K Number
K252550

Validate with FDA (Live)

Date Cleared
2025-11-18

(97 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SAFElife™ T-Cup Multi-Drug Urine Test Cup:

SAFElife™ T-Cup Multi-Drug Urine Test Cup is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine at the cutoff concentrations of:

Drug (Identifier)Cutoff Level
6-Monoacetylmorphine (6-MAM)10 ng/mL
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Fentanyl (FTY)1 ng/mL
Methamphetamine (MET/mAMP)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Opiates (OPI)2000 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFTY)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC)50 ng/mL or 20 ng/mL
Tramadol (TRA)100 ng/mL

SAFElife™ T-Cup Multi-Drug Urine Test Cup offers any combinations from 1 to 19 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx:

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methylenedioxymethamphetamine, Methamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine with below cutoff
concentrations and approximate detection time:

Drug (Identifier)CalibratorCut-off Level
6-Monoacetylmorphine (6-MAM)6-Monoacetylmorphine10 ng/mL
Amphetamine (AMP500)d-Amphetamine500 ng/mL
Amphetamine (AMP1000)d-Amphetamine1000 ng/mL
Secobarbital (BAR)Secobarbital300 ng/mL
Buprenorphine (BUP)Buprenorphine10 ng/mL
Oxazepam (BZO)Oxazepam300 ng/mL
Cocaine (COC150)Benzoylecgonine150 ng/mL
Cocaine (COC300)Benzoylecgonine300 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1,5-dimethyl-3,3-diphenyl-pyrrolidine300 ng/mL
Fentanyl (FTY)Fentanyl1 ng/mL
Methylenedioxymethamphetamine (MDMA)3,4-Methylenedioxymethamphetamine500 ng/mL
Methamphetamine (MET500/mAMP500)D(+)-Methamphetamine500 ng/mL
Methamphetamine (MET1000/mAMP1000)D(+)-Methamphetamine1000 ng/mL
Morphine (MOP300)Morphine300 ng/mL
Opiates (OPI2000)Morphine2000 ng/mL
Methadone (MTD)Methadone300 ng/mL
Norfentanyl (NFTY)Norfentanyl5 ng/mL
Oxycodone (OXY)Oxycodone100 ng/mL
Phencyclidine (PCP)Phencyclidine25 ng/mL
Propoxyphene (PPX)d-Propoxyphene300 ng/mL
Nortriptyline (TCA)Nortriptyline1000 ng/mL
Cannabinoids (THC20)11-nor-Δ9-THC-9-COOH20 ng/mL
Cannabinoids (THC50)11-nor-Δ9-THC-9-COOH50 ng/mL
Tramadol (TRA)Tramadol100 ng/mL

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx offers any combinations from 1 to 19 drugs of abuse tests with or without on-board adulteration/specimen validity test (SVT) but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

Device Description

SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids, 6-Monoacetylmorphine, Fentanyl, Norfentanyl, and Tramadol in human urine. Each SAFElife™ T-Cup Multi-Drug Urine Test Cup or SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx device consists of a test cup and a package insert. Each test cup is sealed with one sachet of desiccant in an aluminum pouch. SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx have a general identical design. The SAFElife™ T-Cup Multi-Drug Urine Test Cup is for over-the-counter (OTC) use and the SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is for professional use.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - SAFElife T-Cup Multi-Drug Urine Test Cup

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

November 18, 2025

Guangzhou Wondfo Biotech Co., Ltd.
Kaiyu Xiao
Senior Regulatory Affairs Manager
No.8 Lizhishan Road, Science City, Huangpu District
Guangzhou, 510663
China

Re: K252550
Trade/Device Name: SAFElife T-Cup Multi-Drug Urine Test Cup; SAFElife T-Cup Multi-Drug Urine Test Cup Dx
Regulation Number: 21 CFR 862.3100
Regulation Name: Amphetamine Test System
Regulatory Class: Class II
Product Code: NFT, NGL, PTH, NFV, NFY, PTG, NGG, LCM, QBF, QAW, NFW
Dated: August 13, 2025
Received: October 14, 2025

Dear Kaiyu Xiao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K252550 - Kaiyu Xiao Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252550 - Kaiyu Xiao Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2025.11.18 08:22:57 -05'00'

Joseph Kotarek
Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 3

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K252550

Device Name: SAFElife T-Cup Multi-Drug Urine Test Cup; SAFElife T-Cup Multi-Drug Urine Test Cup Dx

Indications for Use (Describe)

SAFElife™ T-Cup Multi-Drug Urine Test Cup:

