K Number
K260355

Validate with FDA (Live)

Date Cleared
2026-03-09

(34 days)

Product Code
Regulation Number
862.3100
Age Range
20 - 50
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The WEBEST Multi-Drug Urine Cup is a rapid lateral flow immunoassays for the qualitative detection of D-Amphetamine, Butalbital, Buprenorphine, Oxazepam, Benzoylecgonine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Fentanyl, Methylenedioxymethamphetamine, D-Methamphetamine, Methadone, 6-Monoacetylmorphine, Norfentanyl, Morphine, Oxycodone, Phencyclidine, D-Propoxyphene, Nortriptyline, 11-nor-Δ9-THC-COOH and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/ml)
AMPD-Amphetamine1000 or 500
BARButalbital300
BUPBuprenorphine10
BZOOxazepam300
COCBenzoylecgonine300 or 150
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FYLFentanyl1
MDMAMethylenedioxymethamphetamine500
METD-Methamphetamine1000 or 500
MTDMethadone300
6-MAM6-Monoacetylmorphine10
NFYLNorfentanyl5
MOP/OPIMorphine2000 or 300
OXYOxycodone100
PCPPhencyclidine25
PPXD-Propoxyphene300
TCANortriptyline1000
THC11-nor-Δ9-THC-COOH50
TMLTramadol100

The single or multi-test cup can consist of up to nineteen (19) of the above listed analytes in any combination with or without on-board adulteration/ specimen validity tests (SVT).

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

The Webest Home Multi-Drug Urine Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

CODESUBSTANCECUT-OFF (ng/ml)
AMPD-Amphetamine1000 or 500
BARButalbital300
BUPBuprenorphine10
BZOOxazepam300
COCBenzoylecgonine300 or150
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FYLFentanyl1
MDMAMethylenedioxymethamphetamine500
METD-Methamphetamine1000 or 500
MTDMethadone300
6-MAM6-Monoacetylmorphine10
NFYLNorfentanyl5
MOP/OPIMorphine2000 or300
OXYOxycodone100
PCPPhencyclidine25
PPXD-Propoxyphene300
TCANortriptyline1000
THC11-nor-Δ9-THC-COOH50
TMLTramadol100

This is not a screening device to monitor prescription medication.

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This device provides only a preliminary test result. A more specific alternate method shall be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

Device Description

Webest Multi-Drug Urine Cup and Webest Home Multi-Drug Urine Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

AI/ML Overview

N/A

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.06.09

March 9, 2026

WEBEST Biotech, LLC
℅ Jenny Xia
LSI International Inc
504 East Diamond Ave., Suite H
Gaithersburg, Maryland 20877

Re: K260355
Trade/Device Name: Webest Multi-Drug Urine Cup; Webest Home Multi-Drug Urine Cup
Regulation Number: 21 CFR 862.3100
Regulation Name: Amphetamine test system
Regulatory Class: Class II
Product Code: NFT PTH NGL NFY NFV PTG NGG NGM QBF QAW NFW
Dated: February 3, 2026
Received: February 3, 2026

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K260355- Jenny Shia Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph Kotarek, Ph.D.
Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Digitally signed by JOSEPH A. KOTAREK -S
Date: 2026.03.09 14:31:09 -04'00'

Page 3

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known): K260355

Device Name: Webest Multi-Drug Urine Cup; Webest Home Multi-Drug Urine Cup

Indications for Use (Describe)

The WEBEST Multi-Drug Urine Cup is a rapid lateral flow immunoassays for the qualitative detection of D-Amphetamine, Butalbital, Buprenorphine, Oxazepam, Benzoylecgonine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Fentanyl, Methylenedioxymethamphetamine, D-Methamphetamine, Methadone, 6-Monoacetylmorphine, Norfentanyl, Morphine, Oxycodone, Phencyclidine, D-Propoxyphene, Nortriptyline, 11-nor-Δ9-THC-COOH and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/ml)
AMPD-Amphetamine1000 or 500
BARButalbital300
BUPBuprenorphine10
BZOOxazepam300
COCBenzoylecgonine300 or 150
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FYLFentanyl1
MDMAMethylenedioxymethamphetamine500
METD-Methamphetamine1000 or 500
MTDMethadone300
6-MAM6-Monoacetylmorphine10
NFYLNorfentanyl5
MOP/OPIMorphine2000 or 300
OXYOxycodone100
PCPPhencyclidine25
PPXD-Propoxyphene300
TCANortriptyline1000
THC11-nor-Δ9-THC-COOH50
TMLTramadol100

The single or multi-test cup can consist of up to nineteen (19) of the above listed analytes in any combination with or without on-board adulteration/ specimen validity tests (SVT).

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

The Webest Home Multi-Drug Urine Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

CODESUBSTANCECUT-OFF (ng/ml)
AMPD-Amphetamine1000 or 500
BARButalbital300

Page 4

BUPBuprenorphine10
BZOOxazepam300
COCBenzoylecgonine300 or150
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FYLFentanyl1
MDMAMethylenedioxymethamphetamine500
METD-Methamphetamine1000 or 500
MTDMethadone300
6-MAM6-Monoacetylmorphine10
NFYLNorfentanyl5
MOP/OPIMorphine2000 or300
OXYOxycodone100
PCPPhencyclidine25
PPXD-Propoxyphene300
TCANortriptyline1000
THC11-nor-Δ9-THC-COOH50
TMLTramadol100

This is not a screening device to monitor prescription medication.

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This device provides only a preliminary test result. A more specific alternate method shall be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) SUMMARY

K260355

  1. Date: March 6, 2026

  2. Submitter: WEBEST Biotech, LLC.
    535 Irwindale Ave.
    Irwindale, CA 91706

  3. Contact person: Jenny Xia
    LSI International Inc.
    504 East Diamond Ave., Suite H
    Gaithersburg, MD 20877
    Telephone: 301-525-6856
    Email: jxia@lsi-consulting.org

  4. Device Name: Webest Multi-Drug Urine Cup
    Webest Home Multi-Drug Urine Cup

  5. Classification: Class II

Product CodeTarget DrugRegulation SectionPanel
NFTAmphetamine (AMP)862.3100, Amphetamine Test SystemToxicology
PTHButalbital (BAR)862.3150, Barbiturate Test SystemToxicology
NGLBuprenorphine (BUP), Fentanyl (FYL), Morphine (MOP/OPI), Oxycodone (OXY), 6-Monoacetylmorphine(6-MAM), Norfentanyl (NFYL), Tramadol (TML)862.3650, Opiate Test SystemToxicology
NFVOxazepam (BZO)862.3170, Benzodiazepine Test SystemToxicology
NFYCocaine (COC)862.3250, Cocaine and cocaine metabolite test systemToxicology
PTG2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Methadone (MTD)862.3620, Methadone Test SystemToxicology
NGGMethylenedioxymethamphetamine (MDMA), Methamphetamine (MET)862.3610, Methamphetamine Test SystemToxicology
QBF862.3700 Propoxyphene test system.Toxicology

