K Number
K130858
Date Cleared
2013-10-11

(197 days)

Product Code
Regulation Number
870.1250
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reverse Medical Microcatheter-027 is intended for use in neuro, peripheral and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as occlusion coils.

Device Description

The Reverse Medical Microcatheter-027 is a single lumen, flexible stiffness composite catheter. The catheter shaft has a hydrophilic coating to reduce friction during use. The Reverse Medical Microcatheter-027 dimensions are included on the individual device labels. The Reverse Medical Microcatheter-027 inner lumen can accommodate guidewires up to 0.025 inches inner diameter to access distal tortuous vasculature. Dual radiopaque markers at the distal portion of the catheter facilitate fluoroscopic visualization.

Each Reverse Medical Microcatheter-027 is provided with accessories, which include a shaping mandrel and peel away introducer within a Tyvek™ pouch.

The shaping mandrel allows the catheter tip to be steam shaped by the physician for proper adjustment to the anatomy prior to use. Test data is presented in this submission for both the 027 and 021 Reverse Medical Microcatheters to support the inclusion of the shaping mandrel and the peel away introducer.

The Reverse Medical Microcatheter-027 incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter is provided sterile, non-pyrogenic, and is intended for single use only.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Reverse Medical Microcatheter-027:

1. Table of Acceptance Criteria and Reported Device Performance

The submission primarily focuses on demonstrating substantial equivalence through non-clinical (bench-top) testing. There isn't a comprehensive "device performance" section in the way one might expect for a clinical study with sensitive/specific metrics. Instead, the performance is reported against established engineering or safety criteria.

Test DescriptionAcceptance CriteriaReported Device Performance
Biocompatibility
Cytotoxicity (L929 MEM Elution)Grade 0 (No cell lysis) per ISO 10993-5.Non-Cytotoxic (Grade 0)
Sensitization (Kligman Maximization)Grade 0 (no visible change) per ISO 10993-10.Non-Sensitizing (Grade 0)
Systemic Toxicity (Acute)Score 0, no toxicity or animal weight loss for cottonseed oil and saline extracts per ISO 10993-11.Non-Toxic (Score 0)
Hemocompatibility (Complement Activation)No greater biological response than corresponding control.Met criteria
HemolysisHemolytic grade score of zero, considered non-hemolytic.Non-Hemolytic (Score 0)
Inactivated Partial Thromboplastin TimeMinimal activator of intrinsic coagulation pathway; passed test (clotting time 290 seconds vs. 300 seconds for negative/reference).Non-Activator
In vivo ThrombogenicityNo adverse effects or clinical signs.Non-Thrombogenic
Pyrogenicity (USP Material Mediated)Non-pyrogenic; no individual rabbit temperature rise >= 0.5 deg. C per ISO 10993-11.Non-Pyrogenic
EtO ResidualsMet criteria in accordance with ISO 10993-7 Part 7.Passed, acceptable limits
Genotoxicity (Bacterial Reverse Mutation)No statistically significant increase in revertant colonies compared to negative controls.Non-Mutagenic
Genotoxicity (Mouse lymphoma)Did not produce significantly more revertant colonies than negative controls.Non-Mutagenic
Genotoxicity (In vivo micronucleus)No statistically significant increase in micronucleus in test article extract vs. negative control.Non-Mutagenic
SterilizationSterility Assurance Level (SAL) of 10^-6 according to ANSI / AAMI / ISO 11135.Validated ($10^{-6}$ SAL)
Design Verification (Bench-Top)
Dimensional/Visual InspectionPer test protocol rev B.Pass
Tip Buckling TestPerform the same or better than predicate devices (95/90 confidence/reliability).Pass
Coating Lubricity TestFrictional force; Coating Length (95/90 confidence/reliability).Pass
Flexibility/Shaft Stiffness TestDistal tip and proximal shaft flexibility comparable or better than predicate.Pass
Priming Volume TestFor comparison to predicate; equal to or better than predicate.Pass
Flow Rate TestFor comparison to predicate; equal to or better than predicate.Pass
Guidewire Compatibility TestFree movement of appropriately sized guidewires; Frictional force (95/90 confidence/reliability).Pass
Guide Catheter Compatibility TestFree movement inside appropriately sized Guide Catheter; Frictional force measurement (95/90 confidence/reliability).Pass
Catheter Air Leakage TestPer ISO 10551-1.Pass
Catheter Liquid Leakage TestNo leaks in accordance with protocol.Pass
Dynamic Pressure TestNo leaks due to dynamic pressure test; Peak Pressure.Pass
Static Pressure TestBurst pressure in accordance with protocol (95/90 confidence/reliability).Pass
Kink Resistance TestDistal kink resistance (95/90 confidence/reliability).Pass
Torque Strength TestX revolutions without failure.Pass
Corrosion ResistanceNo corrosion on metallic components.Pass
USP Particulate TestingReport total of particles; Compare to predicate devices (USP 788).Pass
Tensile StrengthDistal/medial; medial/proximal; proximal/hub (95/90 confidence/reliability).Pass
Navigation/Accessibility/PushabilityComparable to predicate devices.Pass
Micro-devices CompatibilityComparable to predicate devices.Pass
With Steam-Shaping Capabilities
Dimensional and Visual InspectionPer test protocol rev B.Pass
Steam Shaping CapabilitiesNo Damage; Shape Inspection (at various angles).Pass
Coating Lubricity Test - Post Steam ShapingFrictional force (Pass/Fail).Pass
Static Pressure Test - Post Steam ShapingBurst pressure (95/90 confidence/reliability).Pass
Tensile Strength - Post Steam ShapingDistal/Medial; Medial/Proximal; Proximal/Hub (95/90 confidence/reliability).Pass
Navigation/Accessibility/PushabilityComparable to predicate.Pass
Micro-devices CompatibilityCan allow uses of other micro-devices, in simulated flow model.Pass

2. Sample Size for Test Set and Data Provenance

  • Sample Size: The document does not specify exact sample sizes for each bench-top test. For some tests (e.g., Tip Buckling, Coating Lubricity, Guidewire Compatibility, Guide Catheter Compatibility, Static Pressure, Kink Resistance, Tensile Strength), the acceptance criteria mention "95/90," which typically refers to a 95% confidence level with 90% reliability, implying a statistically significant sample size was used for those specific tests. However, the absolute number of devices tested is not provided.
  • Data Provenance: All data presented is from pre-clinical (bench-top) testing. The document does not indicate any human data or provide country of origin, as it's not a clinical study. It's internal testing conducted by Reverse Medical Corporation.

3. Number of Experts Used to Establish Ground Truth for Test Set and Qualifications

This information is not applicable to the provided document. The "ground truth" for this submission is based on established engineering standards (ISO, USP, internal protocols) and comparison to predicate devices, not expert consensus on medical images or diagnoses.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies where multiple human readers interpret data, and discrepancies need to be resolved. This document describes bench-top engineering tests.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This submission focuses on the substantial equivalence of a medical device (microcatheter) through non-clinical testing, not on the performance of an AI algorithm or human reader improvement.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No, a standalone study in the context of an AI algorithm was not done. The device is a physical microcatheter, not an algorithm.

7. Type of Ground Truth Used

The "ground truth" for the tests is derived from:

  • Established international and national standards (e.g., ISO, USP).
  • Internal test protocols and specifications.
  • Performance of legally marketed predicate devices, against which the new device's performance is compared for substantial equivalence.

8. Sample Size for the Training Set

This information is not applicable. There is no "training set" as this is not a machine learning or AI-based device.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, for the same reason as point 8.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).