Search Results
Found 80 results
510(k) Data Aggregation
(175 days)
The VersaD Delivery Catheter is intended for use with compatible guide catheters in facilitating the insertion and guidance of catheters into selected blood vessels in the neuro and peripheral vascular systems.
The VersaD™ Delivery Catheter is a single-lumen, variable stiffness catheter with a long, tapered tip delineated by radiopaque markers. The catheter has a polytetrafluoroethylene (PTFE etched liner), the proximal end has a luer hub, and the distal portion has a hydrophilic coating to reduce friction. The delivery catheter is designed specifically for use with compatible catheters.
N/A
Ask a specific question about this device
(69 days)
The Carrier XL Delivery Catheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and to assist in the delivery of interventional devices, such as distal access catheters, in the neurovasculature.
The Carrier XL Delivery Catheter is a single lumen, variable stiffness, composite catheter. The design facilitates the advancement of the catheter and is intended to assist the delivery of interventional devices in the peripheral and neurovasculature. The outer surface of the Carrier XL Delivery Catheter is coated with a hydrophilic coating to increase lubricity. The proximal end of the Carrier XL Delivery Catheter incorporates a luer fitting for the attachment of accessories. Two radiopaque markers at the distal end help to facilitate fluoroscopic visualization. A Steam Shaping Mandrel and Peel-away Introducer Tube are included within the tray. The Carrier XL Delivery Catheter is provided sterile, non-pyrogenic, and is intended for single use only.
N/A
Ask a specific question about this device
(161 days)
The DUO Microcatheter is intended for the selective placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures.
The DUO Microcatheter is a disposable, single use, sterile device. The DUO Microcatheter is a single-lumen, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 80 cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. A steam shaping mandrel and a peel away introducer are provided with each microcatheter.
N/A
Ask a specific question about this device
(30 days)
The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media.
The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature.
The FUBUKI XF-R Neurovascular Long Sheath (FUBUKI XF-R) consists of a long sheath and a dilator. The long sheath is a single lumen neurovascular catheter designed for introduction of interventional devices, such as guidewires and other therapeutic devices. The long sheath consists of three sections: (1) a shaft, (2) a protector and (3) a connector. The distal portion of the shaft consists of a soft tip and a soft tube. The proximal part of the shaft is covered by the protector (strain relief) and the connector is bonded to the proximal end of the shaft.
The subject device is provided sterile, by ethylene oxide, and is intended for single use only by physicians who have been adequately trained in neurointerventional procedures.
The outer surface of the long sheath is coated with a hydrophilic polymer and the inner lumen of the shaft (excluding the connector portion) is lined with a fluoropolymer layer to facilitate movement of the guidewire and other devices.
The dilator consists of two parts: (1) a shaft and (2) a connector.
The FUBUKI XF-R is supplied with a dilator and rotating hemostasis valve (RHV) packed in a sterile package.
The provided FDA 510(k) Clearance Letter for the FUBUKI XF-R Neurovascular Long Sheath describes the device and its demonstrated substantial equivalence to a predicate device. However, it does not contain information about a study involving an AI/algorithm or a multi-reader multi-case (MRMC) comparative effectiveness study.
Therefore, many of the requested details regarding acceptance criteria, ground truth, expert opinions, and sample sizes for AI/algorithm performance studies cannot be extracted from this document because such a study was not conducted or reported for this submission. This is a medical device, specifically a catheter, not an AI/software device.
I will provide the information that is available in the document, framed as if it were a typical medical device clearance, rather than an AI/ML clearance.
Here's the breakdown based on the provided document:
Acceptance Criteria and Study for FUBUKI XF-R Neurovascular Long Sheath
This clearance pertains to a physical medical device (a neurovascular long sheath), not an AI/software device. Therefore, the "acceptance criteria" and "study" refer to non-clinical bench testing and biocompatibility testing to demonstrate the device's physical performance, safety, and substantial equivalence to a predicate device. There is no AI component involved in this device or its clearance documentation.
1. A table of acceptance criteria and the reported device performance:
The document states that the device met all acceptance criteria, but it does not specify the numerical acceptance criteria for each test. Instead, it lists the types of tests performed and the conclusions.
| Acceptance Criteria Category (Test Type) | Reported Device Performance / Conclusion |
|---|---|
| Non-Clinical Bench Testing | |
| Dimensional Verification | Met all acceptance criteria and performed similarly to the predicate. |
| Simulated Use | Met all acceptance criteria and performed similarly to the predicate. |
| Lubricity | Met all acceptance criteria and performed similarly to the predicate. |
| Coating Integrity / Particulate Evaluation | Met all acceptance criteria and performed similarly to the predicate. |
| Appearance | Met all acceptance criteria and performed similarly to the predicate. |
| Biocompatibility Testing | |
| Cytotoxicity (MEM Elution Test) | Non-Cytotoxic |
| Sensitization (Kligman Maximization Test) | Non-Sensitizing |
| Intracutaneous Reactivity (Intracutaneous Injection Test) | Non-Irritant |
| Acute Systemic Toxicity (System Injection Test) | Non-Toxic |
| Material Mediated Pyrogenicity (Rabbit Pyrogen Test) | Non-Pyrogenic |
| Hemocompatibility (Rabbit Blood Hemolysis Test) | Non-Hemolytic |
| Hemocompatibility (Complement Activation Test SC5b-9) | Non-Activator |
| Hemocompatibility (Partial Thromboplastin Time Test PTT) | Non-Activator |
| Thrombogenicity (In Vivo Thrombogenicity Study) | Comparable thromboresistance to the predicate device. |
| Sterilization and Shelf Life | |
| Sterilization Method | Ethylene Oxide Gas (SAL 10⁻⁶) - Unchanged from predicate, no new issues. |
| Shelf Life | 3 years (supported by accelerated aging performance testing). |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document does not specify the exact number of devices tested for each non-clinical bench or biocompatibility test. It only states that testing was "performed." For biocompatibility, it refers to standard ISO test methods (e.g., "Rabbit Pyrogen Test" implies a certain number of rabbits, but the exact count isn't given).
