K Number
K182097
Date Cleared
2018-11-14

(103 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The React™ 71 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

Device Description

The React™ 71 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with various durable polymer outer layers. A lubricious, polytetrafluoroethylene liner and Engage™ inner layer is used to create a structure that has both proximal stiffness and distal flexibility. The React™ 71 Catheter is also designed with an encapsulated radiopaque distal platinum-iridium marker band which is used for visualization under fluoroscopy. The React™ 71 Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. The proximal end of the React™ 71 Catheter is designed with a green thermoplastic elastomer strain relief and a clear hub. The distal end of the React™ 71 Catheter is coated with a Surmodics Serene™ Coating.

AI/ML Overview

The provided document describes the React™ 71 Catheter. Based on the content, here's a breakdown of the acceptance criteria and the study performed:

1. Table of Acceptance Criteria and Reported Device Performance

The document lists various tests, their methods, and states that the React™ 71 Catheter "met the acceptance criteria" or similar statements for each. Specific quantitative acceptance criteria are generally not provided in this summary, but the results confirm compliance.

Test DescriptionTest Method SummaryAcceptance Criteria (Implicit)Reported Device Performance
Biocompatibility
Cytotoxicity (Elution Method)The test article extract showed no evidence of causing cell lysis or toxicity and had a Grade 0 (No Reactivity).Grade < 2 (Mild Reactivity)Met (Grade 0: No Reactivity), considered non-cytotoxic.
Sensitization (Guinea Pig Maximization Test)The test article extracts showed no evidence of causing delayed dermal contact sensitization in the guinea pig.No evidence of sensitizationMet, does not elicit a sensitization response.
Irritation (Intracutaneous Reactivity)The test article met the requirements of the test since the difference between each test article extract overall mean score and corresponding control extract overall mean score was 0.0 and 0.2 for the Sodium Chloride and Sesame Oil test article extracts, respectively.Difference in mean score ≤ a specified threshold (implied to be low)Met, considered a non-irritant.
Acute Systemic Toxicity (Systemic Toxicity)There was no mortality or evidence of systemic toxicity from the extracts injected into mice.No mortality or systemic toxicityMet, does not indicate signs of toxicity.
Hemocompatibility (Hemolysis)The hemolytic index for the test article in direct contact with blood was 0.8%, and the hemolytic index for the test article extract was 0.0%.Hemolytic index below a specified thresholdMet (0.8% and 0.0%), considered non-hemolytic.
Hemocompatibility (Complement Activation)The concentration of SC5b-9 in the test article was not statistically higher than the activated normal human serum control or the negative control.SC5b-9 concentration not statistically higher than controlsMet, not considered a potential activator of the complement system.
Hemocompatibility (Thrombogenicity – Canine Model)The control article had a mean score of 2.0. The test article had a mean score of 1.7 after four (4) hours (±30 minutes) without systemic anticoagulation.Lower thrombogenic potential than predicate/controlMet, demonstrates lower thrombogenic potential compared to ACE 68 Reperfusion Catheter.
Pyrogenicity (Material Mediated)Not a single animal showed a temperature rise of 0.5°C or more above its baseline temperature. The total rise of the rabbits' temperature during the three (3) hours was 0.0°C.Temperature rise < 0.5°C above baselineMet, considered non-pyrogenic.
Bench Performance (Leveraged Testing)