SAFElife™ T-Cup Multi-Drug Urine Test Cup is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine at the cutoff concentrations of:

Drug (Identifier)Cutoff Level
6-Monoacetylmorphine (6-MAM)10 ng/mL
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Fentanyl (FTY)1 ng/mL
Methamphetamine (MET/mAMP)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Opiates (OPI)2000 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFTY)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC)50 ng/mL or 20 ng/mL
Tramadol (TRA)100 ng/mL

SAFElife™ T-Cup Multi-Drug Urine Test Cup offers any combinations from 1 to 19 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx:

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methylenedioxymethamphetamine, Methamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine with below cutoff

Page 5

FORM FDA 3881 (8/23) Page 2 of 3

concentrations and approximate detection time:

Drug (Identifier)CalibratorCut-off Level
6-Monoacetylmorphine (6-MAM)6-Monoacetylmorphine10 ng/mL
Amphetamine (AMP500)d-Amphetamine500 ng/mL
Amphetamine (AMP1000)d-Amphetamine1000 ng/mL
Secobarbital (BAR)Secobarbital300 ng/mL
Buprenorphine (BUP)Buprenorphine10 ng/mL
Oxazepam (BZO)Oxazepam300 ng/mL
Cocaine (COC150)Benzoylecgonine150 ng/mL
Cocaine (COC300)Benzoylecgonine300 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1,5-dimethyl-3,3-diphenyl-pyrrolidine300 ng/mL
Fentanyl (FTY)Fentanyl1 ng/mL
Methylenedioxymethamphetamine (MDMA)3,4-Methylenedioxymethamphetamine500 ng/mL
Methamphetamine (MET500/mAMP500)D(+)-Methamphetamine500 ng/mL
Methamphetamine (MET1000/mAMP1000)D(+)-Methamphetamine1000 ng/mL
Morphine (MOP300)Morphine300 ng/mL
Opiates (OPI2000)Morphine2000 ng/mL
Methadone (MTD)Methadone300 ng/mL
Norfentanyl (NFTY)Norfentanyl5 ng/mL
Oxycodone (OXY)Oxycodone100 ng/mL
Phencyclidine (PCP)Phencyclidine25 ng/mL
Propoxyphene (PPX)d-Propoxyphene300 ng/mL
Nortriptyline (TCA)Nortriptyline1000 ng/mL
Cannabinoids (THC20)11-nor-Δ9-THC-9-COOH20 ng/mL
Cannabinoids (THC50)11-nor-Δ9-THC-9-COOH50 ng/mL
Tramadol (TRA)Tramadol100 ng/mL

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx offers any combinations from 1 to 19 drugs of abuse tests with or without on-board adulteration/specimen validity test (SVT) but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

Page 6

FORM FDA 3881 (8/23) Page 3 of 3

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 7

510(k) SUMMARY

K252550

1. Date:

November 13, 2025

2. Submitter:

Guangzhou Wondfo Biotech Co., Ltd.
No.8 Lizhishan Road, Science City, Huangpu District
Guangzhou, China

3. Contact Person

Kaiyu Xiao
Senior Regulatory Affairs Manager
Tel: +86-15005196892
E-mail: kaiyu.xiao@wondfo.com.cn

4. Device Name

SAFElife™ T-Cup Multi-Drug Urine Test Cup
SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx

5. Classification

Class II

Product Code & Target DrugRegulation SectionPanel
NFT - Amphetamine (AMP)862.3100, Amphetamine Test SystemToxicology (91)
NGL - Buprenorphine (BUP), Morphine (MOP), Opiates (OPI), Oxycodone (OXY), Fentanyl (FTY), Norfentanyl (NFTY), Tramadol (TRA), 6-Monoacetylmorphine(6-MAM)862.3650, Opiate Test SystemToxicology (91)
PTH - Secobarbital (BAR)862.3150, Barbiturate Test SystemToxicology (91)
NFV - Oxazepam (BZO)862.3170, Benzodiazepine Test SystemToxicology (91)
NFY - Cocaine (COC)862.3250, Cocaine Test SystemToxicology (91)
PTG - 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Methadone (MTD)862.3620, Methadone Test SystemToxicology (91)

Page 8

Product Code & Target DrugRegulation SectionPanel
NGG - Methamphetamine (MET), Methylenedioxymethamphetamine (MDMA)862.3610, Methamphetamine Test SystemToxicology (91)
LCM - Phencyclidine (PCP)UnclassifiedToxicology (91)
QBF - Propoxyphene (PPX)862.3700, Propoxyphene test system.Toxicology (91)
QAW - Nortriptyline (TCA)862.3910, Tricyclic antidepressant drugs test systemToxicology (91)
NFW - Cannabinoids (THC)862.3870, Cannabinoids Test SystemToxicology (91)

