K Number
K151211

Validate with FDA (Live)

Date Cleared
2015-06-04

(29 days)

Product Code
Regulation Number
862.3150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AssureTech Secobarbital Strip test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300ng/mL. The test may yield preliminary positive results when prescription drug Secobarbital is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Oxycodone Strip test is an immunochromatographic assay for the qualitative determination of Oxycodone in human urine at a Cut-Off concentration of 100ng/mL. The test may yield preliminary positive results when prescription drug Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Panel Dip test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively. The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Quick Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively. The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Turn Key-Split Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL. respectively. The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

Device Description

Secobarbital Strip. The AssureTech AssureTech Oxycodone Strip. AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of Secobarbital and/or Oxycodone (target analytes) in human urine. The Quick Cup does not contain a turn-key for device activation. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.

AI/ML Overview

The company, AssureTech, has conducted several studies to demonstrate that its urine drug tests (AssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup, and AssureTech Secobarbital/Oxycodone Turn Key-Split Cup) meet acceptance criteria for qualitative determination of Secobarbital and Oxycodone in human urine.

Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria (Implicit) and Reported Device Performance

The document does not explicitly state formal acceptance criteria values (e.g., "sensitivity must be >95%"). Instead, performance is demonstrated by the proportion of correct results at various drug concentrations, particularly around the cut-off. The implicit acceptance criterion is that the device should accurately classify samples as positive or negative relative to the established cut-off values, and that performance near the cut-off should be reasonable, accounting for the inherent variability of immunoassays.

Study TypeAcceptance Criteria (Implicit)Reported Device Performance (Summary)
PrecisionConsistent results across different manufacturing lots and concentrations around the cut-off.Highly Consistent: For both Secobarbital (cutoff 300 ng/mL) and Oxycodone (cutoff 100 ng/mL), across 3 lots and 4 different device configurations, the tests consistently showed:- 100% negative results for concentrations at -100%, -75%, -50%, and -25% of the cut-off.- High positive rates (typically 47/50 to 50/50, or 94-100%) at the cut-off.- 100% positive results for concentrations at +25%, +50%, +75%, and +100% of the cut-off.
Cut-off VerificationAll samples below -25% cut-off should be negative, and all samples above +25% cut-off should be positive.Met: All samples at and below -25% Cut-Off were negative, and all samples at and above +25% Cut-Off were positive for both Secobarbital and Oxycodone.
InterferenceNo significant interference from common physiological/pathological substances or non-target drugs.Met: Numerous common substances (listed over two pages in the document) showed no interference at 100 µg/mL.
SpecificityLimited cross-reactivity with structurally similar compounds, with established approximate percentages relative to the target drug.Met: Cross-reactivity percentages are provided for various related compounds (e.g., Amobarbital at 30% for Secobarbital, Oxymorphone at 40% for Oxycodone).
Effect of Urine Specific Gravity and pHAccurate results across a range of urine specific gravity (1.000-1.035) and pH (4-9).Met: All samples at/above +25% cut-off were positive, and all samples at/below -25% cut-off were negative within these ranges.
Method Comparison (Professional Use)High agreement with GC/MS results, especially for samples far from the cut-off. Any discrepancies near the cut-off should be acceptable due to inherent variability.High agreement shown: - Secobarbital: For most cases, 100% agreement for negative and high positive samples. Some discordant results were observed for specific samples near the cut-off (e.g., GC/MS 286 ng/mL reported as positive by device; GC/MS 306 ng/mL reported as negative by device).- Oxycodone: Similar high agreement, with some discordant results near the cut-off (e.g., GC/MS 96 ng/mL reported as positive by device; GC/MS 102 ng/mL reported as negative by device).
Lay-user Study (OTC Use)High percentage of correct results for lay users, and ease of understanding instructions.High Agreement & Understandability: - Secobarbital: 90-100% correct results across various concentrations, with 90% at -25% cut-off and 100% at +25% cut-off.- Oxycodone: 95-100% correct results, with 95% at -25% cut-off and 95% at +25% cut-off.- All lay users indicated instructions were easy to follow. Flesch-Kincaid reading Grade Level of 7.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Study:

    • For each drug (Secobarbital, Oxycodone) and each device configuration, 8 concentrations were tested (100%, 75%, 50%, 25% below cut-off, cut-off, and 25%, 50%, 75%, 100% above cut-off).
    • For each concentration, tests were performed two runs per day for 25 days, using 3 different lots.
    • Total tests per concentration per lot: 2 runs/day * 25 days = 50 tests.
    • Total tests per concentration for all 3 lots: 50 tests * 3 lots = 150 tests.
    • Total per drug per device: 150 tests/concentration * 8 concentrations = 1200 tests.
    • Data Provenance: The samples were prepared by "spiking drug in negative samples." This indicates laboratory-prepared samples, not clinical samples. The country of origin is not explicitly stated but implied to be related to the submitter (China). This is a prospective experimental study.
  • Cut-off Verification Study:

    • 150 samples were used, equally distributed at concentrations of -50%, -25%, Cut-Off, +25%, +50% Cut-Off.
    • Tested using three different lots of each device by three different operators.
    • Data Provenance: Laboratory-prepared spiked samples. This is a prospective experimental study.
  • Interference and Specificity Studies:

    • Samples were prepared by adding potential interfering substances or cross-reactants to drug-free urine and target drug urine (at concentrations 25% below and 25% above Cut-Off levels).
    • Tested using three batches of each device.
    • Data Provenance: Laboratory-prepared spiked samples. This is a prospective experimental study.
  • Effect of Urine Specific Gravity and pH:

    • Urine samples with specific gravity from 1.000 to 1.035 or pH from 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels.
    • Tested using three batches of each device.
    • Data Provenance: Laboratory-prepared spiked samples. This is a prospective experimental study.
  • Method Comparison Studies (Professional Use):

    • Sample Size: 80 "unaltered clinical samples" for each device (40 negative and 40 positive based on GC/MS). These 80 samples were further categorized into Negative, Low Negative, Near Cutoff Negative, Near Cutoff Positive, and High Positive.
    • Data Provenance: "Unaltered clinical samples." The country of origin is not specified, but these are retrospective clinical samples with established GC/MS results.
  • Lay-user Study (OTC Use):

    • Sample Size: 1060 lay persons.
    • The document implies that each lay person tested one sample, so 1060 samples in total across all participants/devices.
    • The samples used in this study were prepared at various concentrations (-100%, -75%, -50%, -25%, +25%, +50%, +75% of the cutoff) by spiking drug(s) into drug-free-pooled urine specimens.
    • Data Provenance: Laboratory-prepared spiked samples. This is a prospective study involving human participants.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • Precision, Cut-off, Interference, Specificity, Specific Gravity/pH Studies: Ground truth was established by GC/MS (Gas Chromatography/Mass Spectrometry), which is a gold standard analytical method. No human experts are explicitly mentioned for ground truth establishment for these analytical performance studies, as the concentrations are chemically determined.
  • Method Comparison (Professional Use): The ground truth for the 80 "unaltered clinical samples" was established by GC/MS results. No human experts are explicitly mentioned for adjudicating the GC/MS results themselves. The "three laboratory assistants" performed the device tests, not established the ground truth.
  • Lay-user Study: Ground truth was established by GC/MS for the spiked urine samples.

4. Adjudication Method for the Test Set

  • For the analytical studies (Precision, Cut-off, Interference, Specificity, Specific Gravity/pH), the reference method is GC/MS, so there's no human adjudication in the traditional sense. The device's results are compared directly to the quantitative GC/MS values.
  • For the Method Comparison study, the "three laboratory assistants" are described as "operators" who ran the samples. Their interpretations of the device results were then compared to the GC/MS results. It's not an adjudication of conflicting interpretations by different human readers, but rather a comparison of human interpretation of the device against the GC/MS "gold standard."
  • For the Lay-user study, each lay person interpreted their single device result, which was then compared to the GC/MS result of the sample they tested. No inter-reader adjudication took place for the lay users' results.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not performed as described for AI vs. without AI assistance.
  • The studies involved multiple "viewers" or "operators" (three laboratory assistants for the method comparison, and 1060 lay persons for the lay-user study), which could be considered multi-reader elements. However, these studies were not designed to compare reader performance with and without an AI system. Instead, they assessed the device's performance (interpreted by humans) against a gold standard (GC/MS).
  • Therefore, there is no effect size reported for how much human readers improve with AI vs. without AI assistance, as AI is not part of this device.

