K Number
K172505

Validate with FDA (Live)

Date Cleared
2017-12-28

(132 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MIS dental implant system is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore masticatory function. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and the occlusal load is appropriate.

Narrow implants (Ø3.3mm & UNO) are indicated for use in surgical and restorative applications for placement only in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws, to provide support for prosthetic devices such as artificial teeth, in order to restore the patient chewing function. Mandibular central and lateral incisors must be splinted if using two or more narrow implants adjacent to one another.

Device Description

The MIS C1 NP Conical Connection dental implants are manufactured from Ti-6AI-4V ELI. The conical root-shaped, screw-type implants are designed for both two-stage and single stage procedures, with one internal thread for screwed abutment. The implants are self-tapping, root-form with tapered threads. They have a 3.3 mm diameter and the following lengths: 10mm, 11.5mm, 13mm and 16mm. The internal part of the implant and the lower part of the abutment are anodized for coloring purposes for a quick identification of the diameter and to ensure the adequate abutments. The Implants are supplied sterile.

This submission also includes wide platform (WP) conical connection abutments which are compatible to the previously cleared (K112162) MIS conical connection wide platform implants.

Components:
The MIS C1 Conical Connection Dental Implant is to be used in combination with a variety of conical connection abutments (cover screws, healing caps, cement-retained abutments, gold abutments, OT-equators & ball attachments, multi unit abutments, and temporary Ti and PEEK abutments). These abutments are manufactured with a conical connection, ensuring compatibility to the conical connection implants. The abutments in this submission include NP abutments, compatible with the C1 NP implants also submitted here; and WP abutments compatible with previously cleared (K112162) MIS conical connection WP implants.

AI/ML Overview

The acceptance criteria described in this document are for the MIS C1 Narrow Platform Conical Connection Implant System and MIS C1 Wide Platform Conical Connection Abutments.

Here's an analysis of the provided information:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes several non-clinical performance tests conducted to demonstrate substantial equivalence. The "acceptance criteria" are implied by successful completion of these tests in accordance with relevant ISO standards.

Acceptance Criteria (Implied by standard)Reported Device Performance (as stated in the document)
Biocompatibility: Device materials and manufacturing processes are safe for biological use. (Implied: equivalent to predicate with established biocompatibility)"The subject device is manufactured using identical manufacturing methods... and using the same raw material as the previously cleared predicate, K163349. The subject device is sterilized and packaged using identical materials and processing as the predicate. Finally, the subject device has the same intended use, patient contact duration and type as the predicate. For these reasons, biocompatibility testing was not required to support the substantial equivalence of the subject device."
Fatigue Testing (Implants): Withstand 5,000,000 cycles without failure at a load equivalent to predicate devices per ISO 14801:2007."The test articles were able to withstand 5,000,000 cycles without failure at a substantially equivalent load to the cited predicates."
Fatigue Testing (Abutments): Withstand fatigue limits (specific criteria not detailed, but 'met success criteria')."The subject worst case abutment was tested for fatigue limits and met the pre-determined success criteria."
Sterilization (Gamma Irradiation): A Sterility Assurance Level (SAL) of 10⁻⁶ achieved, all testing requirements met per ANSI/AAMI/ISO 11137-1 and 11137-2."Test results have demonstrated that the SAL of 10-6 was achieved and all testing requirements were met."
Sterilization (Steam - User): Steam sterilization parameters validated per ANSI/AAMI/ISO 17665-1:2006 and ANSI/AAMI/ISO 17665-2:2009."The steam sterilization parameters were validated according to ANSI/AAMI/ISO 17665-1:2006 and ANSI/AAMI/ISO 17665-2:2009 for two methods: gravity displacement steam sterilization and pre vacuum steam sterilization."
Endotoxin Limit: Within acceptance criteria per USP 85, USP 161, and ANSI/AAMI/ ST72 for sterile products."For products supplied sterile, a LAL test is conducted periodically to verify the endotoxin limit is within acceptance criteria according to USP 85, USP 161 and ANSI/AAMI/ ST72."
Disinfection Validation: Reduction of at least 10⁶ of microbiological challenge per ANSI/AAMI/ISO 11737-1:2006 (R)2011, AAMI TIR 30:2011, and AAMI TIR 12:2010 for non-sterile abutments."demonstrating a reduction of at least 106 of the microbiological challenge."
Shelf Life Testing: Package integrity demonstrated for the specified shelf life per ISO 11607-1."Test results were successful and supported a 5 year shelf life of the sterilized products."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact numerical sample size for the fatigue testing or sterilization validation. It refers to "the worst case narrow platform implant and abutment" for fatigue testing. For sterilization, it mentions "each group of products" and "abutments supplied non-sterile".
  • Data Provenance: The studies are non-clinical (laboratory testing). The document heavily references ISO and AAMI standards. The manufacturer is "MIS Implants Technologies Ltd." based in "Bar Lev Industrial Park, ISRAEL". This suggests the testing was likely conducted in Israel or by contract labs following international standards. The document does not mention retrospective or prospective data as it pertains to clinical studies, which were not performed.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable as the ground truth for these tests is established by physical measurements and compliance with engineering and sterilization standards, not expert clinical consensus.

