Search Results
Found 6 results
510(k) Data Aggregation
(274 days)
Intended Use: Plate and screw Implants are intended for temporary fixation, correction or stabilization of bones in various anatomical regions.
One-Quarter Tubular Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius). The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Straight Plates 2.4/2.7 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the foot, hand, wrist, and forearm (including radius) and arthrodesis of small bones of the foot and hand.
Reconstruction Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius)
Radius Distal Styloid Plate 2.4/2.7 is indicated for the fixation for fractures of the radial styloid that have metaphyseal comminution or as an adjunct to either volar of dorsal plate fixation when additional support of the radial styloid is required.
The Radius Proximal head plates 2.4/2.7 are indicated for extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures.
The Radius Distal Volar Plates 2.4/2.7 are indicated for the fixation for fractures, fusions, or osteotomies of the distal radius.
One-Third Tubular Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), and fibula. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Reconstruction Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), wrist, and forearm (including radius), fibula, and calcaneus
Straight Plates 3.5 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the elbow (distal humerus and ulna), wrist, forearm (including radius), and fibula.
Tibia T-Plates 3.5 are indicated for buttressing of partial articular fractures and bone fragments of the distal tibia.
The Clavicle Superior Plates 3.5 are indicated for:
- Fractures of the clavicle shaft
- Fractures of the lateral clavicle
- Malunions of the clavicle
- Non-unions of the clavicle
The Clavicle Hook Plate 3.5 is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
The Humerus Proximal Plates 3.5 are indicated for the fixation of fractures, fracture dislocations, non-unions and osteotomies of the proximal humerus.
The Tibia Proximal Lateral Plates 3.5 are indicated for:
- Split-type fractures of the lateral tibial plateau
- Lateral split fractures with associated depressions
- Pure central depression fractures
- Split or depression fractures of the medial plateau
The Tibia Distal Medial Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Tibia Distal Anterolateral Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Humerus Distal Dorsolateral Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Humerus Distal Medial Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Ulna Proximal Olecranon Plates 3.5 are indicated for
- Complex extra- and intra-articular olecranon fractures
- Non-unions of the proximal ulna
- Osteotomies
- Simple olecranon fractures
Broad Curved Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Broad Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Narrow Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, and distal tibia.
T Plate 4.5/5.0 is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
T-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
LL-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Femur Distal Plates 4.5/5.0 are indicated for distal shaft fractures, supracondylar fractures, intraarticular fractures, and periprosthetic fractures of the distal femur
Tibia Proximal Lateral Plates 4.5/5.0 are indicated for proximal shaft fractures, metaphyseal fractures, intra-articular fractures and periprosthetic fractures of the proximal tibia.
The Genostis Osteosynthesis System comprises of fracture fixation plates, corresponding screws and required surgical instruments. Different shapes and sizes are available to accommodate for different anatomical regions and locations of fractures. The plate screw interface is either locking, non-locking or, for some plates, realized by a combination hole offering both options (locking and non-locking).
All plates and screws are made of titanium alloy (Ti 6Al 4V ELI) according to ASTM F136 and ISO 5832-3.
The implants of the Genostis Osteosynthesis System are sterilized by radiation and are delivered sterile.
The provided FDA 510(k) clearance letter for the Genostis Osteosynthesis System does not describe a study involving an AI/software device or human readers. Instead, it describes a traditional medical device (bone fixation plates and screws) and primarily references non-clinical (biomechanical and biocompatibility) testing to demonstrate substantial equivalence to predicate devices.
Therefore, many of the requested criteria related to AI/software performance, human reader studies, and ground truth establishment in a diagnostic context are not applicable to this document.
However, I can extract and structure the information that is relevant to the device's acceptance criteria based on the provided text, particularly concerning non-clinical testing.
