Search Results
Found 129 results
510(k) Data Aggregation
(34 days)
| Toxicology |
| PTG | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Methadone (MTD) | 862.3620
The WEBEST Multi-Drug Urine Cup is a rapid lateral flow immunoassays for the qualitative detection of D-Amphetamine, Butalbital, Buprenorphine, Oxazepam, Benzoylecgonine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Fentanyl, Methylenedioxymethamphetamine, D-Methamphetamine, Methadone, 6-Monoacetylmorphine, Norfentanyl, Morphine, Oxycodone, Phencyclidine, D-Propoxyphene, Nortriptyline, 11-nor-Δ9-THC-COOH and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/ml) |
|---|---|---|
| AMP | D-Amphetamine | 1000 or 500 |
| BAR | Butalbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 300 or 150 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 |
| FYL | Fentanyl | 1 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | D-Methamphetamine | 1000 or 500 |
| MTD | Methadone | 300 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
| NFYL | Norfentanyl | 5 |
| MOP/OPI | Morphine | 2000 or 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | D-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-COOH | 50 |
| TML | Tramadol | 100 |
The single or multi-test cup can consist of up to nineteen (19) of the above listed analytes in any combination with or without on-board adulteration/ specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Webest Home Multi-Drug Urine Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:
| CODE | SUBSTANCE | CUT-OFF (ng/ml) |
|---|---|---|
| AMP | D-Amphetamine | 1000 or 500 |
| BAR | Butalbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 300 or150 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 |
| FYL | Fentanyl | 1 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | D-Methamphetamine | 1000 or 500 |
| MTD | Methadone | 300 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
| NFYL | Norfentanyl | 5 |
| MOP/OPI | Morphine | 2000 or300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | D-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-COOH | 50 |
| TML | Tramadol | 100 |
This is not a screening device to monitor prescription medication.
The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.
This device provides only a preliminary test result. A more specific alternate method shall be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.
Webest Multi-Drug Urine Cup and Webest Home Multi-Drug Urine Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
N/A
Ask a specific question about this device
(111 days)
for Clinical Use |
| Product Codes, Regulation Numbers | • CFR, 21 CFR 862.1345• DJR 21 CFR 862.3620
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.
Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).
Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.
The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.
N/A
Ask a specific question about this device
(111 days)
for Clinical Use |
| Product Codes, Regulation Numbers | • CFR, 21 CFR 862.1345• DJR 21 CFR 862.3620
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.
Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).
Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.
The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.
The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.
N/A
Ask a specific question about this device
(318 days)
21 CFR 862.3170 | Clinical Toxicology (91) |
| DJR | Methadone Test System (Methadone) | II | 21 CFR 862.3620
The Evidence MultiSTAT DOA Urine MultiPlex is intended for use with the Evidence MultiSTAT. The Evidence MultiSTAT is an analyser intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine at the associated cut offs.
The Evidence MultiSTAT DOA Urine MultiPlex detects the following drugs at the following cut offs:
| Analyte | Analyte in Cut Off Material | Cut Off |
|---|---|---|
| Methamphetamine | S-(+)-Methamphetamine | 500ng/ml |
| Noroxycodone | Noroxycodone | 100ng/ml |
| Benzodiazepines 1 | Oxazepam | 200ng/ml |
| Methadone | (+)-Methadone | 300ng/ml |
| Phenobarbital | Phenobarbital | 200ng/ml |
| Tramadol | Tramadol | 200ng/ml |
| Phencyclidine | Phencyclidine | 25ng/ml |
| Buprenorphine | Norbuprenorphine | 5ng/ml |
| 6-Acetylmorphine | 6-Acetylmorphine | 10ng/ml |
| Fentanyl | Fentanyl | 1ng/ml |
| Benzodiazepines 2 | Lorazepam | 200ng/ml |
| Opiates | Morphine | 300ng/ml |
| Benzoylecgonine/Cocaine | Benzoylecgonine | 150ng/ml |
| Cannabinoids (THC) | (-)-11-nor-9-Carboxy-Δ9-THC | 50ng/ml |
| Amphetamine | S-(+)-Amphetamine | 500ng/ml |
The Evidence MultiSTAT DOA Urine Multiplex provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and/or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS) and the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
For In Vitro Diagnostic use only.
FOR PRESCRIPTION USE ONLY
The Evidence MultiSTAT analyzer is a benchtop fully automated Biochip Array System. It performs simultaneous detection of multiple analytes from a single sample. The core technology is the Randox Biochip, a solid-state device containing an array of discrete test regions containing immobilized antibodies specific to different Drugs of Abuse (DOA) compound classes. A competitive chemiluminescent immunoassay is used for the DOA assays with the drug in the specimen and drug labelled with horseradish peroxidase (HRP) being in direct competition for the antibody binding sites. Increased levels of drug in a specimen will lead to reduced binding of drug labelled with HRP and thus a reduction in chemiluminescence being emitted. The light signal generated from each of the test regions on the biochip is detected by a Charge Coupled Device (CCD) camera in the Evidence MultiSTAT system which, together with the analyzer software, is used to quantify the light output and produce meaningful results.
The immunoassay processes are performed automatically in a self-contained and sealed biochip cartridge, which holds the biochips, the reagents, wash buffer and other fluids required for the test to be conducted.
Evidence MultiSTAT assays employ a qualitative reporting method. Each test sample is assayed against the provided Cut Off material of known concentration, which is used to determine the classification of the samples based on the comparison of the signal output.