4. Adjudication Method

This section is not applicable as the "ground truth" (i.e., test results meeting predefined acceptance criteria) is determined by objective physical and biological testing, not by expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device."

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No, a standalone "algorithm only" study was not done. This device is a physical medical implant, not an AI or software-as-a-medical-device (SaMD) product. Performance is determined by non-clinical engineering and biological testing.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical tests described is based on:

  • Engineering Standards: For mechanical fatigue, compliance with ISO 14801:2007.
  • Sterilization and Biocompatibility Standards: For sterilization (ANSI/AAMI/ISO 11137-1, 11137-2, 17665-1, 17665-2), endotoxin testing (USP 85, USP 161, ANSI/AAMI/ ST72), disinfection (ANSI/AAMI/ISO 11737-1:2006 (R)2011, AAMI TIR 30:2011, AAMI TIR 12:2010), and biocompatibility (implied equivalence to previously cleared predicate).
  • Shelf Life Standards: For package integrity, compliance with ISO 11607-1.

These are objective, measurable criteria established by recognized national and international standards.

8. The Sample Size for the Training Set

This section is not applicable. The device is a physical dental implant system and does not involve machine learning or AI, thus there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reasons as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

MIS Implants Technologies Ltd. % Randy Prebula Partner Hogan Lovells US LLP 555 13th Street, NW Washington, District of Columbia 20004 December 27, 2017

Re: K172505

Trade/Device Name: MIS C1 Narrow Platform Conical Connection Implant System MIS C1 Wide Platform Conical Connection Abutments Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: November 30, 2017 Received: November 30, 2017

Dear Randy Prebula:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mary S. Runner -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172505

Device Name

MIS C1 Narrow Platform Conical Connection Implant System MIS C1 Wide Platform Conical Connection Abutments

Indications for Use (Describe)

MIS dental implant system is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore masticatory function. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and the occlusal load is appropriate.

Narrow implants (03.3mm & UNO) are indicated for use in surgical and restorative applications for placement only in the mandibular central. lateral incisor and maxillary lateral incisor regions of partially edentulous jaws, to provide support for prosthetic devices such as artificial teeth, in order to restore the patient chewing function. Mandibular central and lateral incisors must be splinted if using two or more narrow implants adjacent to one another.

Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MIS Implants Technologies Ltd. The logo features the letters "MIS" in a stylized font, with the "M" and "I" connected and the "S" slightly offset. The text "Implants Technologies Ltd." is written in a smaller font to the right of the logo.