Ethylene Oxide ResidualEvaluated per ISO 10993-7.Met acceptance criteria for residualsMet the acceptance criteria.
Ethylene Chlorohydrin ResidualEvaluated per ISO 10993-7.Met acceptance criteria for residualsMet the acceptance criteria.
Bioburden RecoveryEvaluated per ISO 11737-1.Met acceptance criteria for recoveryMet the acceptance criteria.
Bacterial EndotoxinEvaluated per ANSI/AAMI ST72 and USP <161>.Met acceptance criteria for endotoxin levelsMet the acceptance criteria.
Packaging Visual InspectionEvaluated per ASTM F1886.Met acceptance criteria for visual integrityMet the acceptance criteria.
Bubble LeakEvaluated per ASTM F2096.Met acceptance criteria for leak integrityMet the acceptance criteria.
Seal StrengthEvaluated per ASTM F88.Met acceptance criteria for seal strengthMet the acceptance criteria.
Bench Performance (Conducted Testing)
BioburdenEvaluated per ISO 11737-1.Met acceptance criteria for bioburdenMet the acceptance criteria.
Visual InspectionInspected under x2.5 magnification.Met acceptance criteria for visual integrityMet the acceptance criteria.
Dimensional MeasurementsProximal ID, distal ID, proximal OD, distal OD, usable length, total length, coating length, and distal tip length measured.Met acceptance criteria for specified dimensionsMet the acceptance criteria.
Tip BucklingEvaluated for maximum compressive force.Met acceptance criteria for tip buckling forceMet the acceptance criteria.
Kink ResistanceEvaluated for maximum kink diameter.Met acceptance criteria for kink diameterMet the acceptance criteria.
ParticulateEvaluated per USP <788>.Met acceptance criteria for particulate levelsMet the acceptance criteria.
Coating LubricityEvaluated for average frictional forces.Met acceptance criteria for frictional forcesMet the acceptance criteria.
Tensile StrengthEvaluated per ISO 10555-1. Annex B.Met acceptance criteria for tensile strengthMet the acceptance criteria.
Liquid LeakEvaluated per ISO 10555-1. Annex C.Met acceptance criteria for liquid leakMet the acceptance criteria.
Corrosion ResistanceEvaluated per ISO 10555-1. Annex A.Met acceptance criteria for corrosionMet the acceptance criteria.
Hub Aspiration ResistanceEvaluated per ISO 10555-1. Annex D.Met acceptance criteria for hub air aspirationMet the acceptance criteria.
RadiopacityMarkerband length and wall thickness measured; confirmed via fluoroscopy.Met acceptance criteria for radiopacityMet the acceptance criteria.
Luer StandardsEvaluated per ISO 594-1 and ISO 80369-7.Met acceptance criteria for luer standardsMet the acceptance criteria.
CompatibilityInspected for visual damage when delivering and retrieving interventional devices.Met acceptance criteria for no visual damage/functionalityMet the acceptance criteria.
Torque to FailureEvaluated for transmission of proximal torque to the distal tip.Withstood typical clinical torsional forcesWas able to withstand torsional forces typical of clinical use.
Dynamic PressureEvaluated for the amount of pressure it can withstand.Withstood typical clinical pressuresWas able to withstand pressures typical of clinical use.
Coating IntegrityEvaluated for coating coverage and lubricity.Remained coated and lubriciousRemained coated and lubricious.
UsabilityEvaluated for maneuverability and flexibility against a predicate device.Met acceptance criteria for maneuverability and flexibility.Met the acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify exact sample sizes for each test. For the biocompatibility study on thrombogenicity, it mentions a "Canine Model" with "four (4) hours (±30 minutes) without systemic anticoagulation," and comparing a "control article" to the "test article." This implies a very small comparative animal study, rather than a large test set for AI evaluation.