- Data Provenance: Not explicitly stated, but typically, non-clinical lab testing data would originate from the manufacturer's own testing facilities or contract research organizations. No geographic origin is mentioned for the data, nor is it specified if the tests were retrospective or prospective; however, given that these are physical device tests for a 510(k) submission, they would inherently be prospective (i.e., new tests conducted specifically for this submission).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable to this type of device clearance. "Ground truth" in the context of expert consensus, pathology, or outcomes data is typically reserved for diagnostic devices, particularly those involving image interpretation or clinical decision support, or AI/ML-driven devices. For a physical medical device like a catheter, "ground truth" is established through standardized engineering and biological tests (e.g., tensile strength, fluid flow, material composition, biological response), rather than expert clinical interpretation of data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This is not applicable. Adjudication methods are relevant for clinical trials or multi-reader studies where there might be disagreement in expert assessment (e.g., image interpretation). For bench and biocompatibility testing, results are quantitative or qualitative based on predefined scientific standards and validated test methods.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is specific to diagnostic devices, especially those that involve human interpretation assisted by algorithms (AI). This submission is for a physical percutaneous catheter, not a diagnostic or AI-assisted device. Therefore, a discussion of human reader improvement with AI assistance is not relevant to this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
No, a standalone algorithm performance study was not done. This device is a physical medical instrument, not an algorithm or software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this medical device is based on:
- Predefined engineering specifications and performance standards (e.g., dimensions, strength, lubricity).
- Validated test methods outlined in ISO standards (e.g., ISO 10555-1 for catheters, ISO 10993 series for biocompatibility).
- Chemical and material analysis.
- Biological responses observed in animal models (for biocompatibility).
This is fundamentally different from a ground truth established by expert clinical consensus, pathology, or outcomes data for diagnostic devices.
8. The sample size for the training set:
This is not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established:
This is not applicable. There is no "training set" or "ground truth for a training set" as this is not an AI/ML device.
Ask a specific question about this device
(59 days)
The Branchor X Balloon Guide Catheter is indicated for use to facilitate the insertion and guidance of an intravascular catheter into a selected blood vessel in the neuro vasculature, and injection of contrast media.
The balloon provides temporary vascular occlusion during these procedures.
The Branchor X Balloon Guide Catheter can also be used as a conduit for retrieval devices.
The Branchor X Balloon Guide Catheter is a variable stiffness catheter that has a radiopaque marker at the distal end of the balloon to facilitate fluoroscopic visualization and indicate the balloon position, a branched connector at the proximal end, and is equipped with a braid reinforced lumen. A balloon is attached to the distal end, and the dimensions of the balloon guide catheter and recommended balloon injection volume are provided on the product label.
The outer surface of this balloon guide catheter is coated with a hydrophilic coating for enhanced lubricity when the surface is wet. The shaft lumen is provided with PTFE coating, with the exception of the connector section. This allows the guidewire and other devices to easily move through the section.
The Branchor X Balloon Guide Catheter is packaged with a luer-activated valve, a syringe, a three-way stopcock, a rotating hemostasis valve (RHV), and a peel-away accessories.
The provided text is a 510(k) clearance letter for a medical device, the Branchor X Balloon Guide Catheter. This type of regulatory submission primarily focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving novel efficacy or safety through large-scale clinical trials.
As such, the document details non-clinical bench testing and biocompatibility testing to show that the new device performs comparably to the predicate and meets established safety standards. It explicitly states that "Animal study was not deemed necessary to demonstrate substantial equivalence" and "Clinical study was not deemed necessary to demonstrate substantial equivalence."
Therefore, the information regarding acceptance criteria and the "study that proves the device meets the acceptance criteria" will be focused on these non-clinical tests. There is no information about human-in-the-loop studies (MRMC), standalone AI performance, ground truth establishment for a training set (as there's no AI component mentioned), or expert adjudication, simply because these types of studies were not required for this particular 510(k) submission.