6. Standalone (Algorithm Only) Performance

  • Yes, in essence, standalone performance was assessed in the analytical studies (Precision, Cut-off, Interference, Specificity, Specific Gravity/pH). While a human manually reads the test line on the immunochromatographic assay, the "algorithm" is the biochemical reaction itself in the strip/cup. The performance metrics (e.g., 100% negative at -25% cut-off, 100% positive at +25% cut-off) directly reflect the performance of this chemical "algorithm."
  • The "Method Comparison" and "Lay-user" studies assess the performance of the device in the hands of human users, thus incorporating a human-in-the-loop component. However, the foundational analytical performance is "standalone" for the immunoassay.

7. Type of Ground Truth Used

  • The primary ground truth used across all studies is Gas Chromatography/Mass Spectrometry (GC/MS).
  • For the analytical studies, GC/MS confirmed the exact concentrations of drugs in spiked samples.
  • For the method comparison and lay-user studies, GC/MS served as the reference method for confirming the presence and concentration of drugs in clinical and spiked urine samples, respectively.

8. Sample Size for the Training Set

  • This is an immunochromatographic assay, not an AI/machine learning device. Therefore, the concept of a "training set" for an algorithm doesn't directly apply.
  • The "development" or "training" of such a device typically involves optimizing the chemical reagents (antibodies, conjugates, etc.) and physical design to achieve the desired sensitivity and specificity. The document does not provide details on the sample sizes used during the R&D phase for optimizing the assay components. The reported studies are for performance verification of the final product.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned above, there is no "training set" in the context of an AI algorithm.
  • During the development of the immunoassay, the performance of various chemical formulations and designs would have been evaluated against known concentrations of analytes, likely confirmed by a gold standard method like GC/MS, to establish the optimal components and cut-off levels.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 4, 2015

ASSURE TECH. CO., LTD C/O JOE SHIA REGULATORY CONSULTANT 504 E DIAMOND AVE., SUITE F GAITHERSBURG MD 20877

Re: K151211

Trade/Device Name: AssureTech Secobarbital Strip AssureTech Oxycodone Strip AssureTech Secobarbital/Oxycodone Panel Dip AssureTech Secobarbital/Oxycodone Quick Cup AssureTech Secobarbital/Oxycodone Turn Kev-Split Cup Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: II

Product Code: DIS. DJG Dated: April 3, 2015 Received: May 6, 2015

Dear Mr. Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Katherine Serrano -S

For: Courtney H. Lias, Ph.D. Director

Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151211

Device Name

AssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip AssureTech Secobarbital/Oxycodone Quick Cup AssureTech Secobarbial/Oxycodone Turn Key-Split Cup

Indications for Use (Describe)

The AssureTech Secobarbital Strip test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300ng/mL.

The test may yield preliminary positive results when prescription drug Secobarbital is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Oxycodone Strip test is an immunochromatographic assay for the qualitative determination of Oxycodone in human urine at a Cut-Off concentration of 100ng/mL.

The test may yield preliminary positive results when prescription drug Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Panel Dip test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.

The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxyoodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Quick Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.

The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test

{3}------------------------------------------------

provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Turn Key-Split Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL. respectively.

The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

1. Date:June 3, 2015
2. Submitter:Assure Tech. Co., Ltd.Building 1, No.10, Xiyuansan Road, WestlakeEconomic ZoneHangzhou, China, 310030
3. Contact person:Eric LinAssure Tech. Co., Ltd.Building 1, No.10, Xiyuansan Road, WestlakeEconomic ZoneHangzhou, China, 310030Telephone: 510-860-4680Email: customerservice@assurelabs.com.
4. Device Name:AssureTech Secobarbital StripAssureTech Oxycodone StripAssureTech Secobarbital/Oxycodone Panel DipAssureTech Secobarbital/Oxycodone Quick CupAssureTech Secobarbital/Oxycodone Turn Key-Split Cup

Classification:

Product CodeCFRPanel
DIS21 CFR, 862.3150 Barbiturate Test SystemToxicology
DJG21 CFR, 862.3650 Opiate Test SystemToxicology

5. Predicate Devices: K052115

FIRST CHECK DIAGNOSTICS LLC FIRST CHECK MULTI DRUG CUP 12

    1. Intended Use
      The AssureTech Secobarbital Strip test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300ng/mL.

The test may vield preliminary positive results when prescription drug Secobarbital is ingested. even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Oxycodone Strip test is an immunochromatographic assay for the qualitative determination of Oxycodone in human urine at a Cut-Off concentration of 100ng/mL.

{5}------------------------------------------------

The test may yield preliminary positive results when prescription drug Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Panel Dip test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.

