data – Animal Study
An animal study was conducted in compliance with applicable requirements in the GLP regulation (21 CFR Part 58) to evaluate navigation and aspiration safety of the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system as outlined below in Table 3. Additionally, safety and performance with simulated clot retrieval from a GLP study conducted on the AXS Catalyst 5 and 6 Primary Predicate devices, which are considered to be worst-case test articles for thrombus retrieval, have been leveraged for AXS Catalyst 7 Subject device safety and performance.
| Table 3: Performance Data - Animal Testing | |||||
|---|---|---|---|---|---|
| Test | Test Method Summary | Conclusions | |||
| In-vivo Safety Evaluation | Purpose: To assess the vascular safetyprofile of treatment by aspiration andnavigation of the Subject Devicecompared to the Additional PredicateDevice.Method: Expose porcine test subjects toaspiration treatment using the SubjectAXS Catalyst 7 with AXS UniversalAspiration System and the AdditionalPredicate Penumbra System underworst-case aspiration force andtreatment duration conditions. Assessvascular response by contrastangiography and histopathology at botha subacute (~3 days) and chronic timepoints (~30 days). | Subject Device isequivalent to theAdditional PredicateDevice in safety. |
Performance Data - Clinical
No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the Predicate devices. The review demonstrated that equivalent devices published clinical outcomes that supports the proposed line extension to the AXS Catalyst Distal Access Catheter product family. Given the equivalency of the Subject device and Predicate devices, the Subject device is therefore suitable for the proposed indication for use under the associated NRY Product Code.
{21}------------------------------------------------
Shelf Life Testing
The labeled shelf life for the AXS Catalyst 7 Distal Access Catheter is one year. Shelf life testing (product and packaging) and Distribution Shipping Challenge Conditioning and testing were performed on the Subject device and the results met established criteria.
The shelf life for the AXS Universal Aspiration Tubing, AXS Universal Liner Set, and Medela Dominant Flex Pump are established by Medela.
Sterilization
The AXS Catalyst 7 Distal Access Catheter and all system components (including the Rotating Hemostasis Valve, the Tuohy Borst Valve and the Peel-Away Introducer Sheath) are sterilized with 100% Ethylene Oxide. The AXS Catalyst 7 Distal Access Catheter and the accessories packaged with the Catheter (RHV, Tuohy Borst Valve, and the Peel-Away Introducer Sheath) are provided sterile. A sterility assurance level (SAL) of 10-6 has been demonstrated. The AXS Catalyst 7 Distal Access Catheter and all system components meets EO residuals per EN ISO 10993-7 for a limited contact delivery system – externally communicating. The AXS Catalyst 7 Distal Access Catheter and all system components are for single use only.
The AXS Universal Aspiration Tubing sterilization is provided by Medela.
Biocompatibility
The AXS Catalyst™ 7 Distal Access Catheter was assessed for biocompatibility in accordance with EN ISO 10993-1, "Biological evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process". The Subject device is considered an externally communicating medical device with circulating blood contact for less than 24 hours.
Biocompatibility testing previously conducted for AXS Catalyst 5 & 6 Distal Access Catheter product is representative of the AXS Catalyst 7 Distal Access Catheter product due to similarity of materials, manufacturing processes and design. While there are minor differences, the biocompatibility assessment determined that these differences are low risk and do not impact product biocompatibility. Therefore, the AXS Catalyst 7 Distal Access Catheter will have the same or similar toxicological profile as the AXS Catalyst 5 and 6 Distal Access Catheter. This conclusion was confirmed by conducting selected tests in accordance with EN ISO 10993-1 and its applicable sub-parts. The AXS Catalyst 7 Distal Access Catheter, passed all required biocompatibility testing.
The results of the biocompatibility testing are summarized in Table 4 below.
{22}------------------------------------------------
| Table 4: Overview of Biocompatibility Studies Performed on the Subject Device | |||
|---|---|---|---|
| Test Performed /Applicable ISO 10993Part No. | Results | Conclusion | |
| MEM ElutionCytotoxicity/Part 5 | No biological activity (Grade 0) was observedin the L929 mammalian cells at 48 hours postexposure to the test article extract. The testarticle met the requirements of the test sincethe grade was less than grade 2.The observed cellular response obtained fromthe positive control article extract (Grade 4)and negative control article extract (Grade 0)confirmed the suitability of the test system. | PASSNo cytotoxicityor cell lysis | |
| Hemolysis Extract/DirectContact Method/Part 4 | The test article exhibited 0.0% hemolysisabove the level of hemolysis exhibited by thenegative control via the direct method and0.0% hemolysis above the level of hemolysisexhibited by the negative control via theindirect method. | PASSNon-hemolytic | |
| USP Physiochemical<661> / Part 18 | Non-volatile residue: 7 mgResidue on ignition: < 1 mgHeavy metal: < 1 ppmBuffering capacity: < 1.0 ml | PASS | |
| Chemistry (Heptane)Analysis | Below level of detection | PASS | |
| FTIR (ISO 10993-18) | Scan was conducted to establish baseline forthe AXS Catalyst 7 Distal Access Catheterdevice. | PASS | |
| Natural Rubber LatexELISA Inhibition Assay | Below level of detection | PASS | |
| Table 4: Overview of Biocompatibility Studies Performed on the Subject Device | |||
| Test Performed /Applicable ISO 10993Part No. | Results | Conclusion | |
| for Antigenic ProteinASTM D6499-12 |
{23}------------------------------------------------
Conclusion
Stryker Neurovascular has demonstrated the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system is substantially equivalent to the Primary Predicate device (K173841) and the Additional Predicate device (K161640) based on similar intended use / indications for use, same or similar materials, same fundamental design, and the same fundamental operating principles. The conclusions drawn from risk assessments, the bench and testing conducted using the Subject device compared to the Predicate devices, and a review of available real-world evidence from published clinical articles demonstrate that the subject device is suitable for the indication for use with the associated NRY Product Code. Additionally, testing results summarized above along with the risk assessment demonstrate that the benefits of the device outweigh any residual risks when used in accordance with device Instructions for Use.
Stryker Neurovascular has demonstrated that the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system is as safe, as effective, and performs as well as the legally marketed Predicate devices.
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).