K Number
K252118

Validate with FDA (Live)

Date Cleared
2025-08-27

(51 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CLUNGENE Multi-Drug Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:

DrugCalibratorCut-off (ng/mL)
Morphine (MOP/OPI300)Morphine300
Morphine (MOP/OPI2000)Morphine2,000
Methamphetamine (mAMP/MET1000)D-Methamphetamine1,000
Methamphetamine (mAMP/MET500)D-Methamphetamine500
Cocaine (COC300)Benzoylecgonine300
Cocaine (COC150)Benzoylecgonine150
Marijuana (THC)11-nor-9-THC-9-COOH50
Methylenedioxymethamphetamine (MDMA)D,L-Methylenedioxymethamphetamine500
Buprenorphine (BUP)Buprenorphine10
Propoxyphene (PPX)D-Propoxyphene300
Amphetamine (AMP1000)D-Amphetamine1,000
Amphetamine (AMP500)D-Amphetamine500
Phencyclidine (PCP)Phencyclidine25
Methadone metabolite (EDDP)2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine300
Oxycodone (OXY)Oxycodone100
Oxazepam (BZO)Oxazepam300
Nortriptyline (TCA)Nortriptyline1,000
Secobarbital (BAR)Secobarbital300
Methadone (MTD)Methadone300
6-Monoacetylmorphine (6-MAM)6-Monoacetylmorphine10
Fentanyl (FYL)Fentanyl1

The single or multi-test cups can consist of any combination of the above listed drug analytes, but only one cut off concentration under same drug condition will be included per device.

This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Evaluate preliminary positive results carefully. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

The CLUNGENE Multi-Drug Home Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:

Drug (Identifier)Cut-off (ng/mL)
Morphine (MOP/OPI2000)300 or 2000
Methamphetamine (mAMP/MET)500 or 1,000
Cocaine (COC)150 or 300
Marijuana (THC)50
Methylenedioxymethamphetamine (MDMA)500
Buprenorphine (BUP)10
Propoxyphene (PPX)300
Amphetamine (AMP)500 or 1,000
Phencyclidine (PCP)25
Methadone metabolite (EDDP)300
Oxycodone (OXY)100
Oxazepam (BZO)300
Nortriptyline (TCA)1,000
Secobarbital (BAR)300
Methadone (MTD)300
6-Monoacetylmorphine (6-MAM)10
Fentanyl (FYL)1
The single or multi-test cup offers any combination from above 1 to 17 drugs, but only one cut off concentration under same drug condition will be included per device.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. GC/MS or LC-MS/MS is the preferred confirmatory method.

It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.

Device Description

CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

AI/ML Overview

This document provides details on the performance characteristics of the CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup. Since this is an in vitro diagnostic device (specifically, a drug screening test), the acceptance criteria and study design are typically focused on analytical performance (accuracy, precision, analytical specificity) rather than a multi-reader multi-case (MRMC) comparative effectiveness study, which is more common for imaging AI. Similarly, "human readers improving with AI vs without AI" is not applicable here as the device is the test, not an aid to human interpretation of another modality.

Here's the breakdown based on the provided text:

Acceptance Criteria and Reported Device Performance

The core acceptance criterion for qualitative drug screening tests like this is accurate detection around a specific cutoff concentration. The reported performance demonstrates the device's ability to correctly classify samples as positive or negative relative to these cutoffs.

Table of Acceptance Criteria and Reported Device Performance (Analytical Precision/Reproducibility)

The "Acceptance Criteria" column represents the desired performance for a qualitative assay around its cutoff. For positive results, this means detecting drug concentrations above the cutoff, and for negative results, it means not detecting concentrations below the cutoff. The provided precision data shows the number of positive (+) and negative (-) results out of 50 tests for various concentrations relative to the cutoff. An ideal performance would show 100% positive for concentrations above cutoff and 100% negative for concentrations below, with roughly 50/50 split at the cutoff itself (due to inherent variability).

Drug (Cut-off ng/mL)Acceptance Criteria (Implicit for qualitative assay)Reported Device Performance (Accuracy as evidenced by reproducibility at various concentrations) Number of negative/positive results out of 50 tests. Values are aggregated across 3 lots where available.
MOP300All samples >cutoff should test positive; all samples <cutoff should test negative. Around cutoff, results vary.+100% cutoff: 0-/50+ (100% positive)
+75% cutoff: 0-/50+ (100% positive)
+50% cutoff: 0-/50+ (100% positive)
+25% cutoff: 1-2/48-49+ (96-98% positive)
cutoff: 23-27-/23-27+ (46-54% positive)
-25% cutoff: 49-50-/0-1+ (98-100% negative)
-50% cutoff: 50-/0+ (100% negative)
-75% cutoff: 50-/0+ (100% negative)
-100% cutoff: 50-/0+ (100% negative)
MET1000(Same as above)+100% cutoff: 0-/50+ (100% positive)
+75% cutoff: 0-/50+ (100% positive)
+50% cutoff: 0-/50+ (100% positive)
+25% cutoff: 0-1/49-50+ (98-100% positive)
cutoff: 25-26-/24-25+ (48-52% positive)
-25% cutoff: 49-50-/0-1+ (98-100% negative)
-50% cutoff: 50-/0+ (100% negative)
-75% cutoff: 50-/0+ (100% negative)
-100% cutoff: 50-/0+ (100% negative)
COC300(Same as above)+100% cutoff: 0-/50+ (100% positive)
+75% cutoff: 0-/50+ (100% positive)
+50% cutoff: 0-/50+ (100% positive)
+25% cutoff: 1/49+ (98% positive)
cutoff: 23-25-/25-27+ (46-54% positive)
-25% cutoff: 49-50-/0-1+ (98-100% negative)
-50% cutoff: 50-/0+ (100% negative)
-75% cutoff: 50-/0+ (100% negative)
-100% cutoff: 50-/0+ (100% negative)
(Similar detailed tables for all 20 analytes and two configurations would follow the same pattern as the MOP300, MET1000, and COC300 examples shown above, demonstrating consistent reproducibility around the cutoffs.)

Note: The implicit acceptance criterion for a qualitative test like this is generally that samples significantly above the cutoff should consistently yield positive results, samples significantly below the cutoff should consistently yield negative results, and samples near the cutoff (e.g., +/- 25% or 50% of the cutoff) will show varying results due to inherent assay variability, which is considered acceptable.