The test may vield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Quick Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.

The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

The AssureTech Secobarbital/Oxycodone Turn Key-Split Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL. respectively.

The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test

{6}------------------------------------------------

results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.

    1. Device Description
      Secobarbital Strip. The AssureTech AssureTech Oxycodone Strip. AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of Secobarbital and/or Oxycodone (target analytes) in human urine. The Quick Cup does not contain a turn-key for device activation. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
    1. Substantial Equivalence Information
      A summary comparison of features of the candidate devices and the predicate devices is provided in Tables 1 to 4 below.
ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination of drugs ofabuse in human urine.Same (but the number ofdrugs detected is different)
CalibratorSecobarbital (member of Barbiturates drug class)Barbiturates
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values300 ng/mLSame
Intended UseFor over-the-counter and prescription uses.For over-the-counter use.
ConfigurationsStripCup

Table 1: Features Comparison of AssureTech Secobarbital Strip and the Predicate Devices

Table 2: Features Comparison of AssureTech Oxycodone Strip and the Predicate Devices

ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
CalibratorOxycodoneSame

{7}------------------------------------------------

MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values100 ng/mLSame
Intended UseFor over-the-counter and prescription uses.For over-the-counter use.
ConfigurationsStripCup

Table 3: Features Comparison of AssureTech Secobarbital/Oxycodone Panel Dip and the Predicate Devices

ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
CalibratorSecobarbital and OxycodoneSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values300ng/mL for Secobarbital and 100 ng/mL forOxycodoneSame
Intended UseFor over-the-counter and prescription uses.For over-the-counter use.
ConfigurationsPanel DipCup

Table 4: Features Comparison of AssureTech Secobarbital/Oxycodone Cup and the Predicate Devices

ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
CalibratorSecobarbital and OxycodoneSame

{8}------------------------------------------------

MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values300ng/mL for Secobarbital and 100 ng/mL forOxycodoneSame
Intended UseFor over-the-counter and prescription uses.For over-the-counter use.
ConfigurationsCup with or without turn-keyCup

9. Test Principle

AssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are rapid tests for the qualitative detection of Secobarbital and/or Oxycodone in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibodycoated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

1. Analytical Performance

  • a. Precision
    Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables.

{9}------------------------------------------------

Secobarbital

AssureTech Secobarbital Strip
-----------------------------------
Lot-100%-75%-50%-25%cut off+25%+50%+75%+100%
Numbercut offcut offcut offcutoffcut offcut offcut offcut off
Lot 150-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
AssureTech Secobarbital/Oxycodone Panel Dip
Lot-100%-75%-50%-25%cut off+25%+50%+75%+100%
Numbercut offcut offcut offcutoffcut offcut offcut offcut off
Lot 150-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+3-/47+50+/0-50+/0-50+/0-50+/0-
AssureTech Secobarbital/Oxycodone Turn-Key Split Cup
Lot-100%-75%-50%-25%cut off+25%+50%+75%+100%
Numbercut offcut offcut offcutoffcut offcut offcut offcut off
Lot 150-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+3-/47+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
AssureTech Secobarbital/Oxycodone Quick Cup
Lot-100%-75%-50%-25%cut off+25%+50%+75%+100%
Numbercut offcut offcut offcutoffcut offcut offcut offcut off
Lot 150-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+1-/49+50+/0-50+/0-

Oxycodone

AssureTech Oxycodone Strip

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+1-/49+50+/0-50+/0-50+/0-50+/0-
AssureTech Secobarbital/Oxycodone Panel Dip
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+1-/49+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-

AssureTech Secobarbital/Oxycodone Turn-Key Split Cup

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoff+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+50-/0+3-/47+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+1-/49+50+/0-50+/0-50+/0-50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

{10}------------------------------------------------

AssureTech Secobarbital/Oxycodone Quick Cup

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+3-/47+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+2-/48+50+/0-50+/0-50+/0-50+/0-

b. Linearity

Not applicable.

c. Stability

The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 45 °C and real time stability determination at both 4 °C and 30 °C.

d. Cut-off

A total of 150 samples equally distributed at concentrations of -50% Cut-Off: -25% Cut-Off: Cut-Off: +25% Cut-Off: +50% Cut-Off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% Cut-off and all negative at and below -25% Cut-off for both Secobarbital and Oxycodone.

The following cut-off values for the candidate devices have been verified.