Page 6

Propoxyphene(PPX)
NGMPhencyclidine (PCP)UnclassifiedToxicology
QAWNortriptyline (TCA)862.3910 Tricyclic antidepressant drugs test systemToxicology
NFWCannabinoids (THC)862.3870, Cannabinoids Test SystemToxicology
  1. Predicate Devices:
    AllTest Multi-Drug Rapid Test Cup (K244043)

  2. Intended Use

The Webest Home Multi-Drug Urine Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

CODESUBSTANCECUT-OFF (ng/mL)
AMPD-Amphetamine1000 or 500
BARButalbital300
BUPBuprenorphine10
BZOOxazepam300
COCBenzoylecgonine300 or150
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FYLFentanyl1
MDMAMethylenedioxymethamphetamine500
METD-Methamphetamine1000 or 500
MTDMethadone300
6-MAM6-Monoacetylmorphine10
NFYLNorfentanyl5
MOP/OPIMorphine2000 or300
OXYOxycodone100
PCPPhencyclidine25
PPXD-Propoxyphene300
TCANortriptyline1000
THC11-nor-Δ9-THC-COOH50
TMLTramadol100

This is not a screening device to monitor prescription medication.

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This device provides only a preliminary test result. A more specific alternate method shall be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

Page 7

The WEBEST Multi-Drug Urine Cup is a rapid lateral flow immunoassays for the qualitative detection of D-Amphetamine, Butalbital, Buprenorphine, Oxazepam, Benzoylecgonine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Fentanyl, Methylenedioxymethamphetamine, D-Methamphetamine, Methadone, 6-Monoacetylmorphine, Norfentanyl, Morphine, Oxycodone, Phencyclidine, D-Propoxyphene, Nortriptyline, 11-nor-Δ9-THC-COOH and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

TestCalibratorCut-off (ng/mL)
AMPD-Amphetamine1000 or 500
BARButalbital300
BUPBuprenorphine10
BZOOxazepam300
COCBenzoylecgonine300 or 150
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FYLFentanyl1
MDMAMethylenedioxymethamphetamine500
METD-Methamphetamine1000 or 500
MTDMethadone300
6-MAM6-Monoacetylmorphine10
NFYLNorfentanyl5
MOP/OPIMorphine2000 or 300
OXYOxycodone100
PCPPhencyclidine25
PPXD-Propoxyphene300
TCANortriptyline1000
THC11-nor-Δ9-THC-COOH50
TMLTramadol100

The single or multi-test cup can consist of up to nineteen (19) of the above listed analytes in any combination with or without on-board adulteration/ specimen validity tests (SVT).

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

  1. Device Description

Webest Multi-Drug Urine Cup and Webest Home Multi-Drug Urine Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

  1. Substantial Equivalence Information

Similarities

ItemDevicePredicate (K244043)

Page 8

Intended useQualitative detection of drugs of abuse in urine. For over-the-counter useSame.
MethodologyCompetitive binding, lateral flow immunochromatographic assay based on antigen-antibody reactionSame
Type of TestQualitativeSame
Specimen TypeHuman urineSame
Target Drug and Cut Off ValuesTarget Drugs Cutoff (ng/mL)Same except that no Norfentanyl
Amphetamine(AMP) 1000 or 500
Butalbital (BAR) 300
Buprenorphine (BUP) 10
Oxazepam (BZO) 300
Cocaine (COC) 150 or 300
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) 300
Methylenedioxymethamphetamine (MDMA) 500
Methamphetamine (MET) 1000 or 500
Morphine (MOP300/OPI2000) 300 or 2000
Methadone (MTD) 300
Oxycodone (OXY) 100
Phencyclidine (PCP) 25
Propoxyphene(PPX) 300
Nortriptyline (TCA) 1000
Cannabinoids (THC) 50
6-Monoacetylmorphine(6-AM) 10
Fentanyl (FYL) 1
Norfentanyl (NFYL) 5
Tramadol (TML) 100
ConfigurationsTest cupSame
  1. Standard/Guidance Document Reference (if applicable)
    None referenced.

  2. Test Principle

The Webest Multi-Drug Urine Cup or The Webest Home Multi-Drug Urine Cup is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.

Page 9

When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.

When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.

To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.

  1. Performance Characteristics

A. Analytical performance

a. Precision/Reproducibility:

Precision studies were carried out for samples with concentrations of +100% cutoff, +75% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Samples with concentration of -100% cutoff were drug-free urines samples. Other samples were prepared by spiked target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test cups with one operator per lot. The results obtained are summarized in the following tables.

DrugLot Number+100% cutoff+75% cutoff+50% cutoff+25% cutoffCutoff-25% cutoff-50% cutoff-75% cutoff-100% cut-off
6AM 10Lot 10-/50+0-/50+0-/50+0-/50+27+/23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+2-/48+26+/24-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+1-/49+24+/26-50-/0+50-/0+50-/0+50-/0+
AMP 500Lot 10-/50+0-/50+0-/50+0-/50+27+/23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+1-/49+25+/25-49-/1+50-/0+50-/0+50-/0+
BAR 300Lot 10-/50+0-/50+0-/50+1-/49+27+/23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+27+/23-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+26+/24-48-/2+50-/0+50-/0+50-/0+
BUP 10Lot 10-/50+0-/50+0-/50+0-/50+26+/24-48-/2+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+27+/23-48-/2+50-/0+50-/0+50-/0+
BZO 300Lot 10-/50+0-/50+0-/50+0-/50+27+/23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+25+/25-50-/0+50-/0+50-/0+50-/0+