6. Predicate Device

Product name: Wondfo T-Cup® Multi-Drug Urine Test Cup
510(k) Number: K182701

7. Intended Use

SAFElife™ T-Cup Multi-Drug Urine Test Cup

SAFElife™ T-Cup Multi-Drug Urine Test Cup is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine at the cutoff concentrations of:

Drug (Identifier)Cutoff Level
6-Monoacetylmorphine (6-MAM)10 ng/mL
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Fentanyl (FTY)1 ng/mL
Methamphetamine (MET/mAMP)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Opiates (OPI)2000 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFTY)5 ng/mL

Page 9

Drug (Identifier)Cutoff Level
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC)50 ng/mL or 20 ng/mL
Tramadol (TRA)100 ng/mL

SAFElife™ T-Cup Multi-Drug Urine Test Cup offers any combinations from 1 to 19 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is a competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of 6-Monoacetylmorphine, Amphetamine, Secobarbital, Buprenorphine, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Fentanyl, Methylenedioxymethamphetamine, Methamphetamine, Morphine, Opiates, Methadone, Norfentanyl, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids and Tramadol in human urine with below cutoff concentrations and approximate detection time:

Drug (Identifier)CalibratorCut-off Level
6-Monoacetylmorphine (6-MAM)6-Monoacetylmorphine10 ng/mL
Amphetamine (AMP500)d-Amphetamine500 ng/mL
Amphetamine (AMP1000)d-Amphetamine1000 ng/mL
Secobarbital (BAR)Secobarbital300 ng/mL
Buprenorphine (BUP)Buprenorphine10 ng/mL
Oxazepam (BZO)Oxazepam300 ng/mL
Cocaine (COC150)Benzoylecgonine150 ng/mL
Cocaine (COC300)Benzoylecgonine300 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1,5-dimethyl-3,3-diphenyl-pyrrolidine300 ng/mL
Fentanyl (FTY)Fentanyl1 ng/mL
Methylenedioxymethamphetamine (MDMA)3,4-Methylenedioxymethamphetamine500 ng/mL

Page 10

Drug (Identifier)CalibratorCut-off Level
Methamphetamine (MET500/mAMP500)D(+)-Methamphetamine500 ng/mL
Methamphetamine (MET1000/mAMP1000)D(+)-Methamphetamine1000 ng/mL
Morphine (MOP300)Morphine300 ng/mL
Opiates (OPI2000)Morphine2000 ng/mL
Methadone (MTD)Methadone300 ng/mL
Norfentanyl (NFTY)Norfentanyl5 ng/mL
Oxycodone (OXY)Oxycodone100 ng/mL
Phencyclidine (PCP)Phencyclidine25 ng/mL
Propoxyphene (PPX)d-Propoxyphene300 ng/mL
Nortriptyline (TCA)Nortriptyline1000 ng/mL
Cannabinoids (THC20)11-nor-Δ9-THC-9-COOH20 ng/mL
Cannabinoids (THC50)11-nor-Δ9-THC-9-COOH50 ng/mL
Tramadol (TRA)Tramadol100 ng/mL

SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx offers any combinations from 1 to 19 drugs of abuse tests with or without on-board adulteration/specimen validity test (SVT) but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use.

The tests may yield positive results for the prescription drugs Buprenorphine, Fentanyl, Nortriptyline, Oxazepam, Secobarbital, Oxycodone and Tramadol when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) is the recommended confirmatory method.

8. Device Description

SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Opiates, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline, Cannabinoids, 6-Monoacetylmorphine, Fentanyl, Norfentanyl, and Tramadol in human urine. Each SAFElife™ T-Cup Multi-Drug Urine Test Cup or SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx device consists of a test cup and a package insert. Each test cup is sealed with one sachet of desiccant in an aluminum pouch. SAFElife™ T-Cup Multi-Drug Urine Test Cup and SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx have a general identical design. The SAFElife™ T-Cup Multi-Drug Urine Test Cup is for

Page 11

over-the-counter (OTC) use and the SAFElife™ T-Cup Multi-Drug Urine Test Cup Dx is for professional use.

9. Comparison with Predicate

A summary comparison of features of the candidate device with the predicate device is provided in following table.

