K Number
K243996

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-02-07

(43 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AssureTech Panel Dip Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Methamphetamine, Fentanyl, Norfentany], Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations listed. The single or multi-test panels can consist of up to seventeen (17) of the above listed analytes in any combination. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. For in vitro diagnostic use only.

The AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Methamphetamine, Fentanyl, Norfentany], Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations listed. The single or multi-test panels can consist of up to seventeen (17) of the above listed analytes in any combination. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. For in vitro diagnostic use only.

The AssureTech Multi-drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Cocaine, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations listed. The single or multi-test panel can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

The AssureTech Multi-drug Urine Test Cup are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations listed. The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Device Description

The AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Fentanyl, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP. Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single use in vitro diagnostic devices, which come in the formats of Panel Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

The provided document describes the FDA 510(k) premarket notification for AssureTech Panel Dip Tests and Quick Cup Tests, which are in vitro diagnostic devices for qualitative and simultaneous detection of various drugs of abuse in human urine. The document focuses on demonstrating substantial equivalence to a predicate device (K181768).

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for this type of device are primarily related to its analytical performance, specifically its ability to accurately detect the presence or absence of target drugs at specified cutoff concentrations. The device is a qualitative test, meaning it provides a "positive" or "negative" result, rather than a quantitative measurement.

The study demonstrates performance through:

  • Precision: Consistency of results across multiple runs and lots, especially near the cutoff concentrations.
  • Specificity: Ability to react only with the target drug/metabolite and not with other substances or structurally similar compounds.
  • Interference: Lack of false positives/negatives due to common interfering substances in urine, or variations in urine specific gravity and pH.
  • Method Comparison: Agreement of device results with a known, more precise reference method (LC/MS).
  • Lay-user study: Evaluation of the device's performance when used by non-professionals, assessing ease of use and accuracy of interpretation.

Here is a table summarizing the reported device performance for Fentanyl (FYL), Norfentanyl (NFYL), and as an example for another drug, Amphetamine (AMP) from the "Lay-user study" data. The document does not explicitly state numerical "acceptance criteria" for each performance metric, but rather presents the results of the studies conducted to show sufficient performance for regulatory clearance. The implicit acceptance criterion for a qualitative test like this is generally very high accuracy, especially around the cutoff, and a low rate of false positives/negatives.

Table of Performance for Key Drugs (Fentanyl, Norfentanyl, Amphetamine)

Precision Study (Fentanyl - Panel Dip/Quick Cup, Norfentanyl - Panel Dip/Quick Cup):
The precision data is presented for three lots and various concentrations relative to the cutoff. The data shows very high consistency. For instance, for Fentanyl:

  • At -100%, -75%, -50% cut off (negative range), all 50 tests across 3 lots consistently yielded negative results (50-/0+).
  • At +25%, +50%, +75%, +100% cut off (positive range), all 50 tests consistently yielded positive results (50+/0-).
  • At the cutoff concentration, the device shows variability, as expected for tests near the decision threshold. For Fentanyl Panel Dip, results were 28+/22-, 29+/23-, 28+/22- for Lot 1, 2, 3 respectively (meaning some tests were positive and some negative at the cutoff). This variability is inherent for qualitative tests around the cutoff and implies that some samples at the cutoff may read positive and others negative, which is acceptable performance for a qualitative test. Similar patterns are observed for Quick Cup Fentanyl, Panel Dip Norfentanyl, and Quick Cup Norfentanyl.

Method Comparison Study (Fentanyl - Panel Dip/Quick Cup, Norfentanyl - Panel Dip/Quick Cup):
This study compared the device results against LC/MS, the preferred confirmatory method. The results are presented in tables showing agreement across different concentration ranges (Negative, Low Negative, Near Cutoff Negative, Near Cutoff Positive, High Positive).

Example for FYL (Fentanyl) - Panel Dip, Operator 1:

  • Negative (LC/MS 0): Device: 0 Positive, 1 Negative (1 discordant result here, sample 1484, LC/MS 0.78 ng/mL, Device: +)
  • Low Negative (LC/MS < -50%): Device: 0 Positive, 19 Negative
  • Near Cutoff Negative (LC/MS Between -50% and cutoff): Device: 2 Positive, 18 Negative (2 discordant results, samples 1484, 9778, 4576 which are positive by device but negative by LC/MS near the cutoff)
  • Near Cutoff Positive (LC/MS Between cutoff and +50%): Device: 19 Positive, 1 Negative (1 discordant result, sample 5419, LC/MS 1.05 ng/mL, Device: -)
  • High Positive (LC/MS > +50%): Device: 20 Positive, 0 Negative

Lay-User Study (Selected data for AMP, FYL, NFYL):
This study evaluates the percentage of correct results when used by lay persons at various concentrations relative to the cutoff.