CalibratorCut-off (ng/mL)
Secobarbital300
Oxycodone100

e. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different devices.

Secobarbital:

AcetaminophenErythromycinO-Hydroxyhippuric acid
Acetophenetidinβ-EstradiolD,L-Octopamine
Acetylsalicylic acidEstrone-3-sulfateOxalic acid
AminopyrineEthyl-p-aminobenzoateOxazepam
AmitryptylineFenoprofenOxolinic acid
AmoxicillinFurosemideOxycodone
DL-Amphetamine sulfateGentisic acidOxymetazoline
AmpicillinHemoglobinPapaverine
ApomorphineHydralazinePenicillin-G
Ascorbic acidHydrochlorothiazidePentazocaine
AspartameHydrocodonePerphenazine
AtropineHydrocortisonePhencyclidine
Benzilic acidp-HydroxyamphetaminePhenelzine

7

{11}------------------------------------------------

Benzoic acidp-Hydroxymethamphetamineβ-Phenylethlamine
Benzoylecgonine3-HydroxytyraminePhenylpropanolamine
BilirubinIbuprofenPrednisolone
BrompheniramineImipraminePrednisone
Caffeine(-) IsoproterenolProcaine
CannabidiolIsoxsuprinePromazine
CannabinolKetaminePromethazine
ChloralhydrateKetoprofenD,L-Propanolol
ChloramphenicolLabetalolD-Propoxyphene
ChlorothiazideLevorphanolQuinidine
(±) ChlorpheniramineLoperamideQuinine
ChlorpromazineL-PhenylephrineRanitidine
ChlorquineMaprotilineSalicylic acid
CholesterolMeperidineSerotonin
ClomipramineMeprobamateSulfamethazine
ClonidineMorphine-3-β-D glucuronideSulindac
Cocaine hydrochlorideMethadoneTemazepam
CodeineMethamphetamineTetracycline
Cortisone(±) - 3,4-Methylenedioxy-amphetamineTetrahydrozoline
(-) Cotinine(±)-3,4-Methylenedioxy-methamphetaminehydrochlorideThebaine
CreatinineMorphine SulfateThiamine
DeoxycorticosteroneN-AcetylprocainamideThioridazine
DextromethorphanNalidixic acidTriamterene
DiazepamNaloxoneTrifluoperazine
DiclofenacNaltrexoneTrimethoprim
DiflunisalNaproxenTrimipramine
DigoxinNiacinamideTryptamine
DiphenhydramineNifedipineD, L-Tyrosine
DoxylamineNorcodeinUric acid
Ecgonine hydrochlorideNorethindroneVerapamil
Ecgonine methylesterD-NorpropoxypheneZomepirac
(IR 2S)-EphedrineNoscapine

Oxycodone

AcetophenetidinL-EphedrineOxymetazoline
Nalidixic acidEcgonine methylesterPapaverine
Acetylsalicylic acidEthyl-p-aminobenzoatePenicillin-G
AminopyrineB-EstradiolPerphenazine
AmoxicillinEstrone-3-sulfatePhenelzine
AmpicillinErythromycinL-Phenylephrine
L-PhenylephrineFenoprofenB-Phenylethylamine
ApomorphineFurosemidePhenylpropanolamine
AspartameGentisic acidPrednisone
AtropineHemoglobinLoperamide

{12}------------------------------------------------

Benzilic acidHydralazineQuinine
Benzoic acidHydrochlorothiazideQuinidine
BenzphetamineHydrocortisoneRanitidine
BilirubinO-Hydroxyhippuric acidSalicylic acid
Deoxycorticosterone3-HydroxytyramineSerotonin
CaffeineLabetalolSulfamethazine
ChloralhydrateD, L-IsoproterenolSulindac
ChloramphenicolMeprobamateTetracycline
ChlorothiazideMethoxyphenamineTetrahydrocortisone
D,L-ChlolrpheniramineNalidixic acidMorphine-3-β-D-gluc
ChlorpromazineNaloxoneuronide
ChlorquineNaltrexoneTetrahydrozoline
CholesterolNaproxenThiamine
ClonidineNiacinamideThioridazine
L-CotinineNifedipineD,L-Tyrosine
CortisoneIsoxsuprineTolbutamide
CreatinineD,L-PropanololTriamterene
D-PseudoephedrineKetoprofenTrifluoperazine
DextromethorphanNorethindroneTrimethoprim
ß-DglucuronideD-NorpropoxypheneTyramine
DiclofenacNoscapineD,L-Tryptophan
DiflunisalD,L-OctopamineUrine acid
DigoxinOxalic acidVerapamil
DiphenhydramineOxolinic acidZomepirac

f. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below. There were no differences observed for different devices.