Study Details:

  1. Sample Size Used for the Test Set and Data Provenance:

    • Analytical Performance (Precision/Reproducibility): For each drug and each concentration point (9 concentration levels per drug), 50 tests were performed (2 runs per day for 25 days). Given there are 20 analytes (including the alternative cutoffs), this amounts to 20 drugs * 9 concentrations * 50 tests/concentration = 9000 tests.
    • Analytical Specificity/Interference: Not explicitly stated as a "test set" size with a fixed number of samples, but "drug metabolites and other components" were "spiked into drug-free urine" and tested using three lots of the device. For compounds showing no interference, they were tested at a "concentration of 100µg/mL or specified concentrations" in both drug-free urine and urine containing target drugs at +/- 50% cutoff. Over 100 compounds were listed.
    • Method Comparison Study: For each drug, 80 "unaltered urine clinical samples" were used (40 negative and 40 positive). These were "blind labeled." With 20 analytes, this sums to 20 drugs * 80 samples/drug = 1600 clinical samples.
    • Lay Person Study: 280 lay persons participated. Urine samples were prepared at 7 concentration levels (-100%, +/-75%, +/-50%, +/-25% of cutoff). Each participant received 1 blind-labeled sample and 1 device. The tables suggest that for each configuration (1 and 2), for each drug, 20 samples were tested at each concentration level. Thus, for Configuration 1, there are 17 drugs, so 17 drugs * 7 concentrations * 20 samples/conc = 2380 samples. For Configuration 2, there are 17 drugs, so 17 drugs * 7 concentrations * 20 samples/conc = 2380 samples.
    • Data Provenance: The analytical and method comparison studies were performed "in-house." The lay user study was performed "at three intended user sites." The origin of the urine samples (e.g., country of origin) is not specified. It is implied these are prospective tests using prepared or collected samples for the purpose of the study.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Analytical Precision/Reproducibility, Analytical Specificity/Interference, and Method Comparison: The ground truth for these studies was established by LC-MS/MS or GC/MS (or LC-MS/MS), which are reference analytical methods, not human expert consensus. The text states:
      • "Each drug concentration was confirmed by LC-MS/MS" for precision studies.
      • "The samples were…compared to LC-MS/MS results" for the method comparison study.
      • "The concentrations of the samples were confirmed by LC-MS/MS" for the lay person study.
    • Therefore, human experts were not directly establishing the ground truth for classification.
  3. Adjudication Method for the Test Set:

    • Not applicable as the ground truth was established by LC-MS/MS/GC/MS, a definitive chemical analysis method, not by human expert reading requiring adjudication. The device itself is an immunoassay, the results of which are compared to the LC-MS/MS gold standard.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study design is not applicable to a lateral flow immunoassay drug test. The device is a diagnostic test itself, not an AI assisting human interpretation of another modality.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The precision, specificity, and method comparison studies evaluate the performance of the device itself (the immunoassay) against confirmed concentrations (LC-MS/MS), which represents its "standalone" analytical performance. However, there is a human element in reading the qualitative bands (positive/negative line), which is addressed in the lay-person study.
  6. The Type of Ground Truth Used:

    • The primary ground truth used across all analytical studies (precision, specificity, method comparison, lay person study) was LC-MS/MS or GC/MS results. This is considered a highly accurate and definitive chemical confirmatory method for drug concentrations in urine.
  7. The Sample Size for the Training Set:

    • This device is a lateral flow immunoassay, not a machine learning/AI algorithm that requires a "training set" in the computational sense. Its "training" is inherent in its chemical and biological design. Therefore, this question is not applicable.
  8. How the Ground Truth for the Training Set was Established:

    • As this is not an AI/ML device relying on a training set, this question is not applicable. The device's performance is governed by its chemical design (antibodies, reagents) and manufacturing process, which are developed and validated through iterative biochemical and engineering studies, not by a data-driven training process in the AI sense.

FDA 510(k) Clearance Letter - CLUNGENE Multi-Drug Test Easy Cup

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 27, 2025

Hangzhou Clongene Biotech Co.,Ltd.
℅ Jenny Xia
Director
LSI International Inc
504 E Diamond Ave, Suite H
Gaithersburg, Maryland 20877

Re: K252118
Trade/Device Name: CLUNGENE Multi-Drug Test Easy Cup; CLUNGENE Multi-Drug Home Test Easy Cup
Regulation Number: 21 CFR 862.3100
Regulation Name: Amphetamine Test System
Regulatory Class: Class II
Product Code: NFT, PTH, NGL, NFV, NFY, PTG, NGG, QBF, NGM, QAW, NFW
Dated: July 4, 2025
Received: July 7, 2025

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K252118 - Jenny Xia Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252118 - Jenny Xia Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Kotarek -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2025.08.27 12:37:03 -04'00'

Joseph Kotarek, Ph.D.
Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 2

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K252118

Device Name:
CLUNGENE Multi-Drug Test Easy Cup
CLUNGENE Multi-Drug Home Test Easy Cup

Indications for Use (Describe)

The CLUNGENE Multi-Drug Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:

DrugCalibratorCut-off (ng/mL)
Morphine (MOP/OPI300)Morphine300
Morphine (MOP/OPI2000)Morphine2,000
Methamphetamine (mAMP/MET1000)D-Methamphetamine1,000
Methamphetamine (mAMP/MET500)D-Methamphetamine500
Cocaine (COC300)Benzoylecgonine300
Cocaine (COC150)Benzoylecgonine150
Marijuana (THC)11-nor-9-THC-9-COOH50
Methylenedioxymethamphetamine (MDMA)D,L-Methylenedioxymethamphetamine500
Buprenorphine (BUP)Buprenorphine10
Propoxyphene (PPX)D-Propoxyphene300
Amphetamine (AMP1000)D-Amphetamine1,000
Amphetamine (AMP500)D-Amphetamine500
Phencyclidine (PCP)Phencyclidine25
Methadone metabolite (EDDP)2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine300
Oxycodone (OXY)Oxycodone100
Oxazepam (BZO)Oxazepam300
Nortriptyline (TCA)Nortriptyline1,000
Secobarbital (BAR)Secobarbital300
Methadone (MTD)Methadone300
6-Monoacetylmorphine (6-MAM)6-Monoacetylmorphine10
Fentanyl (FYL)Fentanyl1

The single or multi-test cups can consist of any combination of the above listed drug analytes, but only one cut off concentration under same drug condition will be included per device.

This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Evaluate preliminary positive results carefully. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

The CLUNGENE Multi-Drug Home Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:

Drug (Identifier)Cut-off (ng/mL)
Morphine (MOP/OPI2000)300 or 2000
Methamphetamine (mAMP/MET)500 or 1,000

Page 5

FORM FDA 3881 (8/23) Page 2 of 2

DrugCut-off (ng/mL)
Cocaine (COC)150 or 300
Marijuana (THC)50
Methylenedioxymethamphetamine (MDMA)500
Buprenorphine (BUP)10
Propoxyphene (PPX)300
Amphetamine (AMP)500 or 1,000
Phencyclidine (PCP)25
Methadone metabolite (EDDP)300
Oxycodone (OXY)100
Oxazepam (BZO)300
Nortriptyline (TCA)1,000
Secobarbital (BAR)300
Methadone (MTD)300
6-Monoacetylmorphine (6-MAM)10
Fentanyl (FYL)1

The single or multi-test cup offers any combination from above 1 to 17 drugs, but only one cut off concentration under same drug condition will be included per device.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. GC/MS or LC-MS/MS is the preferred confirmatory method.

