K Number
K252867

Validate with FDA (Live)

Date Cleared
2025-10-09

(30 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

AbbreviationAnalyteCalibratorCut-off (ng/mL)
6AM6-Acetylmorphine6-Monoacetylmorphine10
AMPAmphetamined-Amphetamine500 or 1000
BARSecobarbitalSecobarbital300
BUPBuprenorphineBuprenorphine10
BZOOxazepamOxazepam300
COCCocaineBenzoylecgonine150 or 300
EDDPEDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FENFentanylFentanyl1
FENNorfentanylNorfentanyl5
MDMAEcstasyd,l-Methylenedioxymethamphetamine500
METMethamphetamined-Methamphetamine500 or 1000
MORMorphineMorphine300
MTDMethadoned/l-Methadone300
OPIOpiatesMorphine2000
OXYOxycodoneOxycodone100
PCPPhencyclidinePhencyclidine25
PPXPropoxyphenePropoxyphene300
TCANortriptylineNortriptyline1000
THCMarijuana11-nor-Δ9-THC-9 COOH20 or 50

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This in vitro diagnostic device provides only a preliminary test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

DrugCut-off (ng/mL)Abbreviation
6-monoacetylmorphine106AM
Amphetamine500 / 1000AMP
Secobarbital300BAR
Buprenorphine10BUP
Oxazepam300BZO
Cocaine150 / 300COC
EDDP300EDDP
Fentanyl1FEN
Norfentanyl5FEN
Ecstasy500MDMA
Methamphetamine500 / 1000MET
Methadone300MTD
Opiates300 / 2000MOR/OPI
Oxycodone100OXY
Phencyclidine25PCP
Propoxyphene300PPX
Nortriptyline1000TCA
Marijuana20 / 50THC

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This device provides only a preliminary test result. A more specific alternate method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

Device Description

VINScreen Urine Drug Test Cup and VINScreen Urine Drug Home Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - VINScreen Urine Drug Test Cup

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.01

October 9, 2025

Advin Biotech, Inc.
℅ Jenny Xia
Director, Regulatory Consultant
LSI International Inc
504 E Diamond Ave.
Suite H
Gaithersburg, Maryland 20877

Re: K252867
Trade/Device Name: VINScreen Urine Drug Test Cup; VINScreen Urine Drug Home Test Cup
Regulation Number: 21 CFR 862.3100
Regulation Name: Amphetamine test system
Regulatory Class: Class II
Product Code: NFT, PTH, NGL, NFV, NFY, PTG, NGG, NGM, QBF, QAW, NFW
Dated: September 9, 2025
Received: September 9, 2025

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Page 2

K252867 - Jenny Xia Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252867 - Jenny Xia Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2025.10.09 08:08:29 -04'00'

Joseph Kotarek
Branch Chief for Toxicology
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 2

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K252867

Device Name: VINScreen Urine Drug Test Cup; VINScreen Urine Drug Home Test Cup

Indications for Use (Describe)

The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

AbbreviationAnalyteCalibratorCut-off (ng/mL)
6AM6-Acetylmorphine6-Monoacetylmorphine10
AMPAmphetamined-Amphetamine500 or 1000
BARSecobarbitalSecobarbital300
BUPBuprenorphineBuprenorphine10
BZOOxazepamOxazepam300
COCCocaineBenzoylecgonine150 or 300
EDDPEDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
FENFentanylFentanyl1
FENNorfentanylNorfentanyl5
MDMAEcstasyd,l-Methylenedioxymethamphetamine500
METMethamphetamined-Methamphetamine500 or 1000
MORMorphineMorphine300
MTDMethadoned/l-Methadone300
OPIOpiatesMorphine2000
OXYOxycodoneOxycodone100
PCPPhencyclidinePhencyclidine25
PPXPropoxyphenePropoxyphene300
TCANortriptylineNortriptyline1000
THCMarijuana11-nor-Δ9-THC-9 COOH20 or 50

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This in vitro diagnostic device provides only a preliminary test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

DrugCut-off (ng/mL)Abbreviation
6-monoacetylmorphine106AM
Amphetamine500 / 1000AMP
Secobarbital300BAR
Buprenorphine10BUP
Oxazepam300BZO
Cocaine150 / 300COC
EDDP300EDDP
Fentanyl1FEN
Norfentanyl5FEN
Ecstasy500MDMA
Methamphetamine500 / 1000MET

Page 5

FORM FDA 3881 (8/23) Page 2 of 2

DrugCut-off (ng/mL)Abbreviation
Methadone300MTD
Opiates300 / 2000MOR/OPI
Oxycodone100OXY
Phencyclidine25PCP
Propoxyphene300PPX
Nortriptyline1000TCA
Marijuana20 / 50THC

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This device provides only a preliminary test result. A more specific alternate method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 6

510(k) SUMMARY

K252867

  1. Date: October 5, 2025

  2. Submitter: Advin Biotech, Inc.
    10140 Mesa Rim Rd Unit 1
    San Diego, CA 92121

  3. Contact person: Jenny Xia
    LSI International Inc.
    504 East Diamond Ave., Suite H
    Gaithersburg, MD 20877
    Telephone: 301-525-6856
    Email: jxia@lsi-consulting.org

  4. Device Name: VINScreen Urine Drug Test Cup
    VINScreen Urine Drug Home Test Cup

  5. Classification: Class II

Product CodeTarget DrugRegulation SectionPanel
NFTAmphetamine (AMP)862.3100, Amphetamine Test SystemToxicology
PTHSecobarbital (BAR)862.3150, Barbiturate Test SystemToxicology
NGLBuprenorphine (BUP), Fentanyl (FEN), Morphine (MOP/OPI), Oxycodone (OXY), 6-Monoacetylmorphine(6-AM), Norfentanyl (FEN)862.3650, Opiate Test SystemToxicology
NFVOxazepam (BZO)862.3170, Benzodiazepine Test SystemToxicology
NFYCocaine (COC)862.3250, Cocaine and cocaine metabolite test systemToxicology
PTG2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Methadone (MTD)862.3620, Methadone Test SystemToxicology
NGGMethylenedioxymethamphetamine (MDMA), Methamphetamine (MET)862.3610, Methamphetamine Test SystemToxicology
QBFPropoxyphene(PPX)862.3700 Propoxyphene test system.Toxicology

