K Number
K100952

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-08-05

(121 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Synthes USS are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).

When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

When used with the 3.5/6.0 mm parallel connectors, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm Systems. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5 mm Systems. When used with the 5.0/6.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS 6.0 mm rod systems.

When used with the 3.5/6.0 mm and 4.0/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 3.5/5.5 mm and 4.0/5.5 mm tapered rods, Matrix can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 5.5/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the Matrix System.

In addition, Synthes USS 6.0 mm rod systems can be interchanged with all USS 6.0 mm rods and transconnectors.

Device Description

The Synthes Matrix System is an addition to Synthes' existing posterior thoracolumbar spine systems. The Matrix System consist of a family non-cervical spinal fixation devices intended for posterior pedicle screw fixation (TI-S2), posterior hook fixation (TI-L5) or anterolateral fixation (T8-L5). The implants include pedicle bone screws, polyaxial pedicle screws, monoaxial pedicle screws, polyaxial heads, reduction screws, reduction heads, locking caps, transconnectors, transverse bars, rods and hooks. The implants are primarily manufactured from titanium (ASTM F67-06), titanium alloy (ASTM F1295-05), cobalt-chromium-molybdenum alloy (ASTM F1537-08) or nitinol (ASTM F2063-05), similar to the predicates.

AI/ML Overview

The Synthes Matrix System, a spinal fixation device, demonstrated substantial equivalence to its predicate devices through non-clinical performance testing.

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" with numerical targets in the conventional sense. Instead, the performance data focused on demonstrating that the Synthes Matrix System performs "as well as or better than" the predicate devices in various biomechanical tests. The implied acceptance criterion is that the device's performance in these tests is comparable to or exceeds that of the predicate devices.

Test NameStandard AccordanceReported Device Performance
Cantilever Bend in conjunction with Rod Push-throughASTM F1798 – 97Performs as well as or better than predicate devices.
Static Compression BendASTM F1717 – 09Performs as well as or better than predicate devices.
Static TorsionASTM F1717 – 09Performs as well as or better than predicate devices.
Dynamic Compression BendASTM F1717 – 09Performs as well as or better than predicate devices.
Compress and Return TestingASTM F1717 – 09Performs as well as or better than predicate devices.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes (e.g., number of implants tested) for each non-clinical test. The testing was conducted by Synthes, implying an internal or contracted laboratory setting, but the country of origin of the data is not explicitly stated. The nature of these tests (biomechanical/mechanical) means they are retrospective in the sense that they are conducted on manufactured devices in a controlled environment, not on live patients.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This question is not applicable. The study involved non-clinical biomechanical testing of medical devices, not evaluations by human experts to establish "ground truth" for clinical conditions or diagnoses. The ground truth for these tests is defined by the objective measurements obtained from the testing equipment according to the specified ASTM standards.

4. Adjudication Method for the Test Set:

This question is not applicable. Adjudication methods like "2+1" or "3+1" are relevant for expert consensus in clinical or imaging studies, not for non-clinical mechanical testing where performance is determined by objective measurements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. The submission explicitly states, "Clinical data was not needed for this device." The evaluation was based solely on non-clinical performance data.

6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance):

This question is not applicable. This is a physical medical device (spinal fixation system), not a software algorithm or AI-powered system where "standalone" performance would be a relevant metric.

7. Type of Ground Truth Used:

The "ground truth" for the non-clinical tests was established by objective physical measurements obtained from testing the devices according to specific ASTM standards (ASTM F1798-97 and ASTM F1717-09). These standards define the methodology for evaluating the mechanical properties of spinal implants.

8. Sample Size for the Training Set:

This question is not applicable. There is no information about a "training set" as this is a physical medical device, not a machine learning model or algorithm. The non-clinical tests evaluate the physical properties and performance of the manufactured device.

9. How the Ground Truth for the Training Set Was Established:

This question is not applicable. As stated above, there is no "training set" in the context of this device.

