Search Results
Found 2 results
510(k) Data Aggregation
(274 days)
Intended Use: Plate and screw Implants are intended for temporary fixation, correction or stabilization of bones in various anatomical regions.
One-Quarter Tubular Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius). The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Straight Plates 2.4/2.7 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the foot, hand, wrist, and forearm (including radius) and arthrodesis of small bones of the foot and hand.
Reconstruction Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius)
Radius Distal Styloid Plate 2.4/2.7 is indicated for the fixation for fractures of the radial styloid that have metaphyseal comminution or as an adjunct to either volar of dorsal plate fixation when additional support of the radial styloid is required.
The Radius Proximal head plates 2.4/2.7 are indicated for extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures.
The Radius Distal Volar Plates 2.4/2.7 are indicated for the fixation for fractures, fusions, or osteotomies of the distal radius.
One-Third Tubular Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), and fibula. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Reconstruction Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), wrist, and forearm (including radius), fibula, and calcaneus
Straight Plates 3.5 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the elbow (distal humerus and ulna), wrist, forearm (including radius), and fibula.
Tibia T-Plates 3.5 are indicated for buttressing of partial articular fractures and bone fragments of the distal tibia.
The Clavicle Superior Plates 3.5 are indicated for:
- Fractures of the clavicle shaft
- Fractures of the lateral clavicle
- Malunions of the clavicle
- Non-unions of the clavicle
The Clavicle Hook Plate 3.5 is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
The Humerus Proximal Plates 3.5 are indicated for the fixation of fractures, fracture dislocations, non-unions and osteotomies of the proximal humerus.
The Tibia Proximal Lateral Plates 3.5 are indicated for:
- Split-type fractures of the lateral tibial plateau
- Lateral split fractures with associated depressions
- Pure central depression fractures
- Split or depression fractures of the medial plateau
The Tibia Distal Medial Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Tibia Distal Anterolateral Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Humerus Distal Dorsolateral Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Humerus Distal Medial Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Ulna Proximal Olecranon Plates 3.5 are indicated for
- Complex extra- and intra-articular olecranon fractures
- Non-unions of the proximal ulna
- Osteotomies
- Simple olecranon fractures
Broad Curved Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Broad Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Narrow Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, and distal tibia.
T Plate 4.5/5.0 is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
T-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
LL-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Femur Distal Plates 4.5/5.0 are indicated for distal shaft fractures, supracondylar fractures, intraarticular fractures, and periprosthetic fractures of the distal femur
Tibia Proximal Lateral Plates 4.5/5.0 are indicated for proximal shaft fractures, metaphyseal fractures, intra-articular fractures and periprosthetic fractures of the proximal tibia.
The Genostis Osteosynthesis System comprises of fracture fixation plates, corresponding screws and required surgical instruments. Different shapes and sizes are available to accommodate for different anatomical regions and locations of fractures. The plate screw interface is either locking, non-locking or, for some plates, realized by a combination hole offering both options (locking and non-locking).
All plates and screws are made of titanium alloy (Ti 6Al 4V ELI) according to ASTM F136 and ISO 5832-3.
The implants of the Genostis Osteosynthesis System are sterilized by radiation and are delivered sterile.
The provided FDA 510(k) clearance letter for the Genostis Osteosynthesis System does not describe a study involving an AI/software device or human readers. Instead, it describes a traditional medical device (bone fixation plates and screws) and primarily references non-clinical (biomechanical and biocompatibility) testing to demonstrate substantial equivalence to predicate devices.
Therefore, many of the requested criteria related to AI/software performance, human reader studies, and ground truth establishment in a diagnostic context are not applicable to this document.
However, I can extract and structure the information that is relevant to the device's acceptance criteria based on the provided text, particularly concerning non-clinical testing.
Here's the breakdown based on the information available:
Acceptance Criteria and Study for Genostis Osteosynthesis System
The Genostis Osteosynthesis System is a traditional medical device (bone fixation plates and screws), not an AI/software diagnostic device. As such, the "acceptance criteria" and "study" proving it meets them are primarily based on non-clinical engineering and biological tests, not AI performance metrics or human-in-the-loop diagnostic studies.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Test/Standard | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Material Biocompatibility | Cytotoxicity (ISO 10993-5) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. |
Sensitization (ISO 10993-10) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Irritation/Intracutaneous Reactivity (ISO 10993-23) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Chemical Characterization (ISO 10993-18) | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Toxicological Risk Assessment | Satisfactory bio-compatibility endpoints | Performed; results satisfactory. | |
Bacterial Endotoxin Testing | Acceptable endotoxin levels | Performed; results satisfactory. | |
Mechanical Performance | Screw Performance (ASTM F543) | Satisfactory mechanical performance | Performed; results satisfactory. |
Plate Performance (ASTM F382) | Satisfactory mechanical performance | Performed; results satisfactory. | |
Packaging & Sterilization | Sterility (Radiation) | Device must be sterile at delivery | Implants are sterilized by radiation and delivered sterile. Test conducted on final sterile product. |
General | Substantial Equivalence to Predicates | All non-clinical testing demonstrates substantial equivalence. | All testing indicates substantial equivalence to predicates. |
Note: The document states "All results were satisfactory" for all tests, indicating the device met the pre-defined acceptance criteria for each test.