The Evidence MultiSTAT DOA Urine MultiPlex (EV4393) will be supplied as a test kit comprising:
- 12 x Urine Test Cartridges
- 6 x 1 ml Urine Cut Off Material (lyophilized)
- 4 x 1 ml Urine Positive Control Material (lyophilized)
- 2 x 10 ml Reconstitution Buffer
- 1 x Batch Barcodes
Each kit is supplied with the Evidence MultiSTAT Accessory kit (EV4116) which contains:
- 12 x MultiSTAT Tip Cartridges
- 1 x Tip/Waste Cartridge
- 6 x 1000 µl Pipette Tip
- 1 x Liquid Absorber
Reagent Composition
MultiSTAT DOA Urine MultiPlex Assay Diluent
20 mM phosphate buffer, pH 7.0 containing protein, detergents, and preservatives. This is contained within the cartridge.
MultiSTAT DOA Urine MultiPlex Conjugate
20 mM Tris based buffer, pH 7.0 containing protein, preservatives, and horseradish peroxidase - labelled drug derivatives. This is contained within the cartridge.
MultiSTAT DOA Urine MultiPlex Biochip
Solid substrate containing immobilized antibody discrete test regions. This is contained within the cartridge.
MultiSTAT DOA Urine MultiPlex Wash Buffer
20 mM Tris buffered saline, pH 7.4, containing surfactant and preservatives. This is contained within the cartridge.
LUM-EV934/PX
Luminol-EV934 and Peroxide are contained within the cartridge and are mixed in a ratio of 1:1 by the analyser to give the working signal reagent
MultiSTAT DOA Urine MultiPlex Cut Off
Lyophilised, 20 mM phosphate buffer, pH 7.2 containing stabilizers, preservatives and drug concentrations at the assay cut off values (detailed in Table 3 above).
MultiSTAT DOA Urine MultiPlex Positive Control
Lyophilised, 20 mM phosphate buffer, pH 7.2 containing stabilizers, preservatives, and drug concentrations.
MultiSTAT Reconstitution Buffer
A solution at a neutral pH containing preservatives.
N/A
Ask a specific question about this device
(36 days)
Toxicology |
| PTG | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)Methadone (MTD) | 862.3620
The Healgen Accurate Urine Drug Screen Dip Card is a rapid lateral flow immunoassays for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, d/l-Methadone, d-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, THC-COOH, Norfentanyl and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| 6-MAM | 6-Monoacetylmorphine | 10 |
| AMP | d-Amphetamine | 500 or 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 150 or 300 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | d-Methamphetamine | 500 or 1000 |
| MTD | Methadone | 300 |
| OPI | Morphine | 300 or 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | d-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-∆9-THC-COOH | 50 |
| NFYL | Norfentanyl | 5 |
| TML | Tramadol | 100 |
The single or multi-test panels can consist of up to eighteen (18) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Healgen Accurate Home Urine Drug Test Dip Card is a rapid qualitative immunoassay. The device provides preliminary results for the detection of potential abuse of one or more drugs in human urine at the cutoff concentrations of table below.
This is not a screening device to monitor prescription medication.
| CODE | SUBSTANCE | CUT-OFF (ng/mL) |
|---|---|---|
| AMP | Amphetamine | 1000 or 500 |
| BUP | Buprenorphine | 10 |
| BAR | Secobarbital | 300 |
| BZO | Oxazepam | 300 |
| COC | Cocaine | 300 or 150 |
| EDDP | EDDP | 300 |
| MET/mAMP | Methamphetamine | 1000 or 500 |
| MDMA | Ecstasy | 500 |
| OPI | Morphine | 2000 or 300 |
| MTD | Methadone | 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | Marijuana | 50 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
| NFYL | Norfentanyl | 5 |
| TML | Tramadol | 100 |
This drug test dip card may contain any combination of the drug tests listed in the table above but only one cutoff concentration under same drug condition will be included per device.
This test provides only preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. GC/MS or LC/MS is the preferred confirmatory method. Evaluate preliminary positive results carefully.
Healgen Accurate Home Urine Drug Test Dip Card and Healgen Accurate Urine Drug Screen Dip Card are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine. The device is a dip card format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
N/A
Ask a specific question about this device
(97 days)
Toxicology (91) |
| PTG - 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Methadone (MTD) | 862.3620
SAFElife™ T-Cup Multi-Drug Urine Test Cup:
SAFElife™ T-Cup Multi-Drug Urine Test Cup is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cutoff Level |
|---|---|
| 6-Monoacetylmorphine (6-MAM) | 10 ng/mL |
| Amphetamine (AMP) | 1000 ng/mL or 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 300 ng/mL or 150 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Fentanyl (FTY) | 1 ng/mL |
| Methamphetamine (MET/mAMP) | 1000 ng/mL or 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP) | 300 ng/mL |
| Opiates (OPI) | 2000 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Norfentanyl (NFTY) | 5 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Propoxyphene (PPX) | 300 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Cannabinoids (THC) | 50 ng/mL or 20 ng/mL |
| Tramadol (TRA) | 100 ng/mL |
SAFElife™ T-Cup Multi-Drug Urine Test Cup offers any combinations from 1 to 19 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.
The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.
SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx:
SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methylenedioxymethamphetamine, Methamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine with below cutoff
concentrations and approximate detection time:
| Drug (Identifier) | Calibrator | Cut-off Level |
|---|---|---|
| 6-Monoacetylmorphine (6-MAM) | 6-Monoacetylmorphine | 10 ng/mL |
| Amphetamine (AMP500) | d-Amphetamine | 500 ng/mL |
| Amphetamine (AMP1000) | d-Amphetamine | 1000 ng/mL |
| Secobarbital (BAR) | Secobarbital | 300 ng/mL |
| Buprenorphine (BUP) | Buprenorphine | 10 ng/mL |
| Oxazepam (BZO) | Oxazepam | 300 ng/mL |
| Cocaine (COC150) | Benzoylecgonine | 150 ng/mL |
| Cocaine (COC300) | Benzoylecgonine | 300 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 2-ethylidene-1,5-dimethyl-3,3-diphenyl-pyrrolidine | 300 ng/mL |
| Fentanyl (FTY) | Fentanyl | 1 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 3,4-Methylenedioxymethamphetamine | 500 ng/mL |
| Methamphetamine (MET500/mAMP500) | D(+)-Methamphetamine | 500 ng/mL |
| Methamphetamine (MET1000/mAMP1000) | D(+)-Methamphetamine | 1000 ng/mL |
| Morphine (MOP300) | Morphine | 300 ng/mL |
| Opiates (OPI2000) | Morphine | 2000 ng/mL |
| Methadone (MTD) | Methadone | 300 ng/mL |
| Norfentanyl (NFTY) | Norfentanyl | 5 ng/mL |
| Oxycodone (OXY) | Oxycodone | 100 ng/mL |
| Phencyclidine (PCP) | Phencyclidine | 25 ng/mL |
| Propoxyphene (PPX) | d-Propoxyphene | 300 ng/mL |
| Nortriptyline (TCA) | Nortriptyline | 1000 ng/mL |
| Cannabinoids (THC20) | 11-nor-Δ9-THC-9-COOH | 20 ng/mL |
| Cannabinoids (THC50) | 11-nor-Δ9-THC-9-COOH | 50 ng/mL |
| Tramadol (TRA) | Tramadol | 100 ng/mL |
SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx offers any combinations from 1 to 19 drugs of abuse tests with or without on-board adulteration/specimen validity test (SVT) but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.
The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.
SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids, 6-Monoacetylmorphine, Fentanyl, Norfentanyl, and Tramadol in human urine. Each SAFElife™ T-Cup Multi-Drug Urine Test Cup or SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx device consists of a test cup and a package insert. Each test cup is sealed with one sachet of desiccant in an aluminum pouch. SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx have a general identical design. The SAFElife™ T-Cup Multi-Drug Urine Test Cup is for over-the-counter (OTC) use and the SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is for professional use.
N/A
Ask a specific question about this device
(30 days)
| Toxicology |
| PTG | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Methadone (MTD) | 862.3620
The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:
| Abbreviation | Analyte | Calibrator | Cut-off (ng/mL) |
|---|---|---|---|
| 6AM | 6-Acetylmorphine | 6-Monoacetylmorphine | 10 |
| AMP | Amphetamine | d-Amphetamine | 500 or 1000 |
| BAR | Secobarbital | Secobarbital | 300 |
| BUP | Buprenorphine | Buprenorphine | 10 |
| BZO | Oxazepam | Oxazepam | 300 |
| COC | Cocaine | Benzoylecgonine | 150 or 300 |
| EDDP | EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 |
| FEN | Fentanyl | Fentanyl | 1 |
| FEN | Norfentanyl | Norfentanyl | 5 |
| MDMA | Ecstasy | d,l-Methylenedioxymethamphetamine | 500 |
| MET | Methamphetamine | d-Methamphetamine | 500 or 1000 |
| MOR | Morphine | Morphine | 300 |
| MTD | Methadone | d/l-Methadone | 300 |
| OPI | Opiates | Morphine | 2000 |
| OXY | Oxycodone | Oxycodone | 100 |
| PCP | Phencyclidine | Phencyclidine | 25 |
| PPX | Propoxyphene | Propoxyphene | 300 |
| TCA | Nortriptyline | Nortriptyline | 1000 |
| THC | Marijuana | 11-nor-Δ9-THC-9 COOH | 20 or 50 |
The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.
This in vitro diagnostic device provides only a preliminary test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.
The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:
| Drug | Cut-off (ng/mL) | Abbreviation |
|---|---|---|
| 6-monoacetylmorphine | 10 | 6AM |
| Amphetamine | 500 / 1000 | AMP |
| Secobarbital | 300 | BAR |
| Buprenorphine | 10 | BUP |
| Oxazepam | 300 | BZO |
| Cocaine | 150 / 300 | COC |
| EDDP | 300 | EDDP |
| Fentanyl | 1 | FEN |
| Norfentanyl | 5 | FEN |
| Ecstasy | 500 | MDMA |
| Methamphetamine | 500 / 1000 | MET |
| Methadone | 300 | MTD |
| Opiates | 300 / 2000 | MOR/OPI |
| Oxycodone | 100 | OXY |
| Phencyclidine | 25 | PCP |
| Propoxyphene | 300 | PPX |
| Nortriptyline | 1000 | TCA |
| Marijuana | 20 / 50 | THC |
The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.
This device provides only a preliminary test result. A more specific alternate method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.