K172505 510(k) Summary

1. Submitter

MIS Implants Technologies Ltd. P.O. Box 7, Bar Lev Industrial Park 2015600 ISRAEL Telephone: +972-4-9016800 Fax: +972-4-9918623

Contact: Arbel Shezaf +972-4-9016802 arbel@mis-implants.com

Date Prepared: December 28, 2017

2. US Agent: Motti Weisman - VP Marketing

MIS Implants Technologies Inc. 18-00 Fair Lawn Ave Fair Lawn NJ 07410 Phone: (201) 797-9144 Fax: (201) 797-9145

3. Device Identification

Trade/Proprietary Name: MIS C1 Narrow Platform Conical Connection Implant System MIS C1 Wide Platform Conical Connection Abutments Common/Usual Name: Dental Implant Classification Name: Endosseous dental implant Requlation Number: 872.3640; Product Code: DZE, NHA Device Class: Class II Classification Panel: Dental Devices Panel

4. Predicate Device(s)

Primary predicate device: ANKYLOS C/X Implant System cleared under 510(k) K140347. Reference devices:

  • A.B. DENTAL DEVICES® Dental Implants System K162482 .
  • MIS V3 Conical Connection Dental Implant System K163349 .
  • MIS Conical Connection Implants K112162
  • Straumann NNC Protective Caps K113410 .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MIS Implants Technologies Ltd. The logo consists of the letters "MIS" in a stylized font, with the "I" having a vertical line extending above the other letters. To the right of the logo is the text "Implants Technologies Ltd" in a smaller font.

5. Device Description

The MIS C1 NP Conical Connection dental implants are manufactured from Ti-6AI-4V ELI. The conical root-shaped, screw-type implants are designed for both two-stage and single stage procedures, with one internal thread for screwed abutment. The implants are self-tapping, root-form with tapered threads. They have a 3.3 mm diameter and the following lengths: 10mm, 11.5mm, 13mm and 16mm. The internal part of the implant and the lower part of the abutment are anodized for coloring purposes for a quick identification of the diameter and to ensure the adequate abutments. The Implants are supplied sterile.

This submission also includes wide platform (WP) conical connection abutments which are compatible to the previously cleared (K112162) MIS conical connection wide platform implants.

Components:

The MIS C1 Conical Connection Dental Implant is to be used in combination with a variety of conical connection abutments (cover screws, healing caps, cement-retained abutments, gold abutments, OT-equators & ball attachments, multi unit abutments, and temporary Ti and PEEK abutments). These abutments are manufactured with a conical connection, ensuring compatibility to the conical connection implants. The abutments in this submission include NP abutments, compatible with the C1 NP implants also submitted here; and WP abutments compatible with previously cleared (K112162) MIS conical connection WP implants.

  • Cover screws and healing caps are premanufactured prosthetic components directly connected to ı the endosseous dental implants and are indicated as temporary components to allow healing of the soft tissue. They are made of Ti-6Al-4V ELI, and supplied sterile to the user, for single use.
  • । Cement- Retained Abutments are premanufactured dental implant abutments directly connected to the endosseous dental implant by a prosthetic screw. Cement retained abutments are available straight or angulated, in different heights and diameters. The angulated abutments allow a maximum anqulation of 25°. The straight abutments are intended for 0 degree angulation and straight implantation only. They are made of Ti6Al-4V ELI, and supplied non sterile, to be steam sterilized by the user according to the labeling, intended for single use.
  • । CPK abutments are cement retained abutments intended to be used in temporary and permanent prosthetic rehabilitation. CPK abutments are intended for 00 angulation and straight implantation only. They are supplied non sterile, to be steam sterilized by the user according to the labeling, and intended for single use.
  • Plastic healing caps are premanufactured abutments intended to cover the CPK abutment until the permanent restoration is ready. They are made from PEEK and intended to be used for up to 180 days. The plastic healing caps are delivered non sterile to the user, for single use, to be sterilized by steam sterilization. Different caps are available for use with narrow, standard, and wide platforms.
  • Gold Plastic abutments are intended for permanent restoration, for either single or multiple tooth । screw retained restorations. The lower part of the abutment which connects directly to the implant is made of gold alloy, and the upper part is made of plastic part is dissolved once the casting is done. Gold abutments are intended for 0° angulation and straight implantation only. The

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letters 'mis' in blue. The 'm' and 's' are connected by a curved line at the bottom. To the right of the letters is the word 'Implementation' in a smaller font size. The logo is simple and modern.

