K Number
K253491

Validate with FDA (Live)

Manufacturer
Date Cleared
2026-02-12

(111 days)

Product Code
Regulation Number
862.1600
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.

Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.

Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).

Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Description

The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.

Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.

The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.

AI/ML Overview

N/A

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.03

February 12, 2026

Roche Diagnostics
Noel Mencias
Senior Regulatory Affairs Manager
9115 Hague Rd.
Indianapolis, Indiana 46250

Re: K253490
Trade/Device Name: Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutions
Regulation Number: 21 CFR 862.1345
Regulation Name: Glucose test system
Regulatory Class: Class II
Product Code: CFR, DJR, JJE
Dated: November 14, 2025
Received: November 14, 2025

Dear Noel Mencias:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K253490 - Noel Mencias Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K253490 - Noel Mencias Page 3

Sincerely,

THOMAS C. MILLER -S
Date: 2026.02.12
14:43:07 -05'00'

for Paula Caposino, Ph.D.
Deputy Director
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023

Indications for Use

See PRA Statement below.

510(k) Number (if known)
K253490

Device Name
cobas pro integrated solutions;
Glucose HK Gen.3;
ONLINE DAT Methadone II

Indications for Use (Describe)
cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.

Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.

Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).

Methadone II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. GC-MS is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

cobas pro integrated solutions; Glucose HK Gen.3; ONLINE DAT Methadone II (K253490)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92

Submitter NameRoche Diagnostics
Address9115 Hague RoadP.O. Box 50416Indianapolis, IN 46250-0457
ContactNoel MenciasPhone: (463) 336-2546Email: noel.mencias@roche.com
Date PreparedOctober 24, 2025
Proprietary Name• Glucose HK Gen.3• ONLINE DAT Methadone II• cobas pro integrated solutions
Common Name• Glucose HK Gen.3• ONLINE DAT Methadone II• cobas pro integrated solutions
Classification Name• Hexokinase, Glucose• Enzyme Immunoassay, Methadone• Analyzer, Chemistry (photometric, discrete), for Clinical Use
Product Codes, Regulation Numbers• CFR, 21 CFR 862.1345• DJR 21 CFR 862.3620• JJE, 21 CFR 862.2160
Predicate Devices• Glucose HK Gen.3• ONLINE DAT Methadone II• cobas pro integrated solutions
Establishment Registration• Roche Diagnostics GmbH Mannheim, Germany: 9610126• Roche Diagnostics GmbH Penzberg, Germany: 9610529• Roche Diagnostics Indianapolis, IN United States: 1823260

Page 1

Page 6

1. DEVICE DESCRIPTION

cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.

The cobas pro integrated solutions consists of a high throughput core unit (sample supply unit and sample buffer unit) and can be configured with different analytical units for immunology, clinical chemistry, and electrolyte testing. The cobas c 703 analytical unit being added to cobas pro integrated solutions is a new clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.

Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD) to nicotinamide adenine dinucleotide reduced (NADH). One micromole of NADH is produced for each micromole of glucose consumed. The NADH produced absorbs light at 340 nm and can be detected spectrophotometrically as an increased absorbance.

The Methadone assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, soluble drug conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. When a urine sample contains the drug in question, this drug competes with the drug derivative conjugate for microparticle‑bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.

Page 2

Page 7

2. INDICATIONS FOR USE

cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semiquantitative, and quantitative clinical chemistry and immunochemistry assays as well as ion-selective measurements.

Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors.

ONLINE DAT Methadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC‑MS).

3. TECHNOLOGICAL CHARACTERISTICS

Table 1: Comparison of the cobas c 703 in cobas pro integrated solutions versus the predicate cobas c 503 in cobas pro integrated solutions

Predicatecobas c 503 in cobas pro integrated solutions(K191899)Candidatecobas c 703 in cobas pro integrated solutions(K253490)
Intended UseFully automated clinical chemistry analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluidsSame
Measurement PrincipalAbsorbance Photometry (enzyme, substrate, proteins, DAT and TDM)Same
Workflow PrincipleBatch/randomSame
Throughput1000 test/hr2000 test/hr
Sample Handling

Page 3

Page 8

Predicatecobas c 503 in cobas pro integrated solutions(K191899)Candidatecobas c 703 in cobas pro integrated solutions(K253490)
Typical Sample Volumes1.0 – 25.0 μLSame
Sample TypesSerum, plasma, urine, CSF and other depending on the chemistry testSame
Sample Handling systemInput and transport of samples using universal sample racks, modular sample buffer (MSB) input, core/transportation unit and STAT portSame
Sample Capacity Onboard300Same
Sample IdentificationBarcode on the sample or position on the rackSame
Reagent handling
Reagent Volume5-135 μLSame
Reagent Containercobas c pack green (plastic containers with conus screw caps)Same
Reagent AccessReagent cassette caps pierced onboard by the instrumentSame
Onboard Storage Temperature10°C (Range 5-15°C)Same
Reagent bottle/Cassette identificationRFIDSame
On board reagent storage capacity60 positions70 positions
System Cycle time measuring cell3.6 seconds photometric1.8 seconds photometric
Reagent MixUltrasonicSame
Automatic RerunAvailableSame
Application information transfer to instrumentVia remote transfer (cobas Link) or CDSame
Pipetting System
Sample and Reagent SyringesRotation and Z roboticSame
Reagent Probes2 polished steel probes4 polished steel probes
Sample ProbesRoutine: 1 polished steel probeHbA1c in whole blood: 1 polished steel probe2 polished steel probes, no HbA1c
Probe CleaningAutomatic for all probesSame