Page 7

Product CodeTarget DrugRegulation SectionPanel
NGMPhencyclidine (PCP)UnclassifiedToxicology
QAWNortriptyline (TCA)862.3910 Tricyclic antidepressant drugs test systemToxicology
NFWCannabinoids (THC)862.3870, Cannabinoids Test SystemToxicology
  1. Predicate Devices:
    AllTest Multi-Drug Rapid Test Cup (K244043)

  2. Intended Use

The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

DrugCut-off (ng/mL)Abbreviation
6-monoacetylmorphine106AM
Amphetamine500 / 1000AMP
Secobarbital300BAR
Buprenorphine10BUP
Oxazepam300BZO
Cocaine150 / 300COC
EDDP300EDDP
Fentanyl1FEN
Norfentanyl5FEN
Ecstasy500MDMA
Methamphetamine500 / 1000MET
Methadone300MTD
Opiates300 / 2000MOR/OPI
Oxycodone100OXY
Phencyclidine25PCP
Propoxyphene300PPX
Nortriptyline1000TCA
Marijuana20 / 50THC

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This device provides only a preliminary test result. A more specific alternate method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

Page 8

The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

AbbreviationAnalyteCalibratorCutoff (ng/mL)
6AM6-Acetylmorphine6-Monoacetylmorphine10
AMPAmphetamined-Amphetamine500 / 1000
BARSecobarbitalSecobarbital300
BUPBuprenorphineBuprenorphine10
BZOOxazepamOxazepam300
COCCocaineBenzoylecgonine150 / 300
EDDPEDDP2-ethylidene-1,5-dimethyl-3-3-diphenylpyrrolidine300
FENFentanylFentanyl1
FENNorfentanylNorfentanyl5
MDMAEcstasyd,l-Methylenedioxy methamphetamine500
METMethamphetamined-Methamphetamine500 / 1000
MORMorphineMorphine300
MTDMethadoned/l-Methadone300
OPIOpiatesMorphine2000
OXYOxycodoneOxycodone100
PCPPhencyclidinePhencyclidine25
PPXPropoxyphenePropoxyphene300
TCANortriptylineNortriptyline1000
THCMarijuana11-nor-Δ⁹-THC-9-COOH20 / 50

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This in vitro diagnostic device provides only a preliminary test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

  1. Device Description

VINScreen Urine Drug Test Cup and VINScreen Urine Drug Home Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

  1. Substantial Equivalence Information

Similarities

ItemDevicePredicate (K244043)
Intended useQualitative detection of drugs of abuse in urine. For over-the-counter useSame.
MethodologyCompetitive binding, lateral flow immunochromatographic assay based on antigen-antibody reactionSame

Page 9

ItemDevicePredicate (K244043)
Type of TestQualitativeSame
Specimen TypeHuman urineSame
Target Drug and Cut Off ValuesTarget Drugs: Amphetamine(AMP) 1000 or 500, Secobarbital (BAR) 300, Buprenorphine (BUP) 10, Oxazepam (BZO) 300, Cocaine (COC) 150 or 300, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) 300, Methylenedioxymethamphetamine (MDMA) 500, Methamphetamine (MET) 1000 or 500, Morphine (MOP300/OPI2000) 300 or 2000, Methadone (MTD) 300, Oxycodone (OXY) 100, Phencyclidine (PCP) 25, Propoxyphene(PPX) 300, Nortriptyline (TCA) 1000, Cannabinoids (THC) 20 or 50, 6-Monoacetylmorphine(6-AM) 10, Fentanyl (FEN) 1, Norfentanyl (FEN) 5Same except that no Norfentanyl THC20
ConfigurationsTest cupSame
  1. Standard/Guidance Document Reference (if applicable)
    None referenced.

  2. Test Principle

The VINScreen Urine Drug Test Cup or The VINScreen Urine Drug Home Test Cup is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.

When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized

Page 10

in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.

When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.

To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.

  1. Performance Characteristics

A. Analytical performance

a. Precision/Reproducibility:

Precision studies were carried out for samples with concentrations of +100% cutoff, +75% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Samples with concentration of -100% cutoff were drug-free urines samples. Other samples were prepared by spiked target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test cups with one operator per lot. The results obtained are summarized in the following tables.

DrugLot Number+100% cutoff+75% cutoff+50% cutoff+25% cutoffCutoff-25% cutoff-50% cutoff-75% cutoff-100% cut-off
6AM 10Lot 10-/50+0-/50+0-/50+0-/50+8-/42+49-/1+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
AMP 500Lot 10-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+7-/43+50-/0+50-/0+50-/0+50-/0+
BAR 300Lot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
BUP 10Lot 10-/50+0-/50+0-/50+0-/50+12-/38+48-/2+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
BZO 300Lot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
COC 150Lot 10-/50+0-/50+0-/50+0-/50+7-/43+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+7-/43+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+

Page 11

DrugLot Number+100% cutoff+75% cutoff+50% cutoff+25% cutoffCutoff-25% cutoff-50% cutoff-75% cutoff-100% cut-off
EDDP 300Lot 10-/50+0-/50+0-/50+2-/48+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+1-/49+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
FEN 1Lot 10-/50+0-/50+0-/50+2-/48+13-/37+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+2-/48+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+48-/2+50-/0+50-/0+50-/0+
NFEN 5Lot 10-/50+0-/50+0-/50+2-/48+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+49-/1+50-/0+50-/0+50-/0+
MDMA 500Lot 10-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
MET 500Lot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
MOR 300Lot 10-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
MTD 300Lot 10-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
OXY 100Lot 10-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
PCP 25Lot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
PPX 300Lot 10-/50+0-/50+0-/50+0-/50+7-/43+49-/1+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+7-/43+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
TCA 1000Lot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
THC 20Lot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
AMP 1000Lot 10-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
COC 300Lot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+

