Found 21 results
510(k) Data Aggregation
(115 days)
IPDmilled Blanks are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed CAD/CAM customizations for the IPDmilled Blanks are to be designed and manufactured according to digital dentistry workflow or to be sent to an IPD validated milling center for manufacture. The workflow system integrates multiple components of the digital dentistry workflow: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, milling machine and associated tooling and accessories. Compatible Implant Systems: Dentium Co., Ltd Implantium, Friadent Implant Systems, Neoss ProActive Implant, SPI® Dental Implant, Inicell®, BEGO Semados® S-Line, ANKYLOS® C/X Implant System, MIS V3 Conical Connection Dental Implant System, Conical Connection Implants (MIS® C1), MIS C1 Narrow Platform Conical Connection Implant System, MIS C1 Wide Platform Conical Connection Abutments, PRAMA White Implant Systems, Altatec GmbH CAMLOG®/CONELOG® PROGRESSIVE-LINE Implants, Straumann TLX Implant System, Kontact Dental Implant System, ICX-Implant System, Tapered Pro Conical Implant System, Straumann® Tissue Level, Zimmer Tapered Screw-Vent®, Nobel Biocare® Nobel Active®, Straumann® Bone Level Tapered Implants, Neodent Implant System – GM Line, Osstem Implant System, Xpeed AnyRidge Internal Implant System, 3i Osseotite® Certain® Dental Implants, Astra Tech Implant System (Osseospeed®), OsseoSpeed™ Plus, BioHorizons Tapered Internal Implant System, MIS Internal Hex Dental Implant System (MIS® Seven®), Straumann BLX Ø3.5 mm Implants, Straumann® BLX Implant System, ALTATEC Camlog Screwline Implant System, Replace TiUnite Endosseous Implant.
The purpose of this submission is to expand abutment categories of IPD Dental Implant Abutments, which were previously cleared.
IPDmilled Blanks is a dental implant abutment system composed of dental abutments and screws intended to be placed into dental implants to provide support for dental prosthetic restorations. Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems. IPDmilled Blanks have a pre-manufactured connection interface that fits directly with a pre-specified dental implant. The customized shape of the abutment is intended to be designed and manufactured according to a digital dentistry workflow or to be sent to an IPD validated milling center for manufacture. IPDmilled Blank is delivered non-sterile and the final restoration, including the screw, is intended to be sterilized at the dental clinic before it is placed in the patient.
The IPDmilled Blanks are blistered together with their specific screw. The screws are intended to attach the prosthesis to the dental implant.
The metallic components of the subject abutments and screws are made of titanium alloy conforming to ISO 5832-3 "Implants for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy".
The design and manufacturing of the custom abutment will be conducted using a digital dentistry workflow requiring the use of the following equipment and software:
Intraoral Scanner: 3Shape TRIOS A/S Series
Desktop Scanner: 3Shape E Series.
Design Software: 3Shape Abutment Designer Software, K151455.
AbutmentCAD, K193352.
Milling unit: CORiTEC 350i PRO / CORiTEC 350i Loader PRO with MillBox CAM software.
IPDmilled Blank libraries' have built-in design limitations, and the user isn't allowed to exceed these limitations. The custom abutment design limitation specifications are as follows:
Minimum gingival height: 0.5 mm
Maximum gingival height: 6.0 mm
Minimum wall thickness: 0.4 mm
Minimum post height: 4.0 mm
All IPDmilled Blanks are for straight abutments only.
N/A
Ask a specific question about this device
(146 days)
SIMDA Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
It is compatible with the following systems:
Pre-Milled Blank, Patient-specific abutment
| Compatible System | Implant Body Diameter(mm) | Implant Platform |
|---|---|---|
| Astra OsseoSpeed® EV (K120414, K130999) | 3.0, 3.6, 4.2, 4.8, 5.4 | 3.0, 3.6, 4.2, 4.8, 5.4 |
| Astra Tech Implant System (K101732) Astra Tech OsseoSpeed Astra Tech Fixture MicroMacro (New indication) (K024111) | 3.0S, 3.5S, 4.0S, 4.5, 5.0, 5.0S | 3.0(x-small), 3.5/4.0(small), 4.5/5.0(large) |
| BioHorizons Laser-Lok 3.0 Implant System (K093321) BioHorizons Tapered Internal Implant System (K143022, K071638) | 3.0, 3.4, 3.8, 4.6, 5.8 | 3.0, 3.4, 3.8, 4.6, 5.8 |
| Biomet 3i Certain® (K130949) | 3.25, 4.0, 5.0, 6.0 | 3.25, 4.0, 5.0, 6.0 |
| Straumann BLX Implant (K173961, K181703, K191256) | 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5 | RB, WB |
Ti-Base abutment
| Compatible System | Implant Body Diameter(mm) | Implant Platform |
|---|---|---|
| BioHorizons Laser-Lok 3.0 Implant System (K093321) BioHorizons Tapered Internal Implant System (K143022, K071638) | 3.0, 3.4, 3.8, 4.6, 5.8 | 3.0, 3.4, 3.8, 4.6, 5.8 |
| Biomet 3i Certain® (K130949) | 3.25, 4.0, 5.0, 6.0 | 3.25, 4.0, 5.0, 6.0 |
| Straumann BLX Implant (K173961, K181703, K191256) | 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5 | RB, WB |
Multi-unit abutment
| Compatible System | Implant Body Diameter(mm) | Implant Platform |
|---|---|---|
| Straumann BLX Implant (K173961, K181703, K191256) | 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.5 | RB, WB |
Multi-unit Abutment is only compatible with Straumann BLX Implant System.