It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 6

510(k) SUMMARY

K252118

  1. Date: August 3, 2025

  2. Submitter: Hangzhou Clongene Biotech Co., Ltd.
    No.1 Yichuang Road, Yuhang District.
    Hongzhou, 311121
    China

  3. Contact person: Jenny Xia
    LSI International Inc.
    504 East Diamond Ave., Suite H
    Gaithersburg, MD 20877
    Telephone: 301-525-6856
    Email: jxia@lsi-consulting.org

  4. Device Name: CLUNGENE Multi-Drug Test Easy Cup
    CLUNGENE Multi-Drug Home Test Easy Cup

  5. Classification: Class II

Product CodeTarget DrugRegulation SectionPanel
NFTAmphetamine (AMP)862.3100, Amphetamine Test SystemToxicology
PTHSecobarbital (BAR)862.3150, Barbiturate Test SystemToxicology
NGLBuprenorphine (BUP)Morphine (MOP/OPI)Oxycodone (OXY)6-Monoacetylmorphine (6-MAM)Fentanyl (FYL)862.3650, Opiate Test SystemToxicology
NFVOxazepam (BZO)862.3170, Benzodiazepine Test SystemToxicology
NFYCocaine (COC)862.3250, Cocaine and cocaine metabolite test systemToxicology
PTG2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)Methadone (MTD)862.3620, Methadone Test SystemToxicology
NGGMethylenedioxymethamphetamine (MDMA)Methamphetamine (MET)862.3610, Methamphetamine Test SystemToxicology
NGMPhencyclidine (PCP)UnclassifiedToxicology

Page 7

Product CodeTarget DrugRegulation SectionPanel
QBFPropoxyphene (PPX)862.3700 Propoxyphene test system.Toxicology
QAWNortriptyline (TCA)862.3910 Tricyclic antidepressant drugs test systemToxicology
NFWCannabinoids (THC)862.3870, Cannabinoids Test SystemToxicology
  1. Predicate Devices:
    AllTest Multi-Drug Urine Test Cup (K250727)

  2. Intended Use

The CLUNGENE Multi-Drug Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:

DrugCalibratorCut-off (ng/mL)
Morphine (MOP/OPI300)Morphine300
Morphine (MOP/OPI2000)Morphine2,000
Methamphetamine (mAMP/MET1000)D-Methamphetamine1,000
Methamphetamine (mAMP/MET500)D-Methamphetamine500
Cocaine (COC300)Benzoylecgonine300
Cocaine (COC150)Benzoylecgonine150
Marijuana (THC)11-nor-Δ9-THC-9-COOH50
Methylenedioxymethamphetamine (MDMA)D,L-Methylenedioxymethamphetamine500
Buprenorphine (BUP)Buprenorphine10
Propoxyphene (PPX)D-Propoxyphene300
Amphetamine (AMP1000)D-Amphetamine1,000
Amphetamine (AMP500)D-Amphetamine500
Phencyclidine (PCP)Phencyclidine25
Methadone metabolite (EDDP)2-Ethylidine-1,5-dimethyl-3,3-diphenylpyrrolidine300
Oxycodone (OXY)Oxycodone100
Oxazepam (BZO)Oxazepam300
Nortriptyline (TCA)Nortriptyline1,000
Secobarbital (BAR)Secobarbital300
Methadone (MTD)Methadone300
6-Monoacetylmorphine (6-MAM)6-Monoacetylmorphine10
Fentanyl (FYL)Fentanyl1

The single or multi-test cups can consist of any combination of the above listed drug analytes, but only one cut off concentration under same drug condition will be included per device.

Page 8

This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Evaluate preliminary positive results carefully. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

The CLUNGENE Multi-Drug Home Test Easy Cup is a lateral flow immunoassay for the qualitative detection of Morphine, Methamphetamine, Cocaine, Marijuana, Methylenedioxymethamphetamine, Buprenorphine, Propoxyphene, Amphetamine, Phencyclidine, EDDP (Methadone metabolite), Oxycodone, Oxazepam, Nortriptyline, Secobarbital, Methadone, 6-Monoacetylmorphine and Fentanyl in human urine at the following cut off concentrations:

Drug (Identifier)Cut-off (ng/mL)
Morphine (MOP/OPI2000)300 or 2000
Methamphetamine (mAMP/MET)500 or 1,000
Cocaine (COC)150 or 300
Marijuana (THC)50
Methylenedioxymethamphetamine (MDMA)500
Buprenorphine (BUP)10
Propoxyphene (PPX)300
Amphetamine (AMP)500 or 1,000
Phencyclidine (PCP)25
Methadone metabolite (EDDP)300
Oxycodone (OXY)100
Oxazepam (BZO)300
Nortriptyline (TCA)1,000
Secobarbital (BAR)300
Methadone (MTD)300
6-Monoacetylmorphine (6-MAM)10
Fentanyl (FYL)1

The single or multi-test cup offers any combination from above 1 to 17 drugs, but only one cut off concentration under same drug condition will be included per device.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. GC/MS or LC-MS/MS is the preferred confirmatory method.

It is intended for over-the-counter (OTC) use. For in vitro diagnostic use only

  1. Device Description

CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

Page 9

The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

  1. Substantial Equivalence Information
ItemDevicePredicate (K250727)
Intended useQualitative detection of drugs of abuse in urine. For over-the-counter use.Same
MethodologyCompetitive binding, lateral flow immunochromatographic assay based on antigen-antibody reactionSame
Type of TestQualitativeSame
Specimen TypeHuman urineSame
Target Drug and Cut Off ValuesTarget Drugs: Cutoff (ng/mL)Same except for the
Amphetamine(AMP): 1000 or 500TRA
Secobarbital (BAR): 300
Buprenorphine (BUP): 10
Oxazepam (BZO): 300
Cocaine (COC): 150 or 300
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP): 300
Methylenedioxymethamphetamine (MDMA): 500
Methamphetamine (MET): 1000 or 500
Morphine (MOP300/OPI2000): 300 or 2000
Methadone (MTD): 300
Oxycodone (OXY): 100
Phencyclidine (PCP): 25
Propoxyphene(PPX): 300
Nortriptyline (TCA): 1000
Cannabinoids (THC): 50
6-Monoacetylmorphine(6-MAM): 10
Fentanyl (FYL): 1
ConfigurationsTest cupSame
  1. Standard/Guidance Document Reference (if applicable)

Page 10

None referenced.

  1. Test Principle

CLUNGENE Multi-Drug Test Easy Cup or CLUNGENE Multi-Drug Home Test Easy Cup is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody conjugate binding sites.

When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.

When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.

To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.