Table 1: Features Comparison of SAFElife™ T-Cup Multi-Drug Urine Test Cup and the predicate device

ItemCandidate DevicePredicate - K182701
Device Trade NameSAFElife™ T-Cup Multi-Drug Urine Test CupSAFElife™ T-Cup Multi-Drug Urine Test Cup DxWondfo T-Cup® Multi-Drug Urine Test Cup

General Device Similarities

ItemCandidate DevicePredicate - K182701
Indications For UseDrugs of abuse testing in human urineSame
MethodologyCompetitive binding, lateral flow immunochromatographic assay based on antigen-antibody reactionSame
Type of TestQualitativeSame
Specimen TypeHuman urineSame

General Device Characteristic Differences

Analytes and CutoffTarget DrugsCutoffSame
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET/mAMP)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP)300 ng/mL
Opiates (OPI)2000 ng/mL
Methadone (MTD)300 ng/mL

Page 12

ItemCandidate DevicePredicate - K182701
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC)50 ng/mL
6-Monoacetylmorphine (6-MAM)10 ng/mLNot included in K182701
Fentanyl (FTY)1 ng/mL
Norfentanyl (NFTY)5 ng/mL
Cannabinoids (THC)20 ng/mL
Tramadol (TRA)100 ng/mL

10. Test Principle

SAFElife T-Cup Multi-Drug Urine Test Cup is a lateral flow chromatographic immunoassay. When urine sample is added to the cup device, urine is absorbed into the test strip and migrates upwards by capillary action. If the concentration of target drug presented in the urine sample is below the cutoff level, the target drug will not saturate the binding sites of its specific monoclonal antibody-coated particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored band will be formed on the test line region. If the concentration of target is beyond the cutoff level, the target drug will saturate the binding sites of its specific monoclonal antibody-particles, thus the antibody-coated particles will not be captured by immobilized drug-conjugate hence no colored band will be formed on the test line region. A band should be formed on the control line region regardless of the presence of target drug or metabolite in the sample to indicate that the tests have been performed properly.

11. Performance Characteristics

1. Analytical Performance

Performance data of AMP500/1000, BAR300, BUP10, BZO300, COC150/300, EDDP300, MDMA500, MET500/MET1000, MOP300, MTD300, OPI2000, OXY100, PCP25, PPX300, TCA1000 and THC50 were presented in the 510(k)-cleared submission K182701. In this submission, 6-Monoacetylmorphine (6-MAM10), Fentanyl (FTY1), Norfentanyl (NFTY5), Cannabinoids (THC20), and Tramadol (TRA100) study data are summarized and presented below.

a. Precision/Reproducibility

Precision studies were carried out for samples with concentrations of -100% cutoff, -75% cut off, -50% cutoff, -25% cutoff, cutoff, +25% cutoff, +50% cutoff, +75% cutoff and +100% cutoff. Samples with concentration of -100% cutoff were drug-free urine samples. Other samples were prepared by spiking target drug in drug-free urine samples. Each drug concentration was confirmed by LC/MS or GC/MS. For each concentration, tests were

Page 13

performed two runs per day for 25 days using three lots of test cups. The results of a representative lot are summarized in the following tables:

DrugLot Number-100% Cutoff-75% Cutoff-50% Cutoff-25% CutoffCutoff+25% Cutoff+50% Cutoff+75% Cutoff+100% Cutoff
6-MAM 10Lot 150-/0+50-/0+50-/0+50-/0+25-/25+1-/49+0-/50+0-/50+0-/50+
Lot 250-/0+50-/0+50-/0+48-/2+25-/25+1-/49+0-/50+0-/50+0-/50+
Lot 350-/0+50-/0+50-/0+49-/1+26-/24+2-/48+0-/50+0-/50+0-/50+
FTY1Lot 150-/0+50-/0+50-/0+48-/2+23-/27+0-/50+0-/50+0-/50+0-/50+
Lot 250-/0+50-/0+50-/0+48-/2+24-/26+0-/50+0-/50+0-/50+0-/50+
Lot 350-/0+50-/0+50-/0+48-/2+25-/25+0-/50+0-/50+0-/50+0-/50+
NFTY5Lot 150-/0+50-/0+50-/0+49-/1+24-/26+2-/48+0-/50+0-/50+0-/50+
Lot 250-/0+50-/0+50-/0+49-/1+26-/24+1-/49+0-/50+0-/50+0-/50+
Lot 350-/0+50-/0+50-/0+48-/2+24-/26+1-/49+0-/50+0-/50+0-/50+
THC20Lot 150-/0+50-/0+50-/0+48-/2+26-/24+1-/49+0-/50+0-/50+0-/50+
Lot 250-/0+50-/0+50-/0+49-/1+25-/25+0-/50+0-/50+0-/50+0-/50+
Lot 350-/0+50-/0+50-/0+48-/2+23-/27+2-/48+0-/50+0-/50+0-/50+
TRA100Lot 150-/0+50-/0+50-/0+50-/0+24-/26+0-/50+0-/50+0-/50+0-/50+
Lot 250-/0+50-/0+50-/0+50-/0+27-/23+0-/50+0-/50+0-/50+0-/50+
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+0-/50+0-/50+0-/50+0-/50+
b. Linearity/assay reportable range

