Search Results
Found 3 results
510(k) Data Aggregation
(105 days)
The M.U.S.T. MINI Posterior Cervical Screw System is intended to provide immobilization of spinal segments as an adjunct to fusion, in skeletally mature patient, for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine from T1-T3: traumatic spinal fraumatic dislocations; instability or deformity; failed previous (e.g., pseudarthrosis); tumors involving the cervical spine; degenerative disease, including intractable radior myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.
The M.U.S.T. MINI Posterior Cervical Screw System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the M.U.S.T. MINI Posterior Cervical Screw System may be connected to the M.U.S.T. System rods with the M.U.S.T. MINI rod connectors. Transition rods with differing diameters may also be used to connect the M.U.S.T. MINI Posterior Cervical Screw System to the M.U.S.T. System. Refer to the M.U.S.T. System package insert for a list of the M.U.S.T. Indications of Use.
When used with the Occipital Plate, the M.U.S.T MINI Posterior Cervical Screw System is also intended to provide immobilization and stabilization for the occipito-cervico-thoracic junction (occiput - T3) in treatment of the instabilities mentioned above, including occipitocervical dislocation.
The subject M.U.S.T. MINI Extension implants are line extension to the previously cleared Medacta M.U.S.T. MINI Posterior Cervical Screws System (K171369).
The new subject Posterior Occipital-Cervical Screw System is a multi-component device, manufactured from Titanium-based and CoCr materials, consisting of occipital screws and straight and pre-bent rods that longitudinally connect the occiput with the posterior cervical spine. The system is intended to stabilize and fuse the spine in degenerative disc disease, spinal fusion, cervical fractures and in surgically repaired spinal pseudoarthrosis. In addition, it is used in deformity correction e.g. scoliosis to correct and stabilize the spine.
The M.U.S.T. MINI Extension implants have been designed with the same or similar shape, dimensions and materials as the previously cleared Medacta M.U.S.T. MINI Posterior Cervical Screws System (K171369), Synthes Synapse OCT System (K070573 and K142838), DePuy Mountaineer OCT Spinal System (K042508 and K110353) and Alphatec Solanas Avalon Posterior Fixation System (K111076).
The M.U.S.T. MINI Extension implants are manufactured with the same materials of the Medacta predicate device M.U.S.T. MINI Posterior Cervical Screws System (K171369): Ti-6Al-4V ELI (ISO 5832-3 Implants for surgery -- Metallic materials -- Part 3: Wrought titanium 6-aluminium 4-vanadium allov + ASTM F136 Standard Specification for Wrought Titanium-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)) and CoCtMo (ISO 5832-12 Implants for surgery -- Metallic materials -- Part 12: Wrought cobalt-chromium-molybdenum alloy + ASTM F1537 Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539)).
Additionally, the new 4x and 6x package for the already cleared (K171369) M.U.S.T. MINI set screw implant have been introduced.
This document describes the M.U.S.T. MINI Extension, a spinal fixation system, and its substantial equivalence to predicate devices, rather than a diagnostic AI device. Therefore, the requested information about acceptance criteria, study details, expert involvement, and AI-specific performance metrics is not applicable in the context of this 510(k) submission.
The provided text does not contain information about an AI device or a study assessing its performance against acceptance criteria using a test set, ground truth, or human readers.
Instead, the document focuses on demonstrating that the M.U.S.T. MINI Extension is substantially equivalent to existing, legally marketed spinal fixation devices through a comparison of:
- Intended use: The conditions for which the device is designed to be used.
- Design and technological characteristics: Materials, dimensions, and overall structure.
- Performance evaluations: Mechanical tests to ensure the device meets safety and efficacy standards for spinal implants.