{0}------------------------------------------------

Page 1/2

AUG 0 5 2010

12 510(k) Summary

Name of Firm:Synthes Spine1302 Wrights Lane EastWest Chester, PA 19380
510(k) Contact:Jason LipmanSenior Regulatory Affairs SpecialistTelephone: 610-719-5629 Facsimile: 610-719-5102Email: Lipman.jason@synthes.com
Date Prepared:April 6, 2010
Trade Name:Synthes Matrix System
Classification:21 CFR 888.3070 –Pedicle screw spinal systemClass IIIOrthopaedic and Rehabilitation Devices PanelProduct Code: NKB, MNH, MNI, KWQ, KWP
Predicate Devices:Synthes Matrix System, K092929, K100634Synthes USS, K963045, K082572Synthes Click'X, K992739Synthes Pangea, K052123Synthes USS VAS, K002517
Device Description:The Synthes Matrix System is an addition to Synthes' existing posteriorthoracolumbar spine systems. The Matrix System consist of a familynon-cervical spinal fixation devices intended for posterior pedicle screwfixation (TI-S2), posterior hook fixation (TI-L5) or anterolateralfixation (T8-L5). The implants include pedicle bone screws, polyaxialpedicle screws, monoaxial pedicle screws, polyaxial heads, reductionscrews, reduction heads, locking caps, transconnectors, transverse bars,rods and hooks. The implants are primarily manufactured from titanium(ASTM F67-06), titanium alloy (ASTM F1295-05), cobalt-chromium-molybdenum alloy (ASTM F1537-08) or nitinol (ASTMF2063-05), similar to the predicates.
Intended Use /Indications for Use:The Synthes USS are non-cervical spinal fixation devices intended forposterior pedicle screw fixation (TI-S2/ilium), posterior hook fixation (TI-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is limited toskeletally mature patients with the exception of the Small Stature USS,which includes small stature and pediatric patients. These devices areindicated as an adjunct to fusion for all of the following indications:degenerative disc disease (defined as discogenic back pain withdegeneration of the disc confirmed by history and radiographic studies),spondylolisthesis, trauma (i.e., fracture or dislocation), deformities orcurvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann'sDisease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).When treating patients with Degenerative Disc Disease (DDD), transversebars are not cleared for use as part of the posterior pedicle screw construct.When used with the 3.5/6.0 mm parallel connectors, the Synthes USS 6.0mm rod systems can be linked to the CerviFix 3.5 mm Systems. Inwhen used with 3.5/5.0 mm parallel connectors, the Synthes Small
Stature USS can be linked to the CerviFix 3.5 mm Systems. When usedwith the 5.0/6.0 mm parallel connectors, the Synthes Small Stature USS canbe linked to the Synthes USS 6.0 mm rod systems.
When used with the 3.5/6.0 mm and 4.0/6.0 mm tapered rods, the SynthesUSS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mmSystems, respectively. When used with the 3.5/5.5 mm and 4.0/5.5 mmtapered rods, Matrix can be linked to the CerviFix 3.5 mm and 4.0 mmSystems, respectively. When used with the 5.5/6.0 mm tapered rods, theSynthes USS 6.0 mm rod systems can be linked to the Matrix System.
In addition, Synthes USS 6.0 mm rod systems can be interchanged with allUSS 6.0 mm rods and transconnectors.
Synthes USS• 6.0 mm Rod Systems: USS Side-Opening, USS Dual-Opening, USSVAS variable axis components, USS Fracture, Click'X, Click'XMonoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Iliosacral,ClampFix• 5.5 mm Rod System: Matrix• 5.0 mm Rod System: USS Small Stature
CerviFix• 3.5 mm Rod Systems: CerviFix, Axon, Synapse• 4.0 mm Rod System: Synapse
Comparison of thetechnologicalcharacteristics of thedevice to thepredicate device:The design features, material, and indications for use of thesubject Matrix System is substantially equivalent to the predicatedevices identified. Additionally, the safety and effectiveness of thissystem is adequately supported by documentation within this premarketnotification.
Performance Data(Non-clinical and/orClinical)Synthes conducted the following non-clinical testing: cantilever bend inconjunction with rod push-through in accordance with ASTM F1798 –97 and static compression bend, static torsion, dynamic compressionbend, and compress and return testing in accordance with ASTM F1717– 09. The conclusions drawn from testing demonstrate that the MatrixSystem is as safe and effective and performs as well as or better than thepredicate devices identified. Clinical data was not needed for this device.

.

।.

{1}------------------------------------------------

:

.

. . . .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that suggest the shape of the bird's body and wings. The text is in a simple, sans-serif font and is evenly spaced around the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 0 5 2010

Synthes Spine % Mr. Jason Lipman Senior Regulatory Affairs Specialist 1302 Wrights Lane East West Chester, Pennsylvania 19380

Re: K100952

Trade/Device Name: Synthes Matrix System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, KWQ, KWP Dated: July 09, 2010 Received: July 12, 2010

Dear Mr. Lipman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. Jason Lipman

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) { 1 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the · Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Sincerely yours,

Mark N Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

11 Indications for Use Statement

510(k) Number: K100952

Device Name: Synthes Matrix System

The Synthes USS are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scollosis, kyphosis, and/or lordosis, Schevermann's Disease), turnor, stenosis, and failed previous fusion (pseudoarthrosis).

When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

When used with the 3.5/6.0 mm parallel connectors, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm Systems. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5 mm Systems. When used with the 5.0/6.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS 6.0 mm rod systems.

When used with the 3.5/6.0 mm and 4.0/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively, When used with the 3.5/5.5 mm and 4.0/5.5 mm tapered rods, Matrix can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 5.5/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the Matrix System,

In addition, Synthes USS 6.0 mm rod systems can be interchanged with all USS 6.0 mm rods and transconnectors.

Synthes USS

  • 6.0 mm Rod Systems: USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS o Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, ClampFix
  • 5.5 mm Rod System: Matrix . O
  • 5.0 mm Rod System: USS Small Stature O

CerviFix

  • 3.5 mm Rod Systems: CerviFix, Axon, Synapse ಂ
  • 4.0 mm Rod System: Synapse o

Prescription Use X (21 CFR 801 Subpart D)

AND / OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
----------------------------------------------------------

Division Sign-Division of Surgical, Orthopedic, and Restorative Devices

K100952 Synthes Matrix System Special Special Sportiber