All digitally designed abutments for use with SIMDA Abutments are intended to be sent to a SIMDA validated milling center for manufacture.
SIMDA Abutment is intended for use as an aid in prosthetic restoration. It consists of Pre-Milled Blank, Ti-Base, Multi-unit Abutment, Multi-unit Base. It has a pre-manufactured connection interface that fits directly to an endosseous dental implant.
Pre-milled Blank Design
This device is a patient-specific abutment composed of titanium alloy (Ti-6Al-4V ELI, ASTM F136) which is designed by a clinician or lab, and milled at a SIMDA validated milling center.
Pre-Milled Blank Design Limitation for Patient-specific abutment:
| Design parameter (Patient-specific abutment) | Subject System Design Limit |
|---|---|
| Minimum and Maximum Gingival (Cuff) Height | 0.5~5mm |
| Minimum and Maximum diameter at abutment/implant interface | 4.0~8.0 |
| Minimum and Maximum length of abutment | 6~12mm |
| Minimum and Maximum length of abutment post (length above the abutment collar/gingival height) | 4~8mm |
| Minimum wall thickness at abutment/implant interface | 0.32 (Astra EV), 0.32 (Astra TX), 0.39 (BioHorizons), 0.4 (Biomet 3i), 0.42 (Straumann BLX) |
| Minimum and Maximum abutment angle | 0 |
Pre-Milled Blank Design Limitation for Patient-specific abutment:
Astra OsseoSpeed EV, Astra Tech OsseoSpeed TX is minimum and maximum abutment angle 0~15 º
Ti-Base Abutment
Ti Base consists of a two-piece abutment, where the titanium base is a pre-manufactured titanium component that will be used to support a CAD/CAM-designed zirconia superstructure (the second part of the two-piece abutment) that composes the final abutment.
Ti Base are provided non-sterile therefore must be sterilized after the cementation of the patient-specified superstructure.
Raw material blanks
- InCoris Zi (ZrO2) by Sirona Dental Systems GmbH, L size blanks, cleared under K123664.
Cement
- RelyX Unicem 2Automix by 3M ESPE, cleared under K100756.
Design Limitation for Zirconia superstructure:
| Design Parameter | Limit (Min.~Max.) |
|---|---|
| Minimum and Maximum abutment angle | 0~15 |
| Minimum and Maximum Gingival (Cuff) Height | 0.5~5.0 |
| Minimum and Maximum length of abutment post (length above the abutment collar/gingival height) | 4.0~6.0 |
| Minimum and Maximum diameter at abutment/implant interface | 5.0~8.0 |
| Minimum wall thickness at abutment/implant interface | 0.33 |
Multi-unit Abutment and Base
Multi-unit Abutment which are placed into the dental implant to provide support for the prosthetic restoration. The abutments are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F136). The Multi-unit Abutment has a post height less than 4mm and are always combined with Multi-unit Base, increasing the post height to a minimum of 4.5mm.
N/A
Ask a specific question about this device
(163 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
The purpose of this submission is to expand the DESS Dental Smart Solutions for the DESSLoc Attachment system cleared under K170588, K191986, K212628, K222288, K240208, and K242340 and to:
- include OEM platform compatibilities to the DESSLoc Abutment design that have been previously cleared in other DESS Abutment designs,
- include new OEM platform compatibility for MIS C1 Dental Implant System,
- include attachment components (retention inserts and housing) including reprocessing information in labeling.
The DESSLoc Attachment System consists of abutments and device-specific accessories (retention inserts and denture housings) for resilient attachment of prostheses to endosseous dental implants. There have been no changes to the design of the DESSLoc abutments, the design is the same that has been cleared in the above submissions. The abutments are made of titanium alloy and coated with zirconium nitride (ZrN). The nylon retention insert is manufactured from Polynil® (polyamide 6.6) or Vestamid® Care ML GB30 (polyamide 12). The denture housing is made of titanium alloy with a machined surface or anodized surface. The DESSLoc abutment is compatible with OEM implants, as listed below.
N/A
Ask a specific question about this device
(84 days)
DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
The purpose of this submission is to add abutments to the DESS Dental Smart Solutions system, which includes abutments cleared previously in K242340 and K240208. The subject device abutment designs include Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), and abutment screws.