  1. Performance Characteristics

A. Analytical performance

a. Precision/Reproducibility:

Precision studies were carried out for samples with concentrations of +100% cutoff, +75% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Other samples were prepared by spiked target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test cups. The results obtained are summarized in the following tables:

DrugLot Number+100% cut off+75% cut off+50% cut off+25% cut offcut off-25% cut off-50% cut off-75% cut off-100% cut off
MOP300Lot10-/50+0-/50+0-/50+2-/48+23-/27+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+27-/23+49-/1+50-/0+50-/0+50-/0+
MET1000Lot10-/50+0-/50+0-/50+0-/50+25-/25+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+26-/24+50-/0+50-/0+50-/0+50-/0+
COC300Lot10-/50+0-/50+0-/50+1-/49+24-/26+50-/0+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+25-/25+49-/1+50-/0+50-/0+50-/0+

Page 11

DrugLot Number+100% cut off+75% cut off+50% cut off+25% cut offcut off-25% cut off-50% cut off-75% cut off-100% cut off
Lot30-/50+0-/50+0-/50+1-/49+23-/27+49-/1+50-/0+50-/0+50-/0+
THCLot10-/50+0-/50+0-/50+1-/49+24-/26+48-/2+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+26-/24+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+2-/48+24-/26+50-/0+50-/0+50-/0+50-/0+
MDMALot10-/50+0-/50+0-/50+1-/49+25-/25+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+0-/50+25-/25+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+2-/48+24-/26+50-/0+50-/0+50-/0+50-/0+
BUPLot10-/50+0-/50+0-/50+1-/49+25-/25+48-/2+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+27-/23+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+25-/25+49-/1+50-/0+50-/0+50-/0+
PPXLot10-/50+0-/50+0-/50+1-/49+26-/24+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+23-/27+48-/2+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+25-/25+49-/1+50-/0+50-/0+50-/0+
AMP1000Lot10-/50+0-/50+0-/50+2-/48+24-/26+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+22-/28+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+25-/25+48-/2+50-/0+50-/0+50-/0+
PCPLot10-/50+0-/50+0-/50+1-/49+26-/24+48-/2+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+2-/48+26-/24+48-/2+50-/0+50-/0+50-/0+
EDDPLot10-/50+0-/50+0-/50+2-/48+26-/24+48-/2+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+22-/28+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+26-/24+50-/0+50-/0+50-/0+50-/0+
OXYLot10-/50+0-/50+0-/50+1-/49+25-/25+48-/2+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+23-/27+48-/2+50-/0+50-/0+50-/0+
BZOLot10-/50+0-/50+0-/50+1-/49+26-/24+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+2-/48+23-/27+49-/1+50-/0+50-/0+50-/0+
TCALot10-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+21-/29+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+
BARLot10-/50+0-/50+0-/50+1-/49+25-/25+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+24-/26+49-/1+50-/0+50-/0+50-/0+
MTDLot10-/50+0-/50+0-/50+0-/50+24-/26+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+25-/25+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+2-/48+23-/27+50-/0+50-/0+50-/0+50-/0+
6-MAMLot10-/50+0-/50+0-/50+1-/49+26-/24+48-/2+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+2-/48+27-/23+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+22-/28+49-/1+50-/0+50-/0+50-/0+
FYLLot10-/50+0-/50+0-/50+1-/49+22-/28+49-/1+50-/0+50-/0+50-/0+

Page 12

DrugLot Number+100% cut off+75% cut off+50% cut off+25% cut offcut off-25% cut off-50% cut off-75% cut off-100% cut off
Lot20-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+24-/26+50-/0+50-/0+50-/0+50-/0+
MET500Lot10-/50+0-/50+0-/50+1-/49+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+0-/50+26-/24+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
COC150Lot10-/50+0-/50+0-/50+2-/48+26-/24+48-/2+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+2-/48+22-/28+48-/2+50-/0+50-/0+50-/0+
OPI2000Lot10-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+1-/49+23-/27+49-/1+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+23-/27+50-/0+50-/0+50-/0+50-/0+
AMP500Lot10-/50+0-/50+0-/50+1-/49+26-/24+49-/1+50-/0+50-/0+50-/0+
Lot20-/50+0-/50+0-/50+2-/48+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot30-/50+0-/50+0-/50+1-/49+23-/27+48-/2+50-/0+50-/0+50-/0+

b. Linearity/assay reportable range:

Not applicable. This device is intended for qualitative use only.

c. Stability:

The device is stable at 4-30℃ for 24 months based on the accelerated stability study.

d. Analytical specificity/Interference:

To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine. These urine samples were tested using three lots of the device.

Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100.

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
MOP 300Morphine300100%
6-Monoacetylmorphine (6-MAM)300100%
Codeine300100%
Dihydrocodeine1,00030%
Ethylmorphine1,00030%
Codeine-6-β-D-glucuronide50060%
Heroin300100%
Hydrocodone50,0000.6%
Hydromorphone1,00030%

Page 13

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Levorphanol tartrate1,50020%
Nalorphine HCl1,00030%
Norcodeine2,00015%
Normorphine3,00010%
Oxymorphone10,0003%
Morphine-6-β-D-glucuronide1,00030%
Morphine-3-β-D-glucuronide1,00030%
Thebaine3,00010%
6-Acetylcodeine8,0003.8%
Oxycodone100,000 (Negative)Not detected
Procaine100,000 (Negative)Not detected
Norpropoxyphene100,000 (Negative)Not detected
MET 1000D-Methamphetamine1,000100%
(+/-)3,4-Methylenedioxy-nethylamphetamine (MDEA)20,0005%
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)3,00033.3%
L-Methamphetamine20,0005%
p-Hydroxymethamphetamine30,0003.3%
D,L-Methamphetamine1,000100%
β-Phenylethylamine50,0002%
Mephentermine50,0002%
Methoxyphenamine HCl50,0002%
L-Amphetamine75,0001.3%
D-Amphetamine50,0002%
D,L-Amphetamine100,000 (negative)Not detected
Chloroquine100,000 (negative)Not detected
L-Ephedrine100,000 (negative)Not detected
D-Pseudoephedrine100,000 (negative)Not detected
(+/-)3,4-Methylenedioxyamphetamine (MDA)100,000 (negative)Not detected
Trimethobenzamide100,000 (negative)Not detected
L-Phenylephrine100,000 (negative)Not detected
Procaine100,000 (negative)Not detected
Phentermine100,000 (negative)Not detected
Fenflurmaine50,0002%
COC300Benzoylecogonine300100%
Cocaethylene12,5002.4%
Cocaine HCl78038.5%
Ecgonine32,0000.9%
Norcocaine100,000 (Negative)Not detected

Page 14

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Ecgonine methyl ester100,000 (Negative)Not detected
THC 5011-nor-Δ9-THC-9-COOH50100%
11-nor-Δ8-THC -9-COOH30166.7%
(±)-11-nor-9-carboxy-Δ9-THC10050%
11-nor-Δ9-THC -carboxy glucuronide10050%
(±)-11-Hydroxy-Δ9-THC5,0001%
Δ8-Tetrahydrocannabinol1,5003.3%
Δ9-Tetrahydrocannabinol5,0001%
Cannabinol20,0000.3%
Cannabidiol100,000 (Negative)Not detected
MDMA 500(+/-)3,4-Methylenedioxymethamphetamine (MDMA)500100%
(+/-)3,4-Methylenedioxyamphetamine (MDA)3,00016.7%
(+/-)3,4-Methylenedioxy-nethylamphetamine (MDEA)300166.7%
L-Methamphetamine100,000 (Negative)Not detected
D-Methamphetamine100,000 (Negative)Not detected
D,L-Methamphetamine100,000 (Negative)Not detected
D-Amphetamine100,000 (Negative)Not detected
L-Amphetamine100,000 (Negative)Not detected
D,L-Amphetamine100,000 (Negative)Not detected
L-Ephedrine100,000 (Negative)Not detected
D-Pseudoephedrine100,000 (Negative)Not detected
Phentermine100,000 (Negative)Not detected
L-Phenylephrine100,000 (Negative)Not detected
BUP 10Buprenorphine10100%
Buprenorphine -3-D-glucuronide1606.3%
Norbuprenorphine10100%
Norbuprenorphine-3-D-glucuronide2005%
Morphine100,000 (Negative)Not detected
Oxymorphone100,000 (Negative)Not detected
Hydromorphone100,000 (Negative)Not detected
Codeine100,000 (Negative)Not detected
Nalorphine HCl100,000 (Negative)Not detected
PPX 300D-Propoxyphene300100%
D-Norpropoxyphene300100%
AMP 1000D-Amphetamine1,000100%
L-Amphetamine50,0002%
D,L-Amphetamine3,00033.3%
Hydroxyamphetamine5,00020%

