(190 days)
No
The device description focuses on the physical characteristics and materials of the catheter, and there is no mention of AI, ML, or any software-driven analytical capabilities. The performance studies are bench and design validation, not studies typically associated with evaluating AI/ML performance.
No
The device is described as an "Intracranial Support Catheter" indicated for the "introduction of interventional devices" into the vasculature, not for providing therapy itself. Its function is to facilitate other interventional devices.
No
The device is described as an "Intracranial Support Catheter" intended for the "introduction of interventional devices." This indicates a therapeutic or procedural support role, not a diagnostic one.
No
The device description clearly details physical components like catheters, Nitinol structure, radiopaque marker bands, and hydrophilic coating, indicating it is a hardware device, not software-only.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for the introduction of interventional devices into the peripheral and neurovasculature. This is a procedural device used in vivo (within the body) to facilitate other medical interventions.
- Device Description: The description details a catheter designed for physical insertion into blood vessels.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens (like blood, urine, tissue) in vitro (outside the body) to provide information for diagnosis, monitoring, or screening.
IVD devices are specifically designed to perform tests on samples taken from the body to provide diagnostic information. This device is a tool for delivering other devices during a medical procedure.
N/A
Intended Use / Indications for Use
The Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter are indicated for the introduction of interventional devices into the peripheral and neurovasculature.
Product codes
DOY, DQY
Device Description
The Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter are designed for the introduction of interventional devices into the peripheral and neurovasculature. The Arc™ Mini are a single lumen, flexible, variable stiffness composite catheters with a Nitinol structure. A radiopaque marker band on the distal tip of the devices is used for visualization under fluoroscopy. The distal sections of both catheters are coated with a hydrophilic coating, which is used to reduce the overall frictional force during intravascular use. The Arc™ and Arc™ Mini dimensions are included in the individual device label. The devices are supplied sterile and are intended for single-use only.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Fluoroscopy
Anatomical Site
Peripheral and neurovasculature.
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Prescription Use (Part 21 CFR 801 Subpart D)
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The following bench testing was performed in support of the Arc™ and Arc™ Mini Intracranial Support Catheter and to establish substantial equivalence to the ReFlex™ Guide Catheter:
- Lumen Patency: All devices met acceptance criteria.
- Dimensional Inspection: All device met acceptance criteria.
- Distal Tip Buckling: Distal tip buckling force met acceptance criteria.
- Catheter Injection Flow Rate: Injection flow rate met acceptance criteria.
- Catheter Suction Flow Rate: Suction flow rate met acceptance criteria.
- Vacuum Resistance: Device was resistant to vacuum collapse under static conditions.
- Hub Air Aspiration Leak: Hub air aspiration test met acceptance criteria.
- Shaft Peak Tensile Force: Shaft joint peak tensile force met acceptance criteria.
- Hub Peak Tensile Force: Hub-shaft joint peak tensile force met acceptance criteria.
- Coating Integrity – Baseline: Baseline coating integrity characterized with worst-case defect images. Data collected as engineering reference.
- Coating Lubricity/Durability: Coating lubricity and durability testing met acceptance criteria.
- Particulate Testing: Number of particulates generated met acceptance criteria.
- Coating Integrity Simulated Use: Post-simulated use coating integrity characterized with worst-case defect images.
- Kink Resistance: Device was resistant to kinking around small radii.
- Liquid Leakage: Liquid leakage met acceptance criteria.
- Static/Dynamic Burst: Static/dynamic burst testing met acceptance criteria.
- Torque to Failure: Torque to failure testing met acceptance criteria.
- Physician Usability Testing: The device was navigated through a tortuous benchtop model to assess compatibility with accessories, device stability, ability to aspirate, ability to inject saline or contrast, and the user's ability to navigate to the M1 and M2 segment of the MCA and retrieve a mechanical thrombectomy device. All test results met the acceptance criteria.
A Design Validation study was performed on a bench model to assess the usability of the Arc™ and ArcTM Mini Intracranial Support Catheter compared to the previously cleared ReFlex™ Guide Catheter.
