(101 days)
The MOSS 100 Pedicle Screw System is a thoracolumbar pedicle for the following indications of use:
- Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis
- Trauma (i.e. fractures or dislocations)
- Spinal Stenosis
- Curvatures (i.e. scoliosis, kyphoses and/ or lordosis)
- Tumor
- Pseudoarthrosis
- Failed previous fusion
The Biedermann Motech MOSS 100 Pedicle Screw System is a comprehensive thoracolumbar spinal system that offers posterior clinical solutions. The MOSS 100 System is a universal set of instruments and implants that are indicated for the treatment of significant mechanical instability or deformity of the spine which requires fusion with instrumentation.
The MOSS 100 System provides several options for stabilization of the spine. Therefore, the system includes the following implants:
- -Pedicle Screws Ø5-8mm, length 25-80mm
- -Rods Ø5.5mm, length 30-95mm, 300mm and 480mm
- Locking caps -
- Polyaxial Heads -
The Polyaxial Heads are mounted onto the Pedicle Screws intra-operatively. The implants of the MOSS 100 System are single-use only and the system is provided non sterile.
This document is a 510(k) Premarket Notification for a medical device called the "MOSS 100 Pedicle Screw System." It does not describe a study involving an AI/Machine Learning device or analytical performance metrics typically associated with such software. Instead, it focuses on the performance of a physical pedicle screw system through non-clinical bench testing.
Therefore, many of the requested elements for AI/ML device studies are not applicable to this document. I will extract the available information regarding acceptance criteria and performance for the pedicle screw system.
Here's the information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Type | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Static Compression | ASTM F 1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model) | The results indicate that the MOSS 100 system is substantially equivalent to the identified predicate device (Expedium Spine System, K041119). |
| Dynamic Compression | ASTM F 1717 | The results indicate that the MOSS 100 system is substantially equivalent to the identified predicate device (Expedium Spine System, K041119). |
| Static Torsion | ASTM F 1717 | The results indicate that the MOSS 100 system is substantially equivalent to the identified predicate device (Expedium Spine System, K041119). |
| Static axial pull-off | ASTM F 1798 (Standard Test Methods for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants) | The Biedermann Motech device has higher mechanical characteristics than the reference product (Synthes Matrix Spine, K092929), supporting substantial equivalence. |
2. Sample size used for the test set and the data provenance
The document does not specify the exact number of samples (e.g., individual pedicle screw systems or components) used for each bench test. It refers to "worst case mechanical testing" for ASTM F 1717.
Data provenance: This refers to physical bench testing conducted by the manufacturer, Biedermann Motech GmbH & Co. KG, based in Villingen-Schwenningen, Germany.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a physical device being evaluated through mechanical bench testing against established ASTM standards, not against human expert interpretation of data.
4. Adjudication method for the test set
Not applicable. This is a physical device being evaluated through mechanical bench testing against established ASTM standards. The results are compared to predicate devices and standards directly.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a physical medical device, not an AI/ML software device that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a physical medical device, not an AI/ML algorithm.
7. The type of ground truth used
The ground truth or reference for comparison is established by ASTM standards (ASTM F 1717 and ASTM F 1798) and the performance of identified predicate devices (Expedium Spine System, K041119, and Synthes Matrix Spine, K092929).
8. The sample size for the training set
Not applicable. This device does not involve a training set as it is a physical product, not an AI/ML system.
9. How the ground truth for the training set was established
Not applicable. There is no training set for a physical device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2016
Biedermann Motech GmbH & Co. KG Mr. Gerd Federle Director of Quality Management Bertha-von-Suttner-Str. 23 78054 Villingen-Schwenningen GERMANY
Re: K162232
Trade/Device Name: MOSS 100 Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: October 21, 2016 Received: October 24, 2016
Dear Mr. Federle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K162232
Device Name MOSS 100 Pedicle Screw System
Indications for Use (Describe)
The MOSS 100 Pedicle Screw System is a thoracolumbar pedicle for the following indications of use:
-
Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
-
Spondylolisthesis
-
Trauma (i.e. fractures or dislocations)
-
Spinal Stenosis
-
Curvatures (i.e. scoliosis, kyphoses and/ or lordosis)
-
Tumor
-
Pseudoarthrosis
-
Failed previous fusion
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995,
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the text "K162232 - Page 1 of 4" in the upper left corner. To the right of this text is a logo that includes the text "BIEDERMANN" above the text "MOTECH". A black square is located in the upper right corner of the logo. The logo is enclosed in a thin-lined square.