Page 15

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxyamphetamine (MDA)5,00020%
Diethylstilbestrol4,00025%
Phentermine3,00033.3%
β-Phenylethylamine100,0001%
Tyramine100,0001%
D,L-Norephedrine100,0001%
p-Hydroxynorephedrine100,0001%
p-Hydroxyamphetamine100,0001%
D-Methamphetamine100,000 (Negative)Not detected
L-Methamphetamine100,000 (Negative)Not detected
D,L-Methamphetamine100,000 (Negative)Not detected
L-Ephedrine100,000 (Negative)Not detected
D-Pseudoephedrine100,000 (Negative)Not detected
Phenylpropanolamine100,000 (Negative)Not detected
L-Phenylephrine100,000 (Negative)Not detected
L-Epinephrine100,000 (Negative)Not detected
D,L-Epinephrine100,000 (Negative)Not detected
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)100,000 (Negative)Not detected
(+/-)3,4-Methylenedioxy-nethylamphetamine (MDEA)100,000 (Negative)Not detected
PCP 25Phencyclidine25100%
4-Hydroxy phencyclidine12,5000.2%
Pheniramine40,0000.06%
EDDP 3002-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300100%
Methadone100,000 (Negative)Not detected
EMDP100,000 (Negative)Not detected
Doxylamine100,000 (Negative)Not detected
Levacetylmethadol (LAAM)100,000 (Negative)Not detected
Disopyramide100,000 (Negative)Not detected
Alphamethadol100,000 (Negative)Not detected
OXY 100Oxycodone100100%
Hydrocodone10,0001%
Hydromorphone12,5000.8%
Levorphanol tartrate20,0000.5%
Naloxone HCl15,0000.7%
Naltrexone HCl50,0000.2%
Oxymorphone1,5006.7%
Oxymorphone-3-β-D-glucuronide1,00010%

Page 16

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Noroxycodone2,0005%
Noroxymorphone1,00010%
Dihydrocodeine20,0000.5%
Codeine100,000 (Negative)Not detected
Morphine100,000 (Negative)Not detected
Buprenorphine100,000 (Negative)Not detected
Ethylmorphine100,000 (Negative)Not detected
Thebaine100,000 (Negative)Not detected
6-Monoacetylmorphine (6-MAM)100,000 (Negative)Not detected
BZO 300Oxazepam300100%
α-Hydroxyalprazolam1,50020%
Alprazolam200150%
Bromazepam50060%
Chlordiazepoxide1,50020%
Clobazam100300%
Clonazepam80037.5%
Clorazepate dipotassium200150%
Desalkylflurazepam40075%
Diazepam200150%
Estazolam2,50012%
Flunitrazepam40075%
D,L-Lorazepam1,50020%
Midazolam12,5002.4%
Nitrazepam100300%
Norchlordiazepoxide200150%
Nordiazepam40075%
R,S-Lorazepam glucuronide200150%
Temazepam100300%
Triazolam2,50012%
Demoxepam2,00015%
Flurazepam50060%
Delorazepam2,00015%
Lormetazepam150200%
TCA 1000Nortriptyline1,000100%
Amitriptyline1,50066.7%
Chlorpheniramine30,0003.3%
Clomipramine12,5008%
Cyclobenzaprine HCl5,00020%
Desipramine200500%
Doxepin2,00050%

Page 17

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Duloxetine80,0001.3%
Imipramine750133.3%
Norclomipramine12,5008%
Nordoxepine1,000100%
Promazine1,50066.7%
Trimipramine5,00020%
Maprotiline2,00050%
Promethazine25,0004%
BAR 300Secobarbital300100%
Alphenal150200%
Amobarbital300100%
Aprobarbital200150%
Butabarbital100300%
Butethal100300%
Cyclopentobarbital60050%
Pentobarbital250120%
Phenobarbital250120%
Butalbital2,50012%
Barbital300100%
MTD 300Methadone300100%
Doxylamine5,0006%
EDDP100,000 (Negative)Not detected
Levacetylmethadol (LAAM)100,000 (Negative)Not detected
EMDP100,000 (Negative)Not detected
Alphamethadol100,000 (Negative)Not detected
Pheniramine100,000 (Negative)Not detected
Disopyramide100,000 (Negative)Not detected
Esomeprazole100,000 (Negative)Not detected
6-MAM 106-Monoacetylmorphine (6-MAM)10100%
Heroin10100%
Morphine100,0000.01%
Nalorphine HCl5,0000.2%
Hydrocodone50,0000.02%
Hydromorphone10,0000.1%
Oxymorphone50,0000.02%
Procaine50,0000.02%
Thebaine10,0000.1%
Normorphine100,000 (Negative)Not detected
Chlordiazepoxide100,000 (Negative)Not detected
Clobazam100,000 (Negative)Not detected

Page 18

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
D-Amphetamine100,000 (Negative)Not detected
D,L-Amphetamine100,000 (Negative)Not detected
Levorphanol tartrate100,000 (Negative)Not detected
Codeine100,000 (Negative)Not detected
Ethylmorphine100,000 (Negative)Not detected
Morphine-3-β-D-glucuronide100,000 (Negative)Not detected
Norcodeine100,000 (Negative)Not detected
Oxycodone100,000 (Negative)Not detected
6-Acetylcodeine100,000 (Negative)Not detected
Buprenorphine100,000 (Negative)Not detected
Dihydrocodeine100,000 (Negative)Not detected
Dextromethorphan100,000 (Negative)Not detected
Imipramine100,000 (Negative)Not detected
Meperidine100,000 (Negative)Not detected
(±)-Methadone100,000 (Negative)Not detected
Mitragynine (Kratom)100,000 (Negative)Not detected
Morphine-6-β-D-glucuronide100,000 (Negative)Not detected
Naloxone HCl100,000 (Negative)Not detected
Naltrexone HCl100,000 (Negative)Not detected
Naproxen100,000 (Negative)Not detected
Norbuprenorphine100,000 (Negative)Not detected
Norbuprenorphine-3-D-glucuronide100,000 (Negative)Not detected
Noroxycodone HCl100,000 (Negative)Not detected
Noroxymorphone HCl100,000 (Negative)Not detected
(+)-Norpropoxyphene maleate100,000 (Negative)Not detected
Oxymorphone-3-β-D-glucuronide100,000 (Negative)Not detected
Tapentadol HCl100,000 (Negative)Not detected
Tramadol HCl100,000 (Negative)Not detected
Levacetylmethadol (LAAM)100,000 (Negative)Not detected
FYL 1Fentanyl1100%
Acetyl fentanyl1100%
Acrylfentanyl1100%
Isobutyryl fentanyl2.540%
Ocfentanil520%
Butyryl fentanyl520%
Furanyl fentanyl1010%
Valeryl fentanyl520%
(±) β-Hydroxythiofentanyl2.540%
4-Fluoro-isobutyrylfentanyl1010%
Para-fluorobutyryl fentanyl520%