Pre-Milled Blank Abutments are designed for custom abutment fabrication by a CAD-CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. The Pre-Milled Blank Abutments have a maximum (before milling) diameter of 10 mm or 14 mm and a solid cylindrical design with an engaging or non-engaging implant connection. They are available in versions with a straight screw channel and with an angled screw channel for esthetic considerations. The Pre-Milled Blank Abutments are manufactured from titanium alloy (Ti 6Al-4V).
The design parameters for the CAD-CAM fabrication of a custom abutment from the Pre-Milled Blank Abutment are:
- Minimum wall thickness – 0.45 mm
- Minimum cementable post height for single-unit restoration – 4.0 mm (minimum cementable post height for single-unit restoration is defined as the height above the restorative margin)
- Minimum gingival height – 0.5 mm
- Maximum gingival height – 6.0 mm
- Pre-Milled Blanks are for straight abutments only
Multi-Unit Abutments: Straight and Angled are designed for attachment of multi-unit screw-retained restorations and are provided in three (3) designs, straight (0°), angled 17°, and angled 30°. The designs of the subject Multi-Unit Abutments are similar to the designs of Multi-Unit Abutments cleared in K242340 and K240208. All Multi-Unit Abutments are manufactured from titanium alloy (Ti-6Al-4V).
The subject device straight Multi-Unit Abutments have a non-engaging, threaded design that attaches directly to the implant. The subject device straight (0°) Multi-Unit Abutments are included in this submission are similar in design to straight Multi-Unit Abutments cleared in K242340, K230143, and K222288. Subject device straight Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 1 mm to 5.5 mm.
The subject device angled Multi-Unit Abutments are provided only in an engaging design that requires an abutment screw. The subject device Multi-Unit Abutments angled 17° and 30° are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 2.5 to 4.5 mm.
Select straight and angled Multi-Unit Abutments have an added coating of zirconium nitride (ZrN). This coating is identical to the ZrN coating used on abutments with ZrN cleared in K242340.
DESS® Dental Smart Solutions Screws are designed to attach the abutment to the implant or the prosthesis to the abutment. There are a total of seventeen (17) subject device screws compatible with the subject device components or previously cleared components. The new screws have designs that are similar to those of screws cleared in K242340 and K240208. Screws are made of titanium alloy (Ti-6Al-4V). Select subject device screws are available with DLC (Diamond-like Carbon) coating and are identical to that used on screws cleared in K240208.
N/A
Ask a specific question about this device
(173 days)
Bonafix Implant Abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
All digitally designed superstructures, and/or hybrid crowns for use with Titanium Base are to be sent to a Bonafix validated milling center for manufacture.
Bonafix Abutment Solutions are a dental implant abutments system that includes three abutment design types (Temporary, Straight Ti-Base and Multi-Unit), that can be used to support single-unit or multi-unit prosthetic restorations. These abutments incorporating interface features compatible with sixteen (16) endosseous dental implant system platforms (Eight (8) designs from five (5) manufactures). The subject device abutments platform diameters range from 3.5mm to 6.3mm, and the corresponding compatible implant body diameters also range from 3.5 mm to 6.3mm. The system also includes corresponding abutment screws.
Temporary Abutments are manufactured from Titanium Grade 5 and consist of a coronal section, a platform and a connection part. The abutments are provided non-sterile with instructions for end user sterilization. The Temporary Abutments are seated in the implant with a prosthetic screw which is also manufactured from Titanium Grade 5. The prosthetic screw is delivered with the abutment. Angular correction of temporary abutments and placement of implant bodies at an angle is not allowed, they must be used parallel to the direction of occlusal loading forces. Temporary Abutments can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion. Modifications can be made by the end-user only in the minimum post height. Minimum post height – 4.0 mm. Temporary Abutments have a maximum duration of usage of 180 days.
The Bonafix TiBase abutment is composed of two-piece abutment that is a titanium base at the bottom and a zirconia superstructure (CAD/CAM patient specific superstructure) at the top. The superstructure are fabricated using a CAD/CAM process in Zirconia. The apical end is prefabricated to fit the compatible implant platform, as shown above, and is available with implant connections for crowns (with socket) or bridges (without socket). Angular correction of abutments and placement of implant bodies at an angle is not allowed, they must be used parallel to the direction of occlusal loading forces. Each abutment is provided with a screw designed to fit the compatible implant. The design parameters for the fabrication of the zirconia superstructure, which are already locked in the CAD/CAM software, are as follows: Minimum wall thickness – 0.45 mm, Minimum abutment post height – 4.0 mm, Maximum abutment post height – 6.5mm, Maximum gingival height – 5.0 mm, Minimum gingival height – 0.7 mm, Angulation – 0°. All digitally designed superstructures, and/or hybrid crowns for use with Bonafix TiBase abutments are to be sent to a Zentek validated milling center for manufacture. All superstructures are to be manufactured from zirconia conforming to ISO 13356. The required cement for bonding the zirconia superstructure to the Bonafix TiBase abutments to create the final two-piece abutment is G-CEM LinkAce™ cleared in K120243. Bonafix Ti-Base Abutments are made of titanium alloy (Ti-6Al-4V) conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) with the superstructure made of zirconia conforming to ISO 13356 Implants for surgery – Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).