Example for AMP (Amphetamine):

  • Negative (100% below cutoff): 100% correct (0 positive, 20 negative)
  • Low Negative (-75% to -25% Cutoff): 100% correct negative for -75% and -50%, but 0 positive/20 negative for -25% cutoff.
  • Positive (+25% to +75% Cutoff): Generally high correctness (95%-100%). For +25% cutoff, 95% correctness (19 positive, 1 negative).
Drug (Identifier)Cut-off LevelReported Device Performance (Summary)
Fentanyl (FYL)1 ng/mLPrecision: At -100% to -50% cutoff, 100% negative calls (50-/0+ over 3 lots for Panel Dip & Quick Cup). At +25% to +100% cutoff, 100% positive calls (50+/0- over 3 lots for Panel Dip & Quick Cup). At cutoff, performance is mixed (e.g., Panel Dip Lot 1: 28+/22-). Method Comparison: High concordance with LC/MS, especially for samples well above or below cutoff. Some discordant results near cutoff for both negative (e.g., sample 1484, LC/MS 0.78 ng/mL, device +) and positive (e.g., sample 5419, LC/MS 1.05 ng/mL, device -) as expected for qualitative tests. Lay-User Study: All 20 negative samples at -100%, -75%, -50% cutoff were correctly identified as negative (100% correct). At -25% cutoff, 95% correct (19 negative, 1 positive). All 20 positive samples at +50%, +75% cutoff were correctly identified as positive (100% correct). At +25% cutoff, 100% correct (20 positive).
Norfentanyl (NFYL)5 ng/mLPrecision: At -100% to -50% cutoff, 100% negative calls (50-/0+ over 3 lots for Panel Dip & Quick Cup). At +25% to +100% cutoff, 100% positive calls (50+/0- over 3 lots for Panel Dip & Quick Cup). At cutoff, performance is mixed (e.g., Panel Dip Lot 1: 27+/23-). Method Comparison: High concordance with LC/MS, with some discordance near cutoff (e.g., sample 4074, LC/MS 4.39 ng/mL, device +; sample 0687, LC/MS 5.05 ng/mL, device -). Lay-User Study: All 20 negative samples at -100%, -75%, -50% cutoff were correctly identified as negative (100% correct). At -25% cutoff, 95% correct (19 negative, 1 positive). All 20 positive samples at +25%, +50%, +75% cutoff were correctly identified as positive (100% correct).
Amphetamine (AMP)500 ng/mLLay-User Study: All 20 negative samples at -100%, -75%, -50%, -25% cutoff were correctly identified as negative (100% correct). All 20 positive samples at +50%, +75% cutoff were correctly identified as positive (100% correct). At +25% cutoff, 95% correctness (19 positive, 1 negative).

Detailed Study Information:

  1. Sample sizes used for the test set and the data provenance:

    • Precision Study: For Fentanyl and Norfentanyl, the reported data is for 3 lots, with 2 runs per day for 25 days, for 9 concentrations (e.g., -100% cutoff, -75% cutoff, etc.). This implies 50 tests per concentration per lot (2 runs * 25 days), leading to 450 tests per drug type per lot (9 concentrations * 50 tests), and 1350 tests per drug type across all 3 lots. The data provenance implies these samples were prepared by spiking known concentrations of drug into negative samples, indicating a controlled laboratory environment. Data for other analytes was "reported in K181768" (the predicate device documentation), so the exact sample sizes are not explicitly stated in this document but are assumed to be similar.
    • Method Comparison Study: For Fentanyl and Norfentanyl, 80 unaltered clinical samples (40 negative and 40 positive based on LC/MS results) were used per drug. Each sample was tested by three laboratory assistants for each device type (Panel Dip and Quick Cup). This means 80 samples * 3 operators = 240 tests per drug for each device type. The data provenance is "in-house" and "unaltered clinical samples." The document does not specify the country of origin, but given the FDA submission, it's likely US-based or compliant with US standards. The study appears to be retrospective, using already collected clinical samples for comparison.
    • Lay-user Study: 280 lay persons were used for each device format (Panel Dip and Quick Cup, though the results summarized apply to the overall device type). Urine samples were prepared at 7 different concentrations (negative, +/-25%, +/-50%, +/-75%, +/-100% of cutoff). Each participant received one blind-labeled sample and one device. Assuming each person tested one sample, this implies 280 samples were tested for each specific drug evaluated by lay-users on each format. The data provenance: "samples were prepared by spiking drugs into drug-free pooled urine specimens" and confirmed by LC/MS. This is a controlled experimental set-up rather than real-world patient samples.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Precision Study: Ground truth was established by spiking known concentrations of drugs into negative samples and confirmed by LC/MS. No human experts were involved in establishing the ground truth directly for this part.
    • Method Comparison Study: The ground truth was established by LC/MS (Liquid Chromatography-Mass Spectrometry), which is explicitly stated as the "preferred confirmatory method" and is considered a gold standard for drug detection and quantification in urine. No human expert readers established the ground truth; it was a laboratory instrument measurement. The study used three laboratory assistants to read the device results, but they were comparing their readings against the LC/MS truth, not establishing the truth themselves. Their qualifications are not specified beyond "laboratory assistants."
    • Lay-user Study: The ground truth was established by spiking known concentrations of drugs confirmed by LC/MS. No human experts established the ground truth of the samples. The study assessed the lay-users' ability to interpret the device results against this known truth.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Precision Study: No adjudication method mentioned as samples were prepared with known concentrations.
    • Method Comparison Study: No adjudication method was explicitly mentioned for the device results. Each of the three operators performed their own reads, and their individual results were compared to the LC/MS. Discordant results are noted for each operator.
    • Lay-user Study: No adjudication method was mentioned. Each lay user tested one sample against a pre-defined truth.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in this report. This device is a rapid diagnostic test (lateral flow immunoassay), not an AI-assisted diagnostic tool for interpretation of medical images or other complex data. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is a lateral flow immunoassay, which is a physical diagnostic device producing a visual result (colored lines). It does not involve an "algorithm" in the sense of a software-based AI or computational algorithm. The device itself is the "standalone" diagnostic. Its performance characteristics (precision, specificity, interference) are essentially its "algorithm only" performance. The method comparison study is akin to assessing the device's standalone performance against a gold standard. The lay-user study assesses human interpretation of the device's standalone output.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The primary ground truth for the analytical and method comparison studies was Liquid Chromatography-Mass Spectrometry (LC/MS), which is a highly accurate chemical method for detecting and quantifying substances.
    • For the precision and lay-user studies, the ground truth was based on spiked urine samples with known drug concentrations, which were then confirmed by LC/MS.
  7. The sample size for the training set:

    • This document describes a 510(k) premarket notification for an in vitro diagnostic device (lateral flow immunoassay). Unlike AI/ML-driven devices that require extensive training data, such chemical-based devices are developed and optimized through chemical engineering and biological principles, not by "training" on datasets in the AI sense. Therefore, the concept of a "training set" with a statistical sample size as understood in machine learning is not applicable to this type of device. The development process involves chemical formulation and validation, not data training.
  8. How the ground truth for the training set was established:

    • Since the concept of a "training set" as it pertains to AI/ML devices is not applicable, the establishment of ground truth for a training set is also not relevant in this context. The "ground truth" for the performance evaluation of the device relied on LC/MS results and carefully prepared spiked samples with known drug concentrations.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 7, 2025

Assure Tech LLC % Jenny Xia Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K243996

Trade/Device Name: AssureTech Panel Dip Tests; AssureTech Quick Cup Tests; AssureTech Multidrug Urine Test Panel; AssureTech Multi-drug Urine Test Cup Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: NFT, PTH, NGL, NFV, NFY, PTG, NGG, NGM, QAW, NFW, QBF Dated: December 22, 2024 Received: December 26, 2024

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Kotarek Digitally signed by Joseph A. Kotarek -S -5 Date: 2025.02.07 09:19:07 -05'00' Joseph Kotarek, Ph.D. Branch Chief for Toxicology Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243996

Device Name

AssureTech Panel Dip Tests; AssureTech Quick Cup Tests:

AssureTech Multi-drug Urine Test Panel; AssureTech Multi-drug Urine Test Cup

Indications for Use (Describe)

The AssureTech Panel Dip Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Methamphetamine, Fentanyl, Norfentany], Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL
Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panels can consist of up to seventeen (17) of the above listed analytes in any combination. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

The AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Methamphetamine, Fentanyl, Norfentany], Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL

{4}------------------------------------------------

Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panels can consist of up to seventeen (17) of the above listed analytes in any combination. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

The AssureTech Multi-drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Cocaine, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL
Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panel can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only.