Page 10

Lot 30-/50+0-/50+0-/50+1-/49+26+/24-50-/0+50-/0+50-/0+50-/0+
COC 150Lot 10-/50+0-/50+0-/50+0-/50+25+/25-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+2-/48+27+/23-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+26+/24-49-/1+50-/0+50-/0+50-/0+
EDDP 300Lot 10-/50+0-/50+0-/50+2-/48+27+/23-49-/1+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+24+/26-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+25+/25-49-/1+50-/0+50-/0+50-/0+
FYL 1Lot 10-/50+0-/50+0-/50+1-/49+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+2-/48+25+/25-48-/2+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+25+/25-48-/2+50-/0+50-/0+50-/0+
NFYL 5Lot 10-/50+0-/50+0-/50+0-/50+24+/26-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+25+/25-48-/2+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+27+/23-48-/2+50-/0+50-/0+50-/0+
MDMA 500Lot 10-/50+0-/50+0-/50+0-/50+24+/26-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+27+/23-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+1-/49+23+/27-49-/1+50-/0+50-/0+50-/0+
MET 500Lot 10-/50+0-/50+0-/50+0-/50+27+/23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+1-/49+26+/24-48-/2+50-/0+50-/0+50-/0+
MOR 300Lot 10-/50+0-/50+0-/50+0-/50+27+23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+24+/26-50-/0+50-/0+50-/0+50-/0+
MTD 300Lot 10-/50+0-/50+0-/50+1-/49+25+/25-49-/1+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+27+/23-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+24+/26-50-/0+50-/0+50-/0+50-/0+
OXY 100Lot 10-/50+0-/50+0-/50+0-/50+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-48-/2+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+25+/25-49-/1+50-/0+50-/0+50-/0+
PCP 25Lot 10-/50+0-/50+0-/50+1-/49+25+/25-49-/1+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+1-/49+27+/23-49-/1+50-/0+50-/0+50-/0+
PPX 300Lot 10-/50+0-/50+0-/50+1-/49+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+25+/25-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+26+/24-48-/2+50-/0+50-/0+50-/0+
TCA 1000Lot 10-/50+0-/50+0-/50+0-/50+27+/23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-48-/2+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+27+/23-49-/1+50-/0+50-/0+50-/0+
THC 50Lot 10-/50+0-/50+0-/50+0-/50+25+/25-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+1-/49+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 10-/50+0-/50+0-/50+1-/49+26+/24-49-/1+50-/0+50-/0+50-/0+

Page 11

TML 100Lot 20-/50+0-/50+0-/50+1-/49+26+/24-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+24+/26-50-/0+50-/0+50-/0+50-/0+
AMP 1000Lot 10-/50+0-/50+0-/50+0-/50+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+26+/24-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+26+/24-49-/1+50-/0+50-/0+50-/0+
COC 300Lot 10-/50+0-/50+0-/50+1-/49+27+/23-49-/1+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+23+/27-49-/1+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+25+/25-48-/2+50-/0+50-/0+50-/0+
MET 1000Lot 10-/50+0-/50+0-/50+0-/50+25+/25-49-/1+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+26+/24-50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+2-/48+27+/23-49-/1+50-/0+50-/0+50-/0+
OPI 2000Lot 10-/50+0-/50+0-/50+0-/50+27+/23-50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+23+/27-48-/2+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+1-/49+26+/24-50-/0+50-/0+50-/0+50-/0+

b. Linearity/assay reportable range:

Not applicable. This device is intended for qualitative use only.

c. Stability:

The device is stable at 4-30℃ for 36 months based on accelerated stability study.

d. Analytical specificity/Interference:

To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine. These urine samples were tested using three lots of the device. The results obtained are summarized in the following table. Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100.

6-MAM (6-Acetylmorphine)
(6-Acetylmorphine, Cutoff=10ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Heroin6016.7%
Morphine70,0000.014%
Normorphine100,000(Negative)Not detected
NalorphineHCl100,000(Negative)Not detected
Hydrocodone100,000(Negative)Not detected
Hydromorphone100,000(Negative)Not detected
Chlordiazepoxide100,000(Negative)Not detected
Clobazam100,000(Negative)Not detected
D-Amphetamine100,000(Negative)Not detected
(±)-Amphetamine100,000(Negative)Not detected
Levorphanoltartrate100,000(Negative)Not detected
Codeine100,000(Negative)Not detected

Page 12

Ethylmorphine100,000(Negative)Not detected
Morphine3-β-D-glucuronide100,000(Negative)Not detected
Norcodeine100,000(Negative)Not detected
Oxycodone100,000(Negative)Not detected
Oxymorphone100,000(Negative)Not detected
Procainehydrochloride100,000(Negative)Not detected
Thebaine100,000(Negative)Not detected
6-Acetylcodeine100,000(Negative)Not detected
Buprenorphine100,000(Negative)Not detected
Dihydrocodeine100,000(Negative)Not detected
Dextromethorphan100,000(Negative)Not detected
Imipraminehydrochloride100,000(Negative)Not detected
Meperidine100,000(Negative)Not detected
(±)-Methadone100,000(Negative)Not detected
Mitragynine(kratom)100,000(Negative)Not detected
Morphine-6-β-D-glucuronide100,000(Negative)Not detected
Naloxonehydrochloride100,000(Negative)Not detected
Naltrexonehydrochloride100,000(Negative)Not detected
Naproxen100,000(Negative)Not detected
Norbuprenorphine100,000(Negative)Not detected
Norbuprenorphine-3-D-Glucuronide100,000(Negative)Not detected
Noroxycodone HCL100,000(Negative)Not detected
Noroxymorphone HCL100,000(Negative)Not detected
(+)-Norpropoxyphenemaleate100,000(Negative)Not detected
Oxymorphone-3β-D-glucuronide100,000(Negative)Not detected
TapentadolHCl100,000(Negative)Not detected
Tramadol100,000(Negative)Not detected

AMP(Amphetamine) (Amphetamine, Cutoff=500ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Hydroxyamphetamine4,00012.5%
(+/-)-Methylenedioxyamphetamine(MDA)200250%
D,L-Amphetamine500100%
Diethylstilbestrol2,50020%
L-Amphetamine60,0000.8%
Phentermine4,00012.5%
β-Phenylethylamine50,0001%
Tyramine5,00010%
p-Hydroxynorephedrine50,0001%
D,L-Norephedrine50,0001%
(+/-)3,4-Methylenedioxy-n-100,0000.5%
d-Methamphetamine100,000(Negative)Not detected

Page 13

l-Methamphetamine100,000(Negative)Not detected
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)100,000(Negative)Not detected
Ephedrine hydrochloride100,000(Negative)Not detected
Phenylpropanolamine100,000(Negative)Not detected
Benzphetamine100,000(Negative)Not detected
L-Ephedrine100,000(Negative)Not detected
L-Epinephrine100,000(Negative)Not detected
D,L-Epinephrine100,000(Negative)Not detected