Page 4

Page 9

Predicatecobas c 503 in cobas pro integrated solutions(K191899)Candidatecobas c 703 in cobas pro integrated solutions(K253490)
Liquid Level DetectionSamples: capacitanceReagents: calculated level and countingSame
Clot DetectionYesSame
Temperature ControlCirculating water bath at 37°CSame
Reaction volumeMaximum 185 μLSame
Volume ratiosDefined by master applicationsSame
Sample and Reagent SyringesRotation and Z roboticSame
Detection Information
SpectrophotometerGradient photometer with discrete photodiodes in fixed arraySame
Detection UnitPhotometric/Turbidimetric/Fluorescence and PotentiometricSame
Light sourceTungsten/halogenLED
Light path5.0 mmSame
Measuring unit11
Wavelengths (in nm)340,376, 415, 450, 480, 505, 546, 570, 600, 660, 700, 800Same
Calibration and Quality Control
Calibrators & ControlCalSets and ControlsSame
Calibration Modeslinear and non-linear full calibration, 1- and 2-point recalibration, automatic lot calibrationSame
Calibration Stabilityup to the shelf life of the reagentSame
On-board QCOptionalSame
Interfaces
Host InterfaceEthernet (HL7)Same
PrinterLaserSame
DisplayTouch screen without physical keyboard and mouseSame
Software
Softwarecobas pro integrated solutions system softwareSame
ConfigurationSeveral analytical units with one PC and one coreSame

Page 5

Page 10

Predicatecobas c 503 in cobas pro integrated solutions(K191899)Candidatecobas c 703 in cobas pro integrated solutions(K253490)
Units ControlledISE, c 503, e 801ISE, c 503, c 703, e 801
Function performedData Input, sample processing, result calculation, result reporting, quality controlSame
PC (Control Unit) FunctionsData input (keyboard, disc) data output (screen, printer, host)Same
Core Unit Functionscobas pro core with real time database, data input and output (via system software communication), control of sample conveyerSame
Analytical Unit(s) FunctionsControl of analytic processes (pipetting, incubation, detection) Primary Signal processingSame
MSB FunctionsIntermediate storage of sample racksSame
Data StorageReal time database in Core Unit (storage of System and Application parameters, Calibration Data, QC Data, Sample results, Alarm History)Same
Software Controlled test countdownAvailableSame
Result CalculationAutomated measuring of signal for kinetic and endpoint methods according to cycle time and automated calculation of concentrations via calibration curveSame
Flagging of ErrorsAvailableSame

Page 6

Page 11

Table 2: Similarities and Differences between the Glucose HK Gen.3 on cobas c 703 and the Predicate Device

PredicateGlucose HK Gen. 3 on cobas c 503 (K191899)CandidateGlucose HK Gen. 3 on cobas c 703 (K253490)
Proprietary nameGlucose HK Gen. 3Glucose HK Gen. 3
Intended useIn vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems.In vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on cobas c systems.
TechnologyPhotometricSame
Test formatEnzymaticSame
Test typeQuantitativeSame
Assay protocolR1+R3+Diluent+Sample, incubationSame
Pipetting volume sample1.5 μLSame
Pipetting volume R121 μLSame
Pipetting volume R38 μLSame
Sample typesSerum, plasma, urine and CSFSame
Handling of R1 and R3Liquid, ready to useSame
Measuring Range2 – 750 mg/dL (0.11 – 41.6 mmol/L)Same

Table 3: Similarities and Differences between the ONLINE DAT Methadone II on cobas c 703 and the Predicate Device

PredicateONLINE DAT Methadone II on Roche/Hitachi Family (K021505)CandidateONLINE DAT Methadone II on cobas c 703 (K253490)
MethodologyKIMS, Kinetic interaction of microparticles in a solutionSame
Intended UseMethadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on automated clinical chemistry analyzers at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assessMethadone II (MDN2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of methadone in human urine on cobas c systems at a cutoff concentration of 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality

Page 7

Page 12

PredicateONLINE DAT Methadone II on Roche/Hitachi Family (K021505)CandidateONLINE DAT Methadone II on cobas c 703 (K253490)
assay performance as part of a quality control program.control program.Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS).
Sample TypeUrineUrine
ReagentsConjugate Working Solution: Conjugated methadone derivative in buffer with BSA and preservative.Antibody/Microparticle Working Solution: Microparticles coated with methadone monoclonal antibody (mouse) in buffer with BSA and preservative.Same
ControlsControl Set DAT I: includes PreciPos DAT Set I and PreciNeg DAT Set IControl Set DAT Clinical: includes PreciPOS DAT Clinical and PreciNeg DAT ClinicalSame
CalibratorsPreciset DAT Plus I Calibrator SetPreciset DAT Plus I calibratorsC.f.a.s. DAT Qualitative PlusC.f.a.s. DAT Qualitative Plus Clinical
Qualitative and SemiQuantitative cutoffs300 ng/mLSame

Page 8

Page 13

4. NON-CLINICAL PERFORMANCE EVALUATION

The non-clinical performance studies for Glucose HK Gen.3 and ONLINE DAT Methadone II are summarized below.

4.1. Precision

4.1.1. Repeatability and Intermediate Precision

Precision was determined using human samples and controls in accordance with the CLSI (Clinical and Laboratory Standards Institute) EP05‑A3 requirements with repeatability (n = 84) and intermediate precision (2 aliquots per run, 2 runs per day, 21 days). Results for repeatability and intermediate precision were obtained on the cobas c 703 analyzer.

Table 4: Summary of GLUC3 Precision Results – Serum/Plasma

Repeatability

SampleMean (mg/dL)SD (mg/dL)CV (%)
PCCC11040.4000.4
PCCC22541.110.4
Serum 14.0180.09482.4
Serum 267.90.3710.5
Serum 31120.5680.5
Serum 43891.480.4
Serum 57413.840.5

Intermediate Precision

SampleMean (mg/dL)SD (mg/dL)CV (%)
PCCC11040.7820.8
PCCC22541.910.8
Serum 14.040.1042.6
Serum 267.90.4940.7
Serum 31121.071.0
Serum 43891.870.5
Serum 57415.050.7

Page 9

Page 14

Table 5: Summary of GLUC3 Precision Results – Urine

Repeatability

SampleMean (mg/dL)SD (mg/dL)CV (%)
LiQ_UR127.20.2671.0
LiQ_UR22901.750.6
Sample 13.800.2636.9
Sample 213.80.2071.5
Sample 369.70.4070.6
Sample 43751.8020.5
Sample 57303.320.5

Intermediate Precision

SampleMean (mg/dL)SD (mg/dL)CV (%)
LiQ_UR127.20.2971.1
LiQ_UR22902.410.8
Sample 13.800.2636.9
Sample 213.80.2251.6
Sample 369.70.5680.8
Sample 43752.220.6
Sample 57304.090.6

Table 6: Summary of GLUC3 Precision Results – CSF

Repeatability

SampleMean (mg/dL)SD (mg/dL)CV (%)
LiQ_CSF157.30.3030.5
LiQ_CSF229.60.2310.8
Sample 14.470.2545.7
Sample 237.80.2340.6
Sample 370.80.3350.5
Sample 43751.750.5
Sample 57263.080.4

Intermediate Precision

SampleMean (mg/dL)SD (mg/dL)CV (%)
LiQ_CSF157.30.3680.6
LiQ_CSF229.60.2540.9
Sample 14.470.2655.9
Sample 237.80.3350.9
Sample 370.80.4020.6
Sample 43752.780.7
Sample 57264.230.6

Page 10

Page 15

Table 7: Summary of MDN2 Precision Results, Repeatability and Intermediate Precision – Semi Quantitative

Repeatability

SampleMean (ng/mL)SD (ng/mL)CV (%)
Sample -100% (zero)15.113.790.6
Sample -75%1136.125.4
Sample -50%1705.943.5
DAT1N2386.822.9
DATCN2397.693.2
Cutoff29510.23.5
DAT1P40713.23.2
DATCP39312.13.1
Sample +50%48923.24.7
Sample +75%57920.33.5
Sample +100%65228.74.4

Intermediate Precision

SampleMean (ng/mL)SD (ng/mL)CV (%)
Sample -100% (zero)15.316.1105.0
Sample -75%1139.978.8
Sample -50%16512.37.5
DAT1N24210.94.5
DATCN23911.44.8
Cutoff29622.57.6
DAT1P40718.44.5
DATCP40118.24.5
Sample +50%50531.56.2
Sample +75%57930.65.3
Sample +100%63536.75.8

Page 11

Page 16

Table 8: Summary of MDN2 Precision Results, Repeatability and Intermediate Precision – Qualitative

Reagent Lot: Lot 1 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Reagent Lot: Lot 2 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Reagent Lot: Lot 3 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Page 12

Page 17

Table 9: Summary of MDN2 Precision Results, Repeatability and Intermediate Precision – Qualitative Plus

Reagent Lot: Lot 1 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Reagent Lot: Lot 2 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Reagent Lot: Lot 3 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Page 13