The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

The AssureTech Multi-drug Urine Test Cup are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations of:

{5}------------------------------------------------

Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL
Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

510(k) SUMMARY K243996

Assure Tech. LLC.

    1. Date: December 22, 2024
    1. Submitter:
  • 1521 Concord Pike, Suite 201 Wilmington, DE 19803
    1. Contact person: Jenny Xia LSI International Inc. 504E Diamond Ave., Suite H Gaithersburg, MD 20877 Telephone: 301-525-6856 Email: jxia@lsi-consulting.org
    1. Device Name: AssureTech Panel Dip Tests AssureTech Multi-drug Urine Test Panel AssureTech Quick Cup Tests AssureTech Multi-drug Urine Test Cup
Classification:Class 2
Product CodeClassificationRegulation SectionPanel
NFTAmphetamineII21 CFR § 862.3100, AmphetamineTest SystemToxicology (91)
NFWCannabinoidsII21 CFR § 862.3870, CannabinoidsTest SystemToxicology (91)
NFYCocaineII21 CFR § 862.3250, Cocaine andCocaine Metabolites Test SystemToxicology (91)
NGGMethamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology (91)
NGLMorphineII21 CFR § 862.3650, Opiate TestSystemToxicology (91)
NFVOxazepamII21 CFR § 862.3170,Benzodiazepine Test SystemToxicology (91)
NGLOxycodoneII21 CFR § 862.3650, Opiate TestSystemToxicology (91)
PTHSecobarbitalII21 CFR § 862.3150, BarbiturateTest SystemToxicology (91)
NGLBuprenorphineII21 CFR § 862.3650,Opiate Test SystemToxicology (91)
NGLFentanylNorfentanylII21 CFR § 862.3650,Opiate Test SystemToxicology (91)
NGGMethylenedioxy-methamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology (91)
NGMPhencyclidineunclassifiedEnzyme ImmunoassayPhencyclidineToxicology (91)
PTGMethadoneII21 CFR § 862.3620, MethadoneTest SystemToxicology (91)

{7}------------------------------------------------

PTG2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine(EDDP)II21 CFR § 862.3620, MethadoneTest SystemToxicology (91)
QAWNortriptylineII21 CFR, 862.3910 TricyclicAntidepressant Drugs Test SystemToxicology (91)
QBFPropoxypheneII21 CFR, 862.3700 PropoxypheneTest SystemToxicology (91)
    1. Predicate Devices: K181768
      AssureTech Panel Dip Tests and AssureTech Quick Cup Tests
    1. Indications for Use
      The AssureTech Panel Dip Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Fentanyl, Norfentanyl, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:
Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL
Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panels can consist of up to seventeen (17) of the above listed analytes in any combination.

Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

The AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Fentanyl, Norfentanyl, Morphine, Oxycodone, Secobarbital,

{8}------------------------------------------------

Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL
Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination.

Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

For in vitro diagnostic use only.

The AssureTech Multi-drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL
Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL

{9}------------------------------------------------

Phencyclidine(PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana(THC)50 ng/mL

The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

The AssureTech Multi-drug Urine Test Cup are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Fentanyl, Norfentanyl, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline, d-Propoxyphene and adulterants in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)500 ng/mL
Secobarbital (BAR)300 ng/mL
Buprenorphine (BUP)10 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)150 ng/mL
Methadone metabolite (EDDP)300 ng/mL
Fentanyl (FYL)1 ng/mL
Ecstasy (MDMA)500 ng/mL
Methamphetamine (MET)500 ng/mL
Morphine (MOR)300 ng/mL
Methadone (MTD)300 ng/mL
Norfentanyl (NFYL)5 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test cups can consist of up to seventeen (17) of the above listed analytes in any combination. It is for in vitro diagnostic use only.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result. particularly when the preliminary result is positive.