VINScreen Urine Drug Test Cup and VINScreen Urine Drug Home Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
N/A
Ask a specific question about this device
(30 days)
Cocaine (COC) | 862.3250, Cocaine Test System | Toxicology |
| PTG | 2-ethylidene-1,5-dimethyl-3,3- | 862.3620
Wisdiag Multi-Drug Urine Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Marijuana, Tramadol, Fentanyl, 6-Monoacetylmorphine and Norfentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 500 ng/mL or 1000 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL or 300 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Methamphetamine (MET) | 500 ng/mL or 1000 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP 300/OPI 2000) | 300 ng/mL or 2000 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Propoxyphene (PPX) | 300 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| Tramadol (TML) | 100 ng/mL |
| Fentanyl (FYL) | 1 ng/mL |
| 6-Monoacetylmorphine (6-MAM) | 10 ng/mL |
| Norfentanyl (NFYL) | 5 ng/mL |
The single or multi-test cups can consist of up to nineteen (19) of the above listed analytes in any combination but only one cutoff concentration under same drug condition will be included per device with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
Wisdiag Multi-Drug Urine Home Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Marijuana, Tramadol, Fentanyl, 6-Monoacetylmorphine and Norfentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 500 ng/mL or 1000 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL or 300 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Methamphetamine (MET) | 500 ng/mL or 1000 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP 300/OPI 2000) | 300 ng/mL or 2000 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Propoxyphene (PPX) | 300 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| Tramadol (TML) | 100 ng/mL |
| Fentanyl (FYL) | 1 ng/mL |
| 6-Monoacetylmorphine (6-MAM) | 10 ng/mL |
| Norfentanyl (NFYL) | 5 ng/mL |
Wisdiag Multi-Drug Urine Home Test Cup offers any combinations from 1 to 19 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only. It is intended for over-the-counter use.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Wisdiag Multi-Drug Urine Test Cup and Wisdiag Multi-Drug Urine Home Test Cup are rapid, single-use in vitro diagnostic devices. Each test kit contains a test device in one pouch. One pouch contains a test Wisdiag Cup and two desiccants, and a package insert. The device is in a ready-to-use format and no longer requires assembly before use.
N/A
Ask a specific question about this device
(51 days)
Toxicology |
| PTG | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)Methadone (MTD) | 862.3620
The CLUNGENE Multi-Drug Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:
| Drug | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| Morphine (MOP/OPI300) | Morphine | 300 |
| Morphine (MOP/OPI2000) | Morphine | 2,000 |
| Methamphetamine (mAMP/MET1000) | D-Methamphetamine | 1,000 |
| Methamphetamine (mAMP/MET500) | D-Methamphetamine | 500 |
| Cocaine (COC300) | Benzoylecgonine | 300 |
| Cocaine (COC150) | Benzoylecgonine | 150 |
| Marijuana (THC) | 11-nor-9-THC-9-COOH | 50 |
| Methylenedioxymethamphetamine (MDMA) | D,L-Methylenedioxymethamphetamine | 500 |
| Buprenorphine (BUP) | Buprenorphine | 10 |
| Propoxyphene (PPX) | D-Propoxyphene | 300 |
| Amphetamine (AMP1000) | D-Amphetamine | 1,000 |
| Amphetamine (AMP500) | D-Amphetamine | 500 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
| Methadone metabolite (EDDP) | 2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 |
| Oxycodone (OXY) | Oxycodone | 100 |
| Oxazepam (BZO) | Oxazepam | 300 |
| Nortriptyline (TCA) | Nortriptyline | 1,000 |
| Secobarbital (BAR) | Secobarbital | 300 |
| Methadone (MTD) | Methadone | 300 |
| 6-Monoacetylmorphine (6-MAM) | 6-Monoacetylmorphine | 10 |
| Fentanyl (FYL) | Fentanyl | 1 |
The single or multi-test cups can consist of any combination of the above listed drug analytes, but only one cut off concentration under same drug condition will be included per device.
This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Evaluate preliminary positive results carefully. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The CLUNGENE Multi-Drug Home Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:
| Drug (Identifier) | Cut-off (ng/mL) |
|---|---|
| Morphine (MOP/OPI2000) | 300 or 2000 |
| Methamphetamine (mAMP/MET) | 500 or 1,000 |
| Cocaine (COC) | 150 or 300 |
| Marijuana (THC) | 50 |
| Methylenedioxymethamphetamine (MDMA) | 500 |
| Buprenorphine (BUP) | 10 |
| Propoxyphene (PPX) | 300 |
| Amphetamine (AMP) | 500 or 1,000 |
| Phencyclidine (PCP) | 25 |
| Methadone metabolite (EDDP) | 300 |
| Oxycodone (OXY) | 100 |
| Oxazepam (BZO) | 300 |
| Nortriptyline (TCA) | 1,000 |
| Secobarbital (BAR) | 300 |
| Methadone (MTD) | 300 |
| 6-Monoacetylmorphine (6-MAM) | 10 |
| Fentanyl (FYL) | 1 |
| The single or multi-test cup offers any combination from above 1 to 17 drugs, but only one cut off concentration under same drug condition will be included per device. |
The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. GC/MS or LC-MS/MS is the preferred confirmatory method.
It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.
CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
This document provides details on the performance characteristics of the CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup. Since this is an in vitro diagnostic device (specifically, a drug screening test), the acceptance criteria and study design are typically focused on analytical performance (accuracy, precision, analytical specificity) rather than a multi-reader multi-case (MRMC) comparative effectiveness study, which is more common for imaging AI. Similarly, "human readers improving with AI vs without AI" is not applicable here as the device is the test, not an aid to human interpretation of another modality.
Here's the breakdown based on the provided text:
Acceptance Criteria and Reported Device Performance
The core acceptance criterion for qualitative drug screening tests like this is accurate detection around a specific cutoff concentration. The reported performance demonstrates the device's ability to correctly classify samples as positive or negative relative to these cutoffs.