AMP(Amphetamine) (Amphetamine, Cutoff=1000ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Hydroxyamphetamine8,00013%
(+/-)-Methylenedioxyamphetamine(MDA)130077%
D,L-Amphetamine1,000100%
Diethylstilbestrol5,00020%
L-Amphetamine100,0001%
Phentermine100,0001%
β-Phenylethylamine100,0001%
Tyramine10,00010%
p-Hydroxynorephedrine100,0001%
D,L-Norephedrine100,0001%
p-Hydroxyamphetamine100,000(Negative)Not detected
D-Methamphetamine100,000(Negative)Not detected
L-Methamphetamine100,000(Negative)Not detected
Ephedrine hydrochloride100,000(Negative)Not detected
(+/-)3,4- Methylenedioxymethamphetamine100,000(Negative)Not detected
Phenylpropanolamine100,000(Negative)Not detected
Benzphetamine100,000(Negative)Not detected
L-Ephedrine100,000(Negative)Not detected
L-Epinephrine100,000(Negative)Not detected
D,L-Epinephrine100,000(Negative)Not detected
(+/-)3,4-Methylenedioxyethylamphetamine (MDEA)100,000(Negative)Not detected

BAR(Butalbital) (Butalbital, Cutoff=300ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Alphenal250120%
Amobarbital300010%
Aprobarbital200150%
Butabarbital1,00030%

Page 14

Butethal50060%
Cyclopentobarbital300100%
Pentobarbital130023%
Phenobarbital190016%
Secobarbital300100%
Barbital

BUP(Buprenorphine) (Buprenorphine, Cutoff=10ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Buprenorphine -3-D-Glucuronide1566.67%
Norbuprenorphine4025%
Norbuprenorphine-3-D-Glucuronide5002%
Morphine100,000(Negative)Not detected
Oxymorphone100,000(Negative)Not detected
Hydromorphone100,000(Negative)Not detected

BZO(Oxazepam) (Oxazepam, Cutoff=300ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
a-Hydroxyalprazolam2,50012%
Alprazolam120250%
Bromazepam1,50020%
Chlordiazepoxide1,50020%
Clobazam60500%
Clonazepam10,0003%
ClorazepateDipotassium200150%
Desalkylflurazepam1,50020%
Diazepam100300%
Estazolam150200%
Flunitrazepam160187.5%
D,L-Lorazepam1,56019.23%
Midazolam6,0005.00%
Nitrazepam300100%
Norchlordiazepoxide250120%
Nordiazepam40075%
Oxazepamglucuronide50060%
R,S-Lorazepamglucuronide160187.5%
Temazepam100300%
Triazolam2,50012%
Demoxepam2,00015%
Flurazepam50060%
Delorazepam100,000(Negative)Not detected

Page 15

COC(Benzoylecgonine) (Benzoylecgonine, Cutoff=150ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Cocaethylene100,000(Negative)Not detected
Cocaine hydrochloride3,0005%
Ecgonine100,000(Negative)Not detected
Norcocaine100,000(Negative)Not detected
Ecgonine methyl Ester100,000(Negative)Not detected

COC(Benzoylecgonine) (Benzoylecgonine, Cutoff=300ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Cocaethylene100,000(negative)Not detected
Cocaine hydrochloride5,0006%
Ecgonine100,000(negative)Not detected
Norcocaine100,0000.3%
Ecgonine methyl ester100,000(negative)Not detected

EDDP (EDDP, Cutoff=300ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Methadone100,000(negative)Not detected
EMDP100,000(negative)Not detected
Doxylamine100,000(negative)Not detected
Disopyramide100,000(negative)Not detected
LAAM (Levo-alpha-acetylmethadol)100,000(negative)Not detected
Alpha Methadol100,000(negative)Not detected

FYL(Fentanyl) (Fentanyl, Cutoff=1ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Norfentanyl3000.33%
Acetyl fentanyl520%
Acetyl Norfentanyl100,0000.001%
(±)-β-Hydroxythiofentanyl HCl520%
Acryl Fentanyl520%
Butyryl Fentanyl303.33%
Cis-d, I 3-Methylfentanyl25,0000.004%
Furanyl Fentanyl1010%
Para-fluorobutyrl Fentanyl (P-FBF)254%
Para-fluoroFentanyl254%
9-HydroxyRisperidone10,0000.01%
Alfentanil100,000(Negative)Not detected
Isobutyryl Fentanyl520%

Page 16

4-Fluoro-isobutyryl Fentanyl502%
Norcarfentail Oxalate100,000(Negative)Not detected
Remifentanil100,000(Negative)Not detected
Valeryl Fentanyl520%
ThienylFentnayl1001%
Trans-d, I 3-Methylfentanyl1001%
Despropionylfentanyl (4-ANPP)50,0000.002%
MT-45100,000(Negative)Not detected
Ocfentanil1001%
Risperidone50,0000.002%
Sufentanil50,0000.002%
Carfentanil Oxalate100,000(Negative)Not detected
LabetalolHydrochloride100,000(Negative)Not detected
Trazodone50,0000.002%
U-47700100,000(Negative)Not detected
ω-1-Hydroxyfentanyl50,0000.002%
6-Acetylmorphine100,000(Negative)Not detected
Amphetamine100,000(Negative)Not detected
Buprenorphine100,000(Negative)Not detected
Buprenorphine -3-D-Glucuronide100,000(Negative)Not detected
Codeine100,000(Negative)Not detected
Dextromethorphan100,000(Negative)Not detected
Dihydrocodeine100,000(Negative)Not detected
EDDP100,000(Negative)Not detected
EMDP100,000(Negative)Not detected
Fluoxetine100,000(Negative)Not detected
Heroin100,000(Negative)Not detected
Hydrocodone100,000(Negative)Not detected
Hydromorphone100,000(Negative)Not detected
Ketamine100,000(Negative)Not detected
Levorphanoltartrate100,000(Negative)Not detected
Meperidine100,000(Negative)Not detected
Methadone100,000(Negative)Not detected
Morphine100,000(Negative)Not detected
Morphine-3-β-D-glucuronide100,000(Negative)Not detected
Naloxone hydrochloride100,000(Negative)Not detected
Naltrexonehydrochloride100,000(Negative)Not detected
Norbuprenorphine100,000(Negative)Not detected
Norcodeine100,000(Negative)Not detected
Norketamine100,000(Negative)Not detected
Normeperidine100,000(Negative)Not detected
Normorphine100,000(Negative)Not detected
Noroxycodone100,000(Negative)Not detected

Page 17

Oxycodone100,000(Negative)Not detected
Oxymorphone100,000(Negative)Not detected
Pentazocine (Talwin)100,000(Negative)Not detected
Pipamperone100,000(Negative)Not detected
Tapentadolhydrochloride100,000(Negative)Not detected
Thioridazine100,000(Negative)Not detected
Tilidine100,000(Negative)Not detected
Tramadol100,000(Negative)Not detected
Tramadol-O-Desmethyl100,000(Negative)Not detected
Tramadol-N-Desmethyl100,000(Negative)Not detected