7. Device Description

The AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Fentanyl, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP. Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single

{10}------------------------------------------------

use in vitro diagnostic devices, which come in the formats of Panel Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    1. Substantial Equivalence Information
      A summary comparison of features of the AssureTech Panel Dip Tests and AssureTech Quick Cup Tests and the predicate devices is provided in following tables.
ItemDevicePredicate - K181768
Indication(s)for UseFor the qualitative determination of drugs ofabuse in human urine.Same (but the number ofdrugs detected is different)
Calibrator andCut-Off ValuesAmphetamine (AMP): 500 ng/mlOxazepam (BZO):300 ng/mlCocaine(COC): 150 ng/mlMarijuana (THC):50 ng/mlMethamphetamine (MET): 500 ng/mlMorphine (MOR): 300ng/mLOxycodone(OXY) : 100 ng/mlSecobarbital (BAR): 300 ng/mlBuprenorphine (BUP): 10 ng/mlMethylenedioxy-methamphetamine(MDMA):500 ng/mlPhencyclidine (PCP): 25 ng/mlMethadone (MTD): 300 ng/ml2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP): 300 ng/mlNortriptyline (TCA): 1000 ng/mlPropoxyphene (PPX): 300 ng/mlFentanyl (FYL): 1ng/mlNorfentanyl (NFYL): 5ng/mlSame exceptthat no FYLand NFYL
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor over-the-counterSame
ConfigurationsDip Card and CupSame

Table 1: Features Comparison of AssureTech Panel Dip Tests/AssureTech Quick Cup Tests and the Predicate Devices

  1. Test Principle

{11}------------------------------------------------

The AssureTech Panel Dip Tests, and AssureTech Quick Cup Tests are rapid tests for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Fentanyl, Norfentanyl, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and Propoxyphene in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoffconcentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • Precision a.

Precision studies were carried out for samples with concentrations of -100% cut off. -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables for Fentanyl and Norfentanyl. Data supporting precision for the remaining analytes was reported in K181768.

Panel Dip
Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+49-/1+28+/22-50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+49-/1+29+/23-50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28+/22-50+/0-50+/0-50+/0-50+/0-

Fentanyl

Donal D

Quick Cup

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoff+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+49-/1+29+/21-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+27+/23-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+29+/21-50+/0-50+/0-50+/0-

Norfentanyl

Panel Dip

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+49-/1+27+/23-50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+49-/1+28+/22-50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+28+/22-50+/0-50+/0-50+/0-50+/0-

Quick Cup

{12}------------------------------------------------

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+49-/1+29+/21-50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+49-/1+27+/23-50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+28+/22-50+/0-50+/0-50+/0-50+/0-

c. Stability

The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 45 °C and real time stability determination at both 4 °C and 30 °C.

d. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 50% below and 50% above Cut-Off levels. These urine samples were tested using three lots of each device. Compounds that showed no interference at a concentration of 100ug/mL or specified concentrations are summarized in the following table. There were no differences observed for different devices.

11-nor-9 carboxy THC (except forTHC)DL-TyrosineNifedipine
3-HydroxytyramineDopamineNorethindrone
4-Bromo-2,5,DimethoxyphenethylamineDoxepin (except for TCA)Norpropoxyphene (except for PPX)
7-AminoflunitrazepamEcgonine methyl esterNorpseudoephedrine
7-AminonitrazepamEphedrine (except for MET)Nortriptyline (except for TCA)
AcetaminophenErythromycinNoscapine
Acetone (1000 mg/dL)EstradiolOctopamine
AcetophenetidinEstroneO-Hydroxyhippuric acid
Acetylsalicylic Acid (500 µg/mL)Ethanol (1%)Oxalic Acid (100mg/dL)
Albumin (100 mg/dL)Fenfluramine (except for MET)Oxazepam (except for BZO)
AlbuterolFenofibrateOxolinic acid
AminopyrineFenoprofenOxymetazoline
Amitriptyline (except for TCA)FluphenazinePapaverine
Amlodipine besylateFotemustinePenicillin-G
Amobarbital (except for BAR)FurosemidePentazocine
AmoxicillinGalactosePerphenazine
AmpicillinGamma Globulin (500mg/dL)Phencyclidine (except for PCP)
ApomorphineGemfibrozilPhenelzine
Ascorbic AcidGentisic acidPhenobarbital (except for BAR)
AspartameGlucose (3000 mg/dL)Phentermine (except for AMP)
AspirinGuaiacolglyceryl etherPhenylethylamine (except for MET)
AtropineHemoglobinPrednisone
BaclofenHexobarbitalPromazine (except for TCA/EDDP)
Benzilic acidHydralazinePromethazine
Benzocaine6HydrochlorothiazidePropoxyphene (except for PPX)
Benzoic AcidHydrocortisonePropranolol
Benzoylecgonine (except for COC)HydroxytyraminePseudoephedrine
BenzylpiperiazineIbuprofenPyridoxine
BilirubinI-CotininePyrilamine
Boric Acid (1%)I-ErythromycinPyrogallol
BupropionImipramine (except for TCA)Quinidine
Caffeine (500 µg/mL)IsoproterenolQuinine
CarbamazepineIsoxsuprineQuinolinic Acid
CarisoprodolKetamineRanitidine
CetirizineKetoprofenr-Globulin
Chloral hydrateLabetalolRiboflavin
ChloramphenicolLamotrigineSalicylic Acid
Chlordiazepoxide (except for BZO)LidocaineSecobarbital (except for BAR)
ChlorothiazideLisinoprilSerotonin(5-Hydroxytyramine)
ChlorpheniramineLoperamideSodium Azide
ChlorpromazineLoratidineSufentanil Citrate
CholesterolLorazepam (except for BZO)Sulfamethazine