Table of Acceptance Criteria and Reported Device Performance (Analytical Precision/Reproducibility)
The "Acceptance Criteria" column represents the desired performance for a qualitative assay around its cutoff. For positive results, this means detecting drug concentrations above the cutoff, and for negative results, it means not detecting concentrations below the cutoff. The provided precision data shows the number of positive (+) and negative (-) results out of 50 tests for various concentrations relative to the cutoff. An ideal performance would show 100% positive for concentrations above cutoff and 100% negative for concentrations below, with roughly 50/50 split at the cutoff itself (due to inherent variability).
| Drug (Cut-off ng/mL) | Acceptance Criteria (Implicit for qualitative assay) | Reported Device Performance (Accuracy as evidenced by reproducibility at various concentrations) Number of negative/positive results out of 50 tests. Values are aggregated across 3 lots where available. |
|---|---|---|
| MOP300 | All samples >cutoff should test positive; all samples <cutoff should test negative. Around cutoff, results vary. | +100% cutoff: 0-/50+ (100% positive) |
| +75% cutoff: 0-/50+ (100% positive) | ||
| +50% cutoff: 0-/50+ (100% positive) | ||
| +25% cutoff: 1-2/48-49+ (96-98% positive) | ||
| cutoff: 23-27-/23-27+ (46-54% positive) | ||
| -25% cutoff: 49-50-/0-1+ (98-100% negative) | ||
| -50% cutoff: 50-/0+ (100% negative) | ||
| -75% cutoff: 50-/0+ (100% negative) | ||
| -100% cutoff: 50-/0+ (100% negative) | ||
| MET1000 | (Same as above) | +100% cutoff: 0-/50+ (100% positive) |
| +75% cutoff: 0-/50+ (100% positive) | ||
| +50% cutoff: 0-/50+ (100% positive) | ||
| +25% cutoff: 0-1/49-50+ (98-100% positive) | ||
| cutoff: 25-26-/24-25+ (48-52% positive) | ||
| -25% cutoff: 49-50-/0-1+ (98-100% negative) | ||
| -50% cutoff: 50-/0+ (100% negative) | ||
| -75% cutoff: 50-/0+ (100% negative) | ||
| -100% cutoff: 50-/0+ (100% negative) | ||
| COC300 | (Same as above) | +100% cutoff: 0-/50+ (100% positive) |
| +75% cutoff: 0-/50+ (100% positive) | ||
| +50% cutoff: 0-/50+ (100% positive) | ||
| +25% cutoff: 1/49+ (98% positive) | ||
| cutoff: 23-25-/25-27+ (46-54% positive) | ||
| -25% cutoff: 49-50-/0-1+ (98-100% negative) | ||
| -50% cutoff: 50-/0+ (100% negative) | ||
| -75% cutoff: 50-/0+ (100% negative) | ||
| -100% cutoff: 50-/0+ (100% negative) | ||
| (Similar detailed tables for all 20 analytes and two configurations would follow the same pattern as the MOP300, MET1000, and COC300 examples shown above, demonstrating consistent reproducibility around the cutoffs.) |
Note: The implicit acceptance criterion for a qualitative test like this is generally that samples significantly above the cutoff should consistently yield positive results, samples significantly below the cutoff should consistently yield negative results, and samples near the cutoff (e.g., +/- 25% or 50% of the cutoff) will show varying results due to inherent assay variability, which is considered acceptable.
Study Details:
-
Sample Size Used for the Test Set and Data Provenance:
- Analytical Performance (Precision/Reproducibility): For each drug and each concentration point (9 concentration levels per drug), 50 tests were performed (2 runs per day for 25 days). Given there are 20 analytes (including the alternative cutoffs), this amounts to 20 drugs * 9 concentrations * 50 tests/concentration = 9000 tests.
- Analytical Specificity/Interference: Not explicitly stated as a "test set" size with a fixed number of samples, but "drug metabolites and other components" were "spiked into drug-free urine" and tested using three lots of the device. For compounds showing no interference, they were tested at a "concentration of 100µg/mL or specified concentrations" in both drug-free urine and urine containing target drugs at +/- 50% cutoff. Over 100 compounds were listed.
- Method Comparison Study: For each drug, 80 "unaltered urine clinical samples" were used (40 negative and 40 positive). These were "blind labeled." With 20 analytes, this sums to 20 drugs * 80 samples/drug = 1600 clinical samples.
- Lay Person Study: 280 lay persons participated. Urine samples were prepared at 7 concentration levels (-100%, +/-75%, +/-50%, +/-25% of cutoff). Each participant received 1 blind-labeled sample and 1 device. The tables suggest that for each configuration (1 and 2), for each drug, 20 samples were tested at each concentration level. Thus, for Configuration 1, there are 17 drugs, so 17 drugs * 7 concentrations * 20 samples/conc = 2380 samples. For Configuration 2, there are 17 drugs, so 17 drugs * 7 concentrations * 20 samples/conc = 2380 samples.
- Data Provenance: The analytical and method comparison studies were performed "in-house." The lay user study was performed "at three intended user sites." The origin of the urine samples (e.g., country of origin) is not specified. It is implied these are prospective tests using prepared or collected samples for the purpose of the study.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Analytical Precision/Reproducibility, Analytical Specificity/Interference, and Method Comparison: The ground truth for these studies was established by LC-MS/MS or GC/MS (or LC-MS/MS), which are reference analytical methods, not human expert consensus. The text states:
- "Each drug concentration was confirmed by LC-MS/MS" for precision studies.