NFYL(Norfentanyl) (Norfentanyl, Cutoff=5ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Norfentanyl5100%
Fentanyl1,0000.5%
Acetyl fentanyl5001%
Acetyl Norfentanyl5010%
(±)-β-Hydroxythiofentanyl HCl1005%
Acryl Fentanyl1005%
Butyryl Fentanyl5001%
Cis-d, I 3-Methylfentanyl3001.67%
Furanyl Fentanyl1,0000.5%
Para-fluorobutyrl Fentanyl (P-FBF)1005%
Para-fluoroFentanyl1005%
9-HydroxyRisperidone10,0000.05%
Alfentanil100,000(Negative)Not detected
Isobutyryl Fentanyl806.25%
4-Fluoro-isobutyryl Fentanyl2002.5%
Norcarfentail Oxalate10,0000.05%
Remifentanil10,0000.05%
Valeryl Fentanyl5001%
ThienylFentnayl1005%
Trans-d, I 3-Methylfentanyl2002.50%
Despropionylfentanyl (4-ANPP)60,0000.008%
MT-458,0000.06%
Ocfentanil5,0000.1%
Risperidone6000.83%
Sufentanil10,0000.05%
Carfentanil Oxalate10,0000.05%
LabetalolHydrochloride100,000(Negative)Not detected
Trazodone60,0000.008%
U-47700100,000(Negative)Not detected

Page 18

ω-1-Hydroxyfentanyl60,0000.008%
6-Acetylmorphine100,000(Negative)Not detected
Amphetamine100,000(Negative)Not detected
Buprenorphine100,000(Negative)Not detected
Buprenorphine -3-D-Glucuronide100,000(Negative)Not detected
Codeine100,000(Negative)Not detected
Dextromethorphan100,000(Negative)Not detected
Dihydrocodeine100,000(Negative)Not detected
EDDP100,000(Negative)Not detected
EMDP100,000(Negative)Not detected
Fluoxetine100,000(Negative)Not detected
Heroin100,000(Negative)Not detected
Hydrocodone100,000(Negative)Not detected
Hydromorphone100,000(Negative)Not detected
Ketamine100,000(Negative)Not detected
Levorphanoltartrate100,000(Negative)Not detected
Meperidine100,000(Negative)Not detected
Methadone100,000(Negative)Not detected
Morphine100,000(Negative)Not detected
Morphine-3-β-D-glucuronide100,000(Negative)Not detected
Naloxone hydrochloride100,000(Negative)Not detected
Naltrexonehydrochloride100,000(Negative)Not detected
Norbuprenorphine100,000(Negative)Not detected
Norcodeine100,000(Negative)Not detected
Norketamine100,000(Negative)Not detected
Normeperidine100,000(Negative)Not detected
Normorphine100,000(Negative)Not detected
Noroxycodone100,000(Negative)Not detected
Oxycodone100,000(Negative)Not detected
Oxymorphone100,000(Negative)Not detected
Pentazocine (Talwin)100,000(Negative)Not detected
Pipamperone100,000(Negative)Not detected
Tapentadolhydrochloride100,000(Negative)Not detected
Thioridazine100,000(Negative)Not detected
Tilidine100,000(Negative)Not detected
Tramadol100,000(Negative)Not detected
Tramadol-O-Desmethyl100,000(Negative)Not detected
Tramadol-N-Desmethyl100,000(Negative)Not detected

MDMA(Ecstasy) (Ecstasy, Cutoff=500ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity

Page 19

(+/-)3,4-Methylenedioxy-nethylamphetamine(MDEA)1,000167%
(+/-)-Methylenedioxyamphetamine(MDA)8,0006.25%
L-Methamphetamine100,000(Negative)Not detected
7-Aminoclonazepam100,0000.5%
D-Methamphetamine100,000(Negative)Not detected
D-Amphetamine100,000(Negative)Not detected
L-amphetamine100,000(Negative)Not detected

MET(Methamphetamine) (Methamphetamine, Cutoff=500ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA)20,0002.5%
(±)-MDMA2,00025%
L-Methamphetamine5,00010%
Fenfluramine40,0001.25%
p-Hydroxymethamphetamine16,00010%
D,L-Methamphetamine500100%
β-Phenylethylamine25,0002%
Mephetermine25,0002%
Methoxyphenamine hydrochloride25,0002%
L-Amphetamine60,0000.8%
Ephedrine hydrochloride100,000(Negative)Not detected
(1R,2S)-(-)-Ephedrine100,0000.5%
D-Amphetamine100,000(Negative)Not detected
Chloroquine100,000(Negative)Not detected
(+/-)3,4- Methylenedioxyamphetamine(MDA)100,000(Negative)Not detected
L-Phenylephrine100,000(Negative)Not detected
Trimethobenzamide100,000(Negative)Not detected
Procainehydrochloride100,000(Negative)Not detected
d/l-Amphetamine100,000(Negative)Not detected

MET(Methamphetamine) (Methamphetamine, Cutoff=1000ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxy-n-40,0002.5%
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)8,00012.5%
L-Methamphetamine10,00010%
Fenfluramine100,000(Negative)Not detected
p-Hydroxymethamphetamine27,0003.7%
D,L-Methamphetamine1,000100%

Page 20

β-Phenylethylamine50,0002%
Mephetermine50,0002%
Methoxyphenamine hydrochloride50,0002%
L-Amphetamine100,000(Negative)Not detected
D-Amphetamine100,000(Negative)Not detected
D,L-Amphetamine100,000(Negative)Not detected
Chloroquine100,000(Negative)Not detected
Ephedrine hydrochloride100,000(Negative)Not detected
(+/-)3,4-Methylenedioxyamphetamine (MDA)100,000(Negative)Not detected
Trimethobenzamide100,000(Negative)Not detected
l-phenylephrine100,000(Negative)Not detected
(1R,2S)-(-)-Ephedrine100,000(Negative)Not detected
Procainehydrochloride100,000(Negative)Not detected

MOR(Morphine) (Morphine, Cutoff=300ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Codeine300100%
Dihydrocodeine1,00030%
EthylMorphine200150%
Heroin70042.86%
Hydrocodone2,00015%
Hydromorphone4,0007.5%
Levorphanol tartrate12,0002.5%
Nalorphinehydrochloride50,0000.6%
Thebaine90,0000.33%
Morphine-3-β-d-glucuronide1,00030%
6-Monoacetylmorphine (6-MAM)150200%
Codeine-6-β-D-glucuronide150200%
Morphine-6-β-D-glucuronide150200%
6-Acetylcodeine90033.33%
Normorphine100,000(negative)Not detected
Oxycodone100,000(negative)Not detected
Oxymorphone100,000(negative)Not detected
Norcodeine100,000(negative)Not detected
Procaine100,000(negative)Not detected
Norpropoxyphene100,000(negative)Not detected