The provenance of data (e.g., country of origin, retrospective/prospective) is not provided as the tests are primarily bench and animal studies (biocompatibility) conducted in controlled laboratory settings for device performance and safety.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This is a medical device submission based on product performance and biocompatibility testing, not an AI/software submission requiring expert ground truth for classification or diagnosis.

4. Adjudication Method for the Test Set

Not applicable. As this is not an AI/software submission involving human interpretation or adjudication for ground truth, this information is not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No. This is a medical device submission for a catheter, not an AI system that would typically undergo MRMC studies.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is a physical medical device, not an algorithm. The "standalone" performance here refers to the device's physical and biological performance in bench and animal tests.

7. The Type of Ground Truth Used

The "ground truth" for the tests performed is based on established scientific and engineering standards and methods. For example:

  • Biocompatibility: In vitro and in vivo testing (e.g., cell cultures, guinea pigs, mice, canine model) following ISO 10993 standards.
  • Performance Data – Bench: Standardized laboratory tests (e.g., ISO, ASTM, USP) for physical, mechanical, and material properties.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device. There is no concept of a "training set" as understood in AI/machine learning. The device design and manufacturing processes are refined through engineering development, not machine learning training.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for an AI model. For the device itself, the "ground truth" for its design and performance requirements is established through regulatory standards, predicate device characteristics, and engineering specifications.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 14, 2018

Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Ryan Kenney Senior Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618

Re: K182097

Trade/Device Name: React™ 71 Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 12, 2018 Received: October 15, 2018

Dear Ryan Kenney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Xiaolin Zheng -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182097

Device Name React™ 71 Catheter

Indications for Use (Describe)

The React™ 71 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K182097

510(k) Owner:Micro Therapeutics, Inc. d/b/a ev3 Neurovascular9775 Toledo WayIrvine, CA 92618Establishment Registration No.: 2029214
Contact Person:Ryan KenneySenior Regulatory Affairs SpecialistTelephone: (949) 297-5489Email: ryan.j.kenney@medtronic.com
Date Summary Prepared:October 12, 2018
Trade Name of Device:React™ 71 Catheter
Common Name of Device:Percutaneous Catheter
Review Panel:Neurology
Product Code:DQY
Regulation Number:21 CFR 870.1250
Regulation Name:Percutaneous Catheter
Device Classification:Class II
Primary Predicate Device:AXS Catalyst™ Distal Access Catheter510(k)#: K151667
Additional Predicate Devices:ReFlex™ Guide Catheter510(k)#: K110055React™ 68 Catheter510(k)#: K180715

Device Description:

The React™ 71 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with various durable polymer outer layers. A lubricious, polytetrafluoroethylene liner and Engage™ inner layer is used to create a structure that has both proximal stiffness and distal flexibility. The React™ 71 Catheter is also designed with an encapsulated radiopaque distal platinum-iridium marker band which is used for visualization under fluoroscopy. The React™ 71 Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. The proximal end of the React™ 71 Catheter is designed with a green thermoplastic elastomer strain relief and a clear hub. The distal end of the React™ 71 Catheter is coated with a Surmodics Serene™ Coating.

Indications for Use:

The React™ 71 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

{4}------------------------------------------------

Device Comparison:
AXS Catalyst™ DistalAccess Catheter(K151667)ReFlex™ GuideCatheter(K110055)React™ 68 Catheter(K180715)React™ 71 Catheter(K182097)
Indication for Use(IFU) StatementThe AXS Catalyst™Distal Access Catheteris indicated for theinsertion and guidanceof appropriately sizedinterventional devicesinto a selected bloodvessel in theperipheral andneurovascularsystems. The AXSCatalyst™ DistalAccess Catheter isalso indicated for useas a conduit forretrieval devices.The ReFlex™ GuideCatheter is indicatedfor the introductionof interventionaldevices into theperipheral and neurovasculature.The React™ 68Catheter is indicatedfor the introductionof interventionaldevices into theperipheral and neurovasculature.The React™ 71Catheter is indicatedfor the introductionof interventionaldevices into theperipheral and neurovasculature.
Materials
HubNylonTrogamid®Same as K110055Same as K110055
Strain ReliefThermoplastic RubberSantoprene™DynaFlex®Same as K182097
Inner LayerPTFESame as K151667Same as K151667PTFE,Polyolefin
ReinforcementStainless Steel withNitinol and PolymerFiberNitinolSame as K110055Same as K110055
Outer JacketPebax®PolymericGrilamid™Pebax®Polyamide,Polyolefin,Polyurethane
Marker BandPlatinum/IridiumSameSameSame
AXS Catalyst™ DistalAccess Catheter(K151667)ReFlex™ GuideCatheter(K110055)React™ 68 Catheter(K180715)React™ 71 Catheter(K182097)
AdhesiveCyanoacrylateSameSameSame
CoatingHydrophilicSameSameSame
Dimensions
Working Length132 cm130 cmSame as K151667Same as K151667
Inner Diameter0.060"0.072"0.068"0.071"
Proximal OuterDiameter (OD)0.079"0.084"0.083"0.0855" (Max)
Distal OD0.071"0.084"0.083"0.0855" (Max)
Packaged Accessories
Peelable SheathYesSameSameSame
Rotating HemostasisValveYesNoSame as K110055Same as K110055
Packaging
Packaging CardPolyethyleneSameSameSame
Packaging HoopPolyethyleneSameSameSame
Other
SterilizationEthylene Oxide (EO)SameSameSame