- "The samples were…compared to LC-MS/MS results" for the method comparison study.
- "The concentrations of the samples were confirmed by LC-MS/MS" for the lay person study.
- Therefore, human experts were not directly establishing the ground truth for classification.
- Analytical Precision/Reproducibility, Analytical Specificity/Interference, and Method Comparison: The ground truth for these studies was established by LC-MS/MS or GC/MS (or LC-MS/MS), which are reference analytical methods, not human expert consensus. The text states:
-
Adjudication Method for the Test Set:
- Not applicable as the ground truth was established by LC-MS/MS/GC/MS, a definitive chemical analysis method, not by human expert reading requiring adjudication. The device itself is an immunoassay, the results of which are compared to the LC-MS/MS gold standard.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This type of study design is not applicable to a lateral flow immunoassay drug test. The device is a diagnostic test itself, not an AI assisting human interpretation of another modality.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, implicitly. The precision, specificity, and method comparison studies evaluate the performance of the device itself (the immunoassay) against confirmed concentrations (LC-MS/MS), which represents its "standalone" analytical performance. However, there is a human element in reading the qualitative bands (positive/negative line), which is addressed in the lay-person study.
-
The Type of Ground Truth Used:
- The primary ground truth used across all analytical studies (precision, specificity, method comparison, lay person study) was LC-MS/MS or GC/MS results. This is considered a highly accurate and definitive chemical confirmatory method for drug concentrations in urine.
-
The Sample Size for the Training Set:
- This device is a lateral flow immunoassay, not a machine learning/AI algorithm that requires a "training set" in the computational sense. Its "training" is inherent in its chemical and biological design. Therefore, this question is not applicable.
-
How the Ground Truth for the Training Set was Established:
- As this is not an AI/ML device relying on a training set, this question is not applicable. The device's performance is governed by its chemical design (antibodies, reagents) and manufacturing process, which are developed and validated through iterative biochemical and engineering studies, not by a data-driven training process in the AI sense.
Ask a specific question about this device
(25 days)
Toxicology (91) |
| PTG 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) Methadone | II | 21 CFR § 862.3620
The AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, EDDP, Fentanyl, Ecstasy, Propoxyphene, Morphine, Methadone, Phencyclidine, Oxycodone, Norfentanyl, Methamphetamine, Nortriptyline, 6-Monoacetylmorphine, Tramadol and Marijuana in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 500 ng/mL or 300 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL |
| EDDP | 300ng/ml |
| Fentanyl (FYL) | 1 ng/mL |
| Ecstasy (MDMA) | 500 ng/mL |
| Propoxyphene (PPX) | 300 ng/mL |
| Morphine (MOR) | 2000 ng/mL or 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Norfentanyl (NFYL) | 5 ng/mL |
| Methamphetamine (MET) | 500 ng/mL or 300 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| 6-Monoacetylmorphine (6-MAM) | 10 ng/mL |
| Tramadol (TML) | 100 ng/mL |
| Marijuana (THC) | 50 ng/mL or 20 ng/mL |
Configuration of the AssureTech Quick Cup Tests can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs above when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.
For in vitro diagnostic use only.
The AssureTech Multi-drug Urine Test Cup are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, EDDP, Fentanyl, Ecstasy, Propoxyphene, Morphine, Methadone, Phencyclidine, Oxycodone, Norfentanyl, Methamphetamine, Nortriptyline, 6-Monoacetylmorphine, Tramadol and Marijuana in human urine at the cutoff concentrations of:
| Drug (Identifier) | Calibrator | Cut-off level |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 500 ng/mL or 300 ng/mL |
| Secobarbital (BAR) | Secobarbital | 300 ng/mL |
| Buprenorphine (BUP) | BUP-3-D-Glucuronide | 10 ng/mL |
| Oxazepam (BZO) | Oxazepam | 300 ng/mL |
| Cocaine (COC) | Benzoylecgonine | 150 ng/mL |
| EDDP | 2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine | 300ng/ml |
| Fentanyl (FYL) | Fentanyl | 1 ng/mL |
| Ecstasy (MDMA) | 3,4-Methylenediioxy-MET | 500 ng/mL |
| Propoxyphene (PPX) | D-Propoxyphene | 300 ng/mL |
| Morphine (MOR) | Morphine | 2000 ng/mL or 300 ng/mL |
| Methadone (MTD) | Methadone | 300 ng/mL |
| Phencyclidine (PCP) | Phencyclidine | 25 ng/mL |
| Oxycodone (OXY) | Oxycodone | 100 ng/mL |
| Norfentanyl (NFYL) | Norfentanyl | 5 ng/mL |
| Methamphetamine (MET) | Methamphetamine | 500 ng/mL or 300 ng/mL |
| Nortriptyline (TCA) | Nortriptyline | 1000 ng/mL |
| 6-Monoacetylmorphine (6-MAM) | 6-Monoacetylmorphine | 10 ng/mL |
| Tramadol (TML) | Cis-Tramadol | 100 ng/mL |
| Marijuana (THC) | 11-nor-9-THC-9-COOH | 50 ng/mL or 20 ng/mL |
Configuration of the AssureTech Multi-drug Urine Test Cup can consist of any combination of the above listed drug analytes. It is for in vitro diagnostic use only.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The AssureTech Quick Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, EDDP, Fentanyl, Ecstasy, Propoxyphene, Morphine, Methadone, Phencyclidine, Oxycodone, Norfentanyl, Methamphetamine, Nortriptyline, 6-Monoacetylmorphine, Tramadol and Marijuana (target analytes) in human urine. The products are single-use in vitro diagnostic devices. Each test kit contains a Test Device and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.