{13}------------------------------------------------

ClofibrateLSDSulindac
Clomipramine (except for TCA)L-thyroxineTetracycline
ClonidineMaprotiline (except for TCA)Tetrahydrocortisone 3-acetate
CortisoneMeperidineTetrahydrocortisone3-(B-Dglucuronide)
CotinineMeprobamateTetrahydrozoline
Creatine HvdrateMetforminThiamine
CreatinineMethapyrileneThioridazine
Cyclobenzaprine (except for TCA)MethaqualoneTriamterene
Cyclodextrin-rMethoxyphenamine (except forAMP/MET)Trifluoperazine
CyproheptadineMethylphenidateTrifluoromethylphenyl- piperazine
d.l-IsoproterenolMetoprololTrimethoprim
DemoxepamMetronidazole (300 ug/mL)Tryptamine
DeoxycorticosteroneN-AcetylprocainamideTyramine
Desipramine (except for TCA)NaCl (40000 ug/mL)Urea (2000 mg/dL)
DextromethorphanNalidixic AcidUric Acid
DiclofenacNaloxone (except for OXY)Valproic acid (250 ug/mL)
DiflunisalNaltrexoneVenlafaxine
DigoxinNaproxenVerapamil
Dimethyl-aminoantipyrineN-desmethylapentadolZolpidem
DiphenhydramineNiacinamideZolpidem Tartrate
DiphenylhydantoinNicotineZomepirac
DL-TryptophanNicotinic Acidß-Estradiol
B-Hvdroxybutvric Acid

e. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three lots of each device. The lowest concentration that caused a positive result for each compound are listed below for Fentanyl and Norfentanyl. Data supporting specificity for the remaining analytes was reported in K181768. There were no differences observed for different devices.

Fentanyl (Cutoff=1ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(±) β-hydroxythiofentanyl520
(±)-3-cis-methyl fentanyl502
4-Fluoro-isobutyrylfentanyl502
Acetyl fentanyl520
Acryl fentanyl520
Butyryl fentanyl254
Fentanyl1100
Furanyl fentanyl1010
Isobutyryl fentanyl520
Ocfentanil1001
Para-fluoro fentanyl254
Para-fluorobutyryl fentanyl254
Valeryl fentanyl502
ω-1-Hydroxyfentanyl500000.002
Acetyl norfentanyl>100000<0.001
Alfentanil>100000<0.001
Norcarfentanil>10000<0.01
Norfentanyl>100000<0.001
Remifentanil>10000<0.01
Sufentanil>10000<0.01
Carfentanil>10000<0.01

{14}------------------------------------------------

Despropionyl fentanyl (4-ANPP)>50000<0.002
Norfentanyl(Cutoff=5ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(±)-β-Hydroxythiofentanyl806.25
4-Fluoro-isobutyrylFentanyl2502
9-HydroxyRisperidone100000.05
Acetyl Fentanyl5001
Acetyl Norfentanyl5010
Acryl Fentanyl1005
Alfentanil1000000.005
Butyryl Fentanyl5001
(±)-3-cis-methyl fentanyl2002.5
trans-d, 1-3-methylfentanyl2002.5
Fentanyl5001
Furanyl Fentanyl10000.5
Isobutyryl Fentanyl756.67
Labetalol Hydrochloride1000000.005
MT-4580000.06
Norfentanyl5100
Ocfentanil50000.1
Para-fluorobutyryl fentanyl1005
Para-fluoro fentanyl1005
Risperidone8000.63
Thienyl Fentnayl1005
Valeryl Fentanyl5001
Carfentanil>10000<0.05
Despropionyl fentanyl (4-ANPP)>50000<0.01
Norcarfentanil>10000<0.05
Remifentanil>10000<0.05
Sufentanil>10000<0.05
Trazodone>10000<0.05
U-47700>100000<0.005
ω- 1-Hydroxyfentanyl>50000<0.01

Negative results were obtained for all the following opioids compounds tested at 100 µg/mL. There is no cross-reactivity for these compounds.