OPI(Morphine) (Morphine, Cutoff=2000ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Codeine1,000200%
Dihydrocodeine1,500133.3%
EthylMorphine500400%

Page 21

Codeine-6-β-D-glucuronide2,50080%
Heroin5,00040%
Hydrocodone5,00040%
Hydromorphone7,00028.6%
Levorphanol tartrate16,00012.5%
Nalorphinehydrochloride5,00004%
Norcodeine100,000(Negative)Not detected
Normorphine100,000(Negative)Not detected
Oxymorphone100,000(Negative)Not detected
s-Monoacetylmorphine2,000100%
Morphine-6-β-D-glucuronide1,300153.8%
Thebaine100,000(Negative)Not detected
Morphine 3-β-D-glucuronide1,300153.8%
6-Monoacetylmorphine (6-MAM)1,000200%
6-Acetylcodeine10,00020%
Oxycodone100,000(Negative)Not detected
Procaine100,000(Negative)Not detected
Norpropoxyphene100,000(Negative)Not detected

MTD(Methadone) (Methadone, Cutoff=300ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
EDDP100,000(negative)Not detected
EMDP100,000(negative)Not detected
LAAM100,000(negative)Not detected
Alpha Methadol100,000(negative)Not detected
Doxylamine100,000(negative)Not detected

OXY(Oxycodone) (Oxycodone, Cutoff=100ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Ethyl Oxycodone75,0000.13%
Hydrocodone10,0001%
Hydromorphone100,000(negative)Not detected
Levorphanol tartrate25,0000.40%
Naloxone hydrochloride10,0001%
Naltrexone hydrochloride50,0000.2%
Oxymorphone20050%
Oxymorphone-3β-D-glucuronide100,000(negative)Not detected
Noroxycodone1,5006.67%
Noroxymorphone3,0003.33%
Dihydrocodeine100,000(negative)Not detected
Codeine100,000(negative)Not detected
Morphine100,000(negative)Not detected

Page 22

Buprenorphine100,000(negative)Not detected
Ethylmorphine100,000(negative)Not detected
Thebaine100,000(negative)Not detected
6-acetylmorphine100,000(negative)Not detected

PCP(Phencyclidine) (Phencyclidine, Cutoff=25ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
4-Hydroxyphencyclidine12,5000.2%

PPX(Propoxyphene) (Propoxyphene, Cutoff=300ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Norpropoxyphene1,50020%

TCA(Nortriptyline) (Nortriptyline, Cutoff=1000ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Amitriptyline5,00020%
Chlorpheniramine50,0002%
Clomipramine10,00010%
Cyclobenzaprine Hydrochloride5,00020%
Desipramine1,000100%
Doxepine2,00050%
Duloxetine10,00010%
Imipramine1,000100%
Norclomipramine100,000(negative)Not detected
Nordoxepine1,000100%
Promazine50,0002%
Trimipramine10,00010%
Maprotiline100,000(negative)Not detected
Promethazine100,000(negative)Not detected

THC(Marijuana) (Marijuana, Cutoff=50ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
11-nor-Δ8-THC -9-COOH30166.67%
(-)-11-nor-9-carboxy-Δ9-THC50100%
(±)-11-nor-9-Carboxy-Δ9-THC50100%
11-nor-Δ9-THC -carboxyglucuronide10050%
11-hydroxy-Δ9-Tetrahydrocannabinol15,0000.33%
Δ8- Tetrahydrocannabinol8,0000.63%
Δ9- Tetrahydrocannabinol7,0000.71%
Cannabinol100,000(negative)Not detected
Cannabidiol100,000(negative)Not detected

Page 23

(±)-11-Hydroxy-Δ9-THC100,000(negative)Not detected

TML(Tramadol) (Tramadol, Cutoff=100ng/mL)

Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
n-Desmethyl Tramadol40025%
o-Desmethyl Tramadol1,00010%
o-Desmethyl Venlafaxine100,000(negative)Not detected
Venlafaxine HCl100,000(negative)Not detected

To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.

Compounds that show no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following table.

11-nor-9 carboxy THC (except for THC)DiphenylhydantoinNorpropoxyphene (except for PPX)
3-HydroxytyramineDisopyramideNorpseudoephedrine
4-Bromo2,5,DimethoxyphenethylamineDL-TryptophanNortriptyline (except for TCA)
7-Aminoclonazepam (except for MDMA)DL-TyrosineNoscapine
7-AminoflunitrazepamDopamineO-Hydroxyhippuric acid
7-AminonitrazepamDoxepin (except for TCA)Octopamine
AcetaminophenDoxylamineOlanzapine
Acetone (1000 mg/dL)Duloxetine (except for TCA)Omeprazole
AcetophenetidinEcgonine methyl esterOxalic Acid (100mg/dL)
Acetylsalicylic Acid (500 μg/mL)EphedrineOxalinic acid
AcyclovirErythromycinOxazepam (except for BZO)
Albumin(100 mg/dL)Esomeprazole MagnesiumOxolinic acid
AlbuterolEstradiolOxymetazoline
Alpha MethadolEstronePaliperidone
AminophyllineEthanol (1%)Papaverine
AminopyrineFenfluramine (except for MET 500)Penicillin-G
Amitriptyline (except for TCA)FenofibratePenicillinV Potassium
Amlodipine besylateFenoprofenPentazocine
Amobarbital (except for TCA)Fluoxetine HydrochloridePerphenazine
AmoxicillinFluphenazinePhencyclidine (except for PCP)
AmpicillinFotemustinePhenelzine
ApomorphineFurosemidePhenobarbital (except for BAR)
AripiprazoleGabapentinPhentermine (except for AMP)
Ascorbic AcidGalactosePhenylethylamine (except for AMP, MET)
AspartameGamma Globulin (500mg/dL)Phenylpropanolamine
AspirinGatifloxacinPrednisone
AtomoxetineGemfibrozilPregablin
Atorvastatin CalciumGentisic acidProcaine
AtropineGlucose (3000 mg/dL)Promazine (except for TCA)
AzithromycinGuaiacolglyceryl etherPromethazine
BaclofenHemoglobinPropoxyphene (except for PPX)
Benzilic acidHexobarbitalPropranolol
BenzocaineHydralazinePseudoephedrine
Benzoic AcidHydrochlorothiazidePyridoxine
Benzoylecgonine (except for COC)HydrocortisonePyrilamine
BenzphetamineHydroxytyraminePyrogallol