Here's the breakdown based on the information available:
Acceptance Criteria and Study for Genostis Osteosynthesis System
The Genostis Osteosynthesis System is a traditional medical device (bone fixation plates and screws), not an AI/software diagnostic device. As such, the "acceptance criteria" and "study" proving it meets them are primarily based on non-clinical engineering and biological tests, not AI performance metrics or human-in-the-loop diagnostic studies.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Test/Standard | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Material Biocompatibility | Cytotoxicity (ISO 10993-5) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. |
Sensitization (ISO 10993-10) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Irritation/Intracutaneous Reactivity (ISO 10993-23) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Chemical Characterization (ISO 10993-18) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Toxicological Risk Assessment | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Bacterial Endotoxin Testing | Acceptable endotoxin levels | Performed; results satisfactory. | |
Mechanical Performance | Screw Performance (ASTM F543) | Satisfactory mechanical performance | Performed; results satisfactory. |
Plate Performance (ASTM F382) | Satisfactory mechanical performance | Performed; results satisfactory. | |
Packaging & Sterilization | Sterility (Radiation) | Device must be sterile at delivery | Implants are sterilized by radiation and delivered sterile. Test conducted on final sterile product. |
General | Substantial Equivalence to Predicates | All non-clinical testing demonstrates substantial equivalence. | All testing indicates substantial equivalence to predicates. |
Note: The document states "All results were satisfactory" for all tests, indicating the device met the pre-defined acceptance criteria for each test.
2. Sample Sizes Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact number of units or samples used for each biomechanical or biocompatibility test. It generally states that testing was "performed" and "conducted on final sterile product."
- Data Provenance: Not explicitly stated as "country of origin," but testing would be performed in a controlled laboratory environment. The tests are non-clinical (biomechanical and biocompatibility), not data from patients. The study is by nature prospective in that new manufactured devices were tested to meet the standards, not a retrospective analysis of existing data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not Applicable. This is a medical device (hardware) clearance, not an AI/software diagnostic device that requires expert ground truth for interpretation of medical images or other diagnostic data. The "ground truth" here is the adherence to established engineering and biocompatibility standards.
4. Adjudication Method for the Test Set
- Not Applicable. As this is non-clinical mechanical and biocompatibility testing, there is no "adjudication" in the sense of reconciling expert opinions on diagnostic findings. The results are quantitative and objective measurements against established standard thresholds.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- Not Applicable. MRMC studies are for evaluating clinical interpretation of diagnostic results, typically with AI assistance. This device is a physical implant, not a diagnostic tool or AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not Applicable. This pertains to AI algorithm performance. The device is a bone fixation system and does not rely on an algorithm for its primary function.
7. The Type of Ground Truth Used
- Pre-defined Standards and Controls: The "ground truth" for this device's performance is established by well-defined, internationally recognized (ISO) and American (ASTM) standards for material properties, mechanical strength, and biocompatibility. The device must perform within the specified parameters of these standards. For example, screws must meet certain torque resistance (ASTM F543), and plates must meet specific load-sharing characteristics (ASTM F382).
8. The Sample Size for the Training Set
- Not Applicable. This device does not involve a "training set" in the context of machine learning or AI.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. There is no "training set" for this type of medical device clearance. The design and manufacturing processes are established based on engineering principles and existing predicate device designs, with performance validated against the standards mentioned above.
Ask a specific question about this device
(127 days)
DePuy Synthes VOLT™ Mini Fragment Plating System:
The VOLT™ Mini Fragment Plating System is indicated for internal fracture fixation of bones and bone fragments of the appendicular skeleton appropriate for the implant size.
The VOLT™ Mini Fragment Plating System is intended for adults and both children (2-12 years) and adolescents (12-21 years) in which growth plates (physes) have fused or in which unfused growth plates will not be compromised by fixation.
If used in the femur, tibia, humerus, patella, or pelvis the VOLT™ Mini Fragment Plating System can only be used for non-load bearing stabilization and reduction.
DePuy Synthes VOLT™ Small Fragment Plating System:
The VOLT Small Fragment Plating System is indicated for internal fracture fixation of bones and bone fragments of the appendicular skeleton appropriate for the implant size.
The VOLT Small Fragment Plating System is intended for adults and both children (2-12 years) and adolescents (12-21 years) in which growth plates (physes) have fused or in which unfused growth plates will not be compromised by fixation.
If used in the femur the VOLT Small Fragment Plating System can only be used for non-load bearing stabilization and reduction.
DePuy Synthes VOLT™ Mini Fragment Plating System
The DePuy Synthes VOLT™ Mini Fragment Plating System is a family of implantable devices, consisting of non-contoured, non-anatomic straight and shaped plates (Adaption, Adaption Combi, Compact Straight, Straight Condylar, Hook, T-Plate, Tine, Y-Plate and Triangle), with variable angle screw holes, and two screw types (cortex and locking screws). The system is available in three sizes; 2.0mm, 2.4mm and 2.7mm. The Triangle Plate is available in 2.4mm and 2.7mm sizes only.