6-Acetyl morphineHydromorphoneOxycodone
AmphetamineLevorphanolOxymorphone
BuprenorphineMethadonePentazocine (Talwin)
BuprenorphineglucuronideMorphinePipamperone
CodeineMorphine-3-glucuronideTapentadol
DihydrocodeineNorbuprenorphineTilidine
EDDPNorcodeineTramadol
EMDPNorketamineTramadol-O-Desmethyl

{15}------------------------------------------------

FluoxetineNormeperidineTramadol-N-Desmethyl
HeroinNormorphine
HydrocodoneNoroxycodone

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 50% below and 50% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off. There were no differences observed for different devices.

2. Comparison Studies

Method comparison studies for the AssureTech Panel Dip Tests and the AssureTech Quick Cup Tests were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results. The results are presented in the tables below for Fentanyl and Norfentanyl. Data supporting method comparison for the remaining analytes was reported in K181768.

FYL

PanelDipNegativeLow Negative byLC/MS(less than-50%)Near Cutoff Negative byLC/MS(Between-50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator1Positive0021920
Negative1191810
Operator2Positive0021920
Negative1191810
Operator3Positive0021920
Negative1191810

Discordant Results

OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 1, 2, 314840.78+
Operator 197780.95+
Operator 2, 345760.97+
Operator 154191.05-
Operator 2, 394011.00-
Quick CupNegativeLow Negative by LC/MS(less than -50%)Near Cutoff Negative by LC/MS(Between -50% and cutoff)Near Cutoff Positive by LC/MS(Between the cutoff and +50%)High Positive by LC/MS(greater than +50%)

{16}------------------------------------------------

Operator0031920
1Positive0031920
Negative1191710
2Positive0021920
Negative1191810
3Positive0021820
Negative1191820

Discordant Results

OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 1, 2, 314840.78+
Operator 1, 397780.95+
Operator 1, 245760.97+
Operator 1, 354191.05-
Operator 2, 394011.00-

NFYL

PanelDipNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator1Positive0031920
Negative2181810
Operator2Positive0021920
Negative2181910
Operator3Positive0021820
Negative2181920

Discordant Results

OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 140744.39+
Operator 113114.10+
Operator 1, 202424.82+
Operator 2, 359064.68+
Operator 320664.24+
Operator 1, 306875.05-
Operator 2, 380695.15-
QuickCupNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator1Positive0021820
Negative2181820
Operator2Positive0031920
Negative2181710

{17}------------------------------------------------

Operator0021920
3Positive0021920
3Negative2181810
Discordant Results
OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 1, 240744.39+
Operator 1, 302424.82+
Operator 2, 320664.24+
Operator 259064.68+
Operator 1, 2, 306875.05-
Operator 180695.15-

Lay-user study

A lay user study was performed at three intended user sites with 280 lay persons for each device format. The lay users had diverse educational and professional backgrounds and ranged in age from 20 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested. Typical results are shown below.

The results summary for AMP:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correctresults (%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20124.4020100
-50% Cutoff20246.6020100
-25% Cutoff20377.7020100
+25% Cutoff20621.219195
+50% Cutoff20756.6200100
+75% Cutoff20870.5200100

The results summary for FYL:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correctresults (%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff200.24020100
-50% Cutoff200.51020100
-25% Cutoff200.7701995
+25% Cutoff201.26200100
+50% Cutoff201.47200100
+75% Cutoff201.76200100

The results summary for NFYL:

% of CutoffNumber ofDrugLay person ResultsThe percentage
------------------------------------------------------------------

{18}------------------------------------------------

samplesConcentration byLC/MS(ng/mL)No. ofPositiveNo. ofNegativeof correctresults (%)
-100% Cutoff200020100
-75% Cutoff201.25020100
-50% Cutoff202.65020100
-25% Cutoff203.7301995
+25% Cutoff206.09200100
+50% Cutoff207.14200100
+75% Cutoff208.77200100

The results summary for BAR:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correctresults (%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2076.4020100
-50% Cutoff20158.2020100
-25% Cutoff20224.221890
+25% Cutoff20372.519195
+50% Cutoff20440.2200100
+75% Cutoff20537.9200100