The Multi-unit abutments are screw-attached to the implant to restore fully or partially edentulous arches. Their use is not intended for single-unit crowns. They are designed with various gingival heights (1.5, 2.5 and 3.5mm) and can be screwed directly to a compatible implant. They use a titanium Multi-unit sleeve, which is cemented to the prosthesis. Together with a titanium prosthetic screw, the prosthesis is ready to be screwed onto the Multi-unit abutment, which in turn is screwed onto the implant, thus completing the restoration. This multi-unit abutments are not intended to provide angle or divergence correction. For all system, are manufactured from Titanium Grade 5. Multi-unit sleeve can only be modified in the post height and that the minimum post height is 4mm from gingival collar.
The Prosthetic screw is used to seat all abutments for the system, included the temporary abutments and ti-bases, to the dental implant. They are provided along the prosthetic components, but they are also provided as standalone screws. The Abutment screws are manufactured from titanium alloy conforming to ASTM F136.
All system abutments are provided non-sterile with instructions for end user steam sterilization. The subject devices are provided non-sterile and required to be sterilized. All Bonafix TiBase abutments are packaged into PET bag.
N/A
Ask a specific question about this device
(80 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 and K240982 to allow additional options of zirconia material, scanners, CAM software, and milling machines to the digital dentistry workflow. The subject devices are to be sent to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. There are no changes to the abutment design, implant compatibilities, or design parameters. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows (also referred to as point of care) under K221301 and K240982.
The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.
The provided 510(k) summary for DESS Dental Smart Solutions focuses on demonstrating substantial equivalence to predicate devices for dental implant abutments. It primarily addresses the expansion of compatible materials, scanners, CAM software, and milling machines within an existing digital dentistry workflow. The document does not describe an AI/ML-based device that would typically have acceptance criteria related to diagnostic performance.
Therefore, many of the requested items related to AI/ML device performance (like acceptance criteria for diagnostic metrics, sample size for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this submission.
The acceptance criteria and supporting "study" (non-clinical data) for this device are related to its mechanical performance, biocompatibility, and integration within the digital workflow, demonstrating that the expanded components maintain the safety and effectiveness of the previously cleared predicate devices.
Here's a breakdown based on the information provided and the non-applicability of AI/ML-specific questions:
1. A table of acceptance criteria and the reported device performance
Since this is not an AI/ML diagnostic device, the acceptance criteria are not in terms of traditional diagnostic metrics (sensitivity, specificity, AUC). Instead, they are related to material properties, mechanical integrity, and the digital workflow's accuracy.
| Acceptance Criteria Category | Reported Device Performance (Summary from Submission) |
|---|---|
| Mechanical Integrity / Compatibility | - Fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions. (Implies successful completion to similar or better standards than predicate.) |
| Biocompatibility | - Biocompatibility testing according to ISO 10993-5 and ISO 10993-12. (Implies successful demonstration of biocompatibility.) |
| Sterilization | - Sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937. (Implies successful validation for sterilization.) |
| CAD Design Restrictions | - Software verification included testing of restrictions that prevent design of components outside of the stated design parameters. - Abutment design library validated to demonstrate established design limitations are locked and cannot be modified by the user. (Implies successful implementation and verification of design constraints.) |
| CAM Restriction Zones / Manufacturing Accuracy | - Validation testing of CAM restriction zones conducted, including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in CAM software. (Implies successful validation to ensure manufacturing precision and prevent damage.) |
| Material Conformance | - Zirconia materials conform to ISO 6872. - Titanium alloy conforms to ASTM F136. - Co-Cr-Mo alloy conforms to ASTM F15337. (Implies materials meet standards.) |
| Physical Dimensions | - Device encompasses the same range of physical dimensions as the predicate device. (Implies dimensional equivalence.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated in terms of a "test set" for diagnostic performance. The validation involved physical testing of components (e.g., fatigue testing) and software verification. The specific number of abutments or digital design instances used for these non-clinical tests is not detailed in this summary.
- Data Provenance: Not applicable in the context of patient data for an AI/ML device. The "data" here refers to engineering and material testing results, likely conducted in controlled lab environments (implied to be in accordance with international standards like ISO and ASTM). The manufacturer is Terrats Medical SL, in Spain, so testing would likely originate from their facilities or contracted labs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device relates to engineering specifications and material science.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a diagnostic device involving expert review adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For this device, the "ground truth" is based on:
- Engineering Specifications: Defined design parameters (e.g., minimum wall thickness, post height, angulation limits).