2. Sample Sizes Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact number of units or samples used for each biomechanical or biocompatibility test. It generally states that testing was "performed" and "conducted on final sterile product."
- Data Provenance: Not explicitly stated as "country of origin," but testing would be performed in a controlled laboratory environment. The tests are non-clinical (biomechanical and biocompatibility), not data from patients. The study is by nature prospective in that new manufactured devices were tested to meet the standards, not a retrospective analysis of existing data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not Applicable. This is a medical device (hardware) clearance, not an AI/software diagnostic device that requires expert ground truth for interpretation of medical images or other diagnostic data. The "ground truth" here is the adherence to established engineering and biocompatibility standards.
4. Adjudication Method for the Test Set
- Not Applicable. As this is non-clinical mechanical and biocompatibility testing, there is no "adjudication" in the sense of reconciling expert opinions on diagnostic findings. The results are quantitative and objective measurements against established standard thresholds.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- Not Applicable. MRMC studies are for evaluating clinical interpretation of diagnostic results, typically with AI assistance. This device is a physical implant, not a diagnostic tool or AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not Applicable. This pertains to AI algorithm performance. The device is a bone fixation system and does not rely on an algorithm for its primary function.
7. The Type of Ground Truth Used
- Pre-defined Standards and Controls: The "ground truth" for this device's performance is established by well-defined, internationally recognized (ISO) and American (ASTM) standards for material properties, mechanical strength, and biocompatibility. The device must perform within the specified parameters of these standards. For example, screws must meet certain torque resistance (ASTM F543), and plates must meet specific load-sharing characteristics (ASTM F382).
8. The Sample Size for the Training Set
- Not Applicable. This device does not involve a "training set" in the context of machine learning or AI.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. There is no "training set" for this type of medical device clearance. The design and manufacturing processes are established based on engineering principles and existing predicate device designs, with performance validated against the standards mentioned above.
Ask a specific question about this device
(92 days)
The AxSOS 3 Ti Locking Plate System is intended for long bone fracture fixation. Indications include:
- Diapyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures .
- . Non-unions and malunions
- Normal and osteopenic bone .
- Osteotomies .
This Traditional 510(k) submission is being supplied to the U.S. FDA to provide authorization to market the new AxSOS 3 Ti Locking Plate System. The AxSOS 3 Ti Locking Plate System is an internal fixation device that consists of monoaxial plates and various types of screws to fit different types of fractures in the tibia and femur. The subject components will be available sterile and non-sterile. The plates will be available in sizes ranging from 95-415mm in length and the screws will range from 10-150mm in length.
- Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria | Reported Device Performance |
---|---|
Static strength | Substantially equivalent to predicate devices |
Dynamic Fatigue Testing | Substantially equivalent to predicate devices |
Shear-Off Testing | Substantially equivalent to predicate devices |
Insertion Testing | Substantially equivalent to predicate devices |
Pull-Out Testing | Substantially equivalent to predicate devices |
-
Sample size used for the test set and the data provenance:
The provided text does not include information on specific sample sizes for the non-clinical tests. The tests were performed on "AxSOS 3 Ti Locking Plate System components" but the number of components tested for each criterion is not specified.
The data provenance is from non-clinical laboratory testing conducted for the AxSOS 3 Ti Locking Plate System components. The location of the laboratory or the specific country of origin for the testing data is not explicitly stated, but the sponsor is Stryker Trauma AG, Switzerland. The study type is non-clinical (benchtop) testing, not related to patient data, so "retrospective or prospective" does not apply in the typical sense.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable as the study involved non-clinical mechanical testing, not a clinical evaluation requiring expert interpretation of medical images or data. Ground truth was established through standardized engineering tests.
-
Adjudication method for the test set:
This information is not applicable as the study involved non-clinical mechanical testing, not a clinical evaluation requiring human interpretation or adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was done. This submission explicitly states, "Clinical testing was not required for this submission." The study focused solely on non-clinical mechanical performance.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
No standalone (algorithm only) performance study was done. This device is a physical bone fixation system and does not involve AI algorithms.
-
The type of ground truth used:
The ground truth used was based on results from standardized biomechanical and mechanical tests. This includes measurements of static strength, dynamic fatigue, shear-off properties, insertion forces, and pull-out resistance, all compared against established performance benchmarks or predicate devices.
-
The sample size for the training set:
This information is not applicable as this submission is for a physical medical device (bone plates and screws) and does not involve an AI algorithm with a training set.
-
How the ground truth for the training set was established:
This information is not applicable as this submission is for a physical medical device and does not involve an AI algorithm with a training set or associated ground truth establishment.
Ask a specific question about this device
Page 1 of 1