Page 24

BenzylpiperiazineI-CotinineQuetiapine
BilirubinI-ErythromycinQuinidine
Boric Acid (1%)IbuprofenQuinine
BupropionImipramine (except for TCA)Quinolinic Acid
Caffeine (500 μg/mL)Isoproterenolr-Globulin
CannabidiolIsoxsuprineRanitidine
CaptoprilKetamineRiboflavin
CarbamazepineKetoprofenRifampicin
Carfentanil (except for NFYL)L-EpinephrineSalicylic Acid
CarisoprodolL-phenylephrineSecobarbital (except for BAR)
CefradineL-thyroxineSerotonin(5-Hydroxytyramine)
CephalexinLAAM HClSertraline
CetirizineLabetalolSildenafil Citrate
Chloral hydrateLamotrigineSimvastatin
ChloramphenicolLevofloxacin HydrochlorideSodium Azide
Chlordiazepoxide (except for BZO)LevonorgestrelSufentanil Citrate (except for FYL, NFYL)
ChloroquineLevothyroxine SodiumSulfamethazine
ChlorothiazideLidocaineSulindac
Chlorpheniramine (except for TCA)LisinoprilTelmisartan
ChlorpromazineLoperamideTetracycline
CholesterolLoratadineTetrahydrocortisone 3-acetate
Ciprofloxacin HydrochlorideLoratidineTetrahydrocortisone3-(βDglucuronide)
CitalopramLorazepam (except for BZO)Tetrahydrozoline
ClarithromycinLSDTHC (except for THC)
ClofibrateMagnesiumTheophylline
Clomipramine (except for TCA)MaprotilineThiamine
ClonidineMeperidineThioridazine
ClozapineMeprobamateTrazodone Hydrochloride (except for FYL, NFYL)
Conjugated EstrogensMetforminTriamterene
CortisoneMethapyrileneTrifluoperazine
CotinineMethaqualoneTrifluoromethylphenyl- piperazine
Creatine HydrateMethoxyphenamine (except for MET)Trimethobenzamide
CreatinineMethylphenidateTrimethoprim
Cyclobenzaprine (except for TCA)MetoprololTryptamine
Cyclodextrin-rMetronidazole (300 ug/mL)Tyramine (except for AMP)
CyproheptadineMifepristoneUrea (2000 mg/dL)
D,L-EpinephrineN-AcetylprocainamideUric Acid
d,l-IsoproterenolN-desmethyl TapentadolValproic acid (250 ug/mL)
DelorazepamNaCl (40000 ug/mL)Venlafaxine
Demoxepam (except for BZO)Nalidixic AcidVerapamil
DeoxycorticosteroneNaloxone (except for OXY)Vitamin B2
Desipramine (except for TCA)Naltrexone (except for OXY)Zaleplon
DesloratadineNaproxenZolpidem
DextromethorphanNiacinamideZolpidem Tartrate
DiclofenacNicotineZomepirac
DiflunisalNicotinic Acidβ-Estradiol
DigoxinNifedipineβ-Hydroxybutyric Acid
Dimethyl-aminoantipyrineNitroglycerin
DiphenhydramineNorethindrone

Interference by pH and specific gravity were also evaluated using pooled urine specimens with concentrations of 0 (drug-free), at 50% below and 50% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.

B. Method comparison study

Page 25

The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below.

Drug testTest Cup ResultDrug FreeLow Negative by LC-MS/MS (less than -50%)Near Cutoff Negative by LC-MS/MS (Between -50% and the Cutoff)Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%)High Positive by LCMS/MS (greater than +50%)
6-MAM10Operator A- 10161320
+ 0011919
Operator B- 10161310
+ 0012019
Operator C- 10161420
+ 0001919
AMP500Operator A- 10181110
+ 0011524
Operator B- 10181110
+ 0011524
Operator C- 10181110
+ 0011524
AMP1000Operator A- 10141410
+ 0021722
Operator B- 10141510
+ 0011722
Operator C- 10141400
+ 0021822
BAR300Operator A- 10161420
+ 0001721
Operator B- 10161310
+ 0011821
Operator C- 10161420
+ 0001721
BUP10Operator A- 10161420
+ 0001424
Operator B- 10161310
+ 0011524
Operator C- 10161310
+ 0011524
BZO300Operator- 10191010

Page 26

A+0011524
Operator B-10191010
+0011524
Operator C-10191010
+0011524
COC150Operator A-1015142
+0011523
Operator B-10151410
+0011623
Operator C-10151410
+0011623
COC300Operator A-1016131
+0012118
Operator B-10161320
+0012018
Operator C-10161200
+0022218
EDDP300Operator A-1013161
+0011524
Operator B-10131610
+0011524
Operator C-10131710
+0001524
FYL1Operator A-1014140
+0021822
Operator B-10141410
+0021722
Operator C-10141410
+0021722
NFYL5Operator A-1016132
+0011424
Operator B-10161210
+0021524
Operator C-10161210
+0021524
MDMA500Operator A-1016131
+0011821
Operator B-10161310
+0011821
Operator C-10161310
+0011821
MET500Operator-1016131

Page 27

A+0011524
Operator B-10161210
+0021524
Operator C-10161310
+0011524
MET1000Operator A-1016131
+0011821
Operator B-10161320
+0011721
Operator C-10161310
+0011821
MTD300Operator A-1016131
+0011821
Operator B-10161310
+0011821
Operator C-10161430
+0001621
MOP300Operator A-1013171
+0002019
Operator B-10131620
+0011919
Operator C-10131600
+0012119
OPI2000Operator A-1016110
+0032119
Operator B-10161310
+0012019
Operator C-10161310
+0012019
OXY100Operator A-1015151
+0001722
Operator B-10151310
+0021722
Operator C-10151400
+0011822
PCP25Operator A-1016142
+0001622
Operator B-10161410
+0001722
Operator C-10161300
+0011822
PPX300Operator-1017132

Page 28

A+0001721
Operator B-10171310
+0001821
Operator C-10171210
+0011821
TCA1000Operator A-1017121
+0011623
Operator B-10171220
+0011523
Operator C-10171320
+0001523
THC50Operator A-1015141
+0012019
Operator B-10151410
+0012019
Operator C-10151520
+0001919
TML100Operator A-1016132
+0011622
Operator B-10161300
+0011822
Operator C-10161310
+0011722

Discordant Results are summarized below.