- Material Standards: Conformance to international standards such as ASTM F136, ISO 6872.
- Benchmarking/Predicate Equivalence: Performance is assessed against established performance of the predicate devices and OEM implant systems.
- Software Validation Logic: Verification that software correctly enforces design rules and CAD/CAM restrictions.
8. The sample size for the training set
Not applicable. This device does not involve a machine learning training set.
9. How the ground truth for the training set was established
Not applicable. This device does not involve a machine learning training set.
Ask a specific question about this device
(138 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K221301 to add the ability for the subject device Base Abutments and Pre-milled (Blank) Abutments to be designed using AbutmentCAD software in the digital dentistry workflow, as well as add angulation to some of the Pre-Milled (Blank) Abutments. The subject devices are to Terrats Medical validated milling centers for manufacture, or to be designed and manufactured via a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, titanium and ceramic material, milling machine, and associated tooling and accessories. The proposed change is to allow the subject devices to be designed using AbutmentCAD by exocad GmbH, the current workflow allows only the use of 3Shape by 3Shape A/S for the design software. Another purpose of this submission is to expand the design parameters to allow angulation (up to 30°) on Pre-milled (Blank) Abutments that are compatible with Neodent Grand Morse, Nobel Active/Nobel Parallel Conical 3.0 mm, and Straumann BLX implants. There are no changes to the abutment design or implant compatibilities. All part numbers have been cleared for manufacturing via a validated milling center and digital dentistry workflows in under K221301.
The subject device DESS Dental Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a validated milling center or a digital dentistry workflow. A validated milling center will be under FDA quality system regulations. The digital dentistry workflow scans files from intra-oral and lab (desktop) scanners, CAD software, titanium and ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The digital workflow includes the following products (not subject devices of this submission):
- Ceramic material: VITA YZ ST and VITA YZ XT (K180703)
- . Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436)
- . Intraoral Scanner: 3Shape TRIOS A/S Series Intraoral Scanner (510(k) exempt under 21 CFR 872.3661)
- Desktop scanner: 3Shape D900 Dental Lab Scanner (510(k) exempt under 21 CFR 872.3661)
- Abutment design software: 3Shape Abutment Designer Software (K151455) and AbutmentCAD ● (K193352)
- . Milling machine: VHF R5 by vhf camfacture AG with DentalCAM and DentalCNC 7 software
The provided text describes a 510(k) premarket notification for DESS Dental Smart Solutions, which are dental implant abutments. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device's effectiveness through clinical performance studies with specific statistical acceptance criteria for accuracy, sensitivity, or specificity.
Therefore, the document does not contain the information requested regarding:
- A table of acceptance criteria and reported device performance (in terms of clinical metrics like accuracy, sensitivity, specificity).
- Sample size used for the test set or its provenance.
- Number of experts used to establish ground truth or their qualifications.
- Adjudication method for the test set.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study or its effect size.
- Standalone (algorithm only) performance.
- Type of ground truth used (expert consensus, pathology, outcomes data).
- Sample size for the training set.
- How the ground truth for the training set was established.
The study described in this document focuses on non-clinical performance data to demonstrate substantial equivalence, specifically:
- Sterilization validation: According to ISO 17665-1, ISO 17665-2, and ISO 14937.
- Biocompatibility testing: According to ISO 10993-5 and ISO 10993-12.
- Fatigue testing and reverse engineering analysis: Of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility. This includes fatigue testing of OEM implant bodies with patient-specific abutments made at worst-case angled conditions.
- MR Conditional labeling.
- Validation testing of CAM restriction zones: Including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in the CAM software.
- Software verification: Included testing of restrictions that prevent design of components outside of the stated design parameters. The abutment design library was validated to demonstrate that established design limitations and specifications are locked and cannot be modified by the user.
The acceptance criteria and reported "performance" for this submission are based on these engineering and design validations, ensuring the device meets safety and performance standards equivalent to the predicate device, K221301. The key "performance" metrics are about maintaining physical and material integrity and compatibility.
The core of the submission is to expand the DESS Dental Smart Solutions abutment system to:
- Allow design using AbutmentCAD software (in addition to 3Shape software).
- Add angulation (up to 30°) to some Pre-milled (Blank) Abutments for specific implant systems.
The document explicitly states: "No clinical data were included in this submission." and "The subject device, the predicate device, and reference devices have the same intended use, technological characteristics, and are materials. The subject device, the predicate device, and reference devices encompass the same range of physical dimensions, manufactured by similar methods, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above."
Therefore, this FDA submission is for a physical medical device (dental implant abutment) and its manufacturing/design software modifications, not an AI or diagnostic device that would involve clinical performance metrics like sensitivity or specificity.
Ask a specific question about this device
(99 days)
IPD Dental Implant Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single or multiple dental prosthetic restorations.