Page 12

DrugLot Number+100% cutoff+75% cutoff+50% cutoff+25% cutoffCutoff-25% cutoff-50% cutoff-75% cutoff-100% cut-off
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
MET 1000Lot 10-/50+0-/50+0-/50+0-/50+7-/43+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+7-/43+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
OPI 2000Lot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
THC 50Lot 10-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+

b. Linearity/assay reportable range:

Not applicable. This device is intended for qualitative use only.

c. Stability:

The device is stable at 2-30℃ for 24 months based on accelerated stability study.

d. Analytical specificity/Interference:

To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine. These urine samples were tested using three lots of the device. The results obtained are summarized in the following table. Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100.

6-AM (6-Acetylmorphine)
(6-Acetylmorphine, Cutoff=10ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
6-Acetylmorphine10100%
Heroin2540%
Morphine>100,000<0.01%
Normorphine>100,000<0.01%
Nalorphine>100,000<0.01%
Hydrocodone>100,000<0.01%
Hydromorphone>100,000<0.01%
Chlordiazepoxide>100,000<0.01%
Clobazam>100,000<0.01%
D-Amphetamine>100,000<0.01%
(±)-Amphetamine>100,000<0.01%
Levorphanol>100,000<0.01%
Codeine>100,000<0.01%
Ethylmorphine>100,000<0.01%
Morphine3-β-D-glucuronide>100,000<0.01%

Page 13

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Norcodeine>100,000<0.01%
Oxycodone>100,000<0.01%
Oxymorphone>100,000<0.01%
Procaine>100,000<0.01%
Thebaine>100,000<0.01%
6-Acetylcodeine50,0000.02%
Buprenorphine>10,000<0.1%
Dihydrocodeine>100,000<0.01%
Dextromethorphan>100,000<0.01%
Imipramine>100,000<0.01%
Meperidine>100,000<0.01%
(±)-Methadone>100,000<0.01%
Mitragynine(Kratom)>10,000<0.1%
Morphine-6-β-D-glucuronide>10,000<0.1%
Naloxone>100,000<0.01%
Naltrexone>100,000<0.01%
Naproxen>100,000<0.01%
Norbuprenorphine>10,000<0.1%
Norbuprenorphine-3-D-Glucuronide10,0000.1%
Noroxycodone>100,000<0.01%
Noroxymorphone>10,000<0.1%
Norpropoxyphene>100,000<0.01%
Oxymorphone-3β-D-glucuronide>10,000<0.1%
Tapentadol>100,000<0.01%
cis-Tramadol>100,000<0.01%

AMP(Amphetamine) (Amphetamine, Cutoff=500ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-3,4-Methylenedioxyamphetamine (MDA)2,00025%
(±)-Amphetamine1,00050%
D-Amphetamine500100%
Diethylstilbestrol>10,000<5%
L-Amphetamine50,0001%
Phentermine15,0003.33%
β-Phenylethylamine>100,000<0.5%
Tyramine>100,000<0.5%
p-Hydroxynorephedrine>100,000<0.5%
p-Hydroxyamphetamine5,00010%
D-Methamphetamine>100,000<0.5%
L-Methamphetamine>100,000<0.5%
Ephedrine>100,000<0.5%

Page 14

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-3,4-Methylenedioxymethamphetamine (MDMA)>100,000<0.5%
Phenylpropanolamine (Norephedrine)>100,000<0.5%
Benzphetamine>100,000<0.5%
(1R,2S)-(-)-Ephedrine>100,000<0.5%
L-Epinephrine>100,000<0.5%
D,L-Epinephrine>100,000<0.5%
(±)-3,4-Methylenedioxyethylamphetamine (MDEA)>100,000<0.5%

AMP(Amphetamine) (Amphetamine, Cutoff=1000ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-3,4-Methylenedioxyamphetamine (MDA)5,00020%
(±)-Amphetamine2,00050%
D-Amphetamine1,000100%
Diethylstilbestrol>10,000<10%
L-Amphetamine>100,000<1%
Phentermine25,0004%
β-Phenylethylamine>100,000<1%
Tyramine>100,000<1%
p-Hydroxynorephedrine>100,000<1%
p-Hydroxyamphetamine10,00010%
D-Methamphetamine>100,000<1%
L-Methamphetamine>100,000<1%
Ephedrine>100,000<1%
(±)-3,4-Methylenedioxymethamphetamine (MDMA)>100,000<1%
Phenylpropanolamine (Norephedrine)>100,000<1%
Benzphetamine>100,000<1%
(1R,2S)-(-)-Ephedrine>100,000<1%
L-Epinephrine>100,000<1%
D,L-Epinephrine>100,000<1%
(±)-3,4-Methylenedioxyethylamphetamine (MDEA)>100,000<1%

BAR(Secobarbital) (Secobarbital, Cutoff=300ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Alphenal40075%
Amobarbital50060%
Aprobarbital150200%

Page 15

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Barbital1,00030%
Butabarbital70428.57%
Butethal201500%
Cyclopentobarbital1,00030%
Pentobarbital200150%
Phenobarbital250120%
Secobarbital300100%
Butalbital10,0003%

BUP(Buprenorphine) (Buprenorphine, Cutoff=10ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Buprenorphine10100%
Buprenorphine -3-D-Glucuronide2050%
Norbuprenorphine1506.67%
Norbuprenorphine glucuronide5002%
Codeine>100,000<0.01%
Nalorphine>100,000<0.01%
Morphine>100,000<0.01%
Oxymorphone>100,000<0.01%
Hydromorphone>100,000<0.01%