The plates of this system are available in Stainless Steel and Commercially Pure Titanium, whilst the corresponding screws are available in Stainless Steel and Titanium Alloy (TAV) respectively. Plates and screws within the VOLT™ Mini Fragment Plating System are available either sterile or non-sterile and are single-use only.
DePuy Synthes VOLT™ Small Fragment Plating System
The DePuy Synthes VOLT™ Small Fragment Plating System is a family of implantable devices consisting of 3.5mm non-contoured, non-anatomic plates with variable angle screw holes. and 3.5mm locking, 3.5mm cortex, and 4.0mm cancellous screws.
The plates of this system are available in Stainless Steel and Titanium, whilst the corresponding screws are available in Stainless Steel and Titanium Alloy (TAV). Plates and screws within the VOLT™ Small Fragment Plating System are available either sterile or non-sterile and are single-use only
The provided text is a 510(k) summary for the DePuy Synthes VOLT Mini Fragment Plating System and VOLT Small Fragment Plating System. This document describes the device, its indications for use, and a comparison to predicate devices, focusing on non-clinical performance testing (mechanical and engineering analyses) to demonstrate substantial equivalence.
Crucially, this document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence."
Therefore, the requested information regarding acceptance criteria, study details (sample size, data provenance, expert ground truth, adjudication, MRMC, standalone performance, training set details), which are typically found in clinical performance studies validating AI/ML-based medical devices, is not applicable to this submission.
This 510(k) clearance is based on the devices being "substantially equivalent" to existing predicate devices, primarily demonstrated through non-clinical performance testing (mechanical simulations and analyses), rather than clinical trials or studies involving human data and AI algorithm performance.
Therefore, I cannot provide the requested information as it is not present in the provided text. The device in question is a physical orthopedic implant, not an AI/ML software device that would require such performance evaluations.
Ask a specific question about this device
(90 days)
The 2.4mm Secure Locking Variable Angle Distal Radius Plate and Ø2.4 mm Secure Locking Screws, Self-Tapping are intended for fixation of complex intra- and extra-articular fractures arid osteotomies of the distal radius.
The 2.7mm/3.5mm Secure Locking Distal Humerus Medial Plate, 2.7mm/3.5mm Secure Locking Distal Humerus, Dorsolateral Plate, and 2.7mm/3.5mm Secure Locking Distal Humerus Plate, Dorsolateral With Lateral Support are indicated for intra-articular fractures of the distal humerus, comminuted supracondylar fractures, and nonunions of the distal humerus.
The 3.5mm Secure Locking Small Plate, 3.5mm Secure Locking T-Plate, Small with 3 Head Holes, and Ø3.5 mm Secure Locking Screws, Self-Tapping, are indicated for fixation of fractures, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula.
The 3.5 mm Secure Locking Superior-Anterior Clavicle Plate and 3.5 mm Secure Locking Superior-Anterior Clavice Plate With Lateral Extension is indicated for fixation of fractures, malunions, and ostectornies of the clavice in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.
The 3.5mm Secure Locking Olecranon Plate is indicated for fractures, osteotomies, malunions and non-unions of the olecranon.
The 3.5 mm Secure Locking Philos Proximal Humeral Internal Plate is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus.
The 4.5/5.0mm Secure Locking Narrow LC Dynamic Compression Plate is indicated for fixation of various long bones, such as the humerus, femur and tibia and for use in fixation of peri-prosthetic fractures, and fixation of nonunions or malunions in adult patients.
The 4.5mm/5.0mm Secure Locking Distal Femoral Plate is intended for buttressing multifragmentary distal femur fractures including: supracondylar, intra-articular condylar, periprosthetic fractures, nonunions and malunions, and osteotomies of the femur.
The 4.5mm/5.0mm Secure Locking Medial Proximal Tibia Plate is intended to buttress metaphyseal fractures of the medial tibia plateau, split-type fractures of the medial split fractures with associated depressions and split or depression fractures of the medial tibia plateau. Also, for use in the fixation of nonunions of the medial proximal tibia and tibia shaft.
The 4.5mm/5.0mm Secure Locking L Buttress Plate and 4.5mm/5.0mm Secure Locking T Buttress Plate are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial tibia. Also, for use in fixation of non-unions and malunions.
