Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ANTHEM™ Fracture System is indicated for fixation of fractures, arthrodesis and reconstruction of bones for the appropriate size of the device to be used in adult patients, including the clavicle, scapula, humerus, radius, ulpa, small bones (metacarpals, metatarsals, phalanges), wrist, pelvis, fibula, ankle, and foot. Small fragment and distal fibula plates may be used in all pediatric subgroups (except neonates) and small stature adults. Distal radius plates may be used in adolescents (12-21 years of age).

    Device Description

    The ANTHEM™ Fracture System is a family of plates and screws designed to be used for internal bone fixation. The implants are available in various sizes and shapes to accommodate patient anatomy, and may be contoured or straight, sterile and non-sterile, with locking and non-locking screws. ANTHEM™ implants are manufactured from medical grade titanium alloy, cobalt chromium molybdenum alloy, or stainless steel. All implants are for single use only.

    AI/ML Overview

    This document is a 510(k) Summary for the ANTHEM™ Fracture System, a medical device. Based on the provided text, there is no study described that proves the device meets specific acceptance criteria based on AI/algorithm performance, nor is there any mention of an AI device. The document explicitly describes the ANTHEM™ Fracture System as a system of "plates and screws designed to be used for internal bone fixation."

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance related to AI/algorithmic accuracy.
    • Sample size and data provenance for an AI test set.
    • Number of experts and qualifications for AI ground truthing.
    • Adjudication method for an AI test set.
    • MRMC comparative effectiveness study for AI assistance.
    • Standalone AI performance.
    • Type of ground truth used for AI.
    • Sample size for training set for AI.
    • How ground truth for a training set was established for AI.

    The document focuses on the mechanical and material performance of bone fixation devices, not on diagnostic or AI-driven performance.

    Here's a summary of the performance data that is mentioned in the document, which pertains to the physical device itself:

    1. Acceptance Criteria and Reported Device Performance (for the physical device):

    • Acceptance Criteria (Implicit Standard Performance): The device's performance was evaluated in accordance with recognized ASTM standards for bone fixation devices. This implies that the device had to meet performance benchmarks defined by these standards to demonstrate substantial equivalence to predicate devices. While specific numerical acceptance criteria are not explicitly stated (e.g., "minimum bending strength of X N-m"), the adherence to these standards serves as the implicit acceptance criteria for mechanical performance.
    • Reported Device Performance:
      • Tests Conducted: Engineering analysis, bending strength tests (for plates and screws), pullout strength tests (for screws), and insertion/removal torque tests (for screws).
      • Outcome: "Performance data demonstrates substantial equivalence to the predicate devices."
      • Bacterial Endotoxin Testing (BET): Conducted in accordance with ANSI/AAMI ST-72:2011 to ensure sterility or low endotoxin levels.

    2. Sample Size and Data Provenance (for the physical device testing):

    • The document does not specify the sample sizes used for the mechanical performance tests (e.g., how many plates or screws were tested for bending strength).
    • The data provenance is not mentioned (e.g., where the tests were conducted, whether the materials were from specific batches, etc.).

    3. Number of experts used to establish the ground truth... and qualifications of those experts:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm requiring expert review for ground truth.

    4. Adjudication method for the test set:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done...:

    • Not applicable. This document is about a physical medical device (bone plates and screws), not an AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This document describes a physical medical device, not an algorithm.

    7. The type of ground truth used:

    • For the physical device, "ground truth" would be the objective measurements obtained from standardized mechanical tests (e.g., actual force at failure, actual torque values) against the requirements of the ASTM standards.

    8. The sample size for the training set:

    • Not applicable, as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable, as this is not an AI/machine learning device.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NET Brand Small Fragment and Large Fragment Osteosynthesis System is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, distal femur, proximal tibial pilon, fibula, pelvis and acetabulum fractures; periprosthetic fractures; The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine.

    NET Brand DHS/DCS plating system may be used for fixation of the fractures of proximal femur such as femoral neck, trochanteric, pertrochanteric or intertrochanteric zones.