Secobarbital0/0
(Cut-off=300 ng/mL)ResultCross-Reactivity
SecobarbitalPositive at 300 ng/mL100%
AmobarbitalPositive at 1000 ng/mL30%
AlphenolPositive at 62.5 ng/mL480%
AprobarbitalPositive at 250 ng/mL120%
ButabarbitalPositive at 100 ng/mL300%
ButathalPositive at 500 ng/mL60%
ButalbitalPositive at 5000ng/mL6%
CyclopentobarbitalPositive at 500 ng/mL60%
PentobarbitalPositive at 200 ng/mL150%
PhenobarbitalPositive at 300 ng/mL100%
Oxycodone(Cut-off=100 ng/mL)Result%Cross-Reactivity
OxycodonePositive at 100 ng/mL100%
CodeinePositive at 100000 ng/mL0.1%
AcetylmorphinePositive at 100000 ng/mL0.1%
OxymorphonePositive at 250 ng/mL40%

{13}------------------------------------------------

DihydrocodeinePositive at 100000 ng/mL0.1%
HydromorphonePositive at 100000 ng/mL0.1%
HydrocodonePositive at 3125 ng/mL3.2%
MorphinePositive at 100000 ng/mL0.1%
BuprenorphinePositive at 100000 ng/mL0.1%
EthylmorphinePositive at 100000 ng/mL0.1%

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three batches of each device . Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different devices.

2. Comparison Studies

Method comparison studies for the candidate devices were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:

StripNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
Secobarbital
ViewerPositive0011525
ANegative1020900
ViewerPositive0011525
BNegative1020900
ViewerPositive0011525
CNegative1020900

Discordant Results of Secobarbital Strip

ViewerSample NumberGC/MS ResultStrip Viewer Results
Viewer A817130286Positive
Viewer B817130286Positive
Viewer C817130286Positive

{14}------------------------------------------------

PanelDipNegativeLow Negative byGC/MS(less than-50%)Near Cutoff Negative byGC/MS(Between-50% andcutoff)Near Cutoff Positive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
Negative1020910
ViewerBPositive0011525
Negative1020900
ViewerCPositive0011525
Negative1020900

Discordant Results of Secobarbital Panel Dip

ViewerSample NumberGC/MS ResultPanel DipViewer Results
Viewer A817130286Positive
Viewer B817130286Positive
Viewer C817130286Positive
Viewer A933321306Negative
Turn-KeySplitCupNegativeLow Negative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
Negative120910
ViewerBPositive0001425
Negative1201010
ViewerCPositive0011425
Negative120910

Discordant Results of Secobarbital Turn-Key Split Cup

ViewerSample NumberGC/MS ResultTurn-Key SplitCup
Viewer A817130286Positive
Viewer A862382310Negative
Viewer C817130286Positive
Viewer B933321306Negative
Viewer C722825336Negative

{15}------------------------------------------------

QuickCupNegativeLow Negative byGC/MS(less than-50%)Near Cutoff Negative byGC/MS(Between-50% andcutoff)Near Cutoff Positive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
Negative1020910
ViewerBPositive0011525
Negative1020900
ViewerCPositive0011525
Negative1020900

Discordant Results of Secobarbital Quick Cup

ViewerSample NumberGC/MS ResultQuick Cup Viewer Results
Viewer A216041241Positive
Viewer B126635252Positive
Viewer C126635252Positive
Viewer A358232315Negative

Oxycodone (OXY)

StripNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
ViewerANegative1020910
ViewerBPositive0011525
ViewerBNegative1020900
ViewerCPositive0011525
ViewerCNegative1020900

Discordant Results of OXY Strip

ViewerSample NumberGC/MS ResultStrip Viewer Results
Viewer A63880596Positive
Viewer A757887102Negative
Viewer B63880596Positive
Viewer C63880596Positive

{16}------------------------------------------------

PanelDipNegativeLow Negative byGC/MS(less than -50%)Near Cutoff Negative byGC/MS(Between -50% andcutoff)Near Cutoff Positive byGC/MS(Between thecutoff and +50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011525
Negative1020900
ViewerBPositive0001425
Negative10201010
ViewerCPositive0011525
Negative1020900