The 3.5mm/4.5mm/5.0mm Secure Locking Distal Tibia Plates are intended treatment of non-unions, and fractures of the distal tibia, including simple, comminuted, lateral wedge, depression, medial wedge, bicondylar, combinations of lateral wedge and depression, and fractures with associated shaft fractures.
The Ø3.5mm CORTICAL SCREW and Ø4.5mm CORTICAL SCREW are intended for fixation of fractures, osteotomies and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, calcaneous, femur and fibula in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.
The Ø2.7 mm Secure Locking Screws, Self Tapping, and Ø 2.7mm Cortical Screws are intended for fractures and osteotomies of small bone fragments, including the foot, ankle, and hand in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.
The Ø5.0mm Secure Locking Screws, Self-Tapping are intended for fixation of various long bones, such as the humerus, femur and tibia. It is also for use in fixation of non-unions or malunions.
The 6.5mm Cancellous Screw, 16mm Thread, 32 Thread, Ø4.0mm CANCELLOUS SCREW, Partial Thread, Full Thread, and Ø4.0mm SMALL CANCELLOUS CANNULATED SCREW, Partial Thread, Full Thread are indicated for use in hindfoot and midfoot fusions, subtalar fusions, calcaneal osteotomies, midfoot reconstruction, and ankle arthrodeses.
The 6.5mm Cancellous Cannulated Screw, 16mm Thread, 32 Thread, Full Thread, and 7.0mm Cancellous Cannulated Screw, 16mm Thread, 32 Thread, are indicated for fracture fixation of large bone fragments, such as tibial plateau fractures, ankle arthrodeses, intercondylar femur fractures; and subtalar arthrodeses.
HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System consists of various shapes and sizes of plates featuring compression and locking holes, full-threaded-cortical, locking selftapping screws, compression and dynamic screws. The subject device system also consists of a variety of general use instruments (Class I), which include drill bits, forceps, plate benders, and drill guides.
The plates and screws are manufactured from Stainless Steel and Titanium alloy.
The system contains several models based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, rear-foot, ankle, or other bones appropriate for the size of the device. The plate implants are in many models available, such as:, Reconstruction Plates, T-Plates, Anatomical Plates, Clavicle Plates.
These all are mainly divided into:
- . Large Fragment Plates
- Small Fragment Plates ●
- Mini Fragment Plates .
The locking screw implants are offered in corresponding diameter ranges from 2.4mm, 2.7mm, 3.5mm, 5.0mm diameters with lengths varying from a minimum length of 6 mm to maximum length of 90mm.
The non-locking screw implants are offered in 2.7mm, 3.5mm and 4.5mm diameters, with lengths ranqing from 10mm to 80mm.
The cancellous screw implants are 4.0mm and 6.5mm in diameter, with lengths ranging from 10 to 120 mm. The cancellous cannulated screw implants are offered in 4.0mm. 6.5mm and 7.0mm diameters and lengths ranging from 16mm to 130 mm.
HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are provided non-sterile, the products must be sterilized prior to use. All implants are for single use only.
This document describes the 510(k) summary for the HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System. However, it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML device.
The information provided pertains to the substantial equivalence of a medical implant (bone plates and screws) to predicate devices, focusing on material, design, and mechanical performance rather than AI/ML algorithm performance.
Therefore, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth for an AI/ML device from the given text.
The document discusses the following:
- Device Name: HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System
- Intended Use: Fixation of various bone fractures and osteotomies.
- Predicate Devices: A list of Synthes Locking Compression Plate Systems and Screws.
- Non-Clinical Testing:
- Material Standards: ASTM F136, ASTM F138, ASTM F139 (for Stainless Steel and Titanium alloy).
- Performance Standards: ASTM F382 (Metallic Bone Plates), ASTM F384 (Metallic Angled Orthopedic Fracture Fixation Devices), ASTM F543 (Metallic Medical Bone Screws), and FDA guidance documents for orthopedic screws and fracture fixation plates.
- Performance Results: Conforms to Static Four Point Bend Test, Dynamic Four Point Bend Test (for plates), Torsional Properties, Driving Torque, and Pull-out Test (for screws).
- Clinical Evaluation: Not necessary to demonstrate substantial equivalence, as the device is similar in design, pattern, and intended use to predicate devices.
In summary, the provided text does not describe an AI/ML device or its performance evaluation. It details the regulatory clearance process for a traditional medical implant based on substantial equivalence.