Proponies | tri

abutments are supplied non sterile, to be steam sterilized by the user according to the labeling, and intended for single use.

  • । Multi-Unit abutments are indicated for multiple unit reconstructions when screw retained prosthetics is preferred. Multi-unit abutments allow either direct screw of the prosthesis into the multi-unit abutment or connection to a fixed overdenture bar. Multi-units are available in angulations between 17-309. The Multi units are made of Ti-6Al-4V ELI. They are supplied sterile and intended for single use.
  • OT-Equators & Ball Attachments are screw retained dental implant abutments connected directly to । the endosseous dental implant by their lower threaded part, and are used in completely edentulous jaws for anchoring an overdenture to allow its insertion and removal.
  • -Temporary abutments are premanufactured dental implant abutments directly connected to the endosseous dental implant, intended for use as an aid in temporary prosthetic rehabilitation, for a maximum of 180 days. They are available in Ti-6Al-4V ELI and in natural PEEK, and in anti-rotation and free-rotation. The abutments are supplied non sterilized by the user according to the labeling and intended for single use.

6. Indications for Use

MIS dental implant system is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore masticatory function. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and the occlusal load is appropriate.

Narrow implants (Ø3.3mm & UNO) are indicated for use in surqical and restorative applications for placement only in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws, to provide support for prosthetic devices such as artificial teeth, in order to restore the patient chewing function. Mandibular central and lateral incisors must be splinted if using two or more narrow implants adjacent to one another.

7. Substantial Equivalence Discussion

  • a. Implants: The predicate device for the MIS C1 NP conical connection dental implants is the ANKYLOS C/X (K140347). Both devices have similar indications, although the predicate device states that implants 6.6 mm in length are for two-stage surgical procedures and cemented, removable or screw retained restorations. The subject device does not contain this limitation as its shortest device is 10 mm in length. The length range of the subject device is within the range of the predicate. Although the diameter of the subject device is somewhat smaller, 3.3 vs 3.5 mm, it possesses at least equivalent strength as compared to the predicate device. Their geometrical design is similar with minor differences which do not raise different safety or efficacy questions. Fatique testing per ISO 14801:2007 assessed the impact of these differences and demonstrates at least equivalent performance. The predicate device is indicated for all positions in the jaws, while the subject C1 NP implants are indicated for the mandibular central, lateral incisor and maxillary lateral incisor regions, and must be splinted if using two or more narrow implants adjacent to one another. The differences do not alter the intended use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. A reference device, A.B. DENTAL DEVICES® Dental Implants System (K162482) with the same 3.3mm diameter and similar indications for use as the subject device is also

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MIS Implants Technologies Ltd. The logo consists of the letters "MIS" in a stylized, connected font, with the "i" having a dot above it. To the right of the logo is the text "Implants Technologies Ltd.", indicating the company's name and industry.

presented. A comparison of implant characteristics demonstrates substantial equivalence to the predicate device.