Biocompatibility testing, sterilization, and a 6-month accelerated aging study were also performed.
No clinical studies were performed.
No animal study was performed.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
ReFlex™ Guide Catheter, 510(k)#: K110055
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an emblem that resembles an abstract representation of an eagle or a similar bird, with stylized feathers or wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 29, 2015
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Ms. Phuong Chau Senior Regulatory Affairs Product Specialist 9775 Toledo Way Irvine, California 92618
Re: K150107
Trade/Device Name: Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: June 26, 2015 Received: June 29, 2015
Dear Ms. Phuong Chau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
1
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena நி/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K150107
Device Name
ArcTM Intracranial Support Catheter and ArcTM Mini Intracranial Support Catheter
Indications for Use (Describe)
The Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter are indicated for the introduction of interventional devices into the peripheral and neurovasculature.
Type of Use (Select one or both, as applicable)
For Activities Related to CA SB1383 (Organic Waste) |
---|
For Other Composting Activities |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary – K150107
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular
9775 Toledo Way
Irvine, CA 92618
Establishment Registration No. 2029214 | | | |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Contact Person: | Phuong Chau
Senior Regulatory Affairs Product Specialist
Telephone: (949) 297-5487
E-mail: phuong.chau@covidien.com | | | |
| Date Summary
Prepared: | June 25, 2015 | | | |
| Trade Name of
Device: | Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial
Support Catheter | | | |
| Common Name of
Device: | Catheter, Percutaneous | | | |
| Classification of
Device: | 21 CFR 870.1250 – Class II | | | |
| Product Code: | DQY | | | |
| Predicate Device: | ReFlex™ Guide Catheter, 510(k)#: K110055 | | | |
| Performance Data: | The following bench testing was performed in support of the Arc™ and
Arc™ Mini Intracranial Support Catheter and to establish substantial
equivalence to the ReFlex™ Guide Catheter: | | | |
| | • Lumen Patency | | | |
| | • Dimensional Inspection | | | |
| | • Distal Tip Buckling | | | |
| | • Catheter Injection Flow Rate | | | |
| | • Catheter Suction Flow Rate | | | |
| | • Vacuum Resistance | | | |
| | • Hub Air Aspiration Leak | | | |
| | • Shaft Peak Tensile Force | | | |
| | • Hub Peak Tensile Force | | | |
| | • Coating Integrity – Baseline | | | |
| | • Coating Lubricity/Durability | | | |
| | | | | |
4
- . Particulate Testing
- Coating Integrity Simulated Use .
- . Kink Resistance
- Liquid Leakage .
- Static/Dynamic Burst .
- . Torque to Failure
A Design Validation study was performed on a bench model to assess the usability of the Arc™ and ArcTM Mini Intracranial Support Catheter compared to the previously cleared ReFlex™ Guide Catheter. Biocompatibility testing, sterilization, and a 6-month accelerated aging study were also performed. No clinical studies were performed as there is no change to the indications for use or the fundamental scientific technology of the device.
- Conclusion: The ArcTM Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter are substantially equivalent to the currently cleared ReFlex™ Guide Catheter based on the successful completion of nonclinical bench and design validation testing as well as similar principles of design, operation and indications for use.
Device Description:
The Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter are designed for the introduction of interventional devices into the peripheral and neurovasculature. The Arc™ Mini are a single lumen, flexible, variable stiffness composite catheters with a Nitinol structure. A radiopaque marker band on the distal tip of the devices is used for visualization under fluoroscopy. The distal sections of both catheters are coated with a hydrophilic coating, which is used to reduce the overall frictional force during intravascular use. The Arc™ and Arc™ Mini dimensions are included in the individual device label. The devices are supplied sterile and are intended for single-use only.
Indications for Use:
The Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter are indicated for the introduction of interventional devices into the peripheral and neurovasculature.