Here's the breakdown of the available information:
Acceptance Criteria and Device Performance Study for Branchor X Balloon Guide Catheter
The device, Branchor X Balloon Guide Catheter, demonstrates substantial equivalence to its predicate device, the Branchor Balloon Guide Catheter (K221951), through extensive non-clinical bench testing and biocompatibility assessment. The aim of these studies was to confirm that the new device met pre-established acceptance criteria, functioned as intended, and had a safety and effectiveness profile similar to the predicate.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for each test were "All samples met the acceptance criteria," indicating a 100% success rate for the tested samples against predefined engineering and safety specifications.
| Test | Test Method Summary | Reported Device Performance |
|---|---|---|
| Dimensional Verification | Device dimensions were measured to confirm they meet design specifications. | All samples met the acceptance criteria. |
| Distal Tip Visual Inspection | The distal tip was visually inspected for appropriate shape and smoothness. | All samples met the acceptance criteria. |
| Surface Visual Inspection | The catheter surface was checked for cleanliness and absence of defects. | All samples met the acceptance criteria. |
| Radio-Detectability | The device was evaluated for visibility under X-ray imaging. | All samples met the acceptance criteria. |
| Balloon Diameter to Inflation Pressure | The balloon was inflated and its diameter was measured at various inflation levels. | All samples met the acceptance criteria. |
| Freedom from Leakage and | The balloon was repeatedly inflated and deflated to check for leakage or damage. | All samples met the acceptance criteria. |
| Damage on Inflation | TBD | TBD |
| Balloon Maximum Diameter | The balloon was inflated to its maximum volume and checked for integrity. | All samples met the acceptance criteria. |
| Liquid Leakage under Pressure | The catheter was pressurized with liquid and checked for leaks. | All samples met the acceptance criteria. |
| Air Leakage into Hub Assembly during Aspiration | The hub was aspirated and checked for air ingress. | All samples met the acceptance criteria. |
| Peak Tensile Strength | Tensile force was applied to joints to assess mechanical strength. | All samples met the acceptance criteria. |
| Kink Resistance | The catheter was bent to assess resistance to kinking. | All samples met the acceptance criteria. |
| Tip Flexibility | The flexibility of the distal tip was measured. | All samples met the acceptance criteria. |
| Flow Rate | The flow rate through the catheter was measured. | All samples met the acceptance criteria. |
| Burst Pressure under Static Condition | The device was pressurized until failure to assess burst strength. | All samples met the acceptance criteria. |
| Power Injection | The device was tested for performance during high-pressure injection. | All samples met the acceptance criteria. |
| Torque Strength | The device was rotated to assess resistance to torsional stress. | All samples met the acceptance criteria. |
| Coating Integrity/Particulate | The device was tracked in a simulated anatomical model to evaluate coating integrity and particulate release. | All samples met the acceptance criteria. |
| Simulated Use | The device was used in simulated anatomical model to assess overall performance. | All samples met the acceptance criteria. |
| Connector | The connector was tested for leakage, mechanical integrity, and compatibility. | All samples met the acceptance criteria. |
Biocompatibility Testing:
| Test | Test Summary | Conclusion |
|---|---|---|
| Cytotoxicity (MEM Elution Test) | Determine potential cytotoxicity of mammalian cell culture (L929) to test article extract. | Non-cytotoxic |
| Sensitization (Maximization Test) | Allergenic/sensitizing potential evaluated using polar and non-polar extracts in guinea pig. | Non-sensitizing |
| Irritation or Intracutaneous Reactivity (Intracutaneous Injection Test) | Potential irritation effect of extract via intracutaneous injection of polar and non-polar extracts. | Non-irritant |
| Acute Systemic Toxicity (Systemic Injection) | Determine potential toxic effects of test article extract via single-dose systemic injection in mice. | Non-toxic |
| Material Mediated Pyrogenicity (Rabbit Pyrogen Test) | Determine potential presence of material-mediated pyrogen. | Non-pyrogenic |
| Hemocompatibility (Hemolysis) | Determine potential hemolytic activity in rabbit blood (direct and indirect). | Non-hemolytic |
| Hemocompatibility (Complement Activation (SC5b-9)) | Human plasma exposed directly to device to determine potential activation of complement system. | Non-activator |
| Hemocompatibility (Unactivated Partial Thromboplastin Time Assay (UPTT)) | Human plasma exposed directly to device to assess effect on intrinsic coagulation pathway by measuring clotting time. | Non-activator |
| Hemocompatibility (Thrombogenicity) | Compared thrombogenicity properties of direct blood contacting components (in vivo). | Comparable thromboresistance to commercially available comparator devices |
2. Sample Size and Data Provenance
- Sample Size: The document repeatedly states "All samples met the acceptance criteria" for non-clinical tests. However, the exact numerical sample size for each specific test (e.g., how many catheters were tested for burst pressure) is not specified in this 510(k) summary.
- Data Provenance: The device manufacturer is ASAHI INTECC CO., LTD., located in Japan, with a US presence. The testing is non-clinical bench testing and refers to ISO standards. The data is not from human patients and therefore does not have a country of origin in the typical sense (e.g., patient demographics). All testing appears to be prospective as it was conducted specifically to support this 510(k) submission.
3. Number of Experts and Qualifications for Ground Truth
- Not Applicable. This device is a physical medical instrument, not a diagnostic or AI-driven software. The "ground truth" for its performance is established through engineering and material science standards and physical measurements, not interpretation by human experts (e.g., radiologists interpreting images).