BZO(Oxazepam) (Oxazepam, Cutoff=300ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
a-Hydroxyalprazolam5,0006%
Alprazolam250120%
Bromazepam60050%
Chlordiazepoxide50060%
Clobazam300100%
Clonazepam>100,000<0.3%
Clorazepate50060%
Desalkylflurazepam50060%
Diazepam1,00030%
Estazolam>100,000<0.3%
Flunitrazepam1,00030%
Lorazepam1,00030%
Lormetazepam>100,000<0.3%
Midazolam>100,000<0.3%
Nitrazepam250120%
Norchlordiazepoxide70428.57%
Nordiazepam1,00030%

Page 16

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Oxazepam300100%
Oxazepam glucuronide301000%
Lorazepam glucuronide100300%
Temazepam200150%
Triazolam5,0006%
Demoxepam251200%
Flurazepam>100,000<0.3%
Delorazepam>10,000<3%

COC(Benzoylecgonine) (Benzoylecgonine, Cutoff=150ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Benzoylecgonine150100%
Cocaethylene2,0007.5%
Cocaine>100,000<0.15%
Ecgonine>100,000<0.15%
Norcocaine>100,000<0.15%
Ecgonine methyl ester>100,000<0.15%

COC(Benzoylecgonine) (Benzoylecgonine, Cutoff=300ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Benzoylecgonine300100%
Cocaethylene3,00010%
Cocaine>100,000<0.3%
Ecgonine>100,000<0.3%
Norcocaine>100,000<0.3%
Ecgonine methyl ester>100,000<0.3%

EDDP (EDDP, Cutoff=300ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
EDDP300100%
Methadone>100,000<0.3%
EMDP>100,000<0.3%
Doxylamine>100,000<0.3%
Disopyramide>100,000<0.3%
LAAM(Levo-alpha-acetylmethadol) hydrochloride>100,000<0.3%
Alpha Methadol>100,000<0.3%

Page 17

FEN(Fentanyl) (Fentanyl, Cutoff=1ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Norfentanyl25,0000.004%
Fentanyl1100%
Acetyl fentanyl333.33%
Acetyl norfentanyl20,0000.005%
(±)-β-Hydroxythiofentanyl520%
Acryl fentanyl425%
Butyryl fentanyl520%
(±)-cis-3-methyl Fentanyl1100 %
Furanyl fentanyl1010%
para-Fluorobutyryl fentanyl (PFBF)1010%
para-Fluorofentanyl520%
9-Hydroxyrisperidone>100,000<0.001%
Alfentanil>100,000<0.001%
Isobutyryl fentanyl1010%
4-Fluoro-isobutyryl fentanyl520%
Norcarfentanil>10,000<0.01%
Remifentanil5000.2%
Valeryl fentanyl1010%
Thienyl Fentanyl303.33%
(±)-trans-3-methyl Fentanyl520%
Despropionyl fentanyl (4-ANPP)2500.4%
MT-45502%
Ocfentanil250%
Isotonitazene>100,000<0.001%
Cyclopropylfentanyl1010%
AH-7921 HCL25000.04%
Risperidone>100,000<0.001%
Sufentanil205%
Carfentanil520%
Labetalol>100,000<0.001%
Trazodone>100,000<0.001%
U-47700>100,000<0.001%
ω-1-Hydroxyfentanyl250.00%
6-Acetylmorphine>100,000<0.001%
Amphetamine>100,000<0.001%
Buprenorphine>100,000<0.001%
Buprenorphine -3-D-Glucuronide>100,000<0.001%
Codeine>100,000<0.001%
Dextromethorphan>100,000<0.001%

Page 18

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Dihydrocodeine>100,000<0.001%
EDDP>100,000<0.001%
EMDP>100,000<0.001%
Fluoxetine>100,000<0.001%
Heroin>100,000<0.001%
Hydrocodone>100,000<0.001%
Hydromorphone>100,000<0.001%
Ketamine>100,000<0.001%
Levorphanol>100,000<0.001%
Meperidine>100,000<0.001%
Methadone>100,000<0.001%
Morphine>100,000<0.001%
Morphine-3-β-D-glucuronide>100,000<0.001%
Naloxone>100,000<0.001%
Naltrexone>100,000<0.001%
Norbuprenorphine5,0000.02%
Norcodeine>100,000<0.001%
Norketamine>100,000<0.001%
Normeperidine>100,000<0.001%
Normorphine>100,000<0.001%
Noroxycodone>100,000<0.001%
Oxycodone>100,000<0.001%
Oxymorphone>100,000<0.001%
Pentazocine (Talwin)80,0000.00125%
Pipamperone5,0000.02%
Tapentadol>100,000<0.001%
Thioridazine>100,000<0.001%
Tilidine>100,000<0.001%
cis-Tramadol>100,000<0.001%
O-Desmethyl-cis-tramadol80,0000.00125%
N-Desmethyl-cis-tramadol>100,000<0.001%

NFEN(Norfentanyl) (Norfentanyl, Cutoff=5ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Norfentanyl5100 %
Fentanyl10,0000.05%
Acetyl fentanyl>10,000<0.05%
Acetyl norfentanyl2002.5%
(±)-β-Hydroxythiofentanyl>10,000<0.05%
Acryl fentanyl>10,000<0.05%
Butyryl fentanyl>10,000<0.05%