Page 18

Table 10: Summary of MDN2 Precision Results, Repeatability and Intermediate Precision – Qualitative Plus Clinical

Reagent Lot: Lot 1 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Reagent Lot: Lot 2 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Reagent Lot: Lot 3 - 21 Days 2 Runs 2 Aliquots

SpecimenNNEGPOSCrossover (N)Agreement (%)
Sample -100% (zero)848400100
Sample -75%848400100
Sample -50%848400100
DAT1N848400100
DATCN848400100
Cutoff84n.a.n.a.n.a.n.a.
DAT1P840840100
DATCP840840100
Sample +50%840840100
Sample +75%840840100
Sample +100%840840100

Page 14

Page 19

4.2. Analytical Sensitivity

The Limit of Blank (LoB) of the Glucose HK Gen.3 assay on the cobas c 703 analyzer was determined according to CLSI EP17-A2. The LoB is the highest observed measurement value for an analyte-free sample. The Limit of Blank was determined as the 95th percentile of measurements of blank samples.

The Limit of Detection (LoD) of the Glucose HK Gen.3 assay on the cobas c 703 analyzer was determined according to CLSI EP17-A2. The LoD determines the lower limit for samples with analyte. The LoD was determined as the lowest amount of analyte in a sample that can be detected with a 95% probability.

The Limit of Quantitation (LoQ) of the Glucose HK Gen.3 assay was determined according to CLSI EP17-A2. LoQ determines the lowest amount of analyte that can be quantitatively determined with stated accuracy and stated experimental conditions. The LoQ was determined as the lowest concentration of analyte which can be quantified with a total error of no more than 20%.

Table 11: Summary of GLUC3 Analytical Sensitivity Results

Serum

Limit of Blank0.350 mg/dL (0.0194 mmol/L)
Limit of Detection0.487 mg/dL (0.0270 mmol/L)
Limit of Quantitation1.517 mg/dL (0.0842 mmol/L)

Urine

Limit of Blank0.777 mg/dL (0.0431 mmol/L)
Limit of Detection1.160 mg/dL (0.0644 mmol/L)
Limit of Quantitation1.160 mg/dL (0.0644 mmol/L)

CSF

Limit of Blank0.337 mg/dL (0.0187 mmol/L)
Limit of Detection0.506 mg/dL (0.0281 mmol/L)
Limit of Quantitation1.454 mg/dL (0.0807 mmol/L)

Page 15

Page 20

4.3. Linearity/Assay Reportable Range

Linearity studies were conducted to demonstrate that measurements across the claimed measuring range for each parameter are linear. The studies were performed according to CLSI guideline EP06-Ed2. Linearity was confirmed for the following measuring ranges:

GLUC3
Serum, plasma, urine and CSF
0.11‑41.6 mmol/L (2‑750 mg/dL)

MDN2
Urine
39.3 to 2081 ng/mL

Summary of MDN2 Linearity Results According to CLSI EP06-ED2:

Low end of linear range found: 37.0
High end of linear range found: 2255
Intercept: 0
Slope: 1.0040
Pearson's r: 0.9988
R²: 0.991

4.4. Dilution

Post Dilution Check experiments were conducted to verify the automatic rerun function of the cobas c 703 analytical unit. Post Dilution studies for GLUC3 confirmed the 1:2 dilution of samples via rerun function.

4.5. Endogenous Interferences

The effects of interference by endogenous substances on Glucose HK Gen.3 and ONLINE DAT Methadone II were determined on the cobas c 703 analytical unit using, where applicable, human serum/plasma, urine, and CSF spiked with varying levels of interferent.

Page 16

Page 21

Table 12: Summary of GLUC3 Endogenous Interferences – Serum/Plasma

Potential InterferentNo Interference up toObserved Sample Conc (mmol/L)
Albumin [Level 1]63.3 [g/L]2.17
Albumin [Level 2]62.3 [g/L]12.3
Conjugated Bilirubin [Level 1]64 [Index]2.17
Conjugated Bilirubin [Level 2]79 [Index]12.4
Hemolysis [Level 1]1213 [Index]2.14
Hemolysis [Level 2]1228 [Index]12.4
Immunoglobulin G [Level 1]65.1 [g/L]2.17
Immunoglobulin G [Level 2]65.4 [g/L]12.4
Lipemia [Level 1]1134 [Index]2.17
Lipemia [Level 2]1102 [Index]12.3
Unconjugated Bilirubin [Level 1]77 [Index]2.18
Unconjugated Bilirubin [Level 2]90 [Index]12.2