Device Comparison
The table below provides a comparison of the technological characteristics of the Arc™ Intracranial Support Catheter and Arc M Mini Intracranial Support Catheter and the currently cleared ReFlex (M Guide Catheter.
| | ReFlex™ Guide
Catheter
(K110055) | Arc™ Intracranial
Support Catheter | Arc™ Mini
Intracranial
Support Catheter | Rationale for
Difference (If
Present) |
|-----------------------|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------|
| Indication for
Use | The ReFlex™
Guide Catheter is
indicated for the
introduction of | The Arc Intracranial
Support Catheter is
indicated for the
introduction of | The Arc Mini
Intracranial Support
Catheter is indicated
for the introduction | N/A |
5
| | ReFlex™ Guide
Catheter
(K110055) | Arc™ Intracranial
Support Catheter | Arc™ Mini
Intracranial
Support Catheter | Rationale for
Difference (If
Present) |
|--------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | interventional
devices into the
peripheral and
neurovasculature. | interventional
devices into the
peripheral and
neurovasculature. | of interventional
devices into the
peripheral and
neurovasculature. | |
| Materials | | | | |
| Catheter
Shaft
Materials | PTFE lined
polymeric catheter,
with hydrophilic
coating | PTFE lined
polymeric catheter,
with hydrophilic
coating | PTFE lined
polymeric catheter,
with hydrophilic
coating | The material used
for the Arc and
Arc Mini catheters
were shown to be
biocompatible per
ISO 10993 testing.
Materials of this
type are widely
used in similar
medical devices. |
| Catheter
Shaft
Support | Nitinol | Same | Same | N/A |
| Marker
band | Platinum | Same | Same | N/A |
| Dimensions | | | | |
| Usable Length | 90 – 130 cm | 132 – 135 cm | 160 – 163 cm | Longer lengths
provided for
additional
navigability
options to user. |
| Distal ID | 0.046" – 0.072" | 0.061" | 0.035" | Smaller distal ID
provided for
improved
trackability over
guide wire to
smaller vessels. |
| Distal OD
(Max) | 0.058"–0.084" max | 0.071" max | 0.049" | Smaller distal
ID/OD provided
for improved
distal flexibility. |
| Proximal ID | 0.046" – 0.072" | 0.069" | 0.044" | Smaller proximal
ID provided for
improved
trackability over
guide wire to |
6
| | ReFlex™ Guide
Catheter
(K110055) | Arc™ Intracranial
Support Catheter | Arc™ Mini
Intracranial
Support Catheter | Rationale for
Difference (If
Present) |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------------------|-------------------------------------------------------------------------|
| | | | | smaller vessels. |
| Proximal OD
(Max) | 0.058"–0.084" max | 0.0825" max | 0.0620" | N/A |
| Tip
Configuration | Single, straight
flexible tip | Same | Same | N/A |
| Guidewire
Compatibility | Can be navigated
over a guidewire
with a maximum
OD of 0.038 in. | Same | Can be navigated
over a guidewire
with a maximum
OD of 0.032 in. | Largest
compatible
guidewire size
dependent on
smallest ID. |
| Sterilization
Method | Ethylene Oxide | Same | Same | N/A |
| Packaging | Catheter in
polyethylene hoop
attached to
packaging card
inside
PET/PE/Tyvek
pouch inside SBS
carton | Same | Same | N/A |
Sterilization and Shelf Life
The packaged Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter are sterilized using a validated ethylene oxide (EO) sterilization cycle at the Sterigenics US; LLC facility located at 4900 Gifford Avenue, Los Angeles, CA. The sterilization cycle has been validated to ensure a sterility assurance level (SAL) of 10-s in accordance with ISO 11135-1:2007, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
Aging studies for the Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter devices have established the product and packaging remain functional and maintain sterility for up to 6 months. Aging studies for packaging integrity, seal strength, and device functionality were performed and met all acceptance criteria.
Biocompatibility
Biocompatibility testing was performed in compliance with the FDA consensus standard, recognition number 2-156, AAMI/ANSI/ISO 10993-1: 2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process and U.S. Food and Drug Administration (FDA) Blue Book Memorandum G95-1 (1995) guidelines. All studies were conducted in compliance with U.S. Food and Drug Administration Good Laboratory Practice (GLP) regulations set forth in 21 CFR Part 58.