Discordant Results of OXY Panel Dip

ViewerSample NumberGC/MS ResultPanel DipViewer Results
Viewer A60321690Positive
Viewer C63880596Positive
Viewer B757887102Negative
Turn-KeySplit CupNegativeLow Negative byGC/MS(less than -50%)Near Cutoff Negative byGC/MS(Between -50% andcutoff)Near Cutoff Positive byGC/MS(Between the cutoff and+50%)High Positiveby GC/MS(greater than +50%)
ViewerAPositive0011425
Negative120910
ViewerBPositive0011525
Negative120900
ViewerCPositive0001525
Negative1201000

Discordant Results of OXY Turn-Key Split Cup

ViewerSample NumberGC/MS ResultSplit Cup Viewer Results
Viewer A63880596Positive
Viewer A285248110Negative
Viewer B63880596Positive

{17}------------------------------------------------

QuickCupNegativeLow Negative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
Negative1020910
ViewerBPositive0001425
Negative10201010
ViewerCPositive0011425
Negative1020910

Discordant Results of OXY Quick Cup

ViewerSample NumberGC/MS ResultQuick CupViewer Results
Viewer A63880596Positive
Viewer C60321690Positive
Viewer A757887102Negative
Viewer B508044116Negative
Viewer C285248110Negative

Lay-user study

A lay user study was performed at three intended user sites with 1060 lay persons. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested.

Comparison between GC/MS and Lay Person Results for Secobarbital Strip

% of CutoffNumber of samplesSecobarbitalConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200No. ofPositiveNo. ofNegative100
-75% Cutoff2073020100
-50% Cutoff20149020100
-25% Cutoff2022821890
+25% Cutoff20369200100
+50% Cutoff20455200100
+75% Cutoff20530200100

{18}------------------------------------------------

NumberOxycodone ConcentrationLay person resultsThe
% of Cutoffofsamplesby GC/MS (ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2022020100
-50% Cutoff2048020100
-25% Cutoff208011995
+25% Cutoff2012719195
+50% Cutoff20158200100
+75% Cutoff20176200100

Comparison between GC/MS and Lay Person Results for OXY Strip

Comparison between GC/MS and Lay Person Results for BAR/OXY Panel DipCard

% CutoffNo ofsamplesConcentration byGC/MS(ng/mL)Lay personresultsCorrect Results(%)
SecobarbitalOxycodoneBAROXYBAROXY
-100%20000+/20-0+/20-100100
-75%2073220+/20-0+/20-100100
-50%20149480+/20-0+/20-100100
-25%20228801+/19-1+/19-9595
+25%2036912719+/1-18+/2-9590
+50%2045515820+/0-20+/0-100100
+75%2053017620+/0-20+/0-100100

Comparison between GC/MS and Lay Person Results for BAR/OXY Turn-Key Split Cup

% CutoffNo ofsamplesConcentration byGC/MS(ng/mL)Lay personresultsCorrect Results(%)
SecobarbitalOxycodoneBAROXYBAROXY
-100%20000+/20-0+/20-100100
-75%2073220+/20-0+/20-100100
-50%20149480+/20-0+/20-100100
-25%20228802+/18-1+/19-9095
+25%2036912720+/0-19+/1-10095
+50%2045515820+/0-20+/0-100100
+75%2053017620+/0-20+/0-100100

Comparison between GC/MS and Lay Person Results for BAR/OXY Quick Cup

% CutoffNo ofsamplesConcentration byGC/MS(ng/mL)Lay person resultsCorrect Results(%)
SecobarbitalOxycodoneBAROXYBAROXY
-100%20000+/20-0+/20-100100
-75%2073220+/20-0+/20-100100
-50%20149480+/20-0+/20-100100
-25%20228801+/19-1+/19-9595

{19}------------------------------------------------

+25%20૩૯તે રાજ્યના ઉત્તર પ્રદેશના પાકની ખેતી કરવવામાં આવે છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસ12720+/0-20+/0-100100
+50%2045515820+/0-20+/0-100100
+75%2053017620+/0-20+/0-100100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

3. Clinical Studies

Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that Secobarbital Strip, AssureTech Oxycodone the AssureTech Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are substantially equivalent to the predicate.

§ 862.3150 Barbiturate test system.

(a)
Identification. A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.(b)
Classification. Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).