510(k) Summary
1. Submitter, Correspondent and Manufacturer
| Submitter Name: | Biedermann Motech GmbH & Co. KG |
|---|---|
| Submitter Address: | Bertha-von-Suttner-Str. 2378054 Villingen-SchwenningenGermany |
| Contact Person: | Gerd FederleDirector of Quality ManagementPhone: +49 7720 8510-545Fax: +49 7720 8510-66afe@biedermann.com |
| Date Prepared: | 10/21/2016 |
| Manufacturer Name: | Biedermann Motech GmbH & Co. KG |
| Manufacturer Address: | Bertha-von-Suttner-Str. 2378054 Villingen-SchwenningenGermany |
| 2. Device Identification | |
| Device Trade Name: | MOSS 100 Pedicle Screw System |
| Device Common Name: | Pedicle Screw System |
| Classification Name: | Pedicle Screw Spinal fixation |
| Classification Code: | NKB, MNI and MNH (Class III) |
| Classification Panel: | Orthopedic |
| Regulation Number: | 21 CFR section 888.3070 |
| Regulation Name: | Pedicle screw spinal system |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a logo with the text "BIEDERMANN" on the top line and "MOTECN" on the bottom line. The text is in a simple, sans-serif font. To the right of the text is a solid black square, which is partially overlapping the text. The text and the square are enclosed in a thin, black border.
3. Primary Predicate Device
Expedium Spine System, DePuy Spine, Inc.
4. Reference Device
Synthes Matrix Spine, Synthes (USA)
5. Device Description
The Biedermann Motech MOSS 100 Pedicle Screw System is a comprehensive thoracolumbar spinal system that offers posterior clinical solutions. The MOSS 100 System is a universal set of instruments and implants that are indicated for the treatment of significant mechanical instability or deformity of the spine which requires fusion with instrumentation.
The MOSS 100 System provides several options for stabilization of the spine. Therefore, the system includes the following implants:
- -Pedicle Screws Ø5-8mm, length 25-80mm
- -Rods Ø5.5mm, length 30-95mm, 300mm and 480mm
- Locking caps -
- Polyaxial Heads -
The Polyaxial Heads are mounted onto the Pedicle Screws intra-operatively. The implants of the MOSS 100 System are single-use only and the system is provided non sterile.
6. Material
The implants are manufactured from ASTM F 136 implant grade titanium alloy Ti-6Al-4V ELI.
7. Indications for Use
The MOSS 100 Pedicle Screw System is a thoracolumbar pedicle fixation device for the following indications of use:
- Deqenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- -Spondylolisthesis
- Trauma (i.e. fractures or dislocations) -
- -Spinal Stenosis
- Curvatures (i.e. scoliosis, kyphosis and/ or lordosis) -
- -Tumor
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the text "K162232 - Page 3 of 4" on the top left. To the right of this text is a square with a smaller black square inside. Below the square are two lines of text, "BIEDERMANN" on the first line and "MOTECN" on the second line.
- Pseudoarthrosis -
- Failed previous fusion -
8. Technological Characteristics and Substantial Equivalence
Like the predicate devices (Expedium Spine System – K041119, Synthes Matrix Spine - K092929) the MOSS 100 Pedicle Screw Systems is a multiple component device, made from Titanium alloy Ti-6Al-4V that allows the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. The MOSS 100 spinal implant assembly consists of pedicle screws, heads (interconnection mechanisms), locking caps and rods of a similar design as the predicate devices components.
The System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients, as the predicate is.
9. Performance Data
9.1 Non-Clinical Test Summary
The MOSS 100 system was evaluated per the FDA Guidance Document for Industry and FDA Staff; Spinal System 510(k)s.
The following testing was done to show safety and effectiveness of our system:
Bench testing - mechanical
Mechanical testing according to ASTM F 1717 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model standard was performed in order to provide data to support a substantial equivalence determination. These tests were performed to support system related performance under consideration of well-established acceptance criteria. Performance testing included:
- Static Compression
- . Dynamic Compression
- . Static Torsion
Result: The results of the worst case mechanical testing of the MOSS 100 System indicate that the MOSS 100 system is substantially equivalent to the identified predicate device.
Additional Bench testing
Mechanical testing according to ASTM F 1798 (Standard Test Methods for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants) standard was performed in order to provide data to support a substantial equivalence determination. These
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the text "K162232 - Page 4 of 4" on the top left. On the right side of the image, there is a logo with a black square on top of the text "BIEDERMANN" and "MOTECN". The logo is enclosed in a square.
tests are performed to support system related performance under consideration of well-established acceptance criteria. Performance testing included:
- " Static axial pull-off test
Results: Review and comparison of the data shows the subjects Biedermann Motech device has higher mechanical characteristics than the reference product. The test data supports substantial equivalence of performance of the Biedermann Motech subject device and reference product.
9.2 Clinical Test Summary
No clinical tests were performed.
N/A