Page 19

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-cis-3-methyl Fentanyl>10,000<0.05%
Furanyl fentanyl>10,000<0.05%
para-Fluorobutyryl fentanyl (PFBF)10,0000.05%
para-Fluorofentanyl>10,000<0.05%
9-Hydroxyrisperidone>100,000<0.005%
Alfentanil5,0000.1%
Isobutyryl fentanyl>100,000<0.005%
4-Fluoro-isobutyryl fentanyl>100,000<0.005%
Norcarfentail>10,000<0.05%
Remifentanil>10,000<0.05%
Valeryl fentanyl>10,000<0.05%
Thienyl Fentanyl>100,000<0.005%
(±)-trans-3-methyl Fentanyl>10,000<0.05%
Despropionyl fentanyl (4-ANPP)>10,000<0.05%
MT-45>100,000<0.005%
Ocfentanil>10,000<0.05%
Isotonitazene>100,000<0.005%
Cyclopropylfentanyl>10,000<0.05%
AH-7921 HCL>100,000<0.005%
Risperidone>100,000<0.005%
Sufentanil>10,000<0.05%
Carfentanil>10,000<0.05%
Labetalol>100,000<0.005%
Trazodone>100,000<0.005%
U-47700>100,000<0.005%
ω-1-Hydroxyfentanyl>100,000<0.005%
6-Acetylmorphine>100,000<0.005%
Amphetamine>100,000<0.005%
Buprenorphine>100,000<0.005%
Buprenorphine -3-D-Glucuronide>100,000<0.005%
Codeine>100,000<0.005%
Dextromethorphan>100,000<0.005%
Dihydrocodeine>100,000<0.005%
EDDP>100,000<0.005%
EMDP>100,000<0.005%
Fluoxetine>100,000<0.005%
Heroin>100,000<0.005%
Hydrocodone>100,000<0.005%
Hydromorphone>100,000<0.005%
Ketamine>100,000<0.005%
Levorphanol>100,000<0.005%
Meperidine>100,000<0.005%
Methadone>100,000<0.005%

Page 20

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Morphine>100,000<0.005%
Morphine-3-β-D-glucuronide>100,000<0.005%
Naloxone>100,000<0.005%
Naltrexone>100,000<0.005%
Norbuprenorphine>100,000<0.005%
Norcodeine>100,000<0.005%
Norketamine>100,000<0.005%
Normeperidine>100,000<0.005%
Normorphine>100,000<0.005%
Noroxycodone>100,000<0.005%
Oxycodone>100,000<0.005%
Oxymorphone>100,000<0.005%
Pentazocine (Talwin)>100,000<0.005%
Pipamperone>100,000<0.005%
Tapentadol>100,000<0.005%
Thioridazine>100,000<0.005%
Tilidine>100,000<0.005%
cis-Tramadol>100,000<0.005%
O-Desmethyl-cis-tramadol>100,000<0.005%
N-Desmethyl-cis-tramadol>100,000<0.005%

MDMA(Ecstasy) (Ecstasy, Cutoff=500ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-3,4-Methylenedioxyethylamphetamine (MDEA)500100%
(±)-3,4-Methylenedioxyamphetamine (MDA)5,00010%
(±)-3,4-Methylenedioxymethamphetamine (MDMA)500100%
L-Methamphetamine>100,000<0.5%
7-Aminoclonazepam>100,000<0.5%
D-Methamphetamine>100,000<0.5%
D-Amphetamine>100,000<0.5%
L-Amphetamine>100,000<0.5%

MET(Methamphetamine) (Methamphetamine, Cutoff=500ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-3,4-Methylenedioxyethylamphetamine (MDEA)30,0001.67%

Page 21

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-3,4-Methylenedioxymethamphetamine (MDMA)5,00010%
D-Methamphetamine500100%
L-Methamphetamine15,0003.33%
Fenfluramine>100,000<0.5%
p-Hydroxymethamphetamine3,00016.67%
(±)-Methamphetamine500100%
β-Phenylethylamine>100,000<0.5%
Mephetermine10,0005 %
Methoxyphenamine>100,000<0.5%
L-Amphetamine>100,000<0.5%
D-Amphetamine>100,000<0.5%
(±)-Amphetamine>100,000<0.5%
Chloroquine>100,000<0.5%
Ephedrine>100,000<0.5%
(±)-3,4-Methylenedioxyamphetamine (MDA)>100,000<0.5%
Trimethobenzamide>100,000<0.5%
l-phenylephrine>100,000<0.5%
(1R,2S)-(-)-Ephedrine>100,000<0.5%
Procaine>100,000<0.5%

MET(Methamphetamine) (Methamphetamine, Cutoff=1000ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-3,4-Methylenedioxyethylamphetamine>100,000<1%
(±)-3,4-Methylenedioxymethamphetamine (MDMA)10,00010%
D-Methamphetamine1,000100%
L-Methamphetamine25,0004%
Fenfluramine>100,000<1%
p-Hydroxymethamphetamine5,00020%
(±)-Methamphetamine2,00050%
β-Phenylethylamine>100,000<1%
Mephetermine30,0003.33%
Methoxyphenamine>100,000<1%
L-Amphetamine>100,000<1%
D-Amphetamine>100,000<1%
(±)-Amphetamine>100,000<1%
Chloroquine>100,000<1%
Ephedrine>100,000<1%
(±)-3,4-Methylenedioxyamphetamine (MDA)>100,000<1%
Trimethobenzamide>100,000<1%

Page 22

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
l-phenylephrine>100,000<1%
(1R,2S)-(-)-Ephedrine>100,000<1%
Procaine>100,000<1%

MOR(Morphine) (Morphine, Cutoff=300ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
6-Acetylmorphine50060%
Codeine250120%
Dihydrocodeine50060%
Ethylmorphine300100%
Heroin50060%
Hydrocodone50060%
Hydromorphone50060%
Levorphanol5,0006%
Morphine300100%
Nalorphine3,00010%
Thebaine>100,000<0.3%
Morphine-3-β-d-glucuronide300100%
Codeine-6-β-D-glucuronide250120%
Morphine-6-β-D-glucuronide5,0006%
6-Acetylcodeine1,00030%
Normorphine>100,000<0.3%
Oxycodone>100,000<0.3%
Oxymorphone50,0000.6%
Norcodeine>100,000<0.3%
Procaine>100,000<0.3%
Norpropoxyphene>100,000<0.3%

OPI(Morphine) (Morphine, Cutoff=2000ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
6-Acetylmorphine10,00020%
Codeine10,00020%
Dihydrocodeine10,00020%
Ethylmorphine5,00040%
Heroin20,00010%
Hydrocodone25,0008%
Hydromorphone25,0008%
Levorphanol>100,000<2%
Morphine2,000100%