Table 1: Comparison of subject and predicate Implants

Trade NameMIS C1 NP ConicalConnection DentalImplantsANKYLOS C/X ImplantSystemA.B. DENTAL DEVICES®Dental Implants System
510(k) NumberK172505K140347K162482
ManufacturerMIS Implants Technologies Ltd.DENTSPLY ImplantsA.B. Dental Device Ltd.
Device ClassClass IIClass IIClass II
Product Code(s)DZEDZEDZE
RegulationDescriptionEndosseous dental implantEndosseous dental implantEndosseous dental implant
RegulationNumber872.3640872.3640872.3640
Intended use:Intended to be surgically placedin the bone of the upper orlower jaw arches for anchoringor supporting toothreplacement to restorechewing function.Intended to be surgicallyplaced in the bone of theupper or lower jaw arches foranchoring or supporting toothreplacement to restorechewing function.Intended to be surgically placed inthe bone of the upper or lowerjaw arches for anchoring orsupporting tooth replacement torestore chewing function.
Indications for use: MIS dental implant system isintended to be surgically placedin the bone of the upper orlower jaw arches to providesupport for prosthetic devices,such as artificial teeth, in orderto restore masticatory function.When a one-stage surgicalprocedure is applied, theimplant may be immediatelyloaded when good primarystability is achieved and theocclusal load is appropriate.Narrow implants (Ø3.3mm &UNO) are indicated for use insurgical and restorativeapplications for placement onlyin the mandibular central,lateral incisor and maxillarylateral incisor regions ofpartially edentulous jaws, toprovide support for prostheticdevices such as artificial teeth,in order to restore the patientchewing function. Mandibularcentral and lateral incisors mustbe splinted if using two or morenarrow implants adjacent toone another.ANKYLOS® C/X Implants of 8mm in length or longer are forsingle-stage or two-stagesurgical procedures andcemented, removable or screwretained restorations. TheANKYLOS® C/X Implants maybe used for immediateplacement and function onsingle tooth and/or multipletooth applications whenadequate primary stability isachievable, with appropriateocclusal loading, in order torestore chewing function.Multiple tooth applications maybe splinted.ANKYLOS® C/X Implants of 6.6mm in length are for two-stagesurgical procedures andcemented, removable or screwretained restorations. TheANKYLOS® C/X Implants may beused for immediate placement onsingle tooth and/or multiple toothapplications when adequateprimary stability is achievable,with appropriate occlusal loading,in order to restore chewingfunction.A.B. DENTAL DEVICES® DentalImplants System is indicated foruse in surgical and restorativeapplications for placement in thebone of the upper or lower jaw toprovide support for prostheticdevices, such as artificial teeth, inorder to restore the patient'schewing function. A.B. DENTALDEVICES® DentalImplants System is indicatedalso for immediate loading whengood primary stability isachieved and with appropriateocclusal loading.Two Stage Implants: I22, I5, I55,I10.P4 and P14 angled abutments areto be used only with standardplatform implants 3.5 mm indiameter or larger.
Multiple tooth applicationsmay be splinted.
Implant MaterialTitanium 6Al-4V ELI per ASTMF136CP Titanium grade 2Titanium 6Al-4V ELI
Surface TreatmentAnodized ,sand blasted andacid etchedSand blasted and acid etchedSand blasted with CaP, exceptfor 0.5mm distal part of the neckwhich is smooth.
Body designTapered design, threadedTapered design, threadedTapered design, threaded
Connection TypeConical with indexesConical with indexesConical with internal hex
Type of implantBone level implantBone level implantBone level implant
Implant Diameters3.3 mm3.5 mmØ 3.3 mm
Implant Lengths10, 11.5, 13 and 16 mm6.6, 8, 9.5, 11, 14, 17 mm10, 11.5, 13 and 16 mm
Neck DesignCylindricalCylindricalCylindrical
ApexDomed apexDomed apexFlat apex
ThreadDualDualDual
SterilizationMethodRadiationRadiationRadiation

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MIS Implants Technologies Ltd. The logo consists of the letters "mis" in a stylized font, with the "i" having a vertical line extending above the other letters. To the right of the letters is the text "Implants Technologies Ltd." in a smaller font. The logo is simple and modern, and the text is clear and easy to read.

  • b. Abutments: Healing caps, cover screws, cement retained abutments, CPK abutments, gold abutments, multi-units, temporary abutments, OT-equators and ball attachments were compared to equivalent MIS conical connection abutments, which share the same indications, are made of the same materials, manufactured in the same facility with the same manufacturing conditions and undergo the same surface treatments and were cleared under K163349. No new angulations were introduced. The subject worst case abutment was tested for fatigue limits and met the pre-determined success criteria. The differences between subject and predicate devices did not alter the intended use and new issues of safety and effectiveness were not raised.