Not applicable. This device is intended for qualitative use only.

c. Stability

The devices are stable at 2-30℃ for 24 months based on the real-time stability study data.

d. Analytical specificity

Analytical specificity for this device was determined through adding the potential interfering substances to drug-free urine samples. The relative cross-reactivity represents the minimum concentration necessary to yield a result similar to the cutoff level of the respective assay. Percent cross-reactivity, provided in the below table, was calculated as the concentration of analyte tested that yielded a positive result, divided by the cutoff concentration, multiplied by 100; compounds that did not yield a positive result at the highest concentration tested have relative cross reactivity results represented by a dash in the table below.

Drug/CutoffCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
6-MAM10Heroin6016.7%
Morphine750000.01%
Normorphine150000Not detected
Nalorphine HCl150000Not detected
Hydrocodone150000Not detected

Page 14

Drug/CutoffCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Hydromorphone150000Not detected
Chlordiazepoxide150000Not detected
Clobazam150000Not detected
D-Amphetamine150000Not detected
(±)-Amphetamine150000Not detected
Levorphanol tartrate150000Not detected
Codeine150000Not detected
Ethylmorphine150000Not detected
Morphine3-β-D-glucuronide150000Not detected
Norcodeine150000Not detected
Oxycodone150000Not detected
Oxymorphone150000Not detected
Procaine hydrochloride150000Not detected
Thebaine150000Not detected
6-Acetylcodeine150000Not detected
Buprenorphine150000Not detected
Dihydrocodeine150000Not detected
Dextromethorphan150000Not detected
Imipramine hydrochloride150000Not detected
Meperidine150000Not detected
(±)-Methadone150000Not detected
Mitragynine(kratom)150000Not detected
Morphine-6-β-D-glucuonide150000Not detected
Naloxone hydrochloride150000Not detected
Naltrexone hydrochloride150000Not detected
Naproxen150000Not detected
Norbuprenorphine150000Not detected
Norbuprenorphine glucuronide150000Not detected
Noroxycodone HCL150000Not detected
Noroxymorphone HCL150000Not detected
(+)-Norpropoxyphene maleate150000Not detected
Oxymorphone-3β-D-glucuronide150000Not detected
Tapentadol HCl150000Not detected
Tramadol150000Not detected
FTY1Acetyl fentanyl166.25%
Acrylfentanyl1100.00%
ω-1-Hydroxyfentanyl20,0000.005%
Isobutyryl fentanyl1100.00%
Ocfentanil2.343.48%

Page 15

Drug/CutoffCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Butyryl fentanyl250.00%
Furanyl fentanyl1100.00%
Valeryl fentanyl2.540.00%
(±) β-hydroxythiofentanyl2.540.00%
4-Fluoro-isobutyrylfentanyl333.33%
Para-fluorobutyryl fentanyl425.00%
Para-fluoro fentanyl2.540.00%
(+)-3-cis-methyl fentanyl502.00%
Carfentanil250.00%
Sufentanil156.67%
Alfentanil75000.01%
Despropionyl fentanyl (4-ANPP)2,0000.05%
Remifentanil150000Not detected
Norfentanyl150000Not detected
Acetyl norfentanyl150000Not detected
Norcarfentanil150000Not detected
Trazodone250000.004%
NFTY5Fentanyl1050%
Acetyl fentanyl1503.3%
Acetyl Norfentanyl2002.5%
(±)-β-Hydroxythiofentanyl HCl25000.2%
Acryl Fentanyl25000.2%
Butyryl Fentanyl50000.1%
Furanyl Fentanyl100000.05%
Para-fluoro butyrl Fentanyl (P-FBF)800000.006%
Para-fluoro Fentanyl400000.013%
9-HydroxyRisperidone100000.05%
Alfentanil200000.025%
Isobutyryl Fentanyl50000.1%
Remifentanil150000.03%
Valeryl Fentanyl200000.025%
Thienyl Fentanyl5010%
(+)-3-cis-methyl fentanyl5010%
4-Fluoro-isobutyryl Fentanyl30000Not detected
Despropionyl fentanyl (4-ANPP)30000Not detected
MT-45 diHCL150000Not detected
Ocfentanil150000Not detected
Risperidone150000Not detected