Here's a breakdown of the relevant information provided in the document:
1. A table of acceptance criteria and the reported device performance:
The document mentions that "Testing was conducted according to written protocols with acceptance criteria that were based on standards." However, it does not provide a specific table of acceptance criteria and reported numerical performance results for the new device compared to those criteria. It lists the types of mechanical tests performed, which inherently have acceptance criteria defined by the cited ASTM standards, but the specific Pass/Fail outcomes or quantitative measurements are not detailed in this summary.
Examples of tests performed (implying acceptance criteria based on standards):
- Static Compression Bending Test (ASTM F2706-08 (Reapproved 2014))
- Static Torsion Test (ASTM F2706-08 (Reapproved 2014))
- Dynamic Axial Compression Test (ASTM F2706-08 (Reapproved 2014))
- Dynamic Torsion Test (ASTM F2706-08 (Reapproved 2014))
- Axial Gripping Test (ASTM F1798-13)
- Plate Torque to Failure Test
- Screw Torque to Failure Test
- Pyrogenicity (Bacterial Endotoxin Test (LAL test) according to European Pharmacopoeia §2.6.14/USP chapter , and pyrogen test according to USP chapter )
2. Sample size used for the test set and the data provenance:
- Sample size: Not explicitly stated for each mechanical test. Mechanical tests typically use a specific number of samples of the device components.
- Data provenance: Not applicable in the context of a clinical patient dataset. The "data" here refers to the results of mechanical testing on the device itself (implants, rods, screws). These tests are performed in a lab setting, not on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This is not applicable as there is no "ground truth" related to expert assessment of AI output. The "truth" for this device's performance is determined by adherence to engineering standards and mechanical properties.
- There was a "Design Validation Workshop" mentioned, which implies expert review during the design process, but not for establishing ground truth in a diagnostic context.
4. Adjudication method for the test set:
- Not applicable, as there are no expert adjudications in the context of mechanical testing of a spinal implant.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This type of study is relevant for diagnostic imaging AI, where human readers evaluate cases with and without AI assistance. This document is for a physical surgical implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a physical implant, not a standalone algorithm.
7. The type of ground truth used:
- For the mechanical tests, the "ground truth" is defined by the objective physical and mechanical properties required by the cited ASTM standards. It's not expert consensus, pathology, or outcomes data in the typical sense of AI/diagnostic studies, but rather engineering specifications.
8. The sample size for the training set:
- Not applicable. This is not an AI device that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable, as there is no training set for this device.
In summary: The provided document is a 510(k) summary for a physical medical device (spinal implant) and demonstrates its substantial equivalence primarily through comparisons with predicate devices and mechanical performance testing against established engineering standards. It does not involve AI, diagnostic performance, patient data, or expert interpretations in the way the requested questions imply.
Ask a specific question about this device
(157 days)
The BRIDALVEIL Occipital Cervical Thoracic System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (Cl to C7) and the thoracic spine from TI-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions ( e.g., pseudarthrosis ); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The BRIDALVEIL Occipital Cervical Thoracic System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the BRIDALVEIL Occipital Thoracic System may be connected to the OLYMPIC Posterior Spinal Fixation System rods and connectors. Transition rods with differing diameters may also be used to connect the BRIDALVEIL Occipital Cervical Thoracic System to the OLYMPIC Posterior Spinal Fixation System. Refer to the OLYMPIC Posterior Spinal Fixation System package insert for instructions for use and indications for use.
The BRIDALVEIL Occipital Cervical Thoracic System is a spinal fixation system intended to stabilize the uppermost portion of the spine during the fusion process. The system contains a wide variety of implants and instruments which allows for the transition across multiple spinal segments: Occipital Plate with Screws, Cervical Polyaxial Screws, Laminar Hooks, Cross Connectors, Rod Connectors, and Rods manufactured from Ti6Al4V ELI (ASTM F136) and cobalt chrome alloy (ASTM F1537).
The provided text describes the regulatory clearance for the BRIDALVEIL Occipital Cervical Thoracic System, a spinal fixation system. It does not detail acceptance criteria and a study that proves the device meets those criteria in the context of an AI/ML device.