Page 19

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Para-fluoro fentanyl2.540%
Carfentanil502%
Sufentanil254%
Alfentanil75000.01%
ω-1-Hydroxy fentanyl25000.04%
(±)-3-cis-methyl fentanyl751.3%
Despropionyl fentanyl (4-ANPP)20000.05%
β-hydroxyfentanyl1001%
Thiofentanyl502%
Cyclopropyl fentanyl1010%
Trazodone10000.1%
Remifentanil100,000 (Negative)Not detected
Norcarfentanil100,000 (Negative)Not detected
Norfentanyl100,000 (Negative)Not detected
Acetyl norfentanyl100,000 (Negative)Not detected
6-Acetyl morphine100,000 (Negative)Not detected
Amphetamine100,000 (Negative)Not detected
Buprenorphine100,000 (Negative)Not detected
Buprenorphine-3-D-glucuronide100,000 (Negative)Not detected
Codeine100,000 (Negative)Not detected
Dextromethorphan100,000 (Negative)Not detected
Dihydrocodeine100,000 (Negative)Not detected
EDDP100,000 (Negative)Not detected
EMDP100,000 (Negative)Not detected
Fluoxetine100,000 (Negative)Not detected
Heroin100,000 (Negative)Not detected
Hydrocodone100,000 (Negative)Not detected
Hydromorphone100,000 (Negative)Not detected
Ketamine100,000 (Negative)Not detected
Levorphanol100,000 (Negative)Not detected
Meperidine100,000 (Negative)Not detected
Methadone100,000 (Negative)Not detected
Morphine100,000 (Negative)Not detected
Morphine-3-β-D-glucuronide100,000 (Negative)Not detected
Naloxone100,000 (Negative)Not detected
Naltrexone100,000 (Negative)Not detected
Norbuprenorphine100,000 (Negative)Not detected
Norcodeine100,000 (Negative)Not detected
Norketamine100,000 (Negative)Not detected
Normeperidine100,000 (Negative)Not detected

Page 20

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Normorphine100,000 (Negative)Not detected
Noroxycodone100,000 (Negative)Not detected
Oxycodone100,000 (Negative)Not detected
Oxymorphone100,000 (Negative)Not detected
Pentazocine (Talwin)100,000 (Negative)Not detected
Pipamperone100,000 (Negative)Not detected
Risperidone100,000 (Negative)Not detected
Tapentadol100,000 (Negative)Not detected
Thioridazine100,000 (Negative)Not detected
Tilidine100,000 (Negative)Not detected
Tramadol100,000 (Negative)Not detected
Tramadol-N-Desmethyl100,000 (Negative)Not detected
Tramadol-O-Desmethyl100,000 (Negative)Not detected
Isotonitaze100,000 (Negative)Not detected
AH-7921 HCl100,000 (Negative)Not detected
MET 500D-Methamphetamine500100%
(+/-)3,4-Methylenedioxy-nethylamphetamine (MDEA)10,0005%
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)1,50033.3%
L-Methamphetamine10,0005%
p-Hydroxymethamphetamine15,0003.3%
D,L-Methamphetamine500100%
β-Phenylethylamine25,0002%
Mephentermine25,0002%
Methoxyphenamine HCl25,0002%
L-Amphetamine37,5001.3%
D-Amphetamine25,0002%
D,L-Amphetamine100,000 (negative)Not detected
Chloroquine100,000 (negative)Not detected
L-Ephedrine100,000 (negative)Not detected
D-Pseudoephedrine100,000 (negative)Not detected
(+/-)3,4-Methylenedioxyamphetamine (MDA)100,000 (negative)Not detected
Trimethobenzamide100,000 (negative)Not detected
L-Phenylephrine100,000 (negative)Not detected
Procaine100,000 (negative)Not detected
Phentermine100,000 (negative)Not detected
Fenflurmaine25,0002%
COC 150Benzoylecogonine150100%
Cocaethylene6,2502.4%

Page 21

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Cocaine HCl34044.1%
Ecgonine16,0000.9%
Norcocaine100,000 (Negative)Not detected
Ecgonine methyl ester100,000 (Negative)Not detected
OPI 2000Morphine2000100%
6-Monoacetylmorphine (6-MAM)2,000100%
Codeine2,000100%
Dihydrocodeine1,500133.3%
Ethylmorphine3,00066.7%
Codeine-6-β-D-glucuronide3,00066.7%
Heroin2,000100%
Hydrocodone12,50016%
Hydromorphone5,00040%
Levorphanol tartrate75,0002.7%
Nalorphine HCl5,00040%
Norcodeine8,00025%
Normorphine50,0004%
Oxymorphone25,0008%
Morphine-6-β-D-glucuronide2,000100%
Morphine-3-β-D-glucuronide2,000100%
Thebaine5,00040%
6-Acetylcodeine50,0004%
Oxycodone100,000 (Negative)Not detected
Procaine100,000 (Negative)Not detected
Norpropoxyphene100,000 (Negative)Not detected
AMP 500D-Amphetamine500100%
L-Amphetamine25,0002%
D,L-Amphetamine1,50033.3%
Hydroxyamphetamine2,50020%
(+/-)3,4-Methylenedioxyamphetamine (MDA)2,50020%
Diethylstilbestrol2,00025%
Phentermine1,50033.3%
β-Phenylethylamine50,0001%
Tyramine50,0001%
D,L-Norephedrine50,0001%
p-Hydroxynorephedrine50,0001%
p-Hydroxyamphetamine50,0001%
D-Methamphetamine100,000 (Negative)Not detected
L-Methamphetamine100,000 (Negative)Not detected
D,L-Methamphetamine100,000 (Negative)Not detected

Page 22

Drug/Cut offCompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
L-Ephedrine100,000 (Negative)Not detected
D-Pseudoephedrine100,000 (Negative)Not detected
Phenylpropanolamine100,000 (Negative)Not detected
L-Phenylephrine100,000 (Negative)Not detected
L-Epinephrine100,000 (Negative)Not detected
D,L-Epinephrine100,000 (Negative)Not detected
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)100,000 (Negative)Not detected
(+/-)3,4-Methylenedioxy-nethylamphetamine (MDEA)100,000 (Negative)Not detected
Benzphetamine100,000 (Negative)Not detected

To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.

Compounds that show no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following table.

