Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K162232

    Validate with FDA (Live)

    Date Cleared
    2016-11-17

    (101 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K092929

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MOSS 100 Pedicle Screw System is a thoracolumbar pedicle for the following indications of use:

    • Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
    • Spondylolisthesis
    • Trauma (i.e. fractures or dislocations)
    • Spinal Stenosis
    • Curvatures (i.e. scoliosis, kyphoses and/ or lordosis)
    • Tumor
    • Pseudoarthrosis
    • Failed previous fusion
    Device Description

    The Biedermann Motech MOSS 100 Pedicle Screw System is a comprehensive thoracolumbar spinal system that offers posterior clinical solutions. The MOSS 100 System is a universal set of instruments and implants that are indicated for the treatment of significant mechanical instability or deformity of the spine which requires fusion with instrumentation.

    The MOSS 100 System provides several options for stabilization of the spine. Therefore, the system includes the following implants:

    • -Pedicle Screws Ø5-8mm, length 25-80mm
    • -Rods Ø5.5mm, length 30-95mm, 300mm and 480mm
    • Locking caps -
    • Polyaxial Heads -

    The Polyaxial Heads are mounted onto the Pedicle Screws intra-operatively. The implants of the MOSS 100 System are single-use only and the system is provided non sterile.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device called the "MOSS 100 Pedicle Screw System." It does not describe a study involving an AI/Machine Learning device or analytical performance metrics typically associated with such software. Instead, it focuses on the performance of a physical pedicle screw system through non-clinical bench testing.

    Therefore, many of the requested elements for AI/ML device studies are not applicable to this document. I will extract the available information regarding acceptance criteria and performance for the pedicle screw system.

    Here's the information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria (Standard)Reported Device Performance
    Static CompressionASTM F 1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model)The results indicate that the MOSS 100 system is substantially equivalent to the identified predicate device (Expedium Spine System, K041119).
    Dynamic CompressionASTM F 1717The results indicate that the MOSS 100 system is substantially equivalent to the identified predicate device (Expedium Spine System, K041119).
    Static TorsionASTM F 1717The results indicate that the MOSS 100 system is substantially equivalent to the identified predicate device (Expedium Spine System, K041119).
    Static axial pull-offASTM F 1798 (Standard Test Methods for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants)The Biedermann Motech device has higher mechanical characteristics than the reference product (Synthes Matrix Spine, K092929), supporting substantial equivalence.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact number of samples (e.g., individual pedicle screw systems or components) used for each bench test. It refers to "worst case mechanical testing" for ASTM F 1717.

    Data provenance: This refers to physical bench testing conducted by the manufacturer, Biedermann Motech GmbH & Co. KG, based in Villingen-Schwenningen, Germany.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a physical device being evaluated through mechanical bench testing against established ASTM standards, not against human expert interpretation of data.

    4. Adjudication method for the test set

    Not applicable. This is a physical device being evaluated through mechanical bench testing against established ASTM standards. The results are compared to predicate devices and standards directly.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device, not an AI/ML software device that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device, not an AI/ML algorithm.

    7. The type of ground truth used

    The ground truth or reference for comparison is established by ASTM standards (ASTM F 1717 and ASTM F 1798) and the performance of identified predicate devices (Expedium Spine System, K041119, and Synthes Matrix Spine, K092929).

    8. The sample size for the training set

    Not applicable. This device does not involve a training set as it is a physical product, not an AI/ML system.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150294

    Validate with FDA (Live)

    Date Cleared
    2015-06-05

    (119 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DPZ Pedicular Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e., scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The DPZ Pedicular Fixation System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e., fracture or dislocation); spinal stenosis, deformities or curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudarthrosis and failed previous fusion.

    Device Description

    The DPZ Pedicular Fixation System is a top loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to more closely match the patient's anatomy.

    AI/ML Overview

    The provided text describes the DPZ Pedicular Fixation System, its intended use, and the non-clinical tests conducted to demonstrate its equivalence to predicate devices. However, it explicitly states: "No clinical studies were performed."