4. Adjudication Method for the Test Set
- Not Applicable. As the tests are non-clinical, objective measurements against engineering specifications, there is no need for expert adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not Applicable. No MRMC study was conducted. This type of study involves human readers (e.g., radiologists) evaluating cases, often with and without AI assistance, to assess diagnostic performance. This device is a catheter and does not involve image interpretation or AI assistance for human readers.
6. Standalone (Algorithm Only) Performance Study
- Not Applicable. This device is a physical catheter, not an algorithm or AI software. Therefore, no standalone algorithm performance study was performed or is relevant.
7. Type of Ground Truth Used
- The "ground truth" refers to engineering design specifications, material science standards (e.g., tensile strength, burst pressure limits), and relevant international standards (ISO standards) for medical device performance and biocompatibility. For instance, for dimensional verification, the ground truth is the specified engineering drawing dimensions. For biocompatibility, the ground truth is the absence of toxic, sensitizing, irritant, or pyrogenic effects as defined by ISO 10993 series.
8. Sample Size for the Training Set
- Not Applicable. There is no AI component or machine learning model that requires a training set for this device.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As there is no training set, this question is not relevant.
Ask a specific question about this device
(50 days)
The 93 NeuFlex™ Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
The Piraeus Medical 93 NeuFlex™ Catheter is a single lumen guide catheter that provides access to peripheral, coronary and neuro vasculature. The catheter is comprised of a hollow cylindrical tube bonded at the proximal end to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers. The catheter features a hydrophilic coating to enhance tracking through tortuous vasculature and facilitate the introduction of interventional devices to the peripheral, coronary and neuro vasculature. The 93 NeuFlex™ guide catheter is compatible with 0.035" or smaller guidewires. The distal soft tip facilitates tracking past vessel branches. A radiopaque marker provides visual confirmation of the distal tip location under fluoroscopy. The 93 NeuFlex™ Catheter has an inner diameter of 0.093" (compatible with 6F outer diameter catheters), and a maximum outer diameter of 0.110". The catheter is offered in working lengths of 110 cm and 100 cm. The 93 NeuFlex™ guide catheter is packaged with a loading tool to aid in insertion of the catheter into a short sheath.
Accessory devices required, but not supplied include:
- Guidewires
- Support/diagnostic catheters
- Introducer sheaths
- Rotating hemostasis valves (RHVs)
The provided text describes the FDA 510(k) clearance for the "93 NeuFlex Catheter." This is a medical device clearance document for a physical catheter, not a software-based artificial intelligence/machine learning (AI/ML) device. Therefore, the "acceptance criteria" and "study proving the device meets the acceptance criteria" in this document relate to the physical and functional performance of the catheter itself, primarily through non-clinical bench testing, rather than AI/ML performance metrics like sensitivity, specificity, or AUC, which would be derived from a clinical study with human readers or standalone algorithm performance.
The questions you've posed (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set ground truth) are highly specific to AI/ML software validation studies. Since the provided document is for a physical medical device (a catheter) and relies on bench testing for substantial equivalence, most of these questions are not applicable to the data provided.
However, I can extract the relevant "acceptance criteria" (in the context of bench testing) and the "reported device performance" from the document.
Acceptance Criteria and Reported Device Performance for 93 NeuFlex Catheter
The acceptance criteria for this physical medical device are implied by the "Pass" results in the performance testing. These are not statistical metrics related to diagnostic accuracy, but rather engineering and functional performance criteria.
1. Table of Acceptance Criteria and Reported Device Performance
| Test (Imacceptance Criteria) | Test Method Summary (Standard/Guidance) | Reported Device Performance |
|---|---|---|
| Visual Inspection: Catheter free of surface defects, extraneous matter; distal tip smooth, tapered/rounded. | N/A | Pass |
| Dimensional Verification: Catheter outer diameter, inner diameter, usable length, coating length within specifications. | N/A | Pass |
| Delivery, Compatibility, and Retraction (Trackability): Able to be prepped, inserted, tracked, retracted, and removed per IFU in simulated use without damage. | ISO 10555-1, FDA guidance PTA and Specialty Catheters | Pass |
| Flexibility and Kink Resistance: No kinking of shaft (permanent deformation) after simulated use. | FDA guidance PTA and Specialty Catheters | Pass |
| Compatibility with Other Devices (External): Compatible with 8F introducer (short) sheath. | ISO 10555-1, FDA guidance PTA and Specialty Catheters | Pass |
| Guidewire Compatibility: Able to be delivered over maximum size guidewire indicated in labeling. | ISO 10555-1, FDA guidance PTA and Specialty Catheters | Pass |
| Interventional Device Compatibility (Internal): Able to accommodate other interventional devices (e.g., support catheter, diagnostic catheter) up to maximum size indicated in labeling. | ISO 10555-1, FDA guidance PTA and Specialty Catheters | Pass |
| Accessory Compatibility: Compatible with accessories such as RHVs and the accessory loading tool. | N/A | Pass |
| Catheter Bond Strength: Sufficient tensile bond strength to remain intact throughout a procedure. | ISO 10555-1, FDA guidance PTA and Specialty Catheters | Pass |
| Freedom from Leakage – Positive Pressure: No liquid leakage from hub or catheter shaft at 317 kPa (46 psi) for 30 seconds. | ISO 10555-1 | Pass |