Table 13: Summary of GLUC3 Endogenous Interferences – Urine

Potential InterferentNo Interference up toObserved Sample Conc (mmol/L)
Albumin [Level 1]2.70 [g/L]2.28
Albumin [Level 2]3.10 [g/L]12.5
Calcium [Level 1]12.6 [mmol/L]2.29
Calcium [Level 2]13.5 [mmol/L]12.5
Citrate [Level 1]12.0 [mmol/L]2.28
Citrate [Level 2]12.2 [mmol/L]12.5
Creatinine [Level 1]91.9 [mmol/L]2.28
Creatinine [Level 2]89.7 [mmol/L]12.5
Hemolysis [Level 1]946 [Index]2.31
Hemolysis [Level 2]876 [Index]12.1
Immunoglobulin G [Level 1]1.10 [g/L]2.29
Immunoglobulin G [Level 2]1.30 [g/L]12.4
Magnesium [Level 1]27.3 [mmol/L]2.29
Magnesium [Level 2]28.9 [mmol/L]12.5
Oxalate [Level 1]2.70 [mmol/L]2.27
Oxalate [Level 2]2.70 [mmol/L]12.5
Phosphate [Level 1]139 [mmol/L]2.29

Page 17

Page 22

| Phosphate [Level 2] | 137 [mmol/L] | 12.5 |
| Urea [Level 1] | 1836 [mmol/L] | 2.27 |
| Urea [Level 2] | 1825 [mmol/L] | 12.5 |
| Uric acid [Level 1] | 6.00 [mmol/L] | 2.30 |
| Uric acid [Level 2] | 7.50 [mmol/L] | 12.5 |
| Urobilinogen [Level 1] | 0.250 [mmol/L] | 2.30 |
| Urobilinogen [Level 2] | 0.250 [mmol/L] | 12.3 |

Table 14: Summary of GLUC3 Endogenous Interferences – CSF

Potential InterferentNo Interference up toObserved Sample Conc (mmol/L)
Conjugated Bilirubin [Level 1]75 [Index]2.21
Conjugated Bilirubin [Level 2]66 [Index]12.3
Hemolysis [Level 1]1201 [Index]2.21
Hemolysis [Level 2]1259 [Index]12.2

Table 15: Summary of MDN2 Endogenous Interferences – Urine

CompoundConcentration
Acetone1000 mg/dL
Ascorbic acid1500 mg/dL
Albumin500 mg/dL
Bilirubin25 mg/dL
Calcium133 mg/dL
Creatinine500 mg/dL
Ethanol1000 mg/dL
Glucose2000 mg/dL
Hemoglobin750 mg/dL
IgG110 mg/dL
Magnesium238 mg/dL
NaCl5800 mg/dL
Oxalate200 mg/dL
pH3.9–9.1
Phosphate2028 mg/dL
Sodium citrate, dihydrate323.51 mg/dL
Specific Gravity1.001–1.035 g/mL
Urea6000 mg/dL
Uric Acid100.86 mg/dL
Urobilinogen14.82 mg/dL

Page 18

Page 23

4.6. Exogenous Interferences – Drugs

Drug Interference studies were conducted to evaluate drugs for potential interference with Glucose HK Gen.3 and ONLINE TDM Methadone II measured on the cobas c 703 analytical unit. The following interference claims were verified:

Table 16: Drug Interferences GLUC3, Level 1 – Serum/Plasma

DrugDrug Conc 1 [mg/L]Analyte Mean Reference [mmol/L]Analyte Mean Conc of Spiked Sample [mmol/L]Recovery [%]
N-Acetylcysteine1502.212.22100.5
Acetylsalicylic acid302.212.21100.0
Ampicillin-Na752.212.21100.0
Ascorbic acid52.52.212.21100.0
Cefoxitin7502.212.21100.1
Doxycyclin182.212.2199.7
Heparin3300 IU/L2.212.22100.2
Levodopa7.52.212.22100.3
Methyldopa22.52.212.22100.3
Metronidazole1232.212.23100.6
Rifampicin482.212.22100.2
Acetaminophen1562.192.1899.6
Cyclosporine1.82.192.1999.9
Ibuprofen2192.192.1899.7
Theophylline602.192.19100.0
Phenylbutazone3212.152.16100.4

Table 17: Drug Interferences GLUC3, Level 1 – Urine

DrugDrug Conc 1 [mg/L]Analyte Mean Reference [mmol/L]Analyte Mean Conc of Spiked Sample [mmol/L]Recovery [%]
Acetaminophen30002.292.2799.2
N-Acetylcysteine102.292.2799.5
Ascorbic acid40002.292.2899.7
Cefoxitin120002.292.2699.0
Gentamycine sulfate4002.292.2699.0
Levodopa10002.292.29100.1
Methyldopa20002.292.2899.9
Ofloxacine9002.292.2196.9

Page 19

Page 24

| Ibuprofen | 4000 | 2.13 | 2.19 | 103.1 |
| Phenazopyridine | 300 | 2.13 | 2.12 | 99.6 |
| Salicyluric acid | 6000 | 2.13 | 5.01 | 235.3 |
| Tetracycline | 300 | 2.26 | 1.98 | 87.9 |

DrugDrug Conc 2 [mg/L]Analyte Mean Reference [mmol/L]Analyte Mean Conc of Spiked Sample [mmol/L]Recovery [%]
Salicyluric acid1002.272.34103.0
Tetracycline1002.212.21100.0