    Therefore, I cannot provide information on acceptance criteria based on human performance or studies involving human readers/experts, as no such clinical data is presented in this document.

    Here's the information that can be extracted from the provided text regarding the device's acceptance criteria and the study that proves it meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Standard)Reported Device Performance (Result Summary)
    Static compression per ASTM F1717"The results of these evaluations indicate that the DPZ Pedicular Fixation System is equivalent to predicate devices."
    Dynamic compression per ASTM F1717"The results of these evaluations indicate that the DPZ Pedicular Fixation System is equivalent to predicate devices."
    Static torsion per ASTM F1717"The results of these evaluations indicate that the DPZ Pedicular Fixation System is equivalent to predicate devices."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for the non-clinical tests. The document only mentions "The following analyses were conducted."
    • Data Provenance: Not explicitly stated, but it refers to "non-clinical tests" which typically involve laboratory or bench testing of the device hardware.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The "ground truth" for non-clinical mechanical testing is based on the specifications within the ASTM F1717 standard and the measured performance of the device compared to predicate devices. Human experts are not involved in establishing this ground truth for a mechanical test.

    4. Adjudication method for the test set:

    • Not applicable, as this refers to human reader review and consensus, which was not performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done, as "No clinical studies were performed."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a medical device (pedicle fixation system), not an algorithm or AI. The standalone performance refers to the device's mechanical integrity under specified ASTM standards.

    7. The type of ground truth used:

    • The ground truth for the non-clinical tests was based on the performance requirements and methodologies outlined in the ASTM F1717 standard for spinal implant constructs. The device's performance was then compared to that of predicate devices to establish substantial equivalence.

    8. The sample size for the training set:

    • Not applicable. This is a non-AI/algorithm medical device. "Training set" refers to data used to train an AI model.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K123134

    Validate with FDA (Live)

    Device Name
    EASYSPINE SYSTEM
    Manufacturer
    Date Cleared
    2013-02-01

    (119 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K100952, K092929

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Easyspine® Posterior Spinal System is a posterior, non-cervical pedicle and non-pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally-mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine:

    • Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
    • Spondylolisthesis .
    • Trauma (i.e., fracture or dislocation) .
    • Spinal stenosis .
    • Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
    • . Tumor
    • Pseudoarthrosis .
    • Failed previous fusion ●
    Device Description

    The Easyspine Spinal System is a side-loading posterior spinal pedicle fixation system consisting of various implant components. The 4.25 mm pedicle screws and curved rods represent a line extension to the previously cleared Easyspine Spinal System (K043094, K063794, and K082592). The subject devices are identical to their predicates with respect to general technological characteristics and intended use.

    AI/ML Overview

    The provided document describes the LDR Spine Easyspine Posterior Spinal System, a medical device for spinal fixation. However, the document does NOT contain information about acceptance criteria for device performance, nor does it detail a study proving the device meets such criteria in terms of clinical or algorithm performance.

    Instead, the document focuses on:

    • Device Description: The Easyspine Spinal System is a side-loading posterior spinal pedicle fixation system. The K123134 submission is for a line extension of 4.25 mm pedicle screws and curved rods to an existing system.
    • Indications for Use: The system is intended to provide immobilization and stabilization of spinal segments in skeletally-mature patients as an adjunct to fusion for various spinal instabilities.
    • Non-Clinical Testing: This section states that non-clinical testing was performed on the proposed 4.25 mm pedicle screws in accordance with ASTM standards (F-1717 and F-1798) for static and dynamic axial compression, static torsion, axial gripping capacity, and flexion-extension cantilever.
    • Conclusion: The non-clinical testing results demonstrate that the performance of the proposed LDR Spine Easyspine implants is "substantially equivalent" to their legally marketed predicates. This is the basis for the FDA's 510(k) clearance, which is a regulatory pathway to demonstrate that a new device is as safe and effective as a legally marketed predicate device.