The provided FDA 510(k) Clearance Letter details the performance of the AssureTech Quick Cup Tests and AssureTech Multi-drug Urine Test Cup for qualitative and simultaneous detection of various drugs in human urine.
Here's an analysis of the acceptance criteria and the study proving the device meets those criteria:
1. Acceptance Criteria and Reported Device Performance
For in vitro diagnostic devices like these, acceptance criteria are typically related to the accuracy of the qualitative detection (positive vs. negative) compared to a gold standard, particularly around the established cutoff concentrations. The performance is assessed through analytical studies (precision, specificity, interference) and comparison studies with a confirmatory method.
Here's a table summarizing the implicit acceptance criteria based on the precision and lay-user studies, and the reported device performance. The acceptance criterion is inferred as the ideal performance for these types of tests, where results near or above the cutoff should be positive, and results significantly below should be negative. The performance data below is extracted from the "Precision" and "Lay-user study" sections.
Table of Acceptance Criteria and Reported Device Performance
| Drug (Identifier) | Cut-off Level | Implicit Acceptance Criterion (Qualitative) | Reported Performance (Precision Study - Total of 3 Lots, 50 observations per concentration) | Reported Performance (Lay-user study - Agreement %) |
|---|---|---|---|---|
| AMP300 | 300 ng/mL | -100% to -25% Cutoff: All Negative | -100%/-75%/-50%/-25% Cutoff: 100% Negative (50-/0+) | -100% to -25% Cutoff: 95%-100% agreement |
| +25% to +100% Cutoff: All Positive | +25%/+50%/%75%/+100% Cutoff: 100% Positive (50+/0-) | +25% to +75% Cutoff: 95%-100% agreement | ||
| Cutoff: Majority Positive, some Negative acceptable (reflecting assay variability) | Cutoff: 64-68% Positive (30-34+/16-20-) | Not explicitly reported for cutoff in lay-user study | ||
| MET300 | 300 ng/mL | -100% to -25% Cutoff: All Negative | -100%/-75%/-50%/-25% Cutoff: 100% Negative (50-/0+) | -100% to -25% Cutoff: 100% agreement |
| +25% to +100% Cutoff: All Positive | +25%/+50%/%75%/+100% Cutoff: 100% Positive (50+/0-) | +25% to +75% Cutoff: 95%-100% agreement | ||
| Cutoff: Majority Positive, some Negative acceptable (reflecting assay variability) | Cutoff: 66-72% Positive (33-36+/14-17-) | Not explicitly reported for cutoff in lay-user study | ||
| TML100 | 100 ng/mL | -100% to -25% Cutoff: All Negative | -100%/-75%/-50%/-25% Cutoff: 100% Negative (50-/0+) | -100% to -25% Cutoff: 100% agreement |
| +25% to +100% Cutoff: All Positive | +25%/+50%/%75%/+100% Cutoff: 100% Positive (50+/0-) | +25% to +75% Cutoff: 95% agreement | ||
| Cutoff: Majority Positive, some Negative acceptable (reflecting assay variability) | Cutoff: 64-72% Positive (32-36+/14-18-) | Not explicitly reported for cutoff in lay-user study | ||
| All other listed drugs (Configuration 1 & 2) | Various | -100% to -25% Cutoff: All Negative | Data for these drugs were reported in previous 510(k)s (K243996, K201630, K181768, K180349, K170049, K161044, K153465, K152025, K151211). The precision study for AMP300, MET300, TML100 suggests similar performance. | -100% to -25% Cutoff: 90%-100% agreement |
| +25% to +100% Cutoff: All Positive | +25% to +75% Cutoff: 90%-100% agreement |
Note: The precision study for AMP300, MET300, and TML100 used 3 lots, with "50-/0+" meaning 50 negative results and 0 positive results, and "50+/0-" meaning 50 positive results and 0 negative results. For the 'Cutoff' concentration, it shows a mix of positive and negative results, which is expected due to the nature of qualitative assays around the threshold.
2. Sample Sizes and Data Provenance
-
Precision Study:
- For AMP300, MET300, and TML100: Each drug had 8 concentrations (spanning -100% to +100% of cutoff, plus the cutoff). For each concentration, tests were performed two runs per day for 25 days, for 3 different lots.
- This means 50 observations per concentration per lot (2 runs/day * 25 days/run).
- Total observations per drug for all 3 lots: 8 concentrations * 50 observations/concentration * 3 lots = 1200 observations per drug.
- Data for other drugs refer to previous 510(k) clearances (K243996, K201630, K181768, K180349, K170049, K161044, K153465, K152025, K151211).
- Data Provenance: Retrospective, as samples were "prepared by spiking drug in negative samples" and confirmed by LC/MS. No specific country of origin is mentioned, but typically for FDA submissions, studies are conducted under GLP (Good Laboratory Practice) guidelines, often in the US or by international labs adhering to comparable standards.
- For AMP300, MET300, and TML100: Each drug had 8 concentrations (spanning -100% to +100% of cutoff, plus the cutoff). For each concentration, tests were performed two runs per day for 25 days, for 3 different lots.