7
Test | Result | Conclusion |
---|---|---|
Plastics (USP) | Meets USP Physicochemical | |
extraction parameters. | Passes physical chemical | |
characteristics. | ||
L929 MEM Elution Test - ISO | The test article scored "0" at 24, 48 | |
and 72 ± 4 hours and is considered | ||
non-cytotoxic under the conditions | ||
of this test. | Non-cytotoxic | |
Klingman Maximization Test - | ||
ISO (Guinea Pig Sensitization) | Under the conditions of this | |
protocol, the test article did not | ||
elicit a sensitization response. | Non-sensitizer | |
Intracutaneous Injection Test - | ||
ISO | The differences in the mean test | |
and control scores of the extract | ||
dermal observations were less than | ||
1.0, indicating that the requirements | ||
of the ISO Intracutaneous | ||
Reactivity Test have been met by | ||
the test article. | Non-irritant | |
Acute Systemic Injection Test - | ||
ISO | None of the test article extract | |
treated animals were observed with | ||
clinical signs consistent with | ||
toxicity at any of the observation | ||
periods. | Non-cytotoxic | |
Materials Mediated Rabbit | ||
Pyrogen - ISO | This response did not exceed the | |
USP limit and meets the | ||
requirements for this test. Therefore | ||
these results indicate that the test | ||
article was determined to be non- | ||
pyrogenic. | Non-pyrogenic | |
Hemolysis: Direct Contact / | ||
Indirect Extract | There were no significant | |
differences between the test article | ||
extract and negative control article | ||
results. The test article is | ||
considered non-hemolytic | Non-hemolytic | |
Complement activation C3a and | ||
SC5b-9 | The levels of C3a and SC5b-9 of | |
the Dyson catheter are comparable | ||
to the ReFlex and less than that of | ||
the positive control. | Levels of the compliments C3a | |
and SC5b complements were | ||
similar for Arc and control | ||
device | ||
Thrombosis (in vivo) — Canine | ||
(Arc / ReFlex) | The thromboresistance properties | |
of the Arc and Arc Mini | ||
Intracranial Support Catheters are | ||
acceptable in clinical use. | Acceptable, expected to be | |
equivalent to ReFlex in clinical | ||
use | ||
in vitro Hemocompatibility | ||
Assay | Neither the Arc nor the ReFlex | |
Guide Catheter resulted in a | ||
decrease in any blood component | No adverse effect on platelet | |
and leukocyte counts |
8
Test | Result | Conclusion |
---|---|---|
as compared to the reference | ||
material. These results indicate that | ||
the cause of thrombi is not related | ||
to the materials exposed to human | ||
blood during use. | ||
Partial Thromboplastin Time | Clotting times for Arc and the | |
ReFlex (predicate device) test arms | ||
were similar to the negative control | ||
and the reference material (HDPE), | ||
indicating that the device materials | ||
are not an activator of the intrinsic | ||
coagulation pathway. | No adverse effect on | |
prothrombin coagulation time | ||
of human plasma. | ||
Ames bacterial Mutagenicity 4 | ||
salmonella+1e.coli | Based on the criteria and conditions | |
of the study protocol, the test article | ||
is considered non-mutagenic. | Non-mutagenic | |
in vitro Mouse Lymphoma Assay | ||
with Extended Treatment | The test article is considered to be | |
non-mutagenic (non-genotoxic and | ||
non-clastogenic) in this test system. | Non-mutagenic | |
in vivo Mouse Micronucleus | ||
Assay | Based on the criteria of the assay, | |
the test article is considered non- | ||
mutagenic in this test system. | Non-mutagenic |
Performance Testing – Bench
A summary of the pre-clinical bench testing performed for the Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter is presented in the table below.