    Therefore, based on the provided text, I cannot extract the requested information regarding acceptance criteria for device performance and a study proving those criteria are met for a clinical or software-based outcome. The document describes a non-clinical (benchtop) study designed to demonstrate substantial equivalence to predicate devices based on mechanical performance standards, not a study evaluating human-in-the-loop performance, standalone algorithm performance, or using human expert ground truth.

    If this were a submission for a software-as-a-medical-device (SaMD) or AI-based device, the requested information would be pertinent. However, for this spinal implant submission, the "study" is the non-clinical mechanical testing, and the "acceptance criteria" are implicitly defined by the chosen ASTM standards and the demonstration of substantial equivalence to predicate devices, rather than a quantifiable clinical performance metric.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121670

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-01-25

    (233 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medyssey Co. Ltd. Zenius, Iliad and Kora Spinal Systems are intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograff and/or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The Medyssey Co. Ltd., Cannulated Pedicle Screws are intended to be used with their respective (e.g. Zenius) top-loading posterior spinal fixation system which consists of pedicle screws, rods, set screws, and a transverse (cross) linking mechanism. The Medyssey Co. Ltd., Cannulated Pedicle Screw implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. Various sizes of these implants are available. The Medyssey Co. Ltd., Cannulated Pedicle Screws can be used in the posterior plane providing unilateral and bilateral modes of fixation. The Medyssey Co. Ltd., Cannulated Pedicle Screw Spinal design allows adjustment in both the sagittal and coronal planes permitting screw placement according to the best possible anatomic (spinal) location and orientation. This is accomplished by means of a mobile and non-mobile housing component which houses the screw and the rod which accepts a setscrew which tightens against the rod which tightens head of the pedicle screw upon secure tightening interface of the set screw assembly with the rod. The Cobalt Chrome Rods are for use with the Zenius or Iliad System. Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the Medyssey Co. Ltd. Cannulated Pedicle Screw implants.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Medyssey Cannulated Pedicle Screws and Cobalt Chrome Rods:

    1. Acceptance Criteria and Reported Device Performance

    The provided document describes non-clinical (bench) testing for the Medyssey Cannulated Pedicle Screws and Co-Cr-Mo rods. The acceptance criteria are implicitly defined by the chosen ASTM standard and the comparison to predicate devices, which means the new device must perform at least as well as the legally marketed predicates.

    Acceptance Criteria CategorySpecific Criteria (Implicitly from ASTM F1717-04 and Predicate Comparison)Reported Device Performance
    Mechanical Performance- Static Compression Strength"Analysis has shown that the proposed screws and rods are equivalent to the predicate devices." (Implies meeting or exceeding predicate performance for all tests listed below)
    - Static Torsion Strength
    - Static Tension Strength
    - Dynamic Compression Performance (Fatigue)
    Material Compatibility- Use of biocompatible materials (Titanium alloy [Ti-6Al-4V ELI], Co-Cr-Mo) conforming to ASTM F 136."fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136." and "proposed rods are manufactured from Co-Cr-Mo"
    Design Equivalence- Minor differences in design (cannulation of screws, Co-Cr-Mo rods instead of titanium) do not impact safety or effectiveness."Analysis has shown that the proposed screws and rods are equivalent to the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify the exact number of devices (screws/rods) tested for each non-clinical test (static compression, torsion, tension, dynamic compression). Standard test methodologies for medical devices like ASTM F1717-04 typically require a minimum number of samples for statistically valid results, but these specific numbers are not detailed in this summary.
    • Data Provenance: The data is from non-clinical (bench) testing performed by Medyssey Co. Ltd. The country of origin of the testing data is not explicitly stated, but Medyssey is a Korean company with a US-based contact for regulatory purposes.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the study described is non-clinical (bench testing) of physical device properties, not a clinical study involving human patients or expert interpretation of medical data. The "ground truth" here is the measured mechanical performance of the device against established engineering standards and predicate device performance.

    4. Adjudication Method for the Test Set

    This information is not applicable as the study described is non-clinical (bench testing). Adjudication methods are typically employed in clinical studies or studies involving expert review of medical images/data to resolve discrepancies in interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This document describes a 510(k) submission for a physical medical device (spinal pedicle screws and rods), not an AI/software device. Therefore, there is no mention of human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical implant, not an algorithm or software.