| Freedom from Leakage – Negative Pressure: No air leakage into a 20cc syringe when vacuum pulled for 15 seconds. | ISO 10555-1 | Pass |
| Kink Resistance: No kinking of catheter shaft (permanent deformation) after wrapping around anatomically relevant bend radii. | FDA guidance PTA and Specialty Catheters | Pass |
| Pushability: Proximal shaft has sufficient stiffness for easy pushing to target anatomy without buckling. | ISO 10555-1, FDA guidance PTA and Specialty Catheters | Pass |
| Access Force: Does not require excessive force to navigate and track to target anatomy. | ISO 10555-1, FDA guidance PTA and Specialty Catheters | Pass |
| Static Burst: Static burst pressure compared with maximum pressure generated with manual syringe injection. | ISO 10555-1 | Pass |
| Torque Strength: Navigated through clinically relevant tortuous path; proximal end rotated until failure without issues. | FDA guidance PTA and Specialty Catheters | Pass |
| Distal Tip Flexibility: Compressive force exerted by side of catheter at defined distance from distal tip (when tip deflected) compared to predicate. | N/A | Pass |
| Particulate: Amount of particulate matter generated during simulated use compared to predicate. | FDA guidance PTA and Specialty Catheters | Pass |
| Coating Lubricity, Durability, and Integrity: Frictional forces measured (pinch test); hydrophilic coating examined for damage after simulated use. | N/A | Pass |
| Corrosion: Metallic components contacting fluid path evaluated for corrosion resistance. | ISO 10555-1 | Pass |
Inapplicable AI/ML Specific Questions:
The following questions are geared towards AI/ML software performance studies. Since this FDA clearance is for a physical medical device (a catheter) and the supporting evidence is non-clinical bench testing, these questions are not applicable in the context of the provided document.
- Sample sizes used for the test set and the data provenance: Not applicable. The "test set" here refers to physical catheter samples tested in benchtop models, not a dataset of medical images or patient records. The provenance is internal company testing.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical device performance is established through adherence to engineering specifications and standardized test methods, not expert consensus on medical findings.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods are used in interpreting medical images or data where human variability in judgment exists. This is a bench test for a physical device.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the catheter's performance is derived from standardized engineering and material science tests, often defined by international standards (e.g., ISO 10555-1) and FDA guidance for catheter performance, rather than clinical outcomes or expert medical consensus on diagnostic findings.
- The sample size for the training set: Not applicable. This is not an AI/ML device that requires a training set.
- How the ground truth for the training set was established: Not applicable.
Summary of the Study Proving Device Meets Criteria (Based on Provided Text):
The manufacturer, Piraeus Medical, conducted a series of non-clinical bench performance tests on the 93 NeuFlex Catheter. These tests were performed per company-approved protocols, test methods, and performance standards, many of which reference international standards like ISO 10555-1 and FDA guidance for PTA and Specialty Catheters. The results of all listed tests indicated "Pass," meaning the device met the predefined performance specifications for each test. This extensive bench testing, along with comparison to a predicate device (87 NeuGlide Catheter, K240971) which shares similar intended use and basic technological characteristics, formed the basis for demonstrating substantial equivalence and thus FDA clearance. Biocompatibility and sterilization were leveraged from the predicate device's existing validations. Shelf-life was supported by accelerated aging based on ASTM F1980, also leveraging the predicate's packaging configuration.
Ask a specific question about this device
(270 days)
PATH BGC is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the neurovascular system. The balloon provides temporary vascular occlusion during these and other angiographic procedures.
The PATH Balloon Guide Catheter (PATH BGC) is a dual coaxial lumen catheter consisting of an inner coil reinforced variable stiffness lumen and an outer braid reinforced variable stiffness lumen. A radiopaque marker is included at the tip of the catheter and at the distal and proximal ends of the balloon. A compliant balloon is mounted near the distal end of the catheter to provide vascular occlusion during angiographic procedures. The catheter has hydrophilic coating at the distal and proximal end. A bifurcated luer hub on the proximal end allows attachments for flushing and balloon inflation.
The provided document is a 510(k) Clearance Letter for a medical device (PATH BGC) and does not contain information about a study proving the device meets acceptance criteria related to AI/algorithm performance against a clinical ground truth.
Instead, this document describes the device's physical characteristics, intended use, comparison to a predicate device, and bench testing, biocompatibility testing, and shelf-life testing to demonstrate performance and safety. These tests are common for physical medical devices and confirm that the device meets pre-defined engineering specifications, not clinical accuracy in diagnosis or treatment based on AI.
Therefore, I cannot fulfill the request for information regarding AI acceptance criteria and study details based on the provided text. The document explicitly states:
- Clinical Study: "Clinical study was not deemed necessary to demonstrate substantial equivalence." (Page 11)
- Animal Study: "Animal study was not deemed necessary to demonstrate substantial equivalence." (Page 10)
The "acceptance criteria" discussed in the document refer to engineering specifications and performance metrics for a physical catheter, such as dimensional accuracy, kink resistance, lubricity, mechanical strength, and biocompatibility.