IPD Dental Implant Abutments is a dental implant abutment system composed of dental abutments and screws intended to be placed into dental implants to provide support for dental prosthetic restorations. Abutments provide basis for single or multiple tooth prosthetic restorations. They are available in a variety of connection types to enable compatibility with commercially available dental implants systems. IPD Dental Implant Abutments includes the following categories of dental abutment designs: Healing abutments; Temporary abutments; Cementing titanium abutments; Titanium base (interface) abutments; The system also includes the corresponding Titanium Screws intended to attach the prosthesis to the dental implant.
This is a Premarket Notification (510(k)) summary for the IPD Dental Implant Abutments. This document does not describe a study proving the device meets acceptance criteria as would be found in a clinical trial for a novel AI device with specific performance metrics. Instead, it demonstrates substantial equivalence to predicate devices based on bench testing and engineering analysis.
Here's an breakdown of the information provided, tailored to your request, but emphasizing that this is not a typical AI/ML performance study:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a 510(k) for dental abutments, the "acceptance criteria" are not reported as specific performance metrics (like sensitivity/specificity for an AI model), but rather as compliance with recognized standards and demonstration of substantial equivalence to predicate devices. The "reported device performance" refers to successful completion of various non-clinical tests.
| Acceptance Criteria (Demonstrated Equivalence/Compliance) | Reported Device Performance (Test Results) |
|---|---|
| Same intended use as predicate device | Intended use is identical |
| Same operating principle as predicate device | Operating principle is identical |
| Similar design and same device categories | Very similar design, same categories |
| Same materials and surface coatings | Identical materials and surface coatings |
| Same sterilization processes | Identical sterilization processes |
| Sterilization validation (SAL of 1 x 10^-6) | Met ISO 17665-1 requirements |
| Biocompatibility: Cytotoxicity | Non-cytotoxic (ISO 10993-5 met) |
| Biocompatibility: Sensitization (TiN coated) | Met ISO 10993-10 requirements |
| Biocompatibility: Irritation (TiN coated) | Met ISO 10993-23 requirements |
| Mechanical compatibility (with OEM implants) | Confirmed by reverse engineering/dimensional analysis |
| Validation of digital workflow and software system | Ensured design/manufacturing within parameters |
| Static and dynamic fatigue testing | Met ISO 14801 requirements (worst-case configurations) |
| MRI safety (magnetically induced displacement force/torque) | Reviewed using scientific rationale and published literature, parameters addressed per FDA Guidance |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable in the context of human data or AI model test sets. The "test set" here refers to the physical samples of the IPD Dental Implant Abutments (and their components, potentially in various configurations) that were subjected to the specified bench and biocompatibility tests. The exact number of physical abutments, screws, or zirconia superstructures tested for each specific criterion is not detailed in this summary but would be found in the full test reports.
- Data Provenance: Not applicable in the context of country of origin for patient data. The tests were performed on the manufactured device components. The materials are specified (e.g., Titanium alloy conforming to ISO 5832-3).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. "Ground truth" in this context would generally refer to expert-derived labels for clinical data, which is not part of this 510(k) submission. The "truth" for the performance tests is established by adherence to recognized international standards (e.g., ISO, FDA Guidance) and objective measurements from engineering or biological tests.
4. Adjudication method for the test set
Not applicable. Adjudication methods (like 2+1 or 3+1) are used for resolving disagreements among human readers or expert labelers of clinical data, which is not relevant here. The evaluation criteria for the bench tests are objective and defined by established standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. This device is a dental implant abutment, not an AI software. Therefore, MRMC studies and AI assistance effect sizes are not relevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This is a physical medical device (dental abutments), not a standalone algorithm.
7. The type of ground truth used
The "ground truth" for this device's performance is established by:
- Compliance with international standards (e.g., ISO 14801 for fatigue testing, ISO 10993 series for biocompatibility, ISO 17665-1 for sterilization).
- Engineering specifications and dimensional analysis to ensure compatibility with other specified dental implant systems.
- Validated digital dentistry workflow parameters (e.g., minimum gingival height, wall thickness, post height, angulation for zirconia superstructures), which are based on established dental prosthetic principles.
8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. No training set is used for this physical device.
Ask a specific question about this device
(90 days)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.
The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes dental implants, abutments, and prosthetic components cleared previously in various submissions. The previously cleared abutments and prosthetic components are compatible with a variety of original equipment manufacturer (OEM) dental implants as well as DESS Dental Smart Solutions dental implants.
This submission adds various abutments to the DESS and OEM implant lines as summarized on the following pages in Table 1 Summary of Subject Device Abutment Designs, and Table 2 Summary of Subject Device Abutment Sizes.
The subject device abutment designs include Cover Screws, Healing Abutments, Temporary Abutments, Ti Base Abutments, AURUM Base Abutments (Ti Base abutments with a screw channel design that allows easier instrument access to the abutment screw), CoCr Base Abutments, Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), DESSLoc Abutments (Locator-type abutments), and abutment screws.