Page 16

Drug/CutoffCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Sufentanil150000Not detected
Carfentanil150000Not detected
Labetalol Hydrochloride150000Not detected
Trazodone150000Not detected
U-47700150000Not detected
ω-1-Hydroxyfentanyl30000Not detected
6-Acetyl morphine150000Not detected
(±)-Amphetamine150000Not detected
Buprenorphine150000Not detected
Buprenorphine-3ß-D-glucuronide150000Not detected
Codeine150000Not detected
Dextromethorphan150000Not detected
Dihydrocodeine150000Not detected
EDDP150000Not detected
EMDP150000Not detected
Fluoxetine150000Not detected
Heroin150000Not detected
Hydrocodone150000Not detected
Hydromorphone150000Not detected
Ketamine150000Not detected
Levorphanol tartrate150000Not detected
Meperidine150000Not detected
(±)-Methadone150000Not detected
Morphine150000Not detected
Morphine-3-β-D-glucuronide150000Not detected
Naloxone hydrochloride150000Not detected
Naltrexone hydrochloride150000Not detected
Norbuprenorphine150000Not detected
Norcodeine150000Not detected
Norketamine150000Not detected
Normeperidine150000Not detected
Normorphine150000Not detected
Noroxycodone150000Not detected
Oxycodone150000Not detected
Oxymorphone150000Not detected
Pentazocine (Talwin)150000Not detected
Pipamperone150000Not detected
Tapentadol hydrochloride150000Not detected
Thioridazine150000Not detected

Page 17

Drug/CutoffCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Tilidine150000Not detected
Tramadol150000Not detected
O-Desmethyl -cris-Tramadol150000Not detected
N-Desmethyl -cris-Tramadol150000Not detected
Norcarfentanil150000Not detected
THC 20(−)-11-nor-9-carboxy-Delta8-THC20100%
(-)-11-Nor-Δ9-THC- 9-carboxylic acid glucuronide3066.7%
(±)-11-Hydroxy-Δ9-THC20100%
(-)-11-nor-9-carboxy-Δ 9-THC20100%
(-)-Δ9-THC60000.3%
(-)-Δ8-THC40000.5%
Cannabinol80000.25%
Cannabidiol150000Not detected
TRA100n-Desmethyl -cris-Tramadol40025%
o-Desmethyl -cris-Tramadol100010%
o-Desmethyl Venlafaxine15000Not detected
Venlafaxine HCl150000Not detected
e. Interference

Potential interference from compounds chemically dissimilar to the target drugs (6-Monoacetylmorphine (6-MAM10), Fentanyl (FTY1), Norfentanyl (NFTY5), Cannabinoids (THC20), and Tramadol (TRA100)) and from endogenous agents was performed by spiking the substances into pooled urine containing target drugs at near-cutoff concentrations (at +50% and -50% of cutoff). Unless otherwise indicated, substances were tested for potential interference at concentrations of 100 μg/mL.

The following substances demonstrated no positive. The following substances demonstrated no positive or negative interference on the assays encompassed in this submission.

AcetaminophenEffexorNimodipine
AcetophenetidinEnalapril MaleateNitroglycerin
Acetylsalicylic AcidErythromycinNorethindrone
AcyclovirEsomeprazole MagnesiumN-Acetylprocainamide
Afrinβ-EstradiolO-Hydroxyhippuric Acid
Albumin (100mg/dL)1% ethanolOlanzapine
AminophyllineFenofibrateOmeprazole
AminopyrineFenoprofenOxalic Acid
Amiodarone HydrochlorideFentanyl CitrateOxolinic Acid
Amlodipine MesylateFluoxetine HydrochlorideOxymetazoline
AmoxicillinFluvoxamineOndansetran

