Search Results
Found 2 results
510(k) Data Aggregation
(267 days)
The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.
The Millipede 088 Access Catheter consists of the catheter, a rotating hemostasis valve (RHV) and a valve crossing tool. The catheter, RHV and valve crossing tool are provided sterile. They are sterilized by ethylene oxide (EO).
The Millipede 088 Access Catheter is a single lumen, coil-reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The valve crossing tool is used to open the valve of the access sheath and to facilitate insertion of the Millipede 088 Access Catheter through the access sheath without damage. The RHV is assembled onto the hub of the Millipede 088 Access Catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of other devices through the Millipede 088 Access Catheter.
The provided document is a 510(k) summary for the Millipede 088 Access Catheter. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing detailed acceptance criteria and a standalone clinical study to prove the device meets those criteria.
Therefore, the document does not contain the information requested in points 2, 3, 4, 5, 8, and 9, and only partially addresses points 1, 6, and 7.
Here's a breakdown of what can be extracted:
1. Acceptance Criteria and Reported Device Performance
The document lists various performance tests and their conclusions, indicating that the device met "established specifications" or was "suitable for its intended use." However, the specific quantitative acceptance criteria for each test are not provided. The reported device performance is qualitative rather than quantitative in most cases.
Test | Test Method | Reported Device Performance (Conclusions) |
---|---|---|
Dimensional Inspection | Device dimensions were measured to confirm conformance to the specifications. | The device met established specifications. |
Tip Stiffness | Test specimens were tested for tip flexibility and compared to predicate and reference devices. | The device met established specifications. |
Visual Inspection | Device surface characteristics were assessed to confirm freedom from defects. | The device surface characteristics are suitable for its intended use. |
Simulated Use Testing | Deliverability and compatibility with accessory devices were evaluated in a neurovascular model. | The device performs as intended under simulated use conditions. |
Hydrophilic Coating Integrity | The integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles. | The hydrophilic coating integrity is suitable for its intended use. |
Particulate Recovery | Quantify the particulate size and count generated by simulated use of the test article. | The particulate size and count were similar to control devices. |
Tensile Strength | The tensile strength was evaluated for the bonds between sections of the catheter. | The device met established specifications. |
Air Leakage | Tested per ISO 10555-1:2013 Annex D. | The device integrity is suitable for its intended use. |
Liquid Leakage | Tested per ISO 10555-1:2013 Annex C. | The device integrity is suitable for its intended use. |
Static Burst | Tested per ISO 10555-1:2013 Annex F. | The device integrity is suitable for its intended use. |
Luer Integrity | The luers were evaluated for compliance to relevant standards. | The luers on the device are suitable for their intended use. |
Kink Resistance | Test specimen segments were formed into a defined bend diameter to evaluate kink resistance. | The device met established specifications. |
Torque Strength | The test specimens were rotated in a simulated use model to evaluate integrity after rotation. | The device met established specifications. |
Flow Rate Characterization | The flow rate of saline and a contrast-saline solution was characterized when injected through the catheter. | The flow rate was characterized. |
Radiopacity | Radiopacity of the device was evaluated in an animal model under fluoroscopy. | The radiopacity of the Millipede 088 Access Catheter was similar to a control device. |
Biocompatibility (various tests) | ISO 10993-1, 10993-4 (Hemocompatibility), 10993-5 (Cytotoxicity), 10993-10 (Irritation, Sensitization), 10993-11 (Systemic Toxicity, Pyrogenicity) | Non-cytotoxic, no sensitization response, met intracutaneous reactivity, met acute systemic injection, non-pyrogenic, not a complement activator, not an intrinsic coagulation pathway activator, non-hemolytic, similar thromboresistance to controls. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample sizes for the individual performance tests beyond implicitly suggesting multiple samples were tested (e.g., "test specimens," "test articles"). For the Animal Testing, it states "two studies in a porcine model."
The data provenance for the in vitro and animal studies isn't explicitly stated beyond "Good Laboratory Practices" for the animal studies, which is a standard of conduct rather than a geographic origin. No human data is presented.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided as the submission relies on objective performance testing, biocompatibility studies, and animal studies rather than expert-derived ground truth from human data for this type of device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The evaluation methods described are objective performance tests and animal studies, not human data requiring expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an access catheter, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI assistance is not relevant to its evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the performance testing described (mechanical, dimensional, coating integrity, etc.) and the animal studies represent "standalone" evaluations of the device's physical and functional properties without human interpretation of data in a clinical context.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the performance and biocompatibility tests, the "ground truth" is defined by the established specifications, relevant ISO standards, and comparison to control devices or predicate devices. For example:
- "The device met established specifications."