DrugOperatorSample NumberLC/MS/MS Result (ng/mL)Accurate Result
6AMOperator A202502122511.69-
Operator C202411292810.03-
Operator A, C202501082910.37-
Operator B202502281910.14-
Operator A, B20241219289.62+
AMP500Operator A2024072222520.38-
Operator B2024092628506.41-
Operator C2024081607512.57-
Operator A2024120516494.96+
Operator B2024073117473.49+
Operator C2025011427443.03+
AMP1000Operator A20250224061018.68-
Operator B20240827231029.82-
Operator B, C2024082624984.60+
Operator A2024071704989.98+

Page 29

Operator A, C2025021122875.36+
BAR300Operator A2024071518311.29
Operator A2025012211333.75
Operator B2024092411344.57
Operator C2024112222323.25
Operator C2025012217325.28
Operator B2024070127294.61
BUP10Operator A202407232712.38
Operator A202410242010.17
Operator B202502282810.15
Operator C202501172110.01
Operator B20241108289.85
Operator C20240812169.36
BZO300Operator A2025012103305.56
Operator B2024110729301.66
Operator C2024070128311.53
Operator A, C2024110118264.48
Operator B2024093005278.91
COC150Operator A2024112124152.97
Operator B2024123126151.87
Operator A, C2024071513150.11
Operator A, B2024093011149.17
Operator C2025021221147.39
COC300Operator A,B2024101021300.47
Operator B2024082311304.87
Operator C2025010603296.21
Operator A,B,C2024082222299.37
EDDP300Operator A2025012323303.33
Operator B, C2024071623308.15
Operator A2025010618290.68
Operator B2024103023298.10
FYL1Operator B20240926301.03
Operator C20241114111.01
Operator A, C20250108030.99
Operator A20241115010.94
Operator B, C20240709140.89
Operator B20240805090.99
NFYL5Operator A20240821215.93
Operator B,C20240902215.24
Operator A20241203115.09
Operator C20240712094.12
Operator C20240712163.99
Operator B20250228214.11

Page 30

Operator A,B20250218264.62+
MDMA500Operator A2024122504512.65
Operator B2024082902505.45
Operator C2024091212501.55
Operator A2025012429498.12
Operator B2024080521488.93
Operator C2024122412469.24
MET500OperatorA2024121328516.28
Operator B2024111330500.12
Operator C2024081506506.10
Operator A, B2024120413496.93
Operator B2025012023485.93
Operator C2024070214463.37
MET1000Operator B, C20240722011142.94
Operator A20240724211000.87
Operator B20241114141014.70
Operator C2024070803980.34
Operator B2024071613998.49
Operator A2024092417998.48
MTD300Operator A2024070110312.01
Operator C2024090327338.62
Operator B, C2024110616302.80
Operator C2025010714308.66
Operator B2024101503241.25
Operator A2024121005298.81
MOP300Operator A, B2024083003306.99
Operator B2024090405359.23
Operator C2024120925241.77
Operator B2025011722298.07
OPI2000Operator B,C20240926212032.06
Operator A20240814111946.49
Operator A,C20250221181992.26
Operator A,B20240717021890.60
OXY100Operator B2025021017118.39
Operator A2024112817101.02
Operator B202411291793.38
Operator B,C202501090398.65
PCP25Operator A, B202408220325.61
Operator A202412200926.88
Operator C202409190924.51
PPX300Operator A2024102227319.65
Operator B,C2024082228306.89
Operator A2024103108305.09

Page 31

Operator C2024091310298.22+
TCA100Operator B, C20240813281166.52
Operator B20240920231005.22
Operator A20240716251020.54
Operator C20241118221023.53
Operator B2024090426986.95
Operator A2024070321991.10
THC50Operator C202408192850.10
Operator A202501022350.75
Operator B,C202407022351.39
Operator A202407092948.00
Operator B202411152349.29
TML100Operator A2024080113100.64
Operator C2024073110101.15
Operator A2025021106113.40
Operator A202408142395.20
Operator C202407181699.76
Operator B202411200299.32

C. Lay person study

A lay user study was performed at three intended user sites with 280 lay persons. 158 male and 122 female tested the Webest Multi-Drug Urine Cup. They had diverse educational and professional backgrounds and their age range from 20 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free pooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

Result of The Webest Multi-Drug Urine Cup: Configuration 1

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| AMP 500 | | Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 |
| | | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% |
| BAR 300 | | Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 |
| | | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% |
| BUP 10 | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |

Page 32

Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
BZO 300Positive0000182020
Negative20202020200
Total20202020202020
Agreement (%)100%100%100%100%90%100%100%
COC 150Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
EDDP 300Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
FYL 1Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
MDMA 500Positive0002202020
Negative20202018000
Total20202020202020
Agreement (%)100%100%100%90%100%100%100%
MET 500Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
MTD 300Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
6-MAM 10Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
NFYL 5Positive0002202020
Negative20202018000
Total20202020202020
Agreement (%)100%100%100%90%100%100%100%
MOP 300Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
OXY 100Positive0002192020

Page 33

Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
PCP 25Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
PPX 300Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
TCA 1000Positive0000182020
Negative20202020200
Total20202020202020
Agreement (%)100%100%100%100%90%100%100%
THC 50Positive0001182020
Negative20202019200
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
TML 100Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%

Result of the Webest Multi-Drug Urine Cup: Configuration 2

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| AMP 1000 | | Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 |
| | | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | 100% |
| BAR 300 | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% |
| BUP 10 | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| BZO 300 | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |

Page 34

Agreement (%)100%100%100%100%95%100%100%
COC 300Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
EDDP 300Positive0001202020
Negative20202019000
Total20202020202020
Agreement (%)100%100%100%95%100%100%100%
FYL 1Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
MDMA 500Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
MET 1000Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
MTD 300Positive0000192020
Negative20202020100
Total20202020202020
Agreement (%)100%100%100%100%95%100%100%
6-MAM 10Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
NFYL 5Positive0001182020
Negative20202019200
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
OPI 2000Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
OXY 100Positive0001182020
Negative20202019200
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
PCP 25Positive0001192020
Negative20202019100

Page 35

Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
PPX 300Positive0002192020
Negative20202018100
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
TCA 1000Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
THC 50Positive0001202020
Negative20202019000
Total20202020202020
Agreement (%)100%100%100%95%100%100%100%
TML 100Positive0001192020
Negative20202019100
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%

Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

Clinical Studies:
Not applicable.

  1. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that The Webest Multi-Drug Urine Cup and The Webest Home Multi-Drug Urine Cup are substantially equivalent to the predicate device.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).