    7. The Type of Ground Truth Used

    The "ground truth" used for this non-clinical study is:

    • Engineering Standards and Specifications: Adherence to ASTM F1717-04.
    • Predicate Device Performance: The mechanical performance of the new device was compared to that of legally marketed predicate devices. The implicit ground truth is that the predicate devices are safe and effective, and therefore, a device demonstrating equivalent or superior mechanical properties would also be considered safe and effective for the stated indications.

    8. The Sample Size for the Training Set

    This information is not applicable as the study described is non-clinical (bench testing) of a physical device. There is no concept of a "training set" for this type of evaluation.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this non-clinical study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100634

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-01

    (118 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes USS are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).

    When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

    When used with the 3.5/6.0 mm parallel connectors, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm Systems. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5 mm Systems. When used with the 5.0/6.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS 6.0 mm rod systems.

    When used with the 3.5/6.0 mm and 4.0/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 3.5/5.5 mm and 4.0/5.5 mm tapered rods, Matrix can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 5.5/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be lined to the Matrix System.

    In addition, Synthes USS 6.0 mm rod systems can be interchanged with all USS 6.0 mm rods and transconnectors.

    Synthes USS

    • 6.0 mm Rod Systems: USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, ClampFix
    • 5.5 mm Rod System: Matrix
    • 5.0 mm Rod System: USS Small Stature

    CerviFix

    • 3.5 mm Rod Systems: CerviFix, Axon, Synapse
    • 4.0 mm Rod System: Synapse
    Device Description

    The Synthes Matrix System is an addition to Synthes' existing posterior thoracolumbar spine systems. The Matrix System consists of a family of non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2), posterior hook fixation (T1-L5) or anterolateral fixation (T8-L5). The implants include pedicle bone screws, polyaxial pedicle screws, monoaxial pedicle screws, polyaxial heads, reduction screws, reduction heads, locking caps, transconnectors, transverse bars, rods and hooks. The implants are primarily manufactured from titanium (ASTM F67 - 06), titanium alloy (ASTM F1295 - 05), cobalt-chromium-molybdenum alloy (ASTM F1537 - 08) or nitinol (ASTM F2063 - 05), similar to the predicates.

    The subject of this submission is the addition of transverse bars and tapered rods.

    AI/ML Overview

    The document describes the Synthes Matrix System, a spinal fixation device. The study provided focuses on non-clinical performance testing rather than clinical trials with human subjects or AI performance.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from ASTM F1717–09 standards)Reported Device Performance (Synthes Matrix System)
    Static compression bend testingPerforms as well as or better than predicate devices
    Static torsion testingPerforms as well as or better than predicate devices
    Dynamic compression bend testingPerforms as well as or better than predicate devices

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified. The document states "Synthes conducted the following non-clinical testing..." but does not detail the number of units tested for each type of test.
    • Data Provenance: The testing was non-clinical, conducted by Synthes. It does not involve human subjects, so concepts like country of origin or retrospective/prospective don't apply in the medical imaging sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable to the provided document. The study is a non-clinical, mechanical performance test, not a study evaluating human judgment or diagnostic accuracy. The "ground truth" here is compliance with ASTM standards, not expert consensus on medical images or patient outcomes.

    4. Adjudication method for the test set

    • Not applicable. As noted above, this is a mechanical performance test, not a study involving human readers or adjudicators of medical data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a non-clinical performance study of a spinal fixation device, not a study involving AI or human readers of medical images.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No. This is a non-clinical performance study of a spinal fixation device, not an algorithm.

    7. The type of ground truth used

    • The "ground truth" implicitly used for this non-clinical testing is the mechanical performance requirements specified in ASTM F1717–09 standards for static compression bend, static torsion, and dynamic compression bend testing. The device's performance was compared against these established industry standards and against predicate devices.

    8. The sample size for the training set

    • Not applicable. This is a non-clinical performance study of a physical device, not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1