- "The test article is non-cytotoxic."
- "The particulate size and count were similar to control devices."
- "The radiopacity of the Millipede 088 Access Catheter was similar to a control device."
- For animal testing, "Usability, radiopacity, thromboresistance, and vessel injury were assessed," and "The results for the subject device were comparable to a control device."
8. The sample size for the training set
Not applicable. This device is a medical catheter and does not involve a "training set" in the context of machine learning or AI. Its performance is evaluated through engineering and biological testing.
9. How the ground truth for the training set was established
Not applicable (as above).
Ask a specific question about this device
(151 days)
Naviplan is indicated for pre-operative planning for surgical procedures related to hip, such as artificial joint replacement (3D templating of implants).
NAVIPLAN - CT Planning Software for Total Hip Replacement, or "Naviplan", is a software which assists in the pre-operative planning of total hip replacement surgery based on three dimensional CT-images (computed tomography). The Naviplan allows the user to process CT DICOM images which are automatically segmented. The user gets an automatic identification of anatomical landmarks as an output, which is normally used in hip surgery. The Naviplan automatically computes an initial planning (implant type and size selection as well as positioning) which can be inspected and edited by the user until the optimal configuration is established. The user can export the plan as a pdf report and as a navigation archive to be used with the Naviswiss Hip Navigation System. The export steps are optional, whereas the report serves a documentation purpose.
Here's a breakdown of the acceptance criteria and study information for the NAVIPLAN - CT Planning Software for Total Hip Replacement, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" with numerical targets and then directly provide "reported device performance" in a separate section. Instead, it describes a software verification and validation testing approach. The "reported device performance" is summarized by the statement: "Dedicated software verification testing has been conducted to compare manual segmentation and landmark positioning to the NAVIPLAN - CT Planning Software for Total Hip Replacement automated segmentation and planning." And the overall conclusion: "the applicant concludes, that substantial equivalence between the new and the predicate device has been demonstrated and that the new device, NAV-IPLAN - CT Planning Software for Total Hip Replacement, is at least as safe and as effective as the legally marketed predicate device, Zedview (K133022)."
Based on the information, the acceptance criteria are implicitly tied to the performance of manual segmentation and landmark positioning. The device is expected to perform at a level comparable to, or as accurate as, manual methods.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Automated segmentation and landmark positioning performance comparable to manual methods. | "Dedicated software verification testing has been conducted to compare manual segmentation and landmark positioning to the NAVIPLAN - CT Planning Software for Total Hip Replacement automated segmentation and planning." This implies the automated segmentation and planning results were found to be acceptable when compared to manual methods, leading to the overall conclusion of substantial equivalence. |
Device is at least as safe and effective as the legally marketed predicate device (Zedview, K133022). | The study concludes that substantial equivalence has been demonstrated, indicating the device meets this criterion. |
2. Sample Size for Test Set and Data Provenance
The document states: "Dedicated software verification testing has been conducted to compare manual segmentation and landmark positioning to the NAVIPLAN - CT Planning Software for Total Hip Replacement automated segmentation and planning." However, it does not specify the sample size used for this testing. It also does not mention the country of origin of the data or whether it was retrospective or prospective.
3. Number of Experts and Qualifications for Ground Truth
The document does not specify the number of experts used to establish the ground truth for the test set or their specific qualifications (e.g., "radiologist with 10 years of experience"). It only implies the use of "manual segmentation and landmark positioning" as a reference for comparison.
4. Adjudication Method for the Test Set
The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. Therefore, there is no information on the effect size of how much human readers improve with AI vs. without AI assistance. The study described focuses on comparing the automated software's performance to manual methods, not on human reader performance with or without the software.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
Yes, a standalone performance assessment was done. The software verification testing "compare[d] manual segmentation and landmark positioning to the NAVIPLAN - CT Planning Software for Total Hip Replacement automated segmentation and planning." This implies the algorithm's performance in segmentation and planning was evaluated independently against manual ground truth without a human in the loop of the algorithm's output. The device itself is described as primarily standalone for planning, though its output can be used with a navigation system.
7. Type of Ground Truth Used
The ground truth used was based on manual segmentation and landmark positioning. This can be interpreted as a form of expert consensus or expert-derived ground truth, as manual methods in such a clinical context would typically be performed by trained medical professionals or experts.
8. Sample Size for the Training Set
The document does not provide information on the sample size used for the training set.
9. How Ground Truth for the Training Set was Established
The document does not provide information on how the ground truth for the training set was established.
Ask a specific question about this device
Page 1 of 1