K Number
K063794

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-01-24

(33 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LDR Easyspine System is a posterior, noncervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Device Description

Modifications to the Easyspine System that are the submission are confined solely to the addition of 5 mm diameter screws, 8 mm diameter screws, and washers (also referred to as spacers.) No components of the predicate (unmodified) system have been modified or deleted.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called the Easyspine® System, specifically for modifications including the addition of 5mm and 8mm diameter screws and washers. This document details the regulatory nature of the device and its equivalence to previously cleared predicates, but it does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the context of performance metrics like sensitivity, specificity, or reader improvement.

The 510(k) process for devices like the Easyspine System (a pedicle screw spinal system) focuses on demonstrating "substantial equivalence" to a legally marketed predicate device. This typically involves showing that the device has the same intended use, technological characteristics, and similar or identical materials and operating principles, and that any differences do not raise new questions of safety or effectiveness. Performance for such devices is usually assessed through mechanical testing (e.g., fatigue strength, torsional strength) rather than clinical studies that would generate the kind of performance metrics you've asked for.

Therefore, I cannot provide the requested information for the following reasons:

  • No Acceptance Criteria/Performance Data: The document does not specify quantitative acceptance criteria for device performance (e.g., in terms of clinical outcomes, diagnostic accuracy, or reader performance) nor does it report the device's performance against such criteria. The "performance" discussed pertains to material properties and mechanical integrity rather than analytical or clinical efficacy metrics.
  • No Clinical Study for AI/Diagnostic Performance: This device is a surgical implant, not a diagnostic AI tool or a device that would typically undergo studies to assess human reader improvement or standalone diagnostic performance. The 510(k) largely focuses on mechanical and material equivalence.

Given this, I will answer each numbered point based on the information available in the document, noting when the information is not present.


Here's the breakdown based on the provided text:

  1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy). The acceptance criteria for this 510(k) appear to be demonstrating "substantial equivalence" to predicate devices, including having the same intended use, operating principle, basic design, materials, and packaging/sterilization processes.
    • Reported Device Performance:
      • Material Composition: Made of titanium alloy conforming to ASTM F 136, "Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications."
      • Biocompatibility: Titanium material is described as "biocompatible, corrosion-resistant, and not toxic in a biologic environment."
      • Imaging Compatibility: Allows for "artifact-free x-ray imaging, computed tomography (CT) and Magnetic Resonance Imaging (MRI)."
      • Equivalence to Marketed Product: The modified system shares the same intended use, operating principle, basic design, materials, and packaging/sterilization processes as the predicate Easyspine System.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • This information is not provided in the document. The filing is for a modified spinal implant, and the assessment of "performance" typically involves bench testing (mechanical and material analyses) rather than clinical data sets in the diagnostic sense.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not applicable/provided. The device is a surgical implant, not one that requires establishing ground truth by expert interpretation of images or clinical data for performance assessment in the manner of AI/diagnostic devices.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable/provided for the same reasons as point 3.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This information is not applicable/provided. This device is a spinal implant, not an AI-assisted diagnostic tool.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • This information is not applicable/provided. This device is a spinal implant, not a standalone algorithm.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For this type of device (spinal implant), "ground truth" would typically relate to the material specifications and mechanical testing standards (e.g., ASTM standards for titanium alloy, validated mechanical testing protocols to ensure strength and durability). The document indicates compliance with ASTM F 136 for material composition. No clinical "ground truth" in terms of patient outcomes or pathology is referenced in this summary.
  8. The sample size for the training set

    • This information is not applicable/provided. This device is a spinal implant being evaluated for substantial equivalence, not an AI model requiring a training set.
  9. How the ground truth for the training set was established

    • This information is not applicable/provided for the same reasons as point 8.

{0}------------------------------------------------

510(k) Summary

Modifications to Easyspine® System

K063794
Page 1 of 2

Special 510(k): Device Modification 510(k) Summary Easyspine® System

Sponsor Name:LDR Spine USA4030 West Braker Lane, Suite 360Austin, Texas 78759Telephone (512) 344-3333Fax (512) 344-3350JAN 2 4 2007
Official Contact:James Burrows
Representative/Consultant:Floyd G. LarsonPaxMed International, LLC11234 El Camino Real, Suite. 200San Diego, CA 92130Telephone (858) 792-1235Fax (858) 792-1236
DEVICE NAME
Classification Name:Orthosis, Spinal Pedicle Fixation (MNI)Orthosis, Spondylolisthesis Spinal Fixation (MNH)
Trade/Proprietary Name:Easyspine® System
Common Name:Pedicle Screw Spinal System

ESTABLISHMENT REGISTRATION NUMBER

LDR Spine USA has submitted an Establishment Registration to FDA. The Establishment Registration number is 3004903783. The Owner/Operator number is 9068076.

DEVICE CLASSIFICATION

The Easyspine system is a Class II pedicle screw spinal system (21 CFR 888.3070). The product code for Orthosis, Spondylolistines is Spinal Eixation is MNH. The product code for Orthosis, Spinal Pedicle Fixation is MN. These device opinal intellent is MNT. The product code in Crimsels, Species, Species, Branch.

PREDICATE DEVICE INFORMATION

The predicate device for this modification is the Easyspine System cleared by FDA under
K043004 and K064017 K043094 and K061017.

{1}------------------------------------------------

LABELING

Package labels are similar in format and content to previously cleared labels for the unmodified Easyspine System. The system modifications described in this submission do not necessitate changes to the Package Insert.

PACKAGING AND STERILIZATION

Packaging and sterilization will not change with the modifications described in this submission.

INTENDED USE

The Easyspine System is a posterior, noncervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

DEVICE DESCRIPTION

Modifications to the Easyspine System that are the submission are confined solely to the addition of 5 mm diameter screws, 8 mm diameter screws, and washers (also referred to as spacers.) No components of the predicate (unmodified) system have been modified or deleted.

Material Composition and Biocompatibility

The rods, pedicle screws and components, set screws, connectors and washers of the Easyspine System are made of titanium alloy conforming to the requirements of ASTM F 136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. The titanium material is biocompatible, corrosion-resistant, and not toxic in a biologic environment allowing for artifact-free x-ray imaging, computed tomography (CT) and Magnetic Resonance Imaging (MRI).

EQUIVALENCE TO MARKETED PRODUCT

The modified Easyspine System has the following similarities to the predicate Easyspine System:

  • I has the same intended use.
  • . uses the same operating principle.
  • incorporates the same basic design.
  • 트 incorporates the same materials, and
  • . is packaged and sterilized using the same materials and processes.

In summary, the modified Easyspine System described in this submission is, in our opinion, substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, clutching a caduceus in its talons. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

LDR Spine USA % Mr. Floyd G. Larson President PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, California 92130

JAN 2 4 2007

Re: K063794

Trade/Device Name: Easyspine® System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH Dated: December 21, 2006 Received: December 22, 2007

Dear Mr. Larson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, EDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Floyd G. Larson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours,

for
N. M. M.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

Applicant: LDR Spine USA

510(k) Number (if known):

Device Name: Easyspine System

Indications for Use:

The LDR Easyspine System is a posterior, noncervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Pep

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K063794