Table 2 - Comparison of subject and predicate Abutments Characteristics

Trade Name MIS C1 Conical Connection Abutments MIS V3 Conical Connection Abutments
510(k) NumberSubjectK166349
ManufacturerMIS Implants Technologies Ltd.MIS Implants Technologies Ltd.
Device ClassClass IIClass II
Product Code(s)NHANHA
RegulationDescriptionEndosseous dental implant abutmentEndosseous dental implant abutment
RegulationNumber872.3630872.3630
Intended use:Dental implant abutments are intended to be usedin the upper or lower jaw used for supporting toothreplacements to restore chewing function. Theabutments in combination with two-stageendosseous implants are intended to be used as aDental implant abutments are intended to be used inthe upper or lower jaw used for supporting toothreplacements to restore chewing function. Theabutments in combination with two-stageendosseous implants are intended to be used as a

{8}------------------------------------------------

Prises Ltd

MIS C1 Conical Connection Abutments

MIS V3 Conical Connection Abutments

foundation for anchoring tooth replacements in either jaw. Restorations range from replacing one single tooth to fixed partial dentures using cementretained supra-constructions.

foundation for anchoring tooth replacements in either jaw. Restorations range from replacing one single tooth to fixed partial dentures using cementretained supra-constructions.

Indications foruse:MIS dental implant system is intended to besurgically placed in the bone of the upper or lowerjaw arches to provide support for prostheticdevices, such as artificial teeth, in order to restoremasticatory function. When a one-stage surgicalprocedure is applied, the implant may beimmediately loaded when good primary stability isachieved and the occlusal load is appropriate.Narrow implants (Ø3.3mm & UNO) are indicatedfor use in surgical and restorative applications forplacement only in the mandibular central, lateralincisor and maxillary lateral incisor regions ofpartially edentulous jaws, to provide support forprosthetic devices such as artificial teeth, in orderto restore the patient chewing function.Mandibular central and lateral incisors must besplinted if using two or more narrow implantsadjacent to one another.MIS V3 Conical Connection Dental Implant System isintended to be surgically placed in the bone of theupper or lower jaw arches to provide support forprosthetic devices, such as artificial teeth, in order torestore masticatory function. When a one-stagesurgical procedure is applied, the implant may beimmediately loaded when good primary stability isachieved and the occlusal load is appropriate.Narrow implants (Ø3.3mm) are indicated for use insurgical and restorative applications for placementonly in the mandibular central, lateral incisor andmaxillary lateral incisor regions of partiallyedentulous jaws, to provide support for prostheticdevices such as artificial teeth, in order to restorethe patient chewing function. Mandibular centraland lateral incisors must be splinted if using two ormore narrow implants adjacent to one another.
Material(s)Healing caps, cover screws, Prosthetic Screws, BallAttachment/Equators, cement retained abutments:TI 6Al-4V ELI per ASTM F136Gold plastic abutments: Gold AlloyPlastic cylinder: Polyoxymethylene (POM)Temporary plastic abutments and plastic healingcaps: PEEKHealing caps, cover screws, Prosthetic Screws, BallAttachment/Equators, cement retained abutments:TI 6Al-4V ELI per ASTM F136Gold plastic abutments: Gold AlloyPlastic cylinder: Polyoxymethylene (POM)Temporary plastic abutments: PEEK
SurfaceTreatmentTitanium: Polished and Anodized after machiningTitanium: Polished and Anodized after machining
AngulationNP: 0°, 10°, 20°WP: 0°, 15°, 17°, 25°, 30°NP 0°, 10°, 20°SP: 0°, 15°, 17°, 25°, 30°

8. Non-Clinical Performance Data

As part of demonstrating the substantial equivalence of the C1 NP Conical Connection Dental Implant system and the MIS WP conical connection abutments to the predicate devices, MIS Implants Technologies completed a number of non-clinical performance tests:

  • Biocompatibility The subject device is manufactured using identical manufacturing methods, in the . same manufacturing facility, and using the same raw material as the previously cleared predicate, K163349. The subject device is sterilized and packaged using identical materials and processing as the predicate. Finally the subject device has the same intended use, patient contact duration and type as the predicate. For these reasons, biocompatibility testing was not required to support the substantial equivalence of the subject device.
  • . Fatigue Testing – Mechanical testing of MIS C1 implants in accordance to ISO 14801:2007 was conducted. The worst case narrow platform implant and abutment chosen for the

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for MIS Implants Technologies. The logo consists of the letters "MIS" in a stylized, blue font. To the right of the letters, the words "Implants Tech" are written in a smaller, gray font. The logo is simple and modern.

mplants Technologies Ltd

test was the narrowest implant loaded with the abutment which has the greatest angulation. The test articles were able to withstand 5,000,000 cycles without failure at a substantially equivalent load to the cited predicates. Wide platform abutments were not tested for fatigue because they did not pose a new worst case compared to the cleared standard platform abutments in terms of raw material, diameter, wall thickness and moment arm.

  • Sterilization Testing ●
    • For products supplied sterilized by gamma irradiation (implants, cover screws, healing caps, O final drills, plastic cylinders, multi-units): Sterilization tests were conducted on each group of products in compliance with both ANSI/AAMI/ISO 11137-1 and ANSI/AAMI/ISO 11137-2. Test results have demonstrated that the SAL of 10-6 was achieved and all testing requirements were met.
    • For products supplied non-sterile and intended to be steam sterilized by the user (CPK kits o and CPK abutments, cement retained abutments, gold abutments, temporary abutments, ball attachments and OT-equators): The steam sterilization parameters were validated according to ANSI/AAMI/ISO 17665-1:2006 and ANSI/AAMI/ISO 17665-2:2009 for two methods: gravity displacement steam sterilization and pre vacuum steam sterilization.
  • . For products supplied sterile, a LAL test is conducted periodically to verify the endotoxin limit is within acceptance criteria according to USP 85, USP 161 and ANSI/AAMI/ ST72.
  • . Disinfection Validation: for Abutments supplied non-sterile and intended to be steam sterilized by the user, the disinfection procedure was validated in accordance with ANSI/AAMI/ISO 11737-1:2006 (R)2011, AAMI TIR 30:2011 and AAMI TIR 12:2010 by demonstrating a reduction of at least 106 of the microbiological challenge.
  • Shelf Life Testing - for the products which are supplied sterile, shelf life studies were completed by an independent testing laboratory in order to validate the integrity of the final package. The studies were conducted in accordance with ISO 11607-1. Test results were successful and supported a 5 year shelf life of the sterilized products.
  • Risk Analysis - Risk analysis for MIS implants was conducted in accordance with ISO 14971, Medical Devices: Application of Risk Analysis, for medical devices. It was determined by MIS that all risks associated with MIS implants were acceptable and as low as reasonably possible.

9. Clinical Performance Data

There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The non-dinical testing detailed in this submission supports the substantial equivalence of the device.

10. Summary

The comparison between the subject device and the predicate devices has shown that the indications for use, principles of operation, technological characteristics and materials were similar, and that the differences did not raise new safety and effectiveness issues. Furthermore, performance testing showed that the predicate device is at least equivalent to the predicates by means of performance.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for MIS Implants Technologies Ltd. The logo consists of the letters "MIS" in a stylized font, with the "i" having a dot above it. To the right of the letters is the text "Implants Technologies Ltd."

11. Conclusions

The MIS C1 NP conical connection implant system and WP abutments have the same intended use, incorporate the same fundamental technology, and have similar indications for use as the predicate. Test data to verify the performance of the MIS C1 NP conical connection implant system and WP abutments has been provided including: dynamic fatigue, sterilization, shelf life, and the results of this testing, combined with the design and intended use comparison with the predicate device, support substantial equivalence.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.