    Device Description

    NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System consists of plates and screws in a variety of designs and sizes and made from Ti-6Al-4V alloy or stainless steel. Plates are provided in straight designs and in various geometric configurations that are commonly used in trauma and reconstructive surgery. Plates are provided with screw holes to accommodate non-locking and locking screws designs. Screws are provided in, 3.5mm Cortex Self-tapping, 4.5 mm Cortex self-tapping and 2.7mm self-tapping cortex locking, 3.5mm selftapping Cortex Locking, 5.0 mm cortex Locking thread designs in various lengths. This system is not indicated for use in spine.

    NET Brand of DHS/DCS Plating System made from Ti-6Al-4V alloy or stainless steel and consist of DHS/DCS Plates, lag Screw, compression screw, and 4.5 Cortex screw Self Tapping, The DHS plates are available with barrel length 25mm (short barrel) and 38mm (Standard barrel) and barrel angels varies in 130° to 150°. The DCS plate is having angle of 95°

    The DHS/DCS Screw is available in total length from 50 to 145 mm, thread length 22mm, shaft diameter 7 mm and outer diameter of 12.5. The thread of DHS/DCS Screw has a buttress type.

    The DHS/DCS Compression Screw can be used to achieve fracture compression. Its dimension is available with thread length 36mm and outer diameter 4.0 mm.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the "NET Brand Small Fragment and Large Fragment Osteosynthesis Plating System, NET Brand of DHS/DCS Plating System." This document aims to demonstrate substantial equivalence to legally marketed predicate devices, not to establish new performance criteria through a standalone study with acceptance criteria.

    Therefore, the document does not contain the requested information about acceptance criteria and a study proving the device meets those criteria, as it is not a performance study in that sense.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices. This is achieved by comparing the new device's technological characteristics, intended use, materials, and performance (via testing against established ASTM standards) to those of already approved predicate devices.

    Here's an explanation based on the provided text, addressing the points where information is available or noting its absence:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not define specific "acceptance criteria" for clinical performance in terms of sensitivity, specificity, accuracy, or similar metrics for a new diagnostic or prognostic device. Instead, it demonstrates compliance with recognized engineering and material standards to show that the device performs equivalently to previously approved devices.

    The "reported device performance" is essentially a confirmation of conformance to established ASTM standards for bone plates and screws.

    CharacteristicAcceptance Criteria (Standard Compliance)Reported Device Performance
    MaterialASTM F 136, ASTM F 138, ASTM F 139Conforms
    Bone Plates (Static)ASTM F 382, ASTM F 384 (Static Four Point Bend Test)Conforms
    Bone Plates (Dynamic)ASTM F 382, ASTM F 384 (Dynamic Four Point Bend Test)Conforms
    Bone Screws (Torsional)ASTM F 543 (Torsional Properties)Conforms
    Bone Screws (Driving)ASTM F 543 (Driving Torque)Conforms
    Bone Screws (Pull-out)ASTM F 543 (Pull-out Test)Conforms

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify a "sample size" in terms of patient cases or images, as it is a mechanical device submission. The testing involves mechanical specimens (e.g., plates and screws) tested according to the cited ASTM standards. The specific number of specimens tested for each standard is not detailed in this summary.
    • Data Provenance: Not applicable as it's not a clinical or imaging study. The tests are mechanical and presumably conducted in a laboratory setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This is a submission for a bone fixation device, not a diagnostic device requiring expert interpretation for ground truth establishment. Mechanical testing relies on standardized methodologies, not expert consensus on clinical cases.

    4. Adjudication method for the test set:

    Not applicable for a mechanical device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a medical device for bone fixation, not an AI or diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is a medical device for bone fixation, not an algorithm.

    7. The type of ground truth used:

    The "ground truth" in this context is the established performance benchmarks defined by recognized industry standards (ASTM F382, ASTM F384, ASTM F543 for performance, and ASTM F136, ASTM F138, ASTM F139 for materials). The device is deemed acceptable if it conforms to these standards, indicating mechanical properties are comparable to legally marketed devices.