Page 23

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Nalorphine>100,000<2%
Thebaine>100,000<2%
Morphine-3-β-d-glucuronide50,0004%
Codeine-6-β-D-glucuronide2,50080%
Morphine-6-β-D-glucuronide500400%
6-Acetylcodeine10,00020%
Normorphine>100,000<2%
Oxycodone>100,000<2%
Oxymorphone>100,000<2%
Norcodeine>100,000<2%
Procaine>100,000<2%
Norpropoxyphene>100,000<2%

MTD(Methadone) (Methadone, Cutoff=300ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(±)-Methadone300100%
EDDP>100,000<0.3%
EMDP>100,000<0.3%
LAAM (Levo-alpha-acetylmethadol)10,0003%
Alpha Methadol100300%
Doxylamine50,0000.6%

OXY(Oxycodone) (Oxycodone, Cutoff=100ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Hydrocodone20,0000.5%
Hydromorphone25,0000.4%
Levorphanol50,0000.2%
Naloxone30,0000.33%
Naltrexone20,0000.5%
Oxycodone100100%
Oxymorphone20050%
Oxymorphone-3β-D-glucuronide75013.33%
Noroxycodone2,5004%
Noroxymorphone1,00010%
Dihydrocodeine>100,000<0.1%
Codeine>100,000<0.1%
Morphine>100,000<0.1%
Buprenorphine>10,000<0.1%
Ethylmorphine>100,000<0.1%
Thebaine>100,000<0.1%

Page 24

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
6-Acetylmorphine>100,000<0.1%

PCP(Phencyclidine) (Phencyclidine, Cutoff=25ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Phencyclidine25100%
4-Hydroxyphencyclidine>100,000<0.025%

PPX(Propoxyphene) (Propoxyphene, Cutoff=300ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Norpropoxyphene50060%
Propoxyphene300100%

TCA(Nortriptyline) (Nortriptyline, Cutoff=1000ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Amitriptyline10,00010%
Chlorpheniramine>100,000<1%
Clomipramine50,0002%
Cyclobenzaprine>100,000<1%
Desipramine500200%
Doxepine10,00010%
Duloxetine>100,000<1%
Imipramine50,0002%
Norclomipramine>100,000<1%
Nordoxepin5,00020%
Nortriptyline1,000100%
Promazine10,00010%
Trimipramine10,00010%
Maprotiline5,00020%
Promethazine>100,000<1%

THC(Marijuana) (Marijuana, Cutoff=20ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(-)-11-nor-Δ8-THC -9-COOH5040%
(-)-11-nor-9-carboxy-Δ9-THC20100%
(±)-11-nor-9-carboxy-Δ9-THC5040%
11-nor-Δ9-THC-carboxy glucuronide5,0000.4%

Page 25

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Δ8-THC8,0000.25%
Δ9-THC>10,000<0.2%
Cannabinol>10,000<0.2%
Cannabidiol>100,000<0.02%
(±)-11-hydroxy-Δ9-THC8,0000.25%

THC(Marijuana) (Marijuana, Cutoff=50ng/mL)

CompoundMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
(-)-11-nor-Δ8-THC -9-COOH7071.43%
(-)-11-nor-9-carboxy-Δ9-THC50100%
(±)-11-nor-9-carboxy-Δ9-THC15033.33%
11-nor-Δ9-THC-carboxy glucuronide7,50066.67%
Δ8-THC10,0000.5%
Δ9-THC>10,000<0.5%
Cannabinol>10,000<0.5%
Cannabidiol>100,000<0.05%
(±)-11-hydroxy-Δ9-THC10,0000.5%

To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.

Compounds that show no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following table.

CompoundCompoundCompound
(-) CotinineDiclofenacNorfentanyl (except FEN and NFEN tests)
(-)-11-nor-9-carboxy-Δ9-THC (except THC test)Diclofenac sodiumNorethindrone
4-Bromo-2,5-dimethoxyphenethylamineDiflunisalNorpropoxyphene (except PPX test)
7-AminoclonazepamDigoxinNorpseudoephedrine
7-AminoflunitrazepamDiphenhydramineNortriptyline(except TCA test)
7-AminonitrazepamDiphenylhydantoinNoscapine
AcetaminophenDisopyramideO-Hydroxyhippuric acid
Acetone (1000 mg/dL)Dopamine (3-Hydroxytyramine)Olanzapine
AcetophenetidinDoxepine (except TCA test)Omeprazole
AcyclovirDoxylamine(except MTD test)Oxalic acid (100mg/dL)
Albumin (100mg/dL)DuloxetineOxazepam (except BZO test)
AlbuterolEcgonine methyl esterOxazepam Glucuronide (except BZO test)
Alpha Methadol (except MTD test)EMDPOxolinic acid
AminophyllineEphedrineOxymetazoline
Aminopyrine (4-Dimethylaminoantipyrine)ErythromycinPaliperidone
Amitriptyline(except TCA test)Esomeprazole MagnesiumPapaverine