Device Comparison:

{5}------------------------------------------------

{6}------------------------------------------------

Biocompatibility:

Biocompatibility was conducted for the React™ 71 Catheter. The React™ 71 Catheter is categorized as a limited exposure (< 24 hours), external communicating device contacting circulating blood. The following biocompatibility was conducted for the React™ 71 Catheter:

Test DescriptionResultsConclusions
Cytotoxicity(Elution Method)The test article extract showedno evidence of causing celllysis or toxicity and had aGrade 0 (No Reactivity). Thetest article extract met therequirements of the test sincethe grade was < Grade 2(Mild Reactivity).The ReactTM 71 Catheter isconsidered non-cytotoxic.
Sensitization(Guinea Pig Maximization Test)The test article extractsshowed no evidence ofcausing delayed dermalcontact sensitization in theguinea pig.The ReactTM 71 Catheter doesnot elicit a sensitizationresponse.
Irritation(Intracutaneous Reactivity)The test article met therequirements of the test sincethe difference between eachtest article extract overallmean score and correspondingcontrol extract overall meanscore was 0.0 and 0.2 for theSodium Chloride and SesameOil test article extracts,respectively.The ReactTM 71 Catheter isconsidered a non-irritant.
Acute Systemic Toxicity(Systemic Toxicity)There was no mortality orevidence of systemic toxicityfrom the extracts injected intomice.The ReactTM 71 Catheter doesnot indicate signs of toxicity.

{7}------------------------------------------------

Test DescriptionResultsConclusions
Hemocompatibility(Hemolysis)The hemolytic index for thetest article in direct contactwith blood was 0.8%, and thehemolytic index for the testarticle extract was 0.0%.The ReactTM 71 Catheter isconsidered non-hemolytic.
Hemocompatibility(Complement Activation)The concentration of SC5b-9in the test article was notstatistically higher than theactivated normal human serumcontrol or the negative control.The test article wasstatistically lower than thesponsor provided controlarticle.The ReactTM 71 Catheter is notconsidered to be potentialactivator of the complementsystem.
Hemocompatibility(Thrombogenicity – CanineModel)The control article wasevaluated and determined tohave a mean score of 2.0. Thetest article was evaluated anddetermined to have a meanscore of 1.7 after four (4)hours (±30 minutes) withoutsystemic anticoagulation.The ReactTM 71 Catheterdemonstrates lowerthrombogenic potential inarterial vessels compared tothe ACE 68 ReperfusionCatheter.
Pyrogenicity(Material Mediated)Not a single animal showed atemperature rise of 0.5°C ormore above its baselinetemperature. The total rise ofthe rabbits' temperature duringthe three (3) hours was 0.0°C.The ReactTM 71 Catheter isconsidered non-pyrogenic.

The React™ 71 Catheter has been evaluated to meet requirements specified in ISO 10993-1.

Performance Data – Bench:

Non-clinical bench testing was leveraged and conducted to evaluate the performance of the React™ 71 Catheter.

The following non-clinical bench testing was leveraged for the React™ 71 Catheter:

TestTest Method SummaryResults
Microbial
Ethylene Oxide ResidualThe React™ 71 Catheter was evaluated per ISO 10993-7.The React™ 71 Catheter met the acceptance criteria for ethylene oxide residual.
Ethylene Chlorohydrin ResidualThe React™ 71 Catheter was evaluated per ISO 10993-7.The React™ 71 Catheter met the acceptance criteria for ethylene chlorohydrin residual.

{8}------------------------------------------------

TestTest Method SummaryResults
Bioburden RecoveryThe React™ 71 Catheter was evaluated per ISO 11737-1.The React™ 71 Catheter met the acceptance criteria for bioburden recovery.
Bacterial EndotoxinThe React™ 71 Catheter was evaluated per ANSI/AAMI ST72 and USP <161>.The React™ 71 Catheter met the acceptance criteria for bacterial endotoxin.
Packaging
Visual InspectionThe React™ 71 Catheter was evaluated per ASTM F1886.The React™ 71 Catheter met the acceptance criteria for visual inspection.
Bubble LeakThe React™ 71 Catheter was evaluated per ASTM F2096.The React™ 71 Catheter met the acceptance criteria for bubble leak.
Seal StrengthThe React™ 71 Catheter was evaluated per ASTM F88.The React™ 71 Catheter met the acceptance criteria for seal strength.