Table 18: Drug Interferences GLUC3, Level 2 – Urine

DrugDrug Conc 1 [mg/L]Analyte Mean Reference [mmol/L]Analyte Mean Conc of Spiked Sample [mmol/L]Recovery [%]
Acetaminophen300012.212.199.2
N-Acetylcysteine1012.212.299.7
Ascorbic acid400012.212.299.8
Cefoxitin1200012.212.299.8
Gentamycine sulfate40012.212.2100.2
Levodopa100012.212.2100.0
Methyldopa200012.212.299.7
Ofloxacine90012.212.299.5
Ibuprofen400011.311.8103.9
Phenazopyridine30011.311.7103.2
Salicyluric acid600011.314.5127.9
Tetracycline30011.611.599.7
Salicyluric acid10012.212.2100.7

Page 20

Page 25

Table 19: Drug Interferences MDN2, Semi-Quantitative (MD3S0); Qualitative (MD3Q0); Qualitative Plus Clinical (MD3QC); Qualitative Plus (MD3QP) – Urine

Reagent Lot: Lot 1

CompoundConcentration (ng/mL)Sample -25%Sample +25%
Acetaminophen3000000No crossoverNo crossover
Acetyl cysteine10000No crossoverNo crossover
Acetylsalicylic acid100000No crossoverNo crossover
Aminopyrine100000No crossoverNo crossover
Amitriptyline100000No crossoverNo crossover
Ampicillin100000No crossoverNo crossover
Ascorbic acid4000000No crossoverNo crossover
Aspartame100000No crossoverNo crossover
Atropine100000No crossoverNo crossover
Benzocaine100000No crossoverNo crossover
Caffeine100000No crossoverNo crossover
Calcium dobesilate100000No crossoverNo crossover
Calcium hypochlorite100000No crossoverNo crossover
Carbamazepine100000No crossoverNo crossover
Cefoxitin12000000No crossoverNo crossover
Chloroquine100000No crossoverNo crossover
Chlorpheniramine100000No crossoverNo crossover
Cotinine100000No crossoverNo crossover
Cyclobenzaprine100000No crossoverNo crossover
Cyproheptadine100000No crossoverNo crossover
d-Ephedrine100000No crossoverNo crossover
Desipramine100000No crossoverNo crossover
Diphenylhydantoin100000No crossoverNo crossover
dl-Ephedrine100000No crossoverNo crossover
Dopamine100000No crossoverNo crossover
Doxepin100000No crossoverNo crossover
Doxylamine100000No crossoverNo crossover
d-Phenylpropanolamine100000No crossoverNo crossover
d-Pseudoephedrine100000No crossoverNo crossover
Erythromycin100000No crossoverNo crossover
Estriol100000No crossoverNo crossover
Fenoprofen100000No crossoverNo crossover
Fluoxetine100000No crossoverNo crossover
Furosemide100000No crossoverNo crossover
Gentamicin sulfate400000No crossoverNo crossover
Gentisic acid100000No crossoverNo crossover
Guaiacol glycerol ether100000No crossoverNo crossover
Haloperidol100000No crossoverNo crossover
Hydrochlorothiazide100000No crossoverNo crossover
Ibuprofen4000000No crossoverNo crossover
Isoproterenol100000No crossoverNo crossover
Ketamine100000No crossoverNo crossover

Page 21

Page 26

| l-Ephedrine | 100000 | No crossover | No crossover |
| Levodopa | 1000000 | No crossover | No crossover |
| Lidocaine | 100000 | No crossover | No crossover |
| l-Norpseudoephedrine | 100000 | No crossover | No crossover |
| Maprotiline | 100000 | No crossover | No crossover |
| Melanin | 100000 | No crossover | No crossover |
| Methyldopa | 2000000 | No crossover | No crossover |
| Mianserin | 100000 | No crossover | No crossover |
| Naloxone | 100000 | No crossover | No crossover |
| Naltrexone | 100000 | No crossover | No crossover |
| Naproxen | 100000 | No crossover | No crossover |
| Niacinamide | 100000 | No crossover | No crossover |
| Nicotine | 100000 | No crossover | No crossover |
| Nordoxepin | 100000 | No crossover | No crossover |
| Norethindrone | 100000 | No crossover | No crossover |
| Ofloxacine | 90000 | No crossover | No crossover |
| Penicillin G | 100000 | No crossover | No crossover |
| Phenazopyridine | 300000 | No crossover | No crossover |
| Phenetylamine | 100000 | No crossover | No crossover |
| Phenothiazine | 100000 | No crossover | No crossover |
| Phenylbutazone | 100000 | No crossover | No crossover |
| Phenylpropanolamine | 100000 | No crossover | No crossover |
| Procaine | 100000 | No crossover | No crossover |
| Quinidine | 100000 | No crossover | No crossover |
| Quinine | 100000 | No crossover | No crossover |
| Salicyluric acid | 6000000 | No crossover | No crossover |
| Sulindac | 100000 | No crossover | No crossover |
| Tetracycline | 300000 | No crossover | No crossover |
| Tetrahydrozoline | 100000 | No crossover | No crossover |
| Tramadol | 20000 | No crossover | No crossover |
| Trifluoperazine | 100000 | No crossover | No crossover |
| Tyramine | 100000 | No crossover | No crossover |