This document is a 510(k) premarket notification for a traditional medical device (spinal fixation system), not an AI/ML device. Therefore, the details requested in the prompt, such as reported device performance, sample size for test sets, expert qualifications for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment for the training set, are not applicable in the context of this submission.
Instead, the document focuses on demonstrating substantial equivalence to existing predicate devices based on:
- Intended Use: Providing immobilization and stabilization of spinal segments as an adjunct to fusion for various conditions (traumatic fractures/dislocations, instability, tumors, degenerative disease, etc.) in the craniocervical junction, cervical spine (C1-C7), and thoracic spine (T1-T3). It also mentions restoring spinal column integrity for a limited time in advanced stage tumor patients.
- Design: The system includes occipital plates with screws, cervical polyaxial screws, laminar hooks, cross connectors, rod connectors, and rods.
- Materials: Ti6Al4V ELI (ASTM F136), cobalt chrome alloy (ASTM F1537), Elgiloy CoCrNi alloy (ASTM F1058), and Nitinol #1 (ASTM E2063).
- Mechanical Safety and Performance: Evaluated through non-clinical bench testing.
Here's the information extracted from the document, framed in the context of a traditional medical device submission, rather than an AI/ML one:
1. A table of acceptance criteria and the reported device performance
For this traditional medical device (BRIDALVEIL Occipital Cervical Thoracic System), "acceptance criteria" revolve around demonstrating substantial equivalence to predicate devices through conformity to recognized standards for mechanical performance and material properties. "Reported device performance" is derived from the results of these non-clinical tests meeting the requirements of those standards.
Acceptance Criterion (Demonstrating Substantial Equivalence) | Reported Device Performance (Summary from Non-Clinical Tests) |
---|---|
Mechanical Performance: | The BRIDALVEIL Occipital Cervical Thoracic System demonstrated mechanical performance equivalent to the predicate devices across various tests. |
Static Compression Bending (ASTM F1717) | Results indicated equivalence to predicate devices. |
Dynamic Compression Bending (ASTM F1717) | Results indicated equivalence to predicate devices. |
Static Torsion (ASTM F1717) | Results indicated equivalence to predicate devices. |
Static Compression Bending (ASTM F2706) | Results indicated equivalence to predicate devices. |
Dynamic Compression Bending (ASTM F2706) | Results indicated equivalence to predicate devices. |
Static Torsion (ASTM F2706) | Results indicated equivalence to predicate devices. |
Dynamic Torsion (ASTM F2706) | Results indicated equivalence to predicate devices. |
Static Axial Grip (ASTM F1798) | Results indicated equivalence to predicate devices. |
Static Torsional Grip (ASTM F1798) | Results indicated equivalence to predicate devices. |
Static Transverse Moment (ASTM F1798) | Results indicated equivalence to predicate devices. |
Material Biocompatibility: | Materials used (Ti6Al4V ELI, CoCrMo alloy, Elgiloy CoCrNi alloy, Nitinol #1) are commonly accepted in spinal implants and are equivalent to those used in predicate devices, implying established biocompatibility. (No specific biocompatibility tests are listed, but material equivalence is claimed.) |
Sterility: | (Not explicitly detailed in the summary, but implicit for an implantable device that it would meet relevant sterility standards.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not applicable in the context of human data. The "test set" here refers to physical specimens of the device components tested in a laboratory setting. No human patients or retrospective/prospective data were used for performance evaluation.
- Data Provenance: The data provenance is from non-clinical bench testing conducted in a laboratory. The specific country of origin of the lab is not stated in the provided text.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. "Ground truth" in this context refers to the defined parameters and performance standards derived from the ASTM standards for mechanical testing. These standards are established by expert committees in engineering and materials science, but individual experts are not "adjudicating" a test set as they would for clinical images or diagnoses.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. There's no human adjudication process described for the mechanical test results; rather, the results are compared against predefined criteria within the ASTM standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI/ML device, and no MRMC studies or human reader performance evaluations were conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an AI/ML device. The "standalone performance" is the mechanical testing of the device itself against engineering standards.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For this device, the "ground truth" (or reference for evaluation) is established through recognized industry standards for mechanical testing of spinal implants (e.g., ASTM F1717, ASTM F2706, ASTM F1798). The device's performance is compared against the requirements and typical performance of predicate devices as measured under these standardized conditions.