Page 18

AmpicillinFurosemidePaliperidone
ApomorphineGabapentinPantoprazole
AripiprazoleGentisic AcidPapaverine
AspartameGlibenclamideParoxetine Hydrochloride
AtomoxetineGliclazidePenfluridol
Atorvastatin CalciumGlipizidePenicillinV Potassium
AtropineGlucosePenicillin-G
Benzilic AcidHaloperidolPhenelzine
Benzoic AcidHemoglobinPioglitazone Hydrochloride
BilirubinHydrochlorothiazidePiracetam
BupropionHydrocortisonePravastatin Sodium
Captopril3-HydroxytyraminePrednisone
CarbamazepineIsosorbide DinitratePropylthiouracil
CefradineIsoxsuprineQuetiapine Fumarate
CephalexinIbuprofenQuinine
Chloral HydrateKetoconazoleRanitidine
ChloramphenicolKetoprofenRifampicin
ChlorothiazideKetamineRisperidone
CholesterolKratom powderSalicylic Acid
Ciprofloxacin HydrochlorideLabetalolSerotonin
CitalopramLamotrigineSertraline Hydrochloride
ClarithromycinLevofloxacin HydrochlorideSildenafil Citrate
ClonidineLevonorgestrelSimvastatin
Clopidogrel Hydrogen SulphateLevothyroxine SodiumSodium Valproate
ClozapineLidocaine HydrochlorideSpironolactone
Conjugated EstrogensLisinoprilSulfamethazine
CortisoneLithium CarbonateSulindac
CreatinineLiveriteTetracycline
(-) CotinineLoperamideTetrahydrocortisone 3 -acetate
chlorpheniramineLoratadineTetrahydrocortisone 3-(β-D glucuronide)
D,L-OctopamineMagnesiumTetrahydrozoline
D,L-PropranololMeperidineThiamine
D,L-TyrosineMeprobamateThioridazine
DeoxycorticosteroneMetoprolol TartrateTopiramate
DextromethorphanMifepristoneTramadol Hydrochloride
DiclofenacMirtazapineTrazodone Hydrochloride
DiflunisalMontelukast SodiumTriamterene
DigoxinMosapride CitrateTrifluoperazine
DiphenhydramineMinocyclineTrimethoprim

Page 19

DirithromycinNalidixic AcidUric Acid
DomperidoneNaproxenValproate
D-PseudoephedrineNiacinamideVerapamil
DuloxetineNifedipineVitamin B2
DicyclomineNikethamideVitamin C
ChloroquineEcgonine Methyl EsterPromethazine
f. Urine Density&pH

Interference by pH and specific gravity were also evaluated using pooled urine specimens containing target drugs at near-cutoff concentrations (at +50% and -50% of cutoff). The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.

2. Method comparison Studies

Comparison Study Performance data of AMP500/1000, BAR300, BUP10, BZO300, COC150/300, EDDP300, MDMA500, MET500/MET1000, MOP300, MTD300, OPI2000, OXY100, PCP25, PPX300, TCA1000 and THC50 were presented in the cleared 510(k) K182701 submission.

In this submission, the method comparison study for 6-Monoacetylmorphine (6-MAM10), Fentanyl (FTY1), Cannabinoids (THC20), Tramadol (TRA100) and Norfentanyl (NFTY5), was performed b three operators with 80 unaltered urine samples. These samples were blind labeled and compared to LC/MS or GC/MS results. The results are shown in the table below.

DrugOperatorResultsDrug-free by LC/MSLow Neg by LC/MS (less than -50%)Near Cutoff Neg by LC/MS (Between -50% and the Cutoff)Near Cutoff Pos by LC/MS (Between the cutoff and +50%)High Pos by LC/MS (greater than +50%)
6-MAM 10APositive0032118
Negative13141010
BPositive0022118
Negative13141110
CPositive0012118
Negative13141210
FTY 1APositive0022118
Negative12161010
BPositive0012018
Negative12161120
CPositive0022118
Negative12161010
NFTY 5APositive0032316
Negative10161110

Page 20

DrugOperatorResultsDrug-free by LC/MSLow Neg by LC/MS (less than -50%)Near Cutoff Neg by LC/MS (Between -50% and the Cutoff)Near Cutoff Pos by LC/MS (Between the cutoff and +50%)High Pos by LC/MS (greater than +50%)
BPositive0022216
Negative10161220
CPositive0022216
Negative10161220
THC 20APositive0022810
Negative9151420
BPositive0022910
Negative9151410
CPositive0022810
Negative9151420
TRA 100APositive0022712
Negative10181010
BPositive0012712
Negative10181110
CPositive0012712
Negative10181110

Discordant Results are summarized below:

DrugOperatorSample NumberLC/MS Results (ng/mL)Discordant Device Results
6-MAM 10Operator ASU250500457.867Positive
Operator ASU250500578.193Positive
Operator ASU250500759.192Positive
Operator ASU2505002910.863Negative
Operator BSU250500578.193Positive
Operator BSU250500759.192Positive
Operator BSU2505007410.359Negative
Operator CSU250500759.192Positive
Operator CSU2505007410.359Negative
FTY 1Operator ASU250502710.848Positive
Operator ASU250502840.965Positive
Operator ASU250502491.136Negative
Operator BSU250502840.965Positive
Operator BSU250502491.136Negative
Operator BSU250502701.184Negative
Operator CSU250502710.848Positive
Operator CSU250502840.965Positive
Operator CSU250502701.184Negative
NFTY 5Operator ASU250600424.315Positive
Operator ASU250600304.765Positive
Operator ASU250600114.879Positive
Operator ASU250600655.341Negative
Operator BSU250600304.765Positive

Page 21

DrugOperatorSample NumberLC/MS Results (ng/mL)Discordant Device Results
Operator BSU250600114.879Positive
Operator BSU250600755.526Negative
Operator BSU250600735.699Negative
Operator CSU250600424.315Positive
Operator CSU250600114.879Positive
Operator CSU250600655.341Negative
Operator CSU250600735.699Negative
THC 20Operator ASU2505019418.614Positive
Operator ASU2505018119.363Positive
Operator ASU2505016720.499Negative
Operator ASU2505022020.645Negative
Operator BSU2505019418.614Positive
Operator BSU2505018119.363Positive
Operator BSU2505016720.499Negative
Operator CSU2505018518.422Positive
Operator CSU2505018119.363Positive
Operator CSU2505022020.645Negative
Operator CSU2505023021.541Negative
TRA 100Operator ASU2505013098.136Positive
Operator ASU2505009898.772Positive
Operator ASU25050120102.471Negative
Operator BSU2505013098.136Positive
Operator BSU25050144108.094Negative
Operator CSU2505009898.772Positive
Operator CSU25050120102.471Negative

3. Lay-user study

A lay user study was presented in the cleared 510(k) K182701 submission involving a total of 280 participants from 3 sites. 89 males and 51 females tested T-Cup® Multi-Drug Urine Test Cup Configuration 1 (including AMP 500, BAR 300, BUP 10, BZO 300, COC 150, EDDP 300, MDMA 500, MET 500, MOP 300, MTD 300, OXY 100, PCP 25, PPX 300, TCA 1000, THC 50); 84 male and 56 females tested T-Cup® Multi-Drug Urine Test Cup Configuration 2 (Group 2, including AMP 1000, BAR 300, BUP 10, COC 300, EDDP 300, MDMA 500, MET 1000, OPI 2000, MTDD 300, OXY 100, PCP 25, PPX 300, TCA 1000, THC 50).

In this submission, a lay user study was performed involving a total of 140 participants from 3 sites. 76 males and 64 females tested one configuration of SAFElife™ T-Cup Multi-Drug Urine Test Cup (including 6-MAM10, THC20, TRA100, NFTY5, FTY1, and AMP1000, BAR300, BUP10, BZO300, COC300, mAMP1000, MDMA500, MTD300, OPI2000, OXY100). Each participant was provided one package insert, one blind labeled test solution, and one test device. Test solutions were randomly assigned to participants, one for each.

Page 22

Following testing, users completed a study questionnaire to assess usability and user comprehension, and the results from this questionnaire were found to be acceptable. Participants aged 18 and over, with diverse educational backgrounds, were recruited. Results from the lay user testing are provided in the below table:

Drug/ CutoffResultConcentration
-100% cutoff-75% cutoff-50% cutoff-25% cutoff+25% cutoff+50% cutoff+75% cutoff
6-MAM10Negative20202018100
Positive0002192020
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
AMP1000Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
BAR300Negative20202019200
Positive0001182020
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
BUP10Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
BZO300Negative20202019200
Positive0001182020
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
COC300Negative20202018100
Positive0002192020
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
FTY1Negative20202018100
Positive0002192020
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
mAMP 1000Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
MDMA500Negative20202018100
Positive0002192020
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%

Page 23

Drug/ CutoffResultConcentration
-100% cutoff-75% cutoff-50% cutoff-25% cutoff+25% cutoff+50% cutoff+75% cutoff
MTD300Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
NFTY5Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
OPI2000Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95%95%100%100%
OXY100Negative20202018100
Positive0002192020
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%
THC20Negative20202019200
Positive0001182020
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
TRA100Negative20202018100
Positive0002192020
Total20202020202020
Agreement (%)100%100%100%90%95%100%100%

Lay users completed given surveys on the ease of understanding the package insert. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

4. Clinical Study

Not applicable.

12. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison and Lay-user studies of the devices, it's concluded a substantial equivalence decision.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).