-
Comparison Studies (Clinical Samples):
- For AMP300, MET300, and TML100: The tables show data broken down by "Negative" (presumably drug-free), "Low Negative" (< -50% cutoff), "Near Cutoff Negative" (-50% to cutoff), "Near Cutoff Positive" (cutoff to +50%), and "High Positive" (> +50%). The sum of the positive and negative results across these categories for each operator represents the number of clinical samples tested for that drug.
- AMP300: 5 (Negative) + 15 (LN) + 19 (NCN) + 24 (NCP) + 16 (HP) = 79 samples per operator. (Operator 1)
- MET300: 4 (Negative) + 13 (LN) + 23 (NCN) + 20 (NCP) + 20 (HP) = 80 samples per operator. (Operator 1)
- TML100: 2 (Negative) + 18 (LN) + 18 (NCN) + 19 (NCP) + 20 (HP) = 77 samples per operator. (Operator 1)
- Total (approximate, as numbers vary slightly between operators): ~79+80+77 = ~236 clinical samples for AMP300, MET300, TML100 combined.
- Data for other drugs refer to previous 510(k) clearances.
- Data Provenance: Retrospective, using "unaltered clinical samples." No specific country of origin is mentioned.
- For AMP300, MET300, and TML100: The tables show data broken down by "Negative" (presumably drug-free), "Low Negative" (< -50% cutoff), "Near Cutoff Negative" (-50% to cutoff), "Near Cutoff Positive" (cutoff to +50%), and "High Positive" (> +50%). The sum of the positive and negative results across these categories for each operator represents the number of clinical samples tested for that drug.
-
Lay-User Study:
- Sample Size: 280 lay persons tested the device.
- Configuration 1: 66 male + 74 female = 140 lay persons.
- Configuration 2: 87 male + 53 female = 140 lay persons.
- Data Provenance: Retrospective, samples were "prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens." Confirmed by LC/MS. Conducted "at three intended user sites." No specific country of origin is mentioned.
- Sample Size: 280 lay persons tested the device.
3. Number of Experts and Qualifications for Ground Truth (Clinical Samples)
- Ground Truth Establishment for Clinical Samples: LC/MS (Liquid Chromatography/Mass Spectrometry) is stated as the preferred confirmatory method and was used to confirm the concentrations of the samples. This is an objective chemical method, considered the gold standard for drug detection and quantification in urine.
- Experts: The comparison studies were performed "in-house with three laboratory assistants." While these individuals are performing the rapid tests, the ultimate ground truth is established by the LC/MS results. The "laboratory assistants" are not explicitly designated as "experts" in establishing ground truth, but rather as trained users of the device whose results are compared to the LC/MS gold standard.
4. Adjudication Method for the Test Set (Clinical Samples)
- The document states that "Operators ran unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results."
- There were three operators. The "Discordant Results" tables show discrepancies between the rapid test results and the LC/MS results, sometimes across multiple operators for the same sample.
- No explicit adjudication method (e.g., 2+1, 3+1) for the rapid test results themselves is described. The comparison seems to be a direct comparison of each operator's rapid test result against the LC/MS ground truth, and then discrepancies are noted. The LC/MS data serves as the final, objective ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- This document describes performance characteristics of an in-vitro diagnostic device (a qualitative urine drug test cup).
- No MRMC comparative effectiveness study was performed in the context of comparing human readers (e.g., radiologists interpreting images) with and without AI assistance. This type of study design is specific to AI/CADe (Computer-Assisted Detection) or CADx (Computer-Assisted Diagnosis) devices in imaging, which is not applicable to a lateral flow immunoassay like the AssureTech Quick Cup Tests.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Given this is a physical immunoassay test cup, the concept of a "standalone (algorithm only without human-in-the-loop performance)" study does not directly apply in the same way it would for a software-based AI device.
- The "Comparison Studies" with laboratory assistants and the "Lay-user study" assess the device's performance when interpreted by human users. The device itself, by producing a visual result (line/no line), is the "algorithm." Its performance is inherently tied to human interpretation of that visual output. The precision and specificity studies represent the analytical performance of the device itself.
7. Type of Ground Truth Used
- Analytical Performance Studies (Precision, Specificity, pH/SG Effect): The ground truth was established by spiking known concentrations of drugs into negative urine samples, with concentrations confirmed by LC/MS.
- Comparison Studies (Clinical Samples): The ground truth was established by LC/MS results on unaltered clinical urine samples. LC/MS is a highly accurate chemical analytical method.
- Lay-User Study: Ground truth was established by spiking known concentrations of drugs into drug-free pooled urine specimens, confirmed by LC/MS.
8. Sample Size for the Training Set
- This document describes a 510(k) submission for a traditional in-vitro diagnostic device (immunoassay). It does not mention any artificial intelligence (AI) or machine learning (ML) components that would typically require a "training set" in the computational sense.
- The terms "training set" and "test set" are common in AI/ML validation. For a traditional medical device, the studies described (precision, interference, specificity, comparison, lay-user) serve as the "validation set" against pre-defined performance criteria.
- Therefore, N/A for "training set" in the context of AI/ML.
9. How the Ground Truth for the Training Set Was Established
- N/A (as above, no "training set" in the AI/ML context).
- However, if we consider how the device itself was developed, the ground truth for optimizing its performance (e.g., antibody binding, membrane characteristics) would have relied on highly controlled experiments with known concentrations of analytes, likely confirmed by advanced analytical chemistry methods like LC/MS. This process is part of the extensive R&D and quality control that precedes a 510(k) submission.
Ask a specific question about this device
Page 1 of 13