CompoundCompoundCompound
AcetaminophenDoxylamine (except MTD test)Octopamine
Acetone (1000 mg/dL)D-PseudoephedrineO-Hydroxyhippuric acid
AcetophenetidinDuloxetine (except TCA test)Olanzapine
Acetylsalicylic acid (Aspirin)Ecgonine methyl esterOmeprazole
AcyclovirEMDPOxalic acid (100 mg/dL)
Albumin (100 mg/dL )ErythromycinOxolinic acid
Albuterol sulfate (Proair HFA)Esomeprazole magnesiumOxymetazoline
AlphamethadolEstronePaliperidone
AminophyllineEthanol (1%)Papaverine
AminopyrineFenofibratePenicillin G
AmitriptylineFenoprofenPenicillin V potassium
AmoxicillinFentanyl (except FYL test)Perphenazine
AmpicillinFluoxetine HClPhenacetin
AripiprazoleFluphenazinePhenelzine
AspartameFotemustinePhenylpropanolamine
AtomoxetineFurosemidePrednisone
Atorvastatin CalciumGabapentinPregablin
AtropineGalactose (10 mg/dL)Procaine (except 6-MAM test)
AzithromycinGamma globulin (500 mg/dL)Promazine (except TCA test)

Page 23

CompoundCompoundCompound
BaclofenGatifloxacinPromethazine (except TCA test)
Benzilic acidGemfibrozilPropoxyphene (except PPX test)
BenzocaineGentisic acidPropranolol
Benzoic acidGlibenclamidePropylthiouracil
BenzoylecgonineGliclazidePyridoxine
BenzphetamineGlucose (3000 mg/dL)Pyrilamine
BilirubinGuaiacol glyceryl etherPyrogallol
Boric acid (1%)HemoglobinQuetiapine
BupropionHydralazineQuinine
CaffeineHydrochlorothiazideQuinolinic acid
CannabidiolHydrocortisoneR-(-)-Apomorphine
CaptoprilIbuprofenRanitidine
CarbamazepineImipramine (except TCA test)Ribavirin
Carfentanil (except FYL test)IsoxsuprineRifampicin
CarisoprodolKetamineRisperidone
CefradineKetoprofenSalicylic acid
CephalexinLAAM HClSerotonin (5-Hydroxytyramine)
Chloral hydrateLabetalolSertraline
ChloramphenicolL-EphedrineSildenafil citrate
Chlordiazepoxide (except BZO test)L-EpinephrineSimvastatin
ChloroquineLevofloxacin HClSulfamethazine
ChlorothiazideLevonorgestrelSulindac
Chlorpheniramine (except TCA test)Levothyroxine SodiumTelmisartan
ChlorpromazineLidocaine HClTetracycline
CholesterolLisinoprilTetrahydrocortisone 3-(βDglucuronide)
Ciprofloxacin HClLoperamideTetrahydrocortisone 3-acetate
CitalopramLoratadineTetrahydrozoline
ClarithromycinL-PhenylephrineTheophylline
ClonidineMagnesiumThiamine
ClozapineMaprotiline (except TCA test)Thioridazine
Conjugated EstrogensMeperidineTramadol
CortisoneMeprobamateTrazodone (except FYL test)
Creatine HydrateMethapyrileneTriamterene
CreatinineMethaqualoneTrifluoperazine

Page 24

CompoundCompoundCompound
Cyclobenzaprine HCl (except TCA test)Methoxyphenamine (except MET test)Trimethobenzamide
CyclodextrinMethylphenidateTrimethoprim
D,L-EpinephrineMetoprolol tartrateTryptamine
D,L-IsoproterenolMetronidazole (300 μg/mL)Tyramine (except AMP test)
D,L-OctopamineMifepristoneUrea (2000 mg/dL)
D,L-PropranololMontelukast sodiumUric acid
D,L-TryptophanN-AcetylprocainamideValproic acid (250 μg/mL)
D,L-TyrosineNaCl (4000 mg/dL)Venlafaxine HCl
Delorazepam (except BZO test)Nalidixic acidVerapamil
DeoxycorticosteroneNaloxone HCl (except OXY test)Vitamin B2 (Riboflavin)
DesloratadineNaltrexone HCl (except OXY test)Vitamin C (Ascorbic acid)
DextromethorphanNiacinamideZaleplon
DiclofenacNicotineZomepirac sodium salt
Diclofenac sodiumNicotinic acidβ-Estradiol
DiflunisalNifedipineβ-Phenylethylamine (except MET/AMP test)
DigoxinNitroglycerin(-)-Cotinine
Diphenhydramine HClNordoxepin (except TCA test)(+)-Naproxen
DisopyramideNorethindrone3-Hydroxytyramine
Dopamine HClNorfentanyl4-Dimethyl-aminoantipyrine
Doxepin (except TCA test)Noscapine5, 5-Diphenylhydantoin

Interference by pH and specific gravity were also evaluated using pooled urine specimens with drug concentrations at 50% below and 50% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.

B. Method comparison study

The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below:

Page 25

Drug TestOperatorTest ResultDrug-FreeLow Negative by LC-MS/MS (less than -50% cut off)Near cut off Negative by LC-MS/MS (Between -50%cut off and cut off)Near cut off Positive by LC-MS/MS (Between cut off and +50% cut off)High Positive by LC-MS/MS (greater than +50%cut off)
MOP300Operator 1+0011821
-10151410
Operator 2+0001821
-10151510
Operator 3+0011921
-10151400
MET1000Operator 1+0012218
-10141500
Operator 2+0002118
-10141610
Operator 3+0012118
-10141510
COC300Operator 1+0011821
-10151410
Operator 2+0011821
-10151410
Operator 3+0011921
-10151400
THCOperator 1+0002218
-10141600
Operator 2+0002118
-10141610
Operator 3+0002118
-10141610
MDMAOperator 1+0001920
-10151510
Operator 2+0011920
-10151410
Operator 3+0001920
-10151510
BUPOperator 1+0012218
-10141500
Operator 2+0012118
-10141510

Page 26

Drug TestOperatorTest ResultDrug-FreeLow Negative by LC-MS/MS (less than -50% cut off)Near cut off Negative by LC-MS/MS (Between -50%cut off and cut off)Near cut off Positive by LC-MS/MS (Between cut off and +50% cut off)High Positive by LC-MS/MS (greater than +50%cut off)
Operator 3+0012118
-10141510
PPXOperator 1+0011920
-10141510
Operator 2+0011920
-10141510
Operator 3+0011920
-10141510
AMP1000Operator 1+0011920
-10151410
Operator 2+0001920
-10151510
Operator 3+0002020
-10151500
PCPOperator 1+0002118
-10141610
Operator 2+0012118
-10141510
Operator 3+0002118
-10141610
EDDPOperator 1+0012119
-10141500
Operator 2+0002119
-10141600
Operator 3+0002019
-10141610
OXYOperator 1+0011920
-10151410
Operator 2+0012020
-10151400
Operator 3+0011920
-10151410
BZOOperator 1+0011920
-10151410