Test | Method | Conclusions |
---|---|---|
Lumen Patency | The total length of the device | |
must pass a mandrel of the | ||
required size from the proximal | ||
hub to distal tip. | All devices met acceptance | |
criteria. | ||
Dimensional Inspection | The usable length, proximal and | |
distal inner and outer diameters | ||
were measured and recorded. | All device met acceptance | |
criteria. | ||
Distal Tip Buckling | Repeated distal tip buckling force | |
under compressive load was | ||
evaluated for stiffness. | Distal tip buckling force met | |
acceptance criteria. | ||
Catheter Injection Flow Rate | Device flow rate was measured | |
using ISO 10555-1, Annex E test | ||
methods with injection through | ||
proximal hub. | Injection flow rate met | |
acceptance criteria. | ||
Test | Method | Conclusions |
Catheter Suction Flow Rate | Device flow rate was measured | |
using ISO 10555-1, Annex E test | ||
methods with injection through | ||
distal tip. | Suction flow rate met acceptance | |
criteria. | ||
Vacuum Resistance | Device was evaluated for | |
resistance to vacuum collapse | ||
under static conditions with a | ||
60cc syringe. | Device was resistant to vacuum | |
collapse under static conditions. | ||
Hub Air Aspiration Leak | Device was tested for ISO | |
10555-1, Annex D hub air | ||
aspiration leak. | Hub air aspiration test met | |
acceptance criteria. | ||
Shaft Peak Tensile Force | Shaft peak tensile strength was | |
tested to failure for each joint | ||
using ISO 10555-1, Annex B test | ||
methods. | Shaft joint peak tensile force met | |
acceptance criteria. | ||
Hub Peak Tensile Force | Hub-shaft joint peak tensile | |
strength was tested to failure | ||
using ISO 10555-1, Annex B test | ||
methods | Hub-shaft joint peak tensile force | |
met acceptance criteria. | ||
Coating Integrity - Baseline | Fully assembled devices were | |
inspected under a minimum 2.5x | ||
magnification for worst-case | ||
coating defects. | Baseline coating integrity | |
characterized with worst-case | ||
defect images. Data collected as | ||
engineering reference. | ||
Coating Lubricity/Durability | Device coating was evaluated for | |
average frictional force and | ||
durability. | Coating lubricity and durability | |
testing met acceptance criteria. | ||
Particulate Testing | Device was evaluated for | |
particulate generation under | ||
simulated use in a representative | ||
tortuous anatomical model. | Number of particulates generated | |
met acceptance criteria. | ||
Coating Integrity - Simulated | ||
Use | Fully assembled devices were | |
pre-conditioned under simulated | ||
use conditions in a representative | ||
tortuous anatomical model. | ||
Coating was inspected under a | ||
minimum 2.5x magnification and | ||
subsequently tested for average | ||
frictional force. | Post-simulated use coating | |
integrity characterized with | ||
worst-case defect images. | ||
Kink Resistance | Device was wrapped around a | |
rod of known radius and | ||
inspected in-place for any kinks. | Device was resistant to kinking | |
around small radii. | ||
Liquid Leakage | Device was tested for ISO | Liquid leakage met acceptance |
Test | Method | Conclusions |
10555-1, Annex C liquid leakage | ||
testing. | criteria. | |
Static/Dynamic Burst | Device was tested under full- | |
length static conditions to burst | ||
and at pressures experienced | ||
during worst-case dynamic | ||
injections. | Static/dynamic burst testing met | |
acceptance criteria. | ||
Torque to Failure | Device was tested for full-length | |
torque strength to determine | ||
number of rotations to failure. | Torque to failure testing met | |
acceptance criteria. | ||
Physician Usability Testing | The device was navigated | |
through a tortuous benchtop | ||
model to assess compatibility | ||
with accessories, device stability, | ||
ability to aspirate, ability to inject | ||
saline or contrast, and the user's | ||
ability to navigate to the M1 and | ||
M2 segment of the MCA and | ||
retrieve a mechanical | ||
thrombectomy device. | All test results met the | |
acceptance criteria. |
9
10
Performance Testing - Animal
No animal study was performed as there is no change to the indications for use or the fundamental scientific technology for the new devices. Substantial equivalence of the Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter has been established to the predicate device through the results of bench testing.
Performance Testing - Clinical
No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the new devices. Substantial equivalence of the Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter has been established to the predicate device through the results of bench and design validation testing.