    8. The sample size for the training set:

    Not applicable. This is a mechanical device, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device submission. The "ground truth" for demonstrating substantial equivalence rests on adherence to material and performance standards, which are developed and accepted by the scientific and engineering community.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083032
    Date Cleared
    2009-01-07

    (89 days)

    Product Code
    Regulation Number
    888.3030
    Why did this record match?
    Reference Devices :

    K993106, K000682, K000684, K011815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERI-LOC™ Bone Plating and Screw System can be used for adult and pediatric patients, as well as patients with osteopenic bone. PERI-LOC™ bone plates and screws are indicated for fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, calcaneus and clavicle.

    Device Description

    The PERI-LOC™ Bone Plating and Screw System represents titanium versions of stainless steel PERI-LOC™ Periarticular Locked Plating System bone plates and bone screws previously cleared under K082516, K061352, K051735, and K033669. Like the predicate devices, the PERI-LOC™ Bone Plating and Screw System devices feature various lengths of straight and contoured locking bone plates: locking and non-locking bone screws, washers, plus a screw adapter accessory. Bone plates in the PER/-LOC™ Bone Plating and Screw System feature a screw to-plate locking feature, which forms a locked, fixed-angle construct to aid in maintaining fracture reduction.

    AI/ML Overview

    The provided text is a 510(k) summary for the PERI-LOC™ Bone Plating and Screw System. It describes the device, its intended use, and its substantial equivalence to predicate devices. However, this document does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria.

    The 510(k) pathway is primarily focused on demonstrating substantial equivalence to a legally marketed predicate device, rather than requiring new clinical studies to prove performance against specific acceptance criteria like those outlined in your request. While the device's technological characteristics are mentioned as being similar to predicate devices, there's no discussion of quantitative performance metrics, study designs, sample sizes, or ground truth establishment.

    Therefore, I cannot populate the table or answer most of your detailed questions based solely on the provided text. The document focuses on regulatory approval through substantial equivalence, not on a detailed performance study with defined acceptance criteria.

    Here's what I can infer or explicitly state based on the provided text:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in the document. The 510(k) primarily focuses on demonstrating substantial equivalence to predicate devices, rather than establishing new performance acceptance criteria for a novel device.Not specified in the document. The document states that the new device has "similar technological characteristics" and "similarities in design features and overall indications for use" to predicate devices, implying comparable performance without providing specific metrics.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not applicable and not mentioned in the document. The 510(k) submission does not describe a performance test set.
    • Data provenance: Not applicable. No performance data is presented.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. The document does not describe a study involving ground truth establishment by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No test set or adjudication process is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a bone plating and screw system, not an AI-based diagnostic tool for which MRMC studies are typically performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This device is a mechanical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. The document does not reference performance studies that would require ground truth. The substantial equivalence argument relies on the established safety and effectiveness of predicate devices.

    8. The sample size for the training set

    • Not applicable. This submission doesn't describe an algorithm or a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This submission doesn't describe an algorithm or a training set.

    Summary of what the document does communicate regarding evidence for substantial equivalence:

    The "study" referenced in a 510(k) context is often the comparison to legally marketed predicate devices. The device's safety and effectiveness are supported by demonstrating that it has:

    • Similar intended use.
    • Similar technological characteristics (e.g., made of titanium like some predicate devices, screw-to-plate locking feature).
    • No new questions of safety or effectiveness compared to those predicates.

    The acceptance criteria for a 510(k) largely revolve around demonstrating this substantial equivalence, which is typically shown through engineering comparisons, material biocompatibility, and sometimes mechanical testing (though specific results of such tests are not detailed in this summary). The document lists several predicate devices (K082516, K072818, K061352, K051735, K033669, K993106, K000682, K000684, K011815) to which the PERI-LOC™ Bone Plating and Screw System is compared. The "study" proving it meets acceptance criteria is the regulatory review finding of substantial equivalence based on the submitted documentation comparing the device to these predicates.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1