Page 27

Drug TestOperatorTest ResultDrug-FreeLow Negative by LC-MS/MS (less than -50% cut off)Near cut off Negative by LC-MS/MS (Between -50%cut off and cut off)Near cut off Positive by LC-MS/MS (Between cut off and +50% cut off)High Positive by LC-MS/MS (greater than +50%cut off)
Operator 2+0012020
-10151400
Operator 3+0001920
-10151510
TCAOperator 1+0001920
-10141610
Operator 2+0012020
-10141500
Operator 3+0001920
-10141610
BAROperator 1+0022218
-10141400
Operator 2+0002118
-10141610
Operator 3+0012118
-10141510
MTDOperator 1+0012019
-10141510
Operator 2+0012119
-10141500
Operator 3+0002019
-10141610
6-MAMOperator 1+0012118
-10141510
Operator 2+0002118
-10141610
Operator 3+0002118
-10141610
FYLOperator 1+0002118
-10151510
Operator 2+0022218
-10151300
Operator 3+0012018
-10151420

Page 28

Drug TestOperatorTest ResultDrug-FreeLow Negative by LC-MS/MS (less than -50% cut off)Near cut off Negative by LC-MS/MS (Between -50%cut off and cut off)Near cut off Positive by LC-MS/MS (Between cut off and +50% cut off)High Positive by LC-MS/MS (greater than +50%cut off)
MET500Operator 1+0012019
-10141510
Operator 2+0002019
-10141610
Operator 3+0012019
-10141510
COC150Operator 1+0012118
-10151410
Operator 2+0002118
-10151510
Operator 3+0012118
-10151410
OPI2000Operator 1+0012118
-10151410
Operator 2+0002018
-10151520
Operator 3+0012218
-10151400
AMP500Operator 1+0012118
-10151410
Operator 2+0012118
-10151410
Operator 3+0012118
-10151410

Discordant Results are summarized below.

Drug TestOperatorSample NumberLC-MS/MS Result (ng/mL)Test Result
MOP300Operator 11-08269.32Positive
Operator 31-33289.66Positive
Operator 11-60336.93Negative
Operator 21-17310.02Negative
MET1000Operator 14-79992.48Positive
Operator 34-38897.14Positive
Operator 24-631018.46Negative

Page 29

Drug TestOperatorSample NumberLC-MS/MS Result (ng/mL)Test Result
Operator 34-741007.84Negative
COC300Operator 16-32296.35Positive
Operator 26-63295.68Positive
Operator 36-79272.56Positive
Operator 16-51306.47Negative
Operator 26-68300.47Negative
THCOperator 27-0453.73Negative
Operator 37-3257.78Negative
MDMAOperator 28-07496.27Positive
Operator 18-48508.66Negative
Operator 28-20547.97Negative
Operator 38-42511.26Negative
BUPOperator 1,29-459.86Positive
Operator 39-209.66Positive
Operator 29-7210.23Negative
Operator 39-5010.66Negative
PPXOperator 1,311-21297.14Positive
Operator 211-11293.10Positive
Operator 111-02302.98Negative
Operator 311-65322.25Negative
Operator 211-71309.47Negative
AMP1000Operator 113-08870.13Positive
Operator 113-071070.87Negative
Operator 213-231013.55Negative
PCPOperator 214-3124.88Positive
Operator 114-6925.58Negative
Operator 214-4928.12Negative
Operator 314-2725.47Negative
EDDPOperator 115-21296.15Positive
Operator 315-51305.99Negative
OXYOperator 116-2891.36Positive
Operator 216-5199.55Positive
Operator 316-7796.15Positive
Operator 116-74110.47Negative
Operator 316-75104.20Negative
BZOOperator 117-61273.48Positive
Operator 217-04293.44Positive
Operator 117-37307.46Negative
Operator 317-69319.25Negative
TCAOperator 218-65886.47Positive
Operator 1,318-261003.13Negative

Page 30

Drug TestOperatorSample NumberLC-MS/MS Result (ng/mL)Test Result
BAROperator 119-48288.49Positive
Operator 319-76298.54Positive
Operator 119-39294.45Positive
Operator 219-33303.65Negative
Operator 319-62315.16Negative
MTDOperator 120-65292.16Positive
Operator 220-79289.43Positive
Operator 1,320-27303.66Negative
6-MAMOperator 121-568.94Positive
Operator 121-0610.55Negative
Operator 221-2310.55Negative
Operator 321-0811.46Negative
FYLOperator 210-690.95Positive
Operator 210-170.87Positive
Operator 310-680.96Positive
Operator 110-211.04Negative
Operator 310-071.12Negative
Operator 310-201.02Negative
MET500Operator 13-58458.22Positive
Operator 33-68493.26Positive
Operator 13-35515.17Negative
Operator 23-71539.96Negative
Operator 33-78503.12Negative
COC150Operator 1,35-38145.59Positive
Operator 15-07156.87Negative
Operator 25-71151.49Negative
Operator 35-61164.90Negative
OPI2000Operator 12-671834.41Positive
Operator 32-041976.54Positive
Operator 12-502179.28Negative
Operator 22-102039.78Negative
Operator 22-772034.78Negative
AMP500Operator 112-75491.37Positive
Operator 212-04473.85Positive
Operator 312-70489.38Positive
Operator 112-55510.94Negative
Operator 312-79553.25Negative
Operator 212-68506.47Negative

C. Lay person study

Page 31

A lay user study was performed at three intended user sites with 280 lay persons. 176 male and 104 female participated the study. They had diverse educational and professional backgrounds and their age range from 21 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

Result of CLUNGENE Multi-Drug Test Easy Cup Configuration 1:

| Drug | Cut off (ng/mL) | Results | Concentration |||||||||
|------|-----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| MOP | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| MET | 1000 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | 100% |
| COC | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| THC | 50 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 100% | 100% | 100% |
| MDMA | 500 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% |
| BUP | 10 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% |
| PPX | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | 100% |
| AMP | 1000 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |

Page 32

| Drug | Cut off (ng/mL) | Results | Concentration |||||||||
|------|-----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | 100% |
| PCP | 25 | Negative | 20 | 20 | 20 | 20 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| EDDP | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| OXY | 100 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | 100% |
| BZO | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| TCA | 1000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| BAR | 300 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | 100% |
| MTD | 300 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 100% | 100% | 100% |
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% |
| FYL | 1 | Negative | 20 | 20 | 20 | 20 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 100% | 100% | 100% |

Result of CLUNGENE Multi-Drug Test Easy Cup Configuration 2:

| Drug | Cut off (ng/mL) | Results | Concentration |||||||||
|------|-----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| MET | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |

Page 33

| Drug | Cut off (ng/mL) | Results | Concentration |||||||||
|------|-----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| COC | 150 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| THC | 50 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| OPI | 2000 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% |
| MDMA | 500 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% |
| BUP | 10 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| PPX | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| AMP | 500 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | 100% |
| PCP | 25 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | 100% |
| EDDP | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | 100% |
| OXY | 100 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | 100% |

Page 34

| Drug | Cut off (ng/mL) | Results | Concentration |||||||||
|------|-----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| BZO | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| TCA | 1000 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 100% | 100% | 100% |
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |
| MTD | 300 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90% | 100% | 100% | 100% |
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | 100% |
| FYL | 1 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% |

Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 8.

Clinical Studies:
Not applicable.

  1. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that CLUNGENE Multi-Drug Test Easy Cup and CLUNGENE Multi-Drug Home Test Easy Cup are substantially equivalent to the predicate devices.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).