Page 26

CompoundCompoundCompound
AmlodipineEstronePenicillin-G
Amobarbital(except BAR test)Ethanol (1%)PenicillinV Potassium
AmoxicillinFenfluraminePerphenazine
AmpicillinFenofibratePhenacetin
ApomorphineFentanyl (except FEN and NFEN tests)Phencyclidine (except PCP test)
AripiprazoleFenoprofenPhenelzine
Ascorbic acid (Vitamin C)FluoxetinePheniramine
AspartameFluphenazinePhenobarbital (except BAR test)
Aspirin (Acetylsalicylic acid)FotemustinePhentermine (except AMP test)
AtomoxetineFurosemidePhenylethylamine
Atorvastatin CalciumGabapentinPhenylpropanolamine
AtropineGalactosePrednisone
AzithromycinGatifloxacinPregabalin
BaclofenGemfibrozilProcaine
Benzilic acidGentisic acidPromazine (except TCA test)
BenzocaineGuaiacolglyceryl etherPromethazine
Benzoic acidHemoglobinPropoxyphene (except PPX test)
Benzoylecgonine (except COC test)HexobarbitalPropranolol
BenzphetamineHydralazinePseudoephedrine
BenzylpiperiazineHydrochlorothiazidePyridoxine
BilirubinHydrocortisonePyrilamine
Boric Acid (1%)Hydroxybutyric AcidPyrogallol
BrompheniramineIbuprofenQuetiapine
BupropionImipramine (except TCA test)Quinidine
CaffeineIsoxsuprineQuinine
CannabidiolKetamineQuinolinic Acid
CaptoprilKetoprofenRanitidine
CarbamazepineL-EphedrineRiboflavin
Carfentanil (10ug/mL) (except FEN1 test)L-EpinephrineRisperidone
CarisoprodolL-phenylephrineRifampicin
CefradineL-thyroxineSalicylic acid
CephalexinLAAM (except MTD test)Secobarbital(except BAR test)
CetirizineLabetalolSerotonin (5- Hydroxytyramine)
Chloral hydrateLamotrigineSertraline
ChloramphenicolLevofloxacinSildenafil Citrate
Chlordiazepoxide (except BZO test)LevonorgestrelSimvastatin
ChloroquineLevothyroxine sodiumSodium Azide
ChlorothiazideLidocaineSulfamethazine
ChlorpheniramineLisinoprilSulindac
ChlorpromazineLoperamideTelmisartan
CholesterolLoratidineTetracycline
CiprofloxacinLorazepam Glucuronide (except BZO test)Tetrahydrocortisone 3-(βDglucuronide)
CitalopramLSDTetrahydrocortisone, 3-acetate
ClarithromycinMagnesium ChlorideTetrahydrozoline
ClofibrateMaprotiline(except TCA test)Theophylline
Clomipramine (except TCA test)MeperidineThiamine
ClonidineMeprobamateThioridazine

Page 27

CompoundCompoundCompound
ClozapineMetforminTramadol
Conjugated EstrogensMethapyrileneTrazodone
CortisoneMethaqualoneTriamterene
CreatineMethoxyphenamineTrifluoperazine
CreatinineMethylphenidateTrifluoromethylphenylpiperazine
CyclobenzaprineMetoprololTrimethobenzamide
CyproheptadineMetronidazoleTrimethoprim
D-(+)-Glucose (Glucose (3000 mg/dL)MifepristoneTryptamine
D,L-EpinephrineN-AcetylprocainamideTyramine
D,L-IsoproterenolN-desmethyl TapentadolUrea (2000 mg/dL)
D,L-Lorazepam (except BZO test)NaCl(4000 mg/dL)Uric acid
D,L-Octopamine (Octopamine)Nalidixic acidValproic acid (250 μg/mL)
D,L-PropranololNaloxone (except OXY test)Venlafaxine
D,L-TryptophanNaltrexone (except OXY test)Verapamil
D,L-TyrosineNaproxenVitamin B2
Delorazepam (10ug/mL)Nicotinamide (Niacinamide)Zaleplon
Demoxepam (except BZO test)NicotineZolpidem
DeoxycorticosteroneNicotinic AcidZomepirac
Desipramine (except TCA test)Nifedipineβ-Estradiol (Estradiol)
DesloratadineNitroglycerinγ-Cyclodextrin (Cyclodextrin)
DextromethorphanNordoxepin (except TCA test)γ-Globulin (500mg/dL)

Interference by pH and specific gravity were also evaluated using pooled urine specimens with concentrations of 0 (drug-free), at 50% below and 50% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.

B. Method comparison study

The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below.

Drug testTest Cup ResultDrugFreeLow Negative by LC-MS/MS (less than -50%)Near Cutoff Negative by LC-MS/MS (Between -50% and the Cutoff)Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%)High Positive by LCMS/MS (greater than +50%)
6AM10Operator A-1216111
+00115
Operator B-1216110
+00116

Page 28

Drug testTest Cup ResultDrugFreeLow Negative by LC-MS/MS (less than -50%)Near Cutoff Negative by LC-MS/MS (Between -50% and the Cutoff)Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%)High Positive by LCMS/MS (greater than +50%)
Operator C-1216110
+00116
AMP500Operator A-1216121
+00012
Operator B-1216122
+00011
Operator C-1216121
+00012
AMP1000Operator A-1217110
+00015
Operator B-1217111
+00014
Operator C-1217111
+00014
BAR300Operator A-813180
+00116
Operator B-813170
+00216
Operator C-813170
+00216
BUP10Operator A-1514101
+00117
Operator B-1514101
+00117
Operator C-1514101
+00117
BZO300Operator A-817130
+00212
Operator B-817140
+00112
Operator C-817130
+00212
COC150Operator A-1213142
+00113
Operator B-1213141
+00114
Operator C-1213141
+00114
COC300Operator A-1218101
+00015
Operator B-121891
+00115

Page 29

Drug testTest Cup ResultDrugFreeLow Negative by LC-MS/MS (less than -50%)Near Cutoff Negative by LC-MS/MS (Between -50% and the Cutoff)Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%)High Positive by LCMS/MS (greater than +50%)
Operator C-1218100
+00016
EDDP300Operator A-1116121
+00113
Operator B-1116131
+00013
Operator C-1116111
+00213
FEN1Operator A-915140
+00215
Operator B-915141
+00214
Operator C-915130
+00315
NFEN5Operator A-1016110
+00316
Operator B-1016121
+00215
Operator C-1016110
+00316
MDMA500Operator A-1020100
+00012
Operator B-102091
+00111
Operator C-102090
+00112
MET500Operator A-1314131
+00011
Operator B-1314120
+00112
Operator C-1314111
+00211
MET1000Operator A-1316111
+00014
Operator B-1316111
+00014
Operator C-1316111
+00014
MTD300Operator A-1114151
+00012
Operator B-1114140
+00113