This document is a 510(k) summary for the DESS Dental Smart Solutions, a line of endosseous dental implant abutments. It details the device's characteristics, intended use, and a comparison to predicate devices to establish substantial equivalence.
Based on the provided text, the device itself is a physical medical device (dental implant abutments and associated components), not a software or AI-driven system. Therefore, the questions related to AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, and ground truth establishment for AI would not be directly applicable to this product as described.
The document focuses on establishing substantial equivalence to previously cleared predicate devices through:
- Identical Intended Use: The device is intended for "functional and esthetic rehabilitation of the edentulous mandible or maxilla" by providing support for prosthetic restorations, which is the same as the predicate devices.
- Similar Technological Characteristics: The device utilizes similar designs, materials (Ti-6Al-4V, Co-Cr-Mo alloy), manufacturing processes, and sterilization methods as its predicates.
- Performance Data: Non-clinical testing (mechanical testing per ISO 14801, MR environment assessment) and comparison to existing data from predicate devices are used to demonstrate safety and effectiveness.
Therefore, many of the requested elements for describing AI acceptance criteria and studies are not present or applicable in this document.
However, I can extract information relevant to the device's performance assessment and criteria for its type of submission.
Here's an attempt to answer the questions based solely on the provided text, acknowledging that the nature of the device (a physical implant component) means many AI-specific questions will be answered as "Not Applicable" (N/A):
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantifiable "acceptance criteria" in a table format with corresponding "reported device performance" metrics for this specific 510(k) submission. Instead, it relies on demonstrating substantial equivalence to existing predicate devices.
The underlying "acceptance criteria" for demonstrating substantial equivalence for this device type are primarily through:
- Mechanical Testing (ISO 14801): This is a critical performance standard for dental implants and abutments. The document states that mechanical testing was "conducted according to ISO 14801 to support the performance." The acceptance criteria would be successful completion of these tests, demonstrating the device's mechanical strength and fatigue resistance are comparable to or better than predicate devices. The reported performance is simply that the tests supported the performance.
- Material Conformance: Materials must conform to specific ASTM standards (e.g., ASTM F136 for Ti-6Al-4V, ASTM F1537 for Co-Cr-Mo). The reported performance is that the materials conform to these standards.
- Biocompatibility: While not detailed in this excerpt, the mention of "biocompatibility" in relation to predicates implies conformance to relevant biocompatibility standards (e.g., ISO 10993 series). The reported performance is that it is compatible.
- Sterilization Validation: Demonstrated sterility assurance level (SAL) of 10⁻⁶ via validated methods (moist heat or gamma irradiation). The reported performance is that validation was performed and met this SAL.
- Dimensional Compatibility: The abutments must fit the corresponding OEM implants correctly. The reported performance is that reverse engineering dimensional analysis confirmed compatibility.
Due to the nature of the document being a 510(k) summary focusing on substantial equivalence rather than a full study report, specific numerical performance results for the device tests are not provided in this text.
2. Sample sizes used for the test set and the data provenance
- Sample Size for Mechanical Testing: The document states "mechanical testing conducted according to ISO 14801." For such tests, ISO 14801 typically specifies minimum sample sizes (e.g., 10-11 samples for static strength, typically more for fatigue). The exact number of samples used for this specific submission is not explicitly stated, but it would have followed the standard's requirements.
- Data Provenance: The mechanical testing and material analyses are assumed to be "non-clinical data submitted or referenced" by the manufacturer, Terrats Medical SL, based in Barcelona, Spain. The "reverse engineering dimensional analysis" was done by Terrats Medical SL or through contractual agreement. This is prospective testing performed to support the 510(k). The document itself does not specify the country of origin for the underlying OEM implant data used for reverse engineering, although the OEM companies are listed (e.g., Astra Tech AB, BioHorizons).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A. This is a physical device. Ground truth, in the context of AI, refers to validated labels for data used to train and test an algorithm. For a physical device, performance is evaluated through engineering and biocompatibility testing against defined standards. There are no "experts" establishing ground truth in the AI sense. Testing would be performed by qualified engineers and technicians.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
N/A. Adjudication methods are typically used in clinical studies involving interpretation (e.g., by radiologists) to resolve discrepancies. This document describes non-clinical performance testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A. This product is a dental implant abutment, not an AI software intended to assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
N/A. This is a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
N/A. For engineering tests of physical devices, the "ground truth" is typically derived from established engineering principles, international standards (e.g., ISO 14801 for mechanical properties, ASTM for materials), and the physical properties of the materials and designs themselves. There isn't "expert consensus" or "pathology" in the AI or clinical trials sense.
8. The sample size for the training set
N/A. This is a physical device; there's no "training set" in the machine learning sense. The device is manufactured based on established engineering designs and material specifications.