To answer your request, if this were an AI-powered device, the information you're asking for would typically be found in a separate section detailing the AI algorithm's validation study, which is absent here.
Here's a breakdown of why each point of your request cannot be answered from the provided text:
-
A table of acceptance criteria and the reported device performance:
- Not applicable for AI performance. The table provided (Table 2) lists bench tests for the physical catheter (e.g., Visual Inspection, Dimensional Inspection, Kink Resistance, Coating Lubricity, Radiopacity, etc.). The "results" column simply states "Pass – all samples met the pre-determined acceptance criteria," without providing specific numerical performance data against a clinical ground truth.
-
Sample sizes used for the test set and the data provenance:
- Not applicable for AI performance. The "samples" referred to in Table 2 are physical units of the catheter device, not data sets for an AI algorithm. There is no mention of data provenance for AI.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. There is no AI component, and thus no need for experts to establish a "ground truth" for clinical interpretation using the device.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This refers to clinical adjudication of AI outputs, which is not present.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. No MRMC study was performed as no AI assistance is described.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. No standalone algorithm performance is described.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable. No clinical ground truth is established as there is no AI performing diagnostics.
-
The sample size for the training set:
- Not applicable. No AI training set is mentioned.
-
How the ground truth for the training set was established:
- Not applicable. No AI training and ground truth establishment are mentioned.
In summary, the provided document describes a physical medical device (a catheter) and its clearance based on bench testing and biocompatibility, not an AI or software as a medical device (SaMD) that would require the type of clinical performance validation you are inquiring about.
Ask a specific question about this device
(231 days)
The Excelsior XT-27 Microcatheter and Excelsior XT-27 Flex Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
The Excelsior XT-27 Pre-Shaped Microcatheter and Excelsior XT-27 Flex Pre-Shaped Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
The subject devices hereafter referred to as Excelsior® XT-27® Microcatheters are sterile, single lumen, 0.027 inch inner diameter (ID) microcatheters with one tip marker designed to aid the physician in accessing distal neurovasculature when used with a guide catheter and steerable guidewire. Graded shaft stiffness ranging from a highly flexible tip to a semi-rigid proximal section aids the physician in tracking over selectively placed guidewires. A luer fitting located on the microcatheter hub is used for the attachment of accessories. One radiopaque tip marker is positioned at the distal tip of the device to facilitate fluoroscopic visualization® XT-27® Microcatheters are coated on the outer surface with Hydrolene™ coating which reduces friction during manipulation in the vessel.
The Excelsior® XT-27® Microcatheters are available in effective lengths of both 135 cm (53.1 inch) and 150 cm (59.1 inch), with two distal shaft configurations achieved through distal shaft lengths of 6 cm (XT-27 model) and 18 cm (XT-27 Flex model). Both straight tip and pre-shaped versions are available.
The provided text is a 510(k) summary for a medical device (Excelsior XT-27 Microcatheters) and details its acceptance criteria and the study performed to demonstrate substantial equivalence, but it does not describe an AI/ML device. Therefore, the questions related to AI/ML specific criteria (such as sample sizes for test and training sets, number of experts for ground truth, adjudication methods, multi-reader multi-case studies, standalone performance, and data provenance) are not applicable to this document.
However, I can provide the acceptance criteria and study information that is available for this mechanical microcatheter from the document.
1. Table of acceptance criteria and the reported device performance:
| Test | Acceptance Criteria (Inferred from "Pass") | Reported Device Performance |
|---|---|---|
| Guide Catheter Compatibility with 1.17 mm (0.046") ID | Maximum forces required to completely deliver and retrieve the microcatheter inside a 0.046" ID guide catheter with worst-case sized ancillary devices must be acceptable/pass. | Pass |
| Tensile Strength for Joints and Marker Band | Tensile strength after preconditioning by simulated use with worst-case sized ancillary and interventional devices must be acceptable/pass. | Pass |
| Particulate and Coating Integrity | Hydrophilic coating integrity and particulate generation under simulated use conditions with a 0.046" ID guide catheter, and comparison to cleared comparator devices, must be acceptable/pass. | Pass |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size for Test Set: The document does not specify the exact number of samples (e.g., number of microcatheters tested) used for each bench test. It mentions testing with "ancillary devices that represent worst-case sizes" for the Guide Catheter Compatibility test and "worst-case sized ancillary devices and interventional devices" for the Tensile Strength test.
- Data Provenance: The study is described as "Bench testing," indicating it was conducted in a laboratory setting. There is no information regarding country of origin or whether it was retrospective or prospective, as these terms are typically applied to clinical studies involving human or animal subjects.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not applicable as this is a mechanical device performance study, not an AI/ML diagnostic or prognostic study requiring expert opinion for ground truth. The "ground truth" here is the physical performance of the device against engineering specifications.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable for the same reason as point 3. Bench testing results are typically adjudicated by meeting predefined engineering specifications and criteria, not by expert consensus in the same way as diagnostic reads.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for AI-assisted diagnostic devices, which is not what this device is.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is a physical medical device (microcatheter), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" in this context is based on engineering specifications and performance criteria for the physical device. For example, for tensile strength, the ground truth would be a defined minimum force the joints and marker band must withstand without failure. For compatibility, it would be the ability to successfully deliver and retrieve the microcatheter within the specified guide catheter with acceptable force.