Ask a specific question about this device
(308 days)
ARMAR™ / ARTIS™ Small Fragment and Calcaneal Plates & MBOSS™/FIXION™ Screws are indicated for temporary internal fixation and stabilization of osteotomies and fractures, such as:
- comminuted fractures
- supracondylar fractures
- extra-articular fractures
- fractures in osteopenic bone
- nonunions
- malunions
Smaller-sized ARMAR™ / ARTISTM plates are used for small bones and small fragments of the hands and feet.
ARMARTM / ARTIS™ Calcaneal plates are indicated for temporary internal fixation and stabilization of osteotomies and fractures of the calcaneous.
ARMAR™ / ARTIS™ Humerus and Olecranon plates & MBOSSTM/FIXIONT™ Screws are intended for fractures and fracture dislocations, osteotomnies, and non-unions of the distal and proximal humerus, olecranon, and ulna in adult, particularly in osteopenic bone.
Specifically, distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures, osteotomies, malunions and non-unions of the distalhumerus. Olecranon and Proximal ulna plates are indicated for fractures, osteotomies, malunions and non-unions of the olecranon and proximal ulna.
ARMARTM / ARTISTM Tibia and Buttress Plates & MBOSS™/FIXION™ Screws are intended for fixation of the ankle in adults. Specifically,
- Anterolateral Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia.
- Medial Distal and Proximal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia,
- L Buttress and T Buttress Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia, and
- Proximal Lateral Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia, a Distal Tibia T Plates and Distal Tibia L Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia.
ARMAR™ / ARTISTM 2.4mm Plates & MBOSS™/FIXION™ Screws are intended for Fixation of complex intra- and extra-articular fractures and osteotomnies of the distal radius and other small bones in adults.
ARMARTM / ARTISTM Clavical Hook Plate & MBOSS™/FIXION™ Screws is intended for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
ARMARTM / ARTISTM Metaphyseal Plate & MBOSS™/FIXION™ Screws indicated for fractures, osteotomies, and non-unions of the radius and other small bones. Additionally, the Metaphyseal Plate is intended for pelvic and acetabular reconstructive surgery and fracture fixation of the distal humerus, clavicle, and scapula.
ARMARTM / ARTISTM Calcaneal Plate & MBOSS™/FIXION™ Screws is intended for fixation of complex extra-articular and intra-articular fractures and osteotomies of the calcaneus.
The Meril Healthcare Trauma System utilizes screws and plates of various lengths and designs to accommodate a variety of patient anatomies. The Trauma System utilizes medical grade stainless steel and titanium. Trauma System implant are provided clean and non-sterile with instructions for steam sterilization. The instruments are provided non-sterile and are reusable. The implants are provided nonsterile and are single use only.
This document is a 510(k) premarket notification decision letter from the FDA regarding a medical device, specifically bone plates and screws. It does not contain information about an AI/ML device, AI acceptance criteria, or a study proving an AI device meets acceptance criteria.
Therefore, I cannot extract the requested information regarding AI device performance, acceptance criteria, sample sizes, expert ground truth, MRMC studies, or training/test set details from this document. The document describes the substantial equivalence of bone fixation appliances to existing predicate devices based on technological characteristics and mechanical performance data (e.g., static four-point bending, dynamic four-point bending, static torsion, static axial pullout).
Ask a specific question about this device
(240 days)
The Synthes 4.5mm VA-LCP Curved Condylar Plate System is indicated for buttressing multifragmentary distal femur fractures including: supra-condylar; intra-articular and extra-articular condylar fractures, periprosthetic fractures, fractures in normal or osteopenic bone, nonunions and malunions.
The DePuy Synthes Variable Angle Positioning Pins are intended for use with cerclage multifilament cable to augment fracture stabilization with plates used in long bone fixation, when screw placement would be inhibited, as in the presence of intramedullary implant. The Variable Angle Positioning Pins are designed for use with Variable Angle LCP plate implants featuring variable angle locking holes that accept 5.0 mm variable angle bone screws.
The Synthes 4.5mm VA-LCP Curved Condylar Plate System consists of anatomically-contoured, stainless steel and titanium plates and screws featuring variable angle locking and combi-holes designed to provide stable fixation of the distal femur and system-specific instrumentation. The current 510(k) introduces OPTILINK™ Technology stainless steel screws, positioning pins for cerclage cable, and system-specific instrumentation as a line extension to the currently cleared Synthes 4.5mm VA-LCP Curved Condylar Plate System.