Page 30

Drug testTest Cup ResultDrugFreeLow Negative by LC-MS/MS (less than -50%)Near Cutoff Negative by LC-MS/MS (Between -50% and the Cutoff)Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%)High Positive by LCMS/MS (greater than +50%)
Operator C-1114151
+00012
MOR300Operator A-1215120
+00114
Operator B-1215131
+00013
Operator C-1215131
+00013
OPI2000Operator A-1214140
+00018
Operator B-1214141
+00017
Operator C-1214142
+00016
OXY100Operator A-1115131
+00112
Operator B-1115131
+00112
Operator C-1115121
+00212
PCP25Operator A-1315100
+00214
Operator B-1315110
+00114
Operator C-1315111
+00113
PPX300Operator A-1115130
+00113
Operator B-1115130
+00113
Operator C-1115130
+00113
TCA1000Operator A-916140
+00114
Operator B-916130
+00214
Operator C-916140
+00114
THC20Operator A-1315122
+00014
Operator B-1315101
+00215

Page 31

Drug testTest Cup ResultDrugFreeLow Negative by LC-MS/MS (less than -50%)Near Cutoff Negative by LC-MS/MS (Between -50% and the Cutoff)Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%)High Positive by LCMS/MS (greater than +50%)
Operator C-1315111
+00115
THC50Operator A-1117120
+00014
Operator B-1117122
+00012
Operator C-1117121
+00013

Discordant Results are summarized below.

DrugOperatorSample NumberLC/MS/MS Result (ng/mL)Accurate Result
6AMAV6312.050-
BV4049.847+
A, CV9278.340+
AMP500B, CV157521.19-
AV269535.83-
BV326578.93-
AMP1000CV3791148.25-
BV6751107.17-
BAR300A, BV97282.40+
B, CV236287.92+
CV1080291.50+
BUP10A, CV47.940+
B, CV59711.440-
BV10728.880+
AV126711.620-
BZO300CV430252.97+
AV532288.67+
A, CV599296.95+
BV1101276.94+
COC150AV42185.78-
BV147177.26-
CV206141.23+
AV272172.68-
BV1051135.25+
AV1208126.14+
CV1245159.79-
COC300BV31308.60-
AV51319.05-
BV498268.11+
EDDP300AV193312.04-
CV367311.54-

Page 32

DrugOperatorSample NumberLC/MS/MS Result (ng/mL)Accurate Result
BV794308.49-
CV1176281.27+
A, CV1183239.61+
FEN1B, CV1580.908+
AV2530.874+
B, CV3960.911+
BV6631.211-
CV12140.936+
AV12780.910+
NFEN5A, CV3124.929+
A, B, CV3194.995+
A, B, CV10104.998+
BV12715.208-
MDMA500BV254411.27+
CV1275433.84+
BV1177576.19-
MET500A, CV464603.62-
CV713451.20+
B, CV949395.63+
MET1000BV6101095.29-
AV6921094.78-
CV10501139.22-
MTD300CV72327.41-
AV1095324.39-
BV1205235.96+
MOR300CV329345.89-
AV742251.00+
BV899361.11-
OPI2000BV6152073.55-
CV10842016.19-
CV12802099.08-
OXY100BV50488.02+
CV73198.20+
AV76493.25+
CV84291.44+
A, CV1009105.20-
BV1071119.87-
PCP25CV6730.01-
AV48422.14+
A, BV74524.06+
CV104823.30+
PPX300BV268267.51+
A, CV992268.19+

Page 33

DrugOperatorSample NumberLC/MS/MS Result (ng/mL)Accurate Result
TCA100AV105914.22+
CV201886.50+
BV452928.62+
BV861917.240+
THC20A, CV12421.11-
BV13322.19-
AV27122.55-
BV82618.33+
B, CV94318.76+
A, CV12421.11-
THC50BV73454.36-
B, CV84458.23-

C. Lay person study

A lay user study was performed at three intended user sites with 280 lay persons. 126 male and 154 female tested the VINScreen Urine Drug Test Cup. They had diverse educational and professional backgrounds and their age range from 18 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free pooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

Result of The VINScreen Urine Drug Test Cup: Configuration 1

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |
| AMP | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| BUP | 10 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |

Page 34

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| BZO | 300 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95.0% | 100% | 100% |
| COC | 150 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |
| EDDP | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| FEN | 1 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |
| NFEN | 5 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |
| MDMA | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| MET | 500 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 90.0% | 100% | 100% |
| MOP | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| MTD | 300 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 100% | 100% | 100% |
| OXY | 100 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 100% | 100% | 100% |
| PCP | 25 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |

Page 35

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| PPX | 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |
| TCA | 1000 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95.0% | 100% | 100% |
| THC | 20 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 90.0% | 100% | 100% |

Result of The VINScreen Urine Drug Test Cup: Configuration 2

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| AMP | 1000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| BUP | 10 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| BZO | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| COC | 300 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 100% | 100% | 100% |
| EDDP | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |

Page 36

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| FEN | 1 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |
| NFEN | 5 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| MDMA | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| MET | 1000 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 100% | 100% | 100% |
| MTD | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 100% | 100% | 100% |
| OPI | 2000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| OXY | 100 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% |
| PCP | 25 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% |
| PPX | 300 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 90.0% | 100% | 100% |
| TCA | 1000 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 100% | 95.0% | 100% | 100% |

Page 37

| Drug | Cutoff (ng/mL) | Results | Concentration |||||||
|------|----------------|---------|---------------|
| | | | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff |
| THC | 50 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| | | Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 |
| | | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| | | Agreement (%) | 100% | 100% | 100% | 95.0% | 90.0% | 100% | 100% |

Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

Clinical Studies:
Not applicable.

  1. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that The VINScreen Urine Drug Test Cup and The VINScreen Urine Drug Home Test Cup are substantially equivalent to the predicate device.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).