The following non-clinical bench testing was conducted for the React™ 71 Catheter:

TestTest Method SummaryResults
MicrobialBioburdenThe ReactTM 71 Catheter wasevaluated per ISO 11737-1.The ReactTM 71 Catheter metthe acceptance criteria forbioburden.
PerformanceVisual InspectionThe ReactTM 71 Catheter wasinspected under x2.5magnification.The ReactTM 71 Catheter metthe acceptance criteria forvisual inspection.
Dimensional MeasurementsThe proximal ID, distal ID,proximal OD, distal OD, usablelength, total length, coatinglength, and distal tip length ofthe ReactTM 71 Catheter weremeasured.The ReactTM 71 Catheter metthe acceptance criteria fordimensional measurements.
Tip BucklingThe ReactTM 71 Catheter wasevaluated for the maximumcompressive force it canwithstand.The ReactTM 71 Catheter metthe acceptance criteria for tipbuckling.
Kink ResistanceThe ReactTM 71 Catheter wasevaluated for the maximumkink diameter.The ReactTM 71 Catheter metthe acceptance criteria for kinkresistance.
ParticulateThe ReactTM 71 Catheter wasevaluated per USP <788>.The ReactTM 71 Catheter metthe acceptance criteria forparticulate.
TestTest Method SummaryResults
Coating LubricityThe React™ 71 Catheter was evaluated for the average frictional forces.The React™ 71 Catheter met the acceptance criteria for coating lubricity.
Tensile StrengthThe React™ 71 Catheter was evaluated per ISO 10555-1. Annex B.The React™ 71 Catheter met the acceptance criteria for tensile strength at the hub and shaft.
Liquid LeakThe React™ 71 Catheter was evaluated per ISO 10555-1. Annex C.The React™ 71 Catheter met the acceptance criteria for liquid leak.
Corrosion ResistanceThe React™ 71 Catheter was evaluated per ISO 10555-1. Annex A.The React™ 71 Catheter met the acceptance criteria for corrosion resistance.
Hub Aspiration ResistanceThe React™ 71 Catheter was evaluated per ISO 10555-1. Annex D.The React™ 71 Catheter met the acceptance criteria for hub air aspiration.
RadiopacityThe markerband length and wall thickness of the React™ 71 Catheter were measured. In addition, radiopacity was confirmed via fluoroscopy.The React™ 71 Catheter met the acceptance criteria for radiopacity.
Luer StandardsThe React™ 71 Catheter was evaluated per ISO 594-1 and ISO 80369-7.The React™ 71 Catheter met the acceptance criteria for luer standards.
CompatibilityThe React™ 71 Catheter was inspected for visual damage of the catheter when delivering and retrieving interventional devices.The React™ 71 Catheter met the acceptance criteria for delivering and retrieving interventional devices.
Torque to FailureThe React™ 71 Catheter was evaluated for transmission of proximal torque to the distal tip.The React™ 71 Catheter was able to withstand torsional forces that are typical of clinical use.
Dynamic PressureThe React™ 71 Catheter was evaluated for the amount of pressure it can withstand.The React™ 71 Catheter was able to withstand pressures that are typical of clinical use.
Coating IntegrityThe React™ 71 Catheter was evaluated for coating coverage and lubricity.The React™ 71 Catheter remained coated and lubricious.
UsabilityThe React™ 71 Catheter and predicate device were evaluated for maneuverability and flexibility.The React™ 71 Catheter met the acceptance criteria for usability.

{9}------------------------------------------------

{10}------------------------------------------------

Performance Data - Animal:

Not Applicable. A determination of substantial equivalence is based upon successful completion of non-clinical bench testing as there is no change to the intended use or fundamental scientific technology.

Performance Data - Clinical:

Not Applicable. A determination of substantial equivalence is based upon successful completion of non-clinical bench testing as there is no change to the intended use or fundamental scientific technology.

Conclusion:

The design modifications incorporated do not alter the intended use or fundamental scientific technology.

Non-clinical bench testing supports a determination that the subject React™ 71 Catheter is substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).