The provided text is a 510(k) summary for the "Synthes 4.5mm VA-LCP Curved Condylar Plate System Line Extension, Variable Angle Positioning Pins". This document is a premarket notification for a medical device seeking substantial equivalence to existing devices. It focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than proving performance against specific acceptance criteria through a clinical study in the way one might for a novel AI device or a device requiring a PMA.
Therefore, the information typically requested in your prompt (e.g., sample size for test sets, number of experts for ground truth, MRMC studies, standalone algorithm performance, training set details) is not applicable to this type of regulatory submission. This submission primarily relies on non-clinical performance data and material characterization to demonstrate substantial equivalence to predicate devices, and explicitly states that "Clinical data was not needed to demonstrate the safety and effectiveness of the proposed devices."
Since this is a Class II device and a line extension, the focus is on mechanical and material testing to ensure it performs comparably to the predicate.
Here's an attempt to answer your questions based only on the information available in the provided text, acknowledging that many fields will be "Not Applicable" or "Not Provided" in this context:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Test/Evaluation | Reported Device Performance (Summary) |
---|---|---|
Mechanical Performance | Dynamic Fatigue Construct Testing | Subject OPTILINK™ screws met performance requirements. |
Connection Strength Testing | Subject OPTILINK™ screws met performance requirements. | |
Torque-through and screw recess torsion tests | Subject OPTILINK™ screws met performance requirements. | |
ASTM-543-13 Standard Specification and Test Methods for Metallic Medical Bone Screws (Torsional Properties, Insertion/Removal Torque, Pull-out Strength) | Subject OPTILINK™ screws met performance requirements. | |
Biocompatibility | Endotoxin Requirement (LAL test method) | Devices met the specified endotoxin requirement of 20EU/device. |
Material Characterization | Metallurgical Corrosion Behavior (anodic polarization, galvanic and fretting corrosion assessments) | Demonstrated substantial equivalence to the predicate device using both stainless steel and titanium plates. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not provided. The document refers to "testing" and "evaluations" but does not specify the number of samples or specimens used in these mechanical or material tests.
- Data Provenance: Not specified, but generally refers to laboratory testing. Not applicable for retrospective/prospective human data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. Ground truth as typically understood for AI/clinical studies (e.g., expert consensus on image interpretation) is not relevant for mechanical and material testing of an orthopedic implant. Performance is evaluated against engineering standards and specifications.
4. Adjudication method for the test set
- Not applicable. Performance is measured against predetermined test specifications and standards for mechanical and material properties. There is no "adjudication" in the clinical sense.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI device or an imaging device requiring human-in-the-loop performance evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an algorithm or software device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" equivalent in this context would be the established engineering standards and material specifications (e.g., ASTM standards, specified endotoxin limits) that the device must meet, and the performance characteristics of the predicate device for comparison.
8. The sample size for the training set
- Not applicable. This is not a machine learning or AI device.
9. How the ground truth for the training set was established
- Not applicable. This is not a machine learning or AI device.
Summary based on the document:
The device (Synthes 4.5mm VA-LCP Curved Condylar Plate System Line Extension, Variable Angle Positioning Pins) demonstrates its performance and safety through non-clinical testing, material characterization, and comparison to a legally marketed predicate device (Synthes 4.5mm VA LCP Curved Condylar Plate System (K110354)) to establish "substantial equivalence." The evaluations included various mechanical tests (Dynamic Fatigue, Connection Strength, Torque, Torsional Properties, Insertion/Removal Torque, Pull-out Strength) and material tests including endotoxin and corrosion assessments. All performed tests indicated that the device met the required specifications and demonstrated substantial equivalence to the predicate device. No clinical data or human reader studies were deemed necessary for this 510(k) submission.
Ask a specific question about this device
(264 days)
NET Brand Small Fragment and Large Fragment Osteosynthesis System is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, distal femur, proximal tibial pilon, fibula, pelvis and acetabulum fractures; periprosthetic fractures; The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine.
NET Brand DHS/DCS plating system may be used for fixation of the fractures of proximal femur such as femoral neck, trochanteric, pertrochanteric or intertrochanteric zones.
NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System consists of plates and screws in a variety of designs and sizes and made from Ti-6Al-4V alloy or stainless steel. Plates are provided in straight designs and in various geometric configurations that are commonly used in trauma and reconstructive surgery. Plates are provided with screw holes to accommodate non-locking and locking screws designs. Screws are provided in, 3.5mm Cortex Self-tapping, 4.5 mm Cortex self-tapping and 2.7mm self-tapping cortex locking, 3.5mm selftapping Cortex Locking, 5.0 mm cortex Locking thread designs in various lengths. This system is not indicated for use in spine.
NET Brand of DHS/DCS Plating System made from Ti-6Al-4V alloy or stainless steel and consist of DHS/DCS Plates, lag Screw, compression screw, and 4.5 Cortex screw Self Tapping, The DHS plates are available with barrel length 25mm (short barrel) and 38mm (Standard barrel) and barrel angels varies in 130° to 150°. The DCS plate is having angle of 95°
The DHS/DCS Screw is available in total length from 50 to 145 mm, thread length 22mm, shaft diameter 7 mm and outer diameter of 12.5. The thread of DHS/DCS Screw has a buttress type.
The DHS/DCS Compression Screw can be used to achieve fracture compression. Its dimension is available with thread length 36mm and outer diameter 4.0 mm.
The provided document is a 510(k) Premarket Notification for the "NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System." This document aims to demonstrate substantial equivalence to legally marketed predicate devices, not to establish new performance criteria through a standalone study with acceptance criteria.
Therefore, the document does not contain the requested information about acceptance criteria and a study proving the device meets those criteria, as it is not a performance study in that sense.
Instead, the document focuses on demonstrating substantial equivalence to predicate devices. This is achieved by comparing the new device's technological characteristics, intended use, materials, and performance (via testing against established ASTM standards) to those of already approved predicate devices.
Here's an explanation based on the provided text, addressing the points where information is available or noting its absence:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not define specific "acceptance criteria" for clinical performance in terms of sensitivity, specificity, accuracy, or similar metrics for a new diagnostic or prognostic device. Instead, it demonstrates compliance with recognized engineering and material standards to show that the device performs equivalently to previously approved devices.
The "reported device performance" is essentially a confirmation of conformance to established ASTM standards for bone plates and screws.
Characteristic | Acceptance Criteria (Standard Compliance) | Reported Device Performance |
---|---|---|
Material | ASTM F 136, ASTM F 138, ASTM F 139 | Conforms |
Bone Plates (Static) | ASTM F 382, ASTM F 384 (Static Four Point Bend Test) | Conforms |
Bone Plates (Dynamic) | ASTM F 382, ASTM F 384 (Dynamic Four Point Bend Test) | Conforms |
Bone Screws (Torsional) | ASTM F 543 (Torsional Properties) | Conforms |
Bone Screws (Driving) | ASTM F 543 (Driving Torque) | Conforms |
Bone Screws (Pull-out) | ASTM F 543 (Pull-out Test) | Conforms |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document does not specify a "sample size" in terms of patient cases or images, as it is a mechanical device submission. The testing involves mechanical specimens (e.g., plates and screws) tested according to the cited ASTM standards. The specific number of specimens tested for each standard is not detailed in this summary.
- Data Provenance: Not applicable as it's not a clinical or imaging study. The tests are mechanical and presumably conducted in a laboratory setting.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This is a submission for a bone fixation device, not a diagnostic device requiring expert interpretation for ground truth establishment. Mechanical testing relies on standardized methodologies, not expert consensus on clinical cases.
4. Adjudication method for the test set:
Not applicable for a mechanical device submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a medical device for bone fixation, not an AI or diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. This is a medical device for bone fixation, not an algorithm.
7. The type of ground truth used:
The "ground truth" in this context is the established performance benchmarks defined by recognized industry standards (ASTM F382, ASTM F384, ASTM F543 for performance, and ASTM F136, ASTM F138, ASTM F139 for materials). The device is deemed acceptable if it conforms to these standards, indicating mechanical properties are comparable to legally marketed devices.
8. The sample size for the training set:
Not applicable. This is a mechanical device, not a machine learning model requiring a training set.
9. How the ground truth for the training set was established:
Not applicable, as there is no training set for this type of device submission. The "ground truth" for demonstrating substantial equivalence rests on adherence to material and performance standards, which are developed and accepted by the scientific and engineering community.
Ask a specific question about this device
Page 1 of 1