8. The sample size for the training set
- Not applicable. This is not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. This is not an AI/ML device.
Ask a specific question about this device
(118 days)
The Synthes USS are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).
When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.
When used with the 3.5/6.0 mm parallel connectors, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm Systems. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5 mm Systems. When used with the 5.0/6.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS 6.0 mm rod systems.
When used with the 3.5/6.0 mm and 4.0/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 3.5/5.5 mm and 4.0/5.5 mm tapered rods, Matrix can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 5.5/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be lined to the Matrix System.
In addition, Synthes USS 6.0 mm rod systems can be interchanged with all USS 6.0 mm rods and transconnectors.
Synthes USS
- 6.0 mm Rod Systems: USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, ClampFix
- 5.5 mm Rod System: Matrix
- 5.0 mm Rod System: USS Small Stature
CerviFix
- 3.5 mm Rod Systems: CerviFix, Axon, Synapse
- 4.0 mm Rod System: Synapse
The Synthes Matrix System is an addition to Synthes' existing posterior thoracolumbar spine systems. The Matrix System consists of a family of non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2), posterior hook fixation (T1-L5) or anterolateral fixation (T8-L5). The implants include pedicle bone screws, polyaxial pedicle screws, monoaxial pedicle screws, polyaxial heads, reduction screws, reduction heads, locking caps, transconnectors, transverse bars, rods and hooks. The implants are primarily manufactured from titanium (ASTM F67 - 06), titanium alloy (ASTM F1295 - 05), cobalt-chromium-molybdenum alloy (ASTM F1537 - 08) or nitinol (ASTM F2063 - 05), similar to the predicates.
The subject of this submission is the addition of transverse bars and tapered rods.
The document describes the Synthes Matrix System, a spinal fixation device. The study provided focuses on non-clinical performance testing rather than clinical trials with human subjects or AI performance.
Here's an analysis of the provided text in relation to your request:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (from ASTM F1717–09 standards) | Reported Device Performance (Synthes Matrix System) |
---|---|
Static compression bend testing | Performs as well as or better than predicate devices |
Static torsion testing | Performs as well as or better than predicate devices |
Dynamic compression bend testing | Performs as well as or better than predicate devices |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified. The document states "Synthes conducted the following non-clinical testing..." but does not detail the number of units tested for each type of test.
- Data Provenance: The testing was non-clinical, conducted by Synthes. It does not involve human subjects, so concepts like country of origin or retrospective/prospective don't apply in the medical imaging sense.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This question is not applicable to the provided document. The study is a non-clinical, mechanical performance test, not a study evaluating human judgment or diagnostic accuracy. The "ground truth" here is compliance with ASTM standards, not expert consensus on medical images or patient outcomes.
4. Adjudication method for the test set
- Not applicable. As noted above, this is a mechanical performance test, not a study involving human readers or adjudicators of medical data.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a non-clinical performance study of a spinal fixation device, not a study involving AI or human readers of medical images.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- No. This is a non-clinical performance study of a spinal fixation device, not an algorithm.
7. The type of ground truth used
- The "ground truth" implicitly used for this non-clinical testing is the mechanical performance requirements specified in ASTM F1717–09 standards for static compression bend, static torsion, and dynamic compression bend testing. The device's performance was compared against these established industry standards and against predicate devices.
8. The sample size for the training set
- Not applicable. This is a non-clinical performance study of a physical device, not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established
- Not applicable for the same reason as point 8.
Ask a specific question about this device
Page 1 of 1