Reagent Lot: Lot 2

CompoundConcentration (ng/mL)Sample -25%Sample +25%
Amobarbital100000No crossoverNo crossover
Benzoylecgonine100000No crossoverNo crossover
Benzphetamine100000No crossoverNo crossover
Butabarbital100000No crossoverNo crossover
Chlordiazepoxide100000No crossoverNo crossover
Cocaine100000No crossoverNo crossover
Codeine100000No crossoverNo crossover
d-Amphetamin100000No crossoverNo crossover
Dextrometorphan100000No crossoverNo crossover
Dextropropoxyphene100000No crossoverNo crossover
Diazepam100000No crossoverNo crossover
Diphenhydramine100000No crossoverNo crossover
Disopyramide100000No crossoverNo crossover
d-Methamphetamine100000No crossoverNo crossover

Page 22

Page 27

| Ecgonine | 100000 | No crossover | No crossover |
| Ecgonine methyl ester | 100000 | No crossover | No crossover |
| EDDP | 100000 | No crossover | No crossover |
| EMDP | 100000 | No crossover | No crossover |
| Epinephrine | 100000 | No crossover | No crossover |
| Glutethimide | 100000 | No crossover | No crossover |
| Imipramine | 100000 | No crossover | No crossover |
| l-Amphetamine | 100000 | No crossover | No crossover |
| l-Methamphetamine | 100000 | No crossover | No crossover |
| l-Pseudoephedrine | 100000 | No crossover | No crossover |
| LSD | 100000 | No crossover | No crossover |
| MDA | 100000 | No crossover | No crossover |
| MDMA | 100000 | No crossover | No crossover |
| Meperidine | 100000 | No crossover | No crossover |
| Methaqualone | 100000 | No crossover | No crossover |
| Methylphenidate | 100000 | No crossover | No crossover |
| Morphine | 100000 | No crossover | No crossover |
| Nordiazepam | 100000 | No crossover | No crossover |
| Nortriptyline | 100000 | No crossover | No crossover |
| Orphenadrine | 100000 | No crossover | No crossover |
| Oxazepam | 100000 | No crossover | No crossover |
| Pentobarbital | 100000 | No crossover | No crossover |
| Perphenazine | 100000 | No crossover | No crossover |
| Phencyclidine | 100000 | No crossover | No crossover |
| Phenobarbital | 100000 | No crossover | No crossover |
| Phentermine | 100000 | No crossover | No crossover |
| Protriptyline | 100000 | No crossover | No crossover |
| Quetiapine-Carboxylic acid | 500000 | No crossover | No crossover |
| Quetiapine-Fumarat | 750000 | No crossover | No crossover |
| Quetiapine-Sulfoxid | 500000 | No crossover | No crossover |
| Secobarbital | 100000 | No crossover | No crossover |
| THC-9-carboxylic acid | 100000 | No crossover | No crossover |
| Verapamil | 100000 | No crossover | No crossover |

Reagent Lot: Lot 6

CompoundConcentration (ng/mL)Sample -25%Sample +25%
Disopyramide1000000No crossoverNo crossover

Page 23

Page 28

4.7. Cross Reactivity

The effects of interference by endogenous interferences were conducted on one cobas c 703 using one lot, one human sample per concentration level, in one run, ≥2 replicates per sample.

Table 20: Cross Reactivity Results MDN2

Cross ReactantCross Reactivity (%)
Lu AA344437.66
Vortioxetine4.59
Cyamemazine4.12
Methotrimeprazine (Levomepromazine)3.47
Chlorpromazine0.97
Thiothixene0.39
Clomipramine0.17
Thioridazine0.15
Promazine0.14
Promethazine0.074
Trimipramine0.052
Chlorprothixene0.045

4.8. Sample Matrix Comparison

The recovery of analyte values in the presence of anticoagulants with the Glucose HK Gen.3 assay was determined on the cobas c 703 analytical unit by comparing values obtained from samples drawn into serum and plasma collection tubes. The following results were obtained.