9. How the ground truth for the training set was established
N/A. No training set for AI. For device manufacturing, the "ground truth" for design and production parameters comes from established engineering best practices, prior successful device designs (predicate devices), and adherence to quality systems regulations (21 CFR Part 820).
Ask a specific question about this device
(560 days)
Medentika abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Medentika abutments for the Dentsply Sirona Astra Tech OsseoSpeed EV 3.0mm and TX 3.0mm implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Medentika TiBase CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. Medentika TiBase is intended for use with the Straumann® CARES® System. All digitally designed copings and/or crowns are intended to be sent to Straumann for manufacture at a validated milling center.
Medentika abutments for the Nobel Biocare Nobel Active®* 3.0mm, Dentsply Sirona Astra Tech OsseoSpeed EV®* 3.0mm and TX®* 3.0mm implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Medentika PreFace CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Medentika Preface is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center. The final patient matched form is a MedentiCAD abutment.
Medentika abutments for the Dentsply Sirona Astra Tech OsseoSpeed EV 3.0mm implant bodies are indicated for maxillary lateral and mandibular central/lateral incisors only.
Multi-unit abutments are indicated for use with dental implants as a support for multi-unit screw retained bridges and bars in the maxilla or mandible of a partially or fully edentulous patient.
The Medentika abutments include abutments, abutment screws, caps, and bases which are labelled under a specific Medentika series and are compatible with a specified dental implant system. The abutments include sinqle-unit abutments intended for use with dental implants as a support for single or multiple tooth protheses in the maxilla or mandible of a partially or fully edentulous patient. The abutments also include multi-unit abutments indicated for use with dental implants as a support for multi-unit screw retained bridges and bars in the maxilla or mandible of a partially or fully edentulous patient.
The purpose of this premarket notification is to add additional abutments. The subject abutments include abutments compatible with additional dental implant systems forming a new Medentika series (the OT series). The subject abutments also include abutments compatible with new implant diameters in existing Medentika series (E, EV, F, and S). Lastly, the subject abutments include new abutment designs compatible with existing implant diameters in existing Medentika series (R).
This looks like a 510(k) Summary for a medical device (dental abutments), which means the document is about proving "substantial equivalence" to a predicate device, not about proving clinical effectiveness or performance against pre-defined acceptance criteria in the way one might for a novel AI/software medical device.
Therefore, the information requested in your bullet points (e.g., acceptance criteria table, sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set details) is not applicable to this type of regulatory submission because the device is a mechanical one, not an AI/software device. The data presented here is focused on demonstrating physical and mechanical compatibility and equivalence to previously cleared devices.
Here's why each point is not applicable and what information is provided:
-
A table of acceptance criteria and the reported device performance: This document doesn't provide a typical "acceptance criteria" table as would be seen for an AI/software device measuring diagnostic performance (e.g., sensitivity, specificity, AUC). Instead, it relies on demonstrating that the new abutments perform similarly to existing, cleared abutments through "dynamic fatigue testing" and "dimensional analysis and reverse engineering." The performance is implicitly "accepted" if these tests show equivalence to the predicate.
-
Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): No "test set" in the context of diagnostic performance is mentioned. The "testing" refers to non-clinical, physical testing (fatigue, dimensional analysis). There is no patient data involved in this type of submission for a mechanical device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. Ground truth, in the AI/software sense, is not established for this device. The "truth" is based on engineered specifications and physical testing.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No human adjudication of diagnostic output is relevant here.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "ground truth" is engineering specifications and physical measurements.
-
The sample size for the training set: Not applicable. There is no AI/machine learning component to "train."
-
How the ground truth for the training set was established: Not applicable for the same reason as above.
What the document does provide regarding device performance and testing:
-
Type of Testing:
- Dynamic fatigue testing according to FDA guidance and ISO 14801 (Dentistry - implants dynamic loading test for endosseous dental implants).
- Dimensional analysis and reverse engineering of the implant-to-abutment connection platform.
- Sterilization validation (steam and gamma irradiation) referenced from K191123, ISO 17665-1, ISO/TS 17665-2, ISO 11137-1, ISO 11137-2.
- Sterile packaging validation referenced from K191123, ISO 11607-1, ISO 11607-2.
- Biocompatibility evaluations referenced from K142167, K170838, K191123, K150203, K061804 in accordance with ISO 10993-1.
- MR testing referenced from K180564 in accordance with ASTM F2052-15, ASTM F2213-06 (2011), ASTM F2182-11a, and ASTM F2119-13.
-
Conclusion: The tests demonstrated "implant to abutment compatibility" and "established substantial equivalency of the proposed device with predicate devices." This is the "proof" that the device meets the (implicit) acceptance of being substantially equivalent to existing, legally marketed devices.
In summary, this document is for a traditional mechanical medical device, and the regulatory pathway does not involve performance studies in the way you've outlined for AI/software-based devices.
Ask a specific question about this device