The results summary for BUP:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
-100% Cutoff2000
-75% Cutoff202.46020100
-50% Cutoff204.99020100
-25% Cutoff207.4911995
+25% Cutoff2012.5119195
+50% Cutoff2014.88200100
+75% Cutoff2017.53200100

The results summary for BZO:

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff2071.6020100
-50% Cutoff20146.1020100
-25% Cutoff20224.121890
+25% Cutoff20373.8200100
+50% Cutoff20459.7200100
+75% Cutoff20514.0200100

The results summary for COC:

% of CutoffNumber ofDrugLay person ResultsThe percentage
------------------------------------------------------------------

{19}------------------------------------------------

samplesConcentration byLC/MS(ng/mL)No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff2039.1020100
-50% Cutoff2075.6020100
-25% Cutoff20113.011995
+25% Cutoff20188.619195
+50% Cutoff20223.3200100
+75% Cutoff20262.1200100

The results summary for EDDP:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
-100% Cutoff200020100
-75% Cutoff2075.8020100
-50% Cutoff20150.7020100
-25% Cutoff20225.711995
+25% Cutoff20375.719195
+50% Cutoff20450.3200100
+75% Cutoff20525.5200100

The results summary for MDMA:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20126.1020100
-50% Cutoff20250.6020100
-25% Cutoff20361.411995
+25% Cutoff20632.2200100
+50% Cutoff20748.0200100
+75% Cutoff20869.1200100

The results summary for MET:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20122.6020100
-50% Cutoff20248.9020100
-25% Cutoff20376.2020100
+25% Cutoff20628.419195
+50% Cutoff20753.3200100
+75% Cutoff20871.7200100

The results summary for MOR:

% of CutoffNumber ofDrugLay person ResultsThe percentage
------------------------------------------------------------------

{20}------------------------------------------------

samplesConcentration byLC/MS(ng/mL)No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff2076.8020100
-50% Cutoff20150.1020100
-25% Cutoff20225.811995
+25% Cutoff20376.519195
+50% Cutoff20449.1200100
+75% Cutoff20526.3200100

The results summary for MTD:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2075.4020100
-50% Cutoff20157.6020100
-25% Cutoff20225.611995
+25% Cutoff20374.519195
+50% Cutoff20451.6200100
+75% Cutoff20524.2200100

The results summary for OXY:

% of CutoffNumber of samplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsNo. ofPositiveLay person ResultsNo. ofNegativeThe percentageof correct results(%)
-100% Cutoff200200100
-75% Cutoff2026.2200100
-50% Cutoff2050.7020100
-25% Cutoff2075.911995
+25% Cutoff20125.819195
+50% Cutoff20146.7200100
+75% Cutoff20173.7200100

The results summary for PCP:

% of CutoffNumber of samplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
-100% Cutoff200No. ofPositive: 0No. ofNegative: 20100
-75% Cutoff206.42No. ofPositive: 0No. ofNegative: 20100
-50% Cutoff2012.47No. ofPositive: 0No. ofNegative: 20100
-25% Cutoff2019.18No. ofPositive: 2No. ofNegative: 1890
+25% Cutoff2031.63No. ofPositive: 19No. ofNegative: 195
+50% Cutoff2037.73No. ofPositive: 20No. ofNegative: 0100
+75% Cutoff2043.60No. ofPositive: 20No. ofNegative: 0100

The results summary for PPX:

% of CutoffNumber ofDrugLay person ResultsThe percentage
------------------------------------------------------------------

{21}------------------------------------------------

samplesConcentration byLC/MS(ng/mL)No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff2075.0020100
-50% Cutoff20143.7020100
-25% Cutoff20220.611995
+25% Cutoff20375.219195
+50% Cutoff20452.5200100
+75% Cutoff20525.3200100

The results summary for TCA:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20236.9020100
-50% Cutoff20491.6020100
-25% Cutoff20751.311995
+25% Cutoff201254.2200100
+50% Cutoff201493.4200100
+75% Cutoff201750.5200100

The results summary for THC:

% of CutoffNumber of samplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2012.3020100
-50% Cutoff2024.1020100
-25% Cutoff2038.311995
+25% Cutoff2063.319195
+50% Cutoff2076.1200100
+75% Cutoff2087.4200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

  1. Clinical Studies

Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the AssureTech Panel Dip Tests, AssureTech Quick Cup Tests, AssureTech Multi-drug Urine Test Panel and AssureTech Multi-drug Urine Test Cup are substantially equivalent to the predicate.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).