Table 21: Summary of Matrix Comparison Passing/Bablok Results

AnticoagulantSlopeIntercept (mg/dL)Correlation CoefficientConcentration of Samples (mg/dL)
Serum vs. Serum/gel separation tube1.007-0.4321.0003.17-737
Serum vs. K2-EDTA plasma1.003-0.07710.9993.17-737

Page 24

Page 29

AnticoagulantSlopeIntercept (mg/dL)Correlation CoefficientConcentration of Samples (mg/dL)
Serum vs. Li-Heparin plasma1.031-1.590.9993.17-737
Serum vs. NaF/K-Oxalate plasma1.025-1.251.0003.17-737
Serum vs. NaF/Na2-EDTA plasma1.037-1.780.9993.17-737
Serum vs. KF/Na2-EDTA plasma1.028-0.5870.9993.17-737
Serum vs. NaF/Citrate/Na2-EDTA plasma1.0051.870.9992.88-730

4.9. Method Comparison to Predicate

Method Comparison experiments were performed for all sample types using the Glucose HK Gen.3 assay on the cobas c 703 versus the Glucose HK Gen.3 assay on the cobas c 503 to assess the bias between the two analytical units.

The results can be found below:

Table 22: Summary of GLUC3 Method Comparison Data – Serum/Plasma

Quality CharacteristicResult
Slope (Passing/Bablok)0.995
Intercept (Passing/Bablok)-0.0968
Bias (%) at medical decision point 1-0.6
Coefficient of correlation (Pearson (r))1.000
Slope (Linear Regression)0.999
Intercept (Linear Regression)-0.216
Coefficient of correlation (Kendall (Tau))0.990
Sample concentration3.60 - 724
Number of samples (n)74

Page 25

Page 30

Table 23: Summary of GLUC3 Method Comparison Data – Urine

Quality CharacteristicResult
Slope (Passing/Bablok)0.998
Intercept (Passing/Bablok)0.209
Bias (%) at medical decision point 10.9
Coefficient of correlation (Pearson (r))1.000
Slope (Linear Regression)1.001
Intercept (Linear Regression)0.214
Coefficient of correlation (Kendall (Tau))0.950
Sample concentration2.04 - 730
Number of samples (n)67

Table 24: Summary of GLUC3 Method Comparison Data – CSF

Quality CharacteristicResult
Slope (Passing/Bablok)0.987
Intercept (Passing/Bablok)-0.234
Bias (%) at medical decision point 1-1.9
Coefficient of correlation (Pearson (r))1.000
Slope (Linear Regression)0.989
Intercept (Linear Regression)-0.259
Coefficient of correlation (Kendall (Tau))0.984
Sample concentration2.78 - 730
Number of samples (n)75

Page 26

Page 31

87 urine samples, obtained from a clinical laboratory, where they screened negative in a drug test panel, were evaluated with the Methadone II assay. 97.7 % of these normal urines were negative relative to the 300 ng/mL cutoff. 79 urine samples obtained from a clinical laboratory, where they screened preliminary positive with a commercially available immunoassay, were subsequently confirmed by GC‑MS and were evaluated with the Methadone II assay. The following results were obtained with the Methadone II assay on the cobas c 703 analyzer relative to the GC‑MS values.

Table 25: Summary of MDN2 Method Comparison versus GC/MS – Semi-Quantitative

![Table showing GC-MS values comparison with cobas c 703 analyzer results, displaying negative samples, near cutoff samples, and positive samples with corresponding values]

Clinical samples were evaluated with this assay on a cobas c 703 analyzer and on a cobas c 503 analyzer. 50 urine samples screened negative for methadone on a cobas c 503 analyzer were evaluated with the Methadone II assay on a cobas c 703 analyzer. 100 % of these normal urines were negative on both the cobas c 703 analyzer and the cobas c 503 analyzer. 50 urine samples, obtained from a clinical laboratory where they screened preliminary positive with a commercially available immunoassay were subsequently confirmed by GC‐MS and evaluated with the Methadone II assay on a cobas c 703 analyzer. 100 % of the samples were positive on both the cobas c 503 analyzer and the cobas c 703 analyzer.

Page 27

Page 32

Table 26: Summary of MDN2 Method Comparison cobas c 703 versus cobas c 503 – Semi-Quantitative

![Table showing Methadone II correlation with cutoff = 300 ng/mL between cobas c 503 and cobas c 703 analyzers]

4.10. Stability

The stability studies and acceptance criteria have been reviewed and found to be acceptable. The stability data supports Roche Diagnostic's claims as reported in the package labeling.

5. ADDITIONAL INFORMATION

The Glucose HK Gen.3 assay is intended to be used with the following:

• Calibrator f.a.s.
• PreciControl ClinChem Multi 1
• PreciControl ClinChem Multi 2
• Diluent NaCl 9 %

The ONLINE DAT Methadone II assay is intended to be used with the following:

• Preciset DAT Plus I
• C.f.a.s. DAT Qualitative Plus
• C.f.a.s. DAT Qualitative Plus Clinical
• Control Set DAT I
• Control Set DAT Clinical

6. CONCLUSIONS

The analytical performance data for Glucose HK Gen.3 and ONLINE DAT Methadone II on cobas pro integrated solutions support a substantial equivalence decision.

Page 28

§ 862.1600 Potassium test system.

(a)
Identification. A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.(b)
Classification. Class II.