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device, so there is no training set.
9. How the ground truth for the training set was established:
- Not applicable for the same reason as point 8.
Ask a specific question about this device
(90 days)
The AXS Lift Intracranial Base Catheter is introduction of interventional devices into the peripheral and neurovasculature.
The AXS Lift Intracranial Base Catheter ("AXS Lift") is a single lumen, variable stiffness catheter designed for use in facilitating the insertion and guidance of appropriately sized interventional devices into the peripheral and neurovasculature. The distal catheter shaft has a hydrophilic coating to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end. The distal end of the catheter contains bovine derived tallow. It is packaged with one double port rotating hemostasis valve (RHV) and one peel-away introducer sheath. The peel-away introducer sheath is designed to protect the distal tip of the catheter during insertion into the short introducer sheath valve. The AXS Lift Intracranial Base Catheter is compatible with short introducer sheaths with an inner diameter of 7Fr or greater. AXS Lift is supplied sterile, non-pyrogenic, and is intended for single use only.
The provided text, K243593 for the AXS Lift Intracranial Base Catheter, describes the predicate device equivalence testing rather than a study proving the device meets acceptance criteria for an AI/ML algorithm. The document focuses on the mechanical and material performance of a catheter, not the performance of a software algorithm.
Therefore, the requested information regarding acceptance criteria, study design for AI/ML performance (sample size, data provenance, expert ground truth, adjudication, MRMC, standalone performance, training set details), cannot be extracted from this document as it does not contain information about an AI/enabled medical device.
The document definitively states on page 9:
"No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes."
This reinforces that the approval is based on physical and material device characteristics, not an AI/ML algorithm's performance on clinical data or human-in-the-loop studies.
Ask a specific question about this device
(189 days)
The VersaD™ Delivery Catheter is intended for use with compatible guide catheters in facilitating the insertion and guidance of catheters into selected blood vessels in the neuro and peripheral vascular systems.
The VersaD™ Delivery Catheter is a single-lumen, variable stiffness catheter with a long, tapered tip delineated by radiopaque markers. The proximal end has a luer hub, and the distal portion has a hydrophilic coating to reduce friction. The delivery catheter is designed specifically for use with compatible catheters.
The provided text describes the acceptance criteria and the study that proves the Unity Medical VersaD™ Delivery Catheter meets these criteria. It applies to a medical device, not an AI/ML product, so some of the requested information (like multi-reader multi-case studies, expert ground truth for AI, training sets, etc.) is not applicable and thus cannot be provided from this document.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this medical device are primarily defined by successful completion of various performance tests, biocompatibility tests, and sterilization/shelf-life validations. The "Conclusion" column in the tables below indicates whether the device met these criteria.
Performance Testing - Bench
| Test | Test Method Summary | Acceptance Criteria (Implicit) | Reported Device Performance (Conclusion) |
|---|---|---|---|
| Surface Integrity | The surface of the catheter was evaluated for surface defects. | No surface defects present. | The device met the established criteria. |
| Coating Uniformity | The surface of the catheter was evaluated for coating defects and voids. | No coating defects or voids present. | The device met the established criteria. |
| Device Compatibility | A full-length silicone neurovascular model was used to simulate use. The procedure included using a test catheter with worst-case ancillary devices. | Compatible with ancillary devices in simulated use. | The device was found to be compatible with ancillary devices. |
| Push/Track | Simulated use was performed, and the catheter was tracked through a challenge neurovascular model to determine the deliverability of the subject catheter and compatibility with ancillary devices. | Performed as intended under simulated use conditions. | The device performed as intended under simulated use conditions. |
| Torque Response | The catheter was constrained, torqued, and evaluated. | Maintained integrity and functionality under torque. | The device met the established criteria. |
| Particulate | The catheter was tracked through the tortuous neurovascular model with ancillary devices, and particulate generation was measured. | Particle generation similar to predicate device. | Device particulate generation was similar to the predicate device. |
| Coating Adhesion | The surface of the catheter was evaluated for coating voids after simulated use. | No coating voids after simulated use. | The device met the established criteria. |
| Tip Inspection | The distal tip of the catheter was evaluated. | Distal tip met design specifications. | The device met the established criteria. |
| Dimensional Verification | The catheter dimensional attributes were evaluated and measured. | Dimensions met specified requirements. | The device met the established criteria. |
| Kink Resistance | The proximal and distal sections of the catheter were evaluated using a radius apparatus. | Resistant to kinking under specified conditions. | The device met the established criteria. |
| Liquid Leak Under Pressure | The catheter was evaluated by holding hydrostatic pressure. | No liquid leak under pressure. | The device met the established criteria. |
| Air Leak Under Aspiration | The catheter was evaluated for air leakage into the hub assembly during aspiration. | No air leakage into the hub assembly during aspiration. | The device met the established criteria. |
| Burst Strength | The catheter was pressurized to burst with fluid. | Withstood pressure without bursting up to a specified limit. | The device met the established criteria. |
| Tip Stiffness | The tip buckling force of the catheter was evaluated in comparison with the predicate. | Tip buckling force similar to the predicate. | The tip buckling force was similar to the predicate. |
| Tensile Strength (Tip, Mid Joints, and Hub) | The force required to separate the joints in the catheter was evaluated. | Joints maintained integrity under specified tensile force. | The device met the established criteria. |
| Corrosion | The catheter was evaluated for corrosion. | No corrosion observed. | No corrosion was noted on the catheter. |
| Radiopacity | The catheter marker bands visibility were evaluated under fluoroscopy. | Marker bands visible under fluoroscopy. | The device was visible under fluoroscopy. |
| Delivery and Retrieval Force | The forces to deliver and retrieve the catheter within a neurovascular model were measured. | Delivery and retrieval forces comparable to predicate device. | Delivery and retrieval forces were comparable to the predicate device. |
Biocompatibility
| Test Category | Method | Acceptance Criteria (Implicit) | Result (Reported Device Performance) |
|---|---|---|---|
| Cytotoxicity | MEM Elution with L-929 Cells (ISO 10993-5) | No cytotoxic effects. | Non-Cytotoxic |
| Irritation | Intracutaneous Reactivity Test (ISO 10993-23) | No irritant reaction. | Non-Irritant |
| Sensitization | Guinea Pig Maximization Sensitization Test (ISO 10993-10) | No sensitizing reaction. | Non-Sensitizing |
| Systemic Toxicity | Acute Systemic Injection Test (ISO 10993-11) | No evidence of acute systemic toxicity. | No evidence of Acute Systemic Toxicity |
| Material-Mediated Pyrogenicity | Rabbit Pyrogen test (USP <151>) | Non-pyrogenic. | Non-Pyrogenic |
| Complement Activation | SC5b-9 Assay (with comparator control) | No complement activation. | Non- Activator |
| Thrombogenicity - In Vitro Blood Flow Loop Assay | In vitro blood flow loop (ISO 10993-4) | Thromboresistant. | Thromboresistant |
| Thrombogenicity - Coagulation | Partial Thromboplastin Time (PTT) Assay Human Blood | No significant effect on PTT. | No effect on the PTT |
| Thrombogenicity - Platelet Leukocyte | Direct contact Platelet Leukocyte Count Test | No significant effect on platelet and leukocyte count. | No effect on the platelet and leukocyte count |
| Hemolysis | Hemolysis Test (ASTM F756) | Non-hemolytic. | Non- Hemolytic |
Sterilization and Shelf Life
| Test Category | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Sterilization | Minimum sterility assurance level (SAL) of 10⁻⁶. | EO sterilized to achieve a minimum sterility assurance level (SAL) of 10⁻⁶. |
| Shelf Life (12-month) | Product and packaging remain functional and sterile. | Product and packaging remain functional and sterile for 12 months. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each bench test or biocompatibility test. It indicates that testing was performed to demonstrate substantial equivalence.
Regarding data provenance:
- Country of Origin: Not specified. Standard medical device testing labs and animal models would typically be used.
- Retrospective or Prospective: All testing described is prospective bench and lab testing conducted specifically for this 510(k) submission. It's not based on retrospective clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable as the device is a physical medical catheter, not an AI/ML product requiring expert annotation for ground truth. The "ground truth" for this device is defined by passing established engineering specifications, biocompatibility standards, and sterility validations.
4. Adjudication Method for the Test Set
This is not applicable for a physical medical device. Testing results are typically quantitatively measured or visually assessed against predefined acceptance criteria by qualified personnel, not "adjudicated" by multiple experts in a consensus manner as would be done for image annotation in AI/ML.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC study was not conducted because this is a physical delivery catheter, not an AI-assisted diagnostic device. The study described focuses on the physical and biological performance of the catheter itself, demonstrating substantial equivalence to a predicate device through bench, biocompatibility, and sterility testing.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not Applicable. This is a physical medical device, not an algorithm or software. The "standalone" performance refers to the engineering and biological performance of the catheter independent of human user skill in the tests performed (e.g., measuring burst strength of the catheter itself).
7. The Type of Ground Truth Used
The "ground truth" for this medical device is established by:
- Engineering Specifications/Standards: The device must meet predefined dimensional tolerances, mechanical strength requirements (e.g., tensile strength, burst strength), and functional performance (e.g., push/track, torque response, liquid/air leak).
- Biocompatibility Standards (ISO 10993): Adherence to established biological safety profiles (e.g., non-cytotoxic, non-irritant, non-sensitizing).
- Sterilization Standards: Achieving a specified sterility assurance level.
- Comparison to Predicate Device: Many tests (e.g., particulate generation, tip stiffness, delivery/retrieval force) compare the subject device's performance to that of the legally marketed predicate device to demonstrate substantial equivalence.
It does not involve expert consensus, pathology, or outcomes data in the context of diagnostic interpretation, as it's a delivery catheter.
8. The Sample Size for the Training Set
Not Applicable. This is a physical medical device, not an AI/ML system, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As there is no training set for a physical medical device, the concept of establishing ground truth for it does not apply.